throbber
Edwards Lifesciences Corporation, et al. Exhibit 1137, Page 1 of 12
`
`

`

`VOLUME 2
`
`Second Edition
`
`TEXTBOOK
`
`fO
`
`INTERVENTIONAL
`
`CARDIOLOGY
`
`ERIC 1. TOPOL, MD.
`
`Chairman, Department of Cardiology
`Director, Center for Thrombosis and Vascular Biology
`Cleveland Clinic Foundation
`Professor of Medicine
`Cleveland Clinic: Health Sciences Center
`Ohio State University
`Cleveland, Ohio
`
`'
`
`.
`;
`
`W.B. SAUNDERS COMPANY
`A Division of Harcourt Brace 8 Company
`
`Philadelphia, London, Toronto, Montreal, Sydney, Tokyo
`
` .{COW—C”‘V
`lHHHHl
`HHH ‘
`
`
`
`
`Edvvards Lifesciences Corporation, et al. Exhibit 1137, Pagglgwfiéasomwgm
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1137, Page 2 of 12
`
`

`

`
`
`W. B. SAUNDERS COMPANY
`A Division of Harcourt Brace 67 Company
`The Curtis Center
`Independence Square West
`Philadelphia, PA 19106
`
`
`
`Library of Congress Cataloging-in-Publlcatlon Data
`
`93-15130
`
`Textbook of interventional cardiology/ [edited by] Eric J. Topol —
`2nd ed.
`p.
`cm.
`Includes bibliographical references and index.
`ISBN 0-7216-6749-X (V. 1).—ISBN 0-7216-6750-3 (V. 2).—
`lSBN 0-72166722-8 (Set)
`1. Angioplasty.
`2. Cardiovascular system—Diseases—Treatment.
`l. Topol, Eric J.
`[DNLM:
`1. Cardiovascular Diseases—surgery.
`168 T355 1994]
`RD598.5.T49
`1994
`617.4’12059—dc20
`DNLM/DLC
`
`2. Angioplasty. WG
`
`‘I
`
`1
`l
`I
`
`l Vl
`
`Textbook of INTERVENTIONAL CARDIOLOGY
`
`Volume 1 0-7216-6749-X
`Volume 2 0-7216—6750-3
`
`Set ISBN 0-7216-6722-8
`
`Copyright © 1994, 1990 by w. B. Saunders Company.
`
`All rights reserved. No part of this publication may be reproduced or transmitted in any form or by
`any means, electronic or mechanical, including photocopy, recording, or any information storage and
`retrieval system, without permission in writing from the publisher.
`Printed in the United States of America.
`
`Last digit is the print number:
`
`9
`
`8
`
`7
`
`6
`
`5
`
`4
`
`3
`
`2
`
`1
`
`' Edwards Lifesciences Corporation, et al. Exhibit 1137, 'Paggalpfigasomsgso
`
`Edwards Lifesciences Corporation, et al. Exhibit 1137, Page 3 of 12
`
`

`

`
`
`75
`Percutaneous Expandable
`Prosthetic Valves
`'
`
`' Steven R. Bailey
`
`V
`
`The introduction of percutaneous techniques for vascular
`intervention has resulted in the rapid development of new
`devices and techniques for the treatment of coronary and vas—
`cular diseases. One important area of potential interest is the
`application of percutaneous techniques for the treatment of
`valvular heart disease. The exponential growth of percuta—
`neous endovascular procedures is similar to that experienced
`in cardiovascular surgery after the introduction of cardiopul;
`monary bypass. The design of implantable prosthetic heart
`valves to be delivered using a percutaneous approach has
`become an important area for investigation.
`The initial implantation of a prosthetic heartvalve was per-
`formed in 1952 and published in 1953 by Hufnagel and Har-
`vey,1 with a follow-up report in 1954.2 The Hufnagel valve is
`composed of a chamber with a central ball valve as seen in
`~ Figure 75—]. This bulky device was surgically placed in the
`descending aorta distal to the left subclavian artery. Patients
`did surprisingly well clinically despite the fact that only 75
`per cent of the regurgitant volume was diminished using this
`valve in the descending aorta. Subcoronary placement of
`prosthetic valves became possible only after the introduction
`of cardiopulmonary bypass in 1960.3 The past three decades
`have seen significant improvements in the performance of
`mechanical, tissue, and homograft prosthetic valves, with nu-
`merous alterations occurring in the designs of these prosthe-
`ses.4 These changes have improved the functional orifice area
`as well as decreased the complications associated with valve
`replacement such as thrombus formation, embolism, and late
`valve dysfunction.
`Unfortunately, placement of prosthetic heart valves re-
`mains a relatively difficult and often dangerous procedure.
`The surgical risks rise rapidly in patients with serious prob-
`lems such as acute valvular regurgitation and in patients
`whose valvular disease is associated with myocardial
`ischernia?’6
`
`The development of percutaneous catheter—based systems
`for stabilization and treatment of unstable patients with val-
`vular disease, which could be performed at a lower risk to
`the patient, is therefore an important area for research. De-
`veloping a chronically implanted catheter—based valve pros-
`thesis is an exciting new frontier in interventional cardiology.
`1268
`
`HISTORY OF CATHETER-BASED VALVES
`
`The demonstrated success of the Hufnagel valve in treating
`aortic insufficiency precipitated the initial investigation in
`1965 by Hywel Davies7 of a catheter-mounted valve for tem-
`porary relief of aortic insufficiency. This device, although
`crude by today’s standards, was very interesting in design.
`As seen in Figure 75—2 this cone-shaped device was essen—
`tially an inverted parachute. The valve closed during systole
`due to the forward flow of blood out of the ventricle, and
`opened during diastole as the regurgitant flow returned to
`the ventricle. It was anchored onto a 5-H. catheter with thin
`guy wires. No information was provided regarding the type
`of material from which this valve was constructed. Initial an-
`imal experiments were promising, although no human inves-
`tigations were ever reported. One significant problem dis—
`played by this valve was the rapid development of thrombi
`at the base of the cone, a theme common to all prosthetic
`valves. This predisposition was probably enhanced by pro-
`longed stasis of blood in the base of the conical valve.
`Moulopoulos et al.3 reported their investigations into
`catheter-mounted aortic valves in 1971. Utilizing the investi-
`gators’ experience in developing electrocardiographically
`(ECG) triggered intra-aortic balloon pumps, they designed
`and evaluated three separate systems (Fig. 75—3). One was a
`spherical balloon triggered by the ECG to inflate during di-
`astole. The second was a spherical balloon that was pressure
`responsive and deflated when systolic pressure exceeded a
`predetermined value. It inflated when pressure fell below a
`specified diastolic value, resulting in diastasis. Their third
`system was an umbrella-shaped balloon similar to that used
`by Davies et a1. They concluded that the relatively simple
`umbrella system was best, significantly reducing the severity
`of
`acute
`aortic
`insufficiency without major
`acute
`complications.
`1
`There were several disadvantages of a catheter-based valve
`found by these investigators in the chronic animal model, in-
`cluding the development of significant thrombi in the base of
`the umbrella in all animals followed chronically. Most iIn-
`portantly, variable decreases in coronary flow also occurred
`in these animals. The decreased coronary flow was attributed
`
`
`
`wry”?Wfim”?QT”?
`
`' Edwards Lifesciences Corporation, et al. Exhibit 1137, Paggfilpfiéasomégém
`
`Edwards Lifesciences Corporation, et al. Exhibit 1137, Page 4 of 12
`
`

`

`FIGURE 15-l. The Hufnagel valve, when
`assembled, consisted of a hollow plastic
`tube, the fixation rings on each end, and
`
`the polypropylene central ball.
`
`HGIIRE 15-2. The Davis valve was an in-
`vetted plastic cone anchored to the shaft
`of a 5-Fr. catheter. Note the guy wires used
`to stabilize the valve.
`
`75—PERCUTANEOUS EXPANDABLE PROSTHETIC VALVES
`
`1269
`
`
`
`
`FIGURE 75-3. The top section demonstrates the valve motion during systole and diastole. The bottom left
`is a photograph of the umbrella valve and the middle is the balloon valve. The bottom right is a picture of
`the glass cast used to manufacture the umbrella valve.
`
` i|l
`
`i.
`
`
` H»;
`;.
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1137, Paggmlgllasomswsz
`
`Edwards Lifesciences Corporation, et al. Exhibit 1137, Page 5 of 12
`
`

`

`Edwards Lifesciences Corporation, et al. Exhibit 1137, Page 6 of 12
`
`

`

`75—PERCUTANEOUS EXPANDABLE PROSTHETIC VALVES
`
`1171
`
`TABLE 15-11. HIGH-RISK PATIENTS WHO MIGHT BENEFIT
`FROM PERCUTANEOIIS VALVE PLACEMENT
`
`
`Mitral regurgitation due to
`Acute myocardial infarction
`Bacterial endocarditis
`Flail mitral leaflet
`Decompensated chronic mitral regurgitation
`Aortic regurgitation due to
`Bacterial endocardiu‘s
`Decompensated chronic aortic regurgitation
`Tricuspid regurgitation due to
`Bacterial endocarditis
`Pulmonary hypertension
`
`Right ventricular dysfunction
`
`TABLE 75-3. DESIGN PROBLEMS OF PERCIITANEOUS
`VALVES
`___—__.__—__\
`Delivery sheath
`Size of sheath
`Material required to avoid ferchon
`Stiffness of sheath
`Delivery system
`Self-expanding
`Balloon expandable
`Mechanical expansion
`Type of valve
`Bioprosthetic
`Mechanical
`Polymer
`
`to the . valve leaflets covering the coronary ostia during
`diastole.
`
`Phillips at 1511.9 reported a modification of the balloon-
`mounted valve concept in 1976. Figure 75—4 displays the seg-
`mented polyurethane cusp mounted 1 cm from the distal end
`of the catheter that was used in their study. The distal end of
`the catheter contained side holes for monitoring pressure dur-
`ing the procedure. In their dog model they demonstrated a
`significant reduction in aortic diastolic pressure, pulse pres-
`sure, and left ventricular end-diastolic pressure. Coronary ar-
`tery blood flow did not fall and in fact increased slightly, with
`their balloon system.
`investigation of a balloon catheter—
`The most
`recent
`mounted valve was that of Matsubara et al. in 1992.” They
`used the valve shown in Figure 75—5. This somewhat com-
`plicated valve system consists of an 8—Fr. catheter with a latex
`balloon mounted on the catheter. Using hemodynamic eval:
`uation without angiography they demonstrated significant
`decreases in pulse pressure and left ventricular end-diastolic
`pressure as well as increases in aortic diastolic pressure. After
`placement of the device, however, the experimental animals
`still had a significant‘tachycardia and a larger pulse pressure
`than at baseline, suggesting continued aortic regurgitation.
`Unfortunately, the severity of the persistent aortic regurgita-
`tion was not quantitated by the authors.
`Although catheter-based balloon valve technology may of—
`fer some acute relief‘from valvular aortic regurgitation, it is
`not a technology that could feasibly be used over an extended
`period of time as an indwelling device.
`
`BALLOON-EXPANDABLE VALVES
`
`The success of balloon-expandable endovascular stents has
`inspired some investigators to try to develop implantable,
`
`TABLE 75-L LIMITATIONS OF CURRENT VALVE
`PROTI'IESES
`Mechanical
`Reduction in valve orifice
`Hemolysis
`Pannus formation
`Thrombotic
`Systemic emboli
`Stroke
`Complications of anticoagulafion
`Hemorrhagic stroke
`Gastrointestinal bleeding
`Infective endocarditis
`
`a
`balloon-expandable valvular prostheses. Conceptually,
`“stent” should provide a rigid frame, allowing the valve
`struts sufficient strength to remain stable during systole while
`anchoring the valve to the surrounding tissue. Following this
`premise, one should be able to develop a delivery system and
`valve'that can be delivered percutaneously from a large ar~
`tery or even trans-septally from the femoral vein. This
`method of delivery might answer the immediate needs of
`acutely ill patients or those patients at very high surgical risk
`from valvular regurgitation for valve placement (Table 75—1).
`
`DESIGN PROBLEMS OF
`PERCUTANEOIIS VMVES
`
`Current mechanical and prosthetic valves suffer from a
`number of problems (Table 75—2) including the predisposi-
`tion to thrombus formation and embolization, perivalvular
`leak, infection, difficulty sizing valve to annulus, valve de-
`generation, and pannus formation. The designer of any per-
`cutaneously placed valve will need to consider these issues
`during its design and development in order to minimize these
`problems. A high probability exists that percutaneous valves
`will be used, at least initially, as a temporizing measure in
`patients who are not acceptable candidates for surgical valve
`replacement. The long-term problems of valve degeneration
`and calcification will therefore not be as important in the sub-
`set of patients undergoing percutaneous valve insertion as a
`stabilizing procedure.
`The design of a new percutaneous valve is frought with a
`large number of problems as detailed in Table 75—3. The
`problems can be subdivided into three major categories: (1)
`the problem of a delivery sheath that will allow safe and ef-
`fective delivery of a bulky balloon and valve; (2) a balloon
`system that is capable of precisely implanting the valve must
`be manufactured; and (3) modifications of current valves or
`development of new valvular prostheses must occur.
`
`TABLE 75-4. BALLOON~EXPANDABLE VALVE
`
`
`Advantages
`Percutaneous arteriotomy
`Large expansion ratio of stent
`Approach to multiple valves with same system
`Ability to easily increase size of valve
`Disadvantages
`Closure of arteriotomy
`Technically demanding
`Possible dislodgment
`
`Myocardial ischemia
`
`
`
`t al. Exhibit 1137, Pagggvgfigasomw 584
`
`
`
`
`Edwards Lifesciences Corporation, e
`
`a
`
`Edwards Lifesciences Corporation, et al. Exhibit 1137, Page 7 of 12
`
`

`

`
`NH IHH
`
`1272
`
`VIII—VALVULOPLASTY, CONGENITAL AND PERICARDIAL HEART DISEASE
`
`
`
`
`
`HGURE 75-6. A, The stent ring is constructed of stainless-steel
`wires folded in loops. B, A porcine valve was mounted inside the
`rings and attached by means of suture. C, A view from above valve
`illustrating the intact valve with three cusps.
`
`is
`
`L Ilrll
`
`Percutaneous Delivery System
`
`Large lumen sheaths and catheters (up to 22-Fr.) are cur-
`rently used for vascular access for such procedures as cuta-
`neous perfusion devices. Such ‘a large arterial puncture is a
`major concern with any delivery system, due to the potential
`for injury to the artery, bleeding, or pseudoaneurysm for-
`mation. New technologies are emerging for the local repair
`and sealing of large arteriotomies that should aid the percu-
`taneous placement of valves. In the past, up to 25 per cent of
`arteriotomies performed for balloon valvuloplasty or cuta-
`neous perfusion devices required direct surgical closure. This
`repair if necessary, however, could be performed under local
`anesthesia and is a relatively minor surgical procedure com~
`pared to valve replacement.
`Ideally, a successful percutaneous delivery system must be
`no larger than 20 Fr. (6.6 mm) in order to be used in a large
`number of patients. In addition, a removable trochar with a
`soft distal tip Will be necessary to successfully deliver this
`guide sheath to the valve with the smallest amount of arterial
`trauma. Once the delivery sheath is appropriately positioned,
`the trochar can be removed.
`‘
`A protective sleeve covering the balloon and valve will
`
`be required in order to minimize the friction between the
`balloon-valve and the guide catheter. This sleeve might also
`serve the purposes of constraining the valve and preventing
`slippage or embolization of the valve. Alternatively, the bal-
`loon catheter itself might have a sleeve covering just the stent
`to provide similar protection. After expansion of the valve
`the balloon could be retracted into the sheath to avoid arterial
`injury associated with removal of the balloon.
`
`TABLE 75—5. BALLOON DEPLOYABLE VALVE:
`THE OPTIMAL SYSTEM
`
`
`Requirements for valve
`Minimal vascular trauma
`Must not be dislodged
`During delivery
`During balloon withdrawal
`Valve must be matched to annulus
`Valve must remain competent
`Local tissue injury avoided
`Avoid myocardial i'schemia
`
`No significant thrombosis
`
`Edwards Lifesciences Corporation, et al. Exhibit 1137, PaggMfiéfisoo383és5
`
`Edwards Lifesciences Corporation, et al. Exhibit 1137, Page 8 of 12
`
`

`

`75—PERCUTANEOUS EXPANDABLE PROSTHETIC VALVES
`
`1273
`
`FIGURE 75—7. A, The compressed valve and stent are shown from
`the top view. B, The stent and valve have been mounted on the
`delivery balloon. Note the plastic wedges on either side of the stent
`
`to anchor it in place.
`
`
`TABLE 75-6. VALVE CONSTRUCTION DESIGN
`
`REQUIREMENTS
`Orifice
`Primary
`Central blood flow
`Secondary
`Area between Primary orifice and OCCIUder
`Tefiae-zybeMeen ocduder and aortic wall
`Seating Patterns
`Overlappmg
`Full orifice
`valve types
`Mechanical
`Homograft
`Xenograft
`Materials “58d
`Megzwed
`gitaniflsme
`Alloys
`Stemte 21
`Elgiloy
`Pyrollite
`POIYmers
`Tefl‘f“
`
`‘
`
`1
`
`'
`
`.
`
`’
`
`'
`
`Polypropylene
`’
`
`
`DELIVERY BALLOON CHARACTERISTICS
`
`-
`
`An irnportant issue central to the success of percutaneous
`_
`_
`‘
`valve placement Wlll be the deVice or system used to implant
`the valve. The valve could be self-expanding, expandable
`by some mechanical device as yet undefined, or balloon ex-
`pandable. In order to be practical, some form of a balloon-
`expandable system
`probably be required in order for
`such a system to be Widely employed._A balloon-expandable
`system offers the advantages and disadvantages listed in
`Table 75—4. The advantages include the ability to use a rela-
`tively small arterial puncture or arteriotomy in the placement
`of an introducer sheath. Additionally, the use of a balloon
`system will allow expansion of the device from a much
`smaller compressed state to the larger fully deployed state.
`This could result in an expansion ratio between three and ten
`times the compressed diameter. Such an expansion ratio
`might even allow percutaneous placement of large mitral and
`tricuspid valves. Such expansion ratios need not require an
`expandable metal mesh, but might occur in the form of a
`single ring designed much like a hose clamp that opens; even
`self—expanding alloys might be utilized if correctly designed.
`Placement techniques must be devised to assure site—specific
`localization and stabilization of such a device as well as an
`orientation that is coaxial to blood flow. In most cases these
`valves would be placed for severe valvular regurgitation, aS'
`suring a large-volume, high-velocity jet at the site of intended
`
`;
`;
`
`3
`
`5
`
`1_
`_
`:
`5
`;;
`
`E
`:1
`i
`
`z
`
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1137, Paggmlgllasomsfise
`
`Edwards Lifesciences Corporation, et al. Exhibit 1137, Page 9 of 12
`
`

`

`
`
`1274
`
`VIII—VALVULOPLASTY, CONGENITAL AND PERICARDIAL HEART DISEASE
`
`w
`
`
`
`
`FIGURE 15-8. A, The expanded balloon with the stent still in
`place. The arrows indicate the elastic blocks used to keep the valve
`in place. B, The fully expanded valve is seen from the top.
`
`\/
`
`
`
`bv
`
`placement. Such a large blood volume may well require a
`secondary balloon or a novel stabilization method in order to
`assure exact valve placement. In addition, this delivery sys-
`tem will need to be capable of angiographic injections to de-
`termine exact location of the valve and to assure that coro-
`nary ostia are not compromised. Several options are possible
`for such a delivery catheter, including a short segment of
`catheter distal to the valve prosthesis for sampling pressures
`and performing the angiographic injections.
`In order to successfully deploy a percutaneous valve, the
`delivery system will need to meet several criteria (Table 75—
`5). The first will be that the valve is not displaced from the
`balloon during delivery. Second, the balloon must expand to
`a large enough diameter to securely anchor the valve in place.
`Third, retraction of the balloon must not dislodge the valve.
`And fourth, the delivery balloon must not damage surround-
`ing tissue.
`The delivery balloon could be constructed of multiple bal-
`loons placed on a single catheter, although hourglass con-
`struction of the variety currently employed by the lnoue bal-
`loon may be preferable. Multilobed balloons do allow a
`smaller diameter of the deflated balloon but are unstable in
`
`high-flow fields such as one would find in mitral or aortic
`regurgitation.
`The hourglass balloon might have two inflation lumens—
`one for the distal balloon and a second for the proximal bal-
`loon and valve. This technique will allow the valve to be im-
`planted only after the distal balloon has first been anchored
`in place at the site of the valve annulus. This type of balloon
`would also resulbin less damage to surrounding tissue, since
`the largest diameter of such a balloon would be at its center,
`where the valve was situated. Additionally, this single bal-
`loon would be less likely to become entangled in the valve
`and dislodge it. Finally, the hourglass shape is more adapt-
`able to large and small diameters, allowing a single balloon
`design to function as the primary delivery device for all per-
`cutaneous valve implants.
`
`VALVE CONSTRUCTION
`
`The specific material(s) from which a percutaneously
`placed valve is constructed is an area requiring further eval-
`
`Edwards Lifesciences Corporation, et al. Exhibit 1137, Pagemfigasomsfiw
`
`Edwards Lifesciences Corporation, et al. Exhibit 1137, Page 10 of 12
`
`

`

`Edwards Lifesciences Corporation, et al. Exhibit 1137, Page 11 of 12
`
`

`

`
`
`1276
`
`VIII—VALVULOPLASTY, CONGENITAL AND PERICARDIAL HEART DISEASE
`
`that can be “floated” into the ventricle, since it is less likely
`to be caught inbetween ,chordal structures than a system that
`is placed over a wire.
`‘
`
`CURRENT INVESTIGATIONS
`
`The most exciting published work in this area to date is
`the investigationsby Andersen et at. from Denmark pub-
`lished in 1992.13 They reported on their work on developing
`a transluminal' implantable heart valve. They began their in-
`vestigations in 1987 and have developed an exciting catheter—
`based technique that has been initially evaluated in a porcine
`model. Using 0.55-mm surgical grade stainless-steel wire,
`they formed a steel mesh 14 mm in length upon which por-
`cine aortic valves were mounted with suture (Fig. 7546). This
`allowed the stent to be com ressed to 12 mm when folded
`and expanded to 32 mm wia the delivery balloon (Fig. 75—
`7). A 41-Fr. introducer sheath was used to place the valve
`from a surgically created arteriotomy site (Fig. 75—8). The di-
`rect introduction was required because of the small femoral
`arteries of the pigs they used for this evaluation. Placement
`was guided both by fluoroscopy and by echocardiography,
`although transthoracic echocardiography was abandoned be—
`cause the echogenic nature of this valve did not allow accu-
`rate imaging of the valve. Figure 75—9 demonstrates angio-
`graphic images from animals that had subcoronary and
`supracoronary implantation of the valve.
`Although all of the protocol animals survived the initial
`implantation, two of the seven animals studied died of pro-
`gressive left ventricular dysfunction secondary to myocardial
`ischemia. At necropsy no animals demonstrated dissection or
`hematoma. Three of four animals studied did demonstrate
`restriction of coronary flow. Finally, two of the seven valves
`still had mild degrees of aortic regurgitation.
`These investigations also were very creative in the creation
`of a delivery system. They used three balloons, each 15 mm
`in diameter and 70 mm in length, with small blocks placed
`on the balloon to anchor the stent in place.
`Although the above study presents only early data illus-
`trating the large amount of work re uired before a device
`such as this can be used, it illustrates 3121f such devices await
`only the focused evaluation and improvement of a few in-
`vestigators. Without question, within the next few years de-
`vices such as this will be used in selected high-risk patients
`for the treatment of valvular disease.
`
`7 CONCLUSIONS
`
`The development of percutaneous techniques for treatment
`of cardiovascular diseases will continue to provide us with
`
`new methods of stabilizing and treating patients. It is ex-
`tremely likely that one of these technologies will be percu-
`taneously implanted heart valves. Most certainly the treat-
`ment of acute aortic insufficiency will be possible; treatment
`of other regurgitant lesions as well is likely. In 10 years we
`shall very probably look back on the pioneering work de-
`scribed above in the same way we respect the work of Huf~
`nagel, Gruentzig, and Palrnaz today. Certainly, heart valves
`will undergo radical changes in design in the decade.
`
`References
`
`1. Hufnagel CA, Harvey WP: The surgical correction of aortic re-
`gurgitation. Preliminary report. Bull Georgetown Univ Med
`Center 6:3—6, 1953.
`2. Hufnagel CA, Harvey WP, Rabil P], McDermott TF: Surgical
`correction of aortic insufficiency. Surgery 35:673—683, 1954.
`3. HarkenDE, Soroff HS, Taylor W], Lefemine AA, Gupta SK,
`Lunzer» S: Partial and complete prosthesis in aorta insuffi-
`ciency. I Thorac Cardiovasc Surg 40:744—762, 1960.
`4. Grunkemeier GL, Starr A, Rahimtoola SH: Prosthetic heart valve
`performance: Long term follow-up. Curr Prob Cardiol 6:333-
`406, 1992.
`’
`5. Fessatidis I, Hackett D, Oakley CM, et a1: Ten-year clinical eval—
`uation of isolated mitral valve and double valve replacement.
`Ann Thorac Surg 43:368~372, 1987.
`6. Perier P, Deloche A, Chanvaud S, et al: Clinical comparison of
`mitral valve replacement using porcine, Starr and Bjork pros-
`theses. ] Cardiovasc Surg 3:359—362, 1988.
`7. Davies H: Catheter mounted valve for temporary relief of aortic
`insufficiency. Lancet 1:250, 1965.
`8. Moulopolous SD, Anthopolous L, Stamatelopolous S, Stefado-
`rous M: Catheter-mounted aortic valves 11:423—430, 1971.
`9. Phillips 5], Ciborski M, Freed PS, Cascade PN, Jaron D: A tem-
`porary catheter—tip aortic valve: Hemodynarnic effects on ex-
`perimental acute aortic insufficiency. Ann Thorac Surg 21:
`134—137, 1976.
`10. Matsubara T, Yamazoe M, Tamura Y, Ohshima M, Yamazaki Y,
`Suzuki M, lzumi T, Shibata A: Niigata, Japan. Balloon catheter
`with check valves for experimental relief of acute aortic re-
`gurgitation. Am Heart] 124:1002—1008, 1992.
`11. Lo HB, Herold M, Reul H, et al: A tiicuspid polyurethane heart
`valve as an alternative to mechanical prostheses or biopros—
`theses. Trans Am Soc Artif Intern Organs 34:839—844, 1988.
`12. Boretos IW, Poirer RA: Aortic heart valve catheter. United States
`Patent (No. 4.056.854), 1977.
`13. Andersen HR, Knudsen LL, Hasenkam JM: Transluminal im—
`plantation of artificial heart valves. Description of a new ex-
`pandable aortic valve and initial results with implantation by
`catheter technique in closed chest pigs. Eur Heart I 1327047
`708, 1992.
`
`Edwards Lifesciences Corporation, et al. Exhibit 1137, Pagemfiéasomswsg
`
`Edwards Lifesciences Corporation, et al. Exhibit 1137, Page 12 of 12
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket