throbber
o
`
`0
`
`’ Blrd & Blrd (
`
`Bird & Bird LLP
`
`Maximiliansplatz 22
`
`80333 Miinchen
`Germany
`
`Tel +49 (0) 89 3581 6000
`Fax +49 (0) 89 3581 6011
`twobirds.com
`
`BIRD& BIRD LLP - Maximiliansplatz 22 - D<80333 Miinchen
`
`Via facsimile in advance: +49 89 2399 4465 (without exhibits)
`European Patent Office
`Bob—van-Benthem-Platz 1
`80469 Miinchen
`
`Dr. Daniela Kinkeldey
`Ext: +49 (0) 89 3581 6229
`daniela.kinkeldey@twobirds.com
`
`Munich, 17 March 2016
`
`Our ref. (please always state): EDWLI.0054
`
`The facts and arguments (point VII) for the
`
`MOE
`
`filed in the name and on behalf of Edwards Lifesciences Corporation, 1 Edwards Way
`
`Irvine, CA 92614, USA,
`against
`the European
`“Repositionable heart
`valve”
`(European
`patent
`
`patent EP 2 749 254 Bl
`entitled
`application No. 14 161 991.6)
`of
`
`Boston Scientific Scimed, Inc., One Scimed Place, Maple Grove, MN 55311-1566, USA are
`
`presented in the following:
`
`The opposition fee is to be debited from our account 28002046.
`
`A.
`
`Requests
`
`1.
`
`It is requested to revoke the patent in its entirety, based on the grounds that:
`
`1.
`
`2.
`
`3.
`
`its subject matter is not patentable within the terms of Articles 52 to 57 EPC
`(Art. 100(a) EPC), since it is neither novel'nor based on an inventive step;
`
`the opposed patent does not disclose the invention in a manner sufficiently clear
`and complete for
`it
`to be carried out by a person skilled in the art
`(Art. 100(b) EPC, Art. 83 EPC); and
`
`the subject matter of the opposed patent extends beyond the content of the
`application as filed (Art. 100(c) EPC, Art. 76, 123(2) EPC).
`
`Abu Dhabi & Beijing & Bratislava & Brussels & Budapest& Copenhagen & Dubai & Diisseldorf& Frankfurt & The Hague & Hamburg & Helsinki & Hong Kong
`& London & Luxembourg & Lyon 8: Madrid & Milan & Munich & Paris & Prague & Rome & Shanghai & Singapore & Skanderborg & Stockholm & Sydney & Warsaw
`Bird 8: Bird LLP is! einc Parrnerschaft mit beschrfinkrer Haftung nach englischem Recht, eingetragen im Companies House of England and Wales unter der Nummer 0C340318. Der Name Bird &
`Bird bezeichnet eine internationale Anwaltssozietat, bestehend aus Bird & Bird LLP und ihren verbundenen Sozietaten. Bird & Bird praktiziert in den oben angegebenen Standorten. Weitere
`Informationen finden Sie auf unserer Homepage.
`
`Bankverbindung: Deutsche Bank Privat- und Geschaftskunden AG, Kto. Nr. 1066 398, BLZ 300 700 24, SWIFT/BIC Code DEUTDEDBDUE, [BAN Code DE 71 3007 0024 0106 6398 00
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 1 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 1 of 56
`
`

`

`Bird & Bird
`
`17 March 2016
`page 2
`
`II.
`
`In accordance with the Guidelines for Examination, Part E, Section VII, P0int#4
`accelerated processing is requested since an infringement action in respect of the
`European patent is pending before the regional court of Dusseldorf, Germany (case
`number 4aO 137/15).
`'
`
`III. Oral proceedings pursuant to Art. 116 EPC are requested.
`
`B.
`
`Relevant References
`
`1.
`
`The opposition is based on the following prior art:
`
`1331
`
`BB2
`
`BB3
`
`BB4
`
`BB5
`
`BB6
`
`BB7
`
`BB8
`BBQ
`
`Feature analysis of claim 1 of opposed patent
`
`WC 2005/062980 A2 (PCT application of opposed patent)
`
`WO 98/29057 A1
`
`US 4,084,268 A
`
`US 5,855,601 A
`
`W0 03/047468 A1
`
`EP 1 702 247 A0 (parent application of opposed patent)
`
`US 10/971,535 (priority application of opposed patent)
`Assignments filed with the US Patent Office for priority document of
`
`opposed patent
`
`BB10
`
`WO 2005/102015 A2
`
`BBloA
`
`US 60/565,118 (priority application of BB10)
`
`BB11
`
`BB12
`
`BB13
`
`BB14
`
`BB15
`
`BB16
`
`BB17
`
`BB18
`
`WO 98/55047 A1
`
`US 6,015,431 A
`
`W0 03/037222 A2
`
`US 6,729,356 B1
`
`US 2001/0027338 A1
`
`US 2004/0082989 A1
`
`US 2003/0236567 A1
`
`US 5,411,552 A
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 2 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 2 of 56
`
`

`

`Bird 8: Bird
`
`17 March 2016
`page 3
`
`Feature analysis of claim 1 and the alleged invention
`
`C.
`
`2.
`
`The opposed patent comprises 13 claims including independent claim 1. All claims of
`
`the opposed patent relate to an apparatus for endovascularly replacing a patient’s
`
`heart valve.
`
`1.
`
`3.
`
`Below is a feature analysis of independent claim 1 which provides the following
`
`features:
`
`Apparatus for endovascularly replacing a patient’s heart valve,
`comprising:
`
`the apparatus
`
`1.
`
`an expandable anchor (30)
`
`La) supporting a replacement valve (20),
`1b) the anchor having a delivery configuration and a deployed configuration,
`characterized by
`
`2.
`
`a fabric seal (38o)
`
`2.a) extending from the distal end of the valve (20) proximally over the anchor in
`the delivery configuration,
`
`2.b) wherein the seal is bunched up in the deployed configuration.
`
`4.
`
`The above feature analysis is enclosed as a separate document (Exhibit BB1).
`
`II.
`
`5.
`
`In general, the alleged invention relates to an apparatus for endovascularly replacing a
`heart valve. The apparatus claimed by independent claim 1 of the opposed patent
`includes an expandable anchor (feature group 1) and a fabric seal (feature group 2).
`
`6.
`
`According to feature 2.a) of claim 1, the seal extends from the distal end of the valve
`
`(20) proximally over the anchor in the delivery configuration.
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 3 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 3 of 56
`
`

`

`Bird 8: Bird
`
`17 March 2016
`page 4
`
`Feature 2.a) is interpreted to the effect that the seal extends proximally from the distal
`
`end of the valve, i.e. the valve being the reference for the extension of the seal.
`
`Added subject matter (Art. 1001c), 1611! and 1231 2) EPC)
`
`D.
`
`The opposed patent (EP 2 749 254 B1) was filed as a divisional of EP 1 702 247, which
`is based on the PCT application W0 2005/062 980 A2 (hereafter: “W0i98o”,
`Exhibit BB2). The relevant description and the figures of the opposed patent as filed
`are identical. to the description and figures of WO ‘980. Therefore, in the following
`reference will be only made to the disclosure of WO ‘980 and Art 76(1) EPC. However,
`
`the exact same arguments and reasoning also applies for the opposed patent as
`
`originally filed and Art 123(2) EPC.
`
`In accordance with G 1/05 (OJ EPO 2008, 271) and G 1/06 (OJ EPO 2008, 307), it is a
`condition that anything disclosed in the granted patent must be directly and
`unambiguously derivable not only from the original application on which the patent
`has been granted, but also from what was disclosed in each of the preceding
`applications as filed. Accordingly, the subject matter claimed in the granted claims of
`the opposed patent must also be disclosed in WO’980 (Art 76(1) EPC).
`
`10.
`
`11.
`
`In T 219/09 the board of appeal of the EPO held that, according to established case
`law, it will normally not be allowable to base an amended claim on the extraction of
`isolated features from a set of features originally disclosed only in combination, e.g. a
`
`specific embodiment in the description (intermediate generalization).
`
`is only exceptionally
`Omitting features that are originally disclosed in context
`permissible in the absence of any clearly recognizable functional or structural
`relationship among the features of the specific combination originally disclosed
`(cf. decision T 1067/97) and if the omitted feature is not inextricably linked with those
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 4 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 4 of 56
`
`

`

`Bird 8: Bird
`
`17 March 2016
`page 5
`
`features (cf. decision T 714/00). This is not the case here and the claims as granted are
`
`an intermediate generalisation of the disclosure of WO’980.
`
`I.
`
`General Remarks
`
`12.
`
`In WO’980 a heart valve prosthesis is disclosed e.g. in Fig 107C, where the fabric seal
`
`is disclosed to bunch up in the deployed configuration (page 85, 11. 28-29). This
`
`Fig 1070 is pictured below:
`
`70
`
`
`
`13.
`
`The heart valve prosthesis shown above is substantially similar to the prosthesis
`marketed by the patentee as the so-called Lotus Valve. This valve looks as follows as
`may
`be
`taken
`from
`the
`homepage of
`Boston
`Scientific
`(http: //www.bostonscientific.c0m/en—US/products/transcatheter-heart—valve/lotus-
`
`valve-system.htrnl) :
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 5 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 5 of 56
`
`

`

`Bird & Bird
`
`17 March 2016
`page 6
`
`
`
`14.
`
`The heart valve prosthesis disclosed in WO‘980 is a very specific type of heart valve
`prosthesis and the components of this prosthesis are all required in order for the heart
`valve to function properly. The required features,
`including repositionable self—
`expanding frame, active foreshortening of the anchor using actuators, a locking-
`mechanism to maintain the diameter of the foreshortened anchor and additional
`
`features are disclosed as essential for the function of the apparatus as claimed.
`
`15.
`
`The patentee, however, left out numerous essential features in claim 1 that are
`disclosed in the description of WO’980 to stand in clear functional and structural
`
`context with the features specified in claim 1 as granted.
`
`(1)
`
`16.
`
`The introduction part on pages 1-4 of WO’980 sets out which technical problems are
`present in the devices of the prior art. The general part of the description then
`discloses which features of the claimed apparatus allegedly overcome these problems.
`
`17.
`
`First of all, the introductory part of WO’980 (page 3, 11. 21-24) discloses:
`
`“While providing for an aortic approach, devices described in the Garrison
`patent application sufferfrom several drawbacks. First, the stent portion of the
`device is delivered across the native valve as a single piece in a single step,
`which precludes dynamic repositioning of the stent during delivery.”
`[emphasis added]
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 6 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 6 of 56
`
`

`

`Bird 8: Bird
`
`17 March 2016
`page 7
`
`18.
`
`Indeed it
`
`is described throughout
`
`the specification that
`
`the apparatus
`
`is
`
`repositionable, starting already with the title of WO’980 reading “REPOSITIONABLE
`
`HEART VALVE” and including more than 40 instances of
`
`“
`
`reposition” or
`
`“repositioning”.
`
`19.
`
`Thus, the feature of the apparatus to be repositionable is a core element of the
`
`technology disclosed in the opposed patent. Accordingly, the apparatus claimed in the
`
`opposed patent must be repositionable, which is an essential feature that has been
`
`omitted from claim 1.
`
`(2)
`
`20. A further problem of the prior art devices is disclosed on page 2, 11. 11-13 of WO,’980,
`
`namely the relatively large cross-sectional delivery profile of prior art stent valves:
`
`‘iAnother drawback of the PVT device is its relatively large cross-sectional
`delivery profile. The PVT system's stent/valve combination is mounted onto a
`delivery balloon, making retrograde delivery through the aorta challenging.”
`
`[emphasis added]
`
`21. Accordingly, the description teaches in the context of the diameter of the apparatus
`
`(W0’98o, page 23,11. 1-14) that:
`
`“[...] in order to avoid delivery of anchor 30 on a balloonfor balloon expansion,
`a non-hydraulic or non—pneumatic anchor actuator is used.“
`
`[emphasis added]
`
`22.
`
`In other words, leaving out the balloon solves the problem of a relatively large device
`
`diameter. Thus, as will be further outlined in more detail also below, the apparatus
`
`must include a non—hydraulic or non—pneumatic anchor actuator, which is another
`
`essential feature that has been omitted from claim 1.
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 7 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 7 of 56
`
`

`

`Bird & Bird
`
`17 March 2016
`page 8
`
`(3)
`
`23.
`
`A further problem to be solved is disclosed on page 3, 11. 3—5 of WO’980, namely that
`
`anchoring of prior art valve stents t0 the vessel walls is problematic:
`
`“However when the stent has a valve fastened inside it, as is the case in aortic
`valve replacement, the anchoring of the stent to vessel walls is significantly
`challenged during diastole.”
`
`[emphasis added]
`
`24.
`
`Accordingly, WO’980 (page 9,11. 1-5) discloses that this problem is solved by including
`
`leaflet engagement elements:
`
`“The expandable anchor filrther includes a leaflet engagement element on its
`proximal end to engage the leaflets of the patient’s heart valve. [...] Moreover,
`once engaged, the leaflet engagement element prevents the distal movement of
`the anchor.”
`
`[emphasis added]
`
`25.
`
`Consequently, the apparatus must include a leaflet engagement element in order to
`prevent movement of the anchor, yet another essential feature that has been omitted
`from claim 1.
`
`(4)
`
`26.
`
`A further problem to be solved is mentioned on page 2, 11. 29-30 and page 3, 11. 9—11 of
`
`WO’980 that the prior art devices lack radial strength:
`
`“Another drawback of prior art self—expanding replacement heart valve
`systems is their lack of radial strength.”
`
`‘Moreover, a self-expanding stent without sufficient radial force will end up
`dilating and contracting with each heartbeat, thereby distorting the valve,
`aflecting itsfunction and possibly migrating and dislodging completely.”
`
`[emphasis added]
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 8 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 8 of 56
`
`

`

`Bird 8: Bird
`
`17 March 2016
`page 9
`
`27.
`
`In the context of the radial force exerted by the stent, page 3, 11. 5—6 of WO’980 teaches
`
`that it is important that the valve stent must prevent blood from flowing back into the
`
`inside of the ventricle. In other words, the valve stent must have sufficient strength to
`
`provide a good sealing property to avoid regurgitation:
`
`“The force to hold back arterial pressure and prevent blood from going back
`inside the ventricle during diastole will be directly transferred to the
`stent/vessel wall interface. Therefore the amount of radial force required to
`keep the self—expanding stent/valve in contact with the vessel wall and not
`sliding will be much higher than in stents that do not have valves inside of
`them.”
`
`28.
`
`This problem is solved by providing an anchor that is actuated using an external non-
`hydraulic or non—pneumatic force to actively foreshorten the anchor. Page 23, ll. 1—2 of
`
`WO’980 discloses:
`
`“Anchorgo may be actuated using external non-hydraulic or. non-pneumatic
`force to activer foreshorten in order to increase its radial strength.”
`
`[emphasis added]
`
`29.
`
`The fact, that an active foreshortening of the anchor is required to increase radial
`
`strength is further disclosed on page 11, 11. 12—16 of WO’980:
`
`“Another aspect of the invention provides an apparatus for endovascularly
`replacing a patient’s heart valve, including: an anchor having a collapsed
`delivery configuration and an expanded deployed configuration; and a
`replacement valve coupled to the anchor, wherein the anchor comprises
`enhanced radial strength in the expanded deployed configuration as compared
`to the collapsed delivery configuration due to imposed foreshortening.”
`
`[emphasis added]
`
`30.
`
`Therefore, the apparatus must additionally include that the non-hydraulic or non—
`pneumatic actuator actively foreshortens the apparatus during deployment, which is
`another essential feature that has been omitted from claim 1.
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 9 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 9 of 56
`
`

`

`Bird 8: Bird
`
`17 March 2016
`page 10
`
`(5)
`
`31. On page 4, 11. 1-3 of WO’98o it is further emphasized that the apparatus of the claimed
`invention allegedly solves fl those drawbacks mentioned in the context of the prior
`art:
`
`“In view of drawbacks associated with previously known techniques for
`percutaneously replacing a heart valve,
`it would be desirable to provide
`methods and apparatus that overcome those drawbacks.”
`
`Thus, all features discussed above are clearly essential features of the claimed
`apparatus based on the general description alone. But there are in fact numerous
`further additional features disclosed that are required and essential for the function of
`
`the claimed apparatus which becomes clear in the context of the disclosure of the
`
`specification as a whole and will be explained in the following.
`
`II.
`
`Features omitted in the context of the anchor
`
`(1)
`
`Active foreshortening of anchor with actuators
`
`32.
`
`The foreshortening of the anchor is only disclosed as an active foreshortening of the
`anchor using actuators as is further specified with respect to Figure 107. This follows
`
`from the disclosure on page 85, 11. 26—28 of WO’980, which reads:
`
`“Next, Figure 107C, as the apparatus is actively toreshortened using proximal
`(e.g., fingers) and/or distal actuators
`(e.g., elements 55),
`the leaflet
`engagement elements positively register with the native valve leaflets.”
`
`[emphasis added]
`
`33-
`
`Thus, in the context of the claimed device only an active foreshortening of the anchor
`
`using actuators is disclosed.
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 10 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 10 of 56
`
`

`

`Bird 8: Bird
`
`17 March 2016
`page 11
`
`34.
`
`Furthermore, an active foreshortening is required to provide the stent with sufficient
`
`radial strength. To this end, WO’980 discloses on page 23, 11. 1-2:
`
`“Anchor 30 may be actuated using external non—hydraulic or non—pneumatic
`force to activelyforeshorten in order to increase its radial strength.”
`
`is expressly mentioned in the
`35. That the increase in radial strength is essential
`introductory section on page 2,11. 29—30 of WO’980. That the radial strength, achieved
`
`by an active foreshortening of the anchor by actuators, is a requirement also for the.
`
`sealing function is explained on page 27, 11. 3—5 of WO’980 as follows:
`
`“Once properly aligned, actuators 50 are retracted relative to actuators 60 to
`impose foreshortening upon anchor 30 and expand apparatus 10 to the fully
`deployed configuration, as in Figure 5D. Foreshortening increases the radial
`strength ofanchor 30 to ensure prolonged patency ofvalve annulus An, as well
`as to provide a better seal for apparatus 10 that reduces paravalvular
`regurgitation. As
`seen
`in Figure
`5E,
`locks 40 maintain
`imposed
`foreshortening.”
`
`[emphasis added]
`
`36. Thus, the essential feature that the anchor must include actuators in the delivery
`configuration to be actively foreshortenable for deployment has been omitted from
`
`claim 1 as granted.
`
`37.
`
`The patent application as originally filed defines the degree of foreshortening by
`referring to the ratio of deployed length to collapsed/delivery length (cf. BB2, page 50, ’
`
`11. 7—8). The smaller this ratio gets the greater the degree of foreshortening.
`
`38. The claimed apparatus is allegedly based on figures 32-34 and 107 and the respective
`description in the patent application as originally filed. However, every apparatus that
`is disclosed foreshortens significantly, namely by a ratio of 0.5 or below. When
`measured, the ratio of the deployed to collapsed lengths shown in originally filed
`figures 107 B/C, 32 and 33 is in each case about 0.4. Also the description does not
`disclose any foreshortening ratios above 0.5 (cf. BB2, page 50,11. 7-8).
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 11 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 11 of 56
`
`

`

`Bird 8: Bird
`
`17 March 2016
`page 12
`
`'
`
`39-
`
`Thus, it has also been omitted from claim 1 as granted that the anchor foreshortens
`
`significantly, namely by a ratio of the deployed to collapsed lengths of 0.5 or below.
`
`Non-hydraulic or non-pneumatic anchor actuators
`
`(2)
`
`4o.
`
`Additionally, the description does not allow the use of any actuator but requires the
`
`use of specific actuators, namely non—hydraulic or non—pneumatic anchor actuators.
`
`41.
`
`On page 21, 11. 25-32 WO’980 discloses that a deployment tool is used to place the
`apparatus of the alleged invention by applying a non-hydraulically expanding or non-
`
`pneumatically expanding force on the anchor:
`
`invention relates to apparatus (...) for endovascularly or
`“The present
`percutaneously delivering and deploying a prothesis, e.g., an aortic prosthesis,
`(...). A delivery system and/or deployment tool is provided including a sheath
`assembly and a guidewirefor placing the prosthetic apparatus endovascularly
`within the patient and a user control allowing manipulation of the prosthetic
`apparatus from external to the patient through the application at a non—
`h draulicall
`e
`andin
`or
`non— neumaticall
`e
`andin
`orce on the
`anchor.”
`
`[emphasis added]
`
`42.
`
`This force applied by the deployment tool obviates the use of a balloon. This is
`
`disclosed e.g. page 23, 11. 12-14 of WO’980:
`
`“A deployment tool is used to actuate, reposition, lock and/or retrieve anchor
`30. In order to avoid delivery of anchor 30 on a balloonfor balloon expansion,
`a non-hydraulic or non—pneumatic anchor actuator is used.”
`
`43.
`
`is also required to actively
`tool/actuator
`the mentioned deployment
`Further,
`foreshorten the anchor. In this connection, the specification of WO’980 discloses on
`
`page 85, 11. 26-28 that the active foreshortening is effected by using proximal and
`distal actuators:
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 12 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 12 of 56
`
`

`

`Bird & Bird
`
`17 March 2016
`page 13
`
`“Next, Figure 107C, as the apparatus is actively toreshortened using proximal
`Leg,
`fingersZ andgor distal actuators
`(e.g., elements 55),
`the leaflet
`engagement elements positively register with the native valve leaflets.”
`
`,
`
`[emphasis added]
`
`44.
`
`Accordingly, Figure 107 shows this essential feature of the actuators in the context of
`
`the apparatus as claimed.
`
`45.
`
`In the absence of an anchor configured to be controlled by proximal (e.g., fingers)
`
`and/or distal actuators there would be no active foreshortening of the anchor.
`
`Accordingly, a further essential feature has been inadmissibly left out from the claims
`of the opposed patent, namely that the active foreshortening of the anchor is effected
`
`by the use of non-hydraulic or non—pneumatic proximal and/or distal actuators.
`
`.
`
`(3)
`
`Lockable anchor and locks
`
`47.
`
`The anchor of the claimed apparatus must be actively foreshortenable as outlined
`
`above. However, the active foreshortening of the anchor alone is not sufficient for the
`
`apparatus to function properly. Instead, the foreshortened condition of the anchor,
`which follows from an active foreshortening, also needs to be maintained.
`
`This maintenance of the imposed foreshortening is achieved by using a lockable
`
`anchor, which is disclosed on page 11, 11. 16-18 of WO’980 as follows:
`
`“The apparatus may include a locking mechanism [or maintaining imposed
`toreshortening, and it may be configuredfor retrieval prior to actuation of the
`locking mechanism.”
`
`[emphasis added] .
`
`49.
`
`The fact, that a lockable anchor is an essential feature is further confirmed by WO’980
`
`in the section explaining the use of the claimed apparatus. By referring to Figure 107,
`
`WO’980 discloses on page 85, 11. 32-34:
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 13 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 13 of 56
`
`

`

`Bird 8: Bird
`
`17 March 2016
`page 14
`
`“Once the anchor is fully compressed within the aortic valve, the anchor is
`
`locked the fingers and post mandrels are disengaged, and the seal is adapted
`tofurther limit bloodflow around the replacement valve.”
`
`[emphasis added]
`
`50. Thus, the deployed state can only be maintained by locking the foreshortened state of
`
`the anchor. Thus, the lockability of the anchor is an essential element of the claimed
`
`apparatus as well.
`
`51.
`
`In order for the anchor to be lockable it must comprise locks which maintain the
`
`foreshortened shape of the anchor. This is disclosed by WO’980 on page 27, 11. 3-5:
`
`“Once properly aligned, actuators 50 are retracted relative to actuators 60 to
`impose toreshortening upon anchor 30 and expand apparatus 10 to the fully
`deployed configuration, as in Figure 5D. Foreshortening increases the radial
`strength ofanchor 30 to ensure prolonged patency ofvalve annulus An, as well
`as to provide a better seal for apparatus 10 that reduces paravalvular
`regurgitation. As
`seen
`in Figure
`5E,
`locks 40 maintain imposed
`toreshortening. ”
`
`[emphasis added]
`
`52.
`
`The respective locks are described by WO’980 on page 26, 11. 8-14 with reference to
`
`Figure 5E and to Figures 4D-4F as follows:
`
`“In Figure 4D, continuedforeshortening causes male interlocking elements 44
`of locks 40 to engagefemale interlocking elements 42. The male elements mate
`with the female elements, thereby locking apparatus 10 in the foreshortened
`configuration, as seen in Figure 4E. Actuators 50 are then pulled through
`eyelets 45 of male elements 44 to remove the actuatorsfrom apparatus 10, and
`actuators 62 are pulled through the proximal end of anchor 30 to uncouple
`actuators 60from the apparatus, thereby separating delivery system 100from '
`apparatus 10. Fully deployed apparatus 10 is shown in Figure 4F.”
`
`53.
`
`Similarly, WO’980 (page 48, 11. 28—31) provides by reference to Figures 49B, 50B and
`
`50C:
`
`“Anchor 30 comprises a plurality of anchor lock elements 34, e.g., buckles 34,
`attached to its proximal region, onefor each post 32. Posts 32 may comprise a
`lock element that forms a two—part
`locking mechanism with anchor lock
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 14 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 14 of 56
`
`

`

`Bird & Bird
`
`17 March 2016
`page 15
`
`elements 34 for maintaining anchor 30 in a deployed or expanded
`configuration (e.g., as illustrated in Figures 493, 503 and 50C).”
`
`54.
`
`As follows from the above passages of WO’980, the claimed apparatus must be
`
`equipped with both, a lockable anchor and locks, which, therefore, both are essential
`
`features of the apparatus according to the teaching of opposed patent.
`
`(4)
`
`Self-expandable anchor
`
`55.
`
`Using a self—expandable anchor relates to a problem stated in the introductory part of
`WO’980, namely to reduce the cross—sectional delivery profile of the apparatus. This is
`
`achieved by means of a self-expandable anchor since an increase in diameter due to
`
`the extra space that would be required by a balloon can be avoided.
`
`The need to eliminate the use of a balloon for delivery of the anchor is set out on
`
`page 23, 11. 12—14 of WO’980 as follows:
`
`“A deployment tool is used to actuate, reposition, lock and/or retrieve anchor
`30. In order to avoid delivery of anchor 30 on a balloon for balloon expansion,
`a non-hydraulic or non-pneumatic anchor actuator is used.”
`
`[emphasis added]
`
`57-
`
`Thus, the configuration of the anchor needs to allow a delivery of the anchor by means
`of a non-hydraulically expanding or non—pneumatically actuator instead of a balloon
`catheter as already set out above. However, this is not the entire disclosure on the
`deployment of the anchor. It is additionally disclosed in relation to Figures 107A—107C
`on page 84, line 34 — page 85, line 2 that:
`I
`
`“As the sheathing catheter is pulled proximally out of the native valve, anchor
`30 and replacement valve 20 become unsheathed. Immediately the portion of
`the unsheathed anchor 30 dynamically self—expands (...).”
`
`Accordingly, it is also an essential feature that the anchor is self—expandable.
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 15 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 15 of 56
`
`

`

`Bird 8: Bird
`
`17 March 2016
`page 16
`
`(5)
`
`Repositionable and retrievable anchor
`
`59-
`
`The steps shown in Figures 107A-107C reflect what
`
`is also disclosed in the
`
`specification of WO’980 on page 23, 11. 12—14:
`
`“A deployment tool is used to actuate, reposition, lock and/or retrieve anchor
`30. In order to avoid delivery of anchor 30 on a balloon for balloon expansion,
`a non—hydraulic or non—pneumatic anchor actuator is used.”
`
`6o.
`
`Also in Figure 107 such a deployment tool is used. Thus, the anchor 30 must be
`
`configured to be used with actuators in order to be repositionable. This type of anchor
`
`is used throughout the description.
`
`61.
`
`Furthermore, the heart valve prosthesis is only repositionable with the actuators until
`
`the device is locked. Thus,
`
`the locks are also required in conjunction with a
`
`repositionable valve prosthesis once the correct position has been reached (see also
`
`comments above concerning the requirements for locks).
`
`62.
`
`The anchor is only repositionable if it is also retrievable. Thus, it is an essential feature
`
`that the anchor is also retrievable and repositionable.
`
`(6)
`
`Anchor posts on the inside
`
`The anchor must comprise posts on the inside also because the valve needs to be
`
`attached to the posts there. In this context, WO’980 discloses on page 22, 11. 31-33:
`
`“Annular base 22 of replacement valve 20 preferably is coupled to skirt region
`34 of anchor 30, while commissures 24 of replacement valve leaflets 26 are
`coupled to posts 38.”
`
`As shown in Figure 1 of WO’980, posts 38a, 38b and 38c are attached to the anchor.
`Further embodiments of posts are shown in Figure 18 and Figures 50-62 of WO’980.
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 16 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 16 of 56
`
`

`

`Bird & Bird
`
`17 March 2016
`page 17
`
`As such, posts are an essential element and are required for an active foreshortening
`
`of the anchor.
`
`65. According to WO’980 the active foreshortening achieved by the anchor posts and
`
`actuators in turn is required to create a seal. This is disclosed for example on page 27,
`
`11. 1—6:
`
`“During foreshortening, actuators 60 push against lip region 32 of anchor 30,
`while actuators 50 pull on posts 38 of the anchor. Once properly aligned,
`actuators 50 are retracted relative to actuators 60 to impose foreshortening
`upon anchor 30 and expand apparatus 10 to the fiilly deployed configuration,
`as in Figure 5D. Foreshortening increases the radial strength of anchor 30 to
`ensure prolonged patency of valve annulus An, as well as to provide a better
`§e_alfor apparatus 10 that reduces paravalvular regurgitation.”
`
`[emphasis added]
`
`66.
`
`Posts are also disclosed in the Figures of WO’980 showing the claimed apparatus. For
`
`example in Figure 1073 which is pictured below, one post of the anchor is marked by
`an arrow.
`
`
`
`FIG. 107B
`
`67. With respect to the embodiments shown in Figures 107A —107C the specification of
`
`WO’980 (page 85, 11. 32-34) further discloses:
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 17 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 17 of 56
`
`

`

`Bird & Bird
`
`17 March 2016
`page 18
`
`“Once the anchor is fully compressed within the aortic valve, the anchor is
`
`locked the fingers and post mandrels are disengaged, and the seal is adapted
`tofurther limit bloodflow around the replacement valve.”
`
`[emphasis added]
`
`68.
`
`Since the posts comprise the locking mechanism which is essential to maintain an
`
`imposed foreshortening and the radial strength of the anchor (see also WO’980,
`
`page 23, 1. 1-2), the posts themselves must be considered as essential elements. This is
`
`further supported by Figures 50—53 of WO’980, which show anchor posts having a
`
`lock alignment feature (see also WO’980, page 18, l. 26 through page 19, l. 3 and point
`
`3 above).
`
`Besides, the posts cannot be located anywhere, but must be attached on the inside of
`
`the anchor as for instance shown in Figure 1073 pictured above. The specification of
`
`WO’980 (page 22,11. 17-19) discloses:
`
`“Anchor 30 has a lip region 32, a skirt region 34 and a body region 36. First,
`second and third posts 38a, 38b and 380, respectively, are coupled to skirt
`region 34 and extend within lumen 31 of anchor 30.”
`
`7o.
`
`In addition, the valve leaflets are coupled to the posts, which is only possible if the
`
`posts are attached to the inside the anchor (cf. WO’980, page 22, 11. 32-33, page 48,
`
`11. 19-24).
`
`71.
`
`Thus, anchor posts attached on the inside of the anchor are an essential feature of the
`
`apparatus as claimed.
`
`(7)
`.
`Leaflet engagement element of anchor
`
`72.
`
`The essential feature of the leaflet engagement elements addresses the problem set out
`
`in the introduction of WO’980 to effectively anchor the stent to prevent migration and
`
`to decrease the risk of blocking the coronary ostia (cf. WO’980, page 3, 11. 3-6, 11. 28-
`
`30, page 3,11. 26-30).
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 18 of 56
`
`Edwards Lifesciences Corporation, et al. Exhibit 1148, Page 18 of 56
`
`

`

`Bird 8: Bird
`
`17 March 2016
`page 19
`
`73-
`
`The anchor must also comprise a leaflet engagement element to prevent the distal
`
`movement of the anchor. In this context, WO’98

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket