`Estimating Respiratory and Heart Rates from the
`Correntropy Spectral Density of the
`Photoplethysmogram
`
`Ainara Garde1,2*, Walter Karlen1,2, J. Mark Ansermino1,2, Guy A. Dumont1,2
`
`1 Electrical and Computer Engineering in Medicine Group, The University of British Columbia and BC Childrens Hospital, Vancouver, British Columbia, Canada,
`2 Anesthesiology, Pharmacology and Therapeutics, The University of British Columbia and BC Childrens Hospital, Vancouver, British Columbia, Canada
`
`Abstract
`
`The photoplethysmogram (PPG) obtained from pulse oximetry measures local variations of blood volume in tissues,
`reflecting the peripheral pulse modulated by heart activity, respiration and other physiological effects. We propose an
`algorithm based on the correntropy spectral density (CSD) as a novel way to estimate respiratory rate (RR) and heart rate
`(HR) from the PPG. Time-varying CSD, a technique particularly well-suited for modulated signal patterns, is applied to the
`PPG. The respiratory and cardiac frequency peaks detected at extended respiratory (8 to 60 breaths/min) and cardiac (30 to
`180 beats/min) frequency bands provide RR and HR estimations. The CSD-based algorithm was tested against the
`Capnobase benchmark dataset, a dataset from 42 subjects containing PPG and capnometric signals and expert labeled
`reference RR and HR. The RR and HR estimation accuracy was assessed using the unnormalized root mean square (RMS)
`error. We investigated two window sizes (60 and 120 s) on the Capnobase calibration dataset to explore the time resolution
`of the CSD-based algorithm. A longer window decreases the RR error, for 120-s windows, the median RMS error (quartiles)
`obtained for RR was 0.95 (0.27, 6.20) breaths/min and for HR was 0.76 (0.34, 1.45) beats/min. Our experiments show that in
`addition to a high degree of accuracy and robustness, the CSD facilitates simultaneous and efficient estimation of RR and
`HR. Providing RR every minute, expands the functionality of pulse oximeters and provides additional diagnostic power to
`this non-invasive monitoring tool.
`
`Citation: Garde A, Karlen W, Ansermino JM, Dumont GA (2014) Estimating Respiratory and Heart Rates from the Correntropy Spectral Density of the
`Photoplethysmogram. PLoS ONE 9(1): e86427. doi:10.1371/journal.pone.0086427
`
`Editor: Derek Abbott, University of Adelaide, Australia
`
`Received June 17, 2013; Accepted December 10, 2013; Published January 22, 2014
`Copyright: ß 2014 Garde et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits
`unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
`
`Funding: This work was supported in part by The Natural Sciences and Engineering Research Council of Canada (NSERC), and the Canadian Institutes of Health
`Research (CIHR) through Collaborative Health Research Projects Program. The funders had no role in study design, data collection and analysis, decision to
`publish, or preparation of the manuscript.
`
`Competing Interests: The authors have declared that no competing interests exist.
`
`* E-mail: ainara.garde@cw.bc.ca
`
`Introduction
`
`The ability to track multiple vital signs from a simple, low cost,
`and easy to use non-invasive sensor is desirable to facilitate
`physiological tele-monitoring. There is a clear need for reliable
`and simple methods for tracking cardio-respiratory activity over
`time to monitor patients in the intensive care environment or
`patients at home with long-term disease with associated instability
`in respiratory or cardiovascular function. Therefore, the remote
`and automated monitoring of heart rate (HR) and respiratory rate
`(RR) is an important field of research [1].
`An abnormal RR is often an early sign of critical illness. For
`example, an essential criterion integrated in guidelines for the
`diagnosis of pneumonia in children (age 1–5 years)
`is
`the
`assessment of an elevated RR (.40 breaths/min) [2]. However,
`clinical measurement of RR has been shown to have poor
`reliability and repeatability [3]. A reliable estimate of RR assessed
`in an automated way is therefore crucial in the application of
`remote tele-monitoring, where persons with no specialized training
`are conducting the assessment. This would enable early support
`for timely recognition and management of physiological deterio-
`ration of high-risk patient groups [4].
`
`Pulse oximetry is widely used in health facilities to monitor
`physiological vital
`signs.
`It
`is based on the principle of
`photoplethysmography (PPG), an optical technique to measure
`local variations of blood volume in tissues. Two light-emitting
`diodes (LEDs) illuminate the tissue and a photo detector detects
`the light reflected by the tissue. The intensity of the light detected
`varies with each heart beat as the blood volume changes over time
`[5]. Blood oxygen saturation (SpO2) is calculated by measuring
`the difference in absorption of oxygenated and deoxygenated
`hemoglobin at two distinct wavelengths, red (660 nm) and infrared
`(940 nm). Oxygenated blood preferably absorbs infrared light and
`transmits red light and deoxygenated blood has the inverted
`absorption characteristics [4].
`The PPG is a complex signal composed of different but related
`components. The most recognized PPG waveform component is
`the peripheral pulse synchronized to each heart beat
`(AC
`component). This AC component is superimposed and modulated
`by a quasi DC component that varies slowly due to respiration,
`vasomotor activity and vasoconstrictor waves [4]. In addition, an
`autonomic response to respiration causes a variation of HR
`synchronized with RR, referred to as respiratory sinus arrhythmia.
`The PPG signal is also influenced by other mechanisms that are
`not completely understood. However, it is generally accepted that
`
`PLOS ONE | www.plosone.org
`
`1
`
`January 2014 | Volume 9 |
`
`Issue 1 | e86427
`
`

`

`Respiratory & Heart Rate from PPG with Correntropy
`
`Vc(m):e{jvm,
`
`ð1Þ
`
`XN
`
`{1
`
`Pv(v)~
`
`m~{(N{1)
`
`where Vc(m) is the centered correntropy function, in which the
`mean of the transformed data is subtracted so as to reduce the
`effect of output DC bias. It is estimated by Vc(m)~V (m){ VV
`where Vc(m) is the correntropy function and VV the correntropy
`mean, defined as:
`
`ð2Þ
`
`ð3Þ
`
`k(x(n){x(n{m)),
`
`XN
`
`V (m)~
`
`1
`N{mz1
`
`n~m
`
`XN
`XN
`
`k(x(n){x(n{m)):
`
`m~1
`
`n~m
`
`1 N
`
`2
`
`V ~
`
`The sigmoidal, Gaussian, polynomial, and spline kernels are the
`most commonly used symmetric positive definite kernels, applied
`to machine learning, function approximation, density estimation,
`and support vector machine classification [22], [23]. The Gaussian
`kernel function, applied in the present study, is given by
`
`h
`
`i
`
`k(x(n){x(n{m))~
`
`1ffiffiffiffiffiffi
`p
`
`
`2ps
`
`{(x(n){x(n{m))2
`2s2
`
`e
`
`,
`
`ð4Þ
`
`where s is the kernel parameter, here set using Silverman’s rule of
`density estimation [19].
`is a
`[19],
`Correntropy,
`introduced by Santamaria et al.
`similarity measure defined in terms of inner products of vectors
`in a kernel parameter space. It provides information on both the
`time structure and the statistical distribution. In addition, the use
`of kernel methods makes the correntropy computationally efficient
`since it can be computed directly from the data.
`Autoregressive (AR) spectral analysis based on the Yule-Walker
`method was applied to improve spectral resolution compared to
`conventional techniques [18]. The autoregressive coefficients were
`estimated from the correntropy function, using the YuleWalker
`equations [24]. The selection of model order is a trade-off between
`the frequency resolution and the spurious peaks. The optimal
`model order between 5 and 15 was selected using the minimal
`description length criteria defined by Rissanen [25].
`HR and RR estimation. The CSD over time shows both
`respiratory and cardiac frequency peaks reflecting the RR and HR
`respectively (Figure 2). These peaks can be tracked in the region of
`the respiratory and cardiac frequency bands. Reference HR and
`RR ranges were extracted from a review of observational studies
`that used HR data from 143,346 children and RR data from 3,881
`children (from 6 months to 18 years old) [26]. Based on 99th and
`1st centiles for children and young adults, the HR could range
`from 30 to 180 beats/min (0.5 to 3 Hz, respectively) and RR from
`8 to 60 breaths/min (0.14 to 1 Hz, respectively [26]. The range in
`adults is much more restricted but would be included in this range.
`An extreme range may occur in critical illness, such as an elevated
`HR in the presence of an arrhythmia or an elevated RR (w 40
`breaths/min in children with pneumonia [2]) as an early indicator
`of critical illness. However, those pathological or abnormal RR
`and HR values will also be included in this extended HR and RR
`ranges extracted from the review. Therefore, the maximum value
`
`it has potential to provide clinically useful information about the
`cardio-vascular and respiratory system [6] and its SpO2 pattern
`characterization has successfully applied to detect sleep apnea [7].
`Well-established methods have been described for the estima-
`tion of SpO2 and HR from the PPG [8], [9]. In addition, several
`methods based on characterization of the PPG cycles morphology
`in the time domain, using time-frequency analysis [10], [11], [12],
`[13], [14], [15], [16] digital filtering [5], [17] and smart fusion [6]
`have been proposed to estimate RR. However, this estimation of
`RR in pulse oximetry is not yet commercially established. The
`simultaneous estimation of HR and RR from the PPG signal
`would provide a low processing overhead that is desirable for
`simple and low cost physiological monitor. This would reduce vital
`sign monitoring hardware to one peripheral sensor and one signal-
`processing step.
`Correntropy-based spectral density (CSD) has been found to be
`particularly well suited for the characterization of modulated
`signals. This method provides an improved spectral resolution
`compared to conventional techniques like power spectral density
`(PSD)and shows promise in the detection of modulated patterns
`[18]. Correntropy is a generalized correlation function that
`provides information on higher-order statistics. It is able to detect
`nonlinearities that conventional techniques (based on second-order
`statistics), may be unable to detect. Another attractive property of
`the correntropy function is its robustness against impulsive noise
`[19], [20].
`In this paper we propose a novel algorithm based on CSD to
`estimate both RR and HR simultaneously from the PPG signal
`obtained from pulse oximetry. The initial application will be to
`develop an easy-to-use portable device that measures multiple vital
`signs. This algorithm is ideally suited to be implemented on the
`Phone OximeterH, a mobile device that integrates a commercially
`available and federal drug administration (FDA) approved pulse
`oximeter (Xpod) with a mobile phone. The Phone OximeterH enables
`the analysis of vital signs and intuitive display of information to
`health care providers [21]. In addition, Phone Oximeter’s SpO2
`characterization has been successfully applied to detect sleep
`apnea [7].
`This paper is organized as follows; the Materials and Methods
`section describes the dataset used for the development and testing
`of the newly developed algorithm to estimate RR and HR based
`on CSD, and explains the algorithm with brief description of CSD
`and PSD methods. The accuracy of the CSD-based algorithm is
`presented in the Results
`section, which is
`followed by the
`Discussion, Limitations and Conclusion sections.
`
`Materials and Methods
`
`CSD-based Algorithm
`Conventional spectral analysis assumes a stationary signal and is
`therefore unable to identify HR and RR changes over time. An
`approach to account for such changes is to implement a time-
`varying spectral analysis. Firstly, a sliding time window of 60 s or
`120 s with 50% overlap is used to segment PPG signal
`into
`segments assumed to be stationary and suitable for spectral
`analysis. Secondly, the CSD is applied to the signal segments.
`Thirdly, the HR is estimated by detecting the maximum frequency
`peak within the cardiac frequency band and filtered from the
`signal, and lastly the RR is estimated by detecting the maximum
`frequency peak within the respiratory frequency band (see
`Figure 1).
`Correntropy spectral density. The CSD is a generalization
`of the conventional power spectral density. It is based on the
`Fourier transform of the centered correntropy function [18],
`
`PLOS ONE | www.plosone.org
`
`2
`
`January 2014 | Volume 9 |
`
`Issue 1 | e86427
`
`

`

`Respiratory & Heart Rate from PPG with Correntropy
`
`order between 5 and 15 was selected using the minimal description
`length criteria defined by Rissanen [25].
`
`Simulation Database
`To show certain performance properties of the algorithm a
`simulated PPG signal was first produced. Respiration has three
`different effects on the PPG waveform. The first and more
`predominant effect is a shift in the baseline during each breath.
`The second is a change of the amplitude of the pulse beats with
`each breath which implies that the PPG signal
`is subject to
`amplitude modulation (AM) [15]. The third effect is a variation of
`HR due to an autonomic response to respiration and usually
`decreases with age. Based on the first 2 effects for sake of
`simplicity, the PPG signal was simulated using AM and a baseline
`shift as follows:
`

`x(n)~ (1zm cos (vrn)) cos (vcn)
`
`Šzb(n),
`
`ð7Þ
`
`where fc~vc=2p is the cardiac frequency, fr~vr=2p is the
`respiratory frequency, m [½0,1Š is the modulation index and b(n) is
`the baseline shift synchronized with fr. One hundred outliers with
`values between mean + 5 standard deviation of x(n) were
`randomly added to the signal to simulate noise.
`
`Capnobase Database
`Ethics statement. All subjects were studied according to a
`protocol approved by the University of British Columbia and
`Children’s and Women’s Health Centre of British Columbia
`Research Ethics Board. Informed and written consent to be part of
`the research database was obtained for all subjects. For subjects
`under 16 years of age, parental/guardian written consent was
`obtained. Written assent was obtained for all subjects over the age
`of 11 years.
`Database. Capnobase is an on-line database that contains
`physiological
`signals collected during simultaneously elective
`surgery and routine anesthesia for the purpose of development
`of improved monitoring algorithms in adults and children [27].
`The signals were recorded from 59 children (median age: 8.7,
`range 0.8–16.5 years) and 35 adults (median age: 52.4, range
`26.2–75.6 years) receiving general anesthesia at
`the British
`Columbia Children’s Hospital and St. Paul’s Hospital, Vancouver
`BC, respectively. The recordings included ECG with a sampling
`frequency of 300 Hz, capnometry with a sampling frequency of 25
`Hz, and PPG with a sample frequency of 100 Hz. All signals were
`recorded with S/5 Collect software (Datex-Ohmeda, Finland)
`using a sampling frequency of 300 Hz (PPG and capnometry with
`lower sampling rates were automatically up-sampled).
`Capnobase contains a benchmark dataset with forty-two 8-min
`segments from 29 pediatric and 13 adults cases containing reliable
`recordings of spontaneous or controlled breathing. The capno-
`metric waveform was used as
`the reference gold standard
`recording for RR. A research assistant manually labeled each
`breath in the capnogram and pulse peak in the PPG and validated
`the derived instantaneous reference RR and HR. The beginning
`and end of all artifacts in the PPG waveforms were also manually
`labeled and almost 50% of the cases contained artifacts due to
`movements or similar noise. Capnobase also contains a calibration
`dataset with one hundred twenty-four 2-min segments randomly
`selected from the remaining 52 cases. This dataset is particularly
`challenging because it includes other disturbances such as cardiac
`oscillations etc., which influence the respiratory induced param-
`eters and it also contains substantially more movement artifacts
`than the benchmark dataset. Signals with significant apnea have
`
`Figure 1. Overview of the CSD-based algorithm. Initially the PPG
`signal is segmented into windows (60 s or 120 s) with 50% of overlap. In
`the subsequent step the CSD is applied to calculate the spectrum of the
`windowed signals. The HR is estimated by detecting the maximum
`frequency peak within the cardiac frequency band. The signal is then
`low pass filtered and the RR is estimated by detecting the maximum
`frequency peak within the respiratory frequency band.
`doi:10.1371/journal.pone.0086427.g001
`
`peak frequencies in the cardiac frequency band (0.5 to 3 Hz) and
`in the respiratory frequency band (0.14 to 1 Hz) were automat-
`ically extracted, reflecting HR and RR, respectively.
`For improved resolution around the respiratory frequency peak,
`the HR was filtered using a zero-phase 5th order low pass filter
`with a cutoff frequency of 0.1 Hz below the cardiac frequency. In
`addition,
`frequency peaks close to the secondary harmonics
`around HR were excluded when an elevated RR (w 45 breaths/
`min) were detected. An example of the RR and HR extracted
`from the time varying CSD (Figure 2) is illustrated in Figure 3.
`Power spectral density. Following the same concept a PSD-
`based algorithm was implemented. For a parametric PSD, the
`signal x(n) was modeled through an AR model by
`
`a½kŠx(n{k)ze(n),
`
`ð5Þ
`
`Xp
`
`x(n)~{
`
`k~1
`
`e , a½kŠ
`where e(n) denotes zero-mean white noise with variance s2
`the AR coefficients and p the model order. Once the autore-
`gressive coefficients and the variance s2
`e have been estimated, the
`PSD of an autoregressive process is computed by means of
`
`ð6Þ
`
`

`
`2 ,
`
`s2
`e
`~1 a½kŠ:e{j:vkT
`
`P
`
`p k
`
`1z
`
`

`
`Px(v)~
`
`being T the sampling period. As for the CSD, the optimal model
`
`PLOS ONE | www.plosone.org
`
`3
`
`January 2014 | Volume 9 |
`
`Issue 1 | e86427
`
`

`

`Respiratory & Heart Rate from PPG with Correntropy
`
`Figure 2. Time-varying CSD of 8-min PPG signal. Both respiratory and cardiac frequency peaks reflect RR and HR, respectively. Respiratory
`frequency peak is around 0.3 Hz (18 breaths/min) and cardiac frequency peak around 1.25 Hz (75 beats/min).
`doi:10.1371/journal.pone.0086427.g002
`
`been excluded from the analysis. Datasets can be downloaded
`from the on-line database, CapnoBase.org [27]. CSD-based
`algorithm was optimized using the calibration dataset and then
`validated using the benchmark dataset. Both, the calibration and
`benchmark datasets with reference RR and HR have been
`previously used to test RR estimation from PPG [6].
`
`Algorithm Evaluation
`The accuracy of the CSD-based algorithm was evaluated and
`compared to other methods, using the un-normalized root mean
`square (RMS) error. The RMS error was calculated for each
`subject, considering all estimations over time:
`
`s
`
`Xn
`
`ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
`i xest
`(xref
`
`,
`
`ð8Þ
`
`)2
`
`i~1
`
`i
`
`1 n
`
`RMS error~
`
`Validation. The calibrated algorithm was then validated
`using the Capnobase benchmark dataset. All subjects and all signal
`segments with mechanical or spontaneous breathing,
`including
`those with artifacts, were analyzed. The median error and 1st and
`3rd quartiles were calculated to account for a non-normal RMS
`distribution. A Bland-Altman plot was also performed to compare
`the estimated HR and RR to the reference rates.
`In addition, the performance of our algorithm was compared to
`previously proposed methods based on PPG cycles morphology
`[6], time-frequency analysis [15], [7] and digital filtering [17],
`using the Capnobase benchmark dataset. These methods have
`been implemented according to the description included on these
`papers.
`related samples using
`for
`A Wilcoxon signed-rank test
`Bonferroni correction for multiple comparisons was also applied
`to evaluate the statistical significance of our algorithm’s improve-
`ment. The normality of all distributions was tested using One-
`Sample Kolmogorov-Smirnov test.
`
`Results
`
`CSD Output
`The median RR error obtained with the CSD-based algorithm
`applied to the calibration dataset was 4.2 breaths/min when using
`60-s windows and 1.9 breaths/min when using 120-s windows.
`The RMS error significantly (pv 0.05) decreased with longer
`(10sSilverman) reduced the spurious
`windows. A kernel size of
`harmonics and provided more accurate RR estimates (see Figure 4)
`[19]. Therefore, a 10sSilverman was applied to the Capnobase
`Benchmark dataset.
`CSD shows two clear frequency peaks at HR and RR locations,
`for both simulated and in-vivo signals (Figure 5 and 6). As reported
`in our previous work [18], the AM effect is reflected in CSD
`through a frequency peak at its true position. In comparison, the
`AM in PSD is manifested as secondary harmonics surrounding the
`cardiac frequency peak. Further, CSD is more robust to impulsive
`
`and xest
`where n is the number of observations and xref
`are the
`i
`i
`reference and the estimated values, respectively. The median of
`the instantaneous reference RR and HR were compared to the
`estimations for each time window.
`Calibration. The spectral resolution increases with longer
`time-windows with a concomitant reduction in real-time perfor-
`mance (clinicians are required to wait longer for each estimation).
`To investigate the trade-off in window size, the accuracy of the
`algorithm was evaluated with the calibration dataset, using time
`windows of 60 s and 120 s with an overlap of 50%. The statistical
`significance of the error with the different windows was evaluated
`using Wilcoxon signed-rank test to compare related samples.
`The choice of the kernel parameter (s) is trade-off between the
`power of the respiratory peak and the spurious peaks. The power
`of the respiratory peak and spurious harmonics decreases as s
`increases [18]. The CSD-based algorithm’s sensitivity to s was
`evaluated using the calibration dataset. The s calculated according
`to Silverman’s rule (sSilverman) was used as a reference.
`
`PLOS ONE | www.plosone.org
`
`4
`
`January 2014 | Volume 9 |
`
`Issue 1 | e86427
`
`

`

`Respiratory & Heart Rate from PPG with Correntropy
`
`Figure 3. Time-varying estimated and manually labeled reference RR and HR. Estimated (solid blue with * markers) and manually labeled
`(dotted red with+markers) reference RR in (A) and HR in (B). For this subject the RMS errors estimating RR and HR are 0.25 breaths/min and 0.35
`beats/min, respectively.
`doi:10.1371/journal.pone.0086427.g003
`
`noise (see Figure 5.D and Figure 5.F, respectively) compared to the
`PSD approach. This is because the Gaussian kernel makes
`k(x(n){x(n{m))&0 when either x(n) or x(n{m) is an artifact.
`Although the baseline shift is present in both PSD and CSD,
`CSD provides a more robust respiratory modulation frequency
`peak compared to the PSD (see Figure 5.C and Figure 5.E,
`respectively). The enhanced modulation peak is also observed in
`the in-vivo signals (Figure 6.F and Figure 6.G, respectively) where
`the CSD analysis provides a clearer respiratory frequency peak
`compared to the PSD.
`
`Benchmark Accuracy Measurements
`The CSD-based algorithm provided a significantly lower RR
`error compared to the PSD-based algorithm (Table 1). Expanding
`the permitted cardiac and respiratory frequency bands, increased
`the total range of the errors (Figure 7). RR or HR misdetections
`increased the RMS error considerably.
`The median RR error significantly decreased (pv 0.05) with
`longer time windows for using CSD (from 1.77 to 0.95 breaths/
`min) and PSD (from 7.82 to 3.18 breaths/min) approaches.
`However,
`the median error was not statistically different
`in
`estimating HR with longer time windows.
`A Bland-Altman plot (Figure 8) showed good agreement with a
`HR bias of 0.18 and limits of agreement of 21.52 to 1.91, and a
`RR bias of 21.1 and limits of agreement of 26.52 to 4.32.
`The accuracy of the algorithm per subject is illustrated in
`Figure 9, where the estimated RR and HR using a 60-s sliding
`
`window and reference values for each subject are represented.
`When analyzing the estimation for each time-window (Figure 10),
`it can be observed that most of RR errors are accumulated at low
`frequencies (v 15 breaths/min). However, there are some errors
`located out of the normal range because of artifacts and the use of
`extended respiratory and cardiac frequency bands. The number of
`erroneous estimates was reduced when increasing the window
`length, which is reflected by a lower error range when using 120-s
`windows.
`Table 1 illustrates the performance of a number of methods
`using the Capnobase benchmark dataset. CSD-based method
`provided the lowest RR error when using 120-s windows. A
`Wilcoxon signed-rank test for related samples with Bonferroni
`correction for multiple comparisons has shown the significant
`improvement (pv0.05) of our algorithm when compared to PSD
`and the methods proposed by [7] and [17].
`
`Discussion
`
`In this study we have presented a novel methodology to estimate
`both RR and HR simultaneously from pulse oximetry based on
`CSD. The performance of
`the algorithm has been validated
`against a benchmark dataset using RMS error, comparing all
`estimations with reference RR and HR rates manually labeled by
`a research assistant. The algorithm has shown high accuracy and
`robustness estimating RR and HR simultaneously from PPG, even
`when the search is extended to account for pathological and/or
`
`PLOS ONE | www.plosone.org
`
`5
`
`January 2014 | Volume 9 |
`
`Issue 1 | e86427
`
`

`

`Respiratory & Heart Rate from PPG with Correntropy
`
`Figure 4. CSD sensitivity to the kernel parameter. CSD-based algorithm’s perfomance estimating RR is illustrated for the kernel values:
`sSilverman, 10sSilverman, 100sSilverman and 1000sSilverman. The sSilverman is calculated by Silverman’s rule.
`doi:10.1371/journal.pone.0086427.g004
`
`Figure 5. CSD applied to a simulated signal. (A) Simulated signal with 0.2 Hz modulation respiratory frequency (12 breaths/min), 1 Hz cardiac
`frequency (60 beats/min), and m~1, (B) same simulated signal with some outliers randomly added, (C) and (D) the CSD of the simulated signal with
`and without outliers, and (E) and (F) the PSD of the simulated signal with and without outliers, respectively. CSD analysis provides a clearer and more
`robust against outliers respiratory frequency peak than conventional PSD.
`doi:10.1371/journal.pone.0086427.g005
`
`PLOS ONE | www.plosone.org
`
`6
`
`January 2014 | Volume 9 |
`
`Issue 1 | e86427
`
`

`

`Respiratory & Heart Rate from PPG with Correntropy
`
`Figure 6. CSD applied to an in-vivo signal. CSD and PSD performance applied to an in-vivo signal (1 min) of one infant subject: (A) reference HR
`(dotted red line with . markers) and mean HR represented by a dotted grey line, (B) reference RR (dotted red line with . markers) and mean RR
`represented by a dotted grey line, (C) ECG signal, (D) capnometry, (E) PPG signal, (F) CSD and (G) PSD applied to the PPG signal. In addition, the
`average CSD and PSD spectrum of the database’s population is illustrated in the background on (F) and (G) respectively, where the cardiac
`component is represented in dark grey and the filtered signal that corresponds to respiration in light grey.
`doi:10.1371/journal.pone.0086427.g006
`
`abnormal rates to 8 to 60 breaths/min and 30 to 180 beats/min.
`In addition to its accuracy and robustness,
`the RR error
`significantly decreased when longer time windows were used.
`Moreover to generalize our findings broad ranges of subjects
`including children and adults, under controlled ventilation or
`spontaneously breathing over a wide RR ranges were studied.
`
`Using CSD for frequency estimation is preferable to conventional
`PSD functions as it accounts for higher-order moments and is more
`robust to outliers [20]. CSD is particularly useful in signals with
`amplitude modulation like the PPG signal that was analyzed in this
`study. CSD provides the modulation frequency at its actual location
`along the frequency axis [18],
`instead of at
`locations of
`the
`secondary harmonics surrounding the carrier frequency peak. The
`
`PLOS ONE | www.plosone.org
`
`7
`
`January 2014 | Volume 9 |
`
`Issue 1 | e86427
`
`

`

`Respiratory & Heart Rate from PPG with Correntropy
`
`Figure 7. Boxplot of RMS error. Boxplot of the (A) RR and (B) HR RMS error estimated using time windows of 60 s and 120 s and tracked around
`extended RR (from 8 to 60 breaths/min) and HR (from 30 to 180 beats/min). 1st quartile, median, and 3rd quartile values are displayed as bottom,
`middle and top horizontal line of the boxes. Whiskers are used to represent the most extreme values within 3 times the interquartile range from the
`quartile. Outliers (data with values beyond the ends of the whiskers) were displayed as crosses.
`doi:10.1371/journal.pone.0086427.g007
`
`respiratory effect on the PPG respiration is usually present as a
`baseline shift and an AM synchronized with each breath [15]. CSD
`represents both baseline shift and respiratory AM component at the
`same position, whereas a direct PSD of the signal provides only the
`RR derived from the baseline shift at its real position (see Figures 5
`and 6). Thus, in signals with a dominant AM component, CSD will
`provide a more robust respiratory frequency peak.
`
`the CSD is an appropriate
`The results demonstrate that
`technique to provide simultaneous and efficient estimation of
`RR and HR, and will permit to monitor of HR and RR non-
`invasively using only a peripheral sensor. The relevance of this
`algorithm from the clinical perspective is that it facilitates an
`accurate identification of abnormal or pathological rates. Thus,
`this promising algorithm will expand the functionality and
`diagnostic power of pulse oximeters. A number of algorithms
`
`Table 1. RMS error estimating RR and HR with different methods.
`
`Methods
`
`CSD120 s
`
`PSD120 s
`
`Karlen et al. [6]
`
`Garde et al. [7]
`
`Shelley et al. [15]
`
`Nakajima et al. [17]
`
`Median (1st and 3rd quartile)
`
`RR-RMS error (breaths/min)
`
`HR-RMS error (beats/min)
`
`0.95 (0.27, 6.20)
`
`3.18 (1.20, 11.3)*
`
`1.56 (0.60, 3.15)
`
`3.5 (1.1, 11)*
`
`1.91 (0.41, 7.01)
`
`7.47 (0.59, 10.6)*
`
`0.76 (0.34, 1.45)
`
`0.58 (0.21, 1.17)
`
`0.48 (0.37, 0.77)
`
`0.35 (0.2, 0.59)
`
`n/a
`
`n/a
`
`RMS error median (quartiles) estimating RR and HR using different methods. The statistical significant difference (pv0.05) of the RMS error obtained with the CSD-based
`algorithm in comparison to other methods is indicated (asterisk *).
`doi:10.1371/journal.pone.0086427.t001
`
`PLOS ONE | www.plosone.org
`
`8
`
`January 2014 | Volume 9 |
`
`Issue 1 | e86427
`
`

`

`Respiratory & Heart Rate from PPG with Correntropy
`
`Figure 8. Bland-Altman for HR and RR estimation. Bland-Altman plots for comparison of (A) HR and (B) RR to the reference HR and RR manually
`labeled by the research assistant. The bias and 95% of limits of agreement are ploted in solid lines. It showed a bias of 0.18 and limits of agreement of
`21.52 to 1.91 beats/min for the estimated HR versus reference HR and a bias of 21.1 and limits of agreement of 26.52 to 4.32 breaths/min for the
`estimated RR versus reference RR.
`doi:10.1371/journal.pone.0086427.g008
`
`based on the PPG signal morphology [13], [28] time-frequency or
`spectral analysis [11], [12], [14], [15], [17] digital filtering [5],
`[17], and complex demodulation [10] have been proposed to
`detect RR from PPG. Most of these methods are restricted to
`healthy ranges only and many are computationally expensive.
`Moreover, most of
`these methods have been tested only in
`controlled environments (research laboratories), and their robust-
`ness to artifacts and other influences that are very common in the
`ambulatory environment have not typically been demonstrated.
`Therefore, some of these methods have been implemented and
`
`applied to the same Capnobase benchmark dataset. CSD-based
`algorithm has provided lower RR error (0.95 breaths/min) using
`120-s windows. The main limitation of the methods proposed by
`Shelley et al. [15] and Nakajima et al. [17] is that they restrict
`their estimations to RR v 40 breaths/min.
`The Smart Fusion method proposed by Karlen et al [6] is
`computationally efficient and was evaluated in an ambulatory
`environment. It combines the three respiratory induced variations
`(frequency, intensity, and amplitude) using a mean calculation. This
`method improved the robustness of the RR estimation, with a
`
`Figure 9. Scatter plot, error per subject. Scatter plot showing the median value of estimated and reference values of (A) RR and (B) HR for each
`subject using 60-s time window. The respiratory and cardiac frequency peaks are detected around the extended RR and HR range. Observations with
`artifacts are included. The dotted line represents the optimal performance.
`doi:10.1371/journal.pone.0086427.g009
`
`PLOS ONE | www.plosone.org
`
`9
`
`January 2014 | Volume 9 |
`
`Issue 1 | e86427
`
`

`

`Respiratory & Heart Rate from PPG with Correntropy
`
`Figure 10. Scatter plot, er

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket