throbber
Co
`
`
`
`Edwards SAPIEN 3
`Transcatheter Heart Valve with the Edwards Commander Delivery
`System
`
`
`
`Instructions for Use
`
`CAUTION: Federal (USA) law restricts these devices to sale by or on the order of a physician.
`Implantation of the transcatheter heart valve should be performed only by physicians who have
`received Edwards Lifesciences training. The implanting physician should be experienced in
`balloon aortic valvuloplasty.
`Please verify that you have the latest version of the instructions for use prior to using the device
`by visiting http://THVIFU.edwards.com or by calling 1.800.822.9837. In order to access the
`instructions for use, an IFU Code will be required.
`
`STERILE: The THV is supplied sterilized with glutaraldehyde solution. The delivery system,
`eSheath introducer set, and crimper are supplied sterilized with ethylene oxide gas.
`
`
`
`
`
`
`
`
`
`Edwards, Edwards Lifesciences, the stylized E logo, Carpentier-Edwards, Edwards Commander,
`Edwards eSheath, Edwards SAPIEN, Edwards SAPIEN 3, eSheath, Qualcrimp, PARTNER, PARTNER II,
`SAPIEN, SAPIEN 3, TFX and ThermaFix are trademarks of Edwards Lifesciences Corporation.
`All other trademarks are the property of their respective owners.
`1
`
`Page 1 of 30
`
`

`

`1.0 Device Description
`• Edwards SAPIEN 3 Transcatheter Heart Valve- Model 9600TFX (Figure 1)
`The Edwards SAPIEN 3 Transcatheter Heart Valve (THV) is comprised of a balloon-expandable,
`radiopaque, cobalt-chromium frame, trileaflet bovine pericardial tissue valve, and polyethylene
`terephthalate (PET) fabric skirt. The leaflets are treated according to the Carpentier-Edwards ThermaFix
`process.
`
`
`Table 1
`Valve Size
`20 mm
`23 mm
`26 mm
`29 mm
`
`Height
`15.5 mm
`18 mm
`20 mm
`22.5 mm
`
`
`
`
`Table 2
`Native Valve Annulus Size
`(CT)
`
`Area
`
`Native Valve Annulus Size
`(TEE)
`
`THV Size
`
`Area Derived
`Diameter
`273 – 345 mm2
`20 mm
`18.6-21 mm
`16-19 mm
`338 – 430 mm2
`23 mm
`20.7-23.4 mm
`18-22 mm
`430 – 546 mm2
`26 mm
`23.4-26.4 mm
`21-25 mm
`540 – 683 mm2
`29 mm
`26.2-29.5 mm
`24-28 mm
`THV size recommendations are based on native valve annulus size, as measured by transesophageal
`echocardiography (TEE) or computed tomography (CT). Patient anatomical factors and multiple imaging
`modalities should be considered during THV size selection. Note: Risks associated with undersizing and
`oversizing should be considered.
`
`
`
` •
`
` Edwards Commander Delivery System (Figure 2)
`The Edwards Commander delivery system (usable length 105 cm) is used for delivery of the Edwards
`SAPIEN 3 transcatheter heart valve and consists of a Flex Catheter to aid in valve alignment to the
`balloon, tracking, and positioning of the THV. The delivery system includes a tapered tip to facilitate
`crossing of the native valve. The handle contains a Flex Wheel to control flexing of the Flex Catheter, and
`a Balloon Lock and Fine Adjustment Wheel to facilitate valve alignment and positioning of the valve within
`the native annulus. A stylet is included within the guidewire lumen of the delivery system. The Balloon
`Catheter has radiopaque Valve Alignment Markers defining the working length of the balloon. A
`radiopaque Center Marker in the balloon is provided to help with valve positioning. A radiopaque Triple
`Marker proximal to the balloon indicates the Flex Catheter position during deployment. The inflation
`parameters for THV deployment are:
`
`
`Table 3
`Nominal
`Balloon
`Diameter
`20 mm
`23 mm
`26 mm
`29 mm
`
`Nominal
`Inflation
`Volume
`11 mL
`17 mL
`23 mL
`33 mL
`
`Rated Burst
`Pressure
`(RBP)
`7 atm
`7 atm
`7 atm
`7 atm
`
`Model
`
`9600LDS20
`9600LDS23
`9600LDS26
`9600LDS29
`
`
`
`2
`
`Page 2 of 30
`
`

`

`
`
`
`
`
`Figure 2 Edwards Commander Delivery System
`
`
`• Qualcrimp Crimping Accessory (Figure 3)
`The Qualcrimp crimping accessory (packaged with the Edwards Commander delivery system) is used
`during crimping of the THV.
`
`Figure 3
`
`
`
`
`
`Laminated Qualcrimp
`
`• Edwards eSheath Introducer Set
`
`Refer to the Edwards eSheath Introducer Set for device description.
`
`• Edwards Crimper
`
`Refer to the Edwards Crimper instructions for use for device description.
`
`
`
`Indications
`2.0
`The Edwards SAPIEN 3 Transcatheter Heart Valve (THV), model 9600TFX, and accessories are
`indicated for relief of aortic stenosis in patients with symptomatic heart disease due to severe native
`calcific aortic stenosis who are judged by a heart team, including a cardiac surgeon, to be at high or
`greater risk for open surgical therapy (i.e., Society of Thoracic Surgeons operative risk score ≥8% or at a
`≥15% risk of mortality at 30 days).
`
`3
`
`Page 3 of 30
`
`

`

`3.0 Contraindications
`The THV and delivery systems are contraindicated in patients who cannot tolerate an
`anticoagulation/antiplatelet regimen or who have active bacterial endocarditis or other active infections.
`
`4.0 Warnings
`• Observation of the pacing lead throughout the procedure is essential to avoid the potential risk of
`pacing lead perforation.
`
`• There is an increased risk of stroke in transcatheter aortic valve replacement procedures, as
`compared to balloon aortic valvuloplasty or other standard treatments.
`• The devices are designed, intended, and distributed for single use only. Do not resterilize or reuse
`the devices. There are no data to support the sterility, nonpyrogenicity, and functionality of the
`devices after reprocessing.
`• Incorrect sizing of the THV may lead to paravalvular leak, migration, embolization and/or annular
`rupture.
`• Accelerated deterioration of the THV may occur in patients with an altered calcium metabolism.
`• Prior to delivery, the THV must remain hydrated at all times and cannot be exposed to solutions other
`than its shipping storage solution and sterile physiologic rinsing solution. THV leaflets mishandled or
`damaged during any part of the procedure will require replacement of the THV.
`• Caution should be exercised in implanting a THV in patients with clinically significant coronary artery
`disease.
`• Patients with pre-existing mitral valve devices should be carefully assessed prior to implantation of
`the THV to ensure proper THV positioning and deployment.
`• Do not use the THV if the tamper evident seal is broken, the storage solution does not completely
`cover the THV, the temperature indicator has been activated, the THV is damaged, or the expiration
`date has elapsed.
`• Do not mishandle the delivery system or use it if the packaging or any components are not sterile,
`have been opened or are damaged (e.g. kinked or stretched), or the expiration date has elapsed.
`• Use of excessive contrast media may lead to renal failure. Measure the patient’s creatinine level prior
`to the procedure. Contrast media usage should be monitored.
`• Patient injury could occur if the delivery system is not un-flexed prior to removal.
`• Care should be exercised in patients with hypersensitivities to cobalt, nickel, chromium, molybdenum,
`titanium, manganese, silicon, and/or polymeric materials.
`• The procedure should be conducted under fluoroscopic guidance. Some fluoroscopically guided
`procedures are associated with a risk of radiation injury to the skin. These injuries may be painful,
`disfiguring, and long-lasting.
`• THV recipients should be maintained on anticoagulant/antiplatelet therapy, except when
`contraindicated, as determined by their physician. This device has not been tested for use without
`anticoagulation.
`• Do not add or apply antibiotics to the storage solution, rinse solutions, or to the THV.
`
`5.0 Precautions
`• Long-term durability has not been established for the THV. Regular medical follow-up is advised to
`evaluate THV performance.
`• Glutaraldehyde may cause irritation of the skin, eyes, nose and throat. Avoid prolonged or repeated
`exposure to, or breathing of, the solution. Use only with adequate ventilation. If skin contact occurs,
`immediately flush the affected area with water; in the event of contact with eyes, seek immediate
`medical attention. For more information about glutaraldehyde exposure, refer to the Material Safety
`Data Sheet available from Edwards Lifesciences.
`
`
`4
`
`Page 4 of 30
`
`

`

`• To maintain proper valve leaflet coaptation, do not overinflate the deployment balloon.
`
`• Appropriate antibiotic prophylaxis is recommended post-procedure in patients at risk for prosthetic
`valve infection and endocarditis.
`
`• Safety, effectiveness, and durability have not been established for valve-in-valve procedures.
`
`• Safety and effectiveness have not been established for patients with the following
`characteristics/comorbidities:
`o Non-calcified aortic annulus
`o Severe ventricular dysfunction with ejection fraction < 20%
`o Congenital unicuspid or congenital bicuspid aortic valve
`o Mixed aortic valve disease (aortic stenosis and aortic regurgitation with predominant aortic
`regurgitation > 3+)
`o Pre-existing prosthetic heart valve or prosthetic ring in any position
`o Severe mitral annular calcification (MAC), severe (> 3+) mitral insufficiency, or Gorlin
`syndrome
`o Blood dyscrasias defined as: leukopenia (WBC < 3000 cells/mL), acute anemia (Hb < 9
`g/dL), thrombocytopenia (platelet count < 50,000 cells/mL), or history of bleeding
`diathesis or coagulopathy
`o Hypertrophic cardiomyopathy with or without obstruction (HOCM)
`o Echocardiographic evidence of intracardiac mass, thrombus, or vegetation
`o A known hypersensitivity or contraindication to aspirin, heparin, ticlopidine (Ticlid™), or
`clopidogrel (Plavix™), or sensitivity to contrast media, which cannot be adequately premedicated
`o Significant aortic disease, including abdominal aortic or thoracic aneurysm defined as maximal
`luminal diameter 5 cm or greater; marked tortuosity (hyperacute bend), aortic arch atheroma
`(especially if thick [> 5 mm], protruding, or ulcerated) or narrowing (especially with calcification
`and surface irregularities) of the abdominal or thoracic aorta, severe “unfolding” and tortuosity of
`the thoracic aorta
`o Access characteristics that would preclude safe placement of 14F or 16F Edwards eSheath
`Introducer Set, such as severe obstructive calcification, severe tortuosity or diameter less than
`5.5 mm or 6 mm, respectively
`o Bulky calcified aortic valve leaflets in close proximity to coronary ostia
`6.0 Potential Adverse Events
`Potential risks associated with the overall procedure including potential access complications associated
`with standard cardiac catheterization, balloon valvuloplasty, the potential risks of conscious sedation
`and/or general anesthesia, and the use of angiography:
`• Death
`• Stroke/transient ischemic attack, clusters or neurological deficit
`• Paralysis
`• Permanent disability
`• Respiratory insufficiency or respiratory failure
`• Hemorrhage requiring transfusion or intervention
`• Cardiovascular injury including perforation or dissection of vessels, ventricle, myocardium or valvular
`structures that may require intervention
`• Pericardial effusion or cardiac tamponade
`
`5
`
`Page 5 of 30
`
`

`

`• Embolization including air, calcific valve material or thrombus
`• Infection including septicemia and endocarditis
`• Heart failure
`• Myocardial infarction
`• Renal insufficiency or renal failure
`• Conduction system defect which may require a permanent pacemaker
`• Arrhythmia
`• Retroperitoneal bleed
`• AV fistula or pseudoaneurysm
`• Reoperation
`• Ischemia or nerve injury
`• Restenosis
`• Pulmonary edema
`• Pleural effusion
`• Bleeding
`• Anemia
`• Abnormal lab values (including electrolyte imbalance)
`• Hypertension or hypotension
`• Allergic reaction to anesthesia, contrast media, or device materials
`• Hematoma
`• Syncope
`• Pain or changes at the access site
`• Exercise intolerance or weakness
`• Inflammation
`• Angina
`• Heart murmur
`• Fever
`Additional potential risks associated with the use of the THV, delivery system, and/or accessories include:
`• Cardiac arrest
`• Cardiogenic shock
`• Emergency cardiac surgery
`• Cardiac failure or low cardiac output
`• Coronary flow obstruction/transvalvular flow disturbance
`• Device thrombosis requiring intervention
`• Valve thrombosis
`• Device embolization
`• Device migration or malposition requiring intervention
`• Valve deployment in unintended location
`
`6
`
`Page 6 of 30
`
`

`

`• Valve stenosis
`• Structural valve deterioration (wear, fracture, calcification, leaflet tear/tearing from the stent posts,
`leaflet retraction, suture line disruption of components of a prosthetic valve, thickening, stenosis)
`• Device degeneration
`• Paravalvular or transvalvular leak
`• Valve regurgitation
`• Hemolysis
`• Device explants
`• Nonstructural dysfunction
`• Mechanical failure of delivery system, and/or accessories
`• Non-emergent reoperation
`7.0 Directions for Use
`
`7.1 Required Equipment
`
`
`Product Name
`
`20 mm System
`(9600CM20A)
`
`9600TFX (20 mm)
`
`Edwards SAPIEN 3
`Transcatheter Heart Valve
`Edwards Commander Delivery
`System*
`Edwards eSheath Introducer
`Set**
`
`Table 4
`23 mm System
`(9600CM23A)
`
`Model
`9600TFX (23 mm)
`
`26 mm System
`(9600CM26A)
`
`29 mm System
`(9600CM29A)
`
`9600TFX (26 mm)
`
`9600TFX (29 mm)
`
`9600LDS20
`
`9600LDS23
`
`9600LDS26
`
`9600LDS29
`
`914ES
`
`914ES
`
`914ES
`
`916ES
`
`Edwards Balloon Catheter
`
`9350BC16
`
`9350BC20
`
`9350BC23
`
`9350BC25
`
`
`
`
`
`Inflation devices provided by Edwards Lifesciences
`9600CR
`Edwards Crimper
`* Includes the Qualcrimp Crimping Accessory,2-piece Crimp Stopper and loader
`** Or other compatible sheath provided by Edwards Lifesciences
`
`
`Additional Equipment:
`• 20 cc syringe or larger (x2)
`• 50 cc syringe or larger
`• High-pressure 3-way stopcock (x2)
`• Standard cardiac catheterization lab equipment
`• Fluoroscopy (fixed, mobile or semi-mobile fluoroscopy systems appropriate for use in percutaneous
`coronary interventions)
`• Transesophageal or transthoracic echocardiography capabilities
`• Exchange length 0.035 inch (0.89 mm) extra-stiff guidewire
`• Temporary pacemaker (PM) and pacing lead
`• Sterile rinsing basins, physiological saline, heparinized saline, 15% diluted radiopaque contrast
`medium
`• Sterile table for THV and device preparation
`7.2 THV Handling and Preparation
`Follow sterile technique during device preparation and implantation.
`7
`
`Page 7 of 30
`
`

`

`7.2.1 THV Rinsing Procedure
`Before opening the valve jar, carefully examine for evidence of damage (e.g. a cracked jar or lid, leakage,
`or broken or missing seals).
`CAUTION: THVs from containers found to be damaged, leaking, without adequate sterilant, or
`missing intact seals must not be used for implantation.
`
`Step
`
`Procedure
`
`1
`
`2
`
`3
`
`Set up two (2) sterile bowls with at least 500 mL of sterile physiological saline to thoroughly rinse the
`glutaraldehyde sterilant from the THV.
`
`Carefully remove the THV/holder assembly from the jar without touching the tissue. Verify the THV
`serial identification number with the number on the jar lid and record in the patient information
`documents. Inspect the THV for any signs of damage to the frame or tissue.
`Rinse the THV as follows: Place the THV in the first bowl of sterile, physiological saline. Be sure the
`saline solution completely covers the THV and holder. With the THV and holder submerged, slowly
`agitate (to gently swirl the THV and holder) back and forth for a minimum of 1 minute. Transfer the
`THV and holder to the second rinsing bowl of physiological saline and gently agitate for at least one
`more minute. Ensure the rinse solution in the first bowl is not used. The THV should be left in the
`final rinse solution until needed to prevent the tissue from drying.
`
`CAUTION: Do not allow the THV to come into contact with the bottom or sides of the rinse bowl
`during agitation or swirling in the rinse solution. Direct contact between the identification tag and
`THV is also to be avoided during the rinse procedure. No other objects should be placed in the
`rinse bowls. The THV should be kept hydrated to prevent the tissue from drying.
`
`
`7.2.2 Prepare the Components
`Refer to the Edwards eSheath Introducer Set, Edwards Crimper and Edwards Balloon Catheter
`instructions for use for device preparation.
`
`Step
`
`1
`
`2
`3
`
`4
`
`5
`
`6
`
`7
`
`8
`
`9
`
`Procedure
`Visually inspect all components for damage. Ensure the Edwards Commander delivery system is fully
`unflexed and the balloon catheter is fully advanced in the flex catheter.
`
`WARNING: To prevent possible damage to the balloon shaft, ensure that the proximal end of the
`balloon shaft is not subjected to bending.
`Flush the flex catheter.
`Carefully remove the distal balloon cover from the delivery system.
`Remove the stylet from the distal end of the guidewire lumen and set aside. Flush the guidewire
`lumen with heparinized saline and insert the stylet back into the distal end of the guidewire lumen.
`Note: Failure to insert the stylet back into the guidewire lumen may result in damage to the lumen
`during crimping process.
`Place the delivery system into the default position and make sure that the flex catheter tip is covered
`by the proximal balloon cover. Unscrew the loader cap from the loader tube and flush the loader cap.
`Place the loader cap over the proximal balloon cover and onto the flex catheter with the inside of the
`cap oriented towards the distal tip.
`Fully advance the balloon catheter in the flex catheter.
`Peel off the proximal balloon cover over the blue section of the balloon shaft.
`Attach a 3-way stopcock to the balloon inflation port. Partially fill a 50 cc or larger syringe with
`15-20 mL diluted contrast medium and attach to the 3-way stopcock.
`Fill the inflation device provided by Edwards Lifesciences with excess volume relative to the indicated
`inflation volume. Lock the inflation device and attach to the 3-way stopcock.
`Close the 3-way stopcock to the Inflation device provided by Edwards Lifesciences and de-air the
`system using the 50 cc or larger syringe. Slowly release the plunger and leave zero-pressure in the
`system.
`
`8
`
`Page 8 of 30
`
`

`

`Step
`
`Procedure
`Close the stopcock to the delivery system. By rotating the knob of the inflation device provided by
`Edwards Lifesciences, transfer the contrast medium into the syringe to achieve the appropriate
`volume required to deploy the THV, per the following:
`
`10
`
`11
`
`Delivery System
`Model 9600LDS20
`Model 9600LDS23
`Model 9600LDS26
`Model 9600LDS29
`
`THV Size
`20 mm
`23 mm
`26 mm
`29 mm
`
`Inflation Volume
`11 mL
`17 mL
`23 mL
`33 mL
`
`
`
`Close the stopcock to the 50 cc or larger syringe. Remove the syringe. Verify that the inflation volume
`is correct and lock the Inflation device provided by Edwards Lifesciences.
`CAUTION: Maintain the Inflation device provided by Edwards Lifesciences in the locked
`position until THV deployment.
`
`1
`
`2
`
`3
`4
`
`5
`
`6
`
`7
`
`8
`
`9
`
`7.2.3 Mount and Crimp the THV on the Delivery System
`Step
`Procedure
`Set up two (2) additional sterile bowls with at least 100 mL of sterile physiological saline to thoroughly
`rinse the Qualcrimp crimping accessory.
`Completely submerge the Qualcrimp crimping accessory in the first bowl and gently compress it to
`ensure complete saline absorption. Slowly swirl the Qualcrimp crimping accessory for a minimum of
`1 minute. Repeat this process in the second bowl.
`Remove the THV from the holder and remove the ID tag.
`Attach the 2-piece crimp stopper to the base of the crimper and click into place.
`With the crimper in the open position, gently place the THV into the crimper aperture. Gradually crimp
`the THV until it fits into the Qualcrimp crimping accessory.
`Place the Qualcrimp crimping accessory over the THV making sure the THV is parallel to the edge of
`the Qualcrimp.
`Place the THV and Qualcrimp crimping accessory in crimper aperture. Insert the delivery system
`coaxially within the THV on the Valve Crimp Section (2-3 mm distal to the balloon shaft) with the
`inflow (fabric cuff end) of the THV towards the distal end of the delivery system.
`Crimp the THV until it reaches the Qualcrimp Stop located on the 2-piece Crimp Stopper.
`Gently remove the Qualcrimp crimping accessory from the THV. Remove the Qualcrimp Stop from
`the Final Stop, leaving the Final Stop in place.
`Fully crimp the THV until it reaches the Final Stop.
`NOTE: Ensure that the Valve Crimp Section remains coaxial within the THV.
`Repeat the full crimp of the THV two more times for a total of three full crimps.
`Pull the balloon shaft and lock in the default position.
`Flush the loader with heparinized saline. Immediately advance the THV into the loader until the
`tapered tip of the delivery system is exposed.
`CAUTION: To prevent possible leaflet damage, the THV should not remain fully crimped and/or
`in the loader for over 15 minutes.
`Attach the loader cap to the loader, re-flush the delivery system through the flush port and close the
`stopcock to the delivery system.
`Remove the stylet and flush the guidewire lumen of the delivery system.
`CAUTION: Keep the THV hydrated until ready for implantation.
`CAUTION: The physician must verify correct orientation of the THV prior to its implantation;
`its inflow (fabric cuff end) should be oriented distally towards the tapered tip.
`
`10
`
`11
`12
`
`13
`
`14
`
`9
`
`Page 9 of 30
`
`

`

`7.3 Valvuloplasty and THV Delivery
`Valvuloplasty and THV delivery should be performed under conscious sedation and/or general anesthesia
`with hemodynamic monitoring in a catheterization lab/hybrid operating room with fluoroscopic and
`echocardiographic imaging capabilities.
`Administer heparin to maintain the ACT at ≥ 250 sec during the procedure.
`CAUTION: Use of excessive contrast media may lead to renal failure. Measure the patient’s
`creatinine level prior to the procedure. Contrast media usage should be monitored.
`CAUTION: Procedure may require an arterial cut-down with surgical closure of the puncture site
`due to the size of the arteriotomy.
`7.3.1 Baseline Parameters
`Step
`Procedure
`
`1
`
`2
`
`3
`
`4
`
`Perform a supra-aortic angiogram with fluoroscopic view perpendicular to the aortic valve.
`
`Evaluate the distance of the left and right coronary ostia from the aortic annulus in relation to the THV
`frame height.
`
`Introduce a pacemaker (PM) lead until its distal end is positioned in the right ventricle.
`
`Set the stimulation parameters to obtain 1:1 capture, and test pacing.
`
`7.3.2 Valvuloplasty
`
`Refer to Edwards Balloon Catheter Instructions for Use (IFU) for information on device
`preparation and handling.
`Note: Rapid ventricular pacing should be performed when using the Edwards Balloon Catheter for
`valvuloplasty prior to aortic transcatheter valve implantation.
`
`After placement of the balloon at the intended site, begin rapid ventricular pacing. Once the systolic blood
`pressure has decreased to 50 mmHg or below, balloon inflation can commence.
`
`1
`
`2
`
`CAUTION: THV implantation should not be carried out if the balloon cannot be fully inflated during
`valvuloplasty.
`7.3.3 THV Delivery
`Step
`Procedure
`Prepare and insert the Edwards eSheath Introducer Set. Refer to the Edwards eSheath Introducer Set IFU
`for information on device preparation and handling.
`Insert the loader into the sheath until the loader stops.
`Advance the Edwards Commander delivery system, with the Edwards logo facing up, through the sheath
`until the THV exits the sheath. Retract the loader to the proximal end of the delivery system.
`NOTE: Maintain the proper orientation of the flex catheter (with the Edwards logo facing up)
`throughout the procedure.
`CAUTION: If accessing femorally or via the iliac, the THV should not be advanced through the sheath
`if the sheath tip is not past the aortic bifurcation.
`CAUTION: To prevent possible leaflet damage, the THV should not remain in the sheath for over
`5 minutes.
`
`3
`
`10
`
`Page 10 of 30
`
`

`

`Step
`
`4
`
`5
`
`6
`
`7
`
`8
`
`9
`
`10
`
`11
`
`Procedure
`In a straight section of the aorta, initiate valve alignment by disengaging the Balloon Lock and pulling the
`balloon catheter straight back until part of the Warning Marker is visible. Do not pull past the Warning
`Marker.
`WARNING: To prevent possible damage to the balloon shaft, ensure that the proximal end of the
`balloon shaft is not subjected to bending.
`Engage the Balloon Lock.
`Use the Fine Adjustment Wheel to position the THV between the valve alignment markers.
`CAUTION: Do not turn the Fine Adjustment Wheel if the Balloon Lock is not engaged.
`WARNING: Do not position the THV past the distal Valve Alignment Marker. This will prevent proper
`valve deployment.
`CAUTION: Maintain guidewire position in the left ventricle during valve alignment.
`Advance the catheter and use the flex wheel, if needed, and cross the aortic valve.
`NOTE: Verify the Edwards logo is facing up. The delivery system articulates in a direction opposite
`from the flush port.
`
`If additional working length is needed, remove the loader by unscrewing the loader cap and peeling the
`loader tubing from the delivery system.
`
`Disengage the Balloon Lock and retract the tip of the Flex Catheter to the center of the Triple Marker.
`Engage the Balloon Lock.
`Verify the correct position of the THV with respect to the aortic annulus.
`As necessary, utilize the Flex Wheel to adjust the co-axiality of the THV and the Fine Adjustment Wheel to
`adjust the position of the THV.
`Before deployment, ensure that the THV is correctly positioned between the Valve Alignment Markers and
`the Flex Catheter tip is over the Triple Marker.
`Begin THV deployment:
`• Unlock the Inflation device provided by Edwards Lifesciences.
`• Begin rapid pacing; once systolic blood pressure has decreased to 50 mmHg or below, balloon inflation
`can commence.
`• Deploy the THV by inflating the balloon with the entire volume in the Inflation device provided by
`Edwards Lifesciences, hold for 3 seconds and confirm that the barrel of the inflation device is empty to
`ensure complete inflation of the balloon.
`• Deflate the balloon. When the balloon catheter has been completely deflated, turn off the pacemaker.
`
`7.3.4 System Removal
`Step
`Procedure
`Unflex the delivery system while retracting the device, if needed. Verify that the Flex Catheter tip is
`locked over the Triple Marker and remove the delivery system from the sheath.
`CAUTION: Patient injury could occur if the delivery system is not unflexed prior to removal.
`
`1
`
`2
`
`3
`
`Remove all devices when the ACT level is appropriate. Refer to the Edwards Expandable Introducer
`Sheath Set instructions for use for device removal.
`
`Close the access site.
`
`8.0 How Supplied
`STERILE: The THV is supplied sterilized with glutaraldehyde solution. The delivery system is supplied
`sterilized with ethylene oxide gas.
`8.1 Storage
`The THV must be stored at 10 °C to 25 °C (50 °F to 77 °F). Each jar is shipped in an enclosure containing
`a temperature indicator to detect exposure of the THV to extreme temperature.
`The delivery system should be stored in a cool, dry place.
`11
`
`Page 11 of 30
`
`

`

`9.0 MR Safety
`
` MR Conditional
`The Edwards SAPIEN 3 transcatheter heart valve has been determined to be MR Conditional. A patient
`with this device can be safely scanned in an MR system meeting the following conditions:
`
`• Static magnetic field of 1.5 tesla or 3 tesla.
`
`• Maximum spatial gradient field of 2500 gauss/cm (25 T/m) or less.
`
`• Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 2 W/kg
`(Normal Operating Mode)
`Under the scan conditions defined above, the SAPIEN 3 transcatheter heart valve is expected to produce
`a maximum temperature rise of 3.0 ºC after 15 minutes of continuous scanning.
`
`In non-clinical testing, the image artifact caused by the device extends as far as 10 mm from the implant
`for spin echo images and 30 mm for gradient echo images when scanned in a 3.0T MRI system.
`
`The implant has not been evaluated in MR systems other than 1.5 or 3.0T.
`10.0 Patient Information
`Patient education brochures are provided to each site and should be given to the patient to inform them of
`the risks and benefits of the procedure and alternatives in adequate time before the procedure to be read
`and discussed with their physician. A copy of this brochure may also be obtained from Edwards
`Lifesciences by calling 1.800.822.9837. A patient implant card request form is provided with each
`transcatheter heart valve. After implantation, all requested information should be completed on this form.
`The serial number may be found on the package and on the identification tag attached to the transcatheter
`heart valve. The original form should be returned to the Edwards Lifesciences address indicated on the form
`and upon receipt, Edwards Lifesciences will provide an identification card to the patient.
`11.0 Recovered THV and Device Disposal
`The explanted THV should be placed into a suitable histological fixative such as 10% formalin or 2%
`glutaraldehyde and returned to the company. Refrigeration is not necessary under these circumstances.
`Contact Edwards Lifesciences to request an Explant Kit.
`
`Used delivery system may be disposed of in the same manner that hospital waste and biohazardous
`materials are handled. There are no special risks related to the disposal of these devices.
`12.0 Clinical Studies
`SUMMARY OF PRIMARY CLINICAL STUDY
`
`Primary Clinical Study Design
`
`The PIIS3HR Cohort of the PARTNER II trial was a single arm, non-randomized, historical-controlled
`study to compare the third generation Edwards SAPIEN 3 THV system with the first generation Edwards
`SAPIEN THV system in patients who either have high risk for surgery or cannot undergo surgery
`(inoperable). The valve sizes used in the PIIS3HR trial included only the 23, 26 and 29 mm sizes. The 20
`mm valve size was introduced into the trial after enrollment was completed with the three larger sizes,
`thus a separate nested registry, NR7, with identical inclusion/exclusion criteria as the PIIS3HR Cohort
`except for the aortic annulus diameter, was created to collect data for the 20 mm valve. Data from the
`PIIS3HR cohort and NR7 are pooled for the statistical analyses. For convenience, this combined cohort is
`referred to as “PIIS3HR” hereafter.
`
`The database included 583 eligible patients enrolled at 29 investigational sites in the U.S.
`
`The study used an independent Data Safety Monitoring Board (DSMB) that was instructed to notify
`Edwards Lifesciences of any safety or compliance issues, a Clinical Events Committee (CEC) that was
`responsible for adjudicating endpoint related events reported during the trial per a priori established
`
`12
`
`Page 12 of 30
`
`

`

`VARC 2 definitions,[1] an ECG core laboratory for independent analysis of rhythm, and an
`echocardiographic core laboratory for independently analyzing all echocardiograms.
`
`Accountability
`
`All 583 eligible patients were successfully implanted with a SAPIEN 3 THV, which constitutes the Valve
`Implant (VI) population. Among the VI population, 491 patients were implanted via the transfemoral (TF)
`access route, and 92 patients via the transapical (TA) or transaortic (TAo) access route.
`Table 5: Patient Accountability
`Overall
`TF
`491
`583
`
`TA /TAo
`92
`
`
`Eligible Patient
`Population (EPP)
`Valve Implant
`92
`491
`583
`(VI) Population
`Eligible Patient Population (EPP) consists of all enrolled patients who received
`treatment assignment from the database and entered into the catheterization
`laboratory/hybrid suite and who remained eligible to receive the implant.
`
`Valve Implant (VI) Population consists of all enrolled patients who received a
`SAPIEN 3 implant, and retained the valve upon leaving the catheterization
`laboratory/hybrid suite.
`
`
`Study Population Demographics and Baseline Parameters
`
`The demographics of the study population are summarized in Table 6, which are typical of a TAVR study
`performed in the U.S.
`
`
`Table 6: Patient Demographics and Baseline Characteristics - PIIS3HR VI Population
`Overall
`TF
`TA/TAo
`Characteristic
`(N= 583)
`(N= 491)
`(N= 92)
`Age, yr
`82.6 ± 8.1
`82.8 ± 8.2
`81.7 ± 7.5
`Male sex, no. (%)
`338 (58.0%) 277 (56.4%) 61 (66.3%)
`STS score
`8.6 ± 3.7
`8.4 ± 3.5
`10.0 ± 4.3
`New York Heart Association (NYHA) class, no. (%):
`
`
`
` I/II
`58 (9.9%)
`51 (10.4%)
`7 (7.6%)
` III/IV
`525 (90.1%) 440 (89.6%) 85 (92.4%)
`Coronary artery disease, no. (%)
`444 (76.2%) 360 (73.3%) 84 (91.3%)
`Previous myocardial infarction, no. (%)
`117 (20.1%)
`87 (17.7%)
`30 (32.6%)
`Previous intervention, no. (%)
`
`
`
` Coronary-artery bypass grafting (CABG)
`193 (33.1%) 145 (29.5%) 48 (52.2%)
` Percutane

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket