throbber
Journal of Immunological Methods 243 (2000) 77–97
`
`www.elsevier.nl/locate/jim
`
`A practical approach to multicolor flow cytometry for
`immunophenotyping
`
`b ,*
`a ,1
`Nicole Baumgarth , Mario Roederer
`aDepartment of Genetics, Stanford University Medical School, Beckman Center B-007, Stanford, CA 94305-5318,USA
`bDepartment of Stomatology,UCSF, San Francisco, CA,USA
`
`Abstract
`
`Through a series of novel developments in flow cytometry hardware, software, and dye-chemistry it is now possible to
`simultaneously measure up to 11 distinct fluorescences and two scattered light parameters on each cell. Such advanced
`multicolor systems have a number of advantages over current two- and three-color flow cytometric measurements. They
`provide a large amount of novel information for each sample studied, an exquisitely accurate quantitation of even rare cell
`populations, and allow identification and characterization of novel cell subsets. In particular, this technology is proving
`crucial to identifying functionally homogeneous subsets of cells within the enormously complex immune system; such
`identification and enumeration is important for understanding disease pathogenesis. However, multicolor flow cytometry
`comes with a new and sometimes difficult set of technical problems that must be overcome by users to derive meaningful
`results. In this manuscript, we describe the basic aspects of multicolor flow cytometry, including the technical hurdles and
`artefacts that may occur, and provide some suggestions for how to best overcome these hurdles. While inspired by the
`11-color technology that we currently use, these principles apply to all flow cytometric experiments in which more than one
`fluorescent dye is used.
`2000 Elsevier Science B.V. All rights reserved.
`
`Keywords: Flow cytometry; Immunofluorescence; Immunophenotyping; Multiparameter flow cytometry
`
`1. Introduction
`
`Abbreviations: APC, allophycocyanin; A595, Alexa 595; CasB:
`Cascade Blue; CasY, Cascade Yellow; Cy5.5APC, Cy7APC,
`Cy5.5, Cy7 conjugates of allophycocyanin; Cy5PE, Cy5.5PE,
`Cy7PE, Cy5, Cy5.5, and Cy7 conjugates of phycoerythrin;
`Cy5.5PerCP: Cy5.5 tandem conjugate of PerCP; FITC, fluores-
`cein; PBP: phycobiliprotein (i.e., PE or APC); PE, phycoerythrin;
`PerCP: peridinin chlorophyll protein; PFC, polychromatic flow
`cytometry; PMT, photomultiplier tube; TR, Texas Red; TRPE,
`Texas Red-conjugated phycoerythrin; SA, streptavidin
`*Corresponding author. Tel.: 11-415-514-0395; fax: 11-415-
`476-4204.
`E-mail address: roederer@drmr.com (M. Roederer).
`1
`Present address: Center for Comparative Medicine, University
`of California, Davis, CA 95616, USA.
`
`Since the earliest application of flow cytometry to
`the study of cells, there has been a drive to increase
`the number of distinct measurements for each cell.
`This developmental effort blossomed in the late
`1990s, when the number of independently measur-
`able ‘colors’ (each color corresponds to a distinct
`fluorescence-based measurement of a cellular protein
`or function) increased from five to 11 (Roederer et
`al., 1997; Bigos et al., 1999).
`The success of this developmental effort was due
`to the coordinated development of new hardware,
`new fluorochromes, and new software analysis tools
`
`0022-1759/00/$ – see front matter
`PII: S0022-1759( 00 )00229-5
`
`2000 Elsevier Science B.V. All rights reserved.
`
`Page 1 of 21
`
`Fluidigm
`Exhibit 1011
`
`(cid:211)
`(cid:211)
`

`
`78
`
`N. Baumgarth, M. Roederer / Journal of Immunological Methods 243(2000)77–97
`
`that significantly increase the quality and quantity of
`measurements. This increase comes with a price,
`however, as this new technology has its own set of
`technical problems and difficulties that users must be
`aware of and must overcome in order to derive
`meaningful results. Nonetheless, once these hurdles
`are overcome, this new technology is well worth the
`effort, as the information obtained from the measure-
`ments is not only novel but could not be obtained
`otherwise using standard flow cytometric techniques.
`As we outline below, setting up multicolor flow
`cytometry is not simply achieved by adding new
`reagents to existing reagent combinations, but re-
`quires a more involved process of quality control,
`optimization, and ‘debugging’. Therefore,
`to dis-
`tinguish multicolor flow cytometry with its unique
`benefits and technical problems from current stan-
`dard flow cytometric technologies (i.e., using four or
`fewer fluorescent dyes), we refer to it as ‘poly-
`chromatic flow cytometry’ (PFC) and will use the
`term throughout this manuscript.
`Flow cytometers capable of collecting data for
`more than three or four colors are now becoming
`more prevalent, as manufacturers have recognized
`the significant demand for the types of analysis
`afforded by this technology. Given the bewildering
`array of fluorochromes, lasers, hardware, and soft-
`ware that might be used in PFC, we outline here the
`fundamental requirements,
`interactions, and prob-
`lems associated with setting up this technology, so
`that users can make educated decisions about instru-
`ment requirements and the design of their experi-
`ments. Finally, we provide some examples in which
`this technology has been of particular benefit. We
`hope to provide with this brief review some practical
`tips and encouragement for those thinking of expand-
`ing their current flow cytometric measurements. The
`benefits of true multicolor flow cytometry make this
`technique a particularly useful and probably soon an
`irreplaceable tool for the study of cell biology and
`immunology.
`
`2. Fluorochromes
`
`The ability to measure multiple parameters of each
`cell is limited by the number of fluorochromes that
`
`can be simultaneously measured. The 11-color PFC
`that is currently in routine use at Stanford uses dyes
`excited by three different laser lines. The excitation
`and emission spectra of these dyes and the filters that
`were chosen to collect the emitted light from these
`dyes are shown in Fig. 1.
`
`2.1. Characteristics of useful fluorochromes
`
`When designing experiments for the flow cytome-
`ter that include the use of new dyes, careful consid-
`eration must be given to the choice of fluorochromes.
`Desirable fluorochromes for cytometric technologies
`have several properties:
`they (i) are biologically
`inert;
`(ii) have high cell-associated fluorescence
`intensities (‘bright’); (iii) exhibit little spectral over-
`lap amongst each other; and (iv)
`for
`immuno-
`phenotyping, are easily conjugated to monoclonal
`antibodies.
`
`2.1.1. Biological inertness
`Most of the fluorochromes that are currently in use
`are biologically inert:
`i.e.,
`they do not affect
`the
`cells, nor do they bind directly to cellular elements.
`There are, however, exceptions to this:
`the most
`common example is that of the ‘background’ binding
`of Cy5PE (and other Cyanine–PBP tandem dyes) to
`monocytes and B cells. This background binding is
`variable between species and can be extremely high
`in some instances. For example, Cy5PE binds strong-
`ly to B cells in mice with autoimmune disorders
`(e.g., non-obese diabetic mice). While there are
`methods available to reduce this background, it is of
`concern when the particular cell types being studied
`are those that
`interact
`‘nonspecifically’ with the
`fluorochrome.
`
`2.1.2. Cell-associated fluorescence intensity
`With regard to the high fluorescence intensities or
`the ‘brightness’ of a fluorochrome, it should be noted
`that the characterization of a fluorescence signal as
`‘bright’, i.e., the difference between the unstained
`and the stained cells,
`is still empirical. ‘Bright’
`signals result from fluorochromes with the following
`characteristics: (1) a high extinction coefficient; (2) a
`high quantum yield; (3) an emission spectrum over-
`lapping as little as possible with cellular auto-
`
`Page 2 of 21
`
`Fluidigm
`Exhibit 1011
`
`

`
`N. Baumgarth, M. Roederer / Journal of Immunological Methods 243(2000)77–97
`
`79
`
`Fig. 1. Shown are excitation and emission spectra of the 11 dyes that we currently use for polychromatic flow cytometry (PFC). Indicated
`are the wavelengths of the different laser lines used for their excitation (407, 488 and 595 nm; other laser lines that can be used for some of
`the dyes are shown for reference), and the filters chosen for optimal light collection and minimal spillover. The fluorochromes, laser lines
`and filters are currently in routine use on a modified three-laser hybrid Becton-Dickinson/Cytomation flow cytometer, described previously
`(Roederer et al., 1997; Bigos et al., 1999).
`
`Page 3 of 21
`
`Fluidigm
`Exhibit 1011
`
`

`
`80
`
`N. Baumgarth, M. Roederer / Journal of Immunological Methods 243(2000)77–97
`
`fluorescence; (4) measurable with high sensitivity
`detectors; and (5) the ability to conjugate multiple
`fluorochromes to each detecting unit (e.g., a high
`ratio of fluorochrome to antibody).
`The differences in brightness can be quantitated
`experimentally by conjugating the same monoclonal
`antibody to the various fluorochromes at optimized
`ratios. Fig. 2 shows histograms of human peripheral
`blood lymphocytes that were stained with 11 differ-
`ent conjugates of an anti-CD8 monoclonal antibody,
`one conjugate for each of the 11 fluorochromes that
`we currently use in PFC. As shown in this figure, the
`‘brightness’ is affected not only by the intensity with
`which the positive cells are stained, but also by the
`background of the negative cells. Although some of
`the conjugates appear to stain much ‘brighter’ than
`
`others, all of these fluorochromes are useful for
`clearly distinguishing the positively stained cells
`from the unstained cells.
`It is important to point out that the brightness of a
`fluorochrome will differ depending on the instrument
`used. For example, conjugates of antibodies to PE
`generally result in much brighter staining when cells
`are analyzed with typical benchtop cytometers using
`flow cells as compared to cytometers using jet-in-air
`detection. This is due to a number of differences in
`the optical paths that affect the efficiency with which
`emitted light is collected by the detectors; however,
`other differences (notably the laser power) also
`influence sensitivity. Therefore, reagent ‘brightness’
`is a relative measurement and must be assessed by
`users on any given instrument.
`
`Fig. 2. An anti-human CD8 monoclonal antibody was conjugated at optimal fluorochrome–antibody ratios to each of the indicated
`fluorochromes (Roederer, 1997a,b). Human peripheral blood lymphocytes, purified by gradient centrifugation over Ficoll, were stained with
`each of the indicated anti-CD8 conjugates and with an antibody to the pan-T cell marker CD3, conjugated either to FITC (for CasB and
`CasY graphs) or to CasB (all others). Shown are histograms of cells that stained positively for CD3. The dashed line indicates levels of
`staining for cells stained with anti-CD3 only, and the solid line indicates the anti-CD8 fluorescence for cells stained with both antibodies.
`Empirically, the relative ‘brightness’ of any of these fluorophores can be estimated based on the separation of CD8 positive and negative
`cells. The apparent difference in the number of cells between the positive and negative peaks for these dyes is a visual artefact of the scaling
`(note that the ‘CD8-dim’ cells appear to be more prevalent in those plots where the CD8 peak is higher). There was no difference in the
`percent dim or bright-positive cells for any of distributions.
`
`Page 4 of 21
`
`Fluidigm
`Exhibit 1011
`
`

`
`N. Baumgarth, M. Roederer / Journal of Immunological Methods 243(2000)77–97
`
`81
`
`2.1.3. Spectral overlaps
`The spectral overlaps that exist between dyes
`currently represent the biggest hurdle in PFC analy-
`sis. Due to the similarities and overlaps in the
`emission spectra of different dyes (see Fig. 1), it is
`not possible to choose emission filters that uniquely
`measure only one of
`the dyes in a multicolor
`experiment. The appropriate choice of filter, how-
`ever, can greatly reduce collection of light from
`other fluorochromes (i.e., reduce ‘spillover’). Note
`from Fig. 1 that a set of up to four dyes excited by
`three different
`lasers can be chosen that exhibit
`essentially no spectral overlap; thus, four- or fewer-
`color experiments can be carefully designed to avoid
`the problems caused by spillover. However,
`the
`necessity of having three lasers for a four-color
`experiment renders this an impractical solution.
`Because of
`the spectral overlap, each fluoro-
`chrome will contribute a signal to several detectors,
`therefore the contribution in detectors not assigned to
`that fluorochrome must be subtracted from the total
`signal in those detectors. This process, termed ‘com-
`pensation’, is discussed in detail below.
`
`2.1.4. Conjugation to antibodies
`For many applications of flow cytometry, a further
`requirement for a fluorochrome is that
`it can be
`readily conjugated to antibodies. The 11 dyes we
`commonly use (Figs. 1 and 2) meet this criterion.
`There are, however, varying degrees of difficulty
`involved in carrying out the conjugation reactions. It
`is likely that any laboratory carrying out flow
`cytometric analyses with five or more dyes will need
`to conjugate antibodies, as commercial conjugates
`are not yet available for a number of these dyes.
`Even if all of the dyes become available as antibody
`conjugates, it is unlikely that every combination of
`antibody and fluorochrome necessary for each user
`will be available.
`Detailed procedures for conjugating the dyes can
`be found on the web (Roederer, 1997a) or from
`commercial vendors who supply conjugation kits.
`Briefly, the procedures for conjugating these dyes
`fall into three classes. (1) Small organic molecules
`(FITC, TR, A595, A430, CasB, and CasY) are
`conjugated to antibodies in relatively simple, short
`reactions that
`require only one or
`two column
`separations (or other desalting procedures). (2) Sin-
`
`gle protein molecules (PE, APC) require a slightly
`more complex procedure, but reactive fluorochromes
`can be easily prepared, are stable for long periods of
`time, can be used in a simple conjugation procedure.
`2
`(3) Tandem dyes
`(Cy5PE, Cy5.5PE, Cy7PE,
`Cy5.5APC, Cy7APC, Cy5.5PerCP), require the use
`of more complex procedures which involve first the
`generation of the reactive tandems that can then, in a
`second step, be used for conjugation to antibodies.
`Generating the reactive tandems requires careful
`testing of different conditions for each batch of
`tandem dye to optimize the ratio of the two dyes
`used. However, once the chemistry has been opti-
`mized, a large batch of reactive dye can be prepared
`for use in many conjugations.
`
`2.2. Other fluorochromes.
`
`New fluorochromes for immunophenotyping appli-
`cations are constantly being developed. At the time
`of writing, several dyes exist
`that are useful for
`immunophenotyping applications and which can be
`used in addition to or instead of some dyes listed in
`Fig. 1.
`
`2.2.1. PerCP
`This is a chlorophyll-like protein that can be
`directly conjugated to antibodies. It has an emission
`spectrum similar to Cy5PE, but with no excitation in
`the red, and hence very little spillover into the APC
`detector compared to Cy5PE. On the other hand, it
`has several disadvantages: it is easily ‘bleached’ by
`lasers, limiting its utility to low-power instruments
`such as benchtop instruments; it is less bright than
`Cy5PE; and it is only available in conjugates from a
`single vendor (Becton-Dickinson, San Jose, CA).
`
`2.2.2. Cy5.5PerCP
`A relatively new tandem conjugate of PerCP that
`has excitation and emission spectra very similar to
`Cy5.5PE. Unlike PerCP, it can be used in high power
`laser instruments as it does not ‘bleach’ easily, but is
`also only available in conjugated form through
`Becton-Dickinson.
`
`2
`
`Cy5PE is the same as ‘Tricolor’, Cy7APC is the same as
`‘Allo-7’ and ‘PharRed’, and Cy5.5PerCP is the same as ‘TruRed’.
`
`Page 5 of 21
`
`Fluidigm
`Exhibit 1011
`
`

`
`82
`
`N. Baumgarth, M. Roederer / Journal of Immunological Methods 243(2000)77–97
`
`2.2.3. Alexa 430,595
`A series of new dyes termed ‘Alexa dyes’ from
`Molecular Probes (Eugene, OR) includes two that we
`have tested for their usefulness in PFC. A430 is an
`excellent replacement for CasY, in that it has less
`non-specific interaction with CasB and is otherwise
`roughly similar in ‘brightness’ (Baumgarth, unpub-
`lished). A595 is spectrally similar to Texas Red (TR)
`but does not have its non-specific binding charac-
`teristics, leading to a somewhat better separation of
`positive and negative signals than that seen with
`many TR conjugates.
`
`2.2.4. TRPE
`This dye, also called ‘Red 613’ or ‘ECD’ is
`excited by a 488 nm laser line and can replace
`Cy5PE in a multicolor staining combination, as it has
`considerably less overlap into the APC channel than
`Cy5PE. It is however less bright than Cy5PE and it
`has somewhat more overlap into the PE channel; it
`also displays considerably higher background bind-
`ing typical of TR conjugates.
`There are a number of other dyes that are com-
`monly used to directly stain cellular components
`such as DNA (e.g., Hoechst, propidium iodide),
`cellular membranes (e.g., CFSE, nile red), lysosomes
`(e.g., CFDA), and dyes that can measure metabolic
`activity of the cell (e.g., rhodamine 123). It is beyond
`the scope of this manuscript to cover fluorochromes
`that are not used for immunophenotyping. For a
`detailed description of these therefore refer to (Sha-
`piro, 1994). Importantly, however, the use of any
`combination of dyes, irrespective of the measure-
`ment for which it is used, will need to address the
`same issues such as reagent ‘brightness’, spillover,
`and compensation as described here for antibody
`conjugates.
`
`2.3. Fluorochrome combinations
`
`Out of the dozen or more fluorochromes that can
`be used for immunophenotyping (by virtue of their
`spectral and chemical properties discussed above), it
`is important
`to select
`the best combination for a
`particular experiment under consideration. This is
`true whether the experiment requires four or 11
`colors. To a large extent, the constraints on which
`fluorochromes can be used will depend on the
`
`availability of laser lines on the machine in use (i.e.,
`is a 595-nm dye laser line available, or a 407-nm
`violet-enhanced krypton laser line?). The useable set
`of dyes also depends on the number of detectors that
`are available on the instrument to collect the excited
`light (i.e., how many colors can be collected that are
`excited by the first, second, and third laser line?).
`Once those constraints have been determined, there
`can still be a wide range of possible combinations.
`Table 1 lists a number of such combinations, given
`some constraints of laser and detector availability
`often seen in currently available machines.
`
`3. Hardware
`
`3.1. Lasers
`
`One of the single largest component costs of
`cytometry instrumentation is the cost of the lasers.
`Newer diode lasers are becoming prevalent and these
`can be significantly cheaper than the older gas ion
`lasers. When considering a solid-state laser,
`it
`is
`important to choose one that has a long life span and
`provides a consistent power output, which therefore
`excludes the very cheap solid-state lasers that are
`currently available. In addition, for current stream-in-
`air instrumentation it is desirable to have at least 50
`mW of power for each laser line in use, since the
`fluorescence signal (and thus sensitivity) increases
`with laser power. For most purposes,
`it
`is not
`necessary to have much more power
`than this,
`because most fluorochromes will saturate (are maxi-
`mally excited) at 100–150 mW. Indeed, additional
`power may actually reduce relative signal to back-
`ground staining, because the fluorophore will satu-
`rate at much lower powers than background fluores-
`cence. Table 2 lists some of the currently available
`lasers and the lines that are often used in flow
`cytometry, their approximate costs and the dyes that
`they can excite.
`
`3.2. Optical setup
`
`requires
`Sensitive detection of fluorochromes
`selection of appropriate filters that are placed before
`each detector, or photomultiplier tube (PMT). Filters
`must be selected so as to collect as much emitted
`
`Page 6 of 21
`
`Fluidigm
`Exhibit 1011
`
`

`
`N. Baumgarth, M. Roederer / Journal of Immunological Methods 243(2000)77–97
`
`83
`
`Table 1
`Sample fluorochrome combinations
`
`Dyes
`
`a
`
`4
`
`5
`
`6
`
`7
`
`Lasers
`
`b
`
`2 (4881647)
`
`1 (488)
`
`2 (4881595)
`
`2 (4881632)
`
`1 (488)
`
`2 (4881633)
`
`2 (4881595)
`
`1 (488)
`
`2 (4881632)
`
`2 (4881595)
`
`Combinations
`
`c
`
`FITC, PE, Cy5.5PE or Cy7PE, APC
`FITC, PE, Cy5PE, Cy5.5APC or Cy7APC
`FITC, PE, Cy5PE, APC
`FITC, PE, TRPE, APC
`FITC, PE, Cy5PE, Cy5.5PE or Cy7PE
`
`FITC, PE, Cy5.5PE or Cy7PE, TR, APC
`FITC, PE, Cy5PE, TR, Cy5.5APC or Cy7APC
`FITC, PE, Cy5PE, TR, APC
`FITC, PE, Cy7PE, APC, Cy5.5APC
`FITC, PE, Cy5.5PE, APC, Cy7APC
`FITC, PE, Cy5PE, Cy5.5APC, Cy7APC
`FITC, PE, Cy5PE, APC, Cy5.5APC or Cy7APC
`FITC, PE, Cy5PE or Cy7PE, APC, Cy7APC
`FITC, PE, Cy5PE, Cy5.5PE, Cy7PE
`
`FITC, PE, Cy5.5PE, Cy7PE, APC, Cy5.5APC or Cy7APC
`FITC, PE, Cy5PE, Cy5.5PE or Cy7PE, Cy5.5APC, Cy7APC
`FITC, PE, Cy5PE, Cy5.5PE or Cy7PE, APC, Cy5.5APC or Cy7APC
`FITC, PE, Cy5.5PE, Cy7PE, TR, APC
`FITC, PE, Cy5PE, Cy7PE, TR, Cy5.5APC
`FITC, PE, Cy5PE, Cy5.5PE, TR, Cy7APC
`FITC, PE, Cy5PE, Cy5.5PE or Cy7PE, TR, APC
`FITC, PE, Cy5PE, TRPE, Cy5.5PE, Cy7PE
`
`FITC, PE, Cy5.5PE, Cy7PE, APC, Cy5.5APC, Cy7APC
`FITC, PE, Cy5PE, Cy5.5PE or Cy7PE, APC, Cy5.5APC, Cy7APC
`FITC, PE, Cy5.5PE, Cy7PE, TR, APC, Cy5.5APC or Cy7APC
`FITC, PE, Cy5.5PE or Cy7PE, TR, APC, Cy5.5APC, Cy7APC
`
`a Any of the listed combinations can be augmented with a 407 nm Krypton laser and the fluorochromes CasB and CasY (or A430) to
`increase the number of colors by two without significantly affecting compensation requirements.
`b Number of lasers, followed by excitation lines (see footnote a).
`c Combinations are listed in rough order of increasing quality of measurements. i.e., for any given set of combinations, the first ones will
`tend to have lower compensation requirements and higher brightness. There are many more possible combinations than can be listed in this
`table. The order does not take into account commercial availability or ease of conjugate preparation. Substitutions: PerCP can be substituted
`for Cy5PE on lower laser-power instruments; Cy5.5PerCp can be substituted for Cy5.5PE on any system; A595 can be substituted for TR on
`any system. TRPE could be added to most combinations not employing 595 lasers or used as a substitute for Cy5PE. Where two dyes are
`listed as ‘A or B’, either can be used in the combination listed, but ‘A’ is preferred for reasons of brightness or compensation.
`
`light from the primary fluorochrome for high sen-
`sitivity, but as little as possible from other fluoro-
`chromes to reduce the compensation required. In
`general, these two criteria work against each other;
`therefore, for any given set of fluorochromes used in
`an experiment, the optimal filter set may be different.
`However, we have found that the filter set shown in
`Fig. 1 and Table 3 is close to optimal for many
`experimental setups that are currently in use.
`A further
`important selection criterion is the
`ability to block scattered light from the laser lines.
`Modern filters are considerably better than the old
`
`the FITC
`filters, allowing for example to collect
`signals with a much wider bandpass filter than was
`previously possible (old, 530/30; new, 525/ 50)
`while still blocking scattered light from the 488-nm
`laser line.
`For experiments using a small subset of these
`fluorochromes, we found that using the widest
`possible bandpass that still excludes as many other
`fluorescences as possible yields the best
`results.
`Nonetheless, switching to wide-open filters will
`generally only increase the measured signal by 20–
`100% compared to the bandpass filters shown in
`
`Page 7 of 21
`
`Fluidigm
`Exhibit 1011
`
`

`
`84
`
`N. Baumgarth, M. Roederer / Journal of Immunological Methods 243(2000)77–97
`
`Table 2
`Lasers for flow cytometry
`
`a
`
`Laser type
`
`Argon Ion
`
`Cooling
`
`Air
`
`Krypton
`
`Water
`
`Violet-enhanced
`b
`Krypton ion
`
`c
`
`Dye
`
`Doubled Nd-YAG
`
`HeNe
`
`Diode
`
`Water
`
`Water
`
`Air
`
`Air
`
`Air
`
`Line (nm)
`
`Approx. cost
`
`Fluorochromes
`
`488
`
`568
`
`647
`
`4071413
`
`595
`
`532
`
`632
`
`635
`
`$10K
`
`$30K
`
`$50K
`
`$15K
`
`d
`
`$12K
`
`$6K (50 mW)
`
`$1K (10 mW)
`
`FITC, PE, TRPE, PerCP,
`Cy5PE, Cy5.5PE, Cy5.5PerCP,
`Cy7PE, A488
`
`PE, TRPE, PerCP, Cy5PE,
`Cy5.5PE, Cy5.5PerCP, Cy7PE,
`A568, TR, A595
`APC, Cy5APC, Cy5.5APC,
`Cy7APC (Cy5PE, Cy5.5PE,
`e
`Cy7PE)
`
`CasB, CasY, A430
`
`TR, A595, APC, Cy5,
`Cy5.5APC, Cy7APC (TRPE,
`e
`Cy5PE, Cy5.5PE)
`
`PE, TRPE, PerCP, Cy5PE,
`Cy5.5PE, Cy5.5PerCP, Cy7PE
`
`APC, Cy5, Cy5.5APC,
`
`Cy7APC (Cy5PE, Cy5.5PE)
`
`e
`
`a Except as noted, lasers are rated at 50 mW output or greater at the given line.
`b 405-nm diode lasers with sufficient power for cytometry will be available soon; currently, 4-mW lasers cost about $4K.
`c Currently, 5-mW HeNe lasers with a 594-nm line can be obtained for about $3K; this power level is probably too low except for
`bench-top (flow cell) cytometers.
`d A dye laser also requires a 5-W, 488-nm pump laser, adding another $20K to the cost of this laser. However, the 488-nm line of the
`pump laser (running in ‘all lines’ mode) can be split off and used as the primary laser beam in the instrument, obviating the need to purchase
`the primary Argon laser.
`e Dyes in parentheses will also be excited by this laser, although their primary excitation is by other laser lines.
`
`Table 2. For flow cytometric experiments, 2-fold
`increases in signal-to-background levels are marginal
`compared to the optimizations that can be achieved
`in other parts of the experiment.
`
`4. Compensation
`
`4.1. What is compensation?
`
`Compensation is the process by which the spectral
`overlap between different fluorochromes is mathe-
`matically eliminated. The algorithm of compensation
`is a straightforward application of linear algebra, and
`should not be thought of as a subtraction process.
`Compensation between detectors can be performed
`either by hardware, after signal detection but before
`
`logarithmic conversion and /or digitization, or post-
`collection by software.
`While compensation is one of the most important
`processes required for proper data analysis in flow
`cytometry, it is also perhaps the least well under-
`stood. In order to properly design, implement, and
`analyze multicolor experiments, users must be aware
`of the effects of compensation, understand how to
`apply it correctly, and recognize when data are not
`properly compensated. While we will discuss some
`of these topics here in brief, the reader is strongly
`encouraged to read the texts (Roederer, 1999) and
`web pages (Roederer, 1997b) devoted to this topic
`for more in-depth information.
`
`4.2. Compensation complications
`
`Currently, most instruments provide the capability
`
`Page 8 of 21
`
`Fluidigm
`Exhibit 1011
`
`

`
`N. Baumgarth, M. Roederer / Journal of Immunological Methods 243(2000)77–97
`
`85
`
`Table 3
`Optimal interference filters for fluorochromes
`
`a
`
`Fluorochrome
`
`CasB
`CasY, A430
`FITC
`PE
`Cy5PE
`PerCP
`Cy5.5PE, Cy5.5PerCP
`Cy7PE
`TRPE, TR, A595
`APC
`Cy5.5APC
`Cy7APC
`
`Filter
`
`b
`
`440/40
`545/90
`525/50
`575/25
`665/30
`680/30
`720/45
`785/45
`625/40
`660/40
`705/50
`d
`750LP
`
`c
`
`d
`
`c
`
`a Optimization is a function of maximizing the amount of signal
`collected while minimizing the amount of spillover from other
`fluorochromes. Different filters may be optimal for particular
`applications; see text.
`b Filters are given as ‘x/y’, where x is the wavelength in
`nanometers of the center of the bandpass and y is the width in
`nanometers of the band. LP, long pass filter. Filters have an extra
`coating to block scattered light.
`c
`Use of a 720/45 filter for Cy5.5PE is optimal when Cy5PE is
`also to be measured. For measuring Cy5.5PE (or Cy5.5PerCP) in
`the absence of Cy5PE, then a 705/ 50 is better. Because APC has
`a lower emission wavelength than Cy5PE, the 705/50 works well
`for Cy5.5APC in combination with APC. However, if Cy5.5APC
`is to be measured with Cy5, then the 720/45 filter should be
`selected.
`d
`The 750LP and the 785/50 are roughly equivalent
`collecting Cy7 emission; either can be used.
`
`for
`
`for compensating between several spectrally adjacent
`pairs of detectors: for example, between FITC and
`PE, and between PE and Cy5PE. For a two-color
`analysis, such pairwise compensation is sufficient
`and complete. However, the introduction of a third
`fluorochrome introduces the potential for complex
`interactions that cannot by fully corrected by such
`pair-wise compensations. For example, FITC has low
`but measurable emission into the Cy5PE detector.
`This overlap may not be completely corrected by the
`the FITCfi PE and PEfi Cy5PE
`combination of
`compensation settings, causing what appears to be
`artefactual Cy5PE fluorescence when using a very
`bright FITC stain. Obviously, as more than three
`colors are used, these problems increase.
`Indeed, for a typical 11-color stain, one would
`potentially need to use up to 100 pair-wise com-
`pensation controls to fully compensate all colors (in
`
`reality only about 40 of these are non-zero). Further-
`more, because of the multiple interactions between
`these controls, some settings may actually be less
`than zero; something that current hardware controls
`do not allow. Finally, since setting interacting pair-
`wise compensation controls is an iterative process, it
`would take virtually forever to manually set them
`properly.
`An additional complication to compensation is
`brought about by the use of tandem dyes (e.g.,
`Cy5PE, Cy7APC, etc.). Tandem dyes are covalently
`linked combinations of ‘donor’ (e.g., PE) and ‘accep-
`tor’ (e.g., Cy5) dyes. The ratio of acceptor to donor
`is usually in the range of 3:1 to 10:1, depending on
`the optimization strategy used by the manufacturer.
`In fact, different tandem-conjugated antibodies from
`the same manufacturer often have slightly different
`ratios of acceptor to donor. The problem is that the
`compensation required to correct tandem emissions
`in other fluorescence channels strongly depends on
`this ratio. Therefore, a different compensation setting
`may be needed for every different tandem conjugate
`used in a particular experiment. In essence,
`this
`means that
`the compensation setting needs to be
`adjusted for every different staining combination in
`use.
`that
`is apparent
`it
`Given these complications,
`complete compensation on multicolor systems re-
`quires computer intervention. Currently no instru-
`ment is commercially available that provides users
`with the ability to automatically adjust compensation
`settings on a per-stain basis, nor even to provide
`complete compensation among all pairs of fluores-
`cence detectors. Therefore, the only solution as yet is
`the use of software analysis programs that are
`applied after data collection is completed for com-
`pensation of the data. Examples of such software
`include WinList (for PC) and FlowJo (for Macin-
`tosh).
`The use of software to set compensation simplifies
`the process considerably. Computers can set
`the
`compensation correctly (there is no subjective criter-
`ion applied) and can be instructed to appropriately
`vary the compensation matrix according to the
`specific fluorochromes used in any given sample.
`However, there are significant limitations to software
`compensation (see below). In the interim before
`complete automated hardware compensation is avail-
`
`Page 9 of 21
`
`Fluidigm
`Exhibit 1011
`
`

`
`86
`
`N. Baumgarth, M. Roederer / Journal of Immunological Methods 243(2000)77–97
`
`able, the optimal solution is to combine both hard-
`ware and software compensation to minimize the
`limitations of each.
`
`4.3. Limitations of compensation
`
`Both available hardware and software compensa-
`tion have significant limitations that place restrictions
`on the utility of PFC. Hopefully, some of these
`limitations will soon be obviated by new technology
`developments, which would enable the easier set-up
`of new reagent combinations for multicolor flow
`cytometric measurements. Until then, it is important
`to understand these limitations and how they may be
`best addressed.
`One limitation that is shared by both hardware and
`software compensation is due to the low precision of
`measurements in the far-red channels, Cy7APC,
`Cy5.5APC, and to a certain extent APC. In these
`channels, for dimly stained cells, the actual number
`of photoelectrons detected by the PMT can be very
`low, due to the inefficiency of the currently available
`PMT to detect light at this wavelength. Even the
`‘deep red-sensitive’ PMTs now available do not
`overcome this problem. The unavoidable counting
`errors inherent in the measurement can lead to the
`introduction of significant artefacts. For example, if
`only five photoelectrons are collected, the counting
`]˛
`error is
`5, or 45%. This error propagates to other
`channels that require compensation from this detec-
`tor. The result is that the ‘broadening’ of populations
`that occurs due to compensation (Roederer, 1997b,
`1999) is significantly worse for the far-red emitting
`fluorochromes.
`Hardware compensation operates on the analog
`signals collected from the PMT, before further signal
`processing is done. For detectors with sufficient
`si

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket