`Proceedings of the 29th Annual International
`Conference of the IEEE EMBS
`Cité Internationale, Lyon, France
`August 23-26, 2007.
`
`ThD13.1
`
`1-4244-0788-5/07/$20.00 ©2007 IEEE
`
`1528
`
`
`
`
`A Comparative Evaluation of Adaptive Noise Cancellation
`Algorithms for Minimizing Motion Artifacts in a Forehead-Mounted
`Wearable Pulse Oximeter
`Gary Comtois, Member IEEE, Yitzhak Mendelson, Member IEEE, Piyush Ramuka
`
` Abstract— Wearable physiological monitoring using a pulse
`oximeter would enable field medics to monitor multiple injuries
`simultaneously, thereby prioritizing medical intervention when
`resources are limited. However, a primary factor limiting the
`accuracy of pulse oximetry is poor signal-to-noise ratio since
`photoplethysmographic (PPG) signals, from which arterial
`oxygen saturation (SpO2) and heart rate (HR) measurements
`are derived, are compromised by movement artifacts. This
`study was undertaken to quantify SpO2 and HR errors induced
`by certain motion artifacts utilizing accelerometry-based
`adaptive noise cancellation (ANC). Since the fingers are
`generally more vulnerable to motion artifacts, measurements
`were performed using a custom forehead-mounted wearable
`pulse oximeter developed for real-time remote physiological
`monitoring and triage applications. This study revealed that
`processing motion-corrupted PPG signals by least mean
`squares (LMS) and recursive least squares (RLS) algorithms
`can be effective to reduce SpO2 and HR errors during jogging,
`but the degree of improvement depends on filter order.
`Although both algorithms produced similar improvements,
`implementing the adaptive LMS algorithm is advantageous
`since it requires significantly less operations.
`
`T
`
`I. INTRODUCTION
`HE implementation of wearable diagnostic devices
`would enable real-time remote physiological assessment
`and triage of military combatants, firefighters, miners,
`mountaineers, and other individuals operating in dangerous
`and high-risk environments. This, in turn, would allow first
`responders and front-line medics working under stressful
`conditions to better prioritize medical intervention when
`resources are limited, thereby extending more effective care
`to casualties with the most urgent needs.
`Employing commercial off-the-shelf (COTS) solutions,
`for example finger pulse oximeters to monitor arterial blood
`oxygen saturation (SpO2) and heart rate (HR), or adhesive-
`type disposable electrodes for ECG monitoring, are
`
`
`
`Manuscript received April 2, 2007. This work is supported by the U.S.
`Army MRMC under Contract DAMD17-03-2-0006. The views, opinions
`and/or findings are those of the author and should not be construed as an
`official Department of the Army position, policy or decision unless so
`designated by other documentation.
`Y. Mendelson is a Professor in the Department of Biomedical
`Engineering, Worcester Polytechnic Institute, Worcester, MA 01609 USA
`(phone: 508-831-5103; fax: 508-831-5541; email: ym@wpi.edu).
`G. Comtois is a graduate student in the Department of Biomedical
`Engineering, Worcester Polytechnic Institute, Worcester, MA 01609 USA
`(comtoisg@wpi.edu).
`P. Ramuka is a graduate student in the Department of Biomedical
`Engineering, Worcester Polytechnic Institute, Worcester, MA 01609 USA
`(pramuka@wpi.edu).
`
`impractical for field applications because they limit mobility
`and can interfere with regular activity. Equally important,
`since these devices are designed for clinical settings where
`patient movements are relatively constrained, motion
`artifacts during field applications can drastically affect
`measurement accuracy while subjects remain active.
`Practically, the primary factor limiting the reliability of
`pulse oximetry is attributed to poor signal-to-noise ratio
`(SNR) due to motion artifacts. Since photoplethysmographic
`(PPG) signals, which are used to determine SpO2 and HR,
`are obscured during movements, the implementation of a
`robust pulse oximeter for field applications requires
`sophisticated noise
`rejection algorithms
`to eliminate
`erroneous readings and prevent false alarms.
`To minimize the effects of motion artifacts in wearable
`pulse oximeters, several groups proposed various algorithms
`to accomplish adaptive noise cancellation (ANC) utilizing a
`noise reference signal obtained from an accelerometer
`(ACC) that is incorporated into the sensor to represent body
`movements [1]-[3]. These groups demonstrated promising
`feasibility for movement artifact rejection in PPG signals
`acquired from the finger. However, they did not present
`quantifiable data showing whether accelerometry-based
`ANC resulted in more accurate determination of SpO2 and
`HR derived from PPG signals acquired from more motion-
`tolerant body locations that are more suitable for mobile
`applications.
`
`II. BACKGROUND
`Generally, linear filtering with a fixed cut-off frequency is
`not effective in removing in-band noise with spectral overlap
`and temporal similarity that is common between the signal
`and artifact. Thus, we utilized ANC techniques to filter noisy
`PPG waveforms acquired during field experiments. The
`performance of
`this signal processing approach was
`evaluated based on its potential to lower SpO2 and HR
`measurement errors.
`Among the most popular ANC algorithms are the least
`mean squares (LMS) and recursive least squares (RLS)
`algorithms. Briefly,
`to attenuate
`the
`in-band noise
`component in the desired signal, these algorithms assume
`that
`the reference noise received from
`the ACC
`is
`statistically correlated with the additive noise component in
`the corrupted PPG signal, whereas the additive noise is
`uncorrelated with the noise-free PPG signal. An error signal
`is used to adjust continuously the filter’s tap-weights in
`order to minimize the SNR of the noise-corrected PPG
`signal.
`
`IPR2017-00317
`CONDITIONAL MOTION TO AMEND
`
`VALENCELL, INC.
`EXHIBIT 2134 - PAGE 2
`
`

`

`1529
`
`The performance of ANC algorithms is highly dependent
`on various filter parameters, including filter order (M).
`Accordingly, careful consideration must be given to the
`selection of these parameters and the trade-off between
`algorithm complexity and its computation time.
`Although the basic principles of the LMS and RLS
`techniques share certain similarities, the LMS algorithm
`attempts to minimize only the current error value, whereas in
`the RLS algorithm, the error considered is the total error
`from the beginning to the current data point. Furthermore,
`the performance of each algorithm depends on different
`parameters. For example, the step size (µ) has a profound
`effect on the convergence behavior of the LMS algorithm.
`Similarly, the forgetting factor (λ) determines how the RLS
`algorithm treats past data inputs.
`Compared to the LMS algorithm, the RLS algorithm has
`generally a faster convergence rate and smaller error.
`However, this advantage comes at the expense of increasing
`complexity and longer computational time which increases
`rapidly and non-linearly with filter order.
`
`III. METHODS
`To simulate movement artifacts, we performed a series of
`outdoor and indoor experiments that were intended to
`determine the effectiveness of using the accelerometer-based
`ANC algorithms
`in processing motion-corrupted PPG
`signals acquired by a forehead pulse oximeter. The focus of
`this study was to compare the performance of each algorithm
`by quantifying the improvement in SpO2 and HR accuracy
`generated during typical activities that are expected to
`induce considerable motion artifacts in the field.
`Data were collected by a custom forehead-mounted pulse
`oximeter developed in our laboratory as a platform for real-
`time remote physiological monitoring and triage applications
`[4]-[6]. The prototype wearable system is comprised of three
`units: A battery-operated optical Sensor Module (SM)
`mounted on the forehead, a belt-mounted Receiver Module
`(RM) mounted on the subject’s waist, and a Personal Digital
`Assistant (PDA) carried by a remote observer. The red (R)
`and infrared (IR) PPG signals acquired by the small (φ =
`22mm) and lightweight (4.5g) SM are transmitted wirelessly
`via an RF link to the RM. The data processed by the RM can
`be transmitted wirelessly over a short range to the PDA or a
`PC, giving the observer the capability to periodically or
`continuously monitor the medical condition of multiple
`subjects. The system can be programmed to alert on alarm
`conditions, such as sudden trauma, or when physiological
`values are out of their normal range. Dedicated software was
`used to filter the reflected PPG signals and compute SpO2
`and HR based on the relative amplitude and frequency
`content of the PPG signals. A triaxial MEMS-type ACC
`embedded within the SM was used to get a quantitative
`measure of physical activity. The information obtained
`through the tilt sensing property of the ACC is also used to
`determine body posture. Posture and acceleration, combined
`with physiological measurements, are valuable indicators to
`assess the status of an injured person in the field.
`
`Body accelerations and PPG data were collected
`concurrently from 7 healthy volunteers during 32 jogging
`experiments. These jogging experiments comprised 16
`treadmill, 12
`indoor, and 4 outdoor exercises. Each
`experiment comprised a 1-minute free jogging at speeds
`corresponding to 3.75–6.5 mph, framed by 2-minute resting
`intervals. For validation, reference SpO2 and HR were
`acquired concurrently from the Masimo transmission pulse
`oximeter sensor attached to the subject’s fingertip which was
`kept in a relatively stationary position throughout the study.
`We chose the Masimo pulse oximeter because it employs
`unique signal extraction technology (SET®) designed to
`greatly extend its utility into high motion environments. A
`Polar™ ECG monitor, attached across the subject’s chest,
`provided reference HR data.
`The ACC provided reference noise inputs to the ANC
`algorithms. The X, Y, and Z axes of the triaxial ACC were
`oriented according to the anatomical planes as illustrated in
`Fig. 1. Accelerations generated during movement depend
`upon the types of activity performed. Generally, during
`jogging, acceleration is greatest in the vertical direction,
`although the accelerations in the other two orthogonal
`directions are not negligible. Therefore, the noise reference
`input applied to the ANC algorithms was obtained by
`summing all three orthogonal axes of the ACC. By
`combining signals from all three axes, measurements
`become insensitive to sensor positioning and inadvertent
`sensor misalignment that may occur during movements. To
`compensate for differences in response times, the SpO2 and
`HR measurements acquired
`from each device were
`processed using an 8-second weighted moving average.
`
`ACC axial
`orientations
`
`Finger
`Sensor
`
`SM + ACC
`
`Polar HR Monitor
`
`RM
`
`Masimo
`Pulse Oximeter
`
`Fig. 1: Experimental setup for data collection.
`
`The outputs of the MEMS ACC and raw PPG signals were
`acquired in real-time at a rate of 80 s/s using a custom
`written LabVIEW® program. Data were processed off-line
`using Matlab programming. The ANC algorithms were
`implemented in Matlab with parameters optimized for
`computational speed and measurement accuracy. The LMS
`algorithm was implemented using a constant µ of 0.016. The
`
`IPR2017-00317
`CONDITIONAL MOTION TO AMEND
`
`VALENCELL, INC.
`EXHIBIT 2134 - PAGE 3
`
`

`

`1530
`
`
`
`
`selected filter parameters for the RLS algorithm were
`λ = 0.99 and an inverse correlation matrix P = 0.1. These
`filter parameters were found to be optimal in preliminary
`experiments. For comparison, data were processed by each
`algorithm using variable order filters.
`
`IV. RESULTS
`SpO2 and HR data were derived from the R and IR PPG
`signals utilizing custom extraction algorithms. SpO2 root
`mean squared errors (RMSE) were quantified based on the
`differences between the readings measured by the custom
`and Masimo pulse oximeters, whereas HR errors were
`defined with respect
`to
`the Polar HR monitor. For
`comparison, RMSE were determined by processing the PPG
`signals off-line either with or without the ANC algorithms.
`Fig. 2 shows a representative tracing of SpO2 and HR
`measurements obtained from the custom pulse oximeter with
`and without ANC. Reference measurements obtained
`simultaneously from the Masimo pulse oximeter and Polar
`HR monitor during resting and outdoor jogging were also
`included for comparison.
`
`
`Masimo
`
`LMS
`
`RLS
`
`Without ANC
`
`Resting
`
`Jogging
`
`Resting
`
`Without ANC
`
`RLS
`
`Polar
`
`Masimo
`
`LMS
`
`frequency band ranging between 1.5–3.0 Hz. Further
`analysis of the data showed that in 8 out of the 32 jogging
`experiments (25%), the cardiac-synchronized frequencies
`and movement-induced acceleration frequencies shared a
`common band.
`The averaged errors observed from the series of 32
`experiments are summarized in Figures 3 and 4. Analysis of
`the data clearly revealed that utilizing either the LMS or
`RLS algorithm to process the noise-corrupted PPG signals
`can improve both SpO2 and HR accuracy during jogging.
`Although the degree of improvement varied, because
`different methods are employed to compute SpO2 and HR
`from
`the PPG signal,
`these figures show
`that
`the
`performance of both algorithms depends on filter order used
`to implement each algorithm.
`
`LMS
`RLS
`
`20
`
`15
`
`10
`
`05
`
`SpO2 RMSE (%)
`
`0
`
`4
`
`8
`
`12
`Filter Order (M)
`
`Fig. 3. Averaged SpO2 errors for varying filter orders. Error bars indicate
`±1SD. For comparison, M = 0 represents the error obtained without ANC.
`
`
`16
`
`20
`
`24
`
`LMS
`RLS
`
`25
`
`20
`
`15
`
`10
`
`05
`
`HR RMSE (%)
`
`
`
`0
`
`4
`
`8
`
`12
`Filter Order (M)
`
`Fig. 4. Averaged HR errors for varying filter orders. Error bars indicate
`±1SD. For comparison, M = 0 represents the error obtained without ANC.
`
`16
`
`20
`
`24
`
`V. DISCUSSION
`Pulse oximeters are used routinely in many clinical
`settings where patients are at rest. Their usage in other areas
`is limited because of motion artifacts which is the primary
`contributor to errors and high rates of false alarms. In order
`to design wearable cost-effective devices that are suitable for
`field deployment, it is important to ensure that the device is
`robust against motion induced disturbances. PPG signals
`recorded from the forehead are generally less prone to
`movement artifacts compared to PPG signals recorded from
`
`
`
`Resting
`
`Jogging
`
`Resting
`
`Fig. 2. Representative SpO2 (top) and HR (bottom) measurements obtained
`during outdoor jogging. Filter order M = 16.
`Spectral analysis of the data using FFT revealed that
`during jogging frequency components associated with body
`acceleration and the subject’s HR shared a relatively small
`
`IPR2017-00317
`CONDITIONAL MOTION TO AMEND
`
`VALENCELL, INC.
`EXHIBIT 2134 - PAGE 4
`
`

`

`1531
`
`essential in real-time applications. However, this comes at
`the expense of a longer computational time since the RLS
`algorithm requires M2 operations per iteration. Considering
`for example that an implementation based on a 24th-order
`filter would provide an acceptable error reduction, this
`implies that the LMS algorithm would require only 24
`operations compared to 576 operations that will be required
`to
`implement an adaptive RLS algorithm. Table 1
`summarizes the relative execution times of the LMS and
`RLS adaptive algorithms for processing one data point.
`
`Table 1. Execution times for LMS and RLS algorithms
`Filter Order
`LMS (ms)
`RLS (ms)
`2
`1.0
`6.5
`4
`1.8
`18.5
`8
`3.2
`63.0
`16
`6.2
`235.0
`
`VI. CONCLUSIONS
`This study was designed to investigate the performance of
`accelerometry-based ANC implemented using the LMS and
`RLS algorithms as an effective method to minimizing both
`and HR
`errors
`induced during movement.
`SpO2
`Measurements were performed using a custom, forehead-
`mounted wearable pulse oximeter that was developed in our
`laboratory to serve as a platform for real-time remote
`physiological monitoring and triage applications. The results
`obtained in this study revealed that processing motion-
`corrupted PPG signals by the LMS and RLS algorithm can
`reduce HR and SpO2 errors during jogging. Although both
`algorithms
`produced
`similar
`improvements,
`the
`implementation of the adaptive LMS algorithm is preferred
`since it requires significantly less operations.
`
`[1]
`
`REFERENCES
`J. Y. A. Foo, S. J. Wilson, “A computational system to optimize noise
`rejection in photoplethysmography signals during motion or poor
`perfusion states,” Med. Biol. Eng. Comput., 2006, pp. 140-145.
`[2] L. B. Wood, H. H. Asada, “Noise cancellation model validation for
`reduced motion artifact wearable PPG sensors using MEMS
`accelerometers,” in Proc. 28th IEEE/EMBS Ann. Int. Conf., New York,
`2006, pp. 3525-3528.
`[3] A.R. Relente, L.G. Sison, Characterization and adaptive filtering of
`motion artifacts in pulse oximetry using accelerometers, Proc. 2nd
`Joint EMBS/BMES Conf., Houston, TX, 2002, 1769-1770.
`[4] Y. Mendelson and V. Floroff, A PDA based ad-hoc mobile wireless
`pulse oximeter, in Proc. IASTED International Conference Telehealth
`2005, Banff, Canada, 2005, pp. 1-6.
`[5] Y. Mendelson, R. J. Duckworth, G. Comtois, “A wearable reflectance
`pulse oximeter for remote physiological monitoring,” in Proc. 28th
`IEEE EMBS Annual International Conf., New York, 2006, pp. 912-
`915.
`[6] G. Comtois, Y. Mendelson, “A noise reference input to an adaptive
`filter algorithm for signal processing in a wearable pulse oximeter,” in
`Proc. 33rd Annual Northeast Bioengineering Conf., New York, 2007,
`pp. 106-107.
`
`
`
` a
`
` finger. Nonetheless, morphological distortions of the
`underlying PPG waveforms, from which SpO2 and HR
`measurements are derived, could lead to measurement
`errors, false alarms, and frequent dropouts when subjects
`remain active. For example, as shown in Fig. 2, it is evident
`that the Masimo pulse oximeter, which employs advanced
`signal extraction technology designed to greatly extend its
`utility into high motion environments, was clearly unable to
`accurately track SpO2 and HR while the subject was jogging.
`Although
`to a
`lesser extent, we also noticed more
`pronounced fluctuations in SpO2 recorded by the wearable
`forehead pulse oximeter during jogging. These fluctuations
`are likely caused by PPG waveforms obscured by motion
`artifacts associated with heavier breathing.
`To address the need to improve the performance of a
`prototype reflectance pulse oximeter during jogging, we
`investigated the effectiveness of a MEMS ACC as a noise
`reference input to two popular ANC algorithms. We chose
`the LMS and RLS adaptive routines since other investigators
`showed the promising utility of these algorithms to reduce
`errors attributed to motion artifacts in pulse oximeters [1]-
`[3].
`Analysis of the data acquired during jogging experiments
`showed that ANC implemented using the LMS and RLS
`algorithms can help to improve considerably the accuracy of
`a pulse oximeter, as shown in Fig. 2. However, although the
`differences are not considered clinically significant, we
`found that processing the corrupted PPG signals by each
`algorithm produced slightly different improvements. These
`differences are anticipated since different computational
`principles are employed by a pulse oximeter.
`Since ANC-based filtering implements an adaptive notch
`filter with a notch frequency corresponding to the dominant
`frequency of the measured ACC signal, we expected that an
`overlap of the HR and movement-induced ACC frequencies
`would attenuate
`the fundamental cardiac-synchronized
`frequency of the PPG signals and, therefore significantly
`affecting SpO2 and HR measurements. However, separate
`analysis of
`the data from experiments where body
`accelerations and cardiac rhythms were found
`to be
`synchronized confirmed that applying either the LMS or
`RLS algorithm did not adversely impact the ability to obtain
`accurate SpO2 and HR readings while subjects remain
`active.
`As shown in Fig. 3 and Fig. 4, we found that the degree of
`improvement depends on the filter order (M) used to
`implement each adaptive algorithm, however filters order
`greater
`than 24 produced diminished
`improvements.
`Furthermore, we also found that the LMS algorithm was
`slightly more effective in reducing HR errors compared to
`the RLS implementation.
`Given similar performances, it is important to take into
`consideration
`the complexity of
`the LMS and RLS
`algorithms and the trade-off between algorithmic complexity
`and computation
`time. These principal
`tradeoffs are
`important since our goal is to implement ANC to improve
`the performance of a wearable pulse oximeter during
`motion. For example, compared to the LMS algorithm, the
`RLS algorithm has a faster convergence rate which is
`
`IPR2017-00317
`CONDITIONAL MOTION TO AMEND
`
`VALENCELL, INC.
`EXHIBIT 2134 - PAGE 5
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket