throbber
EXHIBIT 2001
`
`EXHIBIT 2001
`
`

`

`Earbud-Based Sensor for the Assessment
`of Energy Expenditure, HR, and VO2max
`
`STEVEN FRANCIS LEBOEUF', MICHAEL E. AUMER', WILLIAM E. KRAUS2, JOHANNA L. JOHNSONZ,
`and BRIAN DUSCHA2
`
`’ Valencell, Inc., Raleigh, NC; and 2Division ofCardiology, Department ofMedicine, Duke University
`School ofMedicine, Durham, NC
`
`ABSTRACT
`
`LEBOEUF, S. F., M. E. AUMER, W. E. KRAUS, J. L. JOHNSON, and B. DUSCHA. Farbud-Based Sensor for the Amessment of
`Energy Expenditure, HR, and V02,,.,,. Med. Sci. Sports Exen-., Vol. 46, No. 5, pp. 1046 1052, 2014. lntroductionllhrrposc: The goal
`of this program was to determine the fmsibility ofa novel noninvasive, highly miniaturized optornechanical earbud sensor for accurately
`estimating total energy expenditure (TEE) and maximum oxygen consurnptim (V()2,,_,,). The optomechanical sensor module, small
`enough to tit inside commercial audio earbuds, was previously developed to provide a seamless way to measure blood flow informa-
`tion durirrg daily life activities. The sensor module was configured to continuously measure physiological information via photo-
`plethysmography and physical activity infomratim via accelerometry. 'Ihis information was digitized and sent to a microprocessor
`where digital signal—processing algorithms extract physiological metrics in real time. These metrics were streamed wirelessly from
`the earbud to a computer. Methods: In this study, 23 subjects of multiple physical habitus were divided into a training group of 14
`subjects and a validaion group of 9 subjects. Each subject underwent the same exercise measurement protocol consisting of treadmill-
`based cardiopulmonary exercise testing to reach \70;»_....,. Benchmark sensors included a 12-lead IKZG sensor for measuring HR, a
`calibrated treadmill for measuring distance and speed, and a gas-exchange analysis instrument for measuring TEE and \-102...... 'Ihe
`earbud sensor was the device under test Benchmark and device under test data collected from the 14-person training data set study
`were integrated into a preconceived statistical model for correlating benchmark data with earbud sensor data. Coefficients were op-
`timized, and the optimized model was validated in the 9-person validation data set. Results: It was observed that the earbud sensor
`estimated TEE and V0-,._,,_, with mean 1 SD percent estimation errors of 0.7 1 7.4% and
`3.2 L 7.3%, respectively. Conclusion: The
`earbud semorcan accurately estimae TEE and V()2,,_,, during cardiopulmomry exercise testing. Key Words: FAR, ACCH..EROMEI'ER,
`PHOTOPLETHYSMOGRAPHY, PULSE
`
`odifiable health risk fitctors, such as high stress,
`poor diet, and sedentary lifestyle, account for 25%
`of all medical expenses and millions of deaths
`per year worldwide (2). The U.S. population is becoming
`increasingly overweight and unhealthy, with an estimated
`66% of adults categorized as obese or overweight by the
`CDC (26). Nonetheless, more than half of American adults
`exercise on a regular basis (1 1), spending more than $55
`billion in weight loss programs and more than $17 billion
`on fitness products (31). The disconnect between dollars
`spent on weight loss and obesity levels may be explained by
`recent findings that traditional diets do not work (24) alone
`
`Adttas fa‘ carespondence: Steven Francis Lfioeut; Ph.D., Valencell, lnc.,
`28(1)-154 Sumner Bhrd., Raldgh, NC 27616; E-mail: LeBoeu@vak:rrceI.corrL
`Submitted for publication March 2013.
`Accepted for publication October 2013.
`0195-9131/I4/4605-1046/0
`MEDICINE & SCIENCE IN SPORTS & EXERCISE,
`Copyright© 2014 by the American College of Sports Medicine
`D01: 10.1249/MSS.(l)00000000000l83
`
`to prevent weight gain and to promote weight loss. Dietary
`measure must be combined with energy expenditure to ac-
`complish long-terrn weight loss and maintenance.
`Weight loss programs aimed at promoting fitness through
`direct measurement of physical activity (PA) via pedometer
`feedback have shown promise. In particular, incorporating
`a pedometer in daily life activities has been shown to result
`in a significant reduction in body mass index (BMI) and
`blood pressure (7). Furthermore, combining engaging feed-
`back with an online user experience correlates with im-
`proved maintenance of weight loss in long-term diet/weight
`management studies (10). These observations indicate that
`even better outcomes may result from a more direct feed-
`back about energy expenditure and aerobic fitness level,
`such as V02, calories burned, and \'/02.“,
`Indeed, there is a clear opportunity to encourage a broader
`population to embrace active lifestyles by integrating mobile
`fitness monitoring devices with compelling user experiences.
`However, compelling user experiences must be meaningfirl,
`and to be meaningful, the fitness monitoring gadgets must
`provide information that is sufliciently accurate to be action-
`able. This goal is challenged by the fact that commercial pe-
`dorneters are inaccurate by greater than i 20% in reporting
`calories burned (8,29).
`
`1046
`
`Copyright 9 2014 by the American Colege of Sports Medicine. Unauthorized reproduction of this article is prohibited.
`
`VALENCELL_0001
`
`<
`
`Ua
`
`n DL
`
`L!
`:1
`CL
`CL
`
`an
`LL!
`
`UZ L
`
`L’
`
`

`

`Wmm
`
`Cm Um Qm Z(
`
`)>
`‘U
`‘U
`
`
`
`FIGURE I The components and size of the device under test (DUT).
`Shown are the earbud and the medallion containing the majority of
`the computational components. Shown for scale is a U.S. quarter. Note
`the pmition of the sensor module at the bottom of the antitragus. The
`sensor module is configured to lit between the concha and the antitragus
`of the ear.
`
`Recent energy expenditure studies, using a wearable Acti-
`Health chest strap monitor for measuring both PA and HR,
`have demonstrated greater accuracy (5,6). These researchers
`achieved such predictive accuracy through branched equation
`modeling, using HR information and accelerornetry infor-
`mation as independent variables. Although these findings are
`quite encouraging, researchers using the ActiHealth monitor
`point out several shortcomings. First, despite the relatively
`high precision achievable through branched equation model-
`ing, poorer accuracy is observed if individual calibrations are
`not used (5,6). This means that the wearable monitors must
`be calibrated for each user, in a process that is both time con-
`suming and burdensome. Furthermore, as audio earbuds are
`packaged with srnartphones and digital media players that
`are sold in volumes of hundreds of millions of units a year
`(16), the audio earbud form factor provides the opportunity
`to reach a larger consumer audience than that of an HR
`chest strap, which is sold in volumes of less than 10 million
`pa year.
`The goal of this study was to determine the feasibility of
`a highly miniaturized, noninvasive optomechanical earbud
`sensor technology for accurately monitoring physiological
`metrics such as HR, total energy expenditure (TEE), and
`
`maximum oxygen consumption 02",“), and this study is
`reported herein.
`
`METHODS
`
`To overcome these reported limitations, an earbud sensor
`module—as opposed to an ActiGraph wrist-, arm-, or leg-
`wom sensor—was selected in this study (Fig. 1). Details
`of the mechanism of operation are described elsewhere
`(17-21), but in summary, the earbud comprised a highly
`integrated sensor module capable of measuring subtle blood
`flow changes via reflective photoplethysrnography (PPG)
`and changes in body motion through a three-axis acceler-
`ometer. This sensor module was designed 1) to capture and
`digitize the optical PPG signal and 2) to send the digitized
`infomration to a digital signal processor (DSP) for remov-
`ing motion artifacts and environmental noise fi'om the PPG
`signal and to continuously generate estimates of HR and
`V02 metrics in real time based on a statistical model com-
`prising PPG and accelerometry information. The DSP was
`in electrical communication with a Bluetooth chipset so that
`the real—tirne metrics could be called upon by a client de-
`vice (such as a laptop or smartphone). A preliminary feasi-
`bility study of this PerformTek® earbud sensor module had
`previously demonstrated accurate HR measurements during
`exercise, thus potentially eliminating the need for an elec-
`tmcardiographic chest strap in many use cases. This was a
`critical finding for the issue of user compliance, as 58% of
`U.S. headphone owners listen to headphones while exercis-
`ing and 34% wear headphones during evayday life activities
`(such as doing work around the house) (13), 10 times greata
`than the number ofAmericans who exercise with chest straps.
`Subjects. In this study, 23 subjects of good physical
`health were divided into a training group of 14 subjects and
`a validation group of 9 subjects. This sample size is justified
`by the high “effect size” observed for calibrated comela-
`tions of V702 and HR (22) and is further supported by the
`va-y high R2 coefficient observed (23) when comparing
`the earbud-deterrnined HR to 12-lead ECG-measured HR
`
`during exercise. The training group (Table 1a) comprised
`12 men and 2 women: age = 39 i 11.8 yr, weight = 73.5 i
`12.2 kg, height = 69 i 2.9 cm, BMI = 23.6 : 2.1 kg-m_2.
`The validation group (Table 1b) comprised five men and
`four women: age = 36 i 6.9 yr, weight = 67.6 i 15.7 kg,
`height = 173 i 7.4 cm, BMI = 22.3 i 4.0 kg-m'2. Each
`
`TABLE 1. Descriptive characteristics (mean + SD) of (a) trairing group and (1)) Validation group.
`(a) Training Group
`
`(Ir) Validation Grnup
`
`Paramehr
`Sex
`Age
`weight
`Height
`Distance
`Energy expendiirre
`Ilaximum V02
`IMI
`
`Value (Mean + SD)
`2 females, 12 males
`30 + 11.8 yr
`735 + 12.2 kg
`175 + 7.4 cm
`2.95 + 0.5 km
`213 + 47.8 km!
`55.9 r 6.5 mLAkg ‘-min ‘
`23.6 + 2.1 kg»m 2
`
`Pararnohr
`Sex
`Age
`Weight
`Height
`Distance
`Energy expenditure
`Maximum V02
`BMI
`
`Value (Mean + SD)
`4 females, 5 males
`36 r ('19 yr
`67.6 + 15.7 kg
`173 + 7.4 cm
`230 + 0.3 km
`178 + 51.5 lrcal
`55.1 + 5.5 nl.-kg ‘min ‘
`223 + 4.0 kgm 2
`
`EARBUD-BASED SEHSOR FOR PHYSICAL ASSESSMENTS
`
`Mediche & Science in Sports & Exercise.-,
`
`1047
`
`Copyright 9 2014 by the American Colege of Sports Medicine. Unauthorized reproduction of this article is prohibited.
`
`VALENCELL_0002
`
`

`

`subject underwent the same exercise measurement proto-
`col, including a treadrnill-based cardiopulmonary exercise
`(CPX) test, at 0° incline, to reach \'/02...“. The achievernernt
`of VO2,,,,,, was determined by reaching at least two of the
`three following criteria: plateau in V02 over the last nninute
`of exercise, achievement of at least 1.10 RER, arnd achieve-
`mernt ofat least 17 in perceived exertion on the Borg scale. The
`mean i SD \'/02.,“ values of the training group and the vali-
`dation group wee 55.9 : 6.5 and 55.1 i 5.5 mL-kg”-min‘ ',
`respectively. Benchmark sernsors included a 12-lead ECG for
`measuring HR, a calibrated treadmill for measuring distance
`traveled, arnd a gas-exchange analysis instrument for mea-
`suring TEE and VO;h,,,,,. The earbud sensor served as the
`device under test. All subjects provided infonned consent
`as approved by the investigational review board of the Duke
`University School of Medicine.
`CPX testing. Subjects began the study by first being
`prepped for wearing the benchmark sensors. A Quintonl2-
`lead ECG system was used as a bernchmark for HR, and a
`TnreMax 2400 Pan/oMedies (ParvoMedics, Sandy, UT) gas-
`exchange analysis mouthpiece was used as a benchmark for
`ernergy expenditure and continuous measures of V02. The
`banclnmark sensors were calibrated according to the stan-
`dard rnainternance guidelines of the manufacturers. The sub-
`jects were then fitted with an earbud sansor (Fig. 1) powered
`by the aforementioned PerforrnTek physiological monitor-
`ing technology. Participants were then asked to sit at rest in
`a supine position in a reclining chair for a few minutes while
`wearing the benchmark equipment and earbud sansor. Afia
`the resting period, subjects were instructed to move fiorn the
`chair to the calibrated treadmill and execute the CPX testing
`with graded irnternsity ranging from 0 to 9.1 mph speeds. The
`protocol used cornsisted of 2-min stages, increasing the work-
`load by approximately one metabolic equivalent per stage.
`Measurements fiom the bernchmark sensors and earbud sen-
`
`sor were collected continuously throughout the treadmill run.
`Participants were asked to continue running during each in-
`creasing speed until they were completely exhausted. The
`last 40 s of bernchnnark gas-exchange analysis data were av-
`eraged to determine measured peak V02.
`Earbud sensor. The novel noninvasive earbud sen-
`
`sor (Fig. 1) used in this study was designed by Valencell,
`lnc. (Raleigh, NC). The earbud sensor comprised a sensor
`module, a microprocessor, and a wireless Bluetooth® chipset.
`The optomechanical sensor module, comprising the sensor
`elements, was embedded within the right audio earbud, as
`shown in Figure 1, such that the sensor module would rest
`between the concha and the arntitragus of each subject upon
`earbud placement. The right arnd the left earbuds were de-
`signed to be pluggable to a wireless Bluetooth “medallion”
`via a detachable connector (as shown in Fig. 1). The medal-
`lion housed the nnicroprocessor and the Bluetooth clnipset.
`At the heart of this noninvasive earbud sensor is a highly
`miniaturized optomechanical module (l7—21,23) that mea-
`sures optical and mechanical inforrrnation from the area of a
`user’s ear between the antitragus and the concha. This novel
`
`sensor module comprises an infi'ared light-ernitting diode,
`a photodetector element, a tluee-axis accelerometer, and an
`optomecharnical housing. Designed to fit flush with the body
`of a standard audio earbud, the earbud essentially maintains
`the form factor of a typical audio earbud and does not re-
`quire an ear clip or an in-ear-canal sensor to firnction.
`The optical and mechanical infonnation collected fiom
`the car are sampled via methods akin to reflective PPG and
`three-axis accelerometry, and this sampled infomnation is
`then processed by novel algorithms (l7,l8) coded on finn-
`ware within the microprocessor for extracting weak blood
`flow signals from excessive motion noise. It is well known
`that motion artifacts are the greatest limiting factor to accu-
`rate vital sigrns monitoring via PPG (3,l4,27). However,
`Valencell’s PerformTek biometric algorithms actively pro-
`cess noisy body sigrnals and extract accurate biometrics even
`during aggressive nnnning and PA (23). These biometric sig-
`nals are then combined with contextual accelerometry infor-
`mation within a statistical model to gernerate assessments of
`HR zorne, calories burned, aerobic capacity (VO2,.,,,,,), and
`otha parameters (17-21). A phorne, computer, or otlner mo-
`bile device can communicate directly with the nnicroproces-
`sor via a Bluetooth link. In this particular study, the earbud
`sensor data wae streanned direcfly to a laptop via Bluetooth.
`Statistical methodology. A multiple linear regression
`model had been developed previously by Valencell to pro-
`vide a linear relation between estimated TEE, as predicted
`by the earbud sensor measurements, and the measured TEE,
`as recorded by the benchmark gas-exchange analysis de-
`vice. This linear model comprised fixed and tinne-varying
`terms. The fixed terms included weight (W), age (A), and
`sex (G) having a binary value of 0/1 for womenlmen,
`respectively. The tirne-varying terms included the earbud-
`estimated TEE GEB TEE) and the linear operations of real-
`time PPG and accelerometry (ACC). Although the details
`of the linear model are outside the scope of this article,
`the formalism of the resulting linear equation may be de-
`scribed by EB TEE = f(g(PPG), h(ACC), W, A, G), where
`g and h are functions of PPG and ACC, respectively. It is
`important to note that this lirnear model was directed to-
`ward estimating TEE, arnd not the individual elements of
`resting energy expenditure (REE) or activity-related ernergy
`expenditure (AEE), as TEE is what is measured by the gas-
`exchange arnalysis.
`A separate model had beern previously developed by
`Valencell
`to estimate V‘/O2,,,,,,, based on the HR and
`accelerometry data collected during several prior rounds of
`CPX testing. The methodology behind this \'/02...,“ esti-
`mation is described elsewhere (18), and the equation
`follows the formalism of EB \'/02.,” = j(Max HR,
`Min HR, k(ACC)), where EB \'/02,.” is the earbud-
`derived VO2.,,,,,, Max HR is the maximum reliable HR
`measured by the earbud sensor, Min HR is the minimum
`reliable HR measured by the earbud sensor, and k is a
`fimction of the accelerometer readings measured through-
`out the CPX testing.
`
`1048 Otficl Journal of the American College of Sports Medicine
`
`http://www.acsm-msse.org
`
`Copyright 9 2014 by the American Colege of Sports Medicine. Unauthorized reproduction of this article is prohibited.
`
`VALENCELL_0003
`
`<12
`
`n QL
`
`L!
`:1
`L
`
`QC
`
`an
`LL!
`
`UZ L
`
`L’
`
`Ua
`
`

`

`--ECGHR
`--Earbud (DUI) HR
`—-Activitv
`
`200
`
`180
`
`150
`140
`
`120
`
`3‘
`a
`53
`0»
`5 100
`a:
`ad
`'0
`
`80
`
`30:»
`
`so
`40
`20
`
`2
`
`L8 >
`0
`1-5 E
`1.4
`‘-'1
`3
`1.2 m
`'5.’
`1 <
`0
`0.8 8
`3
`
`0.6 3
`0.4
`9’
`F
`
`0.2 "' O
`
`0
`
`200
`
`400
`
`600
`
`800
`
`1000
`
`1200
`
`1400
`
`Time (5)
`
`FIGURE 2 CPX testing output from a characteristic test. In this
`characteristic CPX test. the time of the progressive exercise test con-
`ducted on a standard treadmill is shown on the abscissa; the PA in-
`tensity using accelerometer counts in arbitrary units (a.u.; green line) is
`shown on the rightward ordinate; the HR in beats per minute from
`either the l’erformTek earbud device (red) or from the simultaneously
`measured ECG (blue) is shown on the leftward ordinate. The mrbud-
`determined HR and the ECC-measured HR show complete alignment
`in this exemplary characteristic test.
`
`Alter the 14-person training data study, the best-fitting
`coefficients for the TEE and V02“, models were deter-
`mined, and the resulting optimized equations were used in
`the nine-person validation data study to estimate TEE and
`V02“, in real time. The resulting earbud-estimated values
`GEB TEE and EB V03“) were then compared with
`barchmark-measured values in accordance with the Bland-
`
`Altrnan plot (1,4).
`
`RESULTS
`
`HR. As previously described, the earbud measurements
`of HR and PA are part of the foundational formulas for
`EB TEE and EB v02...,,.. Therefore, it is important that
`these measurements are accurate. An exemplary character-
`istic plot of real-tirne ECG, PerformTek HR, and h(ACC) for
`a subject undergoing a CPX test is shown in Figure 2. Note
`
`
`
`
`
`200
`
`R-squared = 0.98
`
`SD = 4.4%
`
`50
`
`.
`100
`
`.
`150
`
`0
`
`200
`
`E g
`
`§ 150
`
`EQ
`
`100
`3 so
`-E(0
`Lu
`
`0
`
`n
`
`‘W
`
`mL
`
`U Cm Um Dm Z(
`
`>1
`
`::
`
`that for this test, the benchmark ECG and the earbud HR
`are nearly identical throughout the urn, such that they com-
`pletely overlap each other. Although complete overlap was
`not always observed, complete overlap was typically ob-
`served. Only rarely did the earbud or the ECG diverge to a
`substantial degree, as exemplified by the fight conelation
`shown in Figure 3. Also, on the rare occasions when diver-
`gence was observed, it was ofien attributable to either the
`earbud moving out of the ear or the ECG leads decoupl-
`ing from the subject’s skin. For the sake of objectivity in
`this study, all HR data points measured by the earbud and
`ECG sensors are shown in Figure 3, even for the case
`where earbud or ECG failures are subjectively believed to
`have occurred.
`
`A Bland—Altman plot comparing earbud—estirnated HR
`(EB HR) versus the benchmark 12-lead ECG measured from
`the l4—pason training group is presented in Figure 3. This
`figure illustrates the excellent agreement between EB HR
`and ECG throughout a full range of activity fiom rest to
`>200 bpm; the mean difference (bias) was -0.2%, the SD
`was i4.4%, and the coefficient of determination (R2) was
`0.98. In contrast with other reported optical HR measure-
`ment devices reported in literature (3,l4,27), the EB HR
`measurement is quite robust throughout a firll range of ac-
`tivity because the PerforrnTek biometric signal extraction
`algoritluns are capable of charactaizing motion noise during
`numerous activities and attenuating motion artifacts fiom the
`optical signal in real time.
`In contrast with hip and pocket-wom pedometer-based
`approaches for calculating distance (8,l5,29), the PA level
`measured by the earbud prototype provides a good reference
`for body displacement during walking, jogging, and nrnning
`without requiring knowledge of the user's sex, height, age,
`weight, or fitness. Furthermore, neither a calibration re-
`gimen nor a GPS is required to tune parameters to the
`wearers’ gait. The earbud prototype distance measurement
`was highly accurate, with a bias of 0.3%, an SD of 4.2%,
`and an R2 of 0.93. This distance mezsurement was ob-
`
`tained through a novel transformation of three-axis acceler-
`ometer data, and its accuracy is aided by the sensor location
`at the ear.
`
`2°
`
`.
`
`3
`2: ‘°.
`I
`C
`8
`0 ‘Q,
`5

`E’ 10
`'
`>6 -
`
`'
`o.‘
`J,
`
`
`HR
`—mean
`do
`.
`V
`V —+/-1.96'sD
`53' , .9
`‘ ‘
`X‘
`..
`.
`..
`,
`'
`.3
`’. "" ".“r°:
`C
`=
`.3.’..‘4
`‘O
`o
`. .'
`,
`e
`0‘ “ZN;
`
`_20
`
`50
`
`.
`'
`100
`
`2
`
`O
`
`‘ .
`150
`
`v
`
`.
`
`200
`
`ECG heart rate (bpm)
`
`average HR (bpm)
`
`FIGURE 3 HR using the earbud (device under test |DUT|) and the simultaneously mmsured ECG benchmark. A. Regression relation comparing
`estimated (earbud) versus measured (ECG) HR for all data points collected for each participant. B. The Bland Altman plot of same. All the data
`points were taken from the training data collected during the Duke CPX test. The mean dilference (bias) was 0.2% and the SD was 4.4%. The mean
`is shown by the green line, and the 1.96 SD (95% limits of agreement) boundaries are shown by the red lines.
`
`EARBUD—BASED SBISOR FOR PHYSICAL ASSESSMENTS
`
`Mediche & Science in Sports & Exaciseg
`
`1049
`
`Copyright 9 2014 by the American Colege of Sports Medicine. Unauthorized reproduction of this article is prohibited.
`
`VALENCELL_0004
`
`

`

`able to provide understandable, actionable, and motivational
`feedback to the user, and 4) autonomous and user-fiiendly.
`Today, a variety of commercially available products offer
`step counting and estimated calorie monitoring. Many of
`these solutions have provided value to researchers study-
`ing energy balance and to fitness professionals and clini-
`cians monitoring the progress of exercise and diet plans.
`However, none of these products satisfy all the previously
`mentioned criteria, limiting the effectiveness and scope of
`l) long-term clinical research on energy balance research
`and 2) health and fitness solutions for end users. In con-
`trast, newly developed earbud sensor technology offers the
`promise of meeting these needs, enabling a truly seamless
`energy balance—monitoring platform for use in clinical re-
`search, consumer fitness, clinical assessment of energy bal-
`ance, and mobile health management.
`The feasibility has been established for the highly minia-
`turized, noninvasive optical earbud sensor technology for
`accurately monitoring physiological metrics such as HR,
`TEE, and maximum oxygen consumption (\'/Ohm) through
`the ear. The earbud sensor accurately predicted HR through-
`out all activity levels investigated, fiom rest to peak perfor-
`mance, with a mean difference (bias) of -0.2% and an SD
`of i4.4% when compared with an ECG benchrmrk device.
`Furthermore, real-time algorithms within the earbud sensor
`accurately predicted (a) TEE with a bias of -0.7% and an
`SD of 417.4% and (b) V02“, with a bias of -3.2% and an
`SD of 417.3%. This particular evaluation did not address
`user comfort or battery life but a commercially available
`Bluetooth audio headset, the iriverONT'“, incorporating the
`evaluated Pa'formTek® biometric sensor teclmology adver-
`tises several hours of measurement time while also support-
`ing music.
`The excellent performance of the earbud sensor for ac-
`curately measuring HR throughout extreme PA is especially
`noteworthy. Motion artifacts have been the greatest limita-
`tion to the continuous monitoring of vital signs during ac-
`tivity (3,l4,27), and the ability to accurately monitor vital
`signs with a consumer-priced audio headset is particularly
`irnpactful to public health.
`
`00O
`
`E 20
`3
`
`+ ‘calories
`mean
`—+/-1.96 ~ so
`
`g 10
`
`0
`
`;:_> -10
`‘E
`
`-20
`100
`
`.
`.
`.
`.
`300
`250
`200
`150
`average calories (kcal)
`
`i
`350
`
`FIGURE 4 Bland Altman plots simultaneously comparing the refer-
`ence CPX data and the earbud (device under test |DUT|) benchmark
`data collective. The Bland Altman plot of the difference between the
`energy expenditure (I-IE; kcal) using the reference CPX test and the
`DUT benchmark EE measurements taken from the 14-person CPX
`training group. The mean is shown by the green line. and the L96 SD
`(95% limits of agreement) boundaries are shown by the red lines.
`
`TEE. The EB TEE closely estimated the benchmark TEE
`for the training group data set, with a bias of -0.7% and
`an SD of i7.4% (Fig. 4). The correlation between the
`EB TEE and the benchmark TEE for the validation group
`data set was identical with that of the training data set, with
`a bias of -0.7%, an SD of i7.4%, and an R2 coefficient of
`0.86 (Fig. 4).
`\'Io,,..,,,. The EB vo,...,, closely estimated the bench-
`mark V02,m, for the training group data set, with a bias
`of -0.1%, an so of :8.7%, and an R2 coefficient of 0.36
`(Fig. 5). The correlation between the EB V02,” and the
`barchmark TEE for the validation group data set was simi-
`lar to that of the training data set, with a bias of -3.2% and
`an SD of i7.3%.
`
`DISCUSSION
`
`To satisfy commercial, clinical, and research oriented
`markets for personal energy balance monitoring, a wear-
`able sensor module must satisfy four key criteria. The sensor
`module must be 1) seamless with daily living (comfortable,
`convenient, and socially acceptable), 2) sufficientiy accu-
`rate for multiple life activities (indoors and outdoors), 3)
`
` 7;
`
`O 8
`
`C9
`
`A
`3
`"V 10.
`0-‘
`g
`%
`'6
`Q)
`
`0-
`
`§_,_1o.

`
`,
`
`,
`~ distance
`. —mean
`.
`—+/- 1.96 SD
`
`
`
`.
`
`‘
`
`t
`
`..
`1 .5
`
`4
`
`2
`
`,
`
`2.5
`
`,
`B
`3 30’
`5 20,
`"
`E
`8 10*
`> 0
`Q)
`'
`
`§_,-1o-
`'3-20:
`40
`
`,
`o V02max
`
`—mean
`.
`—+/- 1.96 SD
`
`60
`
`.
`80
`
`.4
`100
`
`,
`C
`? 3°’
`‘-5
`«n 20'
`31:2
`g 10-
`8
`r
`
`.
`
`§-1o-
`U-20:.
`100
`
`
`
`
`,
`,
`
`« calories
`—mean
`.
`—+/- 1.96 SD
`
`
`
`..
`
`150
`
`200
`
`250
`
`average calories (kcal)
`
`
`300
`
`average distance (miles)
`
`average VO2max (ml/kg/min)
`
`FIGURE 5 Bland Altman plots simultaneously comparing reference CPX data and earbud (device under test |DUT|) benchmark data collective. A.
`Distance traveled. B. V0,.” (mL-kg '-min '); EI-I (kcal). Nine subjects were studied in th CPX validation group study. The mean is shown by the
`green line. and the 1.96 SD (95% limits of agreement) boundaries are shown by the red lines.
`
`1050 Otficl Journal of the American College of Sports Mediche
`
`http://www.acsm-msse.org
`
`Copyright 9 2014 by the American Colege of Sports Medicine. Unauthorized reproduction of this article is prohibited.
`
`VALENCELL_0005
`
`<11
`
`UZ L
`
`L’
`
`Ua
`
`n QL
`
`L!
`:1
`L
`
`QC
`
`an
`LL!
`
`

`

`mm
`
`U Cm Um Qm Z(
`
`>1
`
`::
`
`‘W
`
`When compared with the approaches in HR chest strap
`for estimating energy expenditure (5,6), the earbud sensor
`algorithms for estimating TEE and V/02m, are also note-
`worthy. It is likely that the fixed location of the earbud
`with respect to the spine increases the accuracy of activity
`measurements, which feed the TEE and the V02...” models.
`However, some important limitations of these algorithms
`are of note. First, these algorithms have demonstrated sub-
`stantial efficacy in estimating TEE and \702,,,,,,, under
`walking, jogging, and mmring conditions, conditions com-
`mon for clinical CPX evaluations. However, it is yet to be
`determined how accurate these algorithms will be at esti-
`mating these parameters during everyday life activities and
`other exercising regimens, such as weight lifiing, swimming,
`contact sports, daily household activities, and the like. There
`are several studies emphasizing the importance of caution
`when applying V02 estimation models to universal PA
`(9,12,22,25,28,30,32,33). Moreover, it is not clear how well
`these algorithms will predict REE or energy expended during
`sedentary activity, where caloric expenditure is dominated
`by the metabolic rate of the individual as opposed to PA.
`Second,
`the V02,“ range of participants in the valida-
`tion study was relatively small: fiorn approximately 50 to
`65 mL-kg"-min" (Fig. 5). Indeed, the relatively low R2
`coefficient for estimated V02“, and the ostensible non-
`linear bias of Figure 5B together suggest that the accuracy
`of the V02“, model cannot be affirmed with the current
`data set. Finally, to be clinically relevant, the predictions for
`TE and V02“, would ideally be even more accurate, with
`the goal of an SD of less than 5%.
`To address these areas for improvement, future work
`should evaluate the efiicacy ofthe earbud sensor at estimating
`
`REFERENCES
`
`l. Altman DG, Bland JM. Mersurement in medicine: the analysk of
`method comparison studies. Statistician. l983;32(3):307 l7.
`2. Anderson DR, Whitmer RW, Goetzel RZ, et al. The relationship
`between modifiable health risks and group level health care ex
`penditures. Am J Health Promot. 2000;l5( l):45 52.
`3. Asada HH, Shaltis P, Reisner A, Sokwoo R, Hutchinson RC.
`Mobile monitoring with wearable photoplethysmogaphic bio
`sensors. IEEE Eng Med Biol Mag. 2003;22(3):28 40.
`4. Bland JM, Altman DG. Statistical methotk firr essing agree
`ment between two methods of clinical mezsurement. lancet.
`l986;l(8476):307 10.
`5. Brage S, Brage N, Franks PW, et al. Branched equation modeling
`of simultaneous accelerornetry and heart rate monitoring improves
`estimate of directly measured physical activity energy expenditure.
`J Appl Physiol. 2004;96(l):343 51.
`6. Brage S, Brage N, Franks PW, et al. Hierarchy of individual cali
`bration levels for heart rate and accelerometry to measure physical
`activity. J Appl Physiol. 2007:l03:682 92.
`7. Bravata DM, Smith Spangler C, Sundaram V, et al. Using pe
`dometers to increase physical activity and improve health. JAMA.
`2007;298(l9):2296 304.
`8. Crouter SE, Schneider PL, Karabulut M, Basett DR Jr. Validity of
`10 electronic pedometers for measuring steps, distance, and energy
`cost. Mad Sa‘ Sports Exerc. 2003;35(8): 1455 60.
`
`TEE and VO2,,,,,, in a larger cohort group having a broader
`range of aerobic capacity, ranging from approximately 35
`to 70 mL-kg‘ '-min_'. Furthermore,
`the earbud sensor
`should be put to the test of estimating TEE during a broader
`set of activities than simple treadmill testing, using the
`standard doubly labeled water (DLW) methodology as a
`benchmark. In addition, the ability of the earbud sensor to
`estimate the resting metabolic rate (REE) of subjects should
`also be assessed.
`
`lrnproving the accuracy of these assessments will rely on
`1) optimizing algorithms based on a larger study sample of
`subjects exercising in more diverse enviromnents (such as
`daily life activities) and 2) adding additional biometrics to
`the predictive algorithms for energy expenditure and
`\702,,,,,,. Valencell’s PerfcrmTek earbud sensor is comprised
`mostly of novel optomechanics and signal extraction algo-
`rithms. The accuracy of the HR algorithms is so high,
`approaching machine error, such that it is unlikely that ad-
`ditional improvements can be made in the optomechanical
`sensor module for accuracy. Rather, advancements are likely
`to arise from the development of an optimized statistical
`model that incorporates personalized REE estimations into
`the model. The algorithms for estimating personalized REE
`can be developed by evaluating the PPG profile of subjects
`at rest with a gas-exchange analysis benchmark (REE test-
`ing) and by identifying new blood flow profile features that
`correlate with the gas-exchange analysis data.
`
`The validation testing it this research was funded it part by the
`National Institutes of Health via Phase I SBIR 1R43DKo83141 —01A1.
`There are no conflicts of interest.
`The results of the present study do not constitute endorsement by
`the American College of Sports Mediche.
`
`9. Daniels .IT. A physio1ogist‘s View of running economy. Med Sci
`Sports Exenc. l985;l7(3):332 8.
`10. Frmk KL, Stevens VJ, Appel l.J. Axociations of Internet website
`use with weight change in a long term weight loss maintenance
`program. J Med Intenra Rm. 20l0;l2(3):e29.
`11. Gallup Healthways Well Being Index. 2012. Available from:
`http://www.gallup.oom/poll/lS7505/americans exercising sliyrtly
`20l2.aspx
`12. Harris C, Debeliso M, Adams K]. The effects of rtmning speed
`on the metabolic and mechanical energy costs of nrnning. J Exerr:
`Physiol. 2003;6(3):28 37.
`13. Headphones: Ownership & Application Study, 2012. The NPD
`Group. August 2012. p. 4.
`14. Jiang HH, Asada HH, Gibbs P. Active noise cancellation using
`MEMS accelerometers for motion tolerant wearable biosensors.
`
`ln: Cotfaarce Proceedings ofthe IEEE Engineering in Medicine
`and Biology Soeiay. 2004. pp. 2157 60.
`15. Kong YC, Ming S. lrnproving energy expend

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket