throbber
THE UNITED STATES PHARMACOPEIA
`
`THE NATIONAL FORMULARY
`
`12601 Twinbrook Parkway, Rockville, MD 20852
`
`The designation on the cover of this publication, “USP NF
`2005,” is for ease of identification only. The publication
`contains two separate compendia: The United States
`Pharmacopeia, Twenty-Eighth Revision, and the National
`Formulary, Twenty-Third Edition.
`
`By authority of the United States Pharmacopeial
`Convention, Ina, meeting at Washington, D. C,
`April 12—16, 2000. Prepared by the Council ofExperts
`and published by the Board of Trustees
`
`Oflicz’alfi‘om January 1, 2005
`
`UNITED STATES PHARMACOPEIAL CONVENTION, INC.
`
`Par Pharm., Inc.
`Exhibit 1054
`Page 001
`
`

`

`NOTICE AND WARNING
`
`Concerning US. Parent or Trademark Rig/its
`
`The inclusion in the United States Phannacopeia 01‘ in the National Formulaiy of a monograph on any drug
`in respect to which patent or trademark rights may exist shall not be deemed, and is not intended as, a grant of,
`or authority to exercise, any right or privilege protected by such patent or trademark. All such rights and
`privileges are vested in the patent or trademark owner, and no other person may exercise the same without
`express permission, authority, or license secured from such patent or trademark owner.
`
`Printed in Canada by Webcom Limited, Toronto, Ontario
`
`Concerning Use ofUSP 01' NF Text
`
`Use of the USP~NF is subject to the terms and conditions of the USPJVF License Agreement. Attention
`is called to the fact that USP and NI? text is fully copyrighted. Authors and others wishing to use portions
`of the text should request permission to do so from the Secretary of the USPC Board of Trustees.
`
`Copyright (Q 2004 The United States Pliarrnacopeial Convention, Inc.
`12601 Twinbrook Parkway, Rockville, MD 20852
`All rights reserved.
`ISSN 0195-7996
`ISBN 1—8897884252
`
`Par Pharm., Inc.
`Exhibit 1054
`Page 002
`
`

`

`Basket—rack Assembly—The basket-rack assembly consists of six
`open-ended transparent tubes, each 7.75 i 0.25 cm long and having
`an inside diameter of 20.7 to 23 mm and a wall 1.0 to 2.8 mm thick;
`the tubes are held in a vertical position by two plastic plates, each 8.8
`to 9.2 cm in diameter and 5 to 7 mm in thickness, with six holes, each
`22 to 26 mm in diameter, equidistant from the center of the plate and
`equally spaced from one another. Attached to the under surface of the
`lower plate is a woven stainless steel wire cloth, which has a plain
`square weave with 1.8- to 2.2-mm mesh apertures and with a wire
`diameter of 0.63 i 0.03 mm. The parts of the apparatus are
`assembled and rigidly held by means of three bolts passing through
`the two plastic plates. A suitable means is provided to suspend the
`basket—rack assembly from the raising and lowering device using a
`point on its axis.
`The design of the basket-rack assembly may be varied somewhat
`provided the specifications for the glass tubes and the screen mesh
`size are maintained.
`
`Disks—The use of disks is permitted only where specified in the
`monograph. If specified in the individual monograph, each tube is
`provided with a cylindrical disk 9.5 i 0.15 mm thick and
`20.7 J_r 0.15 mm in diameter. The disk is made of a suitable,
`transparent plastic material having a specific gravity of between 1.18
`and 1.20. Five parallel 2-mm holes extend between the'ends of the
`cylinder. One of the holes is centered on the cylindrical axis. The
`other holes are centered 6 mm from the axis on imaginary lines
`perpendicular to the axis and parallel to each other. Four identical
`trapezoidal—shaped planes are cut into the wall of the cylinder, nearly
`perpendicular to the ends of the cylinder. The trapezoidal shape is
`symmetrical; its parallel sides coincide with the ends of the cylinder
`and are parallel to an imaginary line connecting the centers of two
`adjacent holes 6 mm from the cylindrical axis. The parallel side of the
`trapezoid on the bottom of the cylinder has a length of 1.6 mm, and its
`center lies at a depth of 1.8 mm from the cylinder’s circumference.
`The parallel side of the trapezoid on the top of the cylinder has a
`length of9.4 i 0.2 mm, and its center lies at a depth of2.6 i 0.1 mm
`from the cylinder’s circumference. All surfaces of the disk are
`smooth. If the use Of disks is specified in the individual monograph,
`add a disk to each tube, and operate the apparatus as directed under
`Procedure.
`
`PROCEDURE
`
`Uncoated TabletsiPlace 1 tablet in each of the six tubes of the
`basket and operate the apparatus, using water maintained at 37 i 2“
`as the immersion fluid unless otherwise specified in the individual
`monograph. At the end of the time limit specified in the monograph,
`lift the basket from the fluid, and observe the tablets: all of the tablets
`have disintegrated completely. If 1 or 2 tablets fail to disintegrate
`completely, repeat'the test on 12 additional tablets: not less than 16 of
`the total of 18 tablets tested disintegrate completely.
`Plain Coated Tablets—Apply the test for Uncoated Tablets,
`operating the apparatus for the time specified in the individual
`monograph.
`tablet in
`Delayed-Release (Enteric Coated) Tablets—Place 1
`each of the six tubes of the basket and, if the tablet has a soluble
`external coating, immerse the basket in water at room temperature for
`5 minutes. Then operate the apparatus using simulated gastric fluid
`TS maintained at 37 i 2“ as the immersion fluid. After 1 hour of
`operation in simulated gastric fluid TS, lift the basket from the fluid,
`and observe the tablets: the tablets show no evidence of disintegra-
`tion, cracking, or softening. Operate the apparatus, using simulated
`intestinal fluid TS maintained at 37 i 2° as the immersion fluid, for
`the time specified in the monograph. Lift the basket from the fluid,
`and observe the tablets: all of the tablets disintegrate completely. If 1
`or 2 tablets fail
`to disintegrate completely, repeat the test on 12
`additional tablets: not less than 16 of the total of 18 tablets tested
`disintegrate completely.
`Buccal Tablets—Apply the test for Uncaated Tablets. Afler 4
`hours, lift the basket from the fluid, and observe the tablets: all of the
`tablets have disintegrated. If 1 or 2 tablets fail
`to disintegrate
`
`'
`
`This test is provided to determine compliance with the diss ,
`requirements where stated in the individual monograph for a ‘
`capsule dosage form. Of the types of apparatus described he ,
`the one specified in the individual monograph. Where the la
`that an article is enteric-coated, and a dissolution or disintegra In
`that does not specifically state that it is to be applied to cute
`articles is included in the individual monograph, the test forD
`Release Articles under Drug Release (724)
`is applied Hm
`otherwise specified in the individual monograph. For hard:
`gelatin capsules and gelatin-coated tablets that do not confo
`Dissolution specification, repeat the test as follows. Where W .
`medium with a pH of less than 6.8 is specified as the Mediu
`individual monograph, the same Medium specified may be
`the addition of purified pepsin that results in an activity of
`Units or less per 1000 mL. For media with a pH of 6.8 o
`pancreatin can be added to produce not more than 1750 USP
`protease activity per 1000 mL.
`USP Reference Standards (ll)—USP Prednisoné
`(Dissolution Calibrator; Disintegraring). USP Salicylic Aci.
`RS (Dissolution Calibrator; Nondisinregrating).
`Apparatus l—The assembly consists of the following.
`vessel made of glass or other inert, transparent material‘;
`metallic drive shaft; and a cylindrical basket. The vessel
`immersed in a suitable water bath of any convenient size _
`a heating jacket. The water bath or heating jacket permits 4
`temperature inside the vessel at 37 i 05° during the test an
`the bath fluid in constant, smooth motion. No part of the.
`including the environment
`in which the assembly 15
`contributes significant motion, agitation, or vibration hey
`due to the smoothly rotating stirring element. Apparatus tha
`observation of the specimen and stirring element during
`preferable. The vessel is cylindrical, with a hemisphencal b0
`with one of the following dimensions and capacities: for
`capacity of l L, the height is 160 mm to 210 mmfifld
`diameter is 98 mm to 106 mm; for a nominal capacity 0f“
`height is 280 mm to 300 mm and its inside diameter is 98 I!
`m; and for a nominal capacity of 4 L, the height is 280 mm
`mm and its inside diameter is 145 mm to 155 mm. Its S
`flanged at the top. A fitted cover may be used to retard eVaP
`The shaft is positioned so that its axis is not more than 2 “
`point from the vertical axis of the vessel and rotates 511309
`without significant wobble. A speed-regulating devrqe 11?
`allows the shafi rotation speed to be selected and mall.)1m ‘
`rate specified in the individual monograph, within i4/“-' ‘
`
`
`completely, repeat the test on 12 additional tablets: not less I,
`the total of 18 tablets tested disintegrate completely,
`Sublingual Tablets—Apply the test
`for Uncoated
`Observe the tablets within the time limit specified in th
`monograph: all ofthe tablets have disintegrated. If 1 or 2 tab]
`disintegrate completely, repeat the test on 12 additional ta .
`less than 16 of the total of 18 tablets tested disintegrate I”:
`Hard Gelatin Capsules—Apply the test for Ultcoafed'
`Attach a removable wire cloth, which has a plain square W ”‘
`1.8- to 2.2-mm mesh apertures and with a wire diameter 0 "I
`0.655 mm, as described under BasketrRack Assembly, to
`of the upper plate of the basket-rack assembly. Observe flfe
`within the time limit specified in the individual monograph:
`capsules have disintegrated except for fragments from [h
`shell. If 1 or 2 capsules fail to disintegrate completely, mpg
`on 12 additional capsules: not less than 16 of the total on?
`tested disintegrate completely.
`-
`Soft Gelatin Capsules—Proceed as directed under Hard ,3,“
`Capsules.
`
`<711> DISSOLUTION
`
`l The materials should not sorb, react, or interfere with the Spec.
`tested.
`2 ifa cover is used, it provides sufficient openings to allow ready“
`the thermometer and withdrawal of specimens.
`
`Par Pharm., Inc.
`Exhibit 1054
`Page 003
`
`

`

`and basket components of the stirring element are fabricated
`less steel. type 316 or equivalent, to the specifications shown
`‘4 re 1, Unless otherwise specified in the individual monograph,
`r, mesh cloth. A basket having a gold coating 0.0001 inch (25
`‘ ' k may be used, The dosage unit is placed in a dry basket at
`‘ g of each test. The distance between the inside bottom of
`e1 and the basket is maintained at 25 i 2 mm during the test.
`
`.
`
`‘
`
`E
`
`a
`
`9.4 to 10.1 mm diameter
`before CDfll/Ilg
`
`Io
`
`
`
`removal of dissolved gases may be used.
`
`Apparatus Suitability Test—Individually test 1 tablet of the USP
`Dissolution Calibratol; Disintegraling B’pe and 1
`tablet of USP
`Dissolution Calibrator; Nondisinregrating Type, according to the
`operating conditions specified. The apparatus is suitable if the results
`obtained are within the acceptable range stated in the certificate for
`that calibrator in the apparatus tcstcd.
`Dissolution Medium—Use the solvent specified in the individual
`monograph. If the Dissolution Medium is a buttered solution, adjust
`the solution so that its pH is within 0.05 unit of the pH specified in the
`individual monograph. [NOTEiDissolved gases can cause bubbles to
`form, which may change the results of the test. In such cases,
`dissolved gases should be removed prior to testingfij
`Time—Where a single time specification is given, the test may be
`concluded in a shorter period if the requirement for minimum amount
`dissolved is met, If two or more times are specified, specimens are to
`be withdrawn only at the stated times, within a tolerance of :2%.
`Procedure for Capsules, Uncoated Tablets, and Plain Coated
`Tablets—Place the stated volume of the Dissolution Medium (-_|- 1%)
`in the vessel of the apparatus specified in the individual monograph,
`assemble the apparatus, equilibrate the Dissolution Medium to
`37 i 0.5“, and remove the thermometer. Place 1 tablet or 1 capsule
`in the apparatus, taking care to exclude air bubbles from the surface of
`the dosage-form unit, and immediately operate the apparatus at the
`rate specified in the individual monograph. Within the time interval
`specified, or at each of the times stated, withdraw a specimen from a
`zone midway between the surface of the Dissolution Medium and the
`top of the rotating basket or blade, not less than 1 cm from the vessel
`wall. [NOTE—Replace the aliquots withdrawn for analysis with equal
`volumes of fresh Dissolution Medium at 37“ or, where it can be
`shown that replacement of the medium is not necessary, correct for
`the volume change in the calculation. Keep the vessel covered for the
`duration of the test, and verify the temperature of the mixture under
`test at suitable times] Perform the analysis as directed in the
`
`
`(E
`
`6.3 to 6.5 or
`9.4 to 10.1 mm
`
`Screen OD.
`22.2 t 1.0 mm
`
`Screen with welded seam:
`40 X 40 mesh, 0.25-mm wire
`diameter with wire openings oi
`0.40 1 0.04 mm; where 20-
`mesh screen is specified, use
`20 X 20 mesh, GAO-mm wire
`diameter with wire openings of
`0.90 i 0.09 mm. [NOTE
`After welding. the screen may
`be slightly altered]
`
`Vent hole
`2.0 1 0.5 mm diameter
`
`Retention spring with
`3 iargs on 120“ centers
`
`Clear opening
`202:0.1mrn
`
`37.0 t
`3.0 mm
`
`:t 1.0 mm
`ODE"
`
`.
`
`._
`
`'
`
`'
`
`th the was:
`
`Molt—Maximum allowable
`i
`l
`at'A“ isiiflmmwhen
`-partis rotated on (Eaxis
`H i basket mounted,
`
`;
`
`Fig. 1. Basket Stirring Element
`
`i
`
`raj aratus 2~Use the assembly from Apparatus I, except that a
`ii" formed from a blade and a shaft is used as the stirring element.
`in ft is positioned so that its axis is not more than 2 mm at any
`m'n mthc vertical axis of the vessel and rotates smoothly without
`it ant wobble. The vertical center line of the blade passes
`tip the axis of the shaft so that the bottom of the blade is flush
`“7 he bottom of the shaft. The paddle conforms to the
`cations shown in Figure 2. The distance of 25 i 2 mm
`the blade and the inside bottom of the vessel is maintained
`' the test. The metallic or suitably inert, rigid blade and shalt
`e a single entity. A suitable two-part detachable design may
`*--';.I provided the assembly remains firmly engaged during the
`it- paddle blade and shafi‘ may be coated with a suitable inert
`“" ‘The dosage unit is allowed to sink to the bottom of the vessel
`otation of the blade is started. A small,
`loose piece of
`live material such as not more than a few turns of wire helix
`"‘ attached to dosage units that would otherwise float. Other
`sinker devices may be used.
`
`.
`
`(2) A and 8 dimensions are -
`not to vary more than
`0.5 mm when part is
`rotated on ‘1: axis.
`(3) Tolerances are #0 mm,
`unless otherwise stated.
`
`NOTES—
`{0 Shaft and blade material:
`303 {or equivalent)
`
`stainlesssteel.
`
`41.5 mm lad/us
`1.2 mm «1di
`,_.
`-t\
`‘\‘
`\\B
`
`4 01 1.0mm
`
`L; 74.0 mm to 75.0 mm *4
`Fig. 2. Paddle Stirring Element
`
`3 One method of deaeration is as follows: Heat the medium, while stirring
`gently, to about 41", immediately filter under vacuum using a filter having a
`porosity of 0.45 pm or less, with vigorous stirring. and continue stirring under
`vacuum for about 5 minutes. Other validated deaeration techniques for
`
`Par Pharm., Inc.
`Exhibit 1054
`Page 004
`
`

`

`(721) DISTILLING ;;
`
`To determine the range of temperatures Within which
`liquid distils, or the percentage ofthe material that distils b 'n
`specified temperatures, use Method 1 or Method [1 as d- t v.
`individual monograph. The lower limit of the range is the te
`indicated by the thermometer when the first drop ofcondens
`the tip of the condenser, and the upper limit is the Dry Pom
`temperature at which the last drop of liquid evaporates
`lowest point in the distillation flask, without regard to
`remaining on the side of the flask, or the temperature observ
`the proportion specified in the individual monograph I
`collected.
`
`NOTE~Cool all liquids that distill below 80° to between '3
`15° before measuring the sample to be distilled.
`
`METHOD I
`
`Apparatus—Use apparatus similar to that specified ft)qu ,,
`except that the distilling flask is of 50- to 60—mL capacity in
`neck of the flask is 10 to 12 cm long and 14 to 16 mm {1;
`diameter. The perforation in the upper insulating board, ifone
`should be such that when the flask is set into it, the portion of
`below the upper surface of the insulating material has a capac i
`to 4 mL.
`
`Procedure—Proceed as directed for Method II, but p ‘
`flask only 25 mL of the liquid to be tested.
`
`Acceptance Table
`
`METHOD II
`
`Number
`
`Tested
`Stage
`Acceptance Criteria
`SI
`6
`Each unit is not less than Q + 5%.
`S2
`6
`Average of 12 units (SI + $2) is equal to or
`greater than Q, and no unit is less than Q —
`15%.
`Average of 24 units (SI + S2 + 8,) is equal to
`or greater than Q, not more than 2 units are
`less than Q 7 15%, and no unit is less than Q
`
`— 25%.
`
`12
`
`individual monograph“. Repeat the test with additional dosage form
`units.
`If automated equipment is used for sampling and the apparatus is
`modified, validation of the modified apparatus is needed to show that
`there is no change in the agitation characteristics of the test.
`Where capsule shells interfere with the analysis, remove the
`contents of not fewer than 6 capsules as completely as possible, and
`dissolve the empty capsule shells in the specified volume of
`Dissolution Medium. Perform the analysis as directed in the
`individual monograph. Make any necessary correction. Correction
`factors greater than 25% of the labeled content are unacceptable.
`Procedure for a Pooled Sample for Capsules, Uncoated
`Tablets, and Plain Coated Tablets—Use this procedure where
`Procedure for a Pooled Sample is specified in the individual
`monograph. Proceed as directed under Procedure fiJr Capsules,
`Uncoated Tablets, and Plain Coated Tablets. Combine equal volumes
`of the filtered solutions of the six or twelve individual specimens
`withdrawn, and use the pooled sample as the test solution. Determine
`the average amount of the active ingredient dissolved in the pooled
`sample.
`Interpretation—
`Unit SampleiUnless otherwise specified in the individual
`monograph,
`the requirements are met if the quantities of active
`ingredient dissolved from the units tested conform to the accompa-
`nying Acceptance Table.Continue testing through the three stages
`unless the results conform at either SI or $2. The quantity, Q, is the
`amount of dissolved active ingredient specified in the individual
`monograph, expressed as a percentage of the labeled content; the 5%,
`15%, and 25% values in the Acceptance Table are percentages of the
`labeled content so that these values and Q are in the same terms.
`
`from the condenser, and again when the last drop Ofllqwd
`
`S3
`
`Pooled Sample—Unless otherwise specified in the individual
`monograph,
`the requirements are met if the quantities of active
`ingredient dissolved from the pooled sample conform to the
`accompanying Acceptance Table for a Pooled Sample. Continue
`testing through the three stages unless the results conform at either S1
`or S2. The quantity, Q, is the amount of dissolved active ingredient
`specified in the individual monograph, expressed as a percentage of
`the labeled content.
`
`Acceptance Table for a Pooled Sample
`Number
`
`Stage
`Tested
`Acceptance Criteria
`6
`Average amount dissolved is not less than Q
`+ 10%.
`Average amount dissolved (Sl .+ 8,) is equal
`to or greater than Q + 5%.
`12
`Average amount dissolved (SI + S, + 8,) is
`
`equal to or greater than Q.
`
`6
`
`
`
`lt'test specimens are filtered, use an inert filter that does not cause adsorption
`"l
`of the active ingredient or contain extractable substances that would interfere
`with the analysis.
`
`Apparatus—Use an apparatus consisting of the followin'
`Distilling Flask—A round-bottom distilling flask, ofhea
`glass, of 200—mL capacity, and having a total length of 17"
`and an inside neck diameter of 20 to 22 mm. Attached abo
`on the neck, approximately 12 cm from the bottom of the
`side-arm 10 to 12 cm long and 5 mm in intemal diame
`forms an angle of 70u to 75C with the lower portion of they .
`Condenser-A straight glass condenser 55 to 60 cm in lo,‘
`a water jacket about 40 cm in length, or a condenser of 0
`having equivalent condensing capacity. The lower en‘ '
`condenser may be bent to provide a delivery tube, or it
`connected to a bent adapter that serves as a delivery tube. p .
`Insulating Boards—Two pieces of insulating board,
`thick and 14 to 16 cm square, suitable for confining the
`lower part of the flask. Each board has a hole in its center, :
`boards differ only with respect to the diameter of the 1101
`diameters are 4 and 10 cm, respectively. In use, the boards
`one upon the other, and resting on a tripod or other suitabl
`with the board having the larger hole on top.
`Receivet'iA lOO-mL cylinder graduated in l-mL $1!de
`Thermometer—In order to avoid the necessity for all
`stern correction, an accurately standardized, partia
`thermometer having the smallest practical subdivisions (11
`than 0.2“) is recommended. Suitable thermometers are 2W~
`the ASTM E-l series 37C through 41C, and 102C through. ‘
`Thermometers (21)). When placed in position, the stem 15, [ii
`the center of the neck and the top of the contraction chambfl
`if 37C or 38C is used) is level with the bottom ofthe outlet
`arm.
`
`Heat Source—A small Bunsen burner or an electrit‘:
`mantle capable of adjustment comparable to that P0551b
`Bunsen burner.
`
`Procedure—Assemble the apparatus, and place in the
`mL of the liquid to be tested, taking care not to allow any or
`to enter the side-arm. Insert the thermometer, shield the 61‘
`and flask assembly from external air currents, and “PP
`regulating it so that between 5 and 10 minutes elapse 1361:” ,
`drop of distillate falls from the condenser. Continue the ‘11 '
`a rate of 4 to 5 mL of distillate per minute, collecting me,‘
`the receiver. Note the temperature when the first drop‘0f,
`
`Par Pharm., Inc.
`Exhibit 1054
`Page 005
`
`

`

`wetting agents and/or solid carriers such as talc or colloidal silicas.
`
`Pharmaceutical aerosols are products that are packaged under
`pressure and contain therapeutically active ingredients that are
`released upon activation of an appropriate valve system They are
`intended for topical application to the skin as well as local application
`into thc nosc (nasal aerosols), mouth (lingual aerosols), or lungs
`(inhalation aerosols). These products may be fitted with valves
`enabling either continuous or metered—dose delivery; hence, the terms
`“lDRUtij Motored Topical Aerosols,” “[DRUG] Metered Nasal
`Aerosols,” etc.
`The term “aerosol” refers to the fine mist of spray that results from
`most
`t,
`surized systems. However,
`the term has been broadly
`mist pplied to all selficontainod pressurized products, some of which
`deliver foams or semisolid fluids. in thc case of Inhalation Aerosols,
`the particle size of the delivered medication must be carefully
`controllcd, and the average size of the particles should be under 5 um.
`These products ore also known as motored—dose inhalcrs (MDls).
`Other aerosol sprays may contain particles up to several hundred
`micrometers in diameter.
`The basic components of an aerosol system are the container, the
`propcllant,
`thc concentrate containing the active ingrodiends),
`the
`valve. and thc actuator: The nature oi’thesc components determines
`such characteristics as particle size distribution, uniformity of dose
`for motored valves. delivery ratc, wctncss and temperature of the
`spray, spray pattern and vclocity or plume geometry, foam density,
`and lhiid viscosity.
`
`
`
`
`
`
`
`AEROSOLS
`
`Types of Aerosols
`
`Acrosols constst ot‘two-phasc (gas and liquid) or throophasc (gas,
`liquid. and solid or liquid) systems. The two-phase aerosol consists of
`a
`solution of active ingredients in liquefied propellant and the
`vaporized propcllant. The solvent is composed ofthe propellant or a
`mixture ol the propellant and cosolvcnts such as alcohol, propylene
`glycol. and polyethylene glycols, which are ottcn used to enhance the
`solubility of the active ingredients.
`\
`'
`Tltt‘et:~ph21.sc systems consist of a suspension or emulsion of the
`active ingrediends)
`in addition to the vaporized propellants. A
`suspension consists of the active ingrediends) that may be dispersed
`in the propellant system with the aid of suitable excipients such as
`
`7
`
`_
`
`"of Enfornzutzon
`
`it fit) itiiarinaccuticai Bosage 33mins
`
`'t'nlatt’ l. lutisritational l‘iliinutis: Zones
`
`
`
`_l_3crivcd Data
`is
`RH
`45
`
`.
`
`inbar
`l
`l .2
`
`r npnn
`lnited Kingdom
`ortltern Europe
`anada
`ussia
`Unith States
`(«firm i (air on, MEWW
`United States
`r apart
`Southern Europe
`(Portugal—(J‘reec )
`or
`ran
`raq
`Sudan
`
`N.
`
`Jr:
`Ghana
`ndoncsia
`
`culzitcd as 1‘) L
`l‘)
`Data recorde: as
`cultith mean liinctic tonpcratrrc.
`al pressure oi"
`*zncr vapor.
`
`(l l5 i PHARMAf/‘iillrl‘ltlAll
`DOSAGE litflth/lg
`
`, Dosttge i‘ornis are provided it); most ot‘ the l’lintinacopcial drug
`‘
`.tances. out the processor; for the propuaiion ot‘niuny ot’tliein arc,
`i general, beyond the scopc oi
`the Pharniacopcia,
`in addition to
`tlefm is; the dosage forms, this section presents the general principles
`in lved in the inninititctnrc ot‘sonic ot‘tlteni. particularly on a small
`lc. Other
`inlbrination that
`is given bears on thc usc ot'
`the
`
`dosage forms.
`
`BREAK/Al FLARE iti'l‘Y
`
`tent to \vtiiclithc tlicrapcutic constituent of
`. availability, or tho r
`.pliarmaccntiml rlosagc t‘ortn ir- ‘ndcd for oral or topical use is
`Wailalilc {or absorption. is intlucnc
`by it variety ot‘t‘actors. Among
`herent factors known to ttftce absorption are the method ot‘
`titanutacturc or method ol‘compounding; tho particic size and crystal
`or polymorph oi the drug substance; and tho dilucnts and
`is used in tin-mutating the dos gc form,
`including tillers,
`disintegrating ancnts,
`lubricants, coatings, solvents, Sit,—
`llending agcnts, and dyes. Lub
`ants and coatings are foremost
`‘ong thcse. The maintenance 0: a demonstrably high degrcc of
`Iozivailability requires panicultt‘ nttcnti’in to all aspects of
`I'Qduction and quality control
`that may aticct
`tl
`nature ot‘ the
`fintsiiei. dosage lorin.
`
`'
`
`TERM iNQLGGY
`
`Occasionally it is necessary to add solvent to the contents oi‘
`‘Ontaii ct just prior to use usually bi“ nsc ot’ instaliility'of some
`mils in the diluted 'toriu, Thus, a solid d1 itcd to yield a suspension is
`filled [DRllGl/m- Sn.»‘[mn'lmz; a solid dissolved (and diluted to yield
`Solution is called [URL/G]
`[627' Solution; and a solution or
`‘ dispel
`ion diluted to yield a inorc dilutc form of the drug is called
`DRUG] ()rtrl ('oiir'e'nirutc. Attcr dilution,
`is important that lhc
`111;; be homogeneously dispersed before administration.
`
`Par Pharm., Inc.
`Exhibit 1054
`Page 006
`
`

`

`2702
`
`{l 35?} iliiarmacentical lineage forms ,3 General Iii/bromine!
`
`is an emulsion containing one or more active
`‘ol
`A foam :ier
`ingredients stirritCtants, aqueous or uonaqueous liquids, and the
`propellants. If the propellant is in the internal (discontirmous) phase
`(ie, of the oibirnwatcr type), a stable foam is discharged; and if the
`propellant is in the external (continuous) phase tie, ofthe water—in-
`oil type). a spray or u quiclcbrcaking foam
`.ischargcd.
`
`Manufacture
`
`Aerosols are usually prepared by one of i\‘\{1
`the “cold-till" process.
`the concentrate (gen
`temperature below 0 ‘) and the refrigerated props
`into open containers (usually chilled). The valve-act
`then crimped onto the container to form a pressure
`
`'
`
`Propeliants
`
`The propellant supplies the necessary pressure Within an aerosol
`system to expel material from the container and, in combination with
`other components, to convert the material into the desired physical
`form. Propellants may be broadly classified as
`liquefied or
`compressed gases having vapor pressures generally exceeding
`atmospheric pressure. Propellants within this dcl‘inition include
`various hydrocarbons, especially halogenated derivatives of methane,
`ethane, and propane, low molecular weight hydrocarbons such as the
`butancs and pentanes, and compressed gases such as carbon dioxide,
`nitrogen, and nitrous oxide. Mixtures of propellants are frequently
`used to obtain desirable pressure, delivery. and spray characteristics.
`A good propellant system should have the proper vapor pressure
`characteristics consistent with the other aerosol components.
`
`p ,
`
`.
`
`the concentrate
`In the “pressuredill” method,
`container. and either the propellant is forced under
`the valve orifice after the valve is sealed. or the i
`to flow under the valve cap and then the vr
`(“under—tliewcap” filling). in both cases oftlte “pr
`provision must be made for evacuation of air by in
`displacement with a small amount of propellant wpo
`process controls usually include monitoring of pror
`and propellant fill weight and pressure testing. teal;
`l
`function testing of the finished aerosol. Micr
`should also be controlled.
`
`Extractable Substances
`
`.
`
`tent
`
`Since pressurized inhalers and aerosols are no
`with organic solvents as the propellant or the vc
`extractables from the clastomcric and plzis‘ic rompt
`formulation is a potentially serious problem. Thus.
`.'
`and the quality of materials used in the nituiulhcure
`components (cg, stem, gaskets, housing, etc.) must
`selected and controlled. Their compatibility will
`components should be well established so as to
`of the valve components and to minimize changes in the medic
`delivery, leak rate. and impurity profile ol‘thc drug product ave
`The extractable profiles of a representative sample of each
`elastomeric and plastic components of the valve should
`est
`under specified conditions and should be correlated to the extra
`profile or the aged drug product or placebo, to ensure r“
`quality and purity of the drug product. Extractnh
`include polynuclear aromatics, nitrosamincs. vulcani
`
`Valves
`
`
`
`The primary function of the valve is to regulate the flow of the
`therapeutic agent and propellant
`from the container. “he spray
`characteristics of the aerosol are influenced by orifice timcnsion,
`number. and location. Most aerosol valves provide for continuous
`spray operation and are used on most topical products However,
`pharmaceutical products for oral or nasal
`inhalation of'en utilize
`metered~dose valves that must deliver a uniform quantity of Spray
`upon each valve activation, The accuracy and reproducibility of the
`doses delivered from metering valves are generally good, comparing
`favorably to the uniformity of solid dosage forms such as tablets and
`capsules, However; when aerosol packages are stored improperly, or
`when they have not been used for long periods of time, valves must
`be primed before use. Materials used For the manufacture of valves
`should be inert to the formulations used. Plastic, rubber. aluminum,
`and stainless steel valve components are commonly used. Victor‘ch
`dose valves must deliver an accurate dose within specified tolerances.
`
`Actuators
`
`An actuator is the fitting attached to an aerosol valve stem, which
`when depressed or moved, opens the valve, and directs the spray
`containing the drug preparation to the desired area. The actuator
`usually indi ‘ates the direction in which the preparation is dispensed
`and protects the hand or finger from the refrigerant effects of the
`propellant. Actuators incorporate an orifice that may vary widely in
`size and shape. The size ofthis orifice, the expansion chamber design,
`and the nature of the propellant and lbrmulation influence the
`delivered dose as well as the physical characteristics of the spray,
`foam, or stream of solid particles dispensed. For inhalation aerosols,
`an actuator capable of delivering the medication in the proper particle
`size range and with the appropriate spray pattern and plume geometry
`is utilized.
`
`Containers
`
`Aerosol containers usually are made ofglass, plastic, or metal. or a
`combination oi‘ these materials. Glass containers must be precisely
`engineered to provide the maximum in pressure safety and impact
`resistance. Plastics may be employed to coat glass containers for
`improved safety characteristics, or
`to coat metal containers to
`improve corrosion resistance and enhance stability of the formula-
`tion. Suitable metals include stainless steel, aluminum, and tin~plated
`steel. Extl‘actublcs or leachables (e.g..ydra\ving oils, cleaning agents,
`etc.) and particulates on the internal surfaces 0 l‘ containers should be
`controlled.
`‘
`
`dial
`
`Medicinal aerosols should contain at least the following ‘
`information on the label as
`in accordance with app
`regulations.
`I'VamingiAvoid inhaling. Avoid spraying into eyes or on ,
`mucous membranes
`.
`in
`NO'I‘E- The statement “Avoid inhaling" is not
`preparations specifically designed for use by inhalation. The I
`“or other mucous membranes" is not necessary for prepflfli
`specifically designed for use on mucous membranes.
`A,
`I'Varm'ng Contents under pressure. Do not puncun'e or in
`container. Do not expose to heat or store at temperatures above it
`(49” C). Keep out of reach of children.
`,
`.
`the label t
`i

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket