throbber
Assessment and optimisation of
`dip-coating procedure for the
`preparation of electroenzymic
`glucose transducers
`
`P.A. Rea*
`
`P. Rolfe
`
`P. J. Goddard
`
`Bioengineering Unit, Department of Paediatrics, John Radcliffe Hospital. University of Oxford,
`Oxford OX3 9DU, England
`
`Abstract-The details of construction and the performance characteristics of a dip (cid:173)
`coated electroenzyn:ic glucose transducer comprising an H 2 0 2 electrode coated with a
`layer of glucose ox1dase encapsulated in cellulose acetate and overlaid with a layer of
`polyurethane are presented. The steady-state current increases when the glucose
`oxidase and cellulose acetate concentrations of the dip-coating solutions are increased,
`but high cellulose acetate concentrations yield thick and mechanically unstable
`membranes. A compromise between current yield and mechanical stability can,
`however, be achieved by employing glucose oxidase and cellulose acetate concen (cid:173)
`, respectively. A polyurethane concentration
`trations of 200 mg m1- 1 and 2 · 5 g 100 m1- 1
`of 6 g 1 OOml- 1 is optimal both in terms of the current yield and the linearity of response.
`The relationship between steady-state current and D -glucose concentration is ' ap(cid:173)
`proximately linear over the concentration range 0·5 to 11 ·5 mM, and if correction is
`made for deviations from linearity at higher glucose concentrations, concentrations in
`excess of 20 mM can readily be quantified. The steady-state current is pH and
`temperature dependent, but the dependencies are relatively small in the physiological
`range. The mean rate of decrease of the glucose current during long-term operation of
`the optimised transducer is 0·83 per cent of the initial current per hour at 37°C. The
`hydrated electrodes perform satisfactorily after storage for more than two weeks at room
`temperature. The transducers have a mean response time [t90%]of 50s or less.
`
`Keywords-Dip-coated electroenzymic glucose transducer, In vitro performance,
`Optimisation
`
`Med. & Biol. Eng . & Comput., 1985, 23, 108- 115
`
`1 Introduction
`APPROXIMATELY 0·5 per cent of the population of the
`developed nations suffer from type I insulin-dependent dia(cid:173)
`betes mellitus (IDDM) (CARISTY et al., 1979). Rapid death
`from the disease is uncommon because of the availability of
`monocomponent insulins, but IDDM remains a major
`clinical problem because of the complications that attend the
`condition: blindness, gangrene, renal damage, neuropathy,
`impotence and micovascular disease. These complications
`have a reported incidence of 30-50 per cent among long(cid:173)
`standing sufferers (DECKERT et al., 1978). Animal experi(cid:173)
`ments and epidemiological studies indicate that the compli(cid:173)
`cations oflDDM are associated with abnormal excursions of
`blood glucose and metabolite concentrations as a result of
`the relative infrequency with which diabetics administer
`insulin to themselves during the conventional management
`of the disease (TCHOBROTSKY, 1978; ENGERMAN et al., 1977).
`Thus, in an attempt to regulate the rate of insulin infusion in
`a more controlled and continuous manner, and thereby
`abrogate the long-term complications of IDDM, open-loop
`
`*Present address and address for correspondence: Cell Physiology
`Laboratory, Department of Biology, McGill University, 1205 Avenue
`Docteur Penfield, Montreal, PO H3A 181, Canada
`
`First received 11th November 1983 and in final form 21st May 1984
`© IFMBE: 1985
`
`continuous subcutaneous insulin infusion pumps (CSII)
`have been developed (ALBISSER et al., 1974a; b).
`Precise control of blood glucose concentrations necessi(cid:173)
`tates a closed-loop device which delivers insulin to the
`systemic blood supply in proportion to the difference
`between the desired and measured physiological glucose
`concentration (SoELDNER, 1982). There is therefore a need for
`in vivo glucose transducers for clinical applications.
`A large number of glucose transducers based on the
`'enzyme electrode principle', in which a glucose oxidase
`membrane and an electrochemical detector are combined,
`have been described (SoELDNER, 1982). The detection meth(cid:173)
`ods include potentiometric (MALMSTADT and PARDUE, 1965)
`and amperometric (GUILBAULT, 1980) measurements of the
`H 20 2 and H + ions generated in the glucose oxidase(cid:173)
`mediated oxidation of D-giucose, direct (UPDIKE and HICKS,
`1967) or differential (BESSMAN and ScHULTZ, 1973) measure(cid:173)
`ments of the 0 2 consumed and amperometric measurements
`of the redox state of the flavin moiety of glucose oxidase via
`an intermediary redox couple (ScttLAPFER et al., 1974). The
`enzyme immobilisation methods examined include entrap(cid:173)
`ment of the enzyme solution with a perm-selective mem(cid:173)
`brane (GUILBAULT and LUBRANO, 1973), gel (HICKS and
`UPDIKE, 1966) or microporous matrix (Sttu and WILSON,
`1979), coreticulation (BAUMAN et al., 1965), adsorption
`(SILVER, 1976) or covalent coupling
`to a membrane
`
`108
`
`Medicel & Biological Engineering & Computing
`
`March1985
`
`Dexcom Inc. v. WaveForm Technologies, Inc.
`IPR2017-01051
`Exhibit 1037
`
`

`

`(THEVENOT et al., 1982). However, our experience of these
`techniques is that only a few can readily be adapted to the
`mass fabrication of small transducers: two conditions that
`must be met before in vivo glucose monitoring can become
`widespread. Of the procedures we have investigated, the
`encapsulation of glucose oxidase in a cellulose acetate matrix
`has proved to be one of the most suitable for transducer
`miniaturisation and bulk production. Independently de(cid:173)
`veloped by SmcHIRI et al. (1982; 1983), this method involves
`the dip-coating of an H 20 2 electrode with a glucose
`oxidase/cellulose acetate mixture and polyurethane.
`SmcHIRI et al. (1982), however, provide only brief details of
`their method of transducer fabrication and little indication of
`the influence of the fabrication method on the response
`characteristics of the transducer. Thus in this report we
`describe the details of construction of such a dip-coated
`glucose transducer and summarise the performance charac(cid:173)
`teristics of the optimised system.
`Three features of the dip-coat are examined:
`
`(a) the influence of the composition of the dip-coat on the
`glucose response of the transducer in terms of its current
`yield and linearity of response
`(b) the pH and temperature dependencies of the steady(cid:173)
`state current
`(c) the stability of the transducer during Jong-term storage
`and operation.
`
`2 Materials and methods
`2.1 Electrodes
`The glucose transducer is a conventional glucose oxidase(cid:173)
`coupled electrode in which D-glucose is measured as the
`steady-state rate of formation of H 20 2 upon the glucose
`oxidase-mediated
`oxidation
`of
`D-glucose
`to
`D-gluconolactone.
`
`b-D-glucose+0 2 -----> D-gluconolactone.
`
`. (1)
`
`The rate of formation of H 20 2 is measured as the steady(cid:173)
`state current resultant on its oxidation at a platinum anode.
`
`. (2)
`
`platinum anode
`
`,..... ...... ~-../epoxy resin
`--------- Ag/AgCI cathode
`
`)
`
`polyurethane
`dip- coat
`
`glucose
`oxidase I cellulose
`acetate dip-coat
`
`O·l25mm
`
`5 mm
`
`The experimental electrodes consisted of a 0· 125 mm dia(cid:173)
`meter platinum anode (purity 99·9 per cent, from Good(cid:173)
`fellow Metals Ltd., Cambridge, England) and a 5 mm
`diameter Ag/ AgCI reference electrode of hemispherical
`geometry set into PVC (Fig. 1). The working and reference
`electrodes were insulated with epoxy resin. PVC was chosen
`as the material for the body of the electrode because it bonds
`tightly with polyurethane.
`
`2. 2 Electrode pretreatment
`The uncoated platinum working electrode was pretreated
`before each experiment by applyigg a potential of - 200 m V
`(w.r.t. Ag/AgCl) until the cathodic current had decayed to a
`minimum. The electrode was then polarised at + 50 m V
`(w.r.t. Ag/AgCI) and the current allowed to decay to about
`0 nA. This pretreatment procedure, recommended by GUIL(cid:173)
`BAULT and LUBRANO (1973), was found to yield reproducible
`H 20 2 currents from the onset of measurement.
`The dip-coated electrodes were pretreated before each
`experiment by applying a potential of +600mV (w.r.t.
`Ag/AgCl); only when the current had decayed to a minimum
`were measurements made. The electrodes were rinsed with
`buffer at the end of each experiment to remove unreacted
`glucose and reaction products. Unless stated to the contrary,
`the freshly coated transducers were hydrated in buffer for
`12 h before any measurements were made.
`
`2. 3 Dip-coating
`The glucose transducers were prepared in a manner
`similar to that described by SmcHIRI et al. (1982). The
`electrode tip was lightly polished with 0·25 µm diamond
`paste (Engis Ltd., Maidstone, England) and dipped into a
`suspension of glucose oxidase (18300U/g; type II from A.
`niger, EC 1.1.3.4 from Sigma Chemical Co. Ltd., Poole,
`England) prepared in a solution of cellulose acetate (acetic
`acid content 53·5-54·5 per cent from BDH, Poole, England).
`The lyophilisate of glucose oxidase was milled to a fine
`powder with a mortar and pestle before its suspension in the
`cellulose acetate solution to improve the evenness and
`reproducibility of film formation. The cellulose acetate
`solutions were made up in 100 per cent acetone or a 1 : 1
`mixture of acetone and ethanol. Immediately after their
`withdrawal from the dip-coating solution, the electrodes
`were inverted and allowed to dry for 45 min at room
`temperature. The polyurethane coat was applied. by dipping
`the glucose oxidase/cellulose acetate-coated electrode into a
`solution of polyurethane (Estane 5701 Fl) dissolved in a
`l: 18 (v/v) mixture of dimethylformamide and tetrahydro(cid:173)
`furan. The electrodes were finally left to dry in air at room
`temperature for 12-18 h before hydration.
`For most of the studies described below, the glucose
`oxidase/cellulose acetate dip-coating solution consisted of
`200 mg ml - 1 glucose oxidase and 2·5 g 100 ml - l cellulose
`acetate suspended in a 1 : 1 mixture of acetone and ethanol.
`The polyurethane dip-coating solution consisted of
`6 g lOOm1- 1 polyurethane dissolved in a 1 : 18 (v/v) mixture
`of dimethylformamide and tetrahydrofuran. These com(cid:173)
`positions were found to be optimal.
`
`2 .4 Note on stirring dependence
`The glucose concentration dependence of the steady-state
`current was measured in unstirred media. In the absence of
`diffusion limitation, the concentration dependence of the
`steady-state current should approximate
`the
`intrinsic
`kinetics of the immobilised enzyme such that
`
`Fig. 1 Schematic diagram of experimental dip-coated glucose
`transducer
`
`iss = (imax[D-glucose])/(Km + [D-glucose])
`
`Medical & Biological Engineering & Computing
`
`March 1985
`
`. (3)
`
`109
`
`Dexcom Inc. v. WaveForm Technologies, Inc.
`IPR2017-01051
`Exhibit 1037
`
`

`

`not lower, in the unstirred condition. Indeed, the results of
`rotating disk-electrode experiments where
`the hydro(cid:173)
`dynamics of the bulk medium can be precisely defined
`demonstrate that H 20 2 effiux from the surface of the
`transducer is a more important determinant of the steady(cid:173)
`state current at high rotation speeds*. Thus, to evaluate the
`influences of diffusion limitation and catalytic limitation it is
`necessary to measure the rates of both glucose and H 20 2
`mass transport to and from the transducer surface over a
`wide range of stirring speeds; a procedure which would be
`impracticable for a large number of routine tests. Con(cid:173)
`sequently, for reasons ofreproducibility and practicality, the
`measurements of the glucose current were made in unstirred
`media.
`
`2. 5 Determination of D-glucose concentration dependence
`All the measurements were made in Miller and Golder's
`constant ionic strength buffer. The buffer consisted of
`80 mM NaCl, 5·68 mM Na2 HP0 4 and 3·2 mM NaH 2 P04 .
`Unless stated otherwise, the final pH of the buffer was 7·44
`units and its ionic strength was 0·11; this approximates the
`pH and ionic strength of blood plasma (DAWSON et al., 1959).
`Aliquots of a stock solution of 2 M D-glucose were added
`to the buffer and the medium stirred to ensure homogeneity.
`The steady-state current was then measured 3-5 min after
`the cessation of stirring. The measurements were made in a
`thermojacketed reaction cell and the D-glucose solutions
`were prepared at least 18 h before their use to ensure an
`equilibrium mixture of the a- and b-anomers. All the
`D-glucose concentrations are expressed as total D-glucose
`(a- plus b-anomers). An electrode potential of +600mV
`(w.r.t. Ag(AgCI) was employed throughout.
`
`5
`
`4
`
`""" c
`
`11 mM
`D - glucose
`
`5 mM
`D - glucose
`
`where iss = steady-state current at glucose concentration
`[D-glucose], imax = maximum steady-state current and
`Km = glucose concentration at which iss = imaxf2.
`Multiplying both sides of expr. 3 by issimax and rearranging
`
`(EADIE, 1942; HoFSTEE, 1952)
`
`. (4)
`
`A plot of iss against iss/[D-glucose] should therefore yield a
`straight line of slope - Km and intercept
`imax· Since
`imax = k+ 2 e 0 , where e0 =enzyme activity, and Km is a
`constant, (k _ 1+k+ 2 )/k + 1 , determination of these two para(cid:173)
`meters from the Eadie-Hofstee transformation ( expr. 4)
`should enable the definition of the performance charac(cid:173)
`teristics of the transducer in terms of its maximum glucose
`current imax and the glucose concentration range over which
`linearity applies Km. However, direct measurements of the
`glucose current show that rein. 4 does not hold for dip-coated
`transducers (Fig. 2). Neither in the stirred nor the unstirred
`condition does a plot of iss against iss/[D-glucose] yield a
`straight-line relationship (Fig. 2b ). Although the deviations
`from linearity are greatest in the unstirred condition, inspec(cid:173)
`tion of the arithmetic plots (Fig. 2a) shows that diffusion
`limitation with respect to the mass transport of glucose to the
`transducer surface is not limiting on the measured steady(cid:173)
`state current. The measured steady-state currents are higher,
`
`a Arithmetic Plot
`
`Unstirred
`,,,,,,---0--()--0
`.... ~
`,...
`,...'1
`
`Stirred
`
`3·0
`
`2. 5
`
`2·0
`
`1·5
`
`"""
`c
`
`~
`
`~
`
`l · 0
`
`0·5
`
`/ A
`
`/
`o/
`7
`
`/
`
`/
`
`/
`9/
`I
`I
`I
`
`I t
`
`I
`I
`I
`I
`I
`I
`I
`p
`I
`I
`I
`q
`I
`I •
`I
`I
`l•
`
`b
`
`<!
`c
`in
`Ill
`
`Eadie-Hofstee Plot
`<><>a.
`"-._Unstirred
`"'
`"
`'
`
`II.
`\
`Stirred\
`\
`\
`o\
`~
`
`•
`
`O·O
`
`0 •
`O·l
`0·4
`0·3
`0·2
`i ss ID- glucose, nA mM-1
`
`0
`
`4
`
`6
`
`14
`12
`10
`D- glucose, mM
`
`16
`
`18
`
`20
`
`22
`
`0
`
`50
`
`200
`100
`150
`Glucose oxidase,mg ml-I
`
`Fig. 2
`
`Influence of stirring the bulk medium on the relationship
`between steady-state current iss and D-glucose concent(cid:173)
`ration. The dip-coat was prepared from dip-coating solutions
`containing 100 mg ml- 1 glucose oxidase suspended in
`2·5 g 100 m1- 1 cellulose acetate dissolved in a 1 : 1 mixture of
`acetone and ethanol. A polyurethane concentration of
`2 g 100 m1- 1 was employed. The measurements were made at
`20°C in Miller & Golder's buffer (pH 7·44; o-11). The buffer
`was stirred with a magnetic stirrer set at maximum velocity.
`(a) arithmetic plot (iss against [D-glucose], (b) Eadie(cid:173)
`Hofstee plot (iss against issf[D-glucose])
`
`Fig. 3 Effect of the glucose oxidase concentration of the glucose
`oxidase/cellulose acetate dip-coating solution on the steady(cid:173)
`state current at D-glucose concentrations of l, 5 and 11 mM.
`The dip-coats were prepared from glucose oxidase sus(cid:173)
`pended in 2·5g100 ml- 1 cellulose acetate dissolved in a 1 : l
`mixture of acetone and ethanol. Polyurethane was employed
`at a concentration of 2g lOOm1- 1
`. Conditions were 37°C,
`pH7·44, 0·1 I. Mean ±SE (n = 5) are shown
`
`* Unpublished observations by Rea
`
`110
`
`Medical & Biological Engineering & Computing
`
`March 1985
`
`Dexcom Inc. v. WaveForm Technologies, Inc.
`IPR2017-01051
`Exhibit 1037
`
`

`

`The pH and temperature dependencies of the glucose
`current were assessed in a similar manner.
`
`2.6 Stability to long-term storage and long-term operation
`These were determined as described in the legends of the
`corresponding figures.
`
`3 Results
`3. l Optimisation of composition of dip-coating solutions
`The effects of the glucose oxidase, cellulose acetate and
`polyurethane concentrations of the dip-coating solutions on
`the current yield of the final transducer are summarised in
`Figs. 3, 4 and 5. The influences of each of the three dip-coat
`components were measured at D-glucose concentrations of
`1, 5 and 11 mM to determine the concentration dependence
`as well as the overall current yield of the transducer.
`An increase in the glucose oxidase concentration of the
`2·5 g 100 ml - 1 cellulose acetate dip-coating solution from 50
`to 200mgm1- 1 causes a 7-10-fold increase in the steady(cid:173)
`state current at D-glucose concentrations of 5 and 11 mM
`(Fig. 3). The effects of glucose oxidase concentrations in
`excess of 100 mg ml - 1 are less marked at lower D-glucose
`concentrations; this is assumed to result from diffusion
`limitation with respect to the glucose current at low bulk
`D-glucose concentrations.
`
`11 mM
`D- glucose
`
`D- glucose
`
`5
`
`I
`I
`I
`I
`
`L
`
`/ D- qlucose
`
`/
`
`/
`
`/
`
`/
`
`/
`
`lmM
`D- glucose
`
`0
`
`Fig. 4
`
`4
`Cellulose acetate, g 100 ml-1
`
`Effect of cellulose acetate concentration on steady-state
`current at D-glucose concentrations of 1, 5 and 11 mM.
`Glucose oxidase and polyurethane concentrations of
`200 mg ml- 1 and 2 g 100 ml- 1, respective[ y, were employed
`throughout. Conditions were otherwise as for Fig. 3. Dissol(cid:173)
`ution of cellulose acetate in 1 : 1 mixture of acetone and
`ethanol(---); dissolution of cellulose acetate in 100 per
`cent acetone (---)
`
`The steady-state current increases with cellulose acetate
`concentration, and this is more marked at the higher glucose
`concentrations and when the cellulose acetate dip-coating
`solutions are prepared in 100 per cent acetone rather than a
`1 : 1 mixture of acetone and ethanol (Fig. 4). The more
`positively eccentric relationship between the steady-state
`current and cellulose acetate concentration when the latter is
`dissolved in 100 per cent acetone rather than a 1 : 1 mixture
`of acetone and ethanol was considered to make reproduc(cid:173)
`ibility between transducers more difficult to achieve. Con(cid:173)
`sequently, a 1 : l mixture of acetone and ethanol was
`employed as the solvent.
`
`<(
`c
`.;, 3
`
`~
`
`10
`
`12
`
`14
`
`16
`
`18
`
`Polyurethane, g/lOOml
`
`Fig. 5 Effect of polyurethane concentration of dip-coating solution
`on steady-state current at D-glucose concentrations of 1, 5
`and 11 mM. Glucose oxidase and cellulose acetate concen(cid:173)
`trations of200mgml- 1 and 2·5g lOOml- 1 were employed
`throughout. Conditions were otherwise as for Fig. 3
`
`An important factor when constructing transducers is
`their mechanical stability. Thus, although high concen(cid:173)
`trations of glucose oxidase and cellulose acetate give higher
`current yields, they also yield thicker films which have a
`spongy appearance upon hydration. As these were con(cid:173)
`sidered to be less stable mechanically than thinner films and
`unnecessarily increase the response time, a compromise
`between current yield and mechanical stability was struck by
`employing a glucose oxidase concentration of 200 mg ml - 1
`and a cellulose acetate concentration of 2·5 g 100 ml - i.
`The concentration of polyurethane employed for the
`second dip-coating solution is crucial (Fig. 5). Concen(cid:173)
`trations of polyurethane in excess of 6 g 100 ml - 1 result in a
`severe attenuation of current yield, whereas concentrations
`of less than 6 g 100 ml - l result in large deviations from
`linearity at the higher D-glucose concentrations. A poly(cid:173)
`urethane concentration of 6 g 100 ml - l was
`therefore
`employed.
`
`3. 2 Glucose concentration dependence
`The relationship between steady-state current and
`D-glucose concentration for a glucose transducer prepared
`from dip-coating solutions containing 200 mg ml - i glucose
`oxidase, 2·5 g 100 m1- 1 cellulose acetate and 6 g 100 m1- 1
`polyurethane is shown in Fig. 6. D-glucose concentrations of
`0·5 mM to in excess of20 mM can readily be determined with
`this transducer. If it is assumed that the relationship between
`steady-state current and D-glucose concentration is ap(cid:173)
`proximately linear over the range 0-11 ·5 mM, a slope of
`0·335±0·083nAmM- 1 (r = 0·995) is obtained. Since a
`blood glucose concentration of 8-11 mM is very suggestive
`of diabetes and a concentration of greater than 11 mM is
`
`Medical & Biological Engineering & Computing
`
`March 1985
`
`111
`
`Dexcom Inc. v. WaveForm Technologies, Inc.
`IPR2017-01051
`Exhibit 1037
`
`

`

`almost diagnostic (SINCLAIR, 1979), the glucose concen-
`tration range over which approximate linearity applies
`appears to be sufficiently wide to include both normal and
`pathological glucose concentrations.
`
`5
`
`4
`
`3
`
`2
`
`<(
`c:
`
`~
`~
`
`6
`
`5
`
`4
`
`3
`
`<(
`c:
`;:,,
`
`~
`
`II· 5 mM
`·o - glucose
`
`5 · 5 mM
`D - glucose
`
`2·5 mM
`D - glucose
`
`I
`
`I
`
`! -f I· 0 mM
`
`D - glucose
`
`7. 0
`
`7. 5
`
`5·5
`
`6·0
`
`6·5
`pH
`Fig. 7 pH dependence of steady-state current at D-glucose concen(cid:173)
`trations of 1, 2·5, 5·5 and 11·5 mM. Miller & Golder's buffer
`adjusted to the required pH with NaH 2 P0 4 or Na 2 HP0 4
`was employed throughout. Conditions were otherwise as for
`Fig.6. Mean ±SE (n = 5) are shown
`
`0 I 2 3 4 5 6 7 8 9 10 11 I2 13 I4 15 I6 I7 I8 19 20 21 22
`D-glucose, mM
`
`Fig. 6 Relationship between steady-state current and D-glucose
`concentration for transducer prepared from 200 mg ml- 1
`glucose oxidase, 2·5g100ml- 1 cellulose acetate and
`6g lOOm1- 1 polyurethane. - - - line drawn through data
`points; -
`- -
`line computed by linear regression (least(cid:173)
`squares method) for D-glucose concentrations of 0·5
`11 ·5 mM.
`The
`computed
`regression
`is
`to
`y = (0·335 ±0·083)x +0·338 nA where y = i,, (nA) and
`x = D-glucose concentration (mM). r = 0·995
`
`3. 3 pH and temperature dependence
`The steady-state current of the transducer is maximal in
`the pH range 6·5-7·0 units and decreases when the pH is
`increased above or decreased below this range (Fig. 7). The
`mean ratios of the steady-state currents obtained at pH 6-97
`w.r.t. 7·33 and pH 6-97 w.r.t: 5·87 are 1·1 and 1 ·0, respectively.
`The temperature dependence of the steady-state current is
`dependent on the D-glucose concentration (Fig. 8). The
`approximate mean temperature coefficients are 0·017, 0·017,
`0·009, 0·075, and 0·098 nA 0 c- 1 over the temperature range
`32-44°C for D-glucose concentrations of 1 ·0, 2·5, 5·5, 11-5
`and 19·5 mM, respectively.
`
`3 .4 Response time
`The time required for the current to attain 90 per cent of its
`final value upon transfer of the transducer from buffer
`lacking glucose to buffer containing 5 mM D-glucose is
`48·8 ± 7·38 s (n = 5). A typical response curve is shown in
`Fig.9.
`
`3. 5 Stability of glucose current
`The dip-coated transducers are relatively stable to storage
`in buffer at 21°C. The glucose current decreases from 100 to
`70 per cent of the initial value during the first 3 days of
`storage in buffer but from there on decreases at only
`approximately l per cent per day (Fig. 10). As the percentage
`loss of the initial current is similar for all three of the
`
`19·5 mM
`D - glucose
`
`11· 5 mM
`D - glucose
`
`l55mM
`D - glucose
`
`~2·5mM
`11 OmM
`
`- D - glucose
`
`D - glucose
`
`4
`
`<(
`c:
`
`~
`
`~
`
`t
`t
`r !
`!
`
`f-
`
`30
`
`32
`
`34
`
`40
`38
`36
`Temperature, °C
`Fig. 8 Temperature dependence of steady-state current at
`D-glucose concentrations ofl·O, 2·5, 5·5, 11·5and19·5 mM.
`Conditions were otherwise as for Fig. 6. Mean ±SE (n = 5)
`are shown
`
`42
`
`44
`
`46
`
`112
`
`Medical & Biological Engineering & Computing
`
`March 1985
`
`Dexcom Inc. v. WaveForm Technologies, Inc.
`IPR2017-01051
`Exhibit 1037
`
`

`

`D-glucose concentrations tested ( 1, 5 and 11 mM), single(cid:173)
`point calibrations may be sufficient to correct for the
`atterluation of current yield during storage. Continuous
`
`<l'.
`c:
`
`<l'.
`c::
`"."'
`
`0
`
`40
`
`80
`
`120
`
`160
`Ti me, s
`
`200
`
`240
`
`280
`
`320
`
`Fig.9
`
`Time dependence of the response of the transducer. The
`transducer was tran~ferred from buffer lacking glucose to
`buffer containing 5 mM D-glucose at the time indicated.
`Conditions were otherwise as
`for Fig. 6.
`t9o% =
`48·8 ± 7·38 s (n = 5)
`
`operation of the transducer at 37°C in buffer containing
`5 mM D-glucose results in a 55 per cent decrease in the
`measure current after 50 h (Fig. 11). 17 per cent of the initial
`current is lost during the first 5h of operation, but from this
`time onwards the drift is approximately linear with a mean
`value of0·83 per cent of the initial current per h (r = 0·891).
`
`4 Conclusions and discussion
`The development of a dip-coating procedure employing
`polyurethane for the preparation of electroenzymic trans(cid:173)
`ducers was considered to be of high priority for several
`reasons.
`
`(a) The application of dip-coats rather than preformed
`membranes generally yields a
`smoother surface
`geometry. Since surface irregularities encourage throm(cid:173)
`bus formation and foreign-body responses (WooDWARD,
`1982), dip-coated surfaces might diminish the incidence
`of thrombus formation during intravascular use and
`inflammation during subcutaneous use.
`(b) Dip-coating procedures combine high mechanical
`stability with comparative simplicity of membrane ap-
`
`c::
`"' L
`
`L
`:J
`u
`
`-ro
`·-c:
`
`Ii'-
`
`r·, ::o·,
`
`I
`
`Ill
`
`100
`
`90
`
`80
`
`70
`
`60
`
`50
`
`40
`
`30
`
`20
`
`10
`
`r------------- ~ ·---
`
`=.
`
`·---·----·---
`
`·--·-- ..
`
`.. I
`
`0
`
`2
`
`3
`
`4
`
`5
`6
`7
`Time. days
`
`8
`
`9
`
`10
`
`11
`
`12
`
`13
`
`Fig. IO Stability of hydrated dip-coated trans(cid:173)
`ducers to long-term storage at 21°C.
`The transducers were stored in Miller &
`Golder's buffer (pH 7-44; 0·1 I) and at
`the times indicated were tested to de(cid:173)
`termine their response to D-glucose at
`concentrations of l, 5 and 11 mM. Each
`bar is the mean ±SE (n = 5) of the
`initial current. The smooth line(---)
`is drawn through the mean percentage of
`the initial current calculated from the
`sum of the percentage initial currents for
`each of the three D-glucose concent(cid:173)
`rations. Conditions were otherwise as for
`Fig.6
`
`~ 1 mM D-glucose
`D 5 mM D-glucose
`ml 11 mM D-glucose
`
`100
`90
`
`-:= .. 80
`
`=>
`u
`
`m
`
`70
`60
`50
`40
`c: ·-
`30
`~ 20
`10
`
`(c)
`
`(d)
`
`plication and thereby facilitate the fabrication of small
`electrodes.
`Of the polymers that are available commercially, the
`polyurethanes appear to be among the most suitable for
`in vivo use. They are mechanically stable, elicit a
`comparatively small tissue reaction and are elastomeric
`and consequently do not leach plasticisers.
`The polyurethanes are very resistant to g-irradiation
`(Irradiation Products Ltd., Swindon, England); a
`common procedure for the sterilisation of transducers
`for in vivo applications. (Providing that the dip-coated
`transducers are dry when irradiated, there is only a 20-
`40 per cent loss of enzyme activity after sterilisationt. As
`the response characteristics of sterilised transducers are
`similar to unsterilised ones, irradiation may therefore be
`an appropriate method of sterilisation.)
`
`The results of the in vitro tests of the dip-coated glucose
`trransducers described above are therefore encouraging.
`When the dip-coating solutions are optimised with respect to
`
`t Unpublished observations by Rea
`
`10
`
`15
`
`20
`
`25
`Time. h
`
`30
`
`35
`
`40
`
`45
`
`50
`
`Fig.11 Stability of glucose current during continuous operation of
`the transducers in the presence of 5 mM D-glucose at 37°C.
`The transducers were hydrated for approximately 1 h in
`Miller & Golder's buffer and then transferred to buffer
`containing 5 mM D-glucose. Conditions were otherwise as
`for Fig. 6. Data taken from continuous chart recorder
`recordings, mean ±SE (-.-.); line of best
`fit,
`y = -0·83x+(85-48±6·98) where y =percentage initial
`current and x =time (h), computedfrom 4·5h onwards by
`least-squares method(--) (r = 0·891; n = 5)
`
`Medical & Biological Engineering & Computing
`
`March 1985
`
`113
`
`Dexcom Inc. v. WaveForm Technologies, Inc.
`IPR2017-01051
`Exhibit 1037
`
`

`

`glucose oxidase, cellulose acetate and polyurethane concen(cid:173)
`tration an electroenzymic glucose transducer with satisfac(cid:173)
`tory performance characteristics can readily be fabricated.
`The range over which the steady-state current is propor(cid:173)
`tional to D-glucose concentration encompasses both normal
`and abnormal blood glucose concentrations, thus enabling
`diagnostic applications, and the pH and temperature de(cid:173)
`pendencies of the glucose current are sufficiently small in the
`physiological range for pH and temperature compensation
`to be unnecessary. The range of linearity may have to be
`extended, by, for example, increasing the polyurethane
`content of the second dip-coat, for the continuous monitor(cid:173)
`ing of a known diabetic where blood glucose concentrations
`are frequently in excess of 12 mM (SINCLAIR, 1979). The
`response time of the transducer is sufficiently short to ensure
`continuous and accurate measurements of changes in blood
`glucose concentration. Perhaps the major shortcoming of
`the transducer, and one that is common to most electro(cid:173)
`enzymic transducers (SoELDNER, 1982), is its stability during
`prolonged use. When the transducer is operated continu(cid:173)
`ously in the presence of glucose at body temperature there is
`a steady and persistent decrease in current. As more than half
`of the initial current is lost after 50 h of operation, the
`transducer is clearly not suitable for long-term implantation.
`However, for short-term experimental or diagnostic pur(cid:173)
`poses the transducer is satisfactory on the condition that drift
`can be compensated by periodic calibrations. We are
`currently developing polyurethane bilumen catheter glucose
`transducers for intravascular use, and these are being tested
`in vivo. The flow and p0 2 dependence of the transducer will
`be evaluated in this measurement situation. Since H 20 2
`efflux from the surface of the transducer is an important
`determinant of the steady-state current (Section 2.4), the
`influence of tissue and blood catalase activities, which will
`contribute to the outwardly directed diffusion potential for
`H 20 2 and therefore change the response characteristics of
`the transducer, will also be determined.
`
`Acknowledgments-The authors wish to thank the MRC for
`financial support during the course of this investigation and the two
`anonymous reviewers for useful comments.
`
`References
`ALBISSER, A. M., LEIBEL, B. S., EWART, T. G., DAVIDOVAC, Z.,
`BOTZ, C. K. and ZINGG, W. (1974a) An artificial pancreas.
`Diabetes, 23, 389-396.
`ALBISSER, A. M., LEIBEL, B. S., EWART, T. G., DAVIDOVAC, Z.,
`BOTZ, c. K., ZINGG, W., SCHIPPER, H. and GANDER, R. (1974b)
`Clinical control of diabetes by the artificial pancreas. Ibid., 23,
`397-404.
`BAUMAN, E. K., GOODSON, L. H., GUILBAULT, G. G. and KRAMER,
`D. N. ( 1965) Preparation of immobilized cholinesterase for use in
`analytical chemistry. Anal. Chem., 37, 1378.
`HESSMAN, S. P. and SCHULTZ, R. D. (1973) Prototype glucose
`oxygen sensor for the artificial pancreas. Trans. Am. Soc. Artif
`Int. Organs, 19, 361.
`CARISTY, M. GREEN, A. CHRISTAU, B., KROMANN, H. and NERUP,
`K. (1979) Epidemiological studies of insulin dependent diabetes
`mellitus. Diabetes Care, 2, 127-130.
`DAWSON, R. M. C., ELLIOTT, D. C., ELLIOTT, w. H. and JONES,
`K. M. (1959) Data for Biochemical Research. Oxford University
`Press, London.
`DECKERT, T., POULSON, J.E. and LARSEN, M. (1978) Prognosis of
`diabetics with .diabetes onset before the age of thirty-one. I:
`Survival, causes of death and complications. Diabetologia, 14,
`363-370.
`
`EADIE, G. S. (1942) The inhibition of cholinesterase by physostig(cid:173)
`mine and prostigmine. J. Biol. Chem., 146, 85-93.
`ENGERMAN, R., BLOODWORTH, J.M. B., and NELSON, S. (1977)
`Relationship of microvascular disease in diabetics to metabolic
`control. Diabetes, 26, 760- 769.
`GUILBAULT, G. G. (1980) Use of enzyme electrodes in biomedical
`investigations. In Medical and biological applications of electro(cid:173)
`chemical devices. KORYTA, J. (Ed.), John Wiley & Sons Ltd.,
`London.
`GUILBAULT, G. G. and LUBRANO, G. J. (1973) An enzyme electrode
`for the aperometric determination of glucose. Anal. Chim. Acta,
`64, 439-455.
`HICKS, G. P. and UPDIKE, S. J. (1966) The preparation and
`characterisation of lyophilised poly

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket