`Instruction Sheet
`for the PASCO
`Model OS-9286A
`MERCURY VAPOR LIGHT SOURCE
`
`Rails for Mounting Filter
`
`Remove screws shown in
`black (8), then remove plate
`to inspect or replace lamp.
`
`To
`115 VAC, 60 Hz
`or
`220 VAC, 50Hz
`
`Hg
`
`LIGHT
`SOURCE
`OS-9286
`
`ON
`
`OFF
`
`DISCONNECT INPUT POWER
`BEFORE REPLACING FUSE.
`FOR CONTINUED FIRE PROTECTION
`REPLACE ONLY WITH 4.0A, 250 V,
`TYPE T FUSE, (PER IEC 127-2/III).
`
`LINE VOLTAGE RANGE: 108 - 132 VAC
`POWER: 125 W MAX
`FREQ.: 47 - 63 HZ
`
`~
`115 VOLTS
`
`FOR CONTINUED FIRE PROTECTION
`REPLACE ONLY WITH 2.0A, 250 V,
`TYPE T FUSE, (PER IEC 127-2/III).
`
`LINE VOLTAGE RANGE: 198 - 242 VAC
`POWER: 125 W MAX
`FREQ.: 47 - 63 HZ
`
`~
`220 VOLTS
`
`FOR CONTINUED FIRE PROTECTION
`REPLACE ONLY WITH 2.0A, 250 V,
`TYPE T FUSE, (PER IEC 127-2/III).
`
`LINE VOLTAGE RANGE: 216 - 264 VAC
`POWER: 125 W MAX
`FREQ.: 47 - 63 HZ
`
`~
`240 VOLTS
`
`On/OFF Switch
`
`Fuse:
`115 V – 4 amp, slo-blo
`220 V – 2 amp, slo-blo
`
` Note: For maximum life of the mercury vapor
`(cid:5)
`lamp:
`Always operate the light source in its upright
`position.
`If you are going to use the light source more
`than once during the day, leave it on. Lamp
`wear results more from turning the light source
`on and off than from steady operation.
`
`(cid:1) (cid:2)
`
`Introduction
`The PASCO scientific Model OS-9286A Mercury Vapor
`Light Source provides approximately 3,000 lumens of
`light in the mercury spectrum. The 100 watt light source
`comes ready to use, with a built-in power supply so the
`unit can be powered from a standard 115 VAC, 60 Hz
`outlet (OS-9286A-220 is powered from a 220 VAC, 50
`Hz outlet). Cooling fins and air vents on the sturdy alu-
`minum case ensure cool, safe operation. In addition, rails
`on the front and rear of the case can be used for mounting
`standard 2-inch by 2-inch filters, so that monochromatic
`light can be obtained.
`
`© 1990 PASCO scientific
`This instruction sheet edited by: Dave Griffith
`
`10101 Foothills Blvd. • P.O. Box 619011 • Roseville, CA 95678-9011 USA
`Phone (916) 786-3800 • FAX (916) 786-8905 • email: techsupp@PASCO.com
`
`1
`
`better
`ways to
`teach physics
`ASML 1330
`ASML 1029
`
`
`
`Mercury Vapor Light Source
`
`012-02262H
`
`400
`
`U.V.
`
`480
`
`NOTE: U.V. is
`blocked by the
`outer glass
`envelope.
`
`Green
`
`Blue
`
`Yellow
`
`Violet 1
`
`300
`
`200
`mW
`5 nm
`
`100
`
`Violet 2
`
`300
`
`500
`400
`Wavelength (nm)
`Spectral Power Distribution per 1000 Lumens
`
`600
`
`Color
`
`Frequency (Hz) Wavelength (nm)
`
`Yellow
`Green
`Blue
`Violet
`Ultraviolet
`
`5.18672E+14
`5.48996E+14
`6.87858E+14
`7.40858E+14
`8.20264E+14
`
`578
`546.074
`435.835
`404.656
`365.483
`
`Wavelength of the Mercury Spectral Lines
`All values except wavelength for yellow line are
`from Handbook of Chemistry and Physics, 46th ed.
`The wavelength of the yellow was determined ex-
`perimentally using a 600 line/mm grating.
`NOTE: The yellow line is actually a doublet with
`(cid:5)
`wavelengths of 578 and 580nm.
`
`To Operate the Light Source:
`Simply plug it into a standard, grounded, 115 VAC (or
`220VAC) outlet; then flip the switch on the front of the
`light source to ON.
`
`CAUTION: The outer glass tube of the mercury
`(cid:5)
`vapor lamp blocks harmful ultraviolet radiation
`produced by the lamp. If the outer tube is cracked
`or broken, this radiation can cause severe skin
`burn and produce eye inflammation. Regularly
`inspect the outer tube for cracks, especially if the
`light source has received a significant jolt. If the
`glass bulb is broken, immediately turn lamp off and
`remove it to avoid possible injury. Replace Bulb
`prior to next use.
`
`CAUTION
`RISK OF ELECTRIC SHOCK
`DO NOT OPEN
`
`CAUTION:
`TO PREVENT THE RISK OF ELECTRIC SHOCK, DO
`NOT REMOVE COVER ON UNIT. NO USER SERVICE-
`ABLE PARTS INSIDE. REFER SERVICING TO QUALI-
`FIED SERVICE PERSONNEL.
`
`The lightning flash with arrowhead, within
`an equilateral triangle, is intended to alert
`the user of the presence of uninsulated
`“dangerous voltage” within the product’s
`enclosure that may be of sufficient magni-
`tude to constitute a risk of electric shock to
`persons.
`
`The exclamation point within an equi-
`lateral triangle is intended to alert the
`user of the presence of important oper-
`ating and maintenance (servicing) in-
`structions in the literature accompany-
`ing the appliance.
`
`Maintenance
`The light source lamp can be expected to provide
`24,000+ hours of trouble free operation. However, if the
`light at any time fails to come on, first check the fuse on
`the front of the light source case. If the fuse is not
`blown, you may need to replace the lamp. In rare cases,
`the problem may be with the ballast. However, ballast
`replacement should be performed only by experienced
`personnel. If the fuse is intact and replacing the bulb
`does not restore operation, we recommend you return the
`unit to PASCO scientific for repair.
`
`CAUTION—HIGH VOLTAGE: Do not open the
`(cid:5)
`light source with the unit plugged in.
`
`2
`
`¨
`
`
`
`012-02262H
`
`Mercury Vapor Light Source
`
`Other PASCO Spectral Equipment
`PASCO scientific offers a variety of student spectral
`light sources, and a precision student spectrometer for
`accurate spectral measurements in the student lab. For
`more information, check the current PASCO catalog,
`or call toll-free at 1-800-772-8700. Outside the United
`States call 1-916-786-3800.
`
`To Replace the Mercury Vapor Lamp:
`Unplug the light source. Allow 10 to 15 minutes for
`the lamp to cool.
`Unscrew the eight screws that attach the top plate of
`the light source and remove the top plate.
`Using a glove or several folds of thick cloth to protect
`your hand, unscrew the lamp. Discard it as you would
`any broken glass.
`Replace the lamp and the top plate of the light source.
`(cid:4)
`Replacement Parts
`Only the following replacement parts should be used:
`
`(cid:2)(cid:1)
`
`(cid:3)
`
`Description
`OS-9286A
`Hg Lamp GE HR100A38
`Hg Lamp Ballast
`Fuse (5x20) 4 amp, slo-blo
`
`OS-9286A-220
`Hg Lamp GE HR100A38
`Hg Lamp Ballast
`Fuse (5x20) 2 amp, slo-blo
`
`Pasco #
`
`526-018
`322-019
`530-035
`
`526-018
`322-028
`530-036
`
`OS-9286A
`
`F1
`
`S1
`
`4 A
`
`115 VAC
`
`T1
`
`OS-9286A-220
`
`220 VAC
`
`¨
`
`N/C (240 V TAP)
`F1
`S1
`2 A
`
`T1
`
`Schematic
`
`3
`
`DS1
`100 W
`HG
`
`DS1
`100 W
`HG
`
`
`
`Mercury Vapor Light Source
`
`012-02262H
`
`Warranty and Equipment Return
`
`Limited Warranty
`PASCO scientific warrants this product to be free from
`defects in materials and workmanship for a period of one
`year from the date of shipment to the customer. PASCO
`will repair or replace, at its option, any part of the product
`which is deemed to be defective in material or workman-
`ship. This warranty does not cover damage to the product
`caused by abuse or improper use. Determination of
`whether a product failure is the result of a manufacturing
`defect or improper use by the customer shall be made
`solely by PASCO scientific. Responsibility for the return
`of equipment for warranty repair belongs to the customer.
`Equipment must be properly packed to prevent damage
`and shipped postage or freight prepaid. (Damage caused
`by improper packing of the equipment for return ship-
`ment will not be covered by the warranty.) Shipping
`costs for returning the equipment, after repair, will be
`paid by PASCO scientific.
`
`Equipment Return
`Should this product have to be returned to PASCO scien-
`tific, for whatever reason, notify PASCO scientific by
`letter or phone BEFORE returning the product. Upon
`notification, the return authorization and shipping instruc-
`tions will be promptly issued.
`
`NOTE:
`(cid:5)
`NO EQUIPMENT WILL BE ACCEPTED FOR
`RETURN WITHOUT AN AUTHORIZATION.
`
`When returning equipment for repair, the units must be
`packed properly. Carriers will not accept responsibility
`for damage caused by improper packing. To be certain
`the unit will not be damaged in shipment, observe the fol-
`lowing rules:
`(cid:1) The carton must be strong enough for the item
`shipped.
`(cid:2) Make certain there is at least two inches of packing
`material between any point on the apparatus and the
`inside walls of the carton.
`(cid:3) Make certain that the packing material can not shift
`in the box, or become compressed, thus letting the
`instrument come in contact with the edge of the box.
`
`Address:
`
`Phone:
`FAX:
`
`PASCO scientific
`10101 Foothills Blvd.
`P.O. Box 619011
`Roseville, CA 95678-9011
`(916) 786-3800
`(916) 786-8905
`
`4
`
`¨