throbber
Liquid
`Chromatography
`for the Analyst
`
`Raymond E W. Scott
`Ge rgetown Umve
`y
`
`Marcel Dekker, Inc.
`
`New York-Basel-Hong Kong
`
`FRESENIUS KABI 1024-0001
`
`

`
`Library of Congress Cata1ogirsg—in~Publication Data
`
`Scott, Raymond P. W. (Raymond Peter William)
`Liquid chromatography for the analyst / Raymond P. W. Scott.
`p.
`cm. —— (Chromatographic science series; V. 67)
`Includes bibliograpiical references and index.
`ISBN (L824?-9184-3 (alk. paper)
`1. Liquid chromatograpliy.
`1. Series: Chromatograpliie science; v. 67.
`QD79.C454S37
`199—-
`/,
`5
`
`543 ’ .(}894~—dc20
`‘-‘iii 1?,
`
`93-43899
`CIP
`
`ISBN: U—8247~9184—3
`
`The publisher offers discounts on this book when ordered in bulk quantities.
`For more information, write to Speciai Sales/Professional Marketing at the
`address below.
`
`This book is printed on acid—free paper.
`
`Copyright © 1994 by MARCEL I)EKKER, WC. All R'ghts Reserved.
`
`Neither this book nor any part may be reproduced or transmitted in any form
`or by any means, electronic or mechanical,
`including photocopying, micro»
`filming, and recording, or by any information storage and retrieval system,
`without permission in writing from the publisher.
`
`MARCEL DEKKER, INC.
`2?0 Madison Avenue, New York, New York 10016
`
`Current printing (last digit):
`10 9 8 7 6 5 4 3 2 1
`
`PRINTED IN THE UNITED STATES OF AMERICA
`
`.....
`FRESENIUS KABI 1024-OOO2
`
`
`
`

`
`Contents
`
`Preface
`Acknowledgments
`
`1. An Introduction to Chromatography
`
`A Short History of LC
`The Separation Process
`The Different Forms of Chromatography
`Chromatography Nomenclature
`References
`
`2. Resolution, Retention and Selectivity
`
`The Plate Theory
`The Retention Volume of a Solute
`Factors that Control the Distribution Coefficient of a Solute
`Molecular Interactions
`The Thermodynamic Explanation of Retention
`Factors that Control the Availability of the
`Stationary Phase
`
`iii
`v
`
`1
`
`2
`4
`7
`9
`13
`
`15
`
`17
`21
`23
`23
`29
`
`33
`
`vii
`
`FRESENIUS KABI 1024-0003
`
`

`
`viii
`
`Chromatographic Methods of identification
`The Capacity Ratio of a Solute
`The Separation Ratio
`Column Efficiency
`References
`
`Liquid Chromatography Phase Systems
`
`The Production of Silica Gel
`Solute/Solvent Interactions on the Surface of Silica Gel
`
`Mono—Layer Adsorption of Solvents on the Surface
`of Silica Gel
`Solute Interactions with the Silica Gel Surface
`
`(Mobile Phase n—Heptane/Chloroform)
`Bi~Layer Adsorption of Soivents on the Surface of
`Silica Gel
`Solute Interactions. with the Silica Gel Surface
`
`(Mobile Phase n~HeptanefEthyl Acetate)
`Silica Gel as an Exclusion Medium
`
`Application Areas for Silica Gel as a Stationary Phase
`Silica Gel as a Stationary Phase in Elution
`Chromatography
`Silica Gel as a Stationary Phase in Exclusion
`Chromatography
`B011 ed Phases
`The Different Classes of Bonded Phases
`"Reverse Phases
`
`The Interaction of Reverse Phases with Solvents
`and Solutes
`
`‘3‘aired Ion Reagents
`The Use of Bonded Phases in Elution Chromatography
`Aqueous Solvent Mixtures
`The Use of Macro~Poroos holymers as Stationary Phases
`Polymer Stationary Phases in Elution Chroma-
`tography
`General Advice for the Unknown Sample
`References
`
`40
`4 l
`42
`44
`48
`
`51
`
`55
`S8
`
`58
`
`60
`
`63
`
`65
`67
`69
`
`69
`
`70
`71
`73
`76
`
`7'?
`79
`81
`82
`84
`
`90
`91
`91
`
`FRESENIUS KABI 1024-OOO4
`
`

`
`The Liquid Chromatography Column
`
`The Rate Theory
`The Summation of Variances
`The Maximum Sample Volume
`The Van Deemter Equation
`The Significance of the HETP Equation
`The Reduced Chromatogram
`Resolution
`Calculating the Efficiency Required to Achieve a
`Specific Resolution
`Peak Asymmetry
`Column Selection
`Column Diameter
`Preparative Columns
`Volume Overload
`References
`
`The Liquid Chromatograph
`
`The Basic Liquid Chromatograph
`The Mobile Phase Supply System
`The Gradient Programmer
`The LC Pump
`The Sample Valve
`The Column and Column Oven
`Detectors
`Data Acquisition and Processing
`References
`
`Liquid Chromatography Detectors
`
`Detector Specifications
`Detector Linearity and Response Index (oz)
`Linear Dynamic Range
`Detector Noise Level
`
`ix
`
`93
`
`94
`94
`95
`97
`105
`106
`108
`
`109
`1 11
`114
`116
`117
`118
`121
`
`123
`
`124
`124
`125
`128
`138
`144
`149
`152
`155
`
`157
`
`158
`158
`161
`162
`
`FRESENIUS KABI 1024-0005
`
`

`
`
`
`Measurement of Detector Noise
`Detector Sensitivity or the Miniinurn Detectable
`Concentration
`
`Pressure Sensitivity
`Flow Sensitivity
`Temperature Sensitivity
`The UV Detector
`
`The Fixed Wavelength Detector
`The Multi—WaVeiength Detector
`The Electrical Conductivity Detector
`The Fluorescence Detector
`The Refractive Index Detector
`
`The Tridet Multifunctional Detector
`References
`
`Sample Preparation
`
`Sample Preparation Techniques
`Filtration Techniques
`Centrifugation Techniques
`Concentration and Extraction Techniques
`An Automatic Sample Extraction and Concen-
`tration Procedure
`
`A Sample Separation Protocol
`LC Sample Preparation for Solid Materials
`Materials of interest Present in High Concentration
`Materials of Interest Present in Low Concentration
`The Preparation of Liquid Samples for LC Anaiysis
`Materials of Interest Present in High Concentration
`Materials of Interest Present in Low Concentration
`The Preparation of Liquid/Solid Samples for LC Analysis
`Materials of Interest Present in High Concentration
`Materials Present in Samples at Low Concentration
`General Comments on Choice of Mobile Phase
`Derivatization Techniques
`Pre—Column Derivatization
`Post~Coiuinn Derivatization
`References
`
`63
`
`64
`l65
`65
`65
`167'
`
`76
`180
`184
`189
`193
`
`I95
`
`195
`
`196
`197
`198
`
`205
`210
`212
`213
`217
`221
`221
`225
`228
`229
`232
`235
`237
`238
`
`245
`247
`
`FRESENIUS KABI 1024-OOOH6
`
`

`
`8. Qualitative and Quantitative Analysis
`
`Qualitative Analysis
`The Effect of Temperature on Retention
`Volume Measurement
`
`The Effect of Solvent Composition on Retention
`Volume Measurement
`
`Quantitative Analysis
`Peak Height Measurements
`Peak Area Measurements
`
`Procedures for Quantitative Analysis
`Quantitative Analysis Using Reference Standards
`The Relative Precision of Peak Height and Peak
`Area Measurements
`Peak De—Convolution
`
`Reporting Analytical Results
`References
`
`9. LC Applications
`
`Separations Based on Exclusion Chromatography
`Exclusion Chromatography Employing Silica Gel
`Exclusion Chromatography Employing Micro-
`Reticulated Cross—Linked Polystyrene Gels
`Chiral Separations
`Interactive LC Systems
`Dispersive Interaction Chromatography
`Polar Interaction Chromatography
`Ionic Interaction Chromatography
`Mixed Interaction Chromatography
`Summary
`
`Index
`
`X1
`
`251
`
`252
`
`260
`
`262
`
`265
`265
`266
`
`267
`267
`
`272
`273
`
`277
`279
`
`281
`
`282
`283
`
`286
`290
`296
`297
`304
`309
`314
`319
`
`32 J
`
`FRESENIUS KABI 1024-0007
`
`

`
`
`
`8 Q
`
`ualitative and Quantitative Analysis
`
`The object of an LC analysis is to establish the probable identity, and
`determine the precise amount, of each ofthe pertinent components
`present in the sample. The pertinent components may include all the
`substances present in the mixture or only those of specific interest.
`The identity of the peak is determined from its position on the
`chromatogram, that is, the time required for it to be eluted, whereas,
`the quantity ofa component present is determined from the peak
`height or peak area. It must be emphasized that a single LC analysis on
`a hitherto unknown sample can not unambiguously confirm the
`presence of a particular compound on the basis of retention data alone.
`Retention data, whether it is corrected retention volume, capacity
`factor or the separation ratio of the solute to that of a standard, can
`only indicate the probability of substance identity. Retention data from
`a second analysis, using a different phase system, increases the
`confidence level but absolute verification requires confirmation by
`another analytical technique. This might include infrared spectrometry,
`mass spectrometry or nuclear magnetic resonance spectroscopy. Such
`evidence would be essential for litigation purposes.
`
`There are in—line LC/spectroscopic systems available, but in most
`cases it is easier to carry out a semi—preparative separation, collect the
`material and carry out the spectroscopic examination off—line.
`However, for routine quality control analyses, where the sample
`
`251
`
`FRESENIUS KABI 1024-0008
`
`

`
`252
`
`clzaracteristics are already well established, retention data can be
`
`quite sufficient for identification purpose.
`
`Qualitative Analysis
`
`In LC both quantitative and qualitative accuracy depends heavily on
`the components ofthe sample being adequately resolved from one
`another. The subject of resolution has already been discussed, but it is
`necessary to consider those areas where uncertainty can still arise.
`Unfortunately, unless the analyst is aware of the pitfalls and how to
`deal with them, false assumptions of resolution can be made very
`easily.
`
`Consider the liquid chromatography peak shown in figure l.
`
`Figure 1
`
`A Single LC Peak
`
`1.25
`
`1
`
`E
`g
`I:
`
`8 0.75
`fl
`
`OD g
`
`GW
`
`0.5 _
`
`0.25
`
`0
`
`'
`
`
`
`3500 Plates
`
`2500 Plates
`
`4.8
`
`5
`
`5.2
`
`5.4
`
`.
`
`Volume Flow of Mobile Phase (ml)
`
`It is seen that the peak shown in figure 1 is asymmetric, which is
`typical of many peaks eluted from LC columns. This particular peak is
`not grossly asymmetric, the front half of the peak having an efficiency
`equivalent to 3500 theoretical plates and the latter half an efficiency Of
`
`FRESENIUS KABI 1024-0009
`
`

`
`253
`
`2500 theoretical plates. This situation frequently occurs in LC and can
`be caused by a number of different effects. The two major causes of
`peak asymmetry have already been touched upon and can arise from a
`difference in peak dispersion occurring in the front half ofthe peak
`relative to the rear half or to the solute distribution coefficient being
`different for the two halves of the peak.
`
`Re-iterating the HETP equation (10) given on page 104,
`
`H=2?tdp+
`
`u
`
`27D
`fi(k)d
`f(k')d2
`um + ——E—3u+ —2——5S——fu
`In
`
`2
`
`(1)
`
`Where the symbols have the meanings previously defined.
`
`Now, it was also shown on page 145 that for a given column, solute,
`mobile phase and flow rate, equation (1) can be reduced to an alternative
`abbreviated form which is given as follows,
`
`(2)
`
`On page 6, it was shown that in the front half of the peak, there will
`be a net transfer of solute from the mobile phase to the stationary
`phase and thus the resistance to mass transfer in the mobile phase will
`dominate. At the rear half of the peak there is a net transfer of solute
`from the stationary phase to the mobile phase and in this case the
`resistance to mass transfer in the stationary phase will dominate. Then
`if the resistance to mass transfer in the stationary phase is greater than
`that for the mobile phase, the rear part of the peak will be broader
`than the front half. In which case,
`
`2
`cldp < czdg
`Dm
`DS
`
`(3)
`
`FRESENIUS KABI 1024-0010
`
`

`
`254
`
`Consequently, the peak will exhibit the asymmetry shown in figure 1.
`It is seen that the relative values of the resistance to mass transfer
`2
`2
`
`terms is controlled by the functions, Fig and
`
`Dm
`
`DS
`
`It would seem that, in practice, the inequality defined in (3) can
`frequently occur but the converse does not appear to be true. Thus,
`peak asymmetry (in part or whole) resulting from inequality in mass
`transfer between the two phases manifests itself in the form shown in
`
`figure 1.
`
`Alternatively, peak asymmetry could arise from thermal effects.
`During the passage of a solute along the column the heats of
`adsorption and desorption that are evolved and adsorbed as the solute
`distributes itself between the phases. At the front of the peak, where
`the solute is being continually adsorbed, the heat of adsorption will be
`evolved and thus the front of the peak will be at a temperature above
`
`its surroundings. Conversely, at the rear of the peak, where there will
`be a net desorption of solute, heat will be adsorbed and the
`temperature or the rear of the peak will fall below its surroundings.
`
`Figure 2
`
`Temperature Profile of a Peak Passing Through a Heat of
`Adsorption Detector
`
`Temperature above that of
`the surroundings
`
`A
`
`Temperature
`
`Temperature below that of
`the surroundings
`
`..?___._?__.>
`Time
`
`FRESENIUS KABI 1024-0011
`
`

`
`255
`
`In fact, this phenomenon has been used as the basis of a very sensitive
`detecting system. An example of the temperature profile of an
`adsorbent as a peak passes over it is shown in figure 2. Unfortunately,
`it was found almost impossible to produce a true simulation of the
`concentration profile of the peak from the temperature profile and
`interest in the detector declined.
`
`Nevertheless, it can be seen from figure 2 that the evolution and
`adsorption of heat will cause the front of the peak to travel through
`the column at an elevated temperature relative to the rear of the peak.
`Furthermore as the heat evolved will be proportional to the mass
`adsorbed, the largest temperature changes will occur at the highest
`concentrations in the peak. As a result, the distribution coefficient of
`the solute with respect to the stationary phase will be slightly smaller
`at the front of the peak and at the high concentrations, i.e. the peak
`maximum. Now, it has already been established that the velocity of a
`solute band through a column is inversely proportional to its
`distribution coefficient. Hence, the high concentrations of the peak-
`front will move slightly faster through the column than the lower
`concentrations of the peak. This will cause the peak maximum to move
`towards the front of the peak and thus produce peak asymmetry. The
`effect of the heat of adsorption on peak shape has been treated
`quantitatively for both LC (1) and GC (2).
`
`In practice, it is probable that both of the effects discussed contribute
`to the overall peak asymmetry. Unfortunately, peak asymmetry varies
`in extent from the very obvious to the barely noticeable and because
`of this, peak asymmetry is often dismissed as the normal shape ofa
`single solute peak. Such an assumption can cause serious errors in both
`qualitative and quantitative analysis.
`
`Under circumstances where two solutes are incompletely resolved and
`one of the pair is present at a much lower concentration than the
`other, the profile of the pair often resembles a normal peak with slight
`asymmetry. Consider the combined elution profile of the two peaks
`shown in figure 3.
`
`FRESENIUS KABI 1024-0012
`
`

`
`256
`
`Figure 3
`
`The Combined Elution Profile of Two Unresolved Peaks
`
`1.25
`
`0.75
`
`3 (II Solute
`Concentration
`
`I
`
`.° I») U‘!
`
`0
`

`
`i
`
`4.7 4.8 4.9
`
`5
`
`5.1 5.2 5.3 5.4 5.5
`
`Volume of Mobile Phase (ml)
`
`It is seen that the peak profile shown in figure 1, resulting from an
`asymmetric single peak, is indistinguishable from the asymmetric peak
`shown in figure 3 that results from a 10% impurity eluting relatively
`close to the parent peak. Care must be taken not to interpret peak
`asymmetry as the normal elution profile when, in fact,
`it
`is an
`indication of incomplete resolution. If there is any uncertainty, the
`splvent should be changed and the peak in question carefully
`monitored. It does not matter if, in changing the solvent mixture, the
`resolution of other components is lost, as only the integrity of the peak
`in question is being examined. The composition of the mobile phase
`should be changed at least twice to be sure that only a single solute is
`present. If no second component can be detected from the peak
`profiles, then the original mobile phase can be used for the analysis
`and the integrity of the peak assumed.
`
`FRESENIUS KABI 1024-0013
`
`

`
`257
`
`This an excellent example of the value of the diode array detector. If
`the chromatogram shown in figure 3 was monitored at two different
`wavelengths, then a peak ratio curve would immediately disclose the
`presence of the second peak (see page 175) and it would no longer be
`necessary to resort to changes in mobile phase composition to establish
`the presence of the impurity.
`
`A second component, even at low concentration, will not only give an
`erroneous value for the amount of solute present but, on examining
`figure 3 it is seen that the retention time ofthe major peak is also
`significantly changed. This effect can be used for quantitative
`estimation of the unresolved mixture providing the chromatographic
`sytem can provide the necessary high precision.
`
`Scott and Reese (3) measured the accuracy and precision that could be
`obtained from a well designed liquid chromatograph and their work
`will be referred to again later. In the course of their work they found
`that, providing the retention time of a mixed peak could be measured
`with sufficient accuracy and the retention times of the two pure
`components were known, then the proportion of the two components
`could be determined, even if only a single peak was apparent. The
`theory behind this work is a little complex (4), and it is not
`appropriate to discuss the details here. The basis of the method was to
`calculate the theoretical retention time from the combined elution
`curves for the two solutes over a range of mixtures. The curve
`relating retention time to composition was then used as a calibration
`curve to determine the composition of an unknown mixture from the
`retention time of the combined peak. The calibration curve that was
`calculated and the experimental points they obtained are shown in
`figure 4.
`
`It is seen that the solutes used were nitrobenzene and deutero-
`nitrobenzene. These two solutes are so similar in physical and
`chemical characteristics that neither can be exclusively detected in a
`solvent without interference by the other. It is also seen that the
`retention times of the two solutes are 8.927 minutes and 9.061 minutes
`giving a difference of only 8.04 sec.
`
`FRESENIUS KABI 1024-0014
`
`

`
`_
`
`258
`
`Figure 4
`
`Graph of Retention Time Difference against Sample
`Composition for Mixtures of Nitrobenzene and Deutero-
`Nitrobenzene
`
`Deutero-
`Nitro-
`Benzene
`
`9.061m.
`
`
`
`.\' U1
`
`
`
`RetentionTimeDiff(secs) NV‘‘U10
`
`8.927
`
`0
`
`20
`
`
`60
`40
`0
`80
`100
`% w/w Deutero—Nitrobenzene
`
`It follows that measurements must be made with a precision of about
`0.2 second if quantitative results are to be of any value. It is seen from
`figure 4 that the experimental points lie very close to the line and a
`fairly accurate measurement of the distribution of the two isotopes can
`' be obtained from retention time measurements. This method has very
`limited areas of application and is given here, more to demonstrate the
`effect of unresolved impurities on retention time, than to suggest it as
`an alternative to adequate chromatographic resolution. In some cases,
`however, particularly in the analysis of isotopes, it may be the only
`practical way to obtain a quantitative evaluation of the mixture by a
`liquid chromatographic method.
`
`FRESENIUS KABI 1024-0015
`
`

`
`Finally, it must be emphasized that the uncertainty arising from the
`slight asymmetry shown by the presence of a 10% impurity in figure
`3 does not depend on the magnitude of the impurity. In figure 5 the
`elution curve resulting from a 50% mixture oftwo closely eluting
`components is shown.
`
`259
`
`Figure 5
`
`The Elution Curve ofa Binary Mixture Containing Equal
`Quantities of Each Solute
`
`1.5
`
`Concentration P (II
`Solute
`
`0
`
`4.7 4.8 4.9
`
`5
`
`t
`5.1 5.2 5.3 5.4 5.5
`
`Volume Flow of Mobile Phase (ml)
`
`It is seen that the profile of the combined peaks is perfectly
`symmetrical and displays no hint that there are two solutes present.
`Obviously an absorption ratio curve from a diode array detector
`would quickly disclose the presence of the two components, as would
`an appropriate changes in mobile phase composition. However, there
`would be a further clue for the analyst to follow that would give
`warning of the "duplicity" ofthe peak. The double peak would be
`very broad and be inconsistent with the change in peak width of the
`other solute peaks with retention time. The peak width of a solute
`increases regularly with retention time but, unfortunately, the
`relationship is not smooth. There are good reasons for this, but they
`
`FRESENIUS KABI 1024-0016
`
`

`
`260
`
`are not pertinent to this discussion. However, a peak such as that
`shown in figure 5 would be excessively broad and obviously out of
`order with the peak widths of neighboring solutes. This should alert
`the analyst to the possibility of an unresolved pair of solutes. The
`analyst should always be alert to the width of each peak in relationship
`to its position on the chromatogram, as this is the first indication of
`the presence of a composite peak.
`
`The Effect of Temperature on Retention Volume Measurement
`
`There are two major factors that influence retention volume
`
`measurement and they are temperature and solvent composition. In
`order to measure retention volume with adequate precision it is
`necessary to know the relationship between retention time and
`temperature so that the control limits of the column temperature can
`be specified.
`
`The effect of temperature on retention time was investigated by Scott
`and Reese (3), who measured the retention volume of the solutes o-
`dinitro-benzene, 2—ethoxy naphthalene and p-chlorophenatole over a
`range of temperatures. The chromatographic conditions used are as
`follows,
`
`Chromatographic Conditions
`
`Column
`
`Silerex 1
`
`Column Length
`Column Diameter
`
`25 cm
`4.6 mm
`
`Column Packing
`l\/Iobile Phase
`Flow Rate
`
`Silica Gel (particle size 10pm)
`44% butyl chloride and 56% n—heptane
`1 ml/min.
`
`Detector
`Sample Volume
`
`UV adsorption at 254 nm
`1 ul
`
`Scott and Reese chose to monitor retention volume as opposed to
`retention time, as retention volume is always the primary dependent
`variable in LC. Retention time is not a primary measurement because
`it must also include the reproducibility of the flow—rate delivered by
`
`FRESENIUS KABI 1024-0017
`
`

`
`261
`
`the pump. The retention volume was measured absolutely by means of
`a burette connected to the end of the column. The column was situated
`in a water bath that was controlled to +/- 0.050C. Precautions were taken
`to ensure the mobile phase attained the column temperature before
`entering the column. The results obtained are shown in figure 6A.
`
`Figure 6A
`
`Graphs of Retention Volume against Temperature for Three
`Solutes
`
`0-Dinitrobenzene Mean Retention Volume 17.19 ml
`Slope -0.262 ml/C 1.53%/C
`
`.$__\\\\\
`
`2-Ethoxy Naphthalene Mean Retention Volume 4.93 ml
`Slope -0.070 ml/C 1.44%/C
`
`
`
`p-chlorophenatol Mean Retention Volume 3.077 ml
`Slope -0.044 ml/C 1.42%/C
`
`_____fi“‘____*—“;7_‘-__
`
`t\) G
`
`».—- (J1
`
`anS
`
`
`
`
`
`RetentionVolume(ml)
`
`20
`
`26
`24
`22
`Temperature (C)
`
`28
`
`It is seen that in order to measure retention volumes with a precision
`of 0.1%, the temperature control must be +/- 0.0400 This level of
`temperature control on a thermostat bath is not difficult to achieve but
`it is extremely difficult, if not impossible, to return to a specific
`temperature to within +/- 0.040C after prior change. To achieve a
`preoision of retention volume measurement of 1%, the temperature
`control must be +/- 0.400 This is far more practical as most column
`oven temperature can be set to a given temperature to within +/-
`0.250C. Although the data was obtained for three specific solutes, the
`results can be taken as reasonably representative for all solutes and
`phase systems. In most practical analyses, the precision limits of
`retention volume measurement will be about 1% but this will not
`include the reproducibility of the flow rate given by the pump. As
`
`FRESENIUS KABI 1024-0018
`
`

`
`262
`
`retention times are usually measured in analytical LC, the precision of
`
`measurement may be significantly greater than +/—1%. It follows that
`
`ifthe identity ofthe peak must be confirmed, and retention data is
`
`being used for the purpose, then it is essential that the column is
`
`carefully thermostatted.
`
`The Effect of Solvent Com_p_osition on Retention Volume Measurement
`
`Scott and Reese (3) also measured the change in retention volume with
`
`solvent composition using the same LC apparatus as that used for
`
`investigating the effect of temperature. The column was thermostatted
`
`at 247°C and the results that were obtained are shown in figure 6B.
`
`Figure 6B
`
`Graphs of Retention Volume against Solvent Composition
`for Three Solutes
`
`o—Dinitrobenzene Mean 17.57 ml
`Slope 5.27% per unit change in concentration
`
`9 —&
`E\i
`
`
`
`I5
`
`to
`
`u
`
`E Ec > =
`
`1
`.2
`‘E
`3

`
`0
`
`2—Ethoxy Naphthalene Mean 5.01 ml
`Slope 5.02% per unit change in concentration
`5
`
`—0-—Z0-:-—0:2——o——:-O-
`
`p-chlorophenatole Mean 2.857 ml
`Slope 3.71% per unit change in concentration
`
`43
`
`if
`
`44
`
`45
`
`Solvent Composition % w/v Butyl Chloride
`
`FRESENIUS KABI 1024-0019
`
`

`
`263
`
`It is seen that small changes in solute composition can have a profound
`effect on the retention volume, but it must be borne in mind that the
`magnitude of the effect will vary somewhat, between different solutes
`and different phase systems. The results shown in figure 6B indicate
`that if a retention volume is to be measured with a precision of 0.1%,
`the solvent composition must be maintained constant to within 0.02%
`w/v. Normally, the solvent composition can be kept constant within
`those limits providing a closed solvent system is used to prevent
`evaporation. However, it would be extremely difficult, if not impossible
`to make up another solvent mixture of the same composition within
`those limits of precision. This would be particularly difficult if either
`of the solvents were volatile. If the precision required for retention
`volume measurement was 1%, the solvent composition would have to
`be maintained constant with a precision of +/- 0.2 % w/v which should
`be quite practical. Furthermore, it would be fairly straightforward to
`make up replacement solvents to the same concentration within the
`same precision limits. For a routine analysis, however, it might well
`be advantageous to prepare the mobile phase in large volumes and
`store it in an appropriate manner. Scott and Reese (3) examined the
`repeatability of retention time measurements by two procedures, one
`employing computer processed data and the other by manual
`measurement of the distance in centimeters from the injection point to
`the peak maximum on a potentiometric recorder chart. They carried
`out twelve replicate analyses and the resulting statistical analysis is
`shown in table 1.
`
`Table 1
`
`The Precision of Retention Data Measurement Made
`Manually and by Computer
`
`Parameter
`Capacity factor
`
`Peak 1
`0.22
`
`Mean (cm)
`S.D.(% Mean)
`Mean (min.)
`S.D.(% Mean)
`
`10.17
`0.85
`3.97
`0.31
`
`Peak 2
`0.94
`
`16.0
`0.245
`6.27
`0.20
`
`Peak 3
`1.50
`
`20.59
`0.19
`8.11
`0.17
`
`Peak 4
`5.21
`
`51.14
`0.15
`20.14
`0.33
`
`FRESENIUS KABI 1024-0020
`
`

`
`Figure 7
`
`Tips of Peaks Having Extreme Retention Values
`Reconstructed by the Computer
`
`Peak Crests (between 99.9 and 100% of the peak heights)
`
`Retention Time Difference
`Recorded
`
`,<———4.4 sec. -1-»;1
`
`I l1l i
`
`<—2.1 sec.
`
`Retention Time Difference
`Actual
`
`By taking the mean positions of the peaks, it is seen that they are only
`2.1 sec apart. However, as a result ofa noise spike on the front of the
`
`FRESENIUS KABI 1024-0021
`
`

`
`265
`
`first peak and a similar noise spike on the back of the second peak, the
`computer interpreted each noise spike as the peak maxima, which
`resulted in a retention difference of 4.4 see. It is clear that a
`significant part of the scatter of retention data can be due to detector
`noise, either arising from the detector itself or from changes in
`operating conditions of the chromatograph. This once more emphasizes
`the importance of using a chromatograph that can maintain stable
`operating conditions in conjunction with a detector having as little
`long term noise as possible. The results also indicate that unless the
`maximum sensitivity is essential for the analysis, the sample size
`should be adjusted so that the detector can be operated well below its
`maximum sensitivity to reduce the consequence of the noise.
`
`Quantitative Analysis
`
`Quantitative estimates of the mass of a particular solute present in a
`sample are obtained from either peak height or peak area measurements.
`The values obtained are then compared with the peak height or area of
`a reference solute present in the sample at a known concentration or
`mass. In this chapter quantitative analysis by LC will be discussed but
`the procedures described should not be considered as entirely
`appropriate for other types of chromatographic analysis. Those
`interested in general quantitative chromatographic analysis including
`GC and TLC are referred to the book by Katz (4).
`
`Peak Height Measurements
`
`Most detectors are concentration sensitive devices and thus the peak
`height will be proportional to the maximum concentration in the peak,
`which, in turn, will be proportional to the total area of the peak. The
`total area of the peak is proportional to the total mass of solute
`contained in the peak providing it is not excessively tailing. As the
`peak height is inversely related to the peak width, then, if peak heights
`are to be used for analytical purposes, all parameters that can affect the
`peak width must be held constant. This means that the capacity factor
`of the solute (k') must remain constant and, consequently, the solvent
`
`FRESENIUS KABI 1024-0022
`
`

`
`266
`
`composition is held stable. The temperature must also be held constant
`and an highly repeatable method of sample injection must be used. If
`computer data acquisition and processing are employed, then a direct
`printout of the peak heights is obtained and, with most systems, the
`calculated analysis is also presented. Ifthe peak heights are to be
`measured manually, which even today is carried out in the majority of
`LC analyses, then the base line is produced beneath the peak and the
`height between the extended base line and the peak maximum
`measured. In general, the measurements should be made estimating to
`the nearest 0.1 mm.
`
`Peak Area Measurements
`
`The area of a peak is the integration of the peak height (concentration)
`with respect to time (volume flow of mobile phase) and thus is
`proportional to the total mass of solute eluted. Measurement of peak
`area accommodates peak asymmetry and even peak tailing without
`compromising the simple relationship between peak area and mass.
`Consequently, peak area measurements give more accurate results
`under conditions where the chromatography is not perfect and the
`peak profiles not truly Gaussian or Poisson.
`
`Unfortunately, neither the computer nor the potentiometric recorder
`measures the primary variable, volume of mobile phase, but does
`measure the secondary variable, time. This places stringent demands
`on the LC pump as the necessary accurate and proportional
`relationship between time and volume flow depends on a constant flow
`rate. Thus, peak area measurements should never be made unless a
`,good quality pump is used to control the mobile phase flow rate.
`Furthermore, the pump must be a constant flow pump and not a
`constant pressure pump.
`
`Peak areas can be measured manually in a number of ways, the
`simplest being the product ofthe peak height and the peak width at
`0.6065 of the peak height. This does not give the true peak area but
`providing the peak is Poisson, Gaussian or close to Gaussian it will
`
`FRESENIUS KABI 1024-0023
`
`

`
`267
`
`always give accurately the same proportion of the peak area. Other
`methods involve the use of a planimeter (an instrument that provides a
`numerical value for the area contained within a perimeter traced out
`by a stylus), and the measurement of the product of the peak height
`and the peak width at the base. The first method is extremely tedious
`and the second somewhat inaccurate. The most accurate manual
`method of measuring peak area is to cut the peak out and weigh it. A
`copy of the chromatogram should be taken and the peaks cut out of the
`copy. This procedure is also a little tedious (not as tedious as the use
`of a planimeter) but does provide very accurate values for the peak
`area. It is particularly effective for skewed or malformed peaks where
`other methods of manual peak area measurement (with the exception
`of the planimeter) fail dismally and give very inaccurate results. The
`recommended method is to use the product of the peak height and the
`peak width at 0.6065 of the peak height but this does require adequate
`resolution of the components of the mixture.
`
`Procedures forQuantitative Analysis
`
`There are two basic methods used in quantitative analysis; one uses a
`reference standard with which the peak areas (peak heights) of the
`other solutes in the sample are compared; the other is a normalization
`procedure where the area (height) of any one peak is expressed as a
`percentage ofthe total area (heights) of all the peaks. There are
`certain circumstances where each method is advantageous, and
`providing they are used carefully and appropriately all give
`approximately the same accuracy and precision.
`
`Qugtitative Analysis Using_Reference Standards
`
`Reference standards can be used in two ways; a weighed amount of the
`standard can be added directly to the sample and the area of the peaks
`of interest compared with that of the standard; alternatively, a
`weighed amount of the standard can be made up in a know

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket