`Technology Conference
`Warsaw, Poland 13 May 200
`
`Design of a Low-cost Physiological Parameter Measurement and Monitoring Device
`
`G. Sen Guptal, S.C. Mukhopadhyayl, B.S. Devlinl and S. Demidenkol
`1School of Electrical and Electronic Engineering, Singapore Polytechnic, 500 Dover Road, Singapore
`Zlnstitute of Information Sciences and Technology, Massey University, Palmerston North, New Zealand
`Email: s.c.mukhopadhyay@massey.ac.nz, SenGupta@sp.edu.sg, b.s.devlin@massey.ac.nz, s. demidenko@massey. ac.nz
`
`Abstract — In this paper we present the design of a low-cost system
`that can be used to monitor physiological parameters, such as
`temperature and heart rate, of a human subject. The system consists
`ofan electronic device which is worn on the wrist andfinger, by an
`elderly or at-risk person. Using several sensors to measure different
`vital signs, the person is wirelessly monitored within his own home.
`An impact sensor has been used to detectfalls. The device detects
`a person is medically distressed and sends an alarm to a receiver
`unit that is connected to a computer. This sets off an alarm, allowing
`help to be provided to the patient. The device is battery poweredfor
`use indoors. The device can be easily adapted to monitor athletes
`and infants. The low cost ofthe device will help to lower the cost of
`home monitoring ofpatients recovering from illness. A prototype of
`the device has been fabricated and extensively tested with very good
`results.
`
`— physiological
`Keywords‘
`transmission, home monitoring
`
`parameters,
`
`S6'l’lS0?‘S,
`
`wireless
`
`I.
`
`INTRODUCTION
`
`Many elderly people dread the idea of being forced to live
`with their adult children, or
`in a rest home or in other
`sheltered
`living
`arrangement.
`They want
`to
`live
`independently and keep control of their own lives. Yet at the
`same time they know there is a high risk of injury or even
`death because of a fall or stroke. With the population aging
`in most developing countries, there will be more and more
`elderly people living alone in future. Such people need to be
`monitored continuously and provided with immediate
`medical help and attention when required.
`The cost of hospitalization is ever increasing, so is the cost
`of rehabilitation after a major illness or surgery. Hospitals are
`looking at sending people back as soon as possible to recoup
`at home. During this recovery period several physiological
`parameters need to be
`continuously measured. Hence
`telemedicine and remote monitoring of patients at home are
`gaining added importance and urgency [l—3]. Today,
`the
`progress in science and technology offers miniaturization,
`speed, intelligence, sophistication and new materials at lower
`cost. In this new landscape, micro—technologies, information
`technologies and telecommunications are the key factors in
`inventing devices to assist mankind. Patients are being
`monitored using a network of wireless sensors
`A system
`to monitor the overall health of welfare facility residents, who
`need constant care, has been reported in
`This system [5]
`has been designed with wireless sensors, wireless repeaters
`and a host computer. The system consists of a piezoelectric
`
`1-4244-0589-0/07/$20.00 ©2007 IEEE
`
`1
`
`sensor, a 2 axis accelerometer, a microcontroller and a low
`power
`transceiver.
`It
`records
`respiration activity and
`indicators of posture for 24 hours. These data are transmitted
`to the wireless repeater by the transceiver. The wireless
`repeaters, which are installed throughout the welfare facility,
`send data, including the repeater's ID, to the host computer.
`The ID is used to detect the resident's location in the welfare
`
`facility. The host computer stores the data, which can be used
`to analyze the resident's overall health condition. When the
`resident is in an emergency situation, such as falling or in an
`inactive state for more that
`the allotted time,
`the host
`computer automatically alerts the situation to the care staff by
`an alarm sound and also by mobile phone. However, all
`reported systems are relatively expensive.
`In the back drop of the importance of continuous
`monitoring of vital physiological parameters of a patient, our
`research was undertaken to design a
`low-cost
`smart
`monitoring device. It aims to provide peace-of-mind to users
`who have medical problems, but are not placed in a hospital
`for monitoring. Caregivers, who look after patients with
`mental or physical disabilities, can use this device when they
`are not able to visually supervise the patients.
`Currently there are monitoring products in the market that
`are aimed to provide emergency assistance to senior citizens,
`rehabilitation
`patients,
`and medically
`or
`physically
`challenged individuals, but these have limitations. St Johns
`and Medic Alert’s LifelinkTM [6] allows the user to set off an
`alarm manually if they are under medical stress, which will
`then dial designated contact phone numbers. The fundamental
`problem with this system is that when medical emergencies
`happen to the user, they are often unconscious and unable to
`press an ‘emergency alert button’. There is no product on the
`market which does not require manual activation of the alarm
`and monitors a user’s vital signs smartly. This is the novel
`design goal of the work presented in this paper.
`The reported device consists of a wrist strap and a finger
`glove. This allows the sensors to be mounted around the wrist
`and finger. A battery and a microcontroller, with built-in RF
`transceiver, are mounted within the wrist strap as well. In
`Section II we present the complete system overview. All the
`sensors are explained in Section III. The hardware details are
`in Section IV and the algorithms in Section V. The prototype
`and test results are discussed in Section VI. The paper ends
`with a discussion on future developments
`
`Apple Inc.
`APLl045
`
`U.S. Patent No. 8,923,941
`
`Apple Inc.
`APL1045
`U.S. Patent No. 8,923,941
`
`
`
`II. SYSTEM OVERVIEW
`
`B. HearlRale Sensor
`
`The system has been designed to take several inputs from
`a human subject to measure physiological parameters such as
`temperature and heart rate. Figure 1 shows the functional
`block diagram of the system hardware. The inputs from the
`sensors are processed and the results transmitted to a receiver
`unit, which is connected to a computer placed in the home,
`using Radio Frequency (RF) wireless
`technology. The
`receiver unit decodes and analyses the data. If it is inferred
`that the person is medically distressed, an alarm is generated.
`The design is modular which makes it rather easy and straight
`forward to add extra sensors for measuring and monitoring
`other parameters. The hardware blocks are explained in full
`details in a later section.
`
`A custom heart rate sensor was designed to read the
`patient’s beats per minute (bpm). The designed sensor is very
`small and inexpensive. The technique used to measure the
`hearl: rate is based on near—infrared (NIR) spectroscopy. NIR
`spectroscopy involves using light in the wavelength of 700-
`900nm to measure blood volume. At these wavelengths most
`tissues do not absorb light — other than hemoglobin (which is
`what we are interested in). This allowed for designing a non-
`invasive and low cost method of measuring the pulse. A
`silicon phototransistor and a GaAs infrared emitting diode
`were used in the sensor, moulded into a flat side-facing
`package. The amount of light
`that was detected by the
`phototransistor varied with the patients heart pulse, as the
`amount of absorbed IR light changed with the flow of blood,
`which is directly linked to the heart rate. This signal was then
`amplified,
`filtered, and sent
`to the microcontroller to be
`analyzed. The heart rate sensor was mounted in the finger
`glove as this position proved to give the best response.
`
`An ADXL3ll accelerometer was used as an impact
`sensor. It provides a 2-axis response, measuring accelerations
`up to +/- 2g.
`It was fitted into the wrist
`strap. The
`accelerometer provides an analog voltage, the amplitude of
`which is directly proportional to acceleration. This signal was
`scaled down to bring it within the acceptable input range of
`the micro-controller, and then analyzed. Software algorithms
`were used to detect sharp impacts, while allowing slower
`movements, such as walking, to be ignored. The purpose of
`this sensor was to detect sudden impacts that could indicate
`the patient had fallen over.
`
`IV. HARDWARE DESIGN
`
`The hardware was built in three separate blocks. A sensor
`card was designed to house
`the
`temperature
`sensor,
`accelerometer and the connections for the NIR emitter and
`
`the
`detector. A separate analog card was designed for all
`analog processing circuitry needed for the sensors, primarily
`for processing the heart rate signal. The temperature and
`acceleration output was fed directly to the micro-controller.
`The micro-controller was mounted on a separate card which
`also had the antenna connection. The cards were connected
`
`by ribbon cables within the wrist strap. Figure 2 shows the
`circuit schematics of the sensor units and Figure 3 shows the
`analog processing circuit for measuring heart beat rate.
`
`A. Lock-In Amplzficalionfor heart rate measurement
`
`The micro-controller modulates
`
`the infra-red emitter
`
`signal at I KHZ through a npn transistor (Figure 2). This is
`then mixed with the signal obtained from the IR sensors
`(back scattered light). This technique, known as Lock-In
`Amplification
`involves phase
`sensitive
`detection.
`
`2 4r3Hz RF
`
`J
`
`THHF Mixea—signal
`r
`'
`micrucuritruller
`.
`I‘
`with RF
`#75 Transmitter
`Analog Signals
`
`M &
`
`Impact Sensor
`(Accelerometer)
`C‘""‘W
`
`Temperature
`Sensor
`Circuitry
`
`HEBIT RBTE
`Sensor
`Circuitry
`
`Fig. 1. Functional block diagram ofthc system hardware
`
`III. SENSORS AND INTERFACE
`
`The system consists of three sensors- a temperature sensor,
`heart rate sensor, and an impact sensor. All
`the sensor
`circuitries used in the design generate analog voltages which
`are fed to the ADC (Analog-to-Digital) inputs of the micro-
`controller. The ADC inputs
`are
`time-multiplexed and
`sampled at different rates. The description of individual
`sensors follows.
`
`A. Temperature Sensor
`
`The temperature measurement is done using a LM35 [7]
`precision integrated-circuit temperature sensor. It provides an
`accuracy of +/— 025°C within the desired temperature
`measurement range of 20—40°C. It has a very low current
`drain of 60 i.iA. This sensor is mounted within the wrist strap,
`positioned in such a way that it is in contact with the skin,
`allowing it to measure the extemal temperature of the skin.
`From the skin temperature, the body temperature is estimated.
`Because an exact measurement of body temperature is not
`required, this method is suitable. Rather, relative changes are
`monitored within set thresholds, which set off the alarm. This
`allows the device to detect changes in body temperature that
`could indicate the patient is undergoing any of the following
`conditions:
`trauma,
`injury, heart
`attack,
`stroke, heat
`exhaustion, and burns
`The temperature sensor is sampled
`once every 3 seconds.
`
`Mlflrflcumrullar
`Wm RF
`Receiver
`
`‘i5 PC
`R5232
`
`C. Impact Sensor
`
`
`
`Mflcfiflélfifllfiief
`
`‘Jl
`.~\:~>;L3l1
`
`L7
`
`
`
`;_‘fiis. _Jami
`3
`5
`
`1 gm
`
`.5-.r»=_'ug_‘J*;_Tu:np
`
`um;-.:
`
`ER signs! tmnsmééaer
`and receiver
`
`l KHZ
`Square
`
`SR7
`
`‘J_r_::1
`
`v
`3320
`
`v ' ‘.f'
`
` CT
`Llninamni tn s|a1ng;13»::»3;s3:ig
`
`
`,--' or
`cam
`
`
`
`Fig. 2. Sensor Units
`
`Maioy wmcia ajdemodelatingi
`”~{"(2(_?ca,
`
`<2 1“:
`
`
`
`
`
`Signal Samar
`
`Sigma! E:zuerfE?|x1 Gain;
`
`Ac couple and amaélfiexlian (mas gain}
`
`Fig. 3. Signal conditioning for Heart Rate Monitoring
`
`The signal of interest (1-2 Hz) is moved into a much higher
`frequency band together with the noise that is picked up by
`the sensors. Demodulation is performed by an analog switch,
`which demodulates at the same frequency and in phase with
`the l KHZ modulating signal. The resulting signal is then
`passed through a low-pass filter with a cut off frequency of
`6Hz. This allows the signal of interest to pass through, while
`filtering out all noise that originated from the sensor. This is
`illustrated in figure 4.
`
`B. Controller Cam’
`
`The controller used in the Wrist strap unit (and also in the
`receiver unit) is a Nordic nR_EF24El. This takes inputs from
`
`am
`
`Ear -hrmnaus
`
`El nplteiue
`
`'
`
`.,ew puss fiéaer
`
`E Eiflwsnvsrm!
`E
`
`‘-
`
`Fig. 4. Lock-In Amplification for heart rate measurement
`
`the sensor circuits in the form of analog voltages. Each sensor
`has a dedicated ADC channel which is multiplexed by the
`microcontroller. Each sensor’s signal
`is sampled at a pre-
`defined rate,
`through interrupt-driven algorithms. This
`microcontroller was chosen because of its small footprint
`(6mm X 6mm),
`low power consumption,
`and built-in
`
`
`
`transceiver. The microcontroller is responsible for producing
`the carrier signal that modulates the IR emitting diode, which
`is used in demodulation too. It is powered by a 9V battery,
`which is regulated down to SE’. The controller card contains
`the microcontroller, crystal oscillator, RF antenna connection,
`and header pins which connect analog and digital pcrts to the
`other components ofthe unit
`
`0. Receiver Um":
`
`sent by the nREF24El using a
`is
`The data packet
`technology termed ShockBurstTM. This allows data to be
`clocked into a FIFO buffer at a low data rate, and then
`transmitted all at once at a very high rate. This lowers power
`consumption by minimizing the amount of receiving and
`transmitting time. Data is transmitted by this method at
`1Mbps. This also reduces the risk of a data collisicn with
`other wrist units, or devices operating in the same band.
`
`V.
`
`SOI~T‘NA_t«’._|:, AND ALGORIIHMS
`
`The hardware of the receiver unit is housed in a plastic
`box and consists of a nREF24El microcontroller, antenna, a
`5V AC adaptor and serial interface port. The receiving unit "s
`connected, via a RS232 serial port, to a personal computer
`EC) and is constantly receiving information about
`tl'e
`patient’s medical
`status. A program, running on the PC,
`receives the packetized information from the serial port,
`decodes the packet and then displays this information on tle
`PC monitor. The program offers several opticns to set
`thresholds of the allowable range of heart
`rate,
`ski
`temperature and impact sensor. When the readings from tl'
`patient move outside ofthe set range, an alarm is raised. Tl’
`software can depict
`the change in data over
`time in a
`graphical plot. One receiver unit can be interfaced with
`multiple wrist units. In the software, one can tab through tle
`information received fi‘om each wrist unit.
`
`mmb
`
`D. Commi.mi'r:an'0n
`
`Communication between the wrist units and the receiver
`
`unit is wireless, powered by the nREF24El, and trananitted
`in the unlicensed 2.4GHz fi'e-quency band. Information is
`gathered every 3 seconds fi‘om the sensors and then encoded
`into a packet. Each packet
`is 6 bytes long, composed as
`shown in figure 5.
`
`Start of File
`
`-I—- flit balanced (1‘.}1010?IZI3
`
`the
`on
`implemented
`algorithms were
`Several
`from the
`take inputs
`microcontroller. These algcrithms
`sensors and process them into meaningful information.
`
`A. Heart Rate Algorfihm
`
`The heart rate signal is an oscillating signal ofl to 2Hz. This
`is sampled by the microcontroller at SOHZ, and then averaged
`over 5 samples, which gives the algorithm an effective data
`input rate of l 0 samples per second. This is stored in a buffer
`which is 32 bytes long. Once this buffer is fL1ll, the algorithm
`scans though each value in the buffer to compute the period
`of the signal (see figure 6). Since the signal is not sinusoidal,
`rather has distinct peaks corresponding to the contractions in
`the heart, the period is measured by marking transitions over
`the 25% mark. This is the most reliable part of the received
`signal. Since the sample time is know (Ills), the number of
`samples between these ‘transition points’ is used to calculate
`the period and hence the fi‘equency. The frequency is then
`multiplied by 60 to compute the heart rate (in bpm, beats per
`minute) of the patient
`ls. r.n::-' Il||.'lFu:ClF-
`
`l-R-arm’.
`
`.
`v-l,~’i
`'li'l'M'-e
`1!," 9.ar:r5il;=__".§:'Jn1r.ss|r.-1
`
`
`
`Fig. 6. Heartrate measurement algorithm
`
`5 Bytes tong
`
`Bi‘-.ha|:Incad[1D1D1D1.'}]
`
`Skin Temgmralme
`
`Impact |¢1+v:H'rr-.a1inr'.
`
`End of File
`
`Fig. 5. Datapacket composition
`
`.5’. Impact Algorithm
`
`is in beats per
`The heart rate included in the packet
`minutes (bpm) which is calculated using the Heart Rate
`Algorithm explained in Section V. The temperature data is
`the result ofthe ADC conversion, done once every 3 seconds,
`and is decoded into degrees Celsius at the receiver‘ unit. The
`impact informaticn is the maximum output voltage of the
`accelerometer measured in the past 3 seconds. This is also the
`digital value obtained from the ADC conversion.
`
`is a DC
`The signal received from the accelerometer
`voltage,
`the amplitude of which is directly related to the
`acceleraticn (62In‘Wg). This input is sampled at lOHz, and
`stored in a buffer which is l0 bytes long. Every second, this
`buffer is scanned, and the maximum amplib_1de is recorded.
`At
`the receiver unit,
`this data is compared to recorded
`representative values for walking and jogging. This ensures
`that an alarm is set off only for very sharp, heavy impacts.
`
`
`
`The reliability of impact detection was very good and the
`walking could easily be differentiated from a fall.
`
`with thumb on carotid artery). Figure 9 shows a segment of a
`received heart rate signal.
`
`VI. PROTOTYPE AND EXPERIMENTAL RESULTS
`
`The Nordic micro-controller development board was used
`to build and test the prototype design. The analog processing
`circuitry and the sensors were assembled on PCBs which
`were placed within the wrist strap. Figure 7 shows the
`prototype hardware. The prototype was powered oft‘ a 9V
`battery. The RF transmission has been tested to operate
`successfully at 10 meters range through obstacles such as
`concrete walls. The receiver unit. without the casing, can be
`seen in figure 8.
`
`Fig. 9. Segment of a heart rate signal
`
`B. Impact Sensor Test results
`
`The output of the accelerometer was tested with walking
`and simulated falling. The results showed the difference was
`simple to detect and proved the accuracy of the algorithm.
`Figure I0 shows the impact sen sor output.
`
`
`
`Fig. I0. Impact sensor output for walking and a fall
`
`VILDISCUSSIONS AND FUTURE DEVELPOMENTS
`
`In this paper we have presented the research, of applied
`nature, done to monitor physiological parameters such as skin
`temperature, heart rate and body impact. A prototype was
`successfiilly developed and tested to establish the proof of
`concept. The algorithms were tested and found to be accurate
`and reliable. The novel aspect ofthe design is its low cost and
`detection of medical distress which does not necessitate
`
`pressing any panic button. This is an enormous improvement
`over existing commercial products.
`An important aspect of the design was miniaturization, so
`that the system was as non-intrusive as possible to the wearer.
`This was achieved by the use of surface-mounted devices on
`the PCBs designed. Low power operational amplifiers were
`used to minimize battery consumption. The price of the unit
`currently is $70. The major costs come from the use of
`precision components, accelerometer and temperature sensor.
`With some modification, the system can be made available
`commercially. Future improvements will focus on the use of
`
`Fig. 8. Prototype receiver connected to a PC‘
`
`A. Heart Rate .S'er.'sor Test Results
`
`A new heart rate is calculated every 3.2 seconds. The
`output of the sensor is a 200-400 mV p-p signal, riding a DC
`signal of 500mV. On human tests it is able to measure within
`:l:2 bpm of a rested patient (confirmed by measuring pulse
`
`
`
`flexible PCBs to replace the stiff cards, so that it could be
`moulded around the wrist unit, making it more comfortable
`for the wearer.
`
`The design of the IR sensors could be improved to
`decrease its susceptibility to noise, to a point where it could
`be moved onto the wrist unit. This would provide a much
`more comfortable and less intrusive unit, getting rid of the
`finger glove.
`The addition of a blood-oxygen sensor would allow the
`system to detect medical distress more accurately by
`measuring the amount of oxygen in the blood (HbO). This
`could be implemented by the addition of another emitter
`diode operating at a wavelength which is more readily
`absorbed by oxygen, and measuring the light using photo
`detector. Blood pressure can also be measured by a technique
`known as ‘pulse delay’ [10] which involves calculating the
`time for the heart pulse to travel a known distance. This is
`directly related to the blood pressure of a patient, and allows
`for non—invasive measurement.
`
`The unit was initially designed for use by the elderly,
`within the house, where a caregiver is present but is not able
`to be constantly in visual contact with the subject. The
`receiver unit would be enhanced so that it can connect to
`
`either the local or cellular phone network, and in the case of
`an emergency would contact an ambulance. Beyond the
`application for elderly patients is the use by anyone who is at-
`risk, with a mental or physical disability. The device could be
`applied to prevent cot deaths in babies, by alerting the parents
`when the infant becomes stressed.
`
`[1]
`
`[2]
`
`[3]
`
`[4]
`
`[5]
`
`[6]
`
`[7]
`
`[8]
`
`[9]
`
`[10]
`
`REFERENCES
`
`Ohta S, Nakamoto H, Shinagawa Y, Tanikawa T., “A health
`monitoring system for
`elderly people
`living alone”,
`Joumal of
`Telemedicine and Telecare, Vol 8, No. 3, June 2002, pp. 151-156
`C.,
`Dittrnar
`A,
`Axisa
`F,
`Delhomme
`G,
`Gehin
`“New concepts and technologies in home care and ambulatory
`monitoring”, Studies in health technology and informatics, 2004, pp 9-
`35.
`Fazlur Rahman, Arun Kumar, G Nagendra, and Gourab Sen Gupta.
`"Network Approach for Physiological Parameter Measurement", IEEE
`Transactions on Instrumentation and Measurement, February 2005, Vol
`54, No.1, pp 337-345
`Jovanov E, Raskovic D, Price J, Chapman J, Moore A, Krishnamurthy
`A., “Patient Monitoring Using Personal Area Networks of Wireless
`Intelligent Sensors”, Biomedical Sciences Instrumentation, 2001, pp
`373-378
`
`Button,
`
`Maki H, Yonczawa Y, Ogawa H, Sato H, Hahn AW, Caldwell WM.,
`“A welfare facility resident care support system”, Biomedical Sciences
`Instrumentation, 2004, pp 480-483.
`LifeLink
`Panic
`http:/fwww.bgehome.com/hs_proteCtion.html#lifelink
`National Semiconductor, LM35 — Precision Centigrade Temperature
`Sensor, http://www.national.com/pfi'LM/'l_.M35 .html
`Y.C.Sydney, A-Z Health Guide from WebMD: Medical tests, Body
`Temperature,
`http:/fwww.webmd.com/hw/healthgguide_atoz/hw198785.asp
`Eastern Michigan University, Lock-In Amplification
`http:/fwww.physics.emich.edu/molab/lock-in/indeX.html
`W.D .Peterson., D.A.Sl<_ramsted., D.E.Glumac, Piezo Film Pulse
`Sensor,http:/fwwwphoenix.tc-ieee.orgf004_Piezo_Film_Blood_
`Flow_Sensor/Phoenix_PiezoPulse.htm
`
`overview,