throbber

`Neutral Citation Number: [2009] EWHC 1304 (Pat)
`
`
`IN THE HIGH COURT OF JUSTICE
`CHANCERY DIVISION
`PATENTS COURT
`
`
`Case No: HC08 C 00934
`
`Royal Courts of Justice
`Strand, London, WC2A 2LL
`
`Date: 12 June 2009
`
`Before :
`
`THE HONOURABLE MR JUSTICE KITCHIN
`- - - - - - - - - - - - - - - - - - - - -
`Between :
`
`EDWARDS LIFESCIENCES AG
`(a company incorporated under the laws of
`Switzerland)
`
`- and –
`
`COOK BIOTECH INCORPORATED (a company
`incorporated under the laws of the state of Indiana,
`USA)
`
`Claimant
`
`Defendant
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Roger Wyand QC, Piers Acland and Miles Copeland (instructed by Bird & Bird) for the
`Claimant
`Simon Thorley QC and Adrian Speck (instructed by Marks & Clerk Solicitors) for the
`Defendant
`
`
`Hearing dates: 6-8, 11, 12, 14 and 15 May 2009
`- - - - - - - - - - - - - - - - - - - - -
`Approved Judgment
`I direct that pursuant to CPR PD 39A para 6.1 no official shorthand note shall be taken of this
`Judgment and that copies of this version as handed down may be treated as authentic.
`
`
`.............................
`
`MR JUSTICE KITCHIN
`
`
`
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1038, p. 1 of 43
`
`

`

`
`
`
`
`Mr Justice Kitchin :
`
`
`
`1.
`
`2.
`
`3.
`
`This is a patent action in which the Claimant (“Edwards”) seeks revocation of
`European Patent (UK) 1 255 510 (“the Patent”). The Defendant (“Cook”) is the
`proprietor of the Patent and has counterclaimed for infringement.
`
`Edwards manufactures the SAPIEN artificial heart valve which was launched in
`Europe in 2007. It is designed to be compressed onto a balloon catheter for
`percutaneous delivery via the femoral artery. It can also be delivered transapically
`through the side of the chest and into the apex (the bottom of the left ventricle) of the
`heart in patients with severe aortic stenosis. It is primarily used to replace the aortic
`valve but is also suitable for replacement of the pulmonary valve.
`
`Cook alleges the SAPIEN infringes the following claims of the Patent which are said
`to be independently valid: 1, 12, 15, 22, 23, 28 and 31. Edwards denies infringement
`and challenges the validity of these claims and claims 3 and 8 (which are also said to
`be independently valid but not infringed) on the following grounds:
`
`i)
`
`Lack of novelty under section 2(3) of the Patents Act 1977 (“the Act”) in the
`light of WO 01/19285 published on 22 March 2001 (“Thorpe”);
`
`ii)
`
`Obviousness in the light of:
`
`a)
`
`b)
`
`c)
`
`d)
`
`U.S. Patent 5,411,552 published on 2 May 1995 (“Andersen”);
`
`EP 0 856 300 A1 published on 5 August 1988 (“Moll”);
`
`“Aortic and venous valve for percutaneous insertion” by D. Pavcnik et
`al., published in 2000 (“Pavcnik”);
`
`common general knowledge.
`
`iii)
`
`iv)
`
`Insufficiency. Edwards contends the specification of the Patent does not
`disclose the alleged invention clearly enough or completely enough for it to be
`performed arising from the use in claim 1 of the word “substantially”.
`Essentially this is a question of the proper interpretation of the claim.
`
`Added matter. Edwards contends the matter disclosed in the specification of
`the Patent as granted has been extended over the original disclosure in the
`application for the Patent as filed. There are two aspects to the objection. One
`arises from the use in claim 1 of the word “substantially” and the other turns
`on the proper interpretation of claim 3.
`
`Witnesses
`
`Each of the parties called two expert witnesses, an interventional cardiologist and a
`bioengineer. On behalf of Edwards, I heard evidence from Dr Nigel Buller and Dr
`Rodolfo Quijano.
`
`Dr Buller is a consultant cardiologist in private practice. Until January 2008, he was
`Head of Interventional Cardiology at the Queen Elizabeth Hospital, Birmingham. The
`Queen Elizabeth has one of the leading cardiology departments in the UK and one of
`
`4.
`
`5.
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1038, p. 2 of 43
`
`

`

`
`
`
`
`6.
`
`7.
`
`8.
`
`9.
`
`10.
`
`11.
`
`
`
`only five centres that provide fully comprehensive adult cardiological services. Dr
`Buller has extensive experience of catheterization procedures, including balloon
`angioplasty and stent implantation and throughout his career has had a close working
`relationship with many of the major medical device manufacturers.
`
`Cook does not suggest I should reach a general conclusion adverse to Dr Buller but
`invites me to say that he may have lost total objectivity in a limited number of
`instances. I decline that invitation. Dr Buller was measured, careful and precise in
`expressing his opinions and I have found his evidence of great assistance.
`
`Dr Quijano has been involved in the design and development of biological and
`mechanical replacement heart and venous valves for more than 35 years. Cook makes
`no criticism of Dr Quijano, and rightly so. He clearly has a passion for and a deep
`understanding of the technical issues involved in the design of replacement cardiac
`and venous valves.
`
`On behalf of Cook, I heard evidence from Professor Martin Rothman and Professor
`David Williams.
`
`Professor Rothman is a consultant cardiologist and the Director of Cardiac Research
`& Development at Barts and the London NHS Trust and Honorary Professor of
`Interventional Cardiology at Queen Mary, University of London. Interventional
`cardiology has been the focus of Professor Rothman’s entire career and he is
`recognised as one of its pioneers. He has worked with cardiovascular stents since the
`early 1980s and over the years has advised many different companies operating in the
`pharmaceutical and medical device sectors in relation to a wide range of devices used
`in conjunction with interventional cardiology.
`
`Edwards accepts that Professor Rothman is a skilled and expert cardiologist but
`contends his evidence was partisan, as illustrated by a marked shift in his opinions
`from those he held in an earlier case between Edwards and a company called
`CoreValve. I think it fair to say the opinions expressed by Professor Rothman in his
`reports in the two cases are indeed different in material respects and this formed the
`basis of a good deal of his cross examination. However, as Cook submits, opinions
`may change in the course of a case, particularly after cross examination, and I accept
`that in formulating his reports in this case Professor Rothman may have given further
`consideration to the abilities of the ordinary skilled person. Importantly, I believe
`Professor Rothman answered the questions put to him fairly and frankly and I found
`his opinions cogent and reasonable.
`
`Professor Williams is currently Professor and Director of International Affairs at the
`Wake Forest Institute of Regenerative Medicine in North Carolina. He is also
`Visiting Professor in the Christiaan Barnard Department of Cardiothoracic Surgery at
`the University of Cape Town. His career over the last forty years has been devoted to
`the fields of bioengineering, biomaterials science and regenerative medicine. Among
`his many activities he has been directly concerned with the development of new
`materials for use in surgically implantable heart valves.
`
`12.
`
`Edwards says Professor Williams did not seem to appreciate the role of the skilled
`person in his approach to the prior art and appeared reluctant to attempt to correct
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1038, p. 3 of 43
`
`

`

`
`
`
`
`13.
`
`
`
`deficiencies so as to make it work. I reject this criticism. I found Professor Williams
`to be careful and fair in addressing the questions put to him.
`
`Edwards also adduced evidence of fact from Mr Stanton Rowe, an employee of
`Edwards, who was involved in the development of the SAPIEN. Mr Rowe’s evidence
`was directed to the suggestion made by Professor Rothman in his first report that it
`took ten years of research to develop the ideas described in Andersen into the
`SAPIEN. He was not cross examined and his evidence ultimately played no real part
`in the matters I have to decide.
`
` The skilled person
`
`14.
`
`15.
`
`16.
`
`There was little between the parties as to the identity of those persons to whom the
`Patent is addressed. Professor Rothman and Professor Williams considered the Patent
`is directed towards a skilled team comprising an interventional cardiologist (in so far
`as it concerns heart valves) or a general vascular surgeon (in so far as it concerns vein
`valves) and, in either case, a bioengineer. Professor Rothman considered the team
`might also consult a cardiac surgeon in order to find out about contemporary work
`with surgically implantable replacement heart valves. Professor Williams elaborated,
`and I accept, that in practice a number of engineers might be involved in the team,
`depending on their specific areas of expertise. For example, one might have
`particular experience of stent design, another experience of the design of cardiac
`valve replacements and a third experience of biomaterials. He too considered that a
`cardiac surgeon would be involved in order to provide experience of some of the
`practical problems encountered in using surgically implantable valves.
`
`Dr Buller believed that the team would have included an interventional cardiologist
`and a medical device designer familiar with the design of stents and implantable
`valves and the materials used to make them.
`
`In the light of all this evidence I am content to adopt the formulation of the skilled
`team propounded by Professor Rothman and Professor Williams, subject to the
`following qualification. I am entirely satisfied that the team would have contained or
`at least consulted with a person familiar with the design of implantable surgical heart
`valves.
`
` Common general knowledge
`
`17.
`
`There was no real dispute as to much of the common general knowledge and the
`following description is drawn largely from the reports of the experts.
`
`The cardiovascular system
`
`18.
`
`The cardiovascular system is divided into the pulmonary circulation which supplies
`blood to the lungs and the systemic circulation which supplies blood to the rest of the
`body. The heart lies at the centre of the system. It pumps blood through the blood
`vessels by repeated rhythmic contractions and it consists of four chambers, two atria
`and two ventricles, as shown in the diagram below:
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1038, p. 4 of 43
`
`

`

`
`
`
`
`
`
`
`
`19.
`An enlarged section of the aortic valve may be represented like this:
`
`
`
`20.
`De-oxygenated blood from the body is collected in the right atrium, passes through
`the tricuspid valve into the right ventricle and is then pumped through the pulmonary
`artery into the lungs where carbon dioxide is removed and oxygen absorbed. As the
`right ventricle contracts, the tricuspid valve closes, ensuring that blood is not injected
`back into the right atrium. At the same time the pulmonary valve opens allowing the
`blood to flow from the right ventricle into the pulmonary artery.
`
`Blood returns to the heart from the lungs through the pulmonary vein and it collects in
`the left atrium. From the left atrium the blood flows to the left ventricle through the
`mitral valve. When the left ventricle contracts, the mitral valve closes, the aortic
`valve opens and the blood is duly pumped through the aorta to the body. The
`pulmonary valve and aortic valve prevent blood returning to the ventricles from the
`pulmonary artery and aorta respectively.
`
`The enlarged section of the diagram of the heart set out above depicts the arrangement
`of the aortic valve, a matter of particular importance in this case. The aortic valve sits
`in the aortic valve annulus, a fibrous ring at the junction between the left ventricle and
`
`21.
`
`22.
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1038, p. 5 of 43
`
`

`

`
`
`
`
`23.
`
`
`
`the aorta immediately below the sinuses. The aortic valve itself has three leaflets (or
`cusps) which are half moon shaped. As mentioned, when the left ventricle contracts,
`the pressure inside the ventricle increases until it is greater than in the aorta, at which
`point the aortic valve opens. When the ventricular contraction ends, pressure in the
`left ventricle rapidly drops. When it falls below the pressure in the aorta, the leaflets
`of the aortic valve collapse and come together along their edges (commissures) and
`flow of blood from the aorta back to the heart is prevented. Sitting within the aortic
`sinuses and within a few millimetres of the leaflets are the coronary ostia, which are
`openings that lead to the coronary arteries. It is crucial that these are not blocked
`when a valve is replaced because the coronary arteries provide the heart muscle with
`blood.
`
`Veins in the limbs also have valves, called venous valves, which prevent blood
`flowing backwards and pooling in the extremities due to the effects of gravity.
`Venous valves have two leaflets. Replacement of faulty veins and venous valves has
`been the subject of much experimentation, but is not yet done routinely in clinical
`practice with treatment primarily centering on removal of abnormal veins or systemic
`treatment with anticoagulants. As Professor Rothman explained, there has been little
`commercial or clinical incentive to develop replacement vein valves and such
`development
`that
`there has been has
`lagged behind
`the development of
`percutaneously delivered heart valves.
`
`Cardiac surgery and prosthetic valves
`
`24.
`
`Surgeons have been replacing diseased or malfunctioning heart valves for over 40
`years. They have used for this purpose a range of prosthetic valves, both mechanical
`and biological.
`
`25. Mechanical replacement heart valves are generally made from a combination of
`metal, carbon and plastic and typically provide a valve function through a tilting disc
`or a ball moving within a cage. They have a long life span but patients suffer an
`increased risk of thrombus formation which requires them to undergo life-long
`anticoagulation therapy.
`
`Biologically derived (bioprosthetic) valves attempt to replicate more closely the
`structure and dynamics of a physiological heart valve. They are made of tissue,
`generally mounted on a textile cuff or metallic or plastic frame and fall into three
`categories: homograft (human whole valves), xenograft (animal whole valves) and
`fabricated (valves tailored from animal pericardium, the tissue that covers the outside
`of the heart). The latter two categories are those of most importance in the context of
`the present case.
`
`Xenograft valves are normally of porcine origin, but can also be of equine or bovine
`origin. The valve is physically removed from the animal and treated chemically in
`order to make the biological tissue immunologically inert and sterile and to improve
`its mechanical properties. It is then attached to a textile cuff allowing it to be sutured
`to the heart tissue or mounted on a frame which provides some mechanical support.
`Such a frame is usually referred to as a stent. In 2000, a well established bioprosthetic
`valve using a porcine valve was called the “Hancock”. In this device the valve is
`fixed to a non-collapsible stent covered with fabric, allowing it to be sutured into the
`patient around its circumference. The valve has three leaflets which, when the valve
`
`26.
`
`27.
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1038, p. 6 of 43
`
`

`

`
`
`
`
`28.
`
`
`
`is closed, meet at their free margins, that is to say the edges which are free to move
`from the centre of the valve towards the circumference of the valve when opening and
`free to return to the centre when closing. The line at which any two of these free edges
`meet is referred to as the commissure or line of coaption.
`
`In the case of fabricated valves, leaflets, normally three in number, are fashioned from
`a sheet of pericardium and again attached to a frame and sewing cuff. In 2000, one of
`the most successful fabricated bioprosthetic heart valves was the “Carpentier-Edwards
`pericardial aortic prosthesis”, also known as the “Perimount”. It was implanted as
`early as 1980 in France and approved for use in the US in 1991. It looks like this:
`
`
`
`29.
`Like the Hancock, the Perimount comprises a non-collapsible stent, covered with
`fabric, which provides a means to suture it into the patient. The stent has three
`projecting portions known as commissural posts to which the leaflets are connected at
`the periphery of their commissures. The leaflets are also sutured to the fabric covered
`stent along the entirety of the inflow side of the valve (the margin of attachment) to
`preserve the valvular mechanism geometry and ensure the valve does not leak
`peripherally.
`
`Pericardium is the only natural tissue that has ever been used commercially to
`fabricate a bioprosthetic valve. Its advantages are numerous. It is available in
`relatively large quantities, permits the production of leaflets that have uniform
`thickness, strength and flexibility, can be cut to any desired size and, after suitable
`chemical treatment, possesses physical properties that closely resemble those of the
`leaflets of human valves. Moreover, pericardium is biocompatible and exhibits low
`thrombogenicity. For all these reasons, for many years before 2000, pericardium was
`the only tissue used for the production of commercialised fabricated bioprosthetic
`valves. However, such valves do suffer from the drawback that they have a tendency
`to denature or calcify, which affects their long term performance.
`
`As Professor Williams explained, by 2000, attempts had been made to develop a
`surgically implantable polymer leaflet heart valve which was seen as having the
`potential to avoid the difficulties of thrombosis caused by the mechanical valves and
`
`30.
`
`31.
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1038, p. 7 of 43
`
`

`

`
`
`
`
`
`
`calcification which tends to occur with bioprosthetic valves. However, no
`commercial polymer leaflet surgical valves existed at that time.
`
`Interventional cardiology
`
`32.
`
`33.
`
`Surgical heart valve replacement involves a major operation and is not suitable for all
`patients. However, from the 1960s a new branch of medicine emerged known as
`interventional cardiology. This is the practice of percutaneously treating problems
`within the heart and associated vessels and is the province of physicians rather than
`surgeons. Interventional procedures are carried out using a catheter to access the site
`in the heart or vasculature where the intervention is to be performed.
`
`As Professor Rothman elaborated, in 1977, Andreas Gruentzig performed the first
`human balloon angioplasty procedure in which a catheter carrying a balloon was
`inserted into an occluded human coronary artery and then expanded to force the artery
`open. By 1990, balloon angioplasty and two related techniques called valvuloplasty
`(using inflation of a balloon catheter to try to open up a stenotic (narrowed) heart
`valve and improve blood flow) and atherectomy (using a high speed rotating device or
`a directional slicing device to remove plaque from the inside of an artery) were
`regularly being undertaken.
`
`34. Most interventional cardiology is performed percutaneously using a needle inserted
`into the femoral or radial artery. But it is also possible to cut through the skin over
`the vessel using a procedure known as “cut-down”. Once access to the artery is
`secured, the catheter is passed to the heart against the blood flow in what is known as
`a “retrograde” approach. Access to the heart can also be achieved by means of an
`“antegrade” approach, that is to say passing the catheter in the same direction as the
`blood flow. In this case the catheter is introduced into a peripheral vein and then
`advanced along the vena cava to the right side of the heart. If access to the left side of
`the heart is required then the catheter must be fed through the wall (called the septum)
`which lies between the two atria of the heart. This technique is used to perform mitral
`valvuloplasty.
`
`In the course of the 1980s and 1990s a great deal of work was also being carried out
`into the design of expandable stents for translumenal implantation. These were
`developed to scaffold the internal surface of an artery, initially to prevent an acute
`closure at the time of a balloon angioplasty procedure, particularly of the coronary
`artery. However, in the 1990s two major randomised trials known as Benestent and
`Stress showed that the use of stents also resulted in reduced occurrence of re-
`narrowing (restenosis) compared with patients receiving balloon angioplasty. As a
`result, by 2000, stents were being used electively with balloon angioplasty in the
`majority of cases. They also allowed interventional cardiologists to attempt
`angioplasty in higher risk and more diseased vessels because they knew that stents
`had the capacity to prevent short-term and long-term complications.
`
`Stents essentially fall into two categories, those which are balloon expandable and
`those which are self-expanding. Balloon expandable stents are compressed around a
`balloon and inserted into a peripheral vessel by catheter. Once the balloon
`expandable stent reaches its destination the balloon is expanded to force open the
`stent by plastic deformation. The balloon is then deflated and the catheter withdrawn.
`Some, like the Palmaz-Schatz had a slotted tube design:
`
`35.
`
`36.
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1038, p. 8 of 43
`
`

`

`
`
`
`
`
`
`
`37.
`
`
`Others, such as the Gianturco-Roubin had an expandable wire coil design:
`
`
`
`38.
`Self-expanding stents are made of a spring or of a “memory metal” such as nitinol.
`These require a sheath to maintain the stent in its compressed form during delivery.
`Once the stent reaches its desired location the sheath is withdrawn and the stent
`expands. One of the first self-expanding stents was the Gianturco Z-stent, which was
`first used in the mid-1980s. It has a “zigzag” design and, in a later modification,
`multiple zigzags were joined together by metal struts or monofilament line to provide
`a greater degree of stability:
`
`
`
`
`39.
`By January 2000, many devices consisted of a number of rings joined together and
`those in the art tended to describe the whole of any such device as a single stent,
`irrespective of how many rings it might contain.
`
`40.
`
`Clearly, stents of different lengths and diameters may be required for different
`applications. By 2000, it was the general practice to “size” the stent to a diameter
`approximately 10 to 20% greater than the normal diameter of the treated vessel so as
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1038, p. 9 of 43
`
`

`

`
`
`
`
`41.
`
`
`
`to ensure it would remain in place once deployed and leave the lumen of the vessel
`unobstructed.
`
`Finally, I should mention that for many years prior to 2000 sheaths had been used to
`cover or contain stents prior to deployment. It was also well known to cover the
`outside of stents with bio-compatible material such as Dacron, for example to create
`stent grafts used to support or isolate a weak portion in a vessel, such as an arterial
`aneurysm.
`
`The Patent
`
`The Patent opens with a description of the “Technical field of the invention”. It states
`that the invention includes a medical device and, more specifically, a valve found
`generally within a frame which, in preferred devices, comprises a radially expandable
`stent which can be delivered through a delivery device such as a catheter.
`
`Paragraph [0002] explains that the closest prior art is EP-A-0808614 (the “614
`application”) which, it is said, discloses the preamble to claim 1. The 614 application
`relates to a self-expandable stent with a tri-leaflet or bi-leaflet valve member,
`preferably “made of parts from living organisms such as a valve from a pig or a
`pericardium from a cow”.
`
`The “Background of the invention” is set out from paragraphs [0003] to [0007]. Two
`types of known replacement valves are described: mechanical devices with moving
`ball valves which are susceptible to clot formation and problems associated with long-
`term wear and tear; and biological valves which suffer from a variety of problems
`including the supply of valves, immune response and problems associated with
`positioning. The Patent explains there is therefore a need for alternative and
`improved devices and methods of providing valvular function within vessels of the
`body.
`
`Paragraph [0008] contains a “Summary of the Invention”. It discloses a medical
`device comprising a frame with a valve located within it. The frame comprises a
`radially-expandable stent (including especially a self-expanding stent) which can be
`delivered using a catheter and then deployed and expanded at a target site in a body
`lumen such as an artery or vein. A preferred use is for the treatment of incompetent
`veins in the legs or feet.
`
`There then follows a “Detailed description of the invention”. Paragraph [0010]
`explains that a valve assembly may have two or more leaflets or cusps. The structure
`of a typical stent of the invention is depicted in figures 1-3 and described in
`paragraphs [0011] to [0013]. One embodiment is said to be a self-expanding stent
`such as the Gianturco and the figures depict a simple arrangement of a cylinder
`formed by a wire bent or otherwise formed into a zigzag configuration. The
`specification explains that the bends at the proximal and distal ends of the stent may
`be connected by sutures which can be used to adjust the size of the stent lumen upon
`expansion.
`
`Paragraphs [0014] to [0017] explain how the valve may be fashioned from a sheet of
`valve material draped over the stent lumen and then pushed down into its interior.
`According to the invention, the valve material is said to be a collagen containing bio-
`
`42.
`
`43.
`
`44.
`
`45.
`
`46.
`
`47.
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1038, p. 10 of 43
`
`

`

`
`
`
`
`
`
`material comprising pericardium which is then fixed to the stent frame by a variety of
`well known means including sutures, adhesives and folds. Connection of the valve to
`the frame is shown in figures 6A and 6B and the Patent explains that it may be
`sutured at its distal and proximal ends.
`
`48.
`
`Paragraph [0018] relates that once the sutures are generally in place, the valve sheet
`will form a valve pocket, as shown in figure 6B:
`
`
`
`49.
`The pocket has a valve apex (50) which extends inside the stent lumen and may be
`sutured to the distal end of the stent frame. There is a part of the valve that will form
`a central valve portion (49) that is not directly sutured to the stent, but otherwise the
`valve is sutured around its proximal perimeter to the proximal end of the stent. The
`valve portion (49) forms the valve opening (52) through which fluid can pass as it
`flows from the distal to the proximal end of the device. However, if the flow is
`reversed then the valve pocket (46) fills and the fluid pressure causes the valve
`portion (49) to extend outwards and, when it does so, to contact the other leaflets or
`cusps and so form a seal to stop or impede fluid flow.
`
`50.
`
`Figure 8, reproduced below, illustrates the valve set in the stent with its distal apex
`(50) sewn to a distal bend of the stent with a suture (40). It also shows the proximal
`perimeter of the valve connected to the proximal portion of the stent with two sutures
`(44):
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1038, p. 11 of 43
`
`

`

`
`
`
`
`
`
`
`
`51.
`Paragraph [0021] continues that the valve opening is actually created in the final step
`of preparation of the preferred device. First, a second valve pocket is made by pushing
`the same sheet of valve material down into the interior on the other side of the stent.
`The two valve pockets are now sitting side by side. The opening can then be created
`by cutting a slit in the sheet which can be sized according to the intended flow rate of
`the passing fluid. The Patent also recognises that opening and closing the valve may
`cause increased wear and tear at the corners of the opening and, for this reason,
`reinforcements may be provided in the form of sutures, as illustrated at (53), or by the
`use of adhesives or any other material or mechanism that permits increased structural
`integrity. The Patent also explains in paragraph [0033] that the slit may terminate
`several millimetres (say 1 to 5 mm) before reaching the edge.
`
`52.
`
`There then follows a description of how the devices of the invention may be made to
`different sizes. Paragraph [0022] explains this may be achieved either by elongating
`the length of the struts of a single stent or by joining a number of stents together (by,
`for example, sutures). It is preferred that the overall length of the device provides an
`aspect ratio (length to expanded diameter) sufficiently high to permit proper
`alignment of the device and that aspect ratios of length to expanded diameter of 1:1 or
`greater are preferred. It is to be noted, however, that in devices comprising multiple
`stents there is no requirement that the individual stents should themselves be of any
`particular length. The teaching of the Patent is simply that the length of the whole
`device should be appropriate for its intended application and its aspect ratio should be
`such as to allow proper alignment.
`
`53.
`
`A variety of multiple stent structures are then described and depicted but some are
`said not to be part of the invention, a reflection of the citation by the examiner of the
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1038, p. 12 of 43
`
`

`

`
`
`
`
`
`
`614 application. Thus figure 12, described in paragraph [0028], is said to be a multi-
`stent device of the invention:
`
`
`54.
`As is figure 17:
`
`
`
`
`
`
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1038, p. 13 of 43
`
`

`

`
`
`
`
`55.
`
`Conversely, figure 15, which comprises two stents, with the valve extending half way
`down the upper stent, is expressly said by paragraph [0023] to be not part of the
`invention:
`
`
`
`
`
`56.
`Likewise figure 19, described in paragraph [0037], is not part of the invention. In this
`case the valve (41) begins in the second stent (6) and extends into the third stent (61):
`
`
`
`
`
`
`
`Edwards Lifesciences Corporation, et al. Exhibit 1038, p. 14 of 43
`
`

`

`
`
`
`
`57.
`
`58.
`
`
`
`There is one other element of the description to which I should refer. The Patent
`explains in paragraphs [0029] to [0031] that the outside of the stent may be wholly or
`partly covered by a sheath. So, for example, excess valve material may be folded over
`to increase the structural integrity of the device and to present a smoother surface to
`the body upon implantation. Alternatively the sheath may be made of a synthetic
`material such as Dacron.
`
`Before turning to the claims, I think it is of some note that the Patent contains no
`experimental results and no detailed discussion of how the invention may be
`implemented. It assumes, for example, the skilled person can work out the necessary
`structural characteristics of the stent, such as its length and diameter (both crimped
`and expanded) and the dimensions and configuration of its struts and the thickness,
`strength and resilience of the wire from which it is to be made.
`
`The claims - interpretation
`
`59.
`
`The parties sensibly agreed a breakdown of the integers of claim 1:
`
`[A]
`
`[B]
`
`[C]
`
`[D]
`
`[E]
`
`[F]
`
`A stent valve, suitable for placement in a vessel, the
`vessel further having a diameter (84) and an inner
`lumenal surface, comprising:
`
`a) a radially expandable stent (20) having a proximal
`stent end (31) and a distal stent end (33),
`
`the stent having an expanded diameter (86) sized to
`permit contact with an inner lumenal surface of the
`vessel;
`
`b) a valve (41) having a proximal valve end (48) and a
`distal valve end (50),
`
`the valve being at least partially located within an
`inner portion of the stent,
`
`is formed with a collagen
`the valve
`wherein
`containing bio material (38),
`
`characterised in that
`
`[G]
`
`[H]
`
`[I]
`
`[J]
`
`said collagen containing bio material comprises
`pericardium
`
`and extends within said stent (20) substantially from
`said proximal stent end

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket