throbber
RAPID COMMUNICATIONS IN MASS SPECTROMETRY
`Rapid Commun Mass Spectrom 2006 20 21832189
`Published online in Wiley InterScience wwwintersciencewileycom
`
`DOI 101002rcm2567
`
`Measurement of paclitaxel and its metabolites in human
`plasma using liquid chromatographyion trap mass
`spectrometry with a sonic spray ionization interface
`
`Henrik Green Karin Vretenbrant Bjorn Norander and Curt Peterson
`Link6ping University SE581 85 Link6ping
`Department of Medicine and Care Faculty of Health Sciences
`
`Division of Clinical Pharmacology
`Sweden
`
`Received 1 March 2006 Revised 9 May 2006 Accepted
`
`11 May 2006
`
`A quantitative
`liquid chromatographyion trap mass spectrometry method for the simultaneous
`in human plasma has
`determination of paclitaxel 6ahydroxypaclitaxel and p3hydroxypaclitaxel
`been developed and validated 6ap3Dihydroxypaclitaxel was also quantified using paclitaxel as a
`reference and docetaxel as an internal standard The substances were extracted from 0500 mL plasma
`using solid phase extraction The elution was performed with acetonitrile and the samples were
`reconstituted in the mobile phase Isocratic highperformance liquid chromatography analysis was
`performed by injecting 50 p1 of reconstituted material onto a 100 X 300 mm C12 column with a
`methano11 trifluoroacetic acidammonium trifluoroacetate in H20 7030 mobile phase at 350 p11
`min The MPHJ+ ions generated in the sonic spray ionization interface were isolated and frag
`mented using two serial mass spectrometric methods one for paclitaxel transition 854 > 569
`551
`and the dihydroxymetabolite transition 886 > 585
`567 and one for the hydroxy metabolites
`551 and docetaxel N+Nar transition
`transition 870 > 585
`567 transition 870 > 569
`830 > 550 Calibration curves were created ranging between 05 and 7500 ngmL for paclitaxel 05
`and 750 ngmL for 6ahydroxypaclitaxel and 05 and 400 ngmL for p3hydroxypaclitaxel Adduct
`ion formation was noted and investigated during method development and controlled by mobile
`phase optimization In conclusion a sensitive method for simultaneous quantification of paclitaxel
`and its metabolites suitable for analysis in clinical studies was obtained Copyright
`2006 John
`Sons Ltd
`
`Wiley
`
`Paclitaxel Taxo11 is an anticancer drug with a broad
`spectrum of antitumor activity including breast ovarian
`It was originally isolated from the
`skin and lung cancer
`Pacific yew tree Taxus brevifolia2 and together with the drug
`taxanes The taxanes
`represents the antineoplastic
`docetaxel
`have a unique mechanism of action among chemotherapeu
`the drugs facilitate the formation and
`tic agents in that
`suppress the depolymerization of microtubules Exposed
`cells are blocked in the G2 M phase in the cell cycle and
`undergo
`apoptosis31
`eventually
`this may be a result of different
`apoptosis directly but
`mechanisms at different drug concentrations5
`in the liver by the
`is primarily metabolized
`P450 CYP enzymes CYP3A4 and CYP2C8
`cytochrome
`CYP3A4 metabolizes
`to p3hydroxypaclitaxel
`paclitaxel
`and CYP2C8 converts the drug into 6ahydroxypaclitaxel
`can be further oxidized to 6ap3
`and these metabolites
`dihydroxypaclitaxe16 Other metabolites have been identified
`
`Paclitaxel also induces
`
`Paclitaxel
`
`to H Green Engineering Biology Division of
`Correspondence
`and Care
`of Medicine
`Clinical Pharmacology Department
`Faculty of Health Sciences Linkoping University SE 581 85
`Linkoping Sweden
`Email henrikgreenimvliuse
`Contractgrant sponsor Swedish Cancer Society Gunnar Nils
`sons Cancer Foundation and the County Council
`in Ostergot
`land
`
`in smaller quantities Studies have shown that
`the major
`and its metabolites
`portion of unchanged
`paclitaxel
`excreted in the feces indicating extensive nonrenal clear
`ance8 However only paclitaxel has been shown to be toxic
`and to have an antitumor activity while its metabolites are
`considered to be inactive9
`
`are
`
`Since
`
`the 1990s when Taxol was
`the beginning of
`as an anticancer drug1° an extensive number
`introduced
`studies have been done on the substance
`of pharmacokinetic
`These studies have resulted in the development
`different methods
`for quantification of paclitaxel
`and its
`logical samples The concentrations
`of paclitaxel
`in human plasma are measured preferably by
`metabolites
`highperformance liquid chromatography HPLC or liquid
`chromatography mass spectrometry LCMS although
`
`of several
`
`in bio
`
`other
`
`techniques have also been reported Immunological
`methods
`such as enzyme linked immunosorbent assay
`ELISA12 and radioimmunoassay13
`have been used to
`screen human plasma but
`these methods have a major
`drawback of cross reactions Most frequently used are HPLC
`systems with UV detection although other chromatography
`eg capillary electrophoresis14 and methods
`techniques
`
`tWILEY
`interScience®
`
`MN
`
`Copyright 0 2006 John Wiley
`
`Sons Ltd
`
`Abraxis EX2034
`Actavis LLC v Abraxis Bioscience LLC
`1PR201701101 1PR201701103 1PR201701104
`
`

`

`2184 H Green et al
`
`using other detectors eg fluorescence detectors15 have also
`been reported
`
`Several developments
`
`in the use of paclitaxel have led to
`of the drug and the need for
`lower plasma concentrations
`more sensitive assays First of all clinicians
`have begun to
`investigate weekly therapy as an alternative
`to the every
`three week administration schedule with a reduction
`of
`dose as a result16 In the development of new chemother
`apeutic regimens where paclitaxel
`is used in combination
`with other drugs the dose is often less than the recom
`mended
`phase II dose In addition the investigation
`of
`has also led to the need
`individualization of chemotherapy
`The relative lack of sensitivity
`for more accurate techniques
`of the HPLC methods with UVdetection17
`and specificity
`of methods
`using mass
`has
`led to the development
`and multiple MS for detection of
`the sub
`Recent reports have described paclitaxel quanti
`based on LCMSMS instrumentation Several
`columns
`different extraction methods mobile phases
`interfaces and MS instruments have been used However
`most of
`these methods were developed
`for paclitaxel only
`and not for the metabolites For our purposes of trying to
`predict individual elimination of paclitaxel
`in cancer patients
`
`spectrometry
`
`stances1731
`
`fications
`
`factors affecting the extent of conversion by
`by investigating
`each metabolic enzyme we needed an accurate and sensitive
`method for quantification of paclitaxel and its metabolites in
`human plasma
`The aim of our study was to develop an LCMS method
`for simultaneous quantification of paclitaxel and its metab
`olites in human plasma To our knowledge this is the first
`method
`for quantification of paclitaxel 6ahydroxypacli
`and p36adihydroxypacli
`taxel p3hydroxypaclitaxel
`in human plasma using ion trap mass spectrometry
`ITMS with a sonic spray ionization SSI interface
`
`taxel
`
`EXPERIMENTAL
`
`Materials
`Paclitaxel C47H5iN014 was kindly supplied by BristolMyers
`CT USA p3 Hydroxypaclitaxel
`Squibb Wallingford
`C47H51N015 and 6ahydroxypaclitaxel C47H51N015 were
`purchased from Gentest Woburn MA USA through BD
`Biosciences Docetaxel C431153N014 was kindly supplied by
`Aventis Pharma Vitry Alforville France and used as an
`and
`internal standard HPLCgrade methanol acetonitrile
`hexane were from LabScan Analytical Sciences Dublin
`Ireland Ethanol was from Kemetyl Stockholm Sweden
`Water was obtained from a Milli Q Biocell station Millipore
`AB Stockholm Sweden Trifluoroacetic acid TFA was from
`Sigma Stockholm Sweden Human plasma was delivered
`from the blood bank of Linkoping University Hospital
`
`Instrumentation
`The LCMS system consisted of a LaChrome chromatograph
`pump model L7100 autosampler L7200 UV detector L
`7400 and an interface model D7000 with a Merck Hitachi
`M8000 LCMSn ion trap equipped with a SSI
`interface
`from Merck KgaA Darmstadt Germany The system was
`controlled using a computer equipped with the Chromatog
`raphy Data Station software LC3DQ MS System Manager
`
`Copyright © 2006 John Wiley
`
`Sons Ltd
`
`Instruments
`
`from Merck Darmstadt Germany and Hitachi
`Inc San Jose CA USA An infusion pump model 11
`Harvard Apparatus Holliston MA USA was connected
`to
`the system during some steps in the optimization The
`4 p Max RP 80A
`column
`was
`a Synergi
`100 x 300 mm Phenomenex CA USA and a guard column
`4 x 30 mm with the same stationary phase was used
`
`analytical
`
`coefficients
`
`Preparation of stock solutions and plasma
`samples
`Paclitaxel was dissolved in methanol
`to a concentration of
`20 mg mL and further diluted in methanol
`to give stock
`solutions in the range of 02200 tigmL using glass pipettes
`The internal standard docetaxel was dissolved in ethanol
`to a
`concentration of 20 mg mL and diluted to 20 lagmL p3
`Hydroxypaclitaxel and 6ahydroxypaclitaxel were dissolved
`and methanol respectively to give approxi
`in acetonitrile
`mate concentrations of 20 lagmL Due to the high inaccuracy
`the manufacturers quantity determination the concen
`of
`trations were determined using absorption
`spectroscopy
`The absorbance was measured at 227 nm for paclitaxel and
`230 nm for
`The following molar
`the other
`substances
`were used 298 mM1 cm 1 for
`extinction
`358 mM1 cm 1
`paclitaxel 148 m141 cm 1 for docetaxel
`and 262 mM1 cm 1 for 6a
`for p3hydroxypaclitaxel
`hydroxypaclitaxel according to the manufacturers docu
`mentation
`Standards were prepared by diluting the stock solutions in
`human drug free plasma Nine different standards denoted
`STD1STD9 were prepared with the following concen
`trations 05 2 5 10 50 250 1000 4000 and 7500 ngmL of
`paclitaxel 05 2 5 10 25 50 100 250 and 750 ngmL of 6a
`and 05 2 5 10 25 50 100 200 and
`hydroxypaclitaxel
`400 ngmL of p3hydroxypaclitaxel The standard concen
`trations were based on the concentrations found in human
`plasma after a dose of 175 mg m2 paclitaxel
`infused during 3 h
`Five different quality control QC samples denoted QC1
`QC5 were prepared with the following concentrations
`8
`100 300 1500 and 5000 ngmL of paclitaxel 8 2040 80 and
`400 ngmL of 6ahydroxypaclitaxel
`and 8 20 40 80 and
`300 ngmL of p3hydroxypaclitaxel
`
`to
`
`Solidphase extraction
`The solid phase extraction SPE procedure was adopted
`from Huizing et al32 with modifications Prior to extraction
`the tips in the vacuum box were washed in acetonitrile
`The CN solid phase 100 mg 1 mL
`minimize crossover
`Isolute 1ST UK columns were conditioned
`with 2 mL
`methanol and 2 mL 10 mM ammonium acetate buffer pH 50
`Then 500 µ1 of spiked plasma were mixed with an equal
`volume of 02 M ammonium acetate pH 50 and 50111 of
`internal standard 20 lagmL and applied to the column The
`columns were washed with 2 mL of 10 mM ammonium
`acetate and 1 mL of methanol 10 mM ammonium acetate
`2080 and finally with 1 mL of hexane Prior to and after the
`hexane wash the columns were dried for a minimum of
`2 min The samples were eluted in 2 mL of acetonitrile and
`evaporated under a stream of nitrogen at 35°C The analytes
`were reconstituted in 200 µ1 mobile phase and sonicated for
`5 min to ensure that
`the compounds dissolved completely
`
`Rapid Commun Mass Spectrom 2006 20 21832189
`DOT 101002rcm
`
`

`

`The samples were then placed in the autosampler and 50 pL
`were injected twice from each vial
`
`LCMS optimization and settings
`Several mobile phases were tested to yield a sensitive and
`stable chromatography and MS detection The mobile phase
`finally used for chromatographic
`separation was methanol
`1 TFA ammonium trifluoroacetate
`in H20 pH 77030 at a
`flow rate of 350 laL min and the chromatography
`ran for
`9 min
`The SSI settings were as follows plate temperature 200°C
`aperture 1 temperature 150°C aperture 2 temperature 120°C
`chamber voltage 06 kV drift voltage 35 V and focus voltage
`34V
`In the ion trap mass spectrometer positive ion polarity was
`used the multiplier voltage being set to 400 V and the peak
`threshold to 10 The automatic sensitivity control ASC was
`turned on and for each run three microscans were centroided
`and averaged The mass spectrometer was run in MS2 mode
`using two alternating methods The first MS2 method was
`used for detection of paclitaxel and dihydroxypaclitaxel a
`reference substance was not available for this metabolite so
`paclitaxel was used as reference In the ion accumulation
`step for this method the low mass cutoff LMC was set to 80
`the accumulation mass range was set to 750940 mz the
`voltage was set to 003 V and the sensitivity
`accumulation
`was 200 The settings for the isolation step were as follows
`and 881891 LMC 750
`isolation mass range 849859
`isolation voltage 015 V and isolation time 10 ms During
`the collision induced dissociation CID step the following
`and 876896
`settings were used CID resonance 849863
`LMC 420V CID time 50 ms and CID voltage 015V The
`second MS2 method was used for the detection of both
`and docetaxel The settings
`step were
`as
`follows
`the
`the ion accumulation
`for
`to 766940 mz the
`accumulation mass range was set
`voltage was set to 003 V and the sensitivity
`accumulation
`was 200 During the isolation step the following settings
`were used isolation mass range 825835 and 865875 LMC
`770 isolation voltage 012 V and isolation time 10 ms In the
`CID step LMC was set to 420 CID resonance to 825835 and
`865881 mz CID time to 50 ms and CID voltage to 02 V
`During both methods the MS spectra were scanned for a
`mass to charge ratio between 500 and 600 mz Chromato
`from the MS
`substance were extracted
`for each
`grams
`ions as indicated
`
`chromatograms
`Table 1
`
`using the product
`
`in
`
`hydroxypaclitaxel metabolites
`
`Validation procedure
`The precision and accuracy of the method were evaluated by
`analyzing paclitaxel p3hydroxypaclitaxel and 6a hydro
`
`Measurement of paclitaxel using LC SSIITMS
`
`2185
`
`STD1STD9
`
`xypaclitaxel at five different concentrations QC1QC5 on
`three different days Six QC samples at each concentration
`were used to evaluate within day variation and six were
`used for between day variation To determine the concen
`tration of the analytes in the QC samples one calibration
`of nine different concentrations
`curve consisting
`were analyzed each day Calibration was performed
`peak in the chromato
`using the area of the corresponding
`gram and the calibration curve was created using a cubic fit
`regression model forced through zero
`The precision of the assay was expressed as a coefficient of
`variation CV at each
`concentration
`the
`by calculating
`as a percentage of the mean calculated
`standard deviation
`concentrations while accuracy of the method was evaluated
`the mean calculated
`concentration
`as a
`by expressing
`the nominal concentration
`The percentage
`percentage of
`extraction
`recoveries
`of paclitaxel p3hydroxypaclitaxel
`for the QC samples were deter
`and 6ahydroxypaclitaxel
`mined by comparing extracted blank samples in which the
`amount of
`SPE eluate had been spiked with a corresponding
`
`checked
`
`substance to those obtained by injections of extracted spiked
`plasma samples The stability of paclitaxel p3hydroxypa
`clitaxel and 6ahydroxypaclitaxel
`in the samples was
`in three different experiments one for
`room
`temperature stability one for stability after
`reconstitution
`in the mobile phase and one experiment for freeze thaw
`stability The stability at room temperature was evaluated by
`two different
`letting eight samples of spiked plasma at
`
`concentrations paclitaxel 50 and 1000 ng mL p3hydro
`xypaclitaxel 50 and 200 ng mL and 6ahydroxypaclitaxel 50
`and 200 ng mL stand at room temperature for 24h and
`another set of eight samples for 48 h The concentrations were
`then compared with the QC samples with corresponding
`To check the stability after reconstitution ten
`concentrations
`QC samples were left on the bench after
`injection and re
`injected 4 days later The areas of
`the peaks were then
`compared to the first
`injection Eight plasma samples at the
`same two concentrations
`the room temperature
`stability test were frozen to 70°C and thawed for three
`cycles and then extracted and analyzed The areas of
`the
`peaks were compared to the QC samples spiked at the same
`concentrations
`
`as for
`
`RESULTS AND DISCUSSION
`Method development
`this assay was to
`The major goal
`in the development
`and its
`quantify low concentrations
`of both
`paclitaxel
`hydroxy metabolites in human plasma Due to the fact that
`we only had access to small amounts of the metabolites the
`and then the
`method was first
`for paclitaxel
`
`developed
`
`of
`
`ions and isolation mz values in MS and retention times Rt for the analytes
`Rt min
`Isolation mz
`
`Product
`
`ions
`
`Method
`
`1 2 2 1 2
`
`550+551+568+569
`
`550+551+568+569
`
`566+567+583586
`
`566+567+583586
`
`548551
`
`849859
`
`865875
`
`865875
`
`881891
`
`825835
`
`61
`36
`51
`
`3 7
`
`Rapid Commun Mass Spectrom 2006 20 21832189
`DOT 101002 rcm
`
`Table 1 Product
`
`Substance
`
`Paclitaxel
`
`p3Hydroxypaclitaxel
`
`6aHydroxypac1itaxe1
`
`6ap3Dihydroxypac1itaxe1
`Docetaxel
`
`Copyright 0 2006 John Wiley
`
`Sons Ltd
`
`

`

`2186 H Green et al
`
`settings were adopted to fit
`metabolites and thereafter
`
`the quantification of the available
`finetuned for the best perform
`
`paclitaxel
`
`either
`
`ance
`In most instruments and previously reported methods
`forms the M±H1+ ion or the JM±Nal
`molecular ion To suppress the formation of the Na+ adduct
`in favor of the M±H1 + ion an acidic mobile phase would be
`required However
`in our system we could not get a
`reproducible formation of the JM±H1+ ion between different
`batches of acidic mobile phase due to the formation of a
`lower or higher amount of JM±Nar To stabilize the system
`to form
`the mobile phase was changed to push paclitaxel
`other ions instead of JM±H1+ By adding positive or negative
`ions to the mobile phase the following adducts could be
`detected in the LCMS M±K1+ 2 mM KC1 Me0H 3070
`JM±Nal ± 2 mM NaC1Me0H 3070 and JM+Acetater
`20 mM ammonium acetatepH
`50 Me0H 3070 The
`highest sensitivity of these adducts was reached with the
`the M±H1+ ion showed a higher
`JM±K1+ ion However
`sensitivity although unstable than any of
`the other
`ions
`probably due to fragmentation of the latter in the ion source
`We then tried to stabilize the formation of the M±H1+ ion by
`suppressing the JM+Nal+ ion This was done using crown
`ethers or TFA the latter giving the highest sensitivity and the
`lowest amount of JM±Nal +adducts Although TFA caused a
`the signaltonoise SN
`suppression of the absolute signal
`ratio for paclitaxel was increased and a stable JM±H1+ ion
`formation was achieved
`The mobile phase was then
`and pH Optimum MS2
`optimized for TFA concentration
`conditions for the analyte response were established using a
`continuous
`infusion of a stock solution containing paclitaxel
`while running the chosen mobile phase components Note
`forms the JM±Nal + adduct using this mobile
`that docetaxel
`phase and instrument
`interfaces were
`Several
`to evaluate
`the
`response The M8000 LCMS ion trap was equipped with
`both an SSI
`interface as well as an electrospray
`ionization
`ESI interface The SSI interface was about
`four times more
`
`also tested
`
`efficient
`
`than the ESI one We also had
`in ionizing paclitaxel
`the ionization on a Perkin Elmer
`the opportunity to test
`SCIEX equipped with both an ESI and an atmospheric
`ionization APCI interface In this system
`pressure chemical
`interface was more efficient
`the ESI
`in ionizing paclitaxel
`than the APCI
`interface The SSI
`interface was
`then
`
`paclitaxel
`
`considered the best choice
`The HPLC column was chosen to retain the analytes
`sufficiently to avoid
`ion suppression
`by early eluting
`substances in the sample extract and separate the analytes
`since some of them have identical mass to charge ratios for
`have the
`some fragments The monohydroxy metabolites
`same molecular mass but they fragment into ions with a
`different mz However p3hydroxypaclitaxel dissociates
`into fragments with the same mz as
`from
`fragments
`and 6ahydroxypaclitaxel
`dissociates into frag
`ments with the same mz as dihydroxypaclitaxel Fig 1 To
`two methods were
`achieve single peak mass chromatograms
`run in parallel one for paclitaxel and dihydroxypaclitaxel
`and one for the monohydroxy metabolites Table 1 By
`the HPLC conditions
`making sure that
`yielded peak
`between the analytes was mini
`separation interference
`retention times were 605 min for paclitaxel
`mized Typical
`365 min for p3hydroxypaclitaxel 50 min for 6ahydro
`and 68 min for
`xypaclitaxel 33 min for dihydroxypaclitaxel
`docetaxel After each injection the chromatography was run
`9 min Figure
`2 shows
`chromatograms
`for
`for
`typical
`paclitaxel and its hydroxylated metabolites generated from
`the product ions formed in MS2 mz values shown in Table
`1 Docetaxel was used as an internal standard however it
`did not increase the precision or accuracy of the method and
`was therefore only used to control
`the SPE and autosampler
`Paclitaxel adheres strongly to most surfaces carryover
`effects and ghost peaks were major problems during early
`method development The problems were traced to the SPE
`procedure and solved by replacing the nylon tips in the SPE
`vacuum box with steel needles which were washed between
`
`all extractions
`
`Paclitaxel+1 854 mz
`OHPac+1 870 mz
`
`0
`
`Paclitaxel
`
`6aOHPac
`p30HPac
`
`R1
`H
`H
`OH
`
`R2
`H
`OH
`H
`
`Paclitaxel 569 raiz
`p30HPac+1 569 mz
`585 mz
`6aOHPac+1
`
`H20
`
`Paclitaxel 551 mz
`p30HPac+1 551 mz
`567 mz
`6aOHPac+1
`
`Figure 1 Molecular structure and fragmentation of paclitaxel and its mono
`hydroxy metabolites
`
`Copyright © 2006 John Wiley
`
`Sons Ltd
`
`Rapid Commun Mass Spectrom 2006 20 21832189
`DOT 101002rcm
`
`

`

`1111Mffil
`
`Bann 0
`
`a00110
`
`700176
`
`604104
`
`btUOtl
`
`4119111f
`
`31113011
`
`21113130
`
`111000
`
`ct
`
`2rg
`
``c
`
`eaa
`
`g
`
`3
`
`Tinle min
`
`1000008
`
`900000
`
`100000
`
`700000
`
`600000
`
`500000
`
`4001100
`
`300000
`
`200000
`
`10000C1
`
`Figure 2 Typical
`hydroxy metabolites
`
`of
`
`its
`
`receiving
`
`for
`
`and
`chromatograms
`for paclitaxel
`isolated from plasma 15 min post end
`infusion from a patient with ovarian cancer
`175 mgm2 of paclitaxel Taxo18 The chromatograms were
`ions formed in MS2
`generated using the sum of the product
`550+551+568+569 mz for paclitaxel 550+551+568+569
`mz for p3hydroxypaclitaxel
`and 566+567+583586 mz
`In the chromatogram the baseline
`for 6ahydroxypaclitaxel
`is elevated with 5000 AU and that
`for 6ahydroxypaclitaxel
`paclitaxel with 100 000 AU for
`illustrative purposes In this
`method the ion trap works with two serial methods one
`generating the paclitaxel and the dihydroxymetabolite chro
`matogram and one the hydroxyl metabolites and the internal
`the peak at 61 min in the p3hydro
`standard Note that
`xypaclitaxel chromatogram is from paclitaxel more than 100
`fold intensity in the correct method and that the peak at 7 min
`in the 6ahydroxypaclitaxel chromatogram is from the internal
`standard which has a 13C fragment with the same mz as the
`metabolite
`
`Measurement of paclitaxel using LCSSIITMS
`
`2187
`
`in the trap at high ion concentrations Others
`charge transfer
`have solved this nonlinearity by diluting the samples33
`However we found that in the range presented in this assay
`the use of nonlinear calibration curves worked just as well
`The calibration curves for the metabolites were linear over
`the whole calibration
`range Figure 3 shows a calibration
`curve for paclitaxel obtained during one of the validation
`runs
`The final method
`
`for several
`
`reasons First of all
`
`concentrations
`
`conditions were chosen
`low
`they allow the quantification
`of
`<10 ng mL of
`in human
`the analytes
`plasma Second they allow the determination of paclitaxel
`and its metabolites simultaneously The method presented
`here has some special features especially the use of TFA as a
`Naadduct
`suppressor and the SSI interface both of which
`of the assay
`improved the performance
`
`analyzing six replicates of
`
`of
`
`Precision and accuracy
`the method was evaluated
`The reproducibility
`by
`five different QC samples on
`three different days Extracted standard curves obtained in
`the range of 057500 ngmL showed excellent curve fitting
`with a coefficient of correlation greater than 099 The within
`day and between day variations are reported in Table 2 The
`method was found to be acceptably precise CV <14 and
`accurate range 83113 except for the within day variation
`The mean
`the lowest
`concentration
`of paclitaxel
`extraction recovery assessed at the same concentrations
`range 6374 p3
`the QC samples was a bit low paclitaxel
`hydroxypaclitaxel 6571 and 6ahydroxypaclitaxel 72
`85 The
`and its hydroxy
`of paclitaxel
`quantification
`metabolites was possible within the range of the standard
`curves The lower limit of quantification was set to 05 ngmL
`for paclitaxel within day precision <12 n = 4 and
`between day accuracy 103122 n = 3 and to 2 ngmL
`for the hydroxy metabolites withinday variation <20
`n = 4 and between day accuracy 107125 n = 3
`
`at
`
`as
`
`Each
`sample was injected twice due to spiking of
`the
`chromatograms We were not able to get rid of the spiking
`phenomenon by adjusting the MS conditions or cleaning of
`the mass spectrometer although improving the water quality
`in the mobile phase helped some The spiking was however
`infrequent and readily identified and if one of
`rather
`the
`contained interference the other
`chromatograms
`injection
`was used With more advanced software it would be rather
`easy just to exclude a single data point caused by spiking
`instead of having to run the samples twice
`Due to the wide range of paclitaxel concentrations found in
`plasma the calibration curve for paclitaxel was divided into
`a low 05250 ngmL and high range 2507500 ngmL
`calibration curve The calibration curve was also created
`regression model due to nonlinearity at high
`using a cubic fit
`The response is linear from 05 up
`paclitaxel concentrations
`to 1000 ngmL Proposed causes of nonlinearity at the high
`end of
`the calibration curve have been the formation of
`paclitaxel dimers and trimers at the interface33 as well as
`
`Copyright C 2006 John Wiley
`
`Sons Ltd
`
`30
`
`kr 20
`D
`
`10
`
`0
`
`0
`
`2000
`4000
`6000
`8000
`Paclitaxel concentration ngmL
`
`Figure 3 A nonlinear calibration curve for the higher
`range
`obtained during one of the vali
`of paclitaxel concentrations
`dation runs The concentration of paclitaxel
`in spiked standard
`samples is shown on the xaxis and the response of the mass
`spectrometer as the area of the peak is presented on the y
`axis A cubic fit
`regression model was used for calibration of
`The responses of the metab
`the paclitaxel concentrations
`olites were linear within the calibration ranges
`
`Rapid Commun Mass Spectrom 2006 20 21832189
`DOT 101002rcm
`
`

`

`2188 H Green et al
`
`Table 2 Precision and accuracy of the quantification of paclitaxel and its monohydroxy metabolites
`reconstitution in the mobile phase
`represents the stability of the substances after
`
`in plasma The last column
`
`Concentration
`
`Accuracy
`
`Precision
`
`Reinj
`
`Within day
`
`Between day
`
`Within day
`
`Between day
`
`4 days later
`
`CV
`
`Mean
`
`100
`103
`116
`106
`88
`127
`106
`131
`117
`86
`120
`108
`113
`115
`92
`
`13
`
`68471
`
`0
`10
`10
`
`41
`
`451
`
`251
`
`0
`11
`
`CV
`
`497472697997939
`
`Mean
`
`151b
`84
`87
`96
`92
`106
`108
`95
`86
`93
`113
`113
`105
`102
`103
`
`Mean
`
`93
`91
`100
`100
`87
`95
`97
`83
`84
`97
`101
`99
`90
`93
`92
`
`Substance
`
`Sample
`
`Paclitaxel
`
`6aOHpac
`
`p30Hpac
`
`QS1
`QS2
`Q53
`Q54
`QS5
`QS1
`QS2
`Q53
`Q54
`QS5
`
`QS1
`QS2
`0S3
`Q54
`QS5
`
`ngmL
`
`Plasma
`
`8
`
`100
`
`300
`
`1500
`
`5000
`
`8
`
`20
`
`40
`
`80
`
`400
`
`8
`
`20
`
`40
`
`80
`
`300
`
`a For each concentration six samples were tested for the within day variation for the between day variation two samples on three different days
`n = 6 were used for each concentration
`and the re injection was done on the within day variation samples
`b The within day accuracies on the following three days were 80 103 and 88
`
`Stability experiments
`The results of the stability tests on the samples are presented
`in Tables 2 and 3 Freeze thaw cycling did not affect
`the
`precision or accuracy of the assay The substances in plasma
`were also stable at room temperature for at least 48 h and
`in the mobile phase the substances were
`after reconstitution
`stable for at least 4 days
`
`Discussion
`Several mass spectrometry methods have been developed for
`in different matrixes1731
`of paclitaxel
`the quantification
`In
`the M+Hr ion has been used for MS
`most methods
`analysis which we found to have the highest sensitivity
`as M+NH41+242534
`other
`adducts
`such
`although
`N4 ±Nai +232534 M+M+Na +23 and M+Acetate126 have
`of paclitaxel and its
`been used In the early characterizations
`
`Table 3 The stability
`of paclitaxel p3hydroxypaclitaxel
`and 6ahydroxypaclitaxel during three cycles of freezethaw
`ing to 70°C and room temperature RT stability for 24 and
`48 h
`
`Concentration
`
`RT
`
`RT
`
`ngmL
`
`Plasma
`
`1000
`
`50
`
`200
`
`50
`
`200
`
`50
`
`Freezethawing
`
`24 h
`
`48 h
`
`Mean
`
`103
`117
`89
`117
`116
`115
`
`Mean
`
`Mean
`
`97
`110
`96
`112
`104
`105
`
`98
`116
`91
`114
`109
`101
`
`Substance
`
`Paclitaxel
`
`6aOHpac
`
`p30Hpac
`
`n = 4
`
`Copyright © 2006 John Wiley
`
`Sons Ltd
`
`metabolites fast atom bombardment was used for ionization
`the compounds Most modern methods use the ESI
`of
`interface although the APC12631 and now the SSI interface
`have been shown to be useful Some of the more impressive
`methods developed are those published by Schellen et al26
`which has an analysis time of only 80 s and by Parise et al 1 7
`which has a range of almost 4 decades Although many
`methods have been published for paclitaxel only a few MS
`methods have been developed for the quantification of the
`metabolites283133 Royer et al31 used a MS1 method with an
`APCI interface and Mortier et a28 and Alexander et al3 used
`MS2 methods
`on triple quadrupole instrument with ESI
`
`interfaces as compared to our use of ion trap methodology
`and a SSI
`a method for
`interface Mortier et al developed
`quantification in oral fluids and human plasma with a liquid
`liquid extraction LLE and gradient chromatography The
`calibration ranges in plasma were 4100 ngmL for paclitaxel
`and 2100 ngmL for the metabolites The method developed
`by Alexander et al is used for quantification of the substances
`in dog and human plasma LLE was used for sample
`preparation and 13C6 labeled paclitaxel was used as an
`internal standard The calibration ranges in plasma were 01
`100 ngmL for all compounds We used SPE instead of LLE
`to minimize ion suppression we also had difficulties with
`using LLE by
`the metabolites when
`of
`the recovery
`unwanted
`compounds hopefully giving us a more robust
`method and isocratic chromatography since the mobile
`phase was shown to be of such importance
`for stable ion
`formation Our assay also has a rather wide calibration range
`of 057500 ngmL for paclitaxel 05750 ngmL for 6a
`hydroxypaclitaxel and 05400 ngmL for p3hydroxypa
`clitaxel which were chosen according to the concentrations
`found in human plasma after a 3h infusion of 175 mg m2
`Taxol® To our knowledge the use of TFA as a stabilizer of
`
`Rapid Commun Mass Spectrom 2006 20 21832189
`DOT 101002rcm
`
`

`

`M+111+ formation and an increase in SN ratio has not been
`presented before and the effect is probably due to the Na+
`binding effect of TFA The responses of the metabolites were
`regression model
`linear but in the case of paclitaxel a cubic fit
`was used due to nonlinearity Others have also found this
`nonlinearity and solved the problem by either diluting the
`sample33 or using nonlinear regression28
`
`CONCLUSIONS
`
`reported chromatographic method quantify
`This is the first
`in human plasma using
`ing paclitaxel and its metabolites
`LC SSIITMS methodology Comprehensive method devel
`opment with optimization of the mobile phase to yield as
`stable and marked formation of the M+111+ ion as possible
`led to an assay with a high response The development
`of a
`highly sensitive method with a wide calibration range for
`simultaneous quantification of paclitaxel and its metabolites
`in human plasma was a prerequisite for this project The
`The advantages of
`validation was performed successfully
`this method
`the simultaneous
`are the high sensitivity
`of all compounds and the use of a relatively
`quantification
`low cost SSI
`ion trap instrument compared with a triple
`quadrupole mass spectrometer the drawback is the spiking
`of the chromatograms which may call for reanalysis of single
`samples This analytical method has been used to quantify
`and its metabolites
`in more than 200 plasma
`paclitaxel
`samples from patients with ovarian cancer and will be used
`for individualized
`in the future to evaluate the conditions
`
`chemotherapy with paclitaxel
`
`Acknowledgements
`The authors wish to thank Martin Josefsson PhD National
`of Forensic Chem
`Board of Forensic Medicine Department
`istry Linkoping Sweden
`for fruitful discussions and for
`allowing us use of the Perkin Elmer SCIEX for an evaluation
`of ESI and APCI interfaces This study was supported by the
`Swedish Cancer Society Gunnar Nilssons Cancer Founda
`tion and the County Council in Ostergotland The authors
`also wish to thank Isaac Austin for proofreading the text
`
`REFERENCES
`1 Rowinsky EK Cazenave LA Donehower RC J Natl Cancer
`Inst 1990 82 1247
`2 Wani MC Taylor HL Wall ME Coggon P McPhail AT J
`Am Chem Soc 1971 93 2325
`3 Schiff PB Fant J Horwitz SB Nature 1979 277 665
`
`Measurement of paclitaxel using LCSSIITMS
`
`2189
`
`4 Goldspiel BR Pharmacotherapy 1997 17 110S
`5 Torres K Horwitz SB Cancer Res 1998 58 3620
`6 Sonnichsen DS Liu Q Schuetz EG Schuetz JD Pappo A
`Relling MV J Pharmacol Exp Ther 1995 275 566
`7 Monsarrat B Alvinerie

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket