throbber
INSTITUTE OF PHYSICS PUBLISHING
`
`Phys. Med. Biol. 48 (2003) 1075–1089
`
`PHYSICS IN MEDICINE AND BIOLOGY
`
`PII: S0031-9155(03)57398-5
`
`Inverse planning for intensity-modulated arc therapy
`using direct aperture optimization
`
`M A Earl, D M Shepard, S Naqvi, X A Li and C X Yu
`
`Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore,
`MD 21201, USA
`
`E-mail: mearl001@umaryland.edu
`
`Received 17 December 2002
`Published 1 April 2003
`Online at stacks.iop.org/PMB/48/1075
`
`Abstract
`Intensity-modulated arc therapy (IMAT) is a radiation therapy delivery
`technique that combines gantry rotation with dynamic multi-leaf collimation
`(MLC). With IMAT, the benefits of rotational IMRT can be realized using
`a conventional linear accelerator and a conventional MLC. Thus far, the
`advantages of IMAT have gone largely unrealized due to the lack of robust
`automated planning tools capable of producing efficient IMAT treatment plans.
`This work describes an inverse treatment planning algorithm, called ‘direct
`aperture optimization’ (DAO) that can be used to generate inverse treatment
`plans for IMAT. In contrast to traditional inverse planning techniques where the
`relative weights of a series of pencil beams are optimized, DAO optimizes the
`leaf positions and weights of the apertures in the plan. This technique allows
`any delivery constraints to be enforced during the optimization, eliminating the
`need for a leaf-sequencing step. It is this feature that enables DAO to easily
`create inverse plans for IMAT. To illustrate the feasibility of DAO applied to
`IMAT, several cases are presented, including a cylindrical phantom, a head and
`neck patient and a prostate patient.
`
`1. Introduction
`
`A fundamental goal of radiotherapy is to deliver a conformal dose of radiation to the tumour
`while simultaneously minimizing the dose to healthy tissue. Intensity-modulated radiation
`therapy (IMRT) is an improvement over conventional techniques in that it has the ability to
`better sculpt the dose distribution around critical structures. Consequently, the sparing of
`critical structures and normal tissue is improved, allowing physicians to escalate the dose
`to the tumour. IMRT delivery techniques can be divided into two classes: fixed gantry and
`rotational. Fixed-gantry IMRT is achieved by delivering overlapping fields from fixed beam
`directions. The field shapes either remain constant during the delivery of radiation or can
`
`0031-9155/03/081075+15$30.00 © 2003 IOP Publishing Ltd Printed in the UK
`
`1075
`
`Page 1 of 15
`
`Elekta Exhibit 1009
`
`

`
`1076
`
`M A Earl et al
`
`dynamically change during radiation delivery. Rotational IMRT is achieved by dynamically
`moving the leaves of the MLC and the gantry during radiation delivery.
`Intensity-modulated arc therapy (IMAT) was first proposed by Yu (1995) as an alternative
`rotational IMRT delivery technique to tomotherapy (Mackie et al 1993, 1995, Carol et al
`1997). With tomotherapy, a binary MLC is used to modulate the intensity of radiation as the
`source rotates about the patient. IMAT is performed with a conventional linear accelerator,
`where the leaves of the MLC change shape as the gantry rotates during delivery. The treatment
`is composed of ‘arcs’, each with a start and stop angle. The MLC field shape changes
`continuously during gantry rotation from the beginning to the end of each arc. To achieve
`a modulated intensity distribution from a given delivery angle, multiple overlapping arcs are
`employed. A key advantage of IMAT is that the delivery is achieved using a conventional
`linear accelerator and a conventional MLC. Therefore, IMAT treatments can be delivered
`using existing equipment in most radiation oncology departments. It should be noted that to
`deliver IMAT plans, the linear accelerator must be equipped with the capability for dynamic
`delivery.
`Traditionally, IMRT treatment plans are generated with a two-step process (Bortfeld et al
`1994, Chui et al 1994, Galvin et al 1993, Webb 1994, Xia and Verhey 1998). First, a set of
`pencil beam intensities is optimized for each specified beam direction. The resulting optimized
`intensity maps are then sent to a leaf-sequencer which determines the set of MLC shapes to
`be used for delivery. It is in this step where any delivery constraints of the treatment unit
`are enforced. The output of the two-step procedure is a set of MLC aperture shapes and
`their corresponding weights, or monitor units. For fixed-field IMRT, no further sequencing is
`needed. However, for IMAT additional sequencing is required to produce deliverable plans.
`To make the dose calculation feasible in IMAT, the arcs are approximated by a series
`of equispaced static beams (Li et al 2001). Applying the aforementioned two-step approach
`to IMAT requires a more complex leaf-sequencing algorithm that links the aperture shapes
`between consecutive angles. This significantly complicates the leaf-sequencing step and makes
`production of IMAT treatment plans using this conventional two-step approach extremely
`cumbersome. Generally, eight to ten arcs are required to achieve three intensity levels
`(Yu 1995).
`Multiple investigators have successfully implemented IMAT using a forward planning
`approach (Yu et al 2002, Wong et al 2002, Ramsey et al 2001). Recently, Yu et al (2002)
`reported on a clinical implementation of IMAT whereby field shapes were generated based
`upon beam’s eye view of the tumour and critical structures. The arc weights were then
`manually changed to achieve the desired dose distribution. Ma et al (2001) proposed a more
`sophisticated method whereby the weights of the arcs in prostate cancer treatments were
`optimized using a gradient search method. Cotrutz et al (2000) introduced a similar technique
`where several arcs were defined, each with the same blocking of the central portion of the field,
`but with varying overall aperture size. The weights of the arcs were then optimized to achieve
`the desired dose distribution. Although some inverse planning was employed in the latter two
`approaches for beam weight optimization, the field shapes were still defined manually. To our
`knowledge, there is no fully automated inverse planning system reported for IMAT.
`In this work, we propose a method that simultaneously optimizes both the field shapes
`and the arc weights. We are able to generate inversely planned IMAT treatment plans using a
`technique that we call ‘direct aperture optimization’ (DAO). In a previous work (Shepard et al
`2002), we discussed the application of the DAO technique to fixed-field step-and-shoot IMRT.
`The key feature of DAO is that the MLC leaf positions and beam weights are the parameters
`that are optimized, as opposed to pencil beam weights. Using this technique, all of the delivery
`constraints are enforced during the optimization, thereby eliminating the need for a separate
`
`Page 2 of 15
`
`

`
`Inverse planning for IMAT
`
`1077
`
`Input:
`(1)
`(2)
`(3)
`
`Number and range of arcs
`Prescription
`Optimization parameters
`
`Calculate pencil beams
`
`Start with BEV of target for
`each beam direction
`
`Select change
`
`Does change
`satisfy delivery
`constraints?
`
`Yes
`
`Calculate dose resulting
`from change and the
`corresponding objective
`function
`
`Reject change
`
`No
`
`No
`
`Reached finishing
`criteria?
`
`Yes
`
`Export dose and plan
`for evaluation and
`delivery
`
`Does objective
`function improve?
`
`Yes
`
`Keep change
`
`No
`
`Generate random number
`x between 0 and 1
`
`No
`
`Is x less than
`probability dictated
`by simulated
`annealing?
`
`Yes
`
`Figure 1. Flow chart of optimization procedure.
`
`leaf-sequencing step. The elimination of the leaf-sequencing step significantly simplifies the
`generation of inverse-planned IMAT treatment plans.
`To illustrate the concept, we present the results of an inverse-planned IMAT case for a
`pseudo-prostate target on a cylindrical phantom, a head and neck case and a prostate case.
`In addition, we present results of the delivery of an inverse-planned IMAT plan to a c-shape
`target in a cylindrical phantom.
`
`2. Materials and methods
`
`2.1. Overall scheme
`
`To generate a DAO inverse plan for an IMAT treatment, the user specifies: (1) the number and
`angular range of the arcs, (2) the prescription parameters and (3) the optimization parameters.
`The software will then: (1) calculate the dose for each beam for the patient-specific CT data,
`(2) perform the optimization and (3) output the results for analysis/delivery. Each step is
`described in this section. An overview of the entire procedure is shown in figure 1.
`
`Page 3 of 15
`
`

`
`1078
`
`2.2. Dose calculation
`
`M A Earl et al
`
`◦
`
`. For the cases presented
`The dose is calculated for a set of equispaced beams spanning 360
`◦
`in this work, 36 beams with an equal spacing of 10
`were used. This spacing convention is
`the compromise used in reference (Yu et al 2002) for the clinical implementation of IMAT.
`Each of these beams is divided into 1 cm × 0.5 cm pencil beams. A separate dose calculation
`is performed for each pencil beam, and the 3D dose matrix for each pencil is stored on disk.
`For example, if we are calculating an 8 × 8 field, we would have 8 × 8 × 2 × 36 = 4608
`separate 3D dose matrices. Since the dose calculation is somewhat time consuming, we store
`the dose matrices on disk so that the dose calculation only needs to be performed once. Every
`time an additional optimization is performed, the pencil beam dose distributions are available
`immediately.
`The dose is calculated using a home-grown algorithm in which photon interactions are
`randomly sampled, and energy kernels are randomly propagated.
`In essence, the method
`superimposes energy deposition kernels using the Monte Carlo method. We will give a brief
`description of the technique here, but a more detailed account of the technique will be described
`in a later publication.
`Histories are tracked on a photon-by-photon basis. A history starts by randomly sampling
`the starting position of the photon spatially within the field. The energy of the photon is then
`randomly sampled from the energy spectrum of the beam. An interaction point is randomly
`selected based upon radiological path length, and a monoenergetic dose spread array (DSA)
`corresponding to the energy sampled is propagated from that point along a randomly selected
`direction. The DSAs calculated by Mackie et al (1988) are used. Hundreds of thousands of
`histories are run for each pencil beam. Instead of propagating secondary electrons as in full-
`blown Monte Carlo, the technique propagates the dose kernel. The use of random sampling
`allows easier computer parallelization and explicit sampling of photon energy obviates the
`need for any depth hardening or off-axis corrections. In addition, no TERMA calculation
`is required. Because generation of IMAT treatment plans requires the calculation of a large
`number of pencil beam dose distributions, the calculation time for each pencil beam is a
`significant consideration. Using this technique enables us to generate the required dose
`matrices for IMAT in a reasonable time without a significant sacrifice in dose calculation
`accuracy.
`Because the dose due to each pencil beam has limited spatial extent, many of the voxels
`receive little or no dose. To reduce the amount of hard disk space required to store each beam,
`we keep the dose information only for voxels that receive more than some fraction of the
`maximum dose. Typically, we set this fraction to 0.25%. This compression technique reduces
`the amount of space required to store the pencil beams by approximately a factor of 500. After
`each pencil beam calculation, the data are stored on disk to be used in the optimization phase
`of the procedure. The pencil beam calculation is performed only once. Depending on the
`voxel size, field size and patient size, the time required to compute all of the pencil beams for
`IMAT cases ranges from 3 to 12 h on 12 750 MHz Pentium III CPUs. The disk space required
`to store the beams ranges from 250 to 500 MB.
`It should be noted that the proposed inverse planning algorithm is independent of the
`pencil beam dose calculation engine. A more accurate dose engine, such as Monte Carlo,
`may be used with the cost of longer computing time. For example, we have examined the
`EGS4/DOSXYZ system (Nelson et al 1985, Ma et al 1998) and have found that the computing
`time is increased by a factor of 20 to 30 as compared with our dose engine. An alternative
`Monte Carlo code is the EGS4/MCDOSE system (Ma et al 2002) which has a significant
`computing time reduction compared to DOSXYZ.
`
`Page 4 of 15
`
`

`
`Inverse planning for IMAT
`
`2.3. Optimization
`
`1079
`
`In the prescription, the user specifies the angular range of the arcs to be included in the
`optimization. The only constraint on the start and stop angles of the arc paths is the beam angles
`for which dose was calculated. To produce intensity modulation, two or more overlapping
`arcs for each arc path are needed.
`The optimization is performed using simulated annealing (Kirkpatrick et al 1993). The
`starting MLC shape conforms to the beam’s eye view of the tumour. The optimization is an
`iterative procedure containing many thousands of iterations. At each iteration, a parameter
`is randomly selected. Applying DAO to IMAT, the parameters are the leaf positions and arc
`weights, or beam weights if the gantry speed is allowed to vary. The parameter is then changed
`an amount sampled from a Gaussian distribution, whose width decreases with iteration number
`according to the schedule
`σ = 1 + (A − 1)
`
`(1)
`
`step
`0
`
`1
`(nsucc + 1)1/T
`step
`where A is the step size at the start of optimization, T
`is the parameter that dictates the
`0
`rate at which σ decreases (or cooling rate) and nsucc is the number of successes, which will be
`described later in the text.
`Intuitively, the final field shapes should be significantly different from the initial BEV of
`the tumour. By making large steps initially, the field will deviate from the BEV more quickly
`and, therefore, become closer to the desired solution. At the end of the optimization, smaller
`steps are required to fine-tune the aperture shapes. The magnitude of the initial step size A
`varies according to the starting size of the field. For starting field sizes around 20 × 20 cm2,
`we typically set A to 5 cm. For starting fields around 8 × 8 cm2, A is typically set to 3 cm.
`These are just sample values: other values could be chosen. Since A is simply the starting
`width of the Gaussian distribution, step sizes smaller and larger than A are possible. It should
`be noted that since pencil beams are used, only discrete changes in the MLC leaf positions are
`allowed. The smallest change in leaf position is equal to the resolution of the pencil beams.
`For instance, in the case where 1 cm × 0.5 cm pencil beams were calculated (1 cm corresponds
`to the width of our MLC leaves), steps of 0.5 cm, 1 cm, 1.5 cm, etc. are allowed.
`Since beam weights are continuous variables, it is not necessary to force the change to
`be discrete as we do with changes in leaf position. Instead, the size of the change is sampled
`directly from a Gaussian distribution. At the start of the optimization, the weights of all beams
`are set to 1.0. The starting width of the Gaussian distribution for the weight change is typically
`set 0.20, although other values can be chosen. As with the leaf position variables, the width
`of the Gaussian decreases with increasing iteration in the optimization.
`After each change in leaf position, the new aperture shape is checked to determine if it
`satisfies the delivery constraints. The delivery constraints are divided into two categories:
`MLC constraints and IMAT constraints. The MLC-based constraints are the same as those
`encountered in static beam delivery: they are imposed by the physical limitations of the MLC
`leaves in forming an aperture. IMAT-based constraints are dynamic: they are imposed by the
`limitations of the speed of leaf motion and the speed of gantry rotation. MLCs manufactured
`by different vendors have different constraints. For instance, with Elekta’s MLC, both the
`opposed leaves and the opposed-adjacent leaves cannot come within 0.8 cm of one another.
`As a result, interdigitation of the opposing leaves is not allowed.
`For IMAT, leaves between consecutive angles within an arc are constrained to move a
`distance of
`
`d = vl θ
`

`
`(2)
`
`Page 5 of 15
`
`

`
`1080
`
`M A Earl et al
`
`where vl is the leaf travel, ω is the gantry angular rotation speed and θ is the angular
`◦
`separation between consecutive angles, typically 10
`. In some cases, the gantry speed can
`be slowed and the dose rate decreased to accommodate large leaf steps between consecutive
`angles. Since radiation is delivered continuously between consecutive angles and the leaf
`positions are interpolated between consecutive angles, the calculated and delivered dose
`distributions can be somewhat different. In general, we found that a maximum leaf step of 3 cm
`was sufficient to provide a continuous delivery with no plan degradation.
`The Elekta machines used in our department require a constant dose rate and a constant
`gantry rotation speed in the delivery of IMAT plans. This constraint is imposed in the
`optimization by requiring that all angles composing an arc have the same relative weight.
`However, the full potential of IMAT can be better realized if either the dose rate or the gantry
`rotation speed can be varied during delivery. To illustrate the effect of this constraint, we will
`present two plans for the pseudo-prostate case for each of the two scenarios: equal weight
`constraint and full IMAT potential.
`The selected change is checked against the constraints outlined above and discarded if any
`constraint is violated. If the change satisfies the delivery constraints, the new dose distribution
`is calculated. Because we are employing pre-computed pencil beams stored on disk, the
`calculation of the new dose is a simple addition or subtraction of the affected pencil beams.
`This technique leads to a very fast iteration time: approximately 1000 iterations per minute
`on a 750 MHz Pentium III. The new dose is then used to calculate the objective function.
`Using the newly calculated dose reflecting the change in field shape or weight, the new
`objective function is calculated. The objective function is a single number which represents
`the deviation from the goals of the treatment plan. More details of the objective function will
`be given in the following section. If the new objective function value is smaller (less deviation
`from the goals of the treatment plan) than the previous one, the change is accepted and a new
`change is chosen. However, if the objective function is larger (more deviation from the goals
`of the treatment plan), the change will be accepted with a certain probability given by the
`relation
`
`P = B
`
`1
`(nsucc + 1)1/T
`prob
`is the cooling rate and nsucc is the number of successes.
`where B is the initial probability, T
`0
`A success is defined as either a change that results in a lower objective function value or a
`change that satisfies the probability constraint.
`
`prob
`0
`
`(3)
`
`2.4. Objective function
`
`The objective functions used in this work were based on a simple least-square function. Two
`types of least-square objective functions were used, a ‘range-based’ and a DVH-based.
`(cid:1)
`(cid:2)
`(cid:1)
`(cid:2)
`For the ‘range-based’ objective function, a set of dose ‘ranges’ is defined for each
`(cid:1)
`(cid:2)
`high
`d low
`structure i. A ‘range’ r(i) has a low dose
`, a high dose
`, a prescription dose
`d
`r (i)
`r (i)
`d p
`and a weight wr (i). Voxels belonging to the ith structure, whose dose d falls in range
`r (i)
`high
`r(i) between dlow
`r (i) , contribute
`r (i) and d
`(cid:1)
`(cid:2)2
`d − d p
`wr (i)
`j
`to the objective function. The total objective function is given by
`nstruct(cid:3)
`i(cid:3)
`i(cid:3)
`(cid:1)
`(cid:1)
`(cid:2) × (cid:1)
`nvoxel
`dj − dlow
`− dj
`dj − d p
`
`1
`nvoxel
`i
`
`ranges
`
`n
`
`r (i)=1
`
`i=1
`
`wr (i)
`
`H
`
`j=1
`
`(cid:2) × H
`
`r (i)
`
`high
`r (i)
`
`d
`
`(cid:2)2
`
`r (i)
`
`(4)
`
`(5)
`
`Page 6 of 15
`
`

`
`Inverse planning for IMAT
`
`1081
`
`(6)
`
`1
`
`2
`
`12
`
`
`
`
`H (x) =
`
`where H (x) is the Heaviside step function given by
`0
`: x < 0
`: x = 0
`: x > 0.
`1
`For instance for the target volume with a prescription dose of 1.00 (100%), one might
`= 0.0, d
`= 0.95, d p
`= 1.00 and w1 = 20.0; (2) d low
`=
`high
`specify two dose ranges: (1) d low
`= 1.00 and w2 = 20.0. Range 1 prevents underdosage of the target
`= 5.00, d p
`1
`1
`high
`1.10, d
`2
`2
`high
`while range 2 prevents overdosage. The d
`value of 5.00 in range 2 simply represents some
`2
`dose very much higher than we would deliver, essentially stating that any voxels with a dose
`above 1.10 will be penalized with the parameters specified in range 2.
`For a critical structure such as a spinal cord that has already received significant radiation,
`= 0.0, d
`= 0.20, d p
`= 0.0 and w1 = 1.0;
`high
`we might specify two dose ranges: (1) d low1
`= 0.0 and w1 = 1000.0. Range 1 provides a relatively
`= 0.20, d
`= 5.00, d p
`1
`1
`high
`(2) d low
`1
`1
`1
`small penalty for portions of the cord which receive a dose that is less than 20% of the PTV
`prescription dose. Range 2 provides a very large penalty for any portion of the cord greater
`than 20% of the PTV prescription dose. This type of range preferentially makes voxels inside
`the cord less than 20% of the PTV prescription dose.
`The DVH-based objective function has been outlined in other works (Bortfeld et al 1997),
`so only a brief description will be given here. The DVH constraints take the form ‘no more
`than x% of the volume is to exceed y% of the dose.’ For each DVH constraint, only a subset
`of the voxels is penalized within the given structure. The subset includes those voxels whose
`∗
`∗
`dose falls between Dmax and D
`is the current dose to at least
`, where Dmax is the y% and D
`x% of the volume. The form of the penalty to these voxels is the same as that in equation (4).
`
`1
`
`3. Results and discussion
`
`3.1. Pseudo-prostate
`
`For this case, a PTV and critical structure simulating a prostate case were drawn in a cylindrical
`phantom of 20 cm diameter. We calculated 36 equispaced 8 × 8 beams each composed of 128
`10 mm × 5 mm pencil beams. The phantom was divided into 2 mm × 2 mm × 5 mm voxels.
`The dose calculation time for one beam direction was about 70 min on one 750 MHz Pentium
`III CPU. Using 12 CPUs, the calculation for all 36 beams was about 3.5 h.
`In total, six arcs were employed: three arc paths, each with two overlapping arcs. The
`to −40
`, and −110
`to −170
`◦
`◦
`◦
`◦
`◦
`◦
`arc paths were 170
`to 110
`, 40
`(we use IEC convention in
`◦
`−1 and a maximum leaf-travel speed ofs
`our clinic). A minimum gantry-rotation speed of 10
`
`−1 were applied to the optimization. Figure 2 shows two axial slices of the final plan.
`3 cm s
`The cumulative dose volume histogram (DVH) for the PTV and critical structure is shown in
`figure 3.
`Figure 4 shows a sample sequence of aperture shapes for an arc of this case. The shapes
`◦
`shown represent the apertures belonging to angles separated by 10
`. The leaf travel direction
`is horizontal and each leaf step is 0.5 cm. Examining the figure reveals that the maximum leaf
`travel between consecutive angles is no more than 3 cm.
`
`3.2. Head and neck
`Like the pseudo-prostate case, we calculated 36 equispaced 8 × 8 beams each composed of
`128 10 mm × 5 mm pencil beams. The phantom was divided into 3 mm × 3 mm × 4 mm
`
`Page 7 of 15
`
`

`
`1082
`
`M A Earl et al
`
`Figure 2. Sample axial slices with isodose lines for the pseudo-prostate case.
`
`Figure 3. Dose volume histogram for the pseudo-prostate case.
`
`voxels. The total dose calculation time for all 36 beams was about 3 h on 12 750 MHz Pentium
`III CPUs. Three arc paths were employed, each with two overlapping arcs. The angular range
`to −50
`and −70
`to −170
`◦
`◦
`◦
`◦
`◦
`◦
`of the paths was 170
`to 70
`, 50
`. The minimum gantry-rotation
`◦
`
`−1 and a maximum leaf-travel speed of 3 cm ss −1 was applied.
`speed was 10
`
`There were several clinical goals in this case. First, for the patient to retain the ability to
`salivate after completion of the treatment, the dose to the parotid glands was to be kept below
`30% of the PTV prescription of 70 Gy. Since this was recurrent disease and the patient had
`already received a significant dose to the spinal cord in previous treatments, the cord dose
`
`Page 8 of 15
`
`

`
`Inverse planning for IMAT
`
`1083
`
`Figure 4. Sample sequence of aperture shapes for an arc of the pseudo-prostate case. Steps are
`denoted by the vertical lines and are separated by 0.5 cm. The maximum leaf travel between
`consecutive angles is always 3 cm or less.
`
`was required to be below 20% of the PTV prescription dose. Finally, no more than 30% of
`the mandible was allowed to receive 50% of the PTV prescription dose. Since one of the
`features of IMAT is that the higher dose regions in normal tissue are minimized at the expense
`of creating larger low dose regions, this case was well suited for IMAT.
`An axial and sagittal slices are shown in figure 5. The 95, 50, 30 and 20% isodose lines
`are shown. Also shown are the PTV, the parotid glands, the mandible and the spinal cord. The
`sharp dose fall-off from the 95% isodose line to the 50% isodose line is evident, and the dose
`to the parotid glands is below the specification. A cumulative dose volume historgram for this
`case is shown in figure 6.
`
`3.3. Prostate
`For the prostate case, the CT data were divided into voxels of 4 mm × 4 mm × 3 mm size.
`Two bilateral arcs were applied, each with three overlapping arcs. The angular range of the
`and −30
`to −150
`◦
`◦
`◦
`◦
`paths were 150
`to 30
`. As with the previous two cases, the leaf travel
`−1 and the minimum gantry-rotation speed was set to 10
`◦
`−1.s
`speed was 3 cm s
`
`Figure 7 shows an axial slice of the final dose distribution. As with any prostate IMRT
`case, a specific clinical goal of this case was to keep the dose to the rectum to an acceptable
`level while maintaining uniform dose to the PTV. Figure 8 shows the cumulative DVH for
`the PTV and rectum, illustrating the level to which these clinical goals were achieved. The
`‘hot-spot’ in the rectum was around 98% of the PTV dose prescription and 50% of the entire
`volume received no more than 20% of the prescription dose. For the PTV, 98% of the volume
`received at least 95% of the prescription dose.
`
`Page 9 of 15
`
`

`
`1084
`
`M A Earl et al
`
`Figure 5. Sample axial and sagittal slices with isodose lines for the head and neck case.
`
`Figure 6. Dose volume histogram for the head and neck case.
`
`3.4. Variable gantry rotation or dose rate
`
`For all of the cases presented,we restricted the gantry rotation speed to be constant,or similarly,
`for the dose rate to be constant. This is reflected in the fact that during the optimization, the
`
`Page 10 of 15
`
`

`
`Inverse planning for IMAT
`
`1085
`
`Figure 7. Sample axial slice with isodose lines for the prostate case.
`
`Figure 8. Dose volume histogram for the prostate case.
`
`relative weights of segments within a single arc are forced to be constant. However, changing
`either the gantry speed or dose rate during gantry rotation will help realize the true potential
`for IMAT. To study this, we relaxed the constant weight constraint within a given arc for the
`
`Page 11 of 15
`
`

`
`1086
`
`M A Earl et al
`
`Figure 9. DVH comparison of weight constraint versus no weight constraint for the pseudo-prostate
`case. An improvement is seen for sparing of the critical structure.
`
`pseudo-prostate case. For the ‘full IMAT potential case’, the weights of the beams belonging
`an arc are independent and are allowed to be different from one another. This essentially
`simulates the effect of allowing the gantry speed to vary or the dose rate to change during
`delivery.
`Figure 9 shows the DVH comparison for the pseudo-prostate case. A noticeable
`improvement in the critical structure DVH is evident, while the PTV gains only slightly.
`The relaxed constraint on the weights of the fields within an arc can be compensated in the
`‘equal-weight’ scenario by adding additional arcs. To illustrate this, for the pseudo-prostate
`case we add one additional arc to each arc path. The comparison between the ‘varying-weight’
`and the ‘additional-arc’ cases is shown in figure 10. It is apparent that some of what was gained
`by relaxing the weight constraint can be compensated in the ‘equal-weight’ scenario by adding
`arcs. The cost of this is the additional delivery time required to deliver the additional arcs.
`
`3.5. IMAT delivery: c-shape target
`
`To demonstrate the efficiency and accuracy of delivery, we present an IMAT plan to which
`we applied the DAO algorithm described above. Figure 11 shows the geometry and arc
`configuration for the case. A c-shape target was placed in the centre of a cylindrical volume.
`Four arc-paths were employed to achieve the desired dose distribution. The two arc-paths on
`the left side of the volume were each composed of two overlapping arcs. These two arc paths
`had angular ranges of −30
`to −80
`and −100
`to −150
`◦
`◦
`◦
`◦
`. The arc-paths on the right-hand
`side of the volume were simple arcs with no overlapping segments. The angular ranges of
`◦
`◦
`◦
`◦
`these arcs were 150
`to 100
`and 80
`to 30
`.
`The optimized plan was output to an Elekta SL20 linear accelerator and delivered to
`a cylindrical phantom. The dose distribution of an axial slice was measured with Kodak
`
`Page 12 of 15
`
`

`
`Inverse planning for IMAT
`
`1087
`
`Figure 10. DVH comparison of no-weight constraint versus additional arc and no weight constraint
`for the pseudo-prostate case. Some of the loss that was observed by imposing the equal-weight
`constraint was recovered by adding additional arcs.
`
`Figure 11. Configuration of the arc delivery to the cylindrical phantom. The target has a ‘c’ shape
`and surrounds the critical structure.
`
`EDR2 film. The plan was delivered in approximately 12 min. Figure 12 shows a comparison
`between the calculated and measured doses. For both cases, the 90% and 50% isodose lines
`are plotted. For the calculated distribution, the effect of approximating the dose calculation
`
`Page 13 of 15
`
`

`
`1088
`
`M A Earl et al
`
`Figure 12. Calculated dose distribution and measured dose distribution for c-shape target in
`cylindrical phantom. The 90% and 50% isodose lines are shown as white lines.
`
`with equispaced beams is evident. The film measurement shows a smoother dose distribution
`than the calculated distribution due to the discrete beam approximation in the calculated
`distribution. Despite this, the 90% and 50% isodose lines are in good agreement. The
`◦
`agreement between the two distributions is an indication that the 3 cm separation between 10
`consecutive angles provides an adequate approximation to a continuous delivery.
`
`4. Conclusion
`
`The full potential of IMAT has gone largely unrealized due to the lack of inverse planning
`techniques which can elegantly produce such plans. Current commercially available IMRT
`systems generally require a separate leaf-sequencing step. Since IMAT imposes constraints
`in addition to the physical limitations of the MLC, leaf sequencing becomes extremely
`complicated. To bypass this problem, we have applied the DAO technique to generate inverse
`plans for IMAT. Since DAO incorporates all delivery constraints into the optimization, the
`leaf-sequencing step is eliminated, thereby significantly simplifying the optimization process.
`We have demonstrated the feasibility of applying DAO to IMAT by presenting several
`plans. All plans show that IMAT is able to achieve conformal dose distributions to the tumour
`with only six to ten arcs. The small number of arcs enables a short delivery time of 10 to
`20 min. We have also shown that we can deliver an IMAT plan on a conventional linear
`accelerator which matches the calculated dose distribution.
`
`Acknowledgment
`
`This research was supported by National Institute of Health Grant no. R29CA66075.
`
`References
`
`Bortfeld T, Haler D H, Waldron T J and Boyer A L 1994 X-ray compensation with multileaf collimators Int. J. Radiat.
`Oncol. Biol. Phys. 28 723–39
`Bortfeld T, Stein J and Preiser K 1997 Clinically relevant intensity modulation optimization using physical criteria
`Proc. XII Int. Conf. Use of Computers in Radiation Therapy (Salt Lake City, UT)
`Carol M, Dawson D and Spied R 1997 A binary volume delivery system temporal-based intensity modulation radiation
`therapy Med. Phys. 24 996–7
`
`Page 14 of 15
`
`

`
`Inverse planning for IMAT
`
`1089
`
`Chui C S, LoSasso T and Spirou S 1994 Dose calculation for photon beam with intensity modulation generated by
`dynamic jaw or multileaf collimation Med. Phys. 21 1237–44
`Cotrutz C, Kappas C and Webb S 2000 Intensity modulated arc therapy (IMAT) with centrally blocked fields Phys.
`Med. Biol. 45 2185–206
`Galvin J M, Chen X G and Smith R M 1993 Combining multileaf fields to modulate fluence distributions Int. J.
`Radiat. Oncol. Biol. Phys. 27 697–705
`Kirkpatrick S, Gelatt C D Jr and Vecchi M P 1993 Optimization by simulated annealing Science 220 67

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket