throbber
Peterreins Schley
`P otent- Lind Reentsanwalte GbR
`
`H erm ann-Sack-Straße 3
`80331 München
`Deutschland
`
`Tel: +49 (89) 8091338-00
`Fox: +49 (89) 8091338-88
`Moil: info@ ps-patent.com
`w w w .ps-paten t.com
`
`Dr. Frank Peterreins ‘
`Dr. Jan-M olte Schley^’ -
`Markus Coehn + ^
`Felix Glöckler +
`Dr. Matthias Trout
`Patrick W iedemann + ^
`
`' R G C h t'iQ n v y a lT
`' P o 'e n to r.w o it
`Europecin P o te rt A tiorney
`
`EPO - Munich
`82
`2 Aug. 2016
`
`P E T E R R E I N S S C H L E Y
`P A T E N T - U N D R E C H T S A N W Ä L T E
`
`P eterreinsSchley • Potsnl- ‘ irifl Rftchtsorw.ólte GbP ■ l-lerrucirin-Sock-StrnBe 3 • 8 0 3 3 L Múfr.hpn
`
`European Patent Office
`
`80298 Munich
`
`Munich, August 24, 2016
`Our ref: 31316-00440PI
`
`Opposition against EP 2 749 254 B l (14 161 991.6)
`Proprietor;
`Boston Scientific Scimed, Inc.
`Opponent 1 :
`Lang & Tomerius
`Opponent 2:
`Edwards Lifesciences Corp.
`Opponent 3:
`Medtronic CV Luxembourg S.à.r.l.
`
`In the name and on behalf of the Proprietor, it is requested
`
`1.
`2 .
`
`to reject the oppositions and to maintain the patent as granted;
`to summon for oral proceedings.
`
`A.
`
`Original Disclosure
`
`A.I
`
`Original Disclosure of Granted Claim 1
`
`The subject-matter of granted claim 1 is originally disclosed.
`
`1.
`The features of the preamble of granted claim 1 are - inter alia - disclosed in the PCT
`application WO 2005/0652980 (BB2) on page 1, lines 3/4, on page 17, lines 23 to 33, on page
`21, lines 25/26, and on page 22, lines 4 to 8 . Original claim 1 of the PCT application W O’980
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 1 of 42
`
`

`

`also provides a suitable disclosure for the preamble of granted claim 1, wherein it is clarified in
`the PCT application that using a braid is only an optional feature, see PCT application W O ’980,
`page 5, lines 4/5:
`
`‘‘The anchor includes an expandable anchor such as a braid. ”
`
`and page 22, lines 21-25:
`
`,.Anchor 30 preferably is fabricated by using self-expanding patterns (laser cut or
`chemically milled), braids, and materials, such as stainless steel, nickel-titanium
`( “Nitinol”) or cobalt chromium, but alternatively may be fabricated using balloon-
`expandable patterns where the anchor is desisned to plastically deform to its fin a l shape
`by means o f balloon expansion. ”).
`
`And finally, also original claims 89 and 340 provide a basis for the preamble of granted claim
`
`1.
`
`The characterizing feature of granted claim 1 is disclosed in context with Figures 32-
`2.
`34, see PCT application W O’980, page 34, lines 26-31:
`
`“Figures 32-34 show another way to seal the replacement valve against leakage. A
`fabric seal 380 extends from the distal end o f valve 20 and back proximally over anchor
`30 during delivery. When deployed, as shown in Figures 33 and 34, fabric seal 380
`bunches up to create fabric flaps and pockets that extend into spaces form ed by the
`native valve leaflets 30 382, particularly when the pockets are filled with blood in
`response to backßow blood pressure. This arrangement creates a seal around the
`replacement valve. ”
`
`and PCT application W O’980, page 8 6, lines 22-32:
`
`“Figures 32-34 illustrate the process o f forming a pleated seal around a replacement
`valve to prevent leakage. Figure 32 illustrates a fabric seal 380prior to deployment and
`foreshortening o f the anchor/valve apparatus. In Figure 32, the fabric seal 380 extends
`from the distal end o f valve 20 proximally over anchor 30 during delivery. During
`deployment, as illustrated in Figure 33, anchor 30 foreshortens and the fabric seal 380
`bunches up to create fabric flaps and pockets that extend into spaces form ed by the
`native valve leaflets 382. The bunched up fabric or pleats occur, in particular, when the
`pockets are filled with blood in response to backßow blood pressure. The pleating can
`create a seal around the replacement 30 valve. Figure 34 illustrates anchor 30,
`surrounded by fabric seal 380 in between native valve leaflets 382. In preferred
`embodiments, at least a portion o f a seal is captured between the leaflets and the wall
`o f the heart when the anchor is folly deployed. ’’
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 2 of 42
`
`

`

`■3 -
`
`The Opponents argue that several features have allegedly been described as being
`3.
`essential features, and that these features are missing in granted claim 1. However, none of
`these features are essential for achieving the sealing effect as explained in the application
`documents. Accordingly, none of these features is contained e.g. in claims 1, 89 and 340 of the
`PCT application W O’980. Indeed, many o f the objections raised are not truly objections under
`Article 123(2) EPC, but are in fact inadmissible objections relating to Article 84 EPC with
`respect to essential features allegedly missing.
`
`A.II Active Foreshortening is no Essential Feature
`
`The aspect of an “active foreshortening” is no essential feature as alleged by the Opponents.
`
`First o f all, an active foreshortening is not even mentioned in any o f independent claims
`1.
`1, 67, 101, 122, 142, 151, 182, 184, 255, 268, 340, 350, 367, 368, 387, 403 and 428 o f the PCT
`application W O’980. In view of these independent claims as originally disclosed, it is not
`understandable why the aspect of an active foreshortening should allegedly be essential.
`
`The aspect of an active foreshortening relates to a specific deliverv catheter - disclosed
`2.
`e.g. in context with Figures 3A to 3F - which axially compresses the anchor (without applying
`any radial forces) to achieve a radial expansion. However, the original application documents
`also clarify that other types of delivery catheters may be used to deploy heart valve implants o f
`the present invention, in particular bv balloon inflation:
`
`•
`
`PCT application W O’980, page 8, lines 15/16:
`
`“Expansion o f the anchor and replacement valve may be bv balloon-expansion, self-
`expansion, and combinations thereof ”
`
`•
`
`PCT application W O’980, page 13, lines 25-27:
`
`“The apparatus may also include a deployment tool coupled to the anchor within the
`catheter and an expandable balloon disposed within the deliverv catheter, the balloon
`beins adapted to expand the anchor. ”
`
`•
`
`PCT application W O’980, page 22, lines 21-25:
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 3 of 42
`
`

`

`• 4 -
`
`,,Anchor 30 preferably is fabricated by using self-expanding patterns (laser cut or
`chemically milled), braids, and materials, such as stainless steel, nickel-titanium
`( “N itinol”) or cobalt chromium, but alternatively may be fabricated using balloon-
`expandable patterns where the anchor is designed to plastically deform to its final shape
`by means o f balloon expansion. ”
`
`As a result, the Opponents’ view that the heart valve implant of the present invention can only
`be deployed by an “active foreshortening” (applying axial compression forces), is incorrect.
`
`3.
`In addition, the general disclosure on a bunched-up fabric seal in the PCT application
`W O’980, on page 34, lines 26-31, does not mention any foreshortening:
`
`“Figures 32-34 show another way to seal the replacement yalye against leakage. A
`fabric seal 380 extends from the distal end o f yalye 20 and back proximally oyer anchor
`30 during deliyery. When deployed, as shown in Figures 33 and 34, fabric seal 380
`bunches up to create fabric flaps and pockets that extend into spaces form ed by the
`natiye yalye leaflets 30 382, particularly when the pockets are filled with blood in
`response to backflow blood pressure. This arrangement creates a seal around the
`replacement yalye. ”
`
`4.
`The embodiment of Figures 107A-C on which Opponent 2 is relying its objection, is
`only “another embodiment” as clarified in the PCT application W O ’980 on page 84, line 30.
`However, pointing to an additional alternative embodiment is no suitable argument for an
`essential feature, if the general disclosure of the invention does not require an active
`foreshortening.
`
`But even the embodiment of Figures 107A-C clarifies that achieving a bunched-up shape of the
`seal in the deployed configuration of the heart valve implant does not require any
`foreshortening. The PCT application W O ’980 only mentions on page 85, line 28, that
`“foreshortening can cause seal 60 to bunch up ... ”.
`
`5.
`Opponent 2 assumes that all problems mentioned in the PCT application must be solved
`by the apparatus/implant o f the opposed patent. This view is, however, based on a
`misunderstanding of the original disclosure.
`
`Contrary to Opponent 2 ’s allegation, the PCT application W O’980 does not require on page 4,
`lines 1-3, that all drawbacks mentioned in context with the prior art, must be solved according
`to the present invention (see Opponent 2’s opposition brief, paragraph 31). The words “methods
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 4 of 42
`
`

`

`• 5 -
`
`and apparatus” are mentioned in plural in this text passage. Therefore, the correct understanding
`of this passage is that various different solutions are provided in order to solve certain problems
`of prior art implants. The number and variety o f independent claims as originally filed already
`show that the PCT application discloses various different solutions addressing different
`drawbacks. However, forcing the Proprietor to include all advantageous aspects in an
`independent claim would unduly limit the right to get sufficient protection for an invention.
`
`And finally. Opponent 2 refers to page 27, lines 3-5 of the PCT application W O’980.
`6 .
`However, this passage relates to a different embodiment, namely the embodiment shown in
`Figures 5A to 5F, which does not even have a bunched-up fabric seal, see in particular Figure
`5E. Therefore, also this attempt of Opponent 2 to argue that an active foreshortening is an
`essential feature, must fail.
`
`A.III Non-Hydraullc/Non-Pneumatic Anchor Actuators are no Essential Features
`
`The non-hydraulic or non-pneumatic anchor actuators are no essential features.
`
`First of all, the non-hydraulic or non-pneumatic anchor actuators described in context
`1.
`with some embodiments of the PCT application are features of the delivery/deployment system,
`1.e. o f the catheter, but no features of the heart valve implant. Accordingly, the passage in the
`PCT application W O ’980 on page 21, lines 28-33, only refers to the delivery/deployment
`system, but not to the heart valve implant. Therefore, it is not logical that the non-hydraulic or
`non-pneumatic anchor actuators should allegedly be essential features of the heart valve
`implant.
`
`In addition, the original application documents also clarify that other types of delivery
`2.
`catheters may be used to deploy the heart valve implant o f the present invention, in particular
`bv balloon inflation:
`
`•
`
`PCT application W O ’980, page 8, lines 15/16;
`
`"Expansion o f the anchor and replacement valve may be by balloon-expansion. self­
`expansion, and combinations thereof. ”
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 5 of 42
`
`

`

`•
`
`PCT application W O’980, page 13, lines 25-27:
`
`‘‘The apparatus may also include a deployment tool coupled to the anchor within the
`catheter and an expandable balloon disposed within the deliverv catheter, the balloon
`beins adapted to expand the anchor. ”
`
`•
`
`PCT application W O’980, page 22, lines 21-25:
`
`,,Anchor 30 preferably is fabricated by using self-expanding patterns (laser cut or
`chemically milled), braids, and materials, such as stainless steel, nickel-titanium
`( “Nitinol”) or cobalt chromium, but alternatively may be fabricated using balloon-
`expandable patterns where the anchor is desisned to plastically deform to its final shape
`bv means o f balloon expansion. ”
`
`•
`
`PCT application W O’980, claim 96:
`
`“96. The apparatus o f claim 89 wherein the anchor and the replacement valve are
`adaptedfor expansion from the delivery to the deployed configurations via a mechanism
`chosen from
`the group consisting o f balloon-expansion. self-expansion, and
`combinations thereof”
`
`The embodiment of Figures 32 to 34 does not mention or show any non-hydraulic or
`3.
`non-pneumatic anchor actuators. And as mentioned above, the embodiment o f Figures 107A-
`C is only “another embodiment” (see PCT application W O’980 on page 84, line 30) which
`shows - e contrario - that non-hydraulic or non-pneumatic anchor actuators are no essential
`features of the heart valve implant according to the general teaching of the original application
`documents.
`
`A.IV Lockable Anchor and Locks are no Essential Features
`
`The lockable anchor and locks are no essential features either.
`
`As discussed in sections A.II and A.III above, neither an active foreshortening nor any
`1.
`non-hydraulic or non-pneumatic anchor actuators are essential features, because the anchor may
`- according to the original disclosure - also be balloon-expandable (which may result in some
`foreshortening of the anchor).
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 6 of 42
`
`

`

`■ 7 -
`
`2.
`The passage in the PCT application W O’980 on page 11, lines 16-18, cited by Opponent
`2 again refers to “another aspect o f the invention ” (see page 11, line 12), whereas the locking
`mechanism is explained as being an optional feature:
`
`“The apparatus may include a
`imposed
`locking mechanism fo r maintaining
`foreshortening, and it may be configured fo r retrieval prior to actuation o f the locking
`mechanism. ”
`
`The optional nature of the locking mechanism is also clarified in the summary section in the
`PCT application W O ’980 on page 4, line 31, to page 5, line 1:
`
`“In some embodiments the deplovins step mav include the stev o f exvandins a balloon
`within the anchor, and in some embodiments the deploying step mav include the step
`lockins the anchor in an expanded configuration. “
`
`3.
`The embodiment of Figures 32 to 34 does not mention or show any locking mechanism.
`And as already mentioned above, the embodiment of Figures 107A-C is only another
`embodiment (see PCT application W O’980 on page 84, line 30) which shows - e contrario -
`that the locking mechanism is no essential features of the heart valve implant according to the
`general teaching of the original application documents. Accordingly, the locking mechanism is
`not mentioned in independent claims 1, 89, and 340 as originally filed.
`
`4.
`In this respect. Opponent 2 again refers to page 27, lines 3-5 of the PCT application
`W O’980. However, this passage relates to another different embodiment, namely the
`embodiment shown in Figures 5A to 5F, which does not even have a bunched-up fabric seal,
`see in particular Figure 5E.
`
`A.V
`
`Self-Expandable Anchor is no Essential Feature
`
`The argument that a self-expandable anchor is allegedly an essential feature, is not convincing
`either. Balloon-expandable anchors are disclosed in the PCT application W O’980. Using a self-
`expandable anchor is only an optional feature. This is explicitly clarified in the PCT application
`W O’980 on page 22, lines 21-25:
`
`,,Anchor 30 preferablv is fabricated by using self-expandins patterns (laser cut or
`chemically milled), braids, and materials, such as stainless steel, nickel-titanium
`( “N itinol”) or cobalt chromium, but altemativelv may be fabricated using balloon-
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 7 of 42
`
`

`

`expandable vatterns where the anchor is desisned to plastically deform to its final shape
`bv means o f balloon expansion. ”
`
`A.VI Repositionable/Retrievable Anchor is no Essential Feature
`
`The aspect of repositioning and retrieving the heart valve implant is no essential feature. This
`is only an optional advantageous aspect as e.g. clarified in the PCT application on page 8 , lines
`30-32 ( “The apparatus mav be retrieved, repositioned and redeployed, and the apparatus may
`be locked in its expanded configuration. ”), on page 1 1, line 18 ( “... and it mav be configured
`fo r retrieval ... ”), and on page 13, line 18 ( "... optionallv retrieving ... ”).
`
`The aspect of repositioning and retrieving the heart valve implant is a separate independent
`invention which is defined in original claims 182 and 255 of the PCT application W O’980. But
`this aspect is not contained in any of the other independent claims as originally filed.
`
`A.VII Anchor Posts on the Inside are no Essential Features
`
`Using anchor posts on the inside for coupling the replacement valve to the anchor is no essential
`feature of the present invention.
`
`The disclosure passages in the PCT application W O’980 on page 22, lines 4 to 9, on page 34,
`lines 26 to 31, and on page 8 6, lines 22 to 32 do not even mention posts for connecting the
`replacement valve to the anchor. Original independent claims 1, 67, 89, 101, 122, 142, 151,
`169, 182, 184,255, 268, 340, 368, 387 and 403 do not mention posts on the inside o f the anchor
`either. And finally, the summary section of the PCT application W O’980 on pages 4 to 15 do
`not mention any posts. These facts alone prove that providing the anchor with posts on the
`inside is merely an optional feature.
`
`The sentence in the PCT application W O’980 on page 22, lines 31 to 33, cited by Opponent 2,
`clarifies itself that coupling the replacement valve to posts is only a preferred feature. Posts are
`only an optional feature, if a lockable self-expandable anchor is used. However, as discussed
`above, a lockable self-expandable anchor is not essential for carrying out the claimed invention,
`because a balloon-expandable anchor may altematively be used.
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 8 of 42
`
`

`

`9 -
`
`The passage in the PCT application W O’980 on page 27, lines 1 to 6 , also cited by Opponent
`2, relates to an embodiment which does not even have a bunched-up fabric seal. This again
`proves that posts are not essential in context with a bunched-up fabric seal.
`
`Opponent 2 further argues that the embodiment of Figures 107A-C of the PCT application
`W O’980 can also be locked, and that a post is shown in Figure 107B. But Opponent 2 ignores
`that the embodiment of Figures 32 to 34 of the PCT application W O’980 does not show any
`posts. Accordingly, posts are neither mentioned on page 34, lines 26 to 31, nor on page 8 6 , lines
`22 to 32. Therefore, the embodiment of Figures 107A-C of the PCT application W O ’980 has
`only additional advantageous features compared to the embodiment of Figures 32 to 34. But
`those additional features are certainly not essential for the claimed invention.
`
`A.VIII Leaflet Engagement Elements are no Essential Features
`
`Using leaflet engagement elements is no essential feature of the claimed invention as alleged
`by Opponents 2 and 3. Leaflet engagement elements are only another inventive aspect defined
`in original claims 101 and 122. However, this aspect is not mentioned in any other independent
`claim the PCT application W O’980. Accordingly, the disclosure passages in the PCT
`application W O’980 on page 22, lines 4 to 9, on page 34, lines 26 to 31, and on page 8 6 , lines
`22 to 32 do not even mention leaflet engagement elements, and no leaflet engagement elements
`are shown in Figures 32 to 34 either.
`
`A.IX Braided Anchor is no Essential Feature
`
`Using a braided anchor is no essential feature of the claimed invention. The PCT application
`W O’980 clarifies on page 22, line 5, that a braid is only an optional feature ( “... expandable
`anchor such as a braid. ”). The passages on page 22, lines 21 to 25, and on page 39, lines 18 to
`27, explicitly mention that plastically deforming balloon-expandable anchors may be used
`instead of braided anchors. And the disclosure passages in the PCT application W O’980 on
`page 34, lines 26 to 31, and on page 8 6 , lines 22 to 32 do not even mention a braided anchor.
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 9 of 42
`
`

`

`- 1 0
`
`The aspects of active foreshortening and using a self-expandable braided anchor which can be
`locked in its expanded state are only other inventive aspects which are not essential in context
`with a bunched-up fabric seal.
`
`A.X Commissures are no Essential Features
`
`Commissures are no essential features in context with a bunched-up fabric seal. None o f the
`independent claims as originally filed contains the feature of commissures. The summary
`section on pages 4 to 15 of the PCT applieation W O’980 does not mention commissures either.
`The sentence in the PCT application W O’980 on page 22, lines 31 to 33, cited by Opponent 2,
`also clarifies that commissures may only be “preferably” used.
`
`A.XI Horizontal Pleats, Flaps and/or Pockets are no Essential Features
`
`The Opponents argue that various aspects relating to the shape and to the arrangement o f the
`bunched-up fabric seal are essential.
`
`In this respect, it should be noted first that claim 1 already contains features on the
`1.
`arrangement of the fabric seal, namely that it is “extending from the distal end o f the valve (20)
`proximally over the anchor in the delivery configuration, wherein the seal is bunched up in the
`deployed configuration”. This wording, which is in line with the disclosure in the PCT
`application W O ’980 on page 34, lines 26 to 28, already clarifies that the fabric seal is at least
`partially arranged around the outside of the anchor. Therefore, Opponent 2 ’s submission in
`paragraph 86 of its opposition brief regarding the arrangement of the fabric seal is not
`understandable.
`
`The word “pleats” is not mentioned in the PCT application W O’980 on page 34, lines
`2.
`26 to 31, either, and cannot be regarded as being an essential feature. The passage on page 85,
`lines 28/29, refers to a different advantageous embodiment and clarifies that “pleats” can
`optionally be formed.
`
`Besides that. Opponent 2 ’s argument is in contradiction to its position in context with novelty
`of the granted claims, because Opponent 2 argues in paragraphs 123, 152 and 179 of its
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 10 of 42
`
`

`

`11
`
`-
`
`opposition brief that “bunched up” is used synonymously to “flaps” and “pleats” in the patent
`specification;
`
`123. The terms “flaps”, “pleats” and “bunched up” seal are used in the specification
`synonymously (see BBa, page 8 5,11. 28-29, page 86,11. 22-23, page 8 6 ,11. 27-28 and
`page 34, 11. 28-29).
`
`Opponents 1 and 3 believe that the wording “to create fabric flaps and pockets that
`3.
`extent into spaces form ed by the native valve leaflets” is missing in granted claim 1. However,
`this phrase only describes the effect which can be achieved by the bunched-up fabric seal during
`use. Granted claim 1 is a product claim, and the features of a product claim must be recognizable
`independent from its use. The relevant structural features on the bunched-up fabric seal are
`already contained in granted claim 1. Therefore, there is no need to include additional
`clarifications on the inventive effect during use.
`
`thus confuse the structural features on the one hand, and the
`Opponents 1 and 3
`technical/medical effect on the other hand. The effect is that - during use - flaps and pockets
`can extend into spaces formed by the native valve leaflets. This is achieved by the bunched-up
`fabric seal.
`
`Opponent 1 tries to base its argument on the words “to” and “that” (see opposition brief, page
`5, paragraph 5), and even admits that the phrase starting with “to” describes the “specific
`functional purpose” (i.e. the technical/medical effect). However, Opponent 1 did not recognize
`that the entire phrase ( “to create fabric flaps and pockets that extent into spaces form ed by the
`native valve leaflets”) starts with the word “to”, i.e. the entire phrase relates to the
`technical/medical effeet achieved by the bunched-up fabric seal.
`
`Referring to Figure 33, Opponent 2 further believes that the fabric seal must comprise
`4.
`circumferential and horizontal pleats. However, the wording “circumferential and horizontal”
`is neither mentioned in the PCT application W O’980 on page 34, lines 26 to 31, nor on page
`8 6, lines 22 to 32. Figure 33 shows a schematic drawing to illustrate an example o f a bunched-
`up fabric seal. Therefore, that the Proprietor cannot be forced to introduce this wording in
`granted claim 1.
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 11 of 42
`
`

`

`1 2 -
`
`A.XII Foreshortening is no Essential Feature
`
`The Opponents further believe that the fabric seal and the anchor must foreshorten during
`deployment.
`
`However, a foreshortening of the anchor and/or fabric seal is not mentioned in the relevant text
`passage on PCT application W O’980 on page 34, lines 26 to 31, at all; nothing has been
`“omitted” in this respect. Therefore, it is clear that a foreshortening has not been described as
`being an essential feature.
`
`Figure 33 shows a schematic drawing to illustrate one possible example o f a bunched-up fabric
`seal. Therefore, the Proprietor cannot be forced to introduce such wording in granted claim 1.
`
`It is only essential - as mentioned in granted claim 1 - that the fabric seal is bunched-up in the
`deployed state (which can be seen when the heart valve implant is expanded to its intended
`deployed diameter in air, and which results in that the fabric seal can extend into spaces formed
`by the native valve leaflets when the heart valve implant is implanted into a patient).
`
`The PCT application W O’980 even clarifies on page 85, lines 28/29, that a foreshortening is
`only an optional mechanism to cause bunching up:
`
`“Foreshortenins can cause seal 60 to bunch up and create pleats. ”
`
`As a result, a foreshortening of the fabric seal and/or the anchor is not essential.
`
`A.XIII Shape of the Seal in the Delivery Configuration is no Essential Feature
`
`According to granted claim 1, the fabric seal must have a bunched-up shape in the deployed
`configuration. Contrary to Opponent 2’s submission, it is not essential whether and how much
`the fabric seal is bunched-up in the delivery configuration o f the heart valve implant.
`
`The text passage in the PCT application W O’980 on page 34, lines 26 to 31, is silent on the
`shape of the fabric seal in the delivery configuration o f the heart valve implant. Figure 32 is
`only a schematic drawing illustrating one embodiment of the claimed invention. And the text
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 12 of 42
`
`

`

`13
`
`passage in the PCT application W O’980 on page 85, lines 28/29, mentions that foreshortening
`“can cause seal 60 to bunch up This means that a foreshortening o f the anchor - if present -
`may only be optionally provided for causing the fabric seal to bunch up.
`
`In view of the purpose of the fabric seal to reduce or avoid a reverse blood flow (between the
`implant and the annulus), only the bunched-up shape of the fabric seal in the deployed
`configuration is o f importance, whereas the shape o f the fabric seal in the delivery configuration
`o f the heart valve implant is not relevant for the sealing function.
`
`A.XIV Attachment of the Seal at the Proximal End is no Essential Feature
`
`The relevant text passages describing the fabric seal are silent on whether or not the fabric seal
`is attached to the upper end of the anchor. Opponent 2 ’s argument that this feature missing in
`granted claim 1 allegedly results in an inadmissible intermediate generalization, is not
`understandable. Figure 34 is only a schematic drawing showing the anchor 30 and the native
`leaflets 382. It is clear to the skilled person that such a schematic drawing is not intended to
`show any details on the fixation o f the fabric skirt.
`
`A.XV Wording “proximally over the anchor” is originally disclosed
`
`Opponent 2 believes that the word “back” should be included in granted claim 1. However,
`granted claim 1 already contains the feature that the fabric seal is “extendins from the distal
`end o f the valve (20) proximallv over the anchor
`The wording “back proximally over
`anchor 3 0 ” has the same meaning as the wording “proximally over the anchor", namely that
`the part of the fabric seal which is arranged outside o f the anchor should extend in proximal
`direction.
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 13 of 42
`
`

`

`14
`
`30
`
`■20
`
`FIG. 33
`
`380
`
`380
`
`■20
`
`FIG. 32
`
`As can be seen in Figures 32 and 33 of the PCT application, the fabric seal may have an inner
`end which is arranged within the anchor so that the replacement valve, which is fully arranged
`within the anchor, can be connected to the fabric seal. The inner part of the fabric seal may thus
`first extend in distal direction, and then change its direction at the distal end o f the anchor, so
`that the outer part of the fabric seal can then extend in proximal direction.
`
`A.XVI The Features “bunched-up” and “deployed configuration” have been disclosed in
`combination
`
`Opponent 1 argues that the features “bunched-up” and “deployed configuration” have allegedly
`not been disclosed in combination. This is, however, incorrect.
`
`As already mentioned in Section A.I.2 above, these features are disclosed in combination e.g.
`in the PCT application W O’980, on page 34, lines 28 to 31 ;
`
`“When deployed, as shown in Figures 33 and 34, fabric seal 380 bunches up to create
`fabric flaps and pockets that extend into spaces form ed by the native valve leaflets 30
`382, particularly when the pockets are filled with blood in response to baclflow blood
`pressure. This arrangement creates a seal around the replacement valve. ”
`
`This passage refers to the shape of the fabric seal in the deploved state of the heart valve
`implant, and not to the deployment procedure. As discussed in Section A.XII above, a
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 14 of 42
`
`

`

`- 15 -
`
`foreshortening o f the anchor is only optional. In particular, the text passage in the PCT
`application W O’980 on page 85, lines 28/29, mentions that foreshortening “can cause seal 60
`to bunch up The same applies to the question whether or not the fabric seal must additionally
`bunch up as a result of the deployment procedure.
`
`A.XVII Amendments to the Description are admissible
`
`Mentioning the wording of granted claim 1 in the adapted summary section o f the opposed
`patent is no violation of Art. 123 (2) / 76 EPC, because granted claim 1 is originally disclosed.
`
`Opponent 2 objects that there is allegedly no basis in the original application documents for the
`last sentence o f paragraph [0018] of the patent specification { “The seal can be adapted to
`prevent blood flow around the replacement valve and the anchor when the anchor and the
`replacement valve are fully deployed. ”). However, Opponent 2 obviously missed to note
`original claim 119 which provides literal support for this feature.
`
`A.XVIII Dependent Claims 12 and 13 are originally disclosed
`
`W ith regard to granted claim 12, Opponent 2 alleges that there is no original disclosure in the
`PCT application for a balloon-expandable anchor. Opponent 2 obviously did not note the
`disclosure in the PCT application W O’980 on page 8 , lines 15 to 17, on page 22, lines 23/24,
`on page 30, lines 6 to 10, and in original claim 96.
`
`Opponents 2 and 3 also object the original disclosure of granted claim 13. But the features of
`granted claim 13 are disclosed in the PCT application W O’980 on page 8 6 , lines 31/32.
`Opponent 2 ’s objection mentioned in paragraph 107 of its opposition brief seems to be rather
`an objection under Art. 83 EPC, but not under Art. 123 (2) / 76 EPC.
`
`Edwards Lifesciences Corporation, et al. Exhibit 1131, Page 15 of 42
`
`

`

`- 1 6
`
`B.
`
`Sufficiency of Disclosure
`
`Opponent 2 did not submit any suitable argument why the skilled person should not be able to
`carry out the invention in knowledge o f the teaching of the opposed patent. As explained in
`context with Figs. 22 to 24 as well as in context with Figs. 29A to 29C of the opposed patent,
`the fabric seal is — according to the present invention — bunched-up in the deployed state o f the
`heart valve implant to provide an improved sealing effect. A bunched-up fabric seal is suitable
`to better adapt to the irregular surface o f the surrounding tissue at the implantation site.
`
`The additional objection of Opponent 2 that allegedly some essential features are missing in
`granted claim 1, is no suitable objection under Art. 83 EPC. The Opponents raised the same
`objections already under Art. 123 (2) / 76 EPC. However, as discussed in Section A above in
`detail, the arguments of the Opponents on allegedly missing features are not justified.
`
`C.
`
`Novelty of Granted Claim 1
`
`C.I
`
`Subject-matter o f the present invention
`
`Accord

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket