Br. 7. Anaesth. (1974), 46, 260
`
`INFECTION HAZARD FROM SYRINGES
`
`c. E. BLOGG, M. A. E. RAMS-AY AND J. D. JARVIS
`
`SUMMARY
`Bacterial contamination of the contents of disposable plastic syringes was demonstrated
`experimentally after repeated refilling. Supporting evidence for this mechanism as a
`potential source of bacteraemia was obtained by a spot check of syringes used for
`repeated injections. It is recommended that disposable syringes should be discarded
`after a single injection.
`
`Downloaded from
`
`http://bja.oxfordjournals.org/
`
` at Baylor Medical Ctr on September 30, 2016
`
`the syringe held with sterile rubber gloves. The
`syringes were filled and refilled using the common
`technique of withdrawing the plunger with one
`hand, whilst the other held the bottle from which
`the culture medium (I% peptone water) was aspi(cid:173)
`rated (fig. 1). The culture medium was withdrawn
`from capped bottles through a sterile rubber seal to
`the maximum calibration mark on the syringe and
`then re-injected into the bottle. The specimen was
`labelled "control".
`The gloved right hand was then dipped into a
`culture of Serratia marcescens, used as a marker
`organism. The syringe plunger was pulled back
`with the contaminated right hand, whilst the un(cid:173)
`contaminated
`left hand held only
`the bottle.
`Approximately 25 X 106 organisms per sq.cm were
`present on the glove. This represents a severe chal(cid:173)
`lenge. The procedure was repeated with the same
`syringe until five refills had been made. Separate
`culture bottles were used for each refill.
`
`(B) Investigation of syringes in prolonged use.
`Plastic syringes, which had been used throughout
`operative procedures for the repeated administration
`of drugs, were collected from anaesthetists in the
`operating theatres immediately after use and the
`needle and needle cover removed from the nozzle,
`which was then sealed with a sterile plastic cap
`(Braun). Care was taken not to contaminate the
`nozzle. The contents were then cultured. Similarly
`capped plastic syringes were collected from the ITU,
`where they had been used for the repeated flushing
`of intravascular cannulae.
`In all cases the culture medium was incubated at
`36°C for 48 hours. The bottles were then examined
`for bacterial contamination.
`
`C. E. BLOGG, F.F.A.R.c.s.; M. A. E. RAMSAY, F.F.A.R.c.s.;
`J. D . JARVIS, F.I.M.L.T.; Division of Anaesthesia and
`Department of Clinical Microbiology, The London
`Hospital, Whitechapel, London E.l.
`
`It is now standard practice to use disposable
`syringes for injection. Jn situations where repeated
`injections are required in the same patient-for
`example, by doctors and nurses in intensive therapy
`units (ITU) and by anaesthetists in operating
`theatres-it is common practice for these syringes
`to be refilled. Simple observation demonstrates that
`mishandling of the plunger may readily occur (fig.
`I), thus creating a potential source of contamination.
`This hypothesis has been investigated. In addi(cid:173)
`tion, syringes which had been used over a prolonged
`period in operating theatres and in an ITU were
`examined for bacterial contamination.
`
`FIG. 1. Common method of drawing up with a syr inge
`showing mishandling of che plunger shaft, that leads to
`bacterial contamination of the concems.
`
`METHOD
`(A) Investigation of syringes with a marker organism.
`A variety of sizes of sterile plastic and glass
`syringes were used.• Each syringe pack was opened
`in a sterile, filtered-air, positive-pressure room, and
`
`• P lastic syringes manufactured by Gillette, Beckton &
`Dickinson and D.H.S.S. Glass syringes manufactured by
`Smith & Nephew.
`
`Hospira, Exh. 2010, p. 1
`
`

`
`Downloaded from
`
`http://bja.oxfordjournals.org/
`
` at Baylor Medical Ctr on September 30, 2016
`
`INFECTION HAZARD FROM SYRINGES
`
`261
`
`RESULTS
`(A) From table I it can be seen that contamination
`never occurred at the first refill but by the second
`refill all of the glass syringes and at least 50% of
`the plastic syringes had their contents contaminated.
`Because the likelihood of transfer of bacteria from
`the plunger to the syringe contents was similar with
`all the makes of plastic syringe examined the results
`have been pooled.
`
`TABLE I. Concaminacion of syringe conrencs. Syringe
`plungers mishandled by a hand concaminaced wich a
`marker organism (Serracia marcescens).
`
`No. of syringe contents contaminated/No.
`of syringes tested
`Refills
`
`Syringes
`
`Control
`
`1st
`
`2nd
`
`5th
`
`Plastic
`1 ml
`2 ml
`5 ml
`10 ml
`20 ml
`Glass s ml
`10 ml
`20 ml
`
`0/10
`0/10
`0/15
`0/15
`0/15
`
`0/5
`0/5
`0/5
`
`0/10
`0/10
`0/15
`0/15
`0/15
`
`0/5
`O/S
`0/5
`
`6/10
`5/10
`7/15
`11/15
`6/15
`
`S/5
`5/S
`5/5
`
`10/10
`10/10
`15/15
`15/15
`15/15
`
`5/S
`5/S
`5/S
`
`(B) One hundred plastic syringes were collected
`from the ITU and the operating theatres (table II).
`The contents of eight syringes were contaminated.
`Of these, five had been used in the ITU. The con(cid:173)
`tents grew Staphylococcus epidermidis 1 (3 syringes),
`Staphylococcus aureus2 (1 syringe) and Strepto(cid:173)
`coccus viridans3 (1 syringe).
`The remaining three syringes had been used for
`patients undergoing major surgery. The contents of
`two syringes used in two separate cases of resection
`of an aortic aneurysm grew Escherichia coli4 • The
`third syringe had been used for a patient undergoing
`mitral valve surgery and from its contents Staphylo(cid:173)
`coccus aureus was grown.
`No patients had clinical evidence of septicaemia.
`
`1. Staphylococcus epidermidis: gram positive, fermenta(cid:173)
`tive, catalase positive, slide coagulase negative, D.N.
`ase negative coccus.
`2. Staphylococcus aureus: gram positive fermentative,
`caralase positive, slide coagulase positive, D.N. ase
`positive coccus.
`3. Streptococcus viridans: gram positive, catalase nega(cid:173)
`tive, a-haemolytic (horse blood agar), coccus.
`4. Escherichia coli: gram negative, motile, citrate nega(cid:173)
`tive, K.C.N. negative, Eijkman positive rod.
`
`D
`
`TABLE II. Organisms culcured from che concenrs of 100
`syringes collecced afcer use in che operacing cheatres or
`che Intensive Care Unit.
`
`Organism
`
`From theatres
`Escherichia coli
`Staphylococcus aureus
`From ITU
`Staphylococcus epidermidis
`Staphylococcus aureus
`Streptococcus viridans
`
`No. of
`Syringes
`
`Total
`No. of
`Syringes
`
`2
`1
`
`3
`1
`1
`
`50
`
`so
`
`DISCUSSION
`that contamination of syringe
`It is undesirable
`contents should occur, although such contamination
`cannot necessarily be equated with clinical bacter(cid:173)
`aemia. Our results show that if the plunger of a
`syringe is contaminated by soiled hands, organisms
`can be transferred to the syringe contents.
`Anaesthetists and intensive care nurses frequently
`soil their fingers with tracheobronchial secretions
`and mucopus, and Lowbury and Lilly (1973)
`demonstrated that even a 2-min wash with ordinary
`bar soap does little to reduce the bacterial count on
`hands. Furthermore, it has been suggested (Teres
`et al., 1973) that in the intensive therapy unit, sinks
`provide reservoirs of Pseudomonas infection and
`that the bands of personnel are the vehicles of
`transmission.
`There have been many investigations into the
`sources of bacterial contamination of intravenous
`fluids and cannulae (Banks et al., 1970; Freeman
`and King, 1971; Colvin et al., 1972; Philips, Eykyn
`and Laker, 1972; Department of Health and Social
`Security, 1972; Committee of the Medicines Com(cid:173)
`mission, 1973; Lapage, Johnson and Holmes, 1973;
`Leading Articles, 1972, 1973a,b), but little has
`been reported on the possible introduction of bac(cid:173)
`teria into infusion systems, or directly into patients
`by the mishandling of syringes (Scurr and Edgar,
`1962).
`The reported incidence of bacteria in intravenous
`fluids to which drugs bad been added (D'Arcy and
`Woodside, 1973) may be accounted for by the use
`of contaminated syringes for the addition of the
`drugs. This is especially hazardous as intravenous
`nutrients are excellent culture media. In particular,
`the risk to patients undergoing cardiac and trans(cid:173)
`plant surgery may be considerable.
`The fact that patients are at risk is shown by our
`findings of organisms in the contents of 8 of 100
`
`Hospira, Exh. 2010, p. 2
`
`

`
`Downloaded from
`
`http://bja.oxfordjournals.org/
`
` at Baylor Medical Ctr on September 30, 2016
`
`262
`
`syringes collected at random. Pathogenic organisms
`were present in 4 of these 8 syringes. We have
`demonstrated that the use of disposable syringes for
`more than one injection is potentially hazardous.
`This study demonstrates that as well as taking the
`obvious precautions of avoiding manual contact
`with the needle or nozzle of the syringe, careless
`mishandling of the syringe plunger with repeated
`injections may allow bacteria to pass from the hands
`to the syringe contents via the plunger. Total avoid(cid:173)
`ance of handling the plunger shaft is difficult, especi(cid:173)
`ally when both hands cannot be employed, and
`therefore the only sensible precaution to be taken is
`to confine the use of the syringe to one injection.
`The seal formed between the washer of the syringe
`plunger and the barrel of the syringe cannot be
`considered as a barrier to bacteria.
`We recommend that syringes should not be used
`for more than one injection, and that a fresh sterile
`syringe should always be used whenever it is neces(cid:173)
`sary to draw up and inject a further supply of a drug
`into the same .Patient.
`
`ACKNOWLEDGEMENTS
`We thank Professor B. R. J. Simpson for his assistance
`and guidance in the preparation of this paper, Miss J.
`Tolhurst and Miss S. R. Liddell for secretarial work, and
`Mr R. Ruddick and Miss J. Abbott of the Departments
`of Photography and Medical Illustration for the figure.
`
`BRITISH JOURNAL OF ANAESTHESIA
`
`REFERENCES
`Banks, D. C., Yates, D. B., Cawdrey, H. M., Harries,
`M. G., and Kidner, P. H. (1970). Infection from intra(cid:173)
`venous catheters. Lancet, 1, 443.
`Colvin, M. P., Blogg, C. E., Savege, T. M., Jarvis, J. D.,
`and Strunin, L. (1972). A safe long-term infusion tech(cid:173)
`nique? Lancet, 2, 317.
`Committee of the Medicines Commission (1973). Report
`on the prevention of microbial contamination of medi(cid:173)
`cinal products. H.M. Stationery Office.
`D'Arcy, P. F., and Woodside, W. (1973). Drug additives:
`a potential source of bacterial contamination of infusion
`fluids. Lancet, 2, 96.
`Department of Health and Social Security (1972). Interim
`report on heat sterilised fluids for parenteral administra(cid:173)
`tion. H.M. Stationery Office.
`Freeman, R., and King, B. (1972). Infective complications
`of indwelling intravenous catheters and the monitoring
`of infections by the nitroblue-tetrazolium test. Lancet,
`1, 994.
`Lapage, S. P., Johnson, R., and Holmes, B. (1973). Bac(cid:173)
`teria from intravenous fluids. Lancet, 2, 284.
`Leading Article (1972). Contaminated infusion fluids. Br.
`Med. 1., 3, 190.
`- - (1973a). Particles in veins. Br. Med. 7., 1, 307.
`- - (1973b). Cleaner medicants. Br. Med. 7., 2, 6.
`Lowbury, E. J. L., and Lilley, H. A. (1973). Use of 4%
`chlorhexidine detergent solution (Hibiscrub) and other
`methods of skin disinfection. Br. Med. 1., 1, 510.
`Phillips, I., Eykyn, S., and Laker, M. (1972). Outbreak
`of hospital infection caused by contaminated autoclaved
`fluids. Lancet, 1, 1258.
`Scurr, C. F., and Edgar, W. M. (1962). A possible danger
`of all-glass syringes. Lancet, 1, 1303.
`Teres, D., Schweers, P., Bushnell, L. S., Hedley-Whyte,
`J., and Feingold, D. S. (1973). Sources of Pseudomonas
`aeruginosa infection in a respiratory /surgical I.T.U.
`Lancet, 1, 415.
`
`BOOK REVIEW
`
`Intractable Pain. By Mark Mehta. Published by W. B.
`Saunders Co. Ltd, London. Pp. 304, illustrated,
`indexed. Price £6.00.
`In opting to write a book about intractable pain Dr Mehta
`has chosen one of the most difficult subjects in medicine.
`He has also chosen one of the most misunderstood,
`despite the fact that most doctors encounter pain in one
`form or another almost every day of their lives. Concern
`has been expressed in the medical press with increasing
`frequency that this paradox, namely familiarity with pain
`in the clinical sense on the one hand and ignorance of its
`many physiological and psychological aspects on the other,
`should exist. Most medical undergraduates receive lectures
`concerning probable pathways for noxious impulses fol(cid:173)
`lowing stimuli in various parts of the body, and turgid
`accounts of the properties of hosts of analgesic agents
`with a limited review of indications for their administra(cid:173)
`tion. Few teachers are sufficiently inspired to explain the
`problems of those suffering from pain in the light of
`present-day knowledge of psychological, social, and
`physiological factors. Lack of knowledge of this kind is
`not confined to doctors alone, for others involved in the
`care of patients are often unaware both of their role in
`treating the severely ill, and of the effects of their atti(cid:173)
`tudes towards pain upon the manner in which they treat
`it. With these points in mind it is refreshing to find that
`Dr Mehta has taken steps to correcting some of the defi-
`
`ciencies. At the outset it should be said that his book is
`primarily concerned with methods of treatment for in(cid:173)
`tractable pain, but he does not neglect to refer to the
`nature of pain and to comment upon the importance of
`the knowledge of its psychological aspects. He gives a
`brief review of the current theories which attempt to
`explain how pain becomes a conscious experience, and
`proceeds to give an analysis of conditions which give rise
`to intractable pain. The third section of his book deals
`with methods of treatment. This includes familiar ap(cid:173)
`proaches to treatment with analgesic drugs and various
`forms of injection, both locally and at sites distant from
`the pain. He also considers less commonly used methods
`of
`treating pain,
`including neurosurgical procedures,
`hypnosis, psychotherapy and even acupuncture.
`For the student of pain, intractable or otherwise, Dr
`Mehta presents us with a fascinating description of this
`area of medicine. His writing is clear and there is an even
`quality about the work, which suggests that in most areas.
`particularly those concerning treatment, he has a good
`deal of personal experience. This is not the book, how(cid:173)
`ever, for those who seek detailed accounts of the theoreti(cid:173)
`cal, physical or psychological bases for pain, but it is
`heartily recommended both for undergraduates and post(cid:173)
`graduates, although the price of £6.00 may initially seem
`a little prohibitive.
`
`M. R. Bond
`
`Hospira, Exh. 2010, p. 3

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket