throbber
Int J Radiation Oncology Biol Phys , Vol 57, No 4, pp 1019 –1032, 2003
`Copyright © 2003 Elsevier Inc
`Printed in the USA All rights reserved
`0360-3016/03/$–see front matter
`
`doi:10.1016/S0360-3016(03)00663-1
`
`CLINICAL INVESTIGATION
`
`Ovary
`
`WHOLE ABDOMINOPELVIC RADIOTHERAPY (WAPRT) USING INTENSITY-
`MODULATED ARC THERAPY (IMAT): FIRST CLINICAL EXPERIENCE
`
`WIM DUTHOY, M.D.,* WERNER DE GERSEM, IR.,* KOEN VERGOTE, M.SC.,* MARC COGHE, LIC.,*
`TOM BOTERBERG, M.D., PHD.,* YVES DE DEENE, M.SC., PHD.,* CARLOS DE WAGTER, IR., PHD.,*
`SIMON VAN BELLE, M.D., PHD.,† AND WILFRIED DE NEVE, M.D., PHD.*
`*Division of Radiotherapy and †Department of Medical Oncology, Ghent University Hospital, Ghent, Belgium
`
`Purpose: Whole abdominopelvic radiation therapy (WAPRT) is a treatment option in the palliation of patients
`with relapsed ovarian cancer. With conventional techniques, kidneys and liver are the dose- and homogeneity-
`limiting organs. We developed a planning strategy for intensity-modulated arc therapy (IMAT) and report on the
`treatment plans of the first 5 treated patients.
`Methods and Materials: Five consecutive patients with histologically proven relapsed ovarian cancer were sent
`to our department for WAPRT. The target volumes and organs at risk (OAR) were delineated on 0.5-cm-thick
`CT slices. The clinical target volume (CTV) was defined as the total peritoneal cavity. CTV and kidneys were
`expanded with 0.5 cm. In a preset range of 8° interspaced gantry angles, machine states were generated with an
`anatomy-based segmentation tool. Machine states of the same class were stratified in arcs. The optimization of
`IMAT was done in several steps, using a biophysical objective function. These steps included weight optimization
`of machine states, leaf position optimization adapted to meet the maximal leaf speed constraint, and planner-
`interactive optimization of the start and stop angles. The final control points (machine states plus associated
`cumulative monitor unit counts) were calculated using a collapsed cone convolution/superposition algorithm. For
`comparison, two conventional plans (CONV) were made, one with two fields (CONV2), and one with four fields
`(CONV4). In these CONV plans, dose to the kidneys was limited by cerrobend blocks. The IMAT and the CONV
`plans were normalized to a median dose of 33 Gy to the planning target volume (PTV). Monomer/polymer gel
`dosimetry was used to assess the dosimetric accuracy of the IMAT planning and delivery method.
`Results: The median volume of the PTV was 8306 cc. The mean treatment delivery time over 4 patients was 13.8
`min. A mean of 444 monitor units was needed for a fraction dose of 150 cGy. The fraction of the PTV volume
`receiving more than 90% of the prescribed dose (V90) was 9% higher for the IMAT plan than for the CONV4
`plan (89.9% vs. 82.5%). Outside a build-up region of 0.8 cm and 1 cm away from both kidneys, the inhomogeneity
`in the PTV was 15.1% for the IMAT plans and 24.9% for the CONV4 plans (for CONV2 plans, this was 34.9%).
`The median dose to the kidneys in the IMAT plans was lower for all patients. The 95th percentile dose for the
`kidneys was significantly higher for the IMAT plans than for the CONV4 and CONV2 plans (28.2 Gy vs. 22.2
`Gy and 22.6 Gy for left kidney, respectively). No relevant differences were found for liver. The gel-measured dose
`was within clinical planning constraints.
`Conclusion: IMAT was shown to be deliverable in an acceptable time slot and to produce dose distributions that
`are more homogeneous than those obtained with a CONV plan, with at least equal sparing of the OARs.
`© 2003 Elsevier Inc.
`
`Intensity-modulated arc therapy (IMAT), Whole abdominopelvic radiotherapy (WAPRT), Ovarian cancer.
`
`INTRODUCTION
`
`The treatment options for most patients with relapsed ovar-
`ian cancer are palliative. For patients who relapse within 12
`months after platinum-based schedules, second- or third-
`
`line chemotherapy is often used, but the response rates
`average between 10% and 30% (1).
`A difficult problem to palliate is bowel obstruction.
`Surgery may be attempted but the disease is often multi-
`focal making palliative resection impossible. In reports of
`
`Reprint requests to: Wim Duthoy, Ghent University Hospital,
`Division of Radiotherapy, De Pintelaan 185, 9000 Ghent, Bel-
`gium. Tel: 32-92403074; Fax: 32-92403863; E-mail: wimd
`@krtkg1.rug.ac.be
`The project “Conformal Radiotherapy Ghent University Hospi-
`tal” is supported by the Belgische Federatie tegen Kanker and by
`grants from the Fonds voor Wetenschappelijk Onderzoek (FWO)
`Vlaanderen (G.0183.03),
`the University of Ghent
`(GOA
`
`12050401, BOF 01112300, 011VO497, 011B3300), and the Cen-
`trum voor Studie en Behandeling van Gezwelziekten.
`Acknowledgments—Wim Duthoy is a Research Assistant (Aspir-
`ant) of the FWO. Yves De Deene is Post-Doctoral Research
`Fellow of the FWO.
`Received Jan 16, 2003, and in revised form May 21, 2003.
`Accepted for publication May 27, 2003.
`
`1019
`
`Page 1 of 14
`
`Elekta Exhibit 1005
`
`

`
`1020
`
`I. J. Radiation Oncology ● Biology ● Physics
`
`Volume 57, Number 4, 2003
`
`Redman et al. (2) and Krebs et al. (3), 10 –15% of patients
`died within 8 weeks after surgery and 35–38% had no
`clinical benefit. The results of chemotherapy for bowel
`obstruction are disappointing. In a report by Abu-Rustum et
`al. (4), a response was observed in only 1 of 18 patients
`treated with chemotherapy for bowel obstruction. Radiation
`therapy seems to compare favorably with second- or third-
`line chemotherapy, with symptom response rates between
`63% and 79% with a median duration between 4 and 9
`months (5–7). Given the pattern of spread of ovarian cancer,
`whole abdominopelvic radiation therapy (WAPRT) could
`be the radiation technique of choice, eventually with a boost
`to sites of gross tumor.
`Maximum tolerated dose levels to the whole kidney are
`often set to 20 Gy or less. When the kidneys are blocked
`from radiation, conventional techniques result in underdos-
`age of the peritoneal regions in the blocked areas.
`We investigated the potential of intensity-modulated ra-
`diation therapy (IMRT) to spare kidneys and liver. For
`reasons described in the discussion section, we selected
`intensity-modulated arc therapy (IMAT) (8) as the most
`appropriate IMRT technique for WAPRT.
`This article describes the results of the translational re-
`search that was performed to bring IMAT into the clinic for
`WAPRT. We report on treatment planning and delivery for
`the first 5 patients treated with IMAT-WAPRT. The clinical
`results of the Phase I study will be reported elsewhere.
`
`METHODS AND MATERIALS
`
`Delineation of target volumes and OARs
`Between November 2001 and October 2002, 5 patients
`with a relapse of a histologically proven adenocarcinoma of
`the ovary were treated by IMAT. All patients signed in-
`formed consent for IMAT to the whole abdomen. A plan-
`ning computed tomography (CT) scan was performed with
`the patient lying supine, with both arms resting under the
`head. The upper border of the scanned volume was located
`10 cm cranial to the diaphragm, whereas the lower border
`was defined as 10 cm caudal to the obturator foramina.
`Several 0.5-cm-thick sequential CT slices were acquired
`without contrast enhancement. Clinical
`target volume
`(CTV) was defined as the total peritoneal cavity, with the
`inclusion of iliac and para-aortic lymph node regions. A 0.5-
`cm rim of liver, adjacent to the peritoneum, was also in-
`cluded into the CTV. The upper CTV boundary was defined
`by the highest CT slice on which the diaphragmatic dome
`was visible, and the most caudal drawing of the CTV was on
`the level of the bottom of the obturator foramina. A plan-
`ning target volume (PTV) was made by a 3D expansion of
`the CTV with a margin of 0.5 cm in all directions. The
`kidneys and the liver (hereafter called OARs: organs at risk)
`were drawn as visualized on CT, and the kidneys were
`expanded with 5 mm (kidney_exp_5mm), to account for
`setup inaccuracy and organ motion. PTV and OARs over-
`lap, which would result in a conflicting requirement from
`the optimization algorithm. A similar problem arises with
`
`those parts of the PTV in the build-up region. Both prob-
`lems were solved by defining subvolumes inside the PTV
`that are used as optimization volumes (PTV_optim).
`PTV_optim was nowhere closer than 0.8 cm to the skin
`surface or 0.5 cm to the expanded kidneys. A surrounding
`structure was made by subtracting the PTV from the total
`scanned volume of the patient. This structure (sur_0cm) was
`used to avoid hot spots outside the PTV. A general descrip-
`tion of the use of PTV subvolumes and surrounding struc-
`tures for IMRT plan optimization is found elsewhere (9).
`
`IMAT planning procedure
`Generation of machine states by ABST. An anatomy-
`based segmentation tool (ABST) was developed at our
`institution to create segments for step-and-shoot IMRT (10).
`For IMAT planning, ABST is used to create an initial set of
`segments, which we call machine states. A machine state is
`described by a set of machine parameters that uniquely
`define the beam incidence, aperture, and photon beam qual-
`ity. After definition of the isocenter location and with the
`collimator, table top and isocenter rotations at 0°, ABST
`generated machine states per 8° of gantry rotation. Restric-
`tion of the range of gantry angles was needed to avoid
`beams traversing metal components of the couch before
`entering the patient. The couch on the linear accelerator
`(Elekta, Crawley, UK) used for the IMAT treatments has
`two C-arms, which can be positioned at one of the 30°
`discrete angles. The largest range of possible gantry angles
`was obtained by setting the arms at 120°, measured from
`their lateral position (Fig. 1). For Patients 2, 3, and 5, a class
`solution was used, implying that the initial machine states
`were generated using a fixed set of parameters, including
`start-and-stop gantry angles, widths of the segments, and
`conformal avoidance structure. The arcs used by the class
`solution are shown in Fig. 1. For Patient 4, where the
`craniocaudal extent of the PTV was 42 cm, a second iso-
`center was defined 12 cm from the first isocenter in caudal
`direction. Only a longitudinal table shift is required to
`perform the transition between the two isocenters. An ad-
`ditional “pelvic arc” was made around a structure called
`PTV_pelvis, with the L5-S1 intervertebral space as the
`upper border. The collimator was rotated by 90° to have the
`leaf movements in the craniocaudal direction and allow for
`feathering in the junction region (11).
`For each gantry angle, ABST generates multiple machine
`states that differ only by apertures of the multileaf collima-
`tor (MLC). Each beam’s-eye view (BEV) projection of a
`MLC aperture covers a part of the PTV at one side of the
`anatomic structure that is to be avoided. A margin of 0.8 cm
`around the PTV is used to account for penumbra. For each
`gantry angle, the machine states differ from each other by
`the degree of coverage of the BEV projection of the PTV.
`For a detailed description of ABST, refer to De Gersem et
`al. (10). These machine states are useful to create intensity
`levels that increase with decreasing distance to the anatomic
`structure that is to be spared. It was shown by Brahme and
`others (12, 13) that such intensity profiles are useful to
`
`Page 2 of 14
`
`

`
`IMAT for whole abdominopelvic radiotherapy ● W. DUTHOY et al.
`
`1021
`
`Fig. 1. Class solution for the IMAT plan. A transverse plane through the patient can be appreciated, with the PTV, the
`left kidney (LK) and the right kidney (RK). The arcs are depicted by circle segments. Machine states for one arc (S0R
`LK) are shown every 16°, from ⫺128° to 0°. The thick gray lines represent the jaw position, each small white bar stands
`for one leaf. Abbreviations: S0R LK: Arc composed by the machine states covering the total BEV projection of the PTV
`passing the right side of the LK. S0L RK: Arc composed by the machine states covering the total BEV projection of
`the PTV passing the left side of the RK. S1R LK: Arc composed by the machine states covering a 3 cmwide area of
`the PTV at the right side of the LK. S1L RK: Arc composed by the machine states covering a 3 cmwide area of the
`PTV at the left side of the RK. The metal C-arms are shown as the gray squares, respectively in their most lateral
`position (asterisk) for the sliding window, and in their 120° position (⫹ sign). The large dashed arrow represents the
`beam direction for the delivery of the sliding window (SW).
`
`create homogeneous dose distributions to a concave PTV
`that conformally avoid anatomic structures at risk. The
`machine states were stratified in classes, the first class
`consisting of machine states with the largest area of MLC
`aperture, the second class consisting of machine states with
`the second largest area of MLC aperture, and so on. Because
`of the algorithm inside ABST that creates MLC apertures
`avoiding anatomic structures with—in this case—smooth
`surfaces, MLC apertures, which belong to the same class of
`machine states, do not differ much from one gantry angle to
`the next, their angular separation being only 8°. Hence, the
`leaf travel required when moving from one gantry angle to
`the next is small for machine states of the same class, which
`is preferable for dynamic transitions as in IMAT.
`For Patients 2–5, an additional posterior “sliding win-
`dow” intensity-modulated beam with a 90° collimator rota-
`
`tion was used. For this beam, the table bars were put on their
`most lateral position. Because of the position of these metal
`bars, and the impossibility to prescribe arcs traveling over
`the 180° gantry point, the range for posterior arcs is very
`limited. Therefore, a sliding window intensity-modulated
`beam with static gantry angle was preferred to boost the
`most posterior region of the PTV. The control points for this
`sliding window were made manually. For Patient 4, the
`most caudal isocenter was selected for this sliding window
`beam.
`Creation and optimization of control points. The machine
`instruction file to deliver arc therapy with dynamic MLC
`consists of a sequence of control points (see Fig. 1). A
`control point is defined as a machine state plus a monitor
`unit count (MUC) value. Delivery of a sequence of control
`points implies that the prescribed machine state has to be
`
`Page 3 of 14
`
`

`
`1022
`
`I. J. Radiation Oncology ● Biology ● Physics
`
`Volume 57, Number 4, 2003
`
`Fig. 2. Picture representing the angular delivery rate. One “virtual”
`arc (dashed line) as well as three deliverable arcs (solid lines) are
`shown. The “virtual” arc is not deliverable on the Elekta linac. The
`deliverable arcs approximate the optimized virtual arc. The final
`angular delivery rate (W) of the deliverable arcs is optimized as
`described in the text.
`
`reached at the MUC value for each control point. The
`transition from a control point to the next is slaved by the
`monitor unit (MU) counter; each parameter (leaf positions,
`jaw positions, and gantry angle) that changes between two
`control points is linearly interpolated as function of the
`MUC value. The beam is paused if the control software
`detects that a machine parameter is outside tolerance to the
`(linearly interpolated) position prescribed by the machine
`state. Control point optimization involves the machine
`states—and more precisely, the leaf positions—as well as
`the MUC values and is done by a segment outline and
`weight adapting tool (SOWAT) (14), modified for IMAT
`purposes (SOWAT-IMAT). The main difference between
`SOWAT and SOWAT-IMAT resides in the maximal leaf
`velocity (MLS) constraints. After the control point genera-
`tion and after each leaf position optimization cycle, a leaf
`velocity constrainer (LVC) adapts leaf positions of all con-
`trol points to obey maximum leaf speeds, minimum dis-
`tances to opposed and diagonally opposed leaves, and max-
`imum leaf position extends (method unpublished). MUCs
`are optimized for each step, a step being defined as the
`transition from one control point to the next. The objective
`function on which the optimization is done is a biophysical
`model and has been described and discussed elsewhere (15,
`16).
`Transformation to deliverable arcs. As a result of SO-
`WAT-IMAT, n machine states and n-1 weights are ob-
`tained. These weights are numbers of monitor units that
`have to be delivered while the machine moves from one
`control point to the next. Apart from leaf and jaw travel,
`such motion involves a 8° gantry rotation. Because the
`Elekta SL-series of linear accelerators was designed to
`deliver arcs with a gantry rotation speed directly propor-
`tional to the dose rate, the number of monitor units delivered
`per degree (angular delivery rate) must remain constant over
`the whole arc. This condition is not secured by SOWAT-
`
`Fig. 3. Adapted Rando phantom as used for the gel dosimetry. At
`each side of the Barex cast, three Rando slices were added to
`obtain full scatter conditions. Tape is attached to the phantom to
`draw laser lines in transverse, sagittal, and coronal reference
`planes. Seven markers are attached to the phantom on the laser
`lines to facilitate positioning. The transverse plane indicated in the
`middle of the barex phantom is 13.5 cm cranial to the treatment
`isocenter. A barex screw is used to close the phantom at the place
`where the gel was inserted, visible left to the sagittal plane.
`
`IMAT. In fact, the (requested) angular delivery rate may be
`different for each 8°-sector of gantry rotation. This problem
`is solved by splitting each arc, which features a variable
`angular delivery rate into multiple overlapping arcs each
`with a constant angular delivery rate (Fig. 2). This proce-
`dure also provides the start-and-stop angles of the delivered
`arcs. The plan is finalized by a SOWAT-IMAT optimization
`cycle, which involves optimization of leaf positions and the
`angular delivery rate (equal for all sectors of the arc to keep
`the angular delivery rate constant within each arc).
`Dose prescription and computation. The prescribed dose
`was 33 Gy (median dose in the PTV), given in 22 fractions.
`Except for the first patient, IMAT plans were accepted using
`the following clinical criteria: less than 5% of the PTV
`volume was allowed to receive more than 107% of the
`prescribed dose, and more than 95% of the volume of
`PTV_optim had to receive more than 90% of the prescribed
`dose; less than 5% and 20% of the kidneys_exp_5mm
`should receive more than 30 Gy and 25 Gy, respectively,
`whereas the median dose had to be lower than 18 Gy; the
`median liver dose was constrained to 30 Gy. After clinical
`
`Page 4 of 14
`
`

`
`IMAT for whole abdominopelvic radiotherapy ● W. DUTHOY et al.
`
`1023
`
`Table 1. Details about the intensity-modulated arc therapy treatments
`
`Isocenters
`
`Arcs
`
`Control points
`
`Monitor units
`
`Delivery time (min)
`
`Patient 1
`Patient 2
`Patient 3
`Patient 4
`Patient 5
`
`1
`1
`1
`2
`1
`
`11
`7 ⫹ 1 SW
`7 ⫹ 1 SW
`7 ⫹ 1 SW
`6 ⫹ 1 SW
`
`73
`110
`78
`99
`148
`
`568
`387
`378
`528
`359
`
`⬍20
`11.6 ⫾ 2.5
`15.3 ⫾ 3.6
`18.1 ⫾ 3.6
`13.5 ⫾ 1.3
`
`Abbreviation: SW ⫽ sliding window.
`For the first patient, no systematic measurements were done concerning the delivery time. The monitor units are for
`one fraction of 150 cGy.
`
`constraints were met by the optimization procedure, a final
`dose computation was performed for 18 MV with the col-
`lapsed cone convolution/superposition algorithm from Pin-
`nacle (Philips Medical Systems, Best, The Netherlands). A
`final optimization of the MUs of all arcs was done using the
`results of this dose computation. The value of a MU is such
`that 100 MUs correspond to 1 Gy at reference depth (5 cm
`for 6 MV and 10 cm for 18 MV) for a 10 ⫻ 10 cm field and
`a source-detector distance of 100 cm.
`IMAT treatment delivery. For each arc, a prescription file
`containing the sequence of control points and related mon-
`itor units is generated and networked to an SLiPlus 18-MV
`linear accelerator (Elekta). The IMAT treatment is delivered
`in local service mode using prototype dynamic control
`software (Elekta), operating as described previously (17).
`Delivery of dynamic prescriptions is not possible in clinical
`mode on the Elekta linear accelerators. The local service
`mode is operated in the same interlock class as in clinical
`mode. Therefore, tolerances used by the linac’s control
`system are the same as for clinical mode.
`
`Conventional plans
`For each patient, two different conventional (CONV)
`plans were made. The first plan was the widely used “AP/
`PA” technique, using an anterior and a posterior field
`(CONV2). A second plan used four beams (CONV4): an-
`terior, posterior, and two lateral fields. The field margins
`were drawn with a 1-cm margin around the PTV in all
`directions. Kidney blocks covered the BEV projection of
`the kidneys with a margin of 0.5 cm. For the CONV2 plans,
`the posterior field was duplicated in two segments, an open
`segment and a segment where kidney blocks were inserted.
`Respectively, 6-MV and 18-MV photons were used for the
`anterior and posterior field. For the CONV4 plans, all four
`fields were duplicated in two segments each, an open seg-
`ment and a segment with kidney blocks. Here, 18-MV
`photons were used for all fields.
`Optimization of the relative segment weights was done
`by the planner to reach a median dose to the expanded
`kidneys between 18 and 20 Gy. Median dose to the liver
`was constrained to 30 Gy. Dose computation was done with
`the same collapsed cone convolution/superposition algo-
`rithm.
`
`Treatment evaluation
`Delivery time, defined as the time between the start of the
`first arc and the end of the last arc or sliding window, was
`measured for Patients 2–5. Additionally, the setup time was
`measured from the entrance of the treatment room by the
`patient to the start of the first arc. This includes the time
`necessary to acquire portal images and correction of the
`patient position.
`Comparison of dose distributions obtained with the
`CONV2, CONV4, and the IMAT plan was done after nor-
`malizing the median dose of the PTV to 33 Gy. To evaluate
`the dose homogeneity in the target volumes, an inhomoge-
`neity factor U95/5 was defined as the difference between the
`95th percentile dose (D95) and the 5th percentile dose (D5),
`divided by the median dose (Dmed). We preferred to use the
`D95 and the D5 above the maximum and minimum dose,
`because an underdosage was allowed in the region close to
`the kidneys. Other endpoints for the target volumes were the
`first percentile dose (as a surrogate for minimum dose), the
`99th percentile dose, and the ratio of volume of the target
`structure receiving more than 95% of the prescribed dose
`(V95) over the total volume. For the parallel-element organs
`kidney and liver, the Dmed was used. The dose–volume
`histograms (DVHs) were reconstructed for the 5 patients by
`calculating the mean dose and the standard error of the
`mean at every 5% volume level. The paired Student t test
`was used. All tests were two-tailed and p ⬍ 0.05 was
`considered as statistically significant.
`
`Dosimetric verification of the IMAT treatment
`Monomer/polymer gel dosimetry was used for 3D dose
`verification of the whole IMAT procedure. A deoxygenated
`hydrogel infused with acrylic monomers forms the basis for
`this dosimetric technique (18). Highly reactive radicals,
`formed by radiolysis during irradiation of the gel, initiate a
`polymerization reaction. The amount of polymer formed is
`related to the absorbed dose. Formation of polymer clusters
`in the water-equivalent gel increases the local spin-spin
`relaxation rate (R2), a typical magnetic resonance (MR)
`contrast parameter. Therefore, MR imaging (MRI) can be
`used to visualize the amount of polymer formed and sub-
`sequently the dose distribution in the gel. With monomer/
`polymer gel dosimetry, it is possible to obtain absorbed dose
`information in 3D with high spatial accuracy (19). For a
`
`Page 5 of 14
`
`

`
`1024
`
`I. J. Radiation Oncology ● Biology ● Physics
`
`Volume 57, Number 4, 2003
`
`Fig. 4. (a) Intensity profiles (rescaled) for the delivered plan of Patient 2. The CT slice through the isocentric plane is
`shown, with the dose distribution in Gy. The PTV (dotted line) extends outside the scanned volume (asterisks). (a)
`Intensity profiles, generated at a range of gantry angles from ⫺128° to 128°, are plotted around the CT slice. (b) Intensity
`profiles for the gantry at ⫺104°, 104° and for the posterior beam, delivered as a sliding window (180°).
`
`Page 6 of 14
`
`

`
`IMAT for whole abdominopelvic radiotherapy ● W. DUTHOY et al.
`
`1025
`
`Table 2. Summary of the DVH data showing averages ⫾ standard deviations
`
`IMAT
`
`CONV2
`
`p value
`
`CONV4
`
`p value
`
`PTV
`V95 (%)
`V90 (%)
`V107 (%)
`D1 (Gy)
`D99 (Gy)
`U95/5 (%)
`PTV_optim
`D1 (Gy)
`U95/5 (%)
`Left kidney (expanded)
`Dmed (Gy)
`D95 (Gy)
`Right kidney (expanded)
`Dmed (Gy)
`D95 (Gy)
`Liver
`Dmed (Gy)
`
`82.2 ⫾ 6.5
`89.9 ⫾ 5.7
`2.7 ⫾ 3.4
`19.0 ⫾ 10.8
`35.8 ⫾ 0.9
`28.1 ⫾ 15.8
`
`27.0 ⫾ 2.8
`15.1 ⫾ 5.8
`
`16.1 ⫾ 3.6
`28.2 ⫾ 1.3
`
`13.6 ⫾ 3.9
`26.0 ⫾ 2.9
`
`24.4 ⫾ 6.3
`
`76.8 ⫾ 5.2
`80.1 ⫾ 4.6
`6.4 ⫾ 3.3
`16.4 ⫾ 9.1
`36.1 ⫾ 0.6
`42.3 ⫾ 7.9
`
`21.7 ⫾ 0.9
`34.9 ⫾ 2.5
`
`19.9 ⫾ 0.6
`22.6 ⫾ 0.8
`
`19.3 ⫾ 1.5
`22.8 ⫾ 1.0
`
`22.8 ⫾ 10.1
`
`0.11
`0.01
`0.18
`0.09
`0.53
`0.03
`
`0.02
`⬍0.01
`
`0.11
`⬍0.01
`
`0.02
`0.02
`
`0.44
`
`73.6 ⫾ 5.9
`82.5 ⫾ 6.1
`7.2 ⫾ 4.41
`16.6 ⫾ 9.3
`36.5 ⫾ 0.9
`34.4 ⫾ 10.9
`
`24.5 ⫾ 1.2
`24.9 ⫾ 4.1
`
`19.4 ⫾ 0.4
`22.2 ⫾ 0.4
`
`18.6 ⫾ 1.1
`22.2 ⫾ 1.0
`
`29.2 ⫾ 2.0
`
`0.01
`0.01
`0.05
`0.10
`0.02
`0.10
`
`0.12
`0.01
`
`0.11
`⬍0.01
`
`0.02
`0.08
`
`0.09
`
`Abbreviations: IMAT ⫽ intensity-modulated arc therapy; CONV2 ⫽ conventional plan with an anteroposterior and a
`posteroanterior field; CONV4 ⫽ conventional plan with four-field technique; PTV ⫽ planning target volume; PTV_optim ⫽
`PTV without build-up region of 0.8 cm and with the exclusion of the expanded kidneys with an extra margin of 5 mm; V95,
`V90, and V107: the partial volume (percent) receiving more than 95%, 90%, and 107% of the prescribed dose, respectively; D99
`共D95–D5兲
`⫽ inhomogeneity factor, defined as
`⫽ Dose given to 99% and 1% of the volume, respectively; U95/5
`and D1
`, with
`Dmed
`D95 the 95th percentile dose; D5 the 5th percentile dose, and Dmed the median dose.
`
`more detailed review on this subject, the reader is referred
`to De Deene et al. (20).
`A Barex (Cifra, Chateau Thierry, France) cast was
`vacuum molded on the abdominal region of the Rando
`phantom (Alderson Research Laboratories, Stamford,
`CT). At our laboratory, the maximum amount of mono-
`mer/polymer gel that can be produced in one batch is 10
`L. Hence, the entire volume irradiated with IMAT can
`not be verified by one gel dosimetry experiment. We
`chose to limit
`the phantom geometry to that part of
`RANDO containing the (dosimetrically most interesting)
`region around the kidneys. Supports on the cranial and
`caudal side, marker lines and placement of fiduciary
`markers (Medtronic, Louisville, KY) on the surface fa-
`cilitated a reproducible positioning of the gel phantom
`during CT scanning, IMAT delivery, and MRI (Fig. 3).
`Three supplemental Rando slices were placed alongside
`the phantom on the cranial and caudal side during CT
`scanning and treatment delivery to ascertain full scatter
`conditions in the gel upon irradiation. Spiral CT scans
`(Siemens Somatom Plus 4, Erlangen, Germany) of the
`gel-filled phantom were transferred to the planning sys-
`tem. Volumes of interest (kidneys, liver, and PTV) from
`the first patient were transferred to the CT data set. For
`this setup, an IMAT plan was made by using the methods
`described previously. This resulted in a plan with six arcs
`and one sliding window. To cover the maximum response
`range of the gel while avoiding gradient dependent non-
`linearities in dose response (21), the monitor units were
`multiplied with a factor of five, thus giving a median dose
`
`of 7.5 Gy to the PTV. For calibration purposes, gel-filled
`test tubes were irradiated to known doses (0 –10 Gy,
`every 1 Gy) to establish a dose-R2 relationship. The gel
`dosimeter and test tubes were scanned together in the
`body coil of a 1T MR system (Expert, Siemens, Erlangen,
`Germany). A 26 spin-echo sequence was applied with a
`Carr-Purcell Meiboom-Gill (CPMG) RF pulse encoding
`scheme and equidistant echo spacing (TE ⫽ 40 –1080
`ms). The phantom was scanned in 33 adjacent transverse
`slices each with a slice thickness of 5 mm. A field of view
`of 320 mm and image resolution of 128 ⫻ 128 resulted in
`an in-plane resolution of 2.5 mm. Ideally, a homogeneous
`unirradiated gel phantom should produce identical R2
`readings throughout
`the entire volume when scanned
`with MRI. A homogeneity study was done on the IMAT
`phantom filled with a blank gel and the MRI protocol was
`adjusted to minimize R2 variations related to temperature
`deviations and radiofrequent inhomogeneities.
`A linear regression was used to describe the dose-R2
`relationship. The gel-measured dose grid was transferred to
`the planning system and scaled to the prescription dose (33
`Gy), which allowed the comparison of measured and com-
`puted DVHs of the different structures inside the abdomi-
`nopelvic phantom,
`truncated to the volume of the gel-
`phantom. Low’s ␥-index (22) was calculated in 3D (dose
`difference criterion ⫽ 7.5%, distance to agreement ⫽ 5
`mm) as a guide to pinpoint significant deviations between
`the computed and measured dose matrix. A ␥-index above
`1 indicates that the specified tolerances are not met.
`
`Page 7 of 14
`
`

`
`1026
`
`I. J. Radiation Oncology ● Biology ● Physics
`
`Volume 57, Number 4, 2003
`
`Fig. 5. DVHs compiled from the data of the 5 patients. Solid lines and bold dots represent the IMAT plans (mean ⫾
`standard error of the mean). Dashed lines and circles represent the 2D plans. (a-c) DVHs of the IMAT and CONV2
`plans. (d-f) DVHs of the IMAT and CONV4 plans. (a) ⫹ (d) DVHs of PTV and expanded left kidney. (b) ⫹ (e) DVHs
`for PTV_whbu and expanded right kidney. (c) ⫹ (f) DVHs of PTV_optim and liver.
`
`RESULTS
`
`The median volume of the PTV in the 5 patients was
`8306 cc (range 5717–9054 cc), and the median of the
`craniocaudal length which had to be covered was 36 cm.
`Details on the treatment plans and delivery times are shown
`in Table 1. Mean delivery time for Patients 2–5 was 13.8
`min (range 9.5–24.5 min). Less than 30% of the given
`fractions had a delivery time exceeding 15 min. Of these,
`50% were seen in Patient 4, who had two isocenters, ne-
`cessitating entrance of the treatment room to perform a
`cranial shift of the patient. For 1 patient (chronologically the
`last), the setup time was measured over all the fractions, and
`showed a mean of 8 ⫾ 2.9 min. Though actively asked, no
`patient complained about the rotating gantry. As an exam-
`ple, the obtained intensity profiles for Patient 5 are shown in
`Fig. 4.
`
`DVH analysis and dose distributions
`The DVH data for the 5 patients are summarized in
`Table 2 and graphically displayed in Fig. 5. For both the
`
`CONV4 and IMAT plans, there is a large variation in
`minimal dose (represented by D1) in the PTV (range
`0.2–26.1 Gy), resulting from the PTV extending outside
`the skin in 1 patient. When considering the PTV without
`a build-up region of 8 mm (PTV_whbu), the very low
`doses that are the result of the ICRU PTV definition
`rather than of the planning technique are eliminated
`(range of D1 19.5–26.9 Gy). The homogeneity in the
`PTV_whbu is better for the IMAT plan than for the
`CONV4 plan (U95/5 is 20% and 32%, respectively; p ⫽
`0.01). Because of the strong dose constraint to the ex-
`panded kidneys and the possibility to generate concave
`dose distributions, we expected an underdosage in the
`PTV in the region around the kidneys. For the PTV_op-
`tim, the mean (and the standard deviation) of the V95 was
`78.4% (⫾ 2.2%) for the CONV4 plan, and 88.9% (⫾
`5.1%) for the IMAT plan (p ⫽ 0.01). For the V90, these
`values were 87.7% (⫾ 3.0%) and 95.8% (⫾ 3.5%) for the
`CONV4 and the IMAT plan, respectively (p ⫽ 0.02). The
`comparison between the IMAT plan and the CONV2 plan
`
`Page 8 of 14
`
`

`
`IMAT for whole abdominopelvic radiotherapy ● W. DUTHOY et al.
`
`1027
`
`Fig. 6. Dose distributions for Patient 1, showing the delivered IMAT plan in a transverse (a) and sagittal (d) plane, and
`the conventional comparison plans in (b) ⫹ (e) for the CONV2 and (c) ⫹ (f) for the CONV4 plan. Isodose values are
`in Gy. The PTV is delineated with a dotted line. The kidneys (LK: left kidney and RK: right kidney) are delineated by
`dashed lines, while the liver (L) is circled by a dashed-dotted line. The dotted straight line in (a) and (d) indicates the
`transection planes in (d-f) and (a-c), respectively.
`
`(see Table 1) shows a significant increase in homogeneity
`(expressed by U95/5) for the PTV and for PTV_optim by
`the IMAT plans.
`The median dose to the kidneys was lower for the IMAT
`plan in all patients when compared with both CONV plans.
`The maximal doses to the kidneys were significantly higher
`for the IMAT plan than for the CONV plans, as can be seen
`in Fig. 5. For the liver, no significant differences were found
`between the IMAT and the CONV plans. The higher ho-
`mogeneity by

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket