throbber
2005
`
`THE UNITED STATES PHARMACOPEIA
`USP 28
`
`NE 23
`THE NATIONAL FORMULARY
`
`By authority of the United States Pharmacopeial
`Convention, Inc., meeting at Washington, D.C.,
`April 12-16, 2000. Prepared by the Council ofExperts
`and published by the Board of Trustees
`Official from January 1, 2005
`
`The designation on the coverofthis publication, “USP NF
`2005,” is for ease of identification only. The publication
`contains two separate compendia: The United States
`Pharmacopeia, Twenty-Eighth Revision, and the National
`Formulary, Twenty-Third Edition.
`
`UNITED STATES PHARMACOPEIAL CONVENTION,INC.
`12601 Twinbrook Parkway, Rockville, MD 20852
`
`|9s Par Pharm.,Inc.
`
`Par Pharm., Inc.
`Exhibit 1054
`Page 001
`
`

`

`NOTICE AND WARNING
`
`Concerning U.S. Patent or Trademark Rights
`
`The inclusion in the United States Pharmacopeiaorin the National Formulary of a monographon any drug
`in respect to whichpatentor trademark rights may exist shall not be deemed, andis not intended as, a grant of,
`or authorily to exercise, any right or privilege protected by such patent or trademark. All such rights and
`privileges are vested in the patent or trademark owner, and no other person mayexercise the same without
`express permission, authority, or license secured from such patent or trademark owner.
`
`Concerning Use of USP or NF Text
`Use of the USP-NF is subject to the terms and conditions of the USP-NF License Agreement. Attention
`is called to the fact that USP and NFtext is fully copyrighted. Authors and others wishing to use portions
`of the text should request permission to do so from the Secretary of the USPC Board of Trustees.
`
`Copyright © 2004 The United States Pharmacopeial Convention,Inc.
`12601 Twinbrook Parkway, Rockville, MD 20852
`All rights reserved.
`ISSN 0195-7996
`ISBN 1-889788-25-2
`Printed in Canada by WebcomLimited, Toronto, Ontario
`
`Par Pharm., Inc.
`
`Par Pharm., Inc.
`Exhibit 1054
`Page 002
`
`

`

`Basket-rack Assembly—Thebasket-rack assembly consists of six
`open-endedtransparenttubes, each 7.75 + 0.25 cm long and having
`an inside diameter of 20.7 to 23 mm and a wall 1.0 to 2.8 mm thick;
`the tubesare held in a vertical position by twoplastic plates, each 8.8
`to 9.2 cm in diameter and 5 to 7 mm in thickness,with six holes, each
`22 to 26 mm in diameter, equidistant from the centerof the plate and
`equally spaced from one another. Attached to the undersurface of the
`lowerplate is a woven stainless steel wire cloth, which has a plain
`square weave with 1.8- to 2.2-mm mesh apertures and with a wire
`diameter of 0.63 + 0.03 mm.- The parts of the apparatus are
`assembled and rigidly held by means ofthree bolts passing through
`the two plastic plates. A suitable means is provided to suspend the
`basket-rack assembly from the raising and lowering device using a
`pointon its axis.
`The design of the basket-rack assembly may be varied somewhat
`provided the specifications for the glass tubes and the screen mesh
`size are maintained.
`
`Disks—Theuse of disks is permitted only where specified in the
`monograph. If specified in the individual monograph, each tube is
`provided with a cylindrical disk 9.5 +0.15 mm thick and
`20.7 + 0.15 mm in diameter. The disk is made of .a suitable,
`transparentplastic material having a specific gravity of between 1.18
`and 1.20. Five parallel 2-mm holes extend between the’ends of the
`cylinder. One of the holes is centered on the cylindrical axis. The
`other holes are centered 6 mm from the axis on imaginary lines
`perpendicular to the axis and parallel to each other. Four identical
`trapezoidal-shapedplanesare cut into the wall of the cylinder; nearly
`perpendicular to the ends of the cylinder. The trapezoidal: shape is
`symmetrical; its parallel sides coincide with the ends of the cylinder
`and are parallel to an imaginary line connecting the centers of two
`adjacentholes 6 mm from the cylindricalaxis. The parallel side of the
`trapezoid on the bottom ofthe cylinder has a length of 1.6 mm,andits
`center lies at a depth of 1.8 mmfrom the cylinder’s circumference.
`The parallel side of the trapezoid on the top of the cylinder has a
`length of 9.4 + 0.2 mm,andits centerlies ata depth of 2.6 + 0.1 mm
`from the cylinder’s circumference. All surfaces of the disk are
`smooth.If the use of disks is specified in the individual monograph,
`add a disk to each tube, and operate the apparatus as directed under
`Procedure.
`
`PROCEDURE
`
`Uncoated Tablets—Place 1 tablet in each of the six tubes ofthe
`basket and operate the apparatus, using water maintained at 37 + 2°
`as the immersion fluid unless otherwise specified in the individual
`monograph.Atthe end ofthe time limit specified in the monograph,
`lift the basket from the fluid, and observethetablets: all of the tablets
`have disintegrated completely. If 1 or 2 tablets fail to disintegrate
`completely, repeatthe test on 12 additionaltablets: not less than 16 of
`the total of 18 tablets tested disintegrate completely.
`Plain Coated Tablets—Apply the test for Uncoated Tablets,
`operating the apparatus for the time specified in the individual
`monograph.
`tablet in
`Delayed-Release (Enteric Coated) Tablets—Place |
`cach of the six tubes of the basket and, if the tablet has a soluble
`external coating, immersethe basket in water at room temperature for
`5 minutes. Then operate the apparatus using simulated gastric fluid
`TS maintained at 37 + 2° as the immersion fluid. After 1 hour of
`operation in simulated gastric fiuid TS,lift the basket from the fluid,
`and observethe tablets: the tablets show no evidence of disintegra-
`tion, cracking, or softening. Operate the apparatus, using simulated
`intestinal fluid TS. maintained at 37 + 2° as the immersionfluid, for
`the time specified in the monograph. Lift the basket from the fluid,
`and observethe tablets: all of the tablets disintegrate completely. If |
`or 2 tablets fail
`to disintegrate completely, repeat the test on 12
`additional tablets: not less than 16 of the total of 18 tablets tested
`disintegrate completely.
`Buccal Tablets—Apply the test for Uncoated Tablets. After 4
`hours,lift the basket from the fluid, and observethetablets: all of the
`tablets have disintegrated. If 1 or 2 tablets fail
`to disintegrate
`
`completely, repeat the test on 12 additionaltablets: not legg ha
`the total of 18 tablets tested disintegrate completely.
`Sublingual Tablets—Apply the test
`for Uncoateg
`Observethe tablets within the time limit specified in th
`monograph:all ofthe tablets have disintegrated. If 1 or2 tabl
`disintegrate completely, repeat the test on 12 additional tabja,
`less than 16 ofthe total of 18 tablets tested disintegrate coi
`Hard Gelatin Capsules—Apply the test for Uncoated
`Attach a removable wire cloth, which hasa plain square We
`1.8- to 2.2-mm mesh apertures and with a wire diameter of
`0.655 mm, as described under Basket-Rack Assembly, to th
`of the upperplate of the basket-rack assembly. Observe the
`within the time limit specified in the individual monograph:
`capsules have disintegrated except for fragments from th
`shell. If 1 or 2 capsules fail to disintegrate completely, Tepe
`on 12 additional capsules: not less than 16 ofthe total of18
`tested disintegrate completely.
`.
`Soft Gelatin Capsules—Proceed as directed under Hard Gey,
`Capsules.
`
`(711) DISSOLUTION
`
`This test is provided to determine compliance with the dissoht
`requirements wherestated in the individual monographfora ta
`capsule dosage form. Ofthe types of apparatus describedhere}
`the one specified in the individual monograph. Wherethelabel
`that an article is enteric-coated, and a dissolution or disintegration
`that does not specifically state thatit is to be appliedto enteric-
`articles is includedin the individual monograph,thetest for De
`Release Articles under Drug Release (724)
`is applied unl
`otherwise specified in the individual monograph. Forhard or
`gelatin capsules and gelatin-coated tablets that do not confo’
`Dissolution specification, repeat the test as follows. Where waler6
`medium with a pH ofless than 6.8 is specified as the Medium’
`individual monograph, the same Mediumspecified may be
`the addition of purified pepsin that results in an activity of
`Units or less per 1000 mL. For media with a pH of 6.8 o
`pancreatin can be added to produce not more than 1750 USP
`protease activity per 1000 mL.
`USP Reference Standards (11)—USP Prednisone Table
`(Dissolution Calibrator, Disintegrating). USP Salicylic Acie
`RS (Dissolution Calibrator, Nondisintegrating).
`Apparatus 1—Theassembly consists ofthe following: é
`vessel made of glass or other inert, transparent material!; a
`metallic drive shaft; and a cylindrical basket. The vessel
`immersed in a suitable water bath of any convenientsize ¢
`a heating jacket. The waterbath or heating jacket permits he
`temperature inside the vessel at 37 + 0.5° duringthetest and
`the bath fluid in constant, smooth motion. Nopart ofthe.
`including the environment
`in which the assembly 1s
`contributes significant motion, agitation, or vibration bey!
`due to the smoothly rotating stirring element. Apparatus tha’
`observation of the specimenandstirring element during
`preferable. The vesselis cylindrical, with a hemisphericalbo
`with one of the following dimensions andcapacities: for
`capacity of 1 L, the height is 160 mm to 210 mm and
`diameter is 98 mm to 106 mm; for a nominal capacity of
`height is 280 mm to 300 mm andits inside diameteris 98 m
`mm; and for a nominal capacity of 4 L, the height is 280 mm
`mm andits inside diameter is 145 mm to 155 mm.Its $
`flanged atthe top. A fitted cover maybe usedto retard evap
`The shaft is positioned so thatits axis is not more than 2 ml
`point from the vertical axis of the vessel androtates smoo
`without significant wobble. A speed-regulating device 18
`allows the shaft rotation speed to be selected and mainta “
`rate specified in the individual monograph, within +4%:
`
`
`' The materials should not sorb, react, or interfere’ with the spect
`tested.
`I
`> Ifa coveris used,it provides sufficient openingsto allow ready. f
`the thermometer and withdrawal of specimens.
`
`Par Pharm., Inc.
`
`Par Pharm., Inc.
`Exhibit 1054
`Page 003
`
`

`

`
`
`NOTES—
`
`(1) Shaft and blade material:
`
`303(or equivalent)
`stainless steel.
`
`(2) A and B dimensions are
`not to vary more than
`
`0.5 mm when part is
`rotated on ¢ axis.
`
`(3) Tolerances are 1.0mm,
`
`unlessotherwise stated.
`
`
`41.5 mm radius
`1.2mmriz
`
`Prin
`
`a\
`
`9.4 to 10.1 mm diameter
`before coating
`
`ares»
`
`
`
`A 35.8mm=Vn.
`
`
`
`‘\
`

`
`19.0mm|
`40.5mm
`1
`
`cA
`
`1
`
`Screen with welded seam:
`40 X 40 mesh,0.25-mm wire
`'
`Screen
`diameterwith wire openings of
`f
`byt 42.0apts
`
`t
`0.40 + 0.04 mm; where 20-
`aa
`40410mm
`meshscreenis specified, use
`I
`th the disso uti
`
`“4PoGon 20 X 20 mesh,0.40-mm wire
`er, Si
`A
`diameterwith wireopeningsof
`eT1DE:
`0.90 + 0.09
`mm.
`[N
`
`Ke
`Nore—Maximumallowable
`After weldin: the eama
`re the abe!
`at ‘A* is £ 1.0 mm when
`ightly
`ale
`”
`Pie
`qunout
`beslightly altered.]
`at
`=}
`"
`[ 74.0 mm to 75.0mm Foto
`
`disintegra ‘the part is rotated onGaxis
`
`to enteric-co
`with baskel mounted.
`Fig, 2. Paddle Stirring Element
`
`
`
`not less
`
`and basket componentsofthestirring element are fabricated
`
`snless steel, type 316 or equivalent, to the specifications shown
`sre
`1. Unless otherwise specified in the individual monograph,
`ncoated
`
`-mesh cloth. A basket having a gold coating 0.0001 inch (2.5
`in the
`hick may be used. The dosageunit is placed in a dry basketat
`Or 2 table
`
`ginning of each test, The distance betweenthe inside bottom of
`tional tab
`el and the basketis maintained at 25 + 2 mm duringthetest.
`
`grate eo Mple
`/ncoated |
`quare w
`
`
`
`
`aly, re
`
`.
`
`e
`
`Vent hole
`2.0 + 0.5 mm diameter
`

`Retention spring with
`3 tangs on 120° centers
`
`Clear opening
`20.2 £0.1 mm
`

`
`6.3 to6.5 or
`9.4 to 10.14mm
`
`54+05mm
`
`Screen O.D.
`22.2+1.0mm
`
`
`
`
`
`
`: applied unle
`For hardor
`
`750 USPL
`
`3
`
`Fig. 1. Basket Stirring Element
`
`
`pparatus 2—Use the assembly from Apparatus 1, except that a
`idle formed from a blade anda shaft is usedasthestirring element.
`shaft is positioned so that its axis is not more than 2 mm at any
`
`!
`1
`itfrom the vertical axis of the vessel and rotates smoothly without
`reriall
`icant wobble. The vertical center line of the blade passes
`
`4
`j
`igh the axis of the shaft so that the bottom ofthe blade is flush
`1 the bottom of ‘the shaft. The paddle conforms to.
`the
`
`ications shown in Figure 2. The distance of 25 +2 mm
`
`k
`’
`een the blade andthe inside bottom ofthe vessel is maintained
`
`“oftheassem
`g the test. The metallic or suitably inert, rigid blade and shaft
`‘
`;
`a single entity. A suitable two-part detachable design may
`on
`ed provided the assembly remains firmly engaged during the
`
`
`ratus that
`The paddle blade and shaft may be coated with a suitable inert
`
`during the
`ing The dosage unit is allowed to sink to the bottom ofthe vessel
`erical bottom
`otation of the blade is started. A small,
`loose piece of
`
`as: for a Dl
`live material such as not more than a few turnsof wire helix
`be attached to dosage units that would otherwise float. Other
`sinker devices may be used.
`
`
`
`4
`
`Apparatus Suitability Test—Individually test 1 tablet of the USP
`Dissolution Calibrator, Disintegraling Type and 1
`tablet of USP
`Dissolution Calibrator, Nondisintegrating Type, according to the
`operating conditions specified. The apparatusis suitable if the results
`obtained are within the acceptable rangestated in the certificate for
`that calibrator in the apparatus tested.
`Dissolution Medium—Usethe solventspecified in the individual
`monograph.If the Dissolution Mediumis a buffered solution, adjust
`the solution so that its pH is within 0.05 unit of the pHspecified in the
`individual monograph. [NOTE—Dissolved gases can cause bubbles to
`form, which may change the results of the test. In such cases,
`dissolved gases should be removedpriorto testing.*]
`Time—Wherea single time specificationis given, the test may be
`concluded in a shorter period if the requirement for minimum amount
`dissolved is met. If two or more timesare specified, specimensare to
`be withdrawn only at the stated times, within a tolerance of +2%.
`Procedure for Capsules, Uncoated Tablets, and Plain Coated
`Tablets—Placethe stated volumeofthe Dissolution Medium(+ 1%)
`in the vessel of the apparatus specified in the individual monograph,
`assemble the apparatus, equilibrate the Dissolution Mediwn to
`37 + 0.5°, and remove the thermometer. Place 1 tablet or 1 capsule
`in the apparatus, taking care to exclude air bubbles from the surface of
`the dosage-form unit, and immediately operate the apparatus at the
`rate specified in the individual monograph. Within the time interval
`specified, or at each of the times stated, withdraw a specimen from a
`zone midway betweenthe surface ofthe Dissolution Mediumand the
`top of the rotating basketor blade, not less than 1 cm from the vessel
`wall. [NOTE—Replacethe aliquots withdrawn for analysis with equal
`volumes of fresh Dissolution Medium at 37° or, where it can be
`shown that replacement of the medium is not necessary, correct for
`the volume changein the calculation. Keep the vessel covered for the
`duration ofthe test, and verify the temperature of the mixture under
`test at suitable times.] Perform the analysis as directed in the
`
`3 One methodof deaeration is as follows: Heat the medium, whilestirring
`gently, to about 41°, immediately filter under vacuum usinga filter having a
`porosity of 0.45 um orless, with vigorousstirring, and continuestirring under
`vacuum for about 5 minutes, Other validated deaeration techniques for
`removalof dissolved gases may be used.
`
`Par Pharm., Inc.
`
`Par Pharm., Inc.
`Exhibit 1054
`Page 004
`
`

`

`individual monograph*. Repeat the test with additional dosage form
`units.
`If automated equipmentis used for sampling and the apparatus is
`modified, validation of the modified apparatus is needed to show that
`there is no changein the agitation characteristics ofthetest.
`Where capsule shells interfere with the analysis, remove the
`contents of not fewer than 6 capsules as completely as possible, and
`dissolve the empty capsule shells in the specified volume of
`Dissolution Medium. Perform the analysis as directed in the
`individual monograph. Make any necessary correction. Correction
`factors greater than 25% of the labeled content are unacceptable.
`Procedure for a Pooled Sample for Capsules, Uncoated
`Tablets, and Plain Coated Tablets—Use this procedure where
`Procedure for a Pooled Sample is specified in the individual
`monograph. Proceed as directed under Procedure for Capsules,
`Uncoated Tablets, and Plain Coated Tablets. Combine equal volumes
`of the filtered solutions of the six or twelve individual specimens
`withdrawn,and use the pooled sample asthe test solution. Determine
`the average amountofthe active ingredient dissolved in the pooled
`sample.
`Interpretation—
`Unit Sample—Unless otherwise specified in the individual
`monograph,
`the requirements are met if the quantities of active
`ingredient dissolved from the units tested conform to the accompa-
`nying Acceptance Table.Continue testing through the three stages
`unless the results conform at either S, or S,. The quantity, Q, is the
`amount of dissolved active ingredient specified in the individual
`monograph, expressedas a percentageof the labeled content; the 5%,
`15%, and 25% valuesin the Acceptance Table are percentages of the
`labeled content so that these values and Q are in the same terms.
`
`
`
`
`
`To determine the range of temperatures within which
`liquid distils, or the percentage ofthe materialthatdistil b a!
`specified temperatures, use Mcthod I or MethodII ag direoy:
`individual monograph.The /owerlimit of the rangeis the ten
`indicated by the thermometer whenthefirst drop ofcondensa
`the tip of the condenser, and the upperlimit is the Dry Poin
`temperature at which the last drop of liquid evaporates
`lowest point in the distillation flask, without regard to
`remaining onthe side of the flask, or the temperature obsery,
`the proportion specified in the individual monograph
`collected.
`NOTE—Coolallliquids that distill below 80° to between 1(
`15° before measuring the sample to be distilled,
`
`METHODI
`
`Apparatus—Use apparatussimilarto that specified forMeth
`except that the distilling flask is of 50- to 60-mL capacity,
`neckofthe flask is 10 to 12 cm long and 14 to 16 mm inj
`diameter. The perforation in the upperinsulating board,ifone
`should be such that when the flask is setinto it, the portion of
`belowthe uppersurface of the insulating materialhas a capacity
`to 4 mL.
`Procedure—Proceed as directed for Method IT, but pla
`flask only 25 mL ofthe liquid to be tested.
`
`
`
`Acceptance Table
`
`METHODII
`
`S;
`
`12
`
`Pooled Sample—Unless otherwise specified in the individual
`monograph,
`the requirements are met if the quantities of active
`ingredient dissolved from the pooled sample conform to the
`accompanying Acceptance Table for a Pooled Sample. Continue
`testing through the three stages unlessthe results conform at either S,
`or S,. The quantity, QO, is the amountof dissolved active ingredient
`specified in the individual monograph, expressed as a percentage of
`the labeled content.
`
`Acceptance Table for a Pooled Sample
`Number
`
`Stage
`Tested
`Acceptance Criteria
`Ss;
`6
`Average amountdissolvedis not less than
`+ 10%,
`Average amountdissolved (S, + S,) is equal
`to or greater than O + 5%.
`12
`Average amount dissolved (S, + S, + §;) is
`
`equal to or greater than Q.
`
`S,
`
`S;
`
`Number
`Apparatus—Use an apparatus consisting ofthe followin;
`
`
`Stage_Tested Acceptance Criteria
`Distilling Flask—A round-bottomdistilling flask, of hea’
`S,
`6
`Each unit is not less than Q + 5%.
`glass, of 200-mL capacity, and havingatotal length of 17 t
`and aninside neck diameter of 20 to 22 mm. Attached abo
`S;
`6
`Average of 12 units (S, + S,) is equal to or
`greater than Q, and no unitis less than O —
`on the neck, approximately 12 cm from the bottom ofthe
`15%.
`side-arm 10 to 12 cm long and 5 mm in internal diame!
`Average of 24 units (S, + S, + §,) is equal to
`forms an angle of 70° to 75° with the lower portion ofthe neck
`or greater than Q, not more than 2 units are
`Condenser—Astraight glass condenser 55 to 60 cmin length
`less than Q— 15%,andno unit is less than 0
`a water jacket about 40 cm in length, or a condenserof o
`— 25%,
`having equivalent condensing capacity. The lower endo}
`condenser may be bent to provide a delivery tube,or it
`connected to a bent adapter that serves as a delivery tube.
`Insulating Boards—Twopieces of insulating board,
`thick and 14 to 16 cm square, suitable for confining the
`lowerpart ofthe flask. Each board has a holein its center, and
`boards differ only with respect to the diameter of the hol
`diameters are 4 and 10 cm,respectively. In use, the boards
`one uponthe other, and resting on a tripod or other suitabl
`with the board having the larger hole on top.
`Receiver—A 100-mL cylinder graduated in 1-mL subdi
`Thermometer—In order to avoid the necessity for an
`stem correction, an accurately standardized, partia
`thermometer having the smallest practical subdivisions (1
`than 0.2°) is recommended. Suitable thermometers are ava
`the ASTM E-1 series 37C through 41C, and 102C through 10
`Thermometers (21)). Whenplacedin position, the stem islo
`the centerofthe neck andthetop ofthe contraction chamber (@
`if 37C or 38Cis used)is level with the bottom ofthe outlet to the si®
`arm.
`
`Heat Source—A small Bunsen burner or an electric
`mantle capable of adjustment comparable to that possib
`Bunsen burner.
`Procedure—Assemble the apparatus, and place in the
`mL ofthe liquid to be tested,taking care not to allow any of
`to enter the side-arm.Insert the thermometer, shield the ent
`and flask assembly from external air currents, and apP!
`regulating it so that between 5 and 10 minutes elapse befor
`
`dropofdistillate falls from the condenser. Continuethe dist
`
`arate of 4 to 5 mL ofdistillate per minute, collecting the¢
`
`the receiver. Note the temperature whenthefirst dropofdist
`from the condenser, and again whenthelast drop ofliqu!
`
`
`
`6
`
`
`
`* Iftest specimensarefiltered, use an inert filter that does not cause adsorption
`ofthe active ingredient or contain extractable substances that would interfere
`with the analysis.
`
`
`
`
`
`Par Pharm., Inc.
`
`Par Pharm., Inc.
`Exhibit 1054
`Page 005
`
`

`

`Nicaragua
`
`
`
`
`
`
`al Information
`
`(1151) Pharmaceutical Dosage Forms
`
`2701
`
`‘Table 1.
`
`international Climatic
`
`7AMES
`
`
`
`
`
`_Derived Data.as,
`
`Ye
`
`Co RH
`mbar
`21
`45
`1.2
`
`
`13.5
`25
`60
`19.0
`
`(Temperate
`Japan
`Jnited Kingdom
`orthern Europe
`anada
`ussia
`United States
`
`(Mediterranean,Subtropical
`United States
`Japan
`Southern Europe
`(Portugal-Greece)
`(oiDry
`ran
`raq
`Sudan
`WW,Hot,Humid —
`Brazil
`Ghana
`ndonesia
`
`~
`
`Eg 30
`
`——
`
`35005.0
`
`
`~ 7626.6
`30
`70
`30.0
`
`
`Data recorded as < 19°
`calculated as 19°,
`
`culated mean Kinetic temperature.
`
`
`Partial pressure of water vapor.
`
`
`
`
`
`(1151) PHARMACEUTICAL
`DOSAGE FORMS —
`
`_ Dosage forms are provided tor most of the Pharmacopeial drug
`aibstances, but the processesfor the preparation of manyofthemare.
`in general, beyond the scope of the Pharmacopeia.
`In addition to
`defining the dosage forms, this section presents the general principles
`involved in the manufacture of some ofthem, particularly on a small
`cae. Other
`informauion thal
`is given bears on the use of the
`in extemporaneous compounding of
`Pharmacopeial substances
`dosage forms.
`
`BIOAVAILABILITY
`
`
`
`woavallability, or the e
`tent 1o which the therapeutic constituent of
`| Piarmacentical dosage form intended for oral or topical use is
`
`
`available for absorption, is influenced
`bya variety of factors. Among
`he inherent factors known to affec absorption are
`the method of
`paiacture or method of compounding: the particle size and crystal
`> polymorph of the drug substance; and the diluents and
`scipients used in formulating the dosage form,
`includingfillers,
`
`
`inders, disintegrating agents,
`b ibricants, coatings, solvents, sus-
`
`pending agents, and dyes. Lubricants and coatings are foremost
`
`ong these, The maintenance of a demonsttrably high degree of
`
`loavailability requires particular attention to all aspects of
`loduction and quality control
`that may affect
`the nature of the
`fhished dosage form.
`
`
`
`
`
`TERMINOLOGY
`
`
`Occasionally it is necessary to add solvent to the contents of 4
`
`‘Ontainer just prior to use, usually because ofinstability of some
`tugsin the diluted form. Thus, a solid dilutedlo yield a suspensionis
`
`alled [DRUG] jor Suspension; a solid dissolvedand diluted to yield
`Solution is called [DRUG]
`for Sofution;, and a solution or
`
` : Desnsion diluted to yield a more dilute form of the drug is called
`DRUG] Oral Concentrate. After dilution,
`it
`is important that the
`tug be homogeneously dispersed belure administration.
`
`
`
`
`
`AEROSOLS
`
`Pharmaceutical aerosols are products that are packaged under
`pressure and contain therapeutically active ingredients that” are
`released upon activation of an appropriate valve system. They are
`imtended for topical applicationto the skin as well as local application
`into the nose (nasa! aerosols), mouth (lingual aerosols), or lungs
`(inhalation aerosols). These products may be fitted with valves
`enabling either continuous or metered-dose delivery; hencé; the terms
`“IDRUG| Metered Topical Aerosols,” “[DRUG] Metered Nasal
`Aerosols,” ete.
`The term “aerosol”refers to the fine mist of spraythat results from
`
`most pressurized systems. However,
`the term has been broadly
`misayaplie2d to all self-contained pressurized products, some of which
`deliver foarns or semisolid fluids. In the case of Inhalation Aerosols,
`the particle size of the delivered medication ‘must be carefully
`controlled, andthe average size ofthe particles should be under 5 um.
`These products are also known as metered-dose inhalers (MDI).
`Other aerosol sprays may contain particles up to several hundred
`micrometers in diameter.
`The basic components of an aerosol system are the container, the
`propellant,
`the concentrate containing the active ingredient(s),
`the
`valve, and the actuator. The nature ofthese components determines
`such characteristics as particle size distribution, uniformity of dose
`for metered valves, delivery ratc, wetness and temperature of the
`spray, spray pattem and velocity or plume geometry, foam density,
`and fluid viscosity.
`
`Types of Aerosols
`
`Aerosols consist oftwo-phase (gas and liquid) or three-phase (gas,
`liquid, andsolid orliquid) systems. The two-phase aerosol consists of
`a
`solution of active ingredients in liquefied propellant and the
`vaporized|propellant, The solvent is composed ofthe’propellant or a
`mixture of the propellant and cosolvents such as alcohol, propylene
`glycol, and polyethylene glycols, whichare often used to enhance the
`solubility of the active ingredients.
`Three-phase systems consist of a suspension or emulsion of the
`active ingredient(s)
`in addition to the vaporized propellants. A
`suspension consists ofthe active ingredient(s) that maybe dispersed
`inin
`the propellant system with the aid of suitable excipients such as
`vetting agents and/orsolid carriers such as tale or colloidal silicas.
`
`
`
`
`
`Par
`
`Pharm., Inc.
`Exhibit 1054
`Page 006
`
`Par Pharm., Inc.
`Exhibit 1054
`Page 006
`
`

`

`2702
`
`(1151) Pharmaceutical Dosage Forms / General iaformution
`
`
`
`
`
`Manufacture
`
`Acrosols are usually prepared by one of two e
`the “cold-fll’ process,
`the concentrate (gen
`temperature below 0°) and the refrigerated propellant ar
`into open containers (usually chilled). The valve-actu;
`then crimped onto the containerto forma pressure-tig
`the interval between propellant addition and crimp;
`
`volatilization of propellant occurs to displace air from the eer
`
`In the “pressure-fill” method,
`the concentrate is placed |
`
`contamer, and either the propellant is forced underpr
`ropellant 4
`the valve orifice afterthe valve is sealed, or the p
`to flow under the valve cap and then the valve
`(“under-the-cap” filling). In both cases of the “pr
`provision must be made for evacuation of air by m
`displacement with a srnall amountofpropellant vapo
`process controls usually include monitoring of proper ¢
`and propellantfill weight andpressuretesting, leak testing-ang
`function testing of the finished aerosol. Microbiological aga
`shouldalso be controlled.
`
`|
`ayespot
`dvough ¢
`actory-fi
`banding, i
`
`Extractable Substances
`Since pressurized inhalers and aerosols are no
`with organic solvents as the propellant or the vehic!
`
`extractables from the elastomeric and plastic component
`formulation is a potentially serious problem. Thus,
`
`and the quality of materials used in the manufacture
`
`components should be well established so as to prev
`ofthe valve components and to minimize changes in the medic
`delivery, leak rate, and impurity profile ofthe drug product ove
`
`The extractable profiles of a representative sample of each
`elastomeric andplastic components ofthe valve should be estab
`underspecified conditions and should he correlated to the extra
`
`profile of the aged drug product or placebo, to ensure reprod
`quality and purity of the drug product. Extractab
`i
`
`include polynuclear aromatics, nitrosamines, vulcani
`
`alors, antioxidants, plasticizers, monomers, etc., should bé-i
`and minimized wherever possible.
`:
`Specifications and limits for individual and total extractable.
`
`different valve components may require the use ofdifferent and
`
`methods.
`In addition,
`the standard USP biological testing
`general test chapters Biological Reactivity Tests,
`In Vitro |
`Biological Reactivity Tesis, In Vivo (88)) as well as othersaf
`may be needed.
`:
`
`
`
`
`is an emulsion containing one or more active
`A foam aerosol
`ingredients, surfactants, aqueous or nonaqueous liquids, and -the
`propellants. Hf the propellant is in the intemal (discontinuous) phase
`(Le., of the oil-in-water typc), a stable foamis discharged; andifthe
`propellant is in the external (continuous) phase (i.e., of the water-in-
`
`oil type), a spray or a quick-breaking foamis discharged.
`
`Propellants
`
`The propellant supplies the necessary pressure within an aerosol
`system to expel malerial from the containerand, in combination with
`other components, to convert the material into the desired physical
`form. Propellants may be broadly classified as
`liquefied or
`compressed gases having vapor pressures generally exceeding
`atmospheric pressure. Propellants within this definition include
`various hydrocarbons, especially halogenated derivatives of methane,
`ethane, and propane, low molecular weight hydrocarbons suchas the
`butanes and pentanes, and compressed gases such as carbon divxide,
`nitrogen, and nitrous oxide. Mixtures of propellants are frequently
`used to obtain desirable pressure, delivery, and spray characteristics.
`A good propellant sysiem should have the proper vapor pressure
`characteristics consistent with the other aerosol components.
`
`Valves
`
`
`
`The primary function of the valve is to regulate the flow ofthe
`therapeutic agent and propellant
`from the container, The spray
`characteristics of the acrosol are influenced by orifice dimension,
`number, and location. Most aerosol valves provide for continuous
`spray operation and are used on most topical products. However,
`pharmaceutical products for oral or nasa!
`inhalation often utilize
`metered-dose valves that must deliver a uniform quantity of spray
`uponeach vaive activation. The accuracy and reproducibility of the
`doses delivered from metering valves are generally good, comparing
`favorably to the uniformity of solid dosage forms suchas tablets and
`capsules. However, when aerosol packagesare stored improperly, or
`when they have not been used for long periods of time, valves must
`be primed before use. Matcrials used for the manufacture of valves
`should be inert to the formulations used. Plastic, rubber, aluminum,
`and stainless steel valve components are commonly used. Metered-
`dose valves must deliver an accurate dose within specified tolerances.
`
`Actuators
`
`Au actualoris the fitting attached to an aerasol valve stem, which
`when depressed or moved, opens the valve, and directs. the spray
`containing the drug preparation to the desired area. The actuator
`usually indicates the direction in whichthe preparation is dispensed
`and protects the hand or finger from the refrigerant effects of the
`propellant. Actuators incorporate an orifice that may vary widely in
`size andshape. The size ofthis orifice, the expansion chamberdesign,
`and the nature of the propellant and formulation influence the
`delivered dose as well as the physical characteristics of the spray,
`foam, orstreamofsolid particles dispensed. For inhalation aerosols,
`an actuator capable of delivering the medication in the properparticle
`size range and with the appropriate spray pattern and plume geometry
`is utilized.
`
`Containers
`
`Aerosol containers usually are made ofglass,plastic, or metal, or a
`combination of these materials. Glass containers must be precisely
`engineered to provide the maxi

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket