The trainable hearing aid:
`What will it do for clients and clinicians?
`
`By Harvey Dillon, Justin A. Zakis, Hugh McDermott, Gitte Keidser,
`Wouter Dreschler, and Elizabeth Convery
`
`INTRODUCTION: THE PROBLEM
`The extreme flexibility of modern hearing aids is present-
`ing an increasing challenge to designers and clinicians:
`How should all the adjustable, or potentially adjustable,
`amplification parameters be set to make life best for the
`intended wearer?
`Of course, all hearing aid fittings begin with a pre-
`scription of some sort. We aim to measure something about
`the client (often just the audiogram) and use this infor-
`mation to deduce how each amplification parameter (e.g.,
`gain for low-frequency, high-intensity sounds) should be
`set. Prescriptive formulas are intended to make the pre-
`scription right for the average person with a particular set
`of measured characteristics (such as an audiogram).
`Inevitably, however, some individuals will prefer set-
`tings that are different from average. Also, prescriptive for-
`mulas are derived from some combination of a theoretical
`rationale and experimental data, and possibly some for-
`mulas do not even achieve their goal of prescribing para-
`meters that are right on average, let alone for any given
`individual.1 Some parameters (e.g., speed of automatic
`noise reduction) are not individually prescribed at all. The
`designer of the hearing aid or of the fitting software sets
`them to the same value for all wearers, even though there
`may be little evidence to guide what that value should be.
`For all these reasons, it is relatively common for a hear-
`ing aid to be better matched to the individual needs of its
`wearer if adjustments are made away from the prescribed
`response. That is, most clinicians should consider the pre-
`scribed response to be a reasonable starting point, not the
`end, of the adjustment journey.
`This raises the difficult issue of how further adjustment
`should be performed. Of necessity, the clinician has made
`this adjustment, via the programming software, in the
`clinic. But there are many reasons why such an adjustment
`process may not lead to an optimal fitting:
`❖ The clinic is usually a low-noise, low-reverberation
`environment, and often the only stimulus used is the
`clinician’s voice. At best, such an adjustment can lead
`to an optimal fitting only for environments with these
`same ideal acoustic characteristics. It is well established
`that the optimal amplification characteristics vary with
`the acoustic environment.2
`❖ Other stimuli can be used, but to simulate every sit-
`uation in which people wear hearing aids would require
`a seemingly endless set of combinations. These include
`different voices at different levels, different types of
`
`noise at different levels, different degrees of reverber-
`ation, different types of non-speech sounds of inter-
`est, different directions of incoming sound (both
`sounds of interest and competing sounds), and dif-
`ferent rates of change of any of the preceding items.
`While any individual aid wearer may wish to have the
`hearing aids optimized for operation in only a hand-
`ful of places of particular importance, it would be a
`daunting task to try to approximate even these situa-
`tions in the clinic. Even if it were possible to make the
`acoustic simulation realistic, the visual environment
`would not be, which might well affect the adjustment
`that is optimal.3
`❖ If a small set of realistic environments could be cre-
`ated in the clinic, it is far from straightforward to struc-
`ture the hearing aid adjustment process so that
`adjustments made when the client is listening to one
`sound do not undo those made while the person was
`listening to previous sounds. The result may well be
`an end adjustment that is worse than the prescribed
`starting point.
`❖ Even if a converging adjustment occurs, the clinical
`time involved would be prohibitive. This will exacer-
`bate the coming problem where, due to aging of the
`population in most developed countries, there will be
`more need than ever to streamline the clinical process
`if available clinical resources are to meet the enormous
`growth in demand for services that will occur over the
`next 25 years.
`Currently, hearing aid wearers do not actually know if
`their hearing aids are optimally adjusted in any situation
`in which they use them. If the sound of the aid is so poor
`they find it unacceptable, they will return to the clinic,
`either to return the hearing aid(s) or to ask that they be
`adjusted. The clinician has to infer the acoustics of the sit-
`uation, understand the nature of the client’s dissatisfaction
`with the sound, and deduce (sometimes with software assis-
`tance) the parameter to be adjusted, the direction of the
`adjustment, and the extent of the adjustment.
`Obviously, this process may go wrong or need to be repeated
`several times, especially if the client continues to experience
`new environments with different acoustic characteristics and
`perceive that the sound is not always quite right.
`So, we are left with a problem: How can we optimally
`adjust the hearing aid? The remainder of this article describes
`a new concept, the trainable hearing aid, that can address
`this problem.
`
`30
`
`The Hearing Journal
`
`A trainable hearing aid
`
`April 2006 • Vol. 59 • No. 4
`
`HIMPP 1014
`
`

`

`A greater level of
`sophistication will
`enable the trainable
`aid to match the vol-
`ume setting the client
`selected to the acou-
`stic environment at
`the time the adjust-
`ment was made. For
`example, the hearing
`aid could measure and
`record the overall level
`of sound in the envi-
`ronment for the few
`seconds prior to the
`volume control being
`adjusted. After being
`adjusted several times
`in different environments, the hearing aid
`develops a picture of how much gain the
`client prefers in each environment.
`Figure 1 shows what this might look
`like after just six adjustments of the hear-
`ing aid. As would be expected, the wearer
`generally prefers less gain (averaged across
`frequencies) as the environment gets
`louder. Armed with this information, how
`much gain would the ideal hearing aid
`automatically select if the user were then
`to enter a new environment with an over-
`all level of, say, 80 dB SPL? The hearing
`aid could estimate this by fitting a curve
`(or in this case a two-segment straight line)
`to the data available to it. From the line
`in Figure 1, the hearing aid would deduce
`that the wearer is likely to select a gain of
`about 12 dB in this new situation, even
`though the user has never provided any
`training for this specific input level.
`If the trainable aid does its job well
`enough, the wearer will not need to touch
`the volume control in this situation.
`Although the wearer has had to do only
`a single task—varying the volume con-
`trol—the hearing aid has been able to
`deduce how the gain should vary as the
`input level changes. That is, the individ-
`
`ual gain adjustments have trained the
`device to have an individually optimized
`low-level gain (in this case 24 dB), com-
`pression ratio (1.7), compression thresh-
`old (55 dB SPL), and, therefore, absolute
`gain for any input level.
`Figure 2 shows a general block dia-
`gram for one form of a trainable hearing
`aid. For the relatively simple version of
`the trainable aid discussed above, the
`acoustic measurement module would con-
`sist of a simple sound-level meter that cal-
`culates overall sound pressure level
`averaged over a few seconds. The pro-
`grammable amplifier is no different from
`the (digital processing) amplifier in any
`current non-linear hearing aid. It’s just
`that its programming inputs remain per-
`manently (and internally) connected to
`the learning algorithm module rather than
`being disconnected once the clinician has
`finished programming the hearing aid.
`The simplest way to think about the
`learning algorithm module is that it main-
`tains a record of user adjustments and the
`corresponding acoustic environments,
`along with the appropriate statistical or
`mathematical processes to deduce the set
`of amplification parameters that best
`match the user’s preferences. In practice,
`there are ways to achieve this without
`requiring the learning algorithm to mem-
`orize every single adjustment and corre-
`sponding set of environment acoustics.
`The user control can be a simple rotary
`control, a toggle, or a pair of up-down
`buttons, and the control can be mounted
`on the hearing aid itself or on a remote
`control. The microphone and receiver are
`entirely conventional.
`Note that, based on the data in Figure
`1, the relationship between the gain pre-
`ferred and the input level is imperfect in
`that the points deviate from the fitted
`curve. For example, on the two occasions
`when the overall level in the environment
`was 60 dB SPL, what caused the wearer
`
`April 2006 • Vol. 59 • No. 4
`
`A trainable hearing aid
`
`The Hearing Journal
`
`31
`
`Figure 2. Block diagram for one general form of a trainable hearing aid.
`
`Figure 1. Hypothetical data showing the gain (averaged across
`frequency) preferred by a hearing aid user in six situations
`versus the overall sound pressure level in each situation
`
`THE CONCEPT AND RANGE
`OF TRAINABILITY
`With a trainable hearing aid, the wearer
`teaches the instrument how it should be
`adjusted. The wearer does so by using the
`aid in those situations in which he or she
`would like assistance with hearing. There-
`fore, the process takes place after the client
`leaves the clinic. Once the clinician explains
`the process to the wearer, the clinician need
`not spend further time on the process.
`Trainable hearing aids will vary in their
`complexity and effectiveness. The first
`generation of device will likely be capa-
`ble of learning only one thing: the vol-
`ume control setting preferred by the client.
`A memory inside the instrument will take
`note of the setting last used by the client.
`When the hearing aid is next switched on,
`it will automatically set the volume to
`some combination of this most recent set-
`ting and the settings it remembers from
`the previous switch-on and/or from other
`times the hearing aid was used.
`If the hearing aid simply set the vol-
`ume control to the setting last used, this
`would return volume control adjustment
`to how it was before digital hearing aids.
`Gain was determined by the physical posi-
`tion of the volume control and by the
`operation of any non-linear features (e.g.,
`compression, noise reduction). Instead,
`the first generation of trainable hearing
`aids will give the user the average of the
`volume control settings the person has
`adopted previously. Thus, it will not be
`affected so much by whether the user was
`in a particularly quiet or particularly noisy
`place when he or she last adjusted the hear-
`ing aid.
`
`

`

`to select a 17-dB gain on one occasion
`but 25 dB on the other? Perhaps on the
`first occasion the dominant sound was
`rain on the roof that the person did not
`want to listen to, while on the second it
`was speech that the client particularly
`wanted to understand. Consequently,
`another level of sophistication for train-
`able hearing aids will be to measure much
`more about the environment than just the
`overall level of sound.
`For example, hearing aids could mea-
`sure the spectral shape, the rate and extent
`to which the spectral shape varies, the
`apparent direction of sound, the modu-
`lation of sound within each frequency
`channel, and the way these modulations
`match each other across channels. These
`characteristics could be individually input
`to the learning algorithm, or they could
`be combined to classify environments in
`various ways, and the resulting classifica-
`tions input to the learning algorithm.
`Obviously, a wide range of acoustic
`characteristics of sound might influence
`the type and degree of amplification the
`user would prefer. It is hard to conceptu-
`
`alize the relationships in a simple graph,
`but inferential statistics can be used to
`derive the link between these acoustic
`characteristics and the amount of gain the
`client chose. Although the hearing aid can
`never actually read the client’s mind (at
`least not in this decade), the hearing aid
`is able to “hear” everything that the client
`hears and take into account many of the
`factors that the client uses in deciding how
`far the hearing aid should be turned up
`or down.
`The trick to doing this automatically
`is that the instrument must first learn
`which factors are actually important (in
`that they seem to affect what the client
`prefers) and what the link is between the
`value of each factor and the gain preferred
`by the client. The necessary statistical oper-
`ations are well within the signal process-
`ing capabilities of the hearing aids that will
`be appearing over the next few years. Many
`of the more sophisticated acoustic para-
`meters mentioned here are already being
`measured in current hearing aids, and the
`measurements are used to control ampli-
`fication in a way determined by the hear-
`
`ing aid designer or by the clinician. By
`contrast, in the trainable aid, the client
`determines the way these acoustic para-
`meters affect the amplification provided.
`
`ADJUSTING THE GAIN-
`FREQUENCY RESPONSE
`So far we have talked only about the hear-
`ing aid automatically—but very intelli-
`gently—adjusting the gain in the same
`way as the user would adjust the volume
`control as situations changed. A hearing
`aid that almost always gave the loudness
`that the user preferred would be a great
`advance over existing hearing aids.4-6
`However, as clinicians know only too
`well, there are many other things to adjust
`in a hearing aid. For example, the amount
`of amplification has to vary with fre-
`quency, and the amount at each frequency
`has to vary with input level, usually by
`different degrees. Therefore, the next level
`of sophistication for the trainable hearing
`aid is the individual adjustment of the
`shape of the gain-frequency response.
`At first sight, it might seem necessary
`to present the user with something like
`
`32
`
`The Hearing Journal
`
`A trainable hearing aid
`
`April 2006 • Vol. 59 • No. 4
`
`

`

`an audio mixing console, or at least a set
`of tone controls, to achieve this. How-
`ever, the trainable aid has to be simple
`enough for people to use it with minimal
`instruction.
`One solution, which we have success-
`fully used in some of our experimental
`work with trainable aids, is to have a sin-
`gle control that takes on different func-
`tions from time to time.7 For example, it
`might be a volume control when first
`adjusted, then become a treble control,
`which turns into a bass control, before
`once again becoming a volume control.
`If the user had to understand exactly
`what was occurring at any time, this would
`be a nightmare for them to use. However,
`if the instruction to the wearer is simply,
`“Turn this control to the position where
`the hearing aid sounds best, and then leave
`it there for 10 seconds or more,” the client
`actually has a simple task to perform, even
`though the amplification parameter being
`trained varies from time to time.
`A different solution is to give the user
`a small number of controls, preferably
`mounted on a remote control. The remote
`
`control could be discarded once training
`was completed.
`Our experimentation to date has indi-
`cated that users can reliably operate up to
`three controls that, between them, gov-
`ern the broad shape of the gain-frequency
`response.8 Indeed, there are several viable
`ways that the controls can be linked to
`the shape of the gain-frequency response,
`although our experimental subjects pre-
`ferred some arrangements to others.
`By analogy with the simple volume
`control trainable aid, if the user can adjust
`the gain-frequency response at any one
`input level and if this is repeated in mul-
`tiple environments, each with its own
`overall input level and spectral shape, then
`the learning algorithm is able to deduce,
`within each frequency region, the desired
`absolute gain at each input level, and thus
`the individually preferred compression
`ratios and compression thresholds.
`In principle, there is no amplification
`parameter that cannot be trained by the
`user, provided he or she can hear some-
`thing change in the sound quality as the
`underlying parameter is varied by moving
`
`the control throughout its range. Quan-
`tities such as depth and speed of automatic
`noise suppression, selection and speed of
`adaptive directionality, compression speed,
`spectral enhancement, and frequency trans-
`position can all be subject to training. In
`general, the more parameters we attempt
`to train, the more time the user will need
`to spend training the device.
`Our experimental data with a trainable
`aid that controlled the non-linear gain-fre-
`quency response suggest that, not sur-
`prisingly, the trained hearing aid became
`more strongly preferred to non-trainable
`(but otherwise identical) hearing aids the
`longer the user spent training it.7 A few
`weeks of training is certainly long enough
`to accumulate the necessary number of aid
`adjustments to achieve a significant pref-
`erence for the trained aid. This preference
`was marked and significant, even under
`double-blinded conditions.7
`Clients do typically appear to be will-
`ing to spend a few weeks training their
`hearing aids, though some are willing to
`spend only a few days and others are will-
`ing to spend months.9 The effort may not
`
`April 2006 • Vol. 59 • No. 4
`
`A trainable hearing aid
`
`The Hearing Journal
`
`33
`
`

`

`be any greater than the client would oth-
`erwise have spent adjusting a conventional
`volume control. And, unlike a conven-
`tional hearing aid, the need to adjust
`should decrease as the hearing aid learns.
`Because trainability can be used in so
`many ways, there is unlikely to be a sin-
`gle “trainable hearing aid.” Rather, many
`companies will introduce trainability, and
`different devices will vary in their method
`of user control, the aspects of the acoustic
`environment that are monitored, and the
`amplification parameters that are con-
`trolled.
`Having explained what we mean by a
`trainable hearing aid, let’s distinguish it
`from a data-logging aid. A trainable hear-
`ing aid may well require the hearing aid
`to log data—user inputs, at least, and
`acoustic environment characteristics for
`a more sophisticated device. Furthermore,
`the data logged by a trainable hearing aid
`could subsequently be perused by the clin-
`ician (in person or over the Internet),
`which is another similarity with data-log-
`ging aids. However, if the hearing aid only
`logs data, for later perusal and action by
`the clinician or the fitting software, and
`does not automatically change amplifica-
`tion on the basis of input from the user
`that it logs, it is not user-trainable.
`
`IMPACT IN THE CLINIC
`Trainability is likely to have a profound
`effect on how hearing aids are prescribed,
`measured, and adjusted in the clinic. It
`may also affect how a succession of
`appointments is structured, and signifi-
`cantly decrease the average number and/or
`length of appointments.
`Most obviously, if the purpose of train-
`ability is to enable the client to customize
`the broad shape of the gain-frequency
`response to his or her particular listening
`environments, then there is no point in
`spending valuable clinical time achieving
`a close match to a prescribed response.
`While a good prescription should still be
`used to ensure that the sound is safe and
`reasonable from the first time the hearing
`aid is worn, it does not seem worthwhile
`spending the time to do real-ear mea-
`surements and subsequent adjustments,
`provided that the manufacturers’ software
`is reasonably accurate at approximating
`the prescription from the outset.
`Further time saving should occur
`whenever a client contacts the dispenser
`
`to complain about the sound quality. Sup-
`pose, for example, that the client finds
`that the hearing aid sounds unpleasantly
`loud and rumbly in traffic noise. Instead
`of scheduling the client for an adjustment
`appointment (or worse, advising the client
`that his or her brain will get used to it and
`accommodate!), the clinic will simply
`instruct the client to head for the nearest
`traffic and repeatedly adjust the control
`to whatever sounds best in that situation.
`Assuming the trainable aid enables the
`client to control the gain-frequency
`response, it will progressively turn down
`the low-frequency gain for high-level
`sounds, if that modification will solve the
`problem. If the trainable aid has sophis-
`ticated environment-monitoring capabil-
`ities and if it is only in traffic noise that
`the client benefits from this adjustment,
`then the automatic adjustment will take
`place only in traffic noise and not affect
`other high-intensity,
`low-frequency
`sounds, such as the wearer’s own voice.
`Pretty quickly, the client will learn that
`the permanent solution to unacceptable
`sound in any situation is to use, and for
`a limited time, adjust the hearing aid in
`that situation. In the process, some clients
`will likely begin to feel more “ownership”
`over the adjustment of the hearing aid
`than if they regard it as something beyond
`their control. Other clients will no doubt
`prefer to leave everything to “the expert.”
`A survey of the clinical (adult, hear-
`ing aid wearing) population has demon-
`strated an overwhelmingly positive
`response to the concept of trainability.9
`Our hope is that trainable hearing aids
`will lead to more effectively customized
`hearing aids, as well as to a greater sense
`of ownership of the fitting, and that this
`combination of improvements will reduce
`the incidence of unused and returned
`hearing aids. However, we will have no
`data on this until trainable hearing aids
`are available in a commercial and cos-
`metically acceptable form.
`
`OTHER POTENTIAL BENEFITS
`Trainable hearing aids could also be set to
`detect long-term changes in the gain
`selected by the user. As the most likely
`explanations for this would be deteriora-
`tion of the user’s hearing or the hearing
`aid’s becoming faulty (e.g., progressively
`blocked with wax), it would even be pos-
`sible to program the hearing aid to warn
`
`the wearer that it is time for a return visit
`to the hearing clinic. What clients think
`about their hearing aid talking to them
`(preferably at some quiet moment) has
`not yet been tested by our research team!
`Clinicians have long suspected that
`hearing aid wearers’ amplification require-
`ments change once they acquire experi-
`ence with the hearing aids. It is commonly
`believed that new users prefer less overall
`gain and less high-frequency emphasis
`than experienced users.
`Several manufacturers include acclima-
`tization or adaptation managers in their
`software to modify the standard pre-
`scriptions. While research support for these
`views is minimal,10,11 the trainable aid
`provides an easy way to allow for these
`effects if the concern is real and causes no
`disadvantage if it is not. Users can start
`and stop training the trainable aid any
`time they wish. A reasonable clinical strat-
`egy would therefore be to instruct clients
`to train the aid intensively during the first
`few weeks, then forget about training and
`enjoy the benefits of the trained aid for
`the next few months. At some later stage,
`they could resume training for an addi-
`tional week or two, and if the user’s pref-
`erences have changed, the trainable aid
`will accommodate to such changes with-
`out further involvement by the clinician.
`Clinicians may wonder if they will still
`be needed at all if the trainable hearing
`aid becomes popular. The answer is an
`easy yes, though how they spend their
`time may well change. If they spend less
`time on hearing aid measurement, adjust-
`ment, and re-adjustment, then they will
`have more time for:
`❖ Understanding clients’ needs, beliefs,
`and motivation or lack of motivation
`prior to a fitting. Extensive research
`indicates that a client’s motivation to
`use hearing aids is an important deter-
`minant of success with amplifica-
`tion.12-16 Additional counseling
`aimed at uncovering clients’ beliefs
`and providing additional perspectives
`for them to consider prior to fitting
`may be effective in increasing hear-
`ing aid use and satisfaction.
`❖ Giving information after the hearing
`aid fitting about the use of assistive lis-
`tening devices, communication or envi-
`ronmental strategies in difficult listening
`situations, and telephone usage.
`❖ Assessing and fitting additional
`
`34
`
`The Hearing Journal
`
`A trainable hearing aid
`
`April 2006 • Vol. 59 • No. 4
`
`

`

`clients, given the likely steep rise in
`demand for services.
`❖ Managing future clients who will have
`complex devices that combine the
`functionality of cochlear implants and
`hearing aids.
`❖ Dealing with the logistical issues that
`will arise from the likely merging of
`hearing aids and other communica-
`tion/entertainment/functional devices
`that provide speech to the ear canals
`(for people with mild hearing loss).
`Time saving will occur only with those
`clients who actually receive a trainable aid.
`We know that it will not be suitable for
`all clients (see next section), so the clini-
`cian will have to determine which clients
`should be offered a trainable aid, and what
`complexity of aid to offer them.
`
`IMPACT ON USERS
`We have already mentioned numerous
`positive impacts of trainability on the hear-
`ing aid wearer. These include:
`❖ A hearing aid with amplification cus-
`tomized by the wearer to his or her
`individual preference in the actual
`environments in which the device will
`be worn.
`❖ Fewer visits to the clinic, especially
`for people initially dissatisfied with
`the sound of their hearing aids.
`❖ The ability for wearers to directly re-
`customize their hearing aids if their
`preferences change following experi-
`ence with the hearing aids, if the hear-
`ing aids are worn in new acoustic
`environments, or if the wearer’s hear-
`ing loss fluctuates.
`❖ The ability to have more acoustic
`parameters customized to preferred
`values than is possible within the time
`constraints of clinical appointments.
`❖ Greater psychological “ownership” of
`the hearing aid fitting and adjustment
`process.
`Against these advantages, the aid
`wearer has to spend time training the hear-
`ing aid(s) during the first few weeks of
`use, which may be inconvenient. Based
`on our experience with experimental sub-
`jects who have evaluated a prototype train-
`able aid and based on the views that we
`sought from more than 100 hearing aid
`wearers who were allowed to try the con-
`trols of a highly trainable device, we expect
`that at least half of hearing aid wearers
`will consider that the benefits of train-
`
`ability outweigh the disadvantages. Our
`profile analyses suggest that factors such
`as gender, age, education, and attitude
`toward technology can be used to predict
`if a client is likely to be a candidate.9
`Even wider application will be found
`if the degree of trainability selected matches
`the technical sophistication of the wearer.
`For example, anyone who can physically
`manipulate a volume control (on the hear-
`ing aid or on a remote control) could use
`a trainable aid that simply made the vol-
`ume control operate in a more intelligent
`fashion. By contrast, the use of a two-con-
`trol trainable aid in which each control
`changes its function from time to time
`would require the wearer to possess a
`higher degree of cognitive functioning.17
`There will doubtless continue to be a need
`for accurately prescribed, fully automatic
`hearing aids without controls for those
`clients least able to operate hearing aids.
`One doubt that might be raised about
`the trainable aid is that it provides the
`amplification characteristics the wearer
`prefers, not necessarily those that are best
`for the wearer. We don’t think this is a
`valid criticism. Who is to say, with any
`certainty, what is best for a client? One
`would need to be extremely confident
`about a prescription to assert that it is best
`for the client when the client is asserting
`that, all things considered, some alterna-
`tive response was better.
`If the prescribed responses really were
`best for everyone, hearing aid fitting
`would be a straightforward mechanical
`process in which fine-tuning based on
`client report was never necessary. Pre-
`scriptive procedures usually try to achieve
`just one or two aims. For example, NAL-
`NL1 seeks to maximize intelligibility
`while keeping overall loudness less than
`or equal to normal.18
`A hearing aid wearer is likely to have
`additional goals, such as maximizing nat-
`uralness, maximizing localization ability,
`or minimizing the distracting effects of
`certain sounds. In our view, the client is
`better able to make a balanced choice
`between alternative amplification charac-
`teristics than anyone else. The client can
`self-assess the ease of understanding speech
`with each amplification alternative offered,
`and weigh this in with other considera-
`tions. We hope that when clients reliably
`get amplification adjusted in the way they
`think is best, more of those who need
`
`amplification will appreciate the many
`advantages of hearing aids.
`
`Harvey Dillon, PhD, is Director of Research at National Acoustic
`Laboratories (NAL) and Deputy Director of the Cooperative Research
`Centre for Cochlear Implant and Hearing Aid Innovation (CRCCIHAI).
`Justin A. Zakis, PhD, was Research Scientist at CRCCIHAI and at
`the Bionic Ear Institute and is now at Dynamic Hearing. Hugh McDer-
`mott, PhD, is Associate Professor at CRCCIHAI and at the University
`of Melbourne. Gitte Keidser, PhD, is Senior Research Scientist at
`CRCCIHAI and at NAL. Wouter Dreschler, PhD, was visiting sci-
`entist at CRCCIHAI and is Professor of Clinical and Experimental Audi-
`ology at
`the Academic Medical Center, Amsterdam.
`Elizabeth Convery, MAud, is research audiologist at CRCCIHAI
`and at NAL. Readers may contact Dr. Dillon at Harvey.dillon@nal.gov.au.
`
`REFERENCES
`1. Keidser G, Brew C, Peck A: Proprietary fitting algorithms
`compared with one another and with generic formu-
`las. Hear J 2003;56(3):28-37.
`2. Keidser G, Brew C, Brewer S, et al.: The preferred response
`slopes and two-channel compression ratios in twenty
`listening conditions by hearing-impaired and normal-
`hearing listeners and their relationship to the acoustic
`input. IJA 2005;44(11):656-670.
`3. Smeds K, Keidser G, Zakis J, et al.: Preferred overall
`loudness II. Listening through hearing aids in field
`and laboratory tests. IJA 2006;45:12-25.
`4. Kochkin S: MarkeTrak IV: Ten-year customer satisfac-
`tion trends in the US hearing instrument market. Hear
`Rev 2002;9(10):14-25,46.
`5. Kochkin S: MarkeTrak IV: Isolating the impact of the
`volume control on customer satisfaction. Hear Rev
`2003;10(1):26-35.
`6. Jenstad LM, Van Tasell DJ, Ewert C: Hearing aid trou-
`bleshooting based on patients’ descriptions. JAAA
`2003;14:347-360.
`7. Zakis JA, Dillon H, McDermott H: The design and eval-
`uation of a hearing aid with trainable amplification
`parameters. In preparation.
`8. Dreschler WA, Keidser G, Convery E, et al.: Reliability
`of client-based adjustments of amplification. In prepa-
`ration.
`9. Keidser G, Convery E, Dillon H: A trainable hearing
`aid: Candidacy and perception by a clinical popula-
`tion. Submitted.
`10. Convery E, Keidser G, Dillon H: A review: Does ampli-
`fication experience have an effect on preferred gain
`over time? Aust NZ J Audiol 2005;27(1):18-32.
`11. Keidser G, Carter L, Dillon H: The effect of advanced
`signal processing strategies in hearing aids on user per-
`formance and preference. In Rasmussen AN, Oster-
`hammel PA, Andersen T, Poulsen T, eds., Hearing Aid
`Fitting, 21st Danavox Symposium, in press.
`12. Brooks D: Measures for the assessment of hearing aid
`provision and rehabilitation. Brit J Audiol 1990;
`24(4):229-233.
`13. Erdman S, Crowley J: Considerations in counseling for
`the hearing impaired. Hear Instr 1984;35(11):50-58.
`14. Gatehouse S: Glasgow Hearing Aid Benefit Profile:
`Derivation and validation of a client-centered out-
`come measure for hearing aid services. JAAA
`1999;10(2):80-103.
`15. Hickson L, Hamilton L, Orange SP: Factors associated
`with hearing aid use. Aust J Audiol 1986;8(2):37-41.
`16. Hickson L, Timm M, Worrall L, et al.: Hearing aid
`fitting: Outcomes for older adults. Aust J Audiol
`1999;21(1):9-21.
`17. Lunner T: Cognitive function in relation to hearing aid
`use. IJA 2003;42 (Suppl 1):49-58.
`18. Byrne D, Dillon H, Ching T, et al.: NAL-NL1 proce-
`dure for fitting non-linear hearing aids: Characteris-
`tics and comparisons with other procedures. JAAA
`2001;12(1):37-51.
`
`36
`
`The Hearing Journal
`
`A trainable hearing aid
`
`April 2006 • Vol. 59 • No. 4
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket