throbber
I
`
`'
`
`'
`
`.
`
`‘
`
`——-——
`
`Volume 14 Number 6 August 1998
`
`379—380 B riefReviews
`J. S. Graln’nsmin
`
`
`Original A rticles
`181—384 INFLIIENCE OF THE REFERENCE CA8 OF PAHIMACNETIC
`OHTIIEII HNALTZEHS O" NITROGEN CONCENTHNTIONS
`IIIIIIIHO CLOSED-CIRCUIT INESTHESIO
`jan F. A. Hrndrirhx, MD, Audra/1.]. iJanZnnih'ri. M'D,
`
`PhD, and Aiidi'i' M. rh' fal/hif, MD
`
`385—391 H'MILIIOILITY OF HECOHOS IN MI OIITI’ATIEIIT
`PIIEMIESTHETIC EONLIINTION CLINIC
`Carrion L, Cihini, MD, and I’Viihcm K. Schwah, PhD
`393—402 I” WHO EIIIILIHITIIIII OF II CLOSE” LOOP MONITORING
`STIIATEIIT FOH IIIIIIICEO PHHIILTSIS
`
`Dccpah Ramakrishna, 1148, Khamuu Brhhchani, PhD, PE.
`Kevin Kicia. M D, jqfih’ y ha‘ahhmr. M S,
`H’oif 14/. van i‘u/Iahzahn, PhD, PE, Raier C. Ehrrharr, PhD,
`and Michaci Daiiar, i143
`
`403—412 HEELECTANCE PULSE OXIMETIIY— PRINCIPLES MIC
`CSTETHIC IPFLICHTION IN THE ZIIIIICH “STEIN
`
`Vaihrr Kc'inig, Renate Hnrh, and Aihm Hath
`413—420 ACCURACY OF IJOLIIME MENSOHEMENTS IN MECHANICAL”
`VENTILNTEII NEWBOHNS:
`ABBII‘II'AIIATWE STIIB‘I IIF enmmsncm nemcas
`Kai Raske, Bern'am Faiiziie, Raiami R. I/Vrmcl', and
`Gerri Schmaiisrh
`
`
`421—424 N" EVOLUTIONARY SOLUTION TO ANESTHESIR NUTOHITEO
`HECOIIO KEEPING
`
`Algorithm
`44i—446 II HEM-TIME ALGORITHM TO II'I'II’HOIIE THE RESPONSE TIME
`OF A CLINICIIL MlllTlCilS NNHLTSEII
`Lawriy Wong, MSc, Rnih Hamihon. MB. ChB,
`
`Eiin'n Paiayinm, PhD. and Clive Hahn, D. Phii
`
`.rl
`
`_Cage, MD, and
`Ahn'n A. Bide”, Ph D, John
`
`Pani}. Poppers, M D
`EHHLOHTION OF A PITCT TTPE SPIHOMETEH IN HELIIIIIT/
`OXTOEN MIXTOHES
`Saran Suadei'qaard. MD, Signrbm'gnr Ka'msan. M D,
`
`Srqfaa Lumiin, MT), and Oh: Simqvisf, M'D
`433—459 OETECTIOH OF “INC INJUHT WITH CONVENTIONAL MIC
`HEOHHL NETWORK-BOSE" OHNLI'SIS OF CONTINUOUS [IOTA
`
`jnhha Rast‘incn, MD, and Mauricio A. Lean, MD
`
`Book: Review
`447 L. C. HENSON AND A. C. LEE [EDS]: SIMOLNTOHS IN
`ANESTHESIOLOOY EOIICHTION
`Richard Mani:
`
`
`
`449—450 INFORMATION FOR ICONTHIIIOTOIIS
`
`\
`‘
`*
`
`KLIJWEII nonnemc PunusnEns
`2—5:‘Efiia—iHNAinfifififififl'EREW"m"
`IN ANAESTHESIA nun INTENSIVE cane, ANII or
`THE SOCIETY FOR TECHNOLOGY IN ANESTNESIA
`
`wiser
`
`=
`
`~
`
`1-
`
`' :21
`
`JUL 30 1999
`JOHN ‘
`-.
`.
`
`ow
`Apple Inc.
`
`w. .. ..
`
`US Patent NO' 8=989=83OCODEN JCMOEH
`ISSN 1337—1307
`
`I
`
`Apple Inc.
`APL1014
`U.S. Patent No. 8,989,830
`
`FITBIT, Ex. 1014
`
`

`

`Volume 14 Number6 August 1998
`
`JOURNAL OF OLINIOAI. MONITORING
`AND OONIPUTINO
`FFIOIAI. JOURNAL OF THE EUROPEAN SOCIETY FOII COMPUTING AND TEOHNOLOOI" IN ANAESTIIESIA ANIl INTENSIVE BANE [ESOTAIIII
`NO OF THE SOEIET‘I FOB TEOHNOLOEY IN ANESTIIESIA [STA]
`
`M B
`
`Harvey L.‘ Edmonds. MD
`John H. Eichhom. Ml)
`TammyY. Euliauo, MD
`M. Fisher, MD
`A. Dean Forbes
`
`llflhedr'iir
`
`11
`
`Wolfgang Friesdorf. MD
`Yasuhiro Fukui. PhD
`Leslie ch des. PhD
`Nikolaus Gravenstein, MD
`A. Gerson Greenburg, MD, PhD
`G. l-ledensticrna, MD
`Wolfgang Heinrichs, MD
`R. C. G. Gallandat Huet, MD
`Kazuyuki Ikeda, MD
`Ville Jantti. MD
`Gavin N. C. Kenny. MD
`Samsun Lampotang, PhD
`j.W. R. McIntyre. MD
`James H. Philip, MD
`Michael L. Quinn, PhD
`Pradccp Ramaaya, MD
`Ira]. Rampil, MD
`Gunter Rau. MD
`Michael F. Roizen. MD
`].—A. Remand, MD
`William B. Rnnciman, MD
`Keith Ruskin, MD
`David A. Sainshury, MD
`H. Schillings, MD
`John W. Severinghaus. MD
`M. Michael Shabot, MD
`Dean Forrest Sittig, l’hl)
`Bradley E. Smith, MD
`Richard Tcplick. MD
`Kevin K. Tremper, MD
`jorge Urzua, MD. PhD
`Karel H. Wesseling, PhD
`Dwayne Westcnskow. PhD
`j. C. Wyatt. PhD
`
`TIIE PARENT JUU IINAlS
`lttm'tmtianttiJorn'nnl of Clinical Alut-tit‘ot‘ittg and Computing,
`founded by Omar Prakash in 1984; Reed Gardner, senior
`editor; and the
`tlrl'mtitort‘t-tg, Founded in 1985 by N.‘i"y
`Jianrntttl' qf Clinical
`Smith, Allen K. Ream. and]. S. Gravenstein.
`'
`
`EUITOIIS
`jefi'tcy M. Feldman. MD
`Allegheny University Hospitals. MCP
`Dept. ofAnesthesiology
`3300 Henry Avenue
`Philadelphia, PA19129, U.S.A.
`E—mail: Feldmalij@al.lhs.cdu
`
`J. S. Gravenstein, MD, Coordinating Editor
`UF/College ochdicine
`Dept. ofAnesthesiology
`P.O. Box100254
`Gainesville, FL 32610. U.S.A.
`E-mail: jgravens@ancst2.anest.ufi.edi|
`Ilkka Kalli, MD, PhD
`City Maternity Hospital
`Dept. of Anaesthesiology
`P.O. Box 61
`FIN-00971 Helsinki, Finland
`E~mailz ilkka.kalli@iki.Fi
`Publisher
`Kluwcr Academic Publishers
`9.0. Box 990
`3300 AZ. Dordrecht,The Netherlands
`E-mail: Locs.Wils@ wkap.nl
`
`EUITONIAI. NANII
`Paul G. Barash, MD
`' Gerard Bashcin. MD, PhD
`Charlotte Bell, MD
`Jan E.W. Benekcn, PhD
`J. A. Elem. PhD
`]. G. Bovill. MD. PhD
`Peter Carstcn sen
`William P. Coleman. PhD
`jefl'rey E. Cooper, PhD
`John Doyle. MD
`Ronald Dueck. MD
`Peter C. Duke. MD
`
`ISSN 1387-1307
`
`All rights reserved
`© 1998 Kluwer Academic Publishers
`No part of the material protected by this copyright notice may be
`reproduced or utilised in any form or by any means. electronic or
`mechanical. including photocopying. recording or by any informa—
`tion storage and retrieval system. without written permission from
`the copyright owner.
`
`Printed in the Netherlands
`
`Printed on «drifter paper
`
`JUL 3 O 1999
`June Cl I-
`6m IIFBHLAHU at
`
`S “M”?
`‘immbuw. wt
`._.
`..
`
`‘IIOLUIIIE I4. 8 ISSUES. IOOO
`Issues include:
`Brief Reviews, Editorials. Clinical Articles. Research Ar-
`ticles. Educational Articles. Book Reviews, Correspondence,
`Selections from the Anesthesia Patient Safety Foundation
`Newsletter.
`
`KittmerAmrit-ntic Pnln'isltrts. Dordrecht. Boston. London
`
`Kluwer Academic Publishers Online
`Visit our website at http; H www.wkap.nl
`
`Thcjrnrrnrti of Clinical tlvltlniwring and Computing is indexed in index
`Mrtlt‘rtts. Cum-tn Contents/Clinical Medicine. Ermnntt llaiedim, Current
`Atlmrt'rtt'ss in Biological Scienrrs. and the Biomedical Engineeriigg Citation
`Index. and is indcxedfabsu'acted in Engineering Index, Srirntt' Citation
`Index. Expanded Resenrrt't Alt-t1, Marika Documentation Service.
`
`II
`
`FITBIT, Ex. 1014
`
`

`

`
`
`I This material may be protected by Copyright law [Title 1? us. Code]
`
`
`
`
`BEFLEBTflNBE PULSE UXIMETIW— PRINCIPLES Mill]
`DBSTETHIG APPLIBATIUN IN THE ZUHIEH SYSTEM
`
`Milli-cht'ii-iiq, Renate Hitch, mid/illicit Hl'ffil
`
`From the Perinatal Physiology Research Department. Department
`of Obstetrics, Zurich University Hospital. CH—Stlfll Zurich. Switzer—
`land.
`
`Received Nov I3, 1997, and in revised formjun 23. 1998, Accepted
`for publication Aug 4,1998.
`
`Address correspondence to Volkcr Kiinig, Perinatal Physiology Re—
`search Department. Department of Obstetrics Zurich University
`Hospital. Cl-LRU‘JI Zurich. Switzerland
`E-mail: vkgt'ct fhk.usz.ch
`
`Jonrimiquii'm'm.‘ .iriuam'mn'ng and Continuing 14: 403-412. 1998.
`I998 Kliirilci'zlrmietirir Publishers. Pri'nicil'i'n lilt' a\-'t'tlit'rlmrds.
`
`sors — and a sensitive modern pulse oximctcr. However,
`
`Kiinig V. lluirh R. Huch A. Reflectance pulse oxilnetry — principles
`and obstetric application in the zurich system.
`J Clin Monit 1998: 14:40:57412
`ABSTBAGT. Transmission and reflectance are the two main
`
`modes ofpulsc oximetry. 1n obstetrics. due to the absence ofa
`transilltnninabie fetal part for transmission oxinietry. the only
`feasible option is the reflectance mode, in which sensor and
`detector are located on the same surface of the bod}J part.
`[-lowever. none of the reflectance pulse oxiineters developed
`for intrapartum use are fully satisfactory, as indicated by the
`fact that none have entered routine use. We have designed,
`developed, constructed and tested a reflectance pulse oxiineter
`with the possibility to adjust the electronic circuits and Signal
`processing in order to determine the effects of various para ni~
`eters on signal amplitude and wave—form and to optimize the
`sensitivity and spatial arrangement ofthe optical elements.
`Following an explanation of the principles of reflectance
`pulse oxiinetry, we report our experience with the design,
`development, construction and field-testing of an in—housc
`reflectance pulse oximctry system for obstetric application.
`
`KEY WEEDS. Oxygen saturation, reflectance pulse oximetry,
`intraparttun feta] monitoring.
`
`INTRIJDIIBTIIIN
`
`l’ulsc oximctry is the combination of spectrophotome—
`try and plethysmography. It permits rapid noninvasive
`measurement of arterial oxygen saturation with the
`added advantages ofsiniple sensor application and direct
`measurement, requiring neither calibration not pre-
`adjustincnt. Pulse oxiinetcts are thus in widespread and
`fast—increasing use. e.g. in intensive care, anesthetics and
`neonatology [1]. All these applications employ “trans—
`mission” pulse oximetry, so called because the light
`used to determine blood oxygen saturation is “trans-
`mitted" from a light emitter on one side of the body
`part to a light receiver on the other side; suitable sites
`are the fingers in adults or hands and feet in neonates or
`children, which are said to be “transilluminated.”
`in obstetrics, fetal oxygen status during labor is a
`crucial parameter. However, no transiliuminablc fetal
`part is available. The only option in this case is reflec—
`tance oximeti'y [2], using a sensor with its light emisi
`sion and detection elements on the same surface of
`
`the body part. Various types of such a reflectance pulse
`oximctcr have been developed for intrapartum use at
`various locations. However, fora wide variety of reasons,
`all are still experimental and not in full routine use [3—7].
`Basically a reflectance measurement can be achieved
`using planar sensors — which can he produced, for
`example. by modifying conventional transmission sen-
`
`403
`
`FITBIT, Ex. 1014
`
`

`

`404 jom'iminfCiinimliid'onfmriirqmid Comparing Vm‘H N06 Augiisri998
`
`PBINBII’LES [IF PIILSEIIXIMETIIY
`
`Signal recording
`
`Light is absorbed on passing through matter. The de-
`gree of absorption depends on the nature of the trans-
`illuminated material and the wavelength of the light
`employed. All optical techniques for determining arte—
`rial oxygen saturation use the marked difference in the
`absorption of red light between oxygenated and re—
`duced hemoglobin.
`The absorption of light passing through bone or
`noupulsatile tissue is constant over time. Oxygenated
`and reduced hemoglobin in the arterial vascular bed, on
`the other hand, cause changes in absorption timed by
`the heart rate due to the pulsatile variation in artery
`thickness. The total intensity of the light after passing
`through tissue can be measured, for example, as the
`photocurrent 1(t) ofa photodiode, and is obtained from
`the Lambert—Beer absorption law as:
`
`I“) = I“ ' exp(_5tisslle ' 5)
`' CXP(_(QO2 ' EHIJO +
`(1— so» - an «ion
`
`5i
`H
`
`dc 1 DC : Iiissur + Idir
`
`where
`
`intensity ofincident light
`In
`slim”: mean absorption coefficient oftissue (function of
`wavelength)
`mean thickness oftrausilluminated tissue
`
`.
`
`30;
`
`oxygen saturation to be determined (: HbO}
`(HbO + Hb), i.e. ratio of oxygenated hemoglo-
`bin concentration to sum of oxygenated and
`reduced hemoglobin concentrations)
`absorption coefficient of oxygenated hemoglobin
`(function of wavelength)
`
`and a signal which varies in time with the pulsatile
`change in artery thickness
`
`AC 3 1mm- ' (502 ' EI-ibo + {l — 502) '5Hb) ' dit)
`
`{4)
`
`with amplitude
`
`ac = l(diastole) — [(systole) =
`
`llissue '
`
`' EHhO + (I — 502) ' EHli) ' d‘
`
`at“,
`
`d(t)
`
`absorption coeflicient of reduced hemoglobin
`(function of wavelength)
`time function of mean pulsatile change in artery
`thickness, with amplitude d = d(diastolc) —
`d (systole)
`
`Measurement may be impaired by light transmitted
`directly from the light source to the receiver or light
`which does not pass through arterially perfused tissue.
`Ifthis "direct light” 1,“, is taken into account, Equation
`(1) changes to:
`
`1(t) 2 [tissue ‘ CXP(*(303 ' E‘Hbo-i-
`(I — SOzi'EHbl‘d(tll+1dir
`
`(“l
`
`where
`
`Ilissul: : It! ' expi—Elissn‘c ' 5)
`
`Since the pulsatile component of the absorption is at
`most a few percent, i.e. the exponent of the second e
`function in Equation (1) is very small, we can use the
`approximation:
`
`expfx) = l +x for
`
`(<1
`
`to obtain the very close approximation:
`
`lit) 2
`[tissue ‘ (1 — {502 ' EHho + l1 — 502) ‘ Eula ' ‘13)) + Idir
`
`(lb)
`
`This light intensity is measured in the photodiodes and
`can be broken down electronically into two compo»
`nents, a time—independent signal
`
`DC 2 Ilissut‘ + Idir
`
`with amplitude equal to the value ofthis signal
`
`(2)
`
`(3)
`
`such instruments come with a “black—box” microproces—
`sor—controlled mode ofoperation making constructional
`adjustments to the electronic circuits and signal process~
`ing virtually impossible. As a result, it becomes difficult
`to determine the cfi‘ect of various parameters on signal
`amplitude and wave-form, optimize sensor sensitivity to
`light intensity and the arrangement of the optical ele—
`ments, and hence assess the dependence of arterial oxygen
`saturation measurement on key physical, technical and
`above all physiological variables. This was the aim
`driving our decision to design, develop and construct
`in-house a system dedicated to obstetric applications.
`Following a brief review of the principles of pulse
`oximetry, we report our experience with the develop—
`ment of the new device,
`together with some field-
`testing results.
`
`The ratio between the ac and dc amplitudes is then
`
`FITBIT, Ex. 1014
`
`

`

`Kemp cl al: The Zurm'r Obsrrrrr't Rqfiermnre Pulse ()ximrrrr
`
`405
`
`l' = flC/dc = (Insure/“tissue + Itllr))'
`(502 -
`€Hbo + (l - 502) ‘ EHbl ' d-
`
`6
`
`(
`
`)
`
`this ratio r is
`In the case that “direct light” 1,“, = 0,
`independent of the incident light intensity In and of the
`absorption in the nonpulsating tissue value ltissuc:
`
`However, despite various theoretical models [9, '10],
`the scatter coefiicients of the various tissue types are
`not known with sufficient accuracy to permit exact
`calculation. Experimental calibration thus has to be
`performed by directly comparing the pulse oximeter
`readings with arterial blood sample values.
`
`r = ac/dc =
`
`(50; ' EH50 + (l — 802) ' E's-ml
`
`' (l
`
`for lair = 0.
`
`(6a)
`
`Transmission pufse oxuuetry
`
`The measured AC signals are seine 10 times weaker
`
`0 No light must be measured that has not passed
`through the pulsatile vascular bed e.g. light passing
`directly from light source to receiver (1:5,).
`I The pulsatile changes in artery thickness must be the
`same for both wavelengths,
`i.e. both wavelengths
`must transilluminate the same tissue region.
`0 Valid measurement assumes that the pulsatile signal
`originates only from varying absorption by arterial
`oxygenated and reduced hemoglobin. The results are
`falsified by other causes ofpulsatile changes in optical
`thickness, e.g. hemoglobin derivatives, circulating
`pigments, pulsatile changes in thickness produced
`mechanically in nonarterially perfused tissue by car—-
`diac action, and, above all, venous pulsation.
`To simplify description ofthe principle behind meas—
`urement and its limitations,
`the Lambert-Beer law
`
`This ratio r is then dependent only on the oxygen
`saturation 302 to be determined, the known absorption
`coefficients eHbo and 5H5, and the mean pulsatile change
`d in the thickness of the arterial vessels in the trans-
`
`illuminated region.
`To eliminate this dependence on d, the measurement
`is performed at two wavelengths with maximally dif-
`fering absorption coefficients. On the assumption that
`the d values are the same for both wavelengths, we
`obtain a variable
`
`R = I‘mi/l‘ir = (RC/dclrcd/(M/dclu
`
`(7)
`
`= (502 - sHbo + (1 c 502) ' EHb)rr—d/
`(502 ' EHho + (l — 502) ' 5Hblir
`
`7a
`
`(
`
`}
`
`from which the unknown 50; is readily calculated
`without knowing the incident light intensity or tissue
`thicknesses.
`
`Calculation assumes the following physical prerequi—
`Sites:
`
`was assumed valid for the passage of light through
`tissue. However, as light is not only absorbed in tissue
`but also scattered, the law is oflimited applicability
`[8}. The exact absorption coefficients must be cor—
`rected by taking the scattering effect into account.
`
`The optical elements are located on opposite sides ofa
`body part. The sensors are applied mainly to the fingers
`and toes. Ears and nose are used only rarely due to poor
`perfusion. ln neonates the sensor is applied around the
`hand or foot. This arrangement largely ensures that the
`optical paths are the same for both wavelengths. Never—
`theless, incorrect sensor attachment can give spurious
`results, e.g. if some of the transmitted light reaches the
`receiving diodes around the outside ofa finger as “direct
`light.”
`Signal magnitudes are an important determinant of
`measurement accuracy: in normal fingertips, the ratio
`ofthe signal due to absorption in pulsating blood (ac)
`to the signal due to absorption in total tissue (dc), r =
`acldc, is 0.02—0.05.
`
`Reflectance pulse oximetry
`
`In this method the light backseattered in the body is
`used to determine oxygen saturation. The optical ele-
`ments are thus located on the same plane on the same
`body surface. Reflection originates from nonhomo-
`geneity in the optical path, i.c. at the interfaces between
`materials with different refractive indices. This means
`
`that on physiological grounds, strong reflections can be
`expected on the entry of light into bone. The trans-
`illuminated tissue must also be well perfused to obtain
`as strong a signal as possible. Not all body parts are as
`well perfused as the fingers or hands, but an acldc ratio
`of 0.001e0.005 can be achieved on the forehead. Perfu—
`sion is also good over the sternum. One method of
`signal enhancement is to heat the measurement site to
`induce hyper-perfusion, which can safely be performed
`up to 42°C. A rubefacient, e.g. nicotinic acid (Rubri-
`ment). can also be applied to the measurement site.
`The principal physical limitations are the following:
`
`The sensor design must eliminate “direct light,” i.c.
`light passing directly from the light sources to the
`photodiodes or that is only scattered in the outer part
`of the skin.
`
`FITBIT, Ex. 1014
`
`

`

`406 Journal ofCliniml Monitoring and Computing Val 14 Na 6 August 1998
`
`than in the transmission method. The conditions
`
`governing the heating of the light—emitting diodes
`(LED) limit
`the potential for producing stronger
`signals by increasing the incident light intensity: not
`only can high uncontrolled temperatures damage
`tissue at the measurement site, but the wavelength
`of the emitted light changes as the LEDs become
`warmer. For this reason the photodiode area must be
`as large as possible.
`As in the transmission mode, the principle of meas—
`urement is the determination of absorption. except
`that this now refers to incoming reflected light. The
`light path is less well defined than in transmission
`mode, and thus may differ between the two wave—
`lengths. The effective absorption coeflicients of the
`calibration inserted in the Lambert—Beer law must be
`
`checked and if necessary corrected by comparison
`with photometrically measured arterial blood values.
`
`neonatal head. Experimentation led to the choice ofa
`vacuum system using sensors cast from silicone rub—
`ber, with a suction groove for fixation, a guard ring
`against direct light, and a connector for a suction
`pump. The photoelectric components are identical in
`both types of sensor.
`I Some metal sensors were fitted with a resistance-wire
`
`heating coil of maximal output 200 mW to induce
`local hyperemia. The temperature was monitored by
`a negative temperature coefficient (NTC) resistor in-
`corporated in the sensor unit.
`
`Numerous types of sensor meeting the above require-
`ments were built. A sensor used for intrapartum meas-
`urements — cast from silicone rubber with suction
`
`channel and pump connection — is shown in cross-
`scction in Figure 1. It is attached to the fetal head with a '
`vacuum of approximately 100 mbar.
`
`TIIE lelllllH IIEFLEBTANEE PIILSE lllllllETEB
`
`Electronics
`
`the 1 kHz sampling rate in the equally spaced se—
`
`o For maximal independence from local tissue differ—
`ences, a radially—symmetric pattern was selected for
`the photoelements. The light source e a chip with
`two LEDS for the wavelengths red = 660 nm and
`infrared : 920 nm A was placed in the center of the
`sensor and surrounded by a radial photodiode array
`for detecting the reflected light. To obtain a good
`signal at minimal
`light intensity,
`the area of the
`photodiodes had to be as large as possible. After some
`preliminary experiments,
`six BX33 photodiodes
`(Siemens) were used with a mean radius of ‘17 mm.
`Their connections are led outwards individually, so
`that the sensor as a whole remains operational if a
`wire breaks or an individual diode is lost. This arrange-
`ment gives an external sensor diameter of 22 mm.
`A guard ring around the LEDs acts as a barrier to
`"direct light." The sensor must also fit snugly to the
`skin to minimize the risk of ambient light reaching
`the photodiodcs.
`Unlike with transmission sensors, fixation to the
`measurement site can pose problems. Aluminum sen—
`sor units are readily fixed with double-sided adhesive
`ECG rings. However,
`these rigid sensor heads are
`unsuited to the small radii of curvature of the fetal]
`
`Sensors
`
`In constructing planar reflection sensors, i.e. sensors in
`which the photoelectric emitting and receiving elements
`lie next to each other in virtually the same plane, special
`attention was paid to the following points (Figure 1).
`
`Initially, we decided to separate signal processing in the
`analog part before the analogfdigital converter (ADC).
`including signal amplification, filtering. and separation
`into DC and AC signals, all handled electronically ; the
`post-ADC digital part was handled by software which
`input the data, averaged and evaluated amplitudes, cal—
`culated saturation and heart rate, and produced a some—
`what complicated sereen display. Now, using modern
`techniques, we have a system in preparation in which a
`fast separate microprocessor unit handles most signal
`processing and digital filtering tasks.
`Figure 2 shows the block diagram of the current
`apparatus with the following individual units:
`
`I Time control of measurement
`
`From a rectified mains power supply signal (100 Hz),
`a phase—locked loop (I’LL) — connected as a frequency
`multiplier — generates a square—wave signal of 64
`kHz. Coupling to the line frequency eliminates data
`acquisition faults due to interferences with the line
`frequency. The 64 kHz from the PLL clock drives a
`' 7-bit counter that addresses an EPROM giving a data
`Cycle of] kHz, 64 pulses long. The output pulses of
`the EPROM control
`the entire sequence of light
`emission, signal acquisition and signal processing.
`LED drive
`
`The oppositely poled red and infrared LEDs are
`located in the output circuit of a current-stabilized
`push—pull output stage. They are triggered by digital
`signals from the control unit via analog switches at
`
`FITBIT, Ex. 1014
`
`

`

`Kc'imlg N m': Thr Zurt'rh Obslrrrir Rqflermmr Prdn- er'mz'fc’."
`
`407
`
`Fig. 1‘ Cress-senior: “trough a silirnnu- rrrhbm' rqflccrfon scusorfor immpm'mm mcasmwneur {fox}!ch saturation.
`
`(D Photodiodes
`C2)
`Red and Infrared LED's
`6)
`Light barrier
`
`FITBIT, Ex. 1014
`
`

`

`408 junrrmquCl'iniml Monitoring and Computing V0114 No6 August 1998
`
`Vacuum
`Pump
`and
`Control
`
`P“?
`Amplifier
`
`Computer
`
`producing a maximal —300 nibar vacuum. The pump
`
`quence: infrared-dark—red—dark (overall length: 1 m5).
`Light intensity is determined by the amplitude of the
`signal driving this output stage. Input signal intensity
`can be selected in two ways:
`4 Manually; The red and infrared LED intensities
`can be manually adjusted independently using two
`potentiometers (Helipot). This permits the use of
`any desired light intensity within the limits stipu-
`lated for test and research purposes.
`Automatically: DC voltage as the input signal
`controls the LEDs so that the DC voltages at the
`computer input for both wavelengths are 2.0 :i: 0.5
`V. Outside this range the control circuit changes
`the LED currents to reset the DC voltages to 2.0 V.
`This setting is used for normal clinical applications.
`To prevent skin damage from overheating even in
`the event of electronic component failure, maximal
`LED intensity is limited by an electronic circuit.
`Input amplifier and sample—and—hold stages
`Using operational amplifiers the photocurrent sup-
`plied by the photodiodes is converted toa voltage
`and then amplified. Six switch positions permit am-
`plifications of 50 menA to 5 V/nA. Afterwards,
`three sample—and—hold (8&H)
`stages — switched
`by the corresponding signals from the control stage -
`resolve the signals into the three components infra—
`red, red and dark. The dark currents are then sub—
`tracted from the red and infrared signals in a subtrac—
`tion stage which also eliminates small ambient light
`components that may have reached the photodiodes.
`
`OscillatorfControl Unit
`
`Ext. Device
`
`(HP, Nellcor)
`
`Fig. 2. Schematic representation qucasuremem electronics.
`
`u Filters
`
`Using low-pass filters the discrete—time signals at the
`8&1-1 output are reconverred to continuous—time sig-
`nals and trimmed of high—frequency components
`using eighth-order Bessel
`filters with a cut—off
`frequency of 7 Hz. The DC components are then
`separated using a low-pass filter with a cut-off fre—
`quency of0.1 Hz. The AC components are passed to a
`further Bessel high-pass 0.7 Hz filter for separation of
`slow motion artifacts, and then to a 40—fold amplifi-
`cation stage. The four signals DC”, DCrcd, AG, and
`ACRE. are thus available at the output. As oxygen
`saturation is calculated from the ratios (acjdch, and
`(acfdc)rcd, it was essential to ensure by careful com—
`ponent selection that the amplifications for the am~
`plification and separation stages were as near as possi—
`ble identical for both wavelengths.
`Patient insulation
`
`the LED controller output
`For patient protection,
`stage and photodiode input stage were electrically
`'isolated from the mains using Burr 8: Brown isola-
`tion amplifiers. These units were powered by an
`insulated power pack.
`Heating
`Sensors incorporating a resistance coil were fitted with
`a precision controller stabilizing temperature within
`the range 38.0—41.0 °C to an accuracy of0.] °C.
`Vacuum pump
`The sensor is fixed to the skin using a small pump
`
`FITBIT, Ex. 1014
`
`

`

`Kiiflig M iii: The Zurich Obstetric Reflerram'e Pulse Oximcrer 409
`
`Start: 33. 1.9? 8:19:21 Tot.TiHa: 60 Min
`RFDXLJ.
`Lxxxxxxx Rxxx geh. 20.03.63 US Datum 23.01.97
`
`0 sec File: LRDBl922.PX2
`"Fox
`802 y
`
`|||||||||||||||||l||||
`
`CTG HR bun
`
`|l|l|
`
`IIIIaIIIIIs
`
`12
`
`—_ _—_E
`
`.1
`
`law is limited by scattering in tissue, experimental
`
`(ADC) card (Mctrabyte). A counter on this card
`triggers analogfdigital conversion at 400 Hz, which
`starts an interrupt program in the computer for read-
`ing the 4 measured values DC,“ DCm; and AC,“
`Acred into a cyclic buffer of 5-seconcl length. From
`the signal (AC5.-+AC,cd) the maxima and minima are
`determined for each cardiac cycle, and hence the
`instantaneous heart rate. Oxygen saturations are cal-
`culated from the amplitudes ac and dc of the AC and
`DC signals.
`Saturation is calculated from the Lambert-Beer
`
`Marker: n
`
`Fig. 3. Street! display of a 10-minute measurement. From top to bottom: oxygen saturation, heart rate, hear: rarefrum HP CTG monitor,
`uterine cairrracriaiifi'om HP CTC monitor. Underneath the DC and AC signals for iiy‘imed and red liglitfor the last 5 seconds (More the
`different ordinate Sfal'csfor DC and AC).
`
`is maintained electronically via a manometer at a
`preset negative pressure. In normal medical use, ad-
`equate fixation is achieved with a vacuum of about
`—l00 mbar.
`
`Sqflware
`
`Apart from various test programs, a program for data
`acquisition, display, calculation and storage and a pro-
`gram for subsequent data postproeessing were written
`in PASCAL.
`
`0 Signal acquisition, calculation and display
`For data acquisition a 486 DOS computer was
`equipped with a 12—bit analogfdigital conversion
`
`the
`for
`law using the absorption coefficients [11]
`nominal wavelengths of our LEDs. However, as
`the actual wavelengths may deviate from these nom—
`inal values and the applicability of the Lambert-Beer
`
`FITBIT, Ex. 1014
`
`

`

`410 journal quiiniml Monitoring and Ceiiipiriing Voi14 No 6 August 1998
`
`calibration ofthe measured saturation values is essen-
`tial.
`
`Calibration
`
`Saturation and heart rate can be averaged over 1—9
`cardiac cycles and are displayed every second. Analog
`heart rate and uterine contraction signals from any
`CTG monitor With analog output
`(e.g. Hewlett
`Packard model 8040) can be input to the ADC every
`second and likewise displayed.
`The measurement display (Figure 3) shows, along the
`bottom, the 4 measured values DC,“ DCmd and ACi“
`ACrcd, over the last 5 seconds of measurement. This
`serves to monitor signal quality during acquisition.
`Poor—sign a1 periods can be marked and excluded from
`data processing. Along the top, measured arterial
`oxygen saturation and heart rate values per second are
`displayed cyclically over a 10-minute interval, with
`the CTG heart rate and contraction input under-
`neath. Any time point can be marked for subsequent
`identification and all values and comment stored in a
`
`file at any time.
`Data analysis
`The values from a stored file can be redisplayed in
`measurement mode using an evaluation program.
`Time intervals can be marked with the arrow keys or
`mouse. Means and standard deviations — including
`the CTG data # are then calculated and displayed.
`
`
`
`HEASIIIIEHIEHTS Mil IIISEIJSSIIIN
`
`Clinical application of any new instrument or measure-
`Inent system presumes:
`
`Approximately 100 what is the most suitable pressure
`
`7 mechanical reliability, accuracy and calibration,
`— feasibility in the clinical situation, including accept—
`ance by both medical personnel and patients,
`the ability not only to determine physiological pa-
`rameters not previously measured in both physiolog—
`ical and pathological situations, but also to evaluate
`the diagnostic significance of such parameters, in this
`case oxygen saturation.
`For the first two more technical points is to say, that
`our instrument required calibration before clinical
`use, together with field tests ofthe suction device and
`long—term oximeter performance during birth.
`For
`these points controls and clinical
`trials were
`performed in our own unit and with colleagues in
`Copenhagen (DK), Graz (A), Oulu (SF) and Berlin
`{D}. The major investigations comprised:
`
`To calibrate a pulse oximeter, an oxygen saturation
`value must be assigned to the measured variables
`
`R =
`
`: (adders/(acacia
`
`cf {7)
`
`on the basis of an experimental or theoretical relation-
`ship. Initially we used the absorption coefficients of
`Zijlstra e1 (ii [1 l]. The general problems of calibrating a
`pulse oximeter have been discussed elsewhere [12, 13].
`For fine calibration we performed the following inves-
`tigations:
`
`o Tests in the arterial oxygen saturation range 88—
`lOO% were conducted in 14 healthy adult volunteers
`breathing normal air and then air with approxi-'
`mater 80% normal oxygen content for 10 minutes
`in each case. The reflection sensor was fixed to the
`
`forehead or sternum with an adhesive ring. Owing to
`the invasive nature of arterial catheterization. refer~
`
`ence values were provided by a MINOLTA PUL-
`SOX 8 transmission pulse oximeter attached to the
`index finger. Data analysis [13] showed a 4.5% differ-
`ence in oxygen saturation between the MINOLTA
`and the preliminary results of our reflectance system
`based on the absorption coefiicents onisztra et a].
`At lower saturation levels, measurements were per—
`formed in cyanotie children before surgery. The chil—
`dren had arterial lines, permitting direct comparison
`with arterial bleed readings [14].
`Low saturations in viva can also be measured in the
`
`fetal lamb [15]. We used this method to compare our
`pulse oximeter readings directly with arterial values
`in the oxygen saturation range 10—80% [16].
`
`Preliminary evaluation of these data shows that our
`previous calculations of oxygen saturation have to be
`corrected in the ill—100% saturation range by a factor of
`1.045.
`
`Fixation
`
`The first experiments in sensor fixation to the head and
`other parts of the human body were performed in
`adults [17] and neonates [18].
`Attachment is simple in practice, even during birth.
`After rupture ofthe membranes, the sensor can be fixed
`to the fetal head once the cervix has dilated to at least 2
`cm. Initial fixation takes 30—60 seconds and allows full
`
`freedom ofmovement [19].
`
`FITBIT, Ex. 1014
`
`

`

`. johnson NJohnson VA, Fishcerobbings B, Bannister].
`Lilford R]. Feta] monitoring with pulse oxinietry. Br]
`Obstet Gynaecol 1991; 98: 36—41
`. Knitza R,Buschmann], Rall G. Ein neues Verfahren zut
`kontinuierlichen Messuug der Fetalem Sauerstofl‘séitti-
`gong sub patru. Geburts Frauenheilk 1992; 52: 319—321
`. Dildy GA, Clark SL, Loucks CA. Preliminary experience
`with intrapartnm fetal pulse oxiinetry in humans. Obstet
`Gynecol 1993; 81: 630—635
`. Hypoxische Gef‘a'hrdung des Fetus sub partu. ln: Knitza
`R (ed). Darinstadt: Steinkopf1994
`. Wukitsch MW, I’etterson MT, Tobler DR, PologeJA.
`Pulse oxi merry: Analysis of theory, technology and prac—
`tice.j Clin Monit1988;4: 290—301
`. Graaff R. Tissue optics applied to reflectan

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket