throbber
Apple Inc.
`APL1017
`U.S. Patent No. 8,989,830
`
`0001
`
`

`
`/‘lnestliesioliniy
`
`THE JOURNAL or
`THE AMERICAN SOCIETY or ANESTHESIOLOGISTS, INC.
`Editor-in-Chief
`LAWRENCE]. SAIDMAN, M.D., San Diego. California
`Editors
`
`David E. Longnecker, M.D.
`julien F. Biebuyck, M.B., D.Phil.
`Philadelphia, Pennsylvania
`Hershey, Pennsylvania
`Dennis T. Mangano, Ph.D., M.D.
`john]. Downes, M.D.
`San Francisco, California
`Philadelphia, Pennsylvania
`Kai Rehder, M.D.
`H. Barrie Fairlcy, M.B., B.S.
`Rochester, Minnesota
`Stanford, California
`Donald R. Stanski, MD.
`Simon Gelman, M.D., Ph.D.
`Stanford, California
`Birmingham, Alabama
`Michael M. Todd, M.D.
`Carol A. Hirshman. M.D.
`Iowa City, Iowa
`Baltimore, Maryland
`Warren M. Zapol, M.D., Boston, Massachusetts
`
`Associate Editors
`
`Charles W. Buffington, M.D.
`Pittsburgh, Pennsylvania
`David H. Chestnut, M.D.
`Iowa City, Iowa
`_[efi‘rey E. Cooper, Ph.D.
`Boston, Massachusetts
`Dennis M. Fisher. M.D.
`San Francisco. Callf01‘nia
`Thomas F. Hornbein, M.D.
`Seattle. Washington
`
`Carl Lynch III, M.D., Ph.D.
`Charlottesville, Virginia
`Mervyn Maze, M.B., Ch.B.
`Stanford, California
`Henry Rosenberg, M.D.
`Philadelphia, Pennsylvania
`Gary R. Strichartz, Pli.D.
`Boston, Massachusetts
`
`Tony L. Yaksh, Ph.D.
`San Diego, California
`
`correspondence relating to editorial nianageiiient. and Letters to the Editor should be mailed to Lawrence
`Manuscripts for publication.
`_]_ soidman‘ M_D__ Anesthesiology, T-015, University of California—-San Diego, San Diego, California 92093. Books for review should
`be mailed to John J_ DOwne5_ M_D,, Department of Anesthesia. The Children's Hospital of Phiiadelphia, One Children's Plaza, 34th
`Street and Civic comer Boulevard. Philadelphia. PA 19104.
`_
`I
`‘
`-
`Au ariieies are accepted for publication with the understanding that they are contributed exclusively to this journal and become the
`property of the American society of Anesthesiologists, Inc. Stateriients or opinions eitpressed in the journal reflect the views of the
`Elu[hor(g) and do no; i-oprcscnt official policy of the American Society of Anesthesiologists unless so stated.
`For details and reeemmondations concerning the preparation of manuscripts see GUIDE FOR AUTHORS. which appears in the January
`and
`]
`’
`.
`_
`_
`_
`A-‘<]2l:l't3yrliis::1:and related correspondence should be addressed to Advertising Manager. ANBSTHESIOLOGV, J. B. Lippincott Company,
`Ea,“ Washington 5qua,.e_ piii|;idg|phia, Pa. 19105. Publication ofan advertisement in Ai~ir.sTHE.siot.ocr does not constitute endorsement
`by ANESTHESIOLOGY of the product described therein or of any representations made by the advertiser with respect to the product.
`SN 00033022) is published 13 times a year {monthly except September. in which two issues are published)
`ANESTHESIOLOGY (IS
`incott Company. at Downsville Pilte, Route 3, Box 20-B, Hagerstown,
`I‘ An sthesiOl03'l5l-5v 1"‘3-- by J‘ 3' UPI’
`.
`.
`.
`,
`‘'0
`h A
`'
`'
`'5 mcncan 30'3"‘-‘l-'9" °
`C
`East Washington Square, Philadelphia, PA 19105. Printed in the U. S. A. ((3 Copyright 1939
`T I
`MD 21740. B ‘
`e located at
`flfi
`,
`.
`.
`.
`by the AmericL;Srl.1'l§c§’5c;::IYCE:‘:{nenh¢5io]°gi5r5_ Inc. Second class postage paid at Hagerstown. MD. and at additional mailing offices.
`Notice of Chen
`of Address: All correspondence related to rnembershipand change of address should be mailed to the Executive
`EC
`515 Bussc Highway, Park Ridge. Illinois 60068. Correspondence regarding nonmember
`S
`-
`'
`fA sthesiologists.
`,
`_
`ecretary, American S0C|€1Y°
`“C
`'ncott/Harper. Downsville Pilte, Route 3, Box 20-B, Hagersrowm
`5
`l3
`’
`'
`ii
`1 t
`ournal Fulfillment Department. Lippi
`_
`I:I|a:;:‘;pltciio;i:7s‘l‘i{i;iuid C:"seI'18o3__;38_3030; in Ma,.Yiand_ call collect 301-824-7300. Correspondence regarding noninember subscriptions
`i
`_
`b “I: subscription rates and subscription orders, should be sent to Woodbell Incorporated, 4-22~l 1. Kitaltasai,
`_
`.
`.
`.
`]::"&lc?:-é3:‘:1;li‘r:lclt'lIti;:)gc:n;i3u‘;ri.tI::Sta: In India_ Nepal. Bangladesh, and Sri Lanlta contact Universal Subscription Agency Pvt. Ltd., 101-102
`c
`'
`.r. s l<
`, New Delhi—1 10017. India.
`_
`e
`_
`,
`_
`A:t::;li::i,5E:in;:i:i£iFroi}o5;aufts. $40.00 individual. $50.00 institution; all other countries exceptjapan. India. Nepal, Bangladesh, and
`,
`.
`-

`-
`‘
`‘l delive
`.
`,
`‘
`,
`_
`'
`available u
`it re uest.
`S
`La ya,
`_
`o
`1, 335.00 institution. Single copies $8 00 Rates for air mail
`ry _
`_
`_
`po
`q
`_
`-
`“Tithe gociefzF:)‘iil'i)c:1:i'vlhlemPublisher receives a request for replacement within 60 days of the mailing in the U. S. or within 90 days in
`A

`I
`-
`‘
`oiui-nod by the postal service as undeliverable, the issue will be replaced.
`.
`an other counmes and the Ongma Issue tlisiznestliesiology, Downsville Pike, Route 3, Box 20-B. I-lagerstown, MD 217-10.
`iei-no] or personal use. or the internal or personal use of specific clients. is granted by_The
`A h
`,
`,
`.
`Ag:-cggzztgglgiéf F;};oAt?]::F_¥e;Eo]ogi5|:§ for libraries and other users registered with the Copyright‘Clc-arance Center (CCC) Transactional
`R
`rtin S
`rovided that the base fee of $00-30 Per ‘3°PY- Plus 500-05 P" Page is Paid d'“°'c"lY ‘° Ccci 21 C°“ST955 3'-~ Salem-
`vicy
`Meiimiliigjgo iilgggeéggg/39 3{)o_30 + $00.05. This does not extend to other kinds of copying. such as copying for general distribution,
`for advertising or Promotional Purposes, for creating “'3” °°"“’°‘i"“’ “"°'k" °' f°' resale‘
`0002
`0002
`
`

`
`Jlnenfiesiollm
`
`January 1989
`
`CONTENTS
`
`EDITORIAL VIEWS
`
`A Change in Format for ANESTHESIOLOGY.
`Lawrence _]. Saidman
`
`Changing Perspectives in Monitoring Oxygenation
`H. Barrie Fairley
`
`Studies in Animals Should Precede Human Use of Spinally Administered Drugs
`Tony L. Yaltslz and j. G. Collins
`
`CLINICAL INVESTIGATIONS
`
`The Influence o
`
`Pipecuroniu
`Oxide.
`
`f Renal Failure on the Pharmacokinetics and Duration of Action of
`In Bromide in Patients Anesthetized with Halothane and Nitrous
`
`well, P. Claver Canjell, Kay P. Caszagnoli, Daniel P. Lynam Mark R
`James E. Cold
`Fahey, Dennis M. Fisher, and Ronald D. Miller
`Effect of Intercostal Nerve Blockade on Respiratory Mechanics and CO2 Chemosen-
`sitivity at Rest and Exercise.
`Bernice R. Heclzer, Robert Bjurstrom, and Robert B. Schoene
`mic, Electrocardiographic, Mechanical, and Metabolic In-
`Comparison of Hernodynative Myocardial Ischemia in Vascular Surgical Patients with
`dicators of Intraopera
`Coronary Artery Disease.
`r, Margarela Ostrnan, Arnold Friedman, George Diamond,
`Slfren Hdggmark, Per Hohne
`Edward Lowenstein, and Sebastian Rel:
`mall Doses of Sufentanil or Fentanyl: Dose Versus EEG
`d Thiopental Requirement.
`
`Induction of Anesthesia with S
`
`Relationship of Mivacurium Chloride in Humans during Nitrous
`The Dose-Response
`ide—Enflurane Anesthesia.
`Oxide—FentanyI or Nitrous Ox
`n B. Kills, Tom Heéer, Marl: R. Fahey, Daniel P. Lynam, and
`james E. Caldwell, joh
`Ronald D. Miller
`
`janlce Mei-Ll Wang, Donnell ]. Creel, and K. C. Wong
`
`(Gontinuetl on page 5)
`
` —
`
`0003
`
`

`
`January 1989—ANESTI-IESIOLOGY
`
`CONTENTS
`
`eal Contractility Predicts Movement during Skin Incision in Patients
`Lower Esophag
`h Halothane, but Not with Nitrous Oxide and Alfentanil.
`Anesthetized wit
`Daniel I. Sessler, Randi Swen, Christine 1. Olcfsson, and Franklin Chow
`
`42
`
`Determination of Intra-abdominal Pressure Using a Transurethral Bladder Catheter:
`Clinical Validation of the Technique.
`Thomas], mam‘, Charles E. Lieber, and Ernest Benjamin
`
`(Continued from page 3)
`
`LABORATORY INVESTIGATIONS
`
`Epidural Clonicline Analgesia in Obstetrics: Sheep Studies.
`james C. Eisenach, Maria I. Castro, David M. Dewan, andjames C. Rose
`
`The Enhancement of Proton/Hydroxyl Flow across Lipid Vesicles by Inhalation An-
`esthetics.
`
`Dgugfas E. Raine: and David S. Cafiso
`The Influence of Dextrose Administration on Neurologic Outcome after Temporary
`al Cord Ischemia in the Rabbit.
`Spin
`john C. Dzramanand and Suzanne S. Moore
`Tachyphylaxis to Local Anesthetics Does Not Result from Reduced Drug Effectiveness
`at the Nerve Itself.
`pm,» Lgpfgrz, Halger Halthnse
`Comparison of the Effects of I-Ialothane on Skinned Myocardial Fibers from Newborn
`and Adult Rabbit. I. Effects on Contractile Proteins.
`Elliot]. Ksrane andjudy 1’. 31%
`
`n, and joachirn 0. Anna:
`
`Effects of “Nitrendipine" on Nitrous Oxide Anesthesia, Tolerance, and Physical De-
`pendence.
`S. Dalin and H. LEW?
`
`MEDICAL INTELLIGENCE ARTICLE
`
`Pulse Oximetry.
`Kevin K. Tremper and Steven Bari???
`
`_
`
`(Cmllilillcd on page 7}
`
`0004
`
`

`
`.]'=1nu:lI‘y' I98£)—/\NES'I"l-IESI(')L()GY
`
`CONTENTS
`
`(Continued from page 5)
`
`LABORATORY REPORTS
`
`Laudanosine Does Not Displace Receptor-specific Ligands from the Benzodiazepinergic
`or Muscarinic Receptors.
`Yeshayahu Kat: and Masha Gavish
`
`Effects of Methemoglobinemia on Pulse Oxirnetry and Mixed Venous Oximetry
`Steven j. Barker, Kevin K. Trernper, and john Hyagg
`Hyperbilirubinemia Does Not Interfere with Hemoglobin Saturation Measured by Pulse
`Oximetry.
`Francis Veyckemans, Philippe Basie, ]. E. Guillaume, Eric Wiliems, Anm-‘g Roger; and
`Thierry Clerbaux
`
`Evaluation of a Blood G35 and Chemistry Monitor for Use during Surgery
`G. Bashein, Wesley K. Greydanns, and Margaret A. Kenny
`
`A Model for Determining the Influence of Hepatic Uptake of Nondepolarizing Muscie
`Relaxants in the Pig.
`Johann Motsch, Pim Hennis, Franz Alta Zimmermann, and Sander Agoszon
`Transtracheal Doppler: A New Procedure for Continuous Cardiac Output Measure-
`ment.
`
`jerome H. Abrams, Roland E. Weber, and Kenneth D. Holman
`
`CASE REPORTS
`
`tracranial Pressure during Hepatic Transp]an_
`
`D. Brajzbard, R. 1'. Parks, M. A. Ramsay,
`and G. B. Klintmalm
`Treatment of Isorhythmic A-V Dissociation during General Anesthesia with Propran-
`010].
`
`A. W. Paulsen, T. R. Vaiek, T. H. Swygerz,
`
`Russell F. Hill
`
`Fiberoptic Endobronchial Intubation for Resection ofan Anterior Mediastinal Mass.
`Dirk Yaunker, Randal! Clark, and Lewis Coveler
`
`o
`
`0005
`
`146
`
`(Continued on page 9)
`
`

`
`January 1989-AN ESTH F.SI(')[.OGY
`
`CONTENTS
`
`Caudal Epidural Anesthesia in an Infant with Epidermolysis Bullosa
`
`'
`
`((3ontiI1ned from page 7;
`
`149
`
`Lawffflfif L Yr‘-’€. joei B. Garner, and Charies B. Maniey
`
`Recurrent Respiratory Depression after Alfentanil Administration
`Rory S. jafe and Dennis Coaison
`
`1 ocame.
`Pain of Delayed Traumatic Splenic Rupture Masked by Intrapleural L.d
`.
`Gregory A. Weaver
`0’
`Omen” 8’ Slang H’ Thong’ James 31- Raflfldiak, and
`Wiiiiam W. Pond, Grego M. 3
`‘,5;
`r
`Dose-res onse Relationshi
`for S
`'
`1 h '
`-
`-
`.
`.
`LOVE Plasma Cholinegteraseuffdiiiiriiyfmne In a Patient with Geneu':allY Detemliflfid
`Charles E. Smith, Geraint Lewis, Francois Donati, and David R Bevan
`
`CORRESPONDENCE
`
`Determination of Decay Constants from Time-varying Pressure Data
`Charies Beat:-ie, Linda S. Humphrey, and Gary Mafugchak
`Reply. Ciifiird R. Swanson and Wiiiiarn W. Muir 11;
`
`IOU.
`Use Caution when Extrapolating from a Small Sample Size to the General Populat.
`David]. Benefiei, Edward A. Eisier, and Rodger Shepherd
`Reply. Irnad H. Abdul-Rasool, Daniel H. Sears, and Ronaid L. Katz
`
`Succinylcholine and Trismus.
`Frederic A. Berry and Carl Lynch HI
`
`An Infant Model to Facilitate Endotracheal Tube Fixation in the Pediatric ICU Patient
`Patrick K. Birmingham and Babette Horn
`
`An Alternative Method for Management of Accidental Dural Puncture for Labor and
`Delivery.
`Shani Cohen, Jonathan S. Daitch, and Paul L. Goidiner
`
`High-press
`Michae
`
`ure Uterine Displacement.
`i
`Dorsey and Waiter L. Miiiar
`
`Calculating the Potency of Mivacurium.
`Aaron F. Koprnan
`Reply. john]. Savarese
`
`166
`
`166
`
`(Continued on page 1 1)
`
`e_—
`
`0006
`
`

`
`january ]989——A N ICSTI-l ESIO LOGY
`
`CONTENTS
`
`Midazolam in a Malignant Hyperthermia-susceptible Patient.
`juliana H.j. Breaks
`
`(Cmitiliued from page 9)
`167
`
`Exchange Autotransfusion Using the Cell Saver during Liver Transplantation
`Mare R. Brown, Michael A. E. Ramsay, and Thomas H. Swygen:
`
`Air Entrainment Through a Multiport Injection System.
`Dean Gilbert, Theodore]. Sanford, _]r., and Brian L. Partridge
`Reply. Thelma Masada
`
`A Tracheal Tube Extension for Emergency Tracheal Reanastomosis.
`Robert S. Holzman
`
`The Relationship Between Malignant Hyperthernia and Neuroleptic Malignant Syn-
`drome.
`
`Haggai Hernzesh, Dov Aizenberg, Marga Lapidot, and Hanan Munit:
`Reply. Stanley N. Carefi Stephan C. Mann, Henry Rosenberg, jejfrey E. Ftetche-r, and Terry
`D. Heiman-Patterson
`
`Appropriate Facilitation of Intravenous Regional Techniques in RSD.
`Kevin Foley, Linda Schatz, and Randall L. Martin
`
`REPORT OF SCIENTIFIC MEETING
`
`ANNOUNCEMENT
`
`GU11)!-I 'l'(_) AU'I‘l'I()R5
`
`The Guide for Authors is published in the January and July issues. It may be found on page 33A of this
`issue.
`
`A N.'i.‘§.»!V it re rrirlur mm?!‘ ("rar.I'mI"J us.-'.r.I" try the (Ihe':'m'rnl' .-H1\'.'r(:rr .‘s'rr::fr:* tn iI'r!'m.".._",-‘j' HIP ;'nm'nal'.
`
`-
`
`0007
`
`

`
`This material may be protected by Copyright law (Title 17 U.S. Code)
`
`0008
`
`

`
`Anesthesiology
`\"".l'0, No l,_]an 1989
`
`PULSE OXIMIETRY
`
`became a standard clinical and laboratory tool in pul-
`monary medicine. Although it was demonstrated to be
`accurate for intraoperative monitoring.” its size and ex-
`pense. and the cumbersome nature of the ear probe pre-
`vented its acceptance as a routine monitor. At this time.
`all oximeters produced various light source wavelengths
`by filtering white light. The filtered light was then trans—
`mitted to and from the tissue through fiberoptic cables.
`In the mid 1970s, Takuo Aoyagi. an engineer working
`for Nihon Kohden Corporation. made an ingenious dis-
`covery regarding oximetry. He was developing a tnethod
`to estimate cardiac output semi-noninvasively by detecting
`the washout curve of dye injected into a peripheral vein
`as it perfused the car. This washout curve was measured
`in the ear with a red and infrared light densitometer sim-
`ilar to the Millikan ear oximeter. He noticed that his
`
`washout curves contained pulsations due to the arterial
`pulse in the ear. To more easily analyze the dye washout
`curve, he subtracted these pulsations from the curve, and
`in doing so he discovered that the absorbance ratio of the
`pulsations at the two wavelengths changed with arterial
`hemoglobin saturation. He soon realized that he could
`build an ear oximeter that measured arterial hemoglobin
`saturation without heating the ear by analyzing pulsatile
`light absorbances.2 This first pulse oximeter, developed
`by Nihon Kohden, used filtered light sources similar to
`MiIlikan’s ear oximeter. The device was evaluated clini-
`cally in the mid 1970s and marketed with little success.2
`In the late 1970s. Scott Wilber in Boulder. Colorado,
`developed the first clinically accepted pulse oximeter by
`making two modifications of
`the Nihon Kohden
`method.“:{: First, he produced a lightweight sensor by
`using light emitting diodes (LEDs} as light sources and
`photodiodes as detectors. Consequently, the instrument
`was connected to its earciip sensor only by a small electrical
`cable. Wilber also improved the saturation estimates by
`using a digital microprocessor to store a complex calibra-
`tion algorithm based on human volunteer data”: This
`method will be discussed in more detail below. This device
`
`was developed by Biox Corporation of Boulder. Colorado,
`and was successfully marketed to pulmonary function lab-
`oratories in the early 1980s.
`The clinical utility of the noninvasive oximeter in the
`operating room was rediscovered in the 19805 by William
`New, an anesthesiologist at Stanford University. Realizing
`that a continuous, noninvasive monitor of oxygenation
`would be useful to anesthesiologists, New developed and
`marketed a pulse oximeter to this group.” The Nellcor
`model N100 had by 1985 become almost synonymous
`with the term "pulse oximeter.“
`
`i Wilber 5: Blood constituent measuring device and method. US
`Patent #4, 407. 290 April 1. 1981.
`
`%0‘sJt‘rt.mAr.«a,v
`
`FIG. 1. Thisfigure is from a 1951 article in ANES'I‘HESi0l.0GY. It
`reveals dramatic desaturation in a 4-yr-old patient during a tonsillec.
`‘°mY- Repmducfd from 3l¢3Ph€I1 RC. Slater HM. Johnson AL, Sekelj
`P: The oximeter—-A technical aid for the anesthesiologist. ANE.5THt-:-
`SIO1-OGV 125543 1951. Wlth permission.
`
`The Ph}’3iC3 and Physiology of Pulse Oximetry
`
`BEE.R’S Law
`
`In the 19305, Matthes used spectrophotometry to dc-
`termine hemoglobin oxygen saturation? This method is
`based on the Beer-Lambert law, which relates the con-
`centration ofa solute to the intensity oflight transmitted
`through a solution.
`
`I-trans = Iine-A
`
`A = DC:
`
`(1)
`
`(13)
`
`Where [trans = intensity oftransmitted light; I,,. = intensity
`of incident light; A = absorption; D = distance light is
`transmitted through the liquid (path length); (3 = Con-
`centration of solute (hemoglobin); e = extinction coeffi-
`cient of the solute (a constant for a given solute at a spec-
`ified wavelength). Thus, if a known solute is in a clear
`solution in a cuvette of known dimensions, the solute con.
`centration can be calculated from measurements of the
`incident and transmitted light intensity at a known wave.
`length. The extinction coefficient E is a property of light
`absorption for a specific substance at a specified wave-
`length. In a one-component system. the absorption A is
`the product of the path length, the concentration, and
`the extinction coefficient. equation la. If multiple solutes
`
`0009
`
`

`
`K. K. TREMPER AND S.
`
`BARKER
`
`Anesthesiolo
`v to, No 1. _]an 1939
`
`By the above definition of oxygen saturation, the two
`forms of hemoglobin that do not bind oxygen (C01-lb
`and MetHb) are not included. This is the origin of what
`is now referred to as “functional hemoglobin saturation,”
`defined as (Severinghaus JW, personal communication);
`
`O21-lb
`2
`Functional S2102 = 6—~l_-IE-;TI-6
`
`x 100%.
`
`(2)
`
`With the advent of multiwavelength oxitneters that can
`measure all four species of hemoglobin, “fractional sat-
`uration” has been defined as the ratio of oxyhemoglobin
`to total hemoglobin:
`
`HEMOGLOBIN EXTINCTION cu fives
`
`Infrpred
`I
`
`methemoglobin
`
`I i
`
`oxyhemoglobin
`nu.--.....,___.
`
`reduced
`hemoglobin
`
`IIIII
`
`
`
`ExtinctionCoeifieient
`
`sue am can 720
`
`160
`
`son
`
`aau
`
`530
`
`920
`
`sen
`
`Wavelength A (nm}
`
`Fractional S3102
`
`FIG. 2. Transmitted light absorbance spectra of four hemoglobin
`species: oxyhemogiobin. reduced hemoglobin, carboxyhemoglobin, and
`methemoglobin. Adapted from Barker 5] and Tremper KK: Pulse
`Oximetry: Applications and limitations, Advances in Oxygen Moni-
`toring. International Anesthesiology Clinics. Boston. Little, Brown and
`Company. 1987, pp. 155-175.
`
`are present, A is the sum of similar expressions for each
`solute. The extinction coefficient can vary dramatically
`with the wavelength of the light. The extinction coeffi-
`cients for various hemoglobin species in the red and in-
`frared wavelength range are shown in figure 2.
`Laboratory oximeters use this principle to determine
`hemoglobin concentration by measuring the intensity of
`light transmitted through a cuvette filled with a hemo-
`globin solution produced from lysed red blood cells.”
`For Beer's law to be valid, both the solvent and the cuvette
`
`must be transparent at the wavelength used, the light path
`length must be known exactly, and no absorbing species
`can be present in the solution other than the known solute.
`It is difficult to fulfill these requirements in clinical devices;
`therefore, each instrument theoretically based on Beer’s
`law also requires empirical corrections to improve accu-
`racy.
`
`HEMOGLOBIN SATURATION DEFINITIONS
`
`Adult blood usually contains four species of hemoglo-
`bin: oxyhernoglobin (OgHb), reduced hemoglobin (Hb),
`methemoglobin (MetHb), and carboxyhemoglobin (C0-
`Hb) (fig. 2). The last two species are in small concentra-
`tions, except in pathologic conditions. There are several
`definitions of hemoglobin saturation. Historically, “oxy-
`gen saturation” was first defined as the oxygen content
`expressed as a percentage of the oxygen capacity. The
`oxygen content (cc of oxygen per 100 cc of blood) was
`measured volumetrically by the method of Van Slylte and
`Neill (1924).15 The oxygen capacity was defined as the
`oxygen content after the blood sample had been equili-
`brated with room air (158 mmHg oxygen at sea level).
`
`O2Hb
`= :—————j x 100 .
`0211!: + Hb + COHb + MetHb
`%
`
`(3)
`
`The fractional hemoglobin saturation is also called the
`“oxyhemoglobin fraction,” or “oxyhemoglobin %.“”
`When oximetry is used to measure hemoglobin satu-
`ration, Beer‘s law must be applied to a solution containing
`four unknown species: O-gHb, Hb, COHb, and MetHb.
`Expanding equation 1a to a four-component system results
`in an absorption given by:
`
`A = D1C1€1
`
`‘l' Dgcgfg ‘l' D3_C3€3
`
`'l" D4C4E4.
`
`The subscripts 1 through 4 correspond to the four he-
`moglobin species. If the path lengths are the same, then
`D can be factored out:
`
`A = D(C1E1 ‘l’ C263 "l' C363 'l' C464).
`
`The extinction coeflicients £1 through 54 are constants at
`a given wavelength X (fig. 2). The absorption defined in
`equation 1c is determined from equation 1 by measuring
`the incident and transmitted light intensities. From equa-
`tion lc, we see that four wavelengths of light are needed
`to produce four equations to solve for the unknown con-
`centrations C1 through C4. If COHb and MetHb were
`not present, their contributions to the absorption would
`be zero and functional hemoglobin saturation could be
`determined by a two-wavelength oximeter (two equations
`and two unknowns). If two wavelengths existed for which
`the extinction coefficients for COHb and MetHb were
`
`zero, then these absorption terms would again be zero
`and a two-wavelength oximeter could measure functional
`saturation. Unfortunately, as illustrated in figure 2, the
`extinction coefficients for CO1-lb and MetHb are not zero
`
`in the red and infrared range, and their presence will,
`therefore, contribute to the absorption. Even though the
`definition of functional hemoglobin saturation involves
`only two hemoglobin species (O2Hb and Hb), when
`_ MetHb and C01-lb are present, four wavelengths are re-
`quired to determine either functional or fractional he-
`moglobin saturation.”
`
`0010
`
`

`
`Atiesibesiultigjr
`V 70. No |.]an 1989
`
`PULSE OxI:viF.'1‘Rv
`
`Noninvasive oximeters measure red and infrared light
`transmitted through a tissue bed, effectively using the fin-
`ger or ear as a cuvette containing hemoglobin. There are
`several technical problems in accurately estimating SaO2
`by this method. First, there are many absorbers in the
`light path other than arterial hemoglobin, including skin,
`soft tissue, and venous and capillary blood. The early ox-
`imeters subtracted the tissue absorbance by compressing
`the tissue during calibration to eliminate all the blood,
`and using the absorbance of bloodless tissue as the base-
`line. These oximeters also heated the tissue to obtain a
`
`signal related to arterial blood with minimum influence
`of venous and capillary blood.
`Pulse oximeters deal with the effects of tissue and ve-
`
`nous blood absorbances in a completely different way.
`Figure 3 schematically illustrates the series of absorbers
`in a living tissue sample. At the top of the figure is the
`pulsatile or AC. component, which is attributed to the
`pulsating arterial blood. The baseline or DC component
`represents the absorbances of the tissue bed, including
`venous blood, capillary blood, and nonpulsatile arterial
`blood. The pulsatile expansion of the arteriolar bed pro-
`duces an increase in path length (see equation lb), thereby
`increasing the absorbance. All pulse oxiineters assume that
`the only pulsatile absorbance between the light source
`and the photodetector is that of arterial blood. They use
`two wavelengths of light: 660 nanometers (red) and 940
`nanometers (near infrared). The pulse oximeter first de-
`termines the AC component of absorbance at each wave-
`length and divides this by the corresponding DC coin-
`ponent to obtain a “ptilse-added" absorbance that is in-
`dependent ofthe incident light intensity. it then calculates
`the ratio (R) of these pulse-added absorbances, which is
`empirically related to SaO2:
`
`_ Acsso/Dcaso
`R _
`AC940/DC940
`
`.
`
`4
`
`(
`
`)
`
`Figure 4 is an example of a pulse oximeter calibration
`curve.” The actual curves used in commercial devices
`
`are based on experimental studies in human volunteers.
`Note that when the ratio of red to infrared absorbance
`
`is one, the saturation is approximately 85%. This fact has
`clinical implications to be discussed later.
`It is a fortuitous coincidence of technology and phys-
`iology that allowed the development of solid-state pulse
`oximeter sensors.” Light emitting diodes (LEDs) are
`available over a relatively narrow range of the electro-
`magnetic spectrum. Among the available wavelengths are
`some that not only pass through skin but also are absorbed
`by both oxyhemoglobin and reduced hemoglobin. For
`best sensitivity, the difference between the ratios of the
`absorbances of OgHb and I-lb at the two wavelengths
`
`PULSI-I (JXIl~il“.‘l'RY
`
`101
`
`LightAbsorption
`
`Absorption due to pulsatile arterial blood
`Absorption due to noi'1—pulsaIile arterial blood
`Absorption clue to venous and capillary blood
`
`'
`
`as itititititiitifititi
`T‘?00O0O0O9O060'O O O A.
`0000000020
`AALL.LA.A. A
`Time
`
`masorptioii due to tissue
`
`FIG. 3. This figure schematically illustrates the light absorption
`through living tissue. Note that the AC signal is due to the pulsatile
`component of the arterial blood while the DC signal is comprised of
`all the nonpnlsatile absorbers in the tissue; nonpulsatile arterial blood,
`venous and capillary blood, and all other tissues. Adapted from Ohmedzi
`Pulse Oximcter Model 3700 Service Manual, 1986, p. 22.
`
`should be maximized. As we see in figure 2, at 660 nano-
`meters. reduced hemoglobin absorbs about ten times as
`mtich light as oxyhemoglobin. (Note that the extinction
`coefficients are plotted on a logarithmic axis.) At the in-
`frared wavelength of 940 nanometers, the absorption
`coefficient of Oglrlb is greater than that of Hb.
`
`Engineering Design and Physiologic Limitations
`
`‘ Although the theory on which pulse oximetry is based
`is relatively straightforward. the appiication of this theory
`to the production ofa clinically useful device involves a
`
`0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.0 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4
`
`Fl : T
`A Cfifilig D C 560
`Ac9.io’lDc one
`
`FIG. 4. This is a typical pulse oximeter calibration curve. Note that
`the SaOg estimate is determined from the ratio (R) of the pulse-added
`red absorbance at 660 nanometers to pulse-added infrared absorbance
`at 940 nanometers. The ratios of red to infrared absorbances vary
`from approximately .4 at I00‘/'6 saturation to 3.4 at 0% saturation.
`Note that the ratio of red to infrared absorbance is one at a saturation
`
`of approximately 85%. This curve can be approximately determined
`on a theoretical basis but, for accurate predictions of SpOg. experi-
`mental data are required. Adapted from _]A Pologe: Pulse oitimetry:
`Technical aspects of machine design, International Anesthesiology
`Clinics, Advances in Oxygen Monitoring. Edited by Tremper 1-(K,
`Barker S]. Boston, Little. Brown and Company, 1987. p 142.
`
`0011
`
`

`
`102
`
`K. K. TREMPER AND S. J. BARKER
`
`Ancslhtsiology
`V '.l'(l.No1.jan19S9
`
`significant engineering effort. This section will present in
`general terms the clinical and physiologic problems of
`oximeter design and their engineering solutions. The dis-
`cussion is divided into four areas: dyshemoglobins and
`dyes, LED center wavelength variability, signal artifact
`management, and accuracy and response. The reader
`should be aware that these problems can interact with
`one another.
`
`DYSHEMOGLOBINS AND DYES
`
`Being two-wavelength devices, pulse oximeters can deal
`with only two hemoglobin species. As noted above, this
`would be adequate to measure functional SaOg if MetHb
`and COI-Ib did not absorb red or infrared light at the
`wavelengths used. Unfortunately, this is not the case, and
`therefore both MetHb and COHb will cause errors in the
`
`pulse oximeter reading. It is not intuitively obvious how
`a pulse oximeter will behave in the presence of dyshem-
`oglobins. With respect to carboxyhemoglobin, we can gain
`some insight from the extinction curves of figure 2. In
`the infrared range (940 nm), COHb absorbs very little
`light: whereas, in the red range (660 nm), it absorbs as
`much light as does 02Hb. This is clinically illustrated by
`the fact that patients with carboxyhemoglobinemia appear
`red. Therefore, to the pulse oximeter, COHb looks like
`O21-lb at 660 nm; while, at 940 nm COHb is relatively
`transparent. The effect of COI-Ib on pulse oximeter values
`has been evaluated experimentally in dogs.” In this study,
`the pulse oximeter saturation (SpOg) was found to be given '
`approximately by:
`
`4, an absorbance ratio of one corresponds to a saturation
`of 85% on the calibration curve. Pulse oximeter error
`
`during methemoglobinernia has also been confirmed in a
`clinical report.”
`In neonatal blood, a fifth type of hemoglobin is present,
`fetal hemoglobin (I-lbF). I-IbF differs from adult Hb in
`the amino acid sequences of two of the four globin sub.
`units. Adult Hb has two 0c- and two ,6-globin chains, while
`I-IbF has two at and two f chains. This difference in globin
`chains has little effect on the extinction curves and there-
`
`fore should not affect pulse oximeter readings.§'ll This is
`indeed fortunate because the fraction of HbF present in
`neonatal blood is a function of gestational age and cannot
`be accurately predicted. HlJF does produce a small error
`in in vitro laboratory oximeters; O«_;HbF may be inter-
`preted as consisting partially of COHb.2°
`The absorbance ratio R (equation 4) may be affected
`by any substance present in the pulsatile blood that absorbs
`light at 660 or 940 nm and was not present in the same
`concentration in the volunteers used to generate the cal-
`ibration curve (fig. 4). Intravenous dyes provide a good
`example of this principle.” '22 Scheller at all. evaluated the
`effects of bolus doses of methylene blue, indigo carmine,
`and indocyanine green on pulse oximeters in human vol-
`unteers.“ They found that methylene blue caused a fall
`in SpOg to approximately 65% for 1-2 min. Indigo car-
`mine produced a very small drop in saturation, while in-
`docyanine green had an intermediate effect. The detec-
`tion of intravenous dyes by pulse oximeters should not
`be surprising. because it was this effect that led Aoyagi
`to the invention of pulse oximetry.2
`
`SP0” :
`
` ra X ,,,,,%_
`total Hb
`
`(5)
`
`LED CENTER WAVELENGTH VARIABILITY
`
`The effects of methemoglobinemia on pulse oximetry
`are also partially predictable from the extinction curves
`(fig. 2). MetHb has nearly the same absorbance as reduced
`hemoglobin at 560 nm, while it has a greater absorbance
`than the other hemoglobins at 940 nm. This is consistent
`with the clinical observation that methemoglobinemia
`produces very dark, brownish blood. Thus, it would be
`expected to produce a large pulsatile absorbance signal
`at both wavelengths. The effect of Metl-Ib on pulse ox-
`imeter readings has also been measured in dogs.” As
`methemoglobin levels increased, the pulse oximeter sat-
`uration (SPO2) tended toward 85% and eventually became
`almost independent of the actual SaOg .13 In other words,
`in the presence of high levels of MetHb, SpOg is erro-
`neously low when S2102 is above 85%, and erroneously
`high when SaOg is below 85%. This may be explained by
`the fact that MetHb causes a large pulsatile absorbance
`at both wavelengths, thereby adding to both the numer-
`ator and denominator of the absorbance ratio R (equation
`4) and forcing this ratio toward unity. As shown in figure
`
`The LEDS used in pulse oximeter sensors are not ideal
`monochromatic light sources: there is a narrow spectral
`range over which they emit light. The center wavelength
`of the emission spectrum varies even among diodes of the
`same type from the same manufacturer. This variation
`can be $15 nanometers.” As seen in figure 2, a shift in
`LED center wavelength will change the measured ex-
`tinction coefficient and thus produce an error in the sat-
`uration estimate. This source wavelength effect will be
`greatest for the red (560 nm) wavelength, because the
`extinction curves have a steeper slope at this wavelength.
`Manufacturers have found two approaches to this prob-
`lem. Some test all the LEDs and reject those that are out
`of their specified wavelength range, e.g., 660 i 5 nano-
`meters. This is expensive due to the number of LEDs
`
`.
`
`§ Pologe _]A, Raley DM: Effects of fetal hemoglobin on pulse ox-
`imetry._I Perinat VII:324-525, 1987.
`ll Anderson _]V: The accuracy of pulse oximetry in neonates: Effects
`of fetal hemoglobin and bilirubin.] Perinat VII:323. 1987.
`
`0012
`
`

`
`Anesthesiology
`V 1'0. No|.]an1‘.IB9
`
`PULSE OXIMETRY
`
`103
`
`rejected; i.r., narrower acceptable range yields improved
`accuracy but also more rejected LEDs. Alternatively,
`other manufacturers program the pulse oximeter to ac-
`cept several ranges of LED center wavelengths for both
`the red and infrared, allowing the device to correct in-
`terna

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket