throbber
Journal of Microscopy, Val. 1 7 0 , Pt 3 , June 1 9 9 3 , pp. 2 2 1 -228.
`Received 2 January 1 9 9 2 ; accepted 1 7 December 1 9 9 2
`
`Methods for large data volumes from confocal
`scanning laser microscopy of lung
`
`E. H. OLDMIXON* & K. C A R L S S O N t
`*Departments of Medicine, Brown University and Memorial Hospital of Rhode Island,
`Pawtucket. R1 0 2 8 6 0 . U.S.A.
`tPhysics IV, Royal institute of Technology. Stockholm S-10044, Sweden
`
`Key words. Confocal scanning laser microscopy, dog lung parenchyma, intensity
`correction, volume fusion, three-dimensional reconstruction, alveolar ductal unit.
`
`Summary
`
`Confocal scanning laser microscopy makes it possible to
`obtain series of optical sections in precise registration.
`Certain studies of lung parenchyma, however, require both
`the fine resolution obtainable with high-numerical-aperture
`(NA) objectives and the extensive fields of view that usually
`would be achieved only with low-NA objectives. This
`article presents a technique that resolves this conflict by
`using a sequence of operations: (i) to correct intensity
`variations on individual sections due to non-uniform
`illumination/detection characteristics of the microscope;
`(ii) to correct intensity variations between successive
`sections in a series due to, for example, depth-related
`absorption or step changes in detector sensitivity; (iii) to
`adjust adjacent, overlapping stacks of sections to a common
`intensity level; and (iv) to fuse a group of such overlapping
`stacks into a single series of larger sections. This resulting
`stack may contain, for example, a complete cross-section of
`an alveolar ductal unit about 500 pm or more in diameter
`at about 1-pm pixel resolution.
`
`Introduction
`
`This paper differs from much of the image-registration
`literature in addressing the problem of registering two or
`more images of the same scene obtained by different
`imaging techniques (e.g. fusion of multimodality images) or
`registration of images of the same subject obtained by one
`method but at different
`times and under different
`conditions, say of position and illumination. Here we
`explore a procedure for enlarging the volume from a
`biological sample which can be visualized using a confocal
`microscope.
`The method was developed to enable us to study certain
`architectural aspects of the dominant subunit of mamma-
`lian lung parenchyma, the alveolar ductal unit. These units
`are approximately cylindrical, with a relatively long,
`
`0 1993 The Royal Microscopical Society
`
`central lumen around which are placed alveoli opening
`onto the lumen. In fully inflated, canine lung parenchyma.
`the ductal unit diameter, measured at 90' to the lumen axis
`from a back wall of an alveolus on one side to a back wall
`on the other, is 500pm or more. One research goal is to
`measure quantitatively how the parenchymal surface.
`essential both in gas exchange and lung mechanical
`behaviour, is disposed within the alveolar ductal units as
`a function of radial distance from the lumen axis.
`Parenchymal tissue was stained generally with a
`fluorescent dye, while the volumes previously occupied by
`air or blood remained non-fluorescent. To locate and
`measure the parenchymal surface with sufficient accu-
`racy, it was necessary to record the fluorescence intensity in
`1 x 1 x 1-pm volume elements throughout a volume large
`enough to contain the complete cross-section of an alveolar
`ductal unit and over a length of the ductal unit as long as
`obtainable, preferably at least 200 pm. The required data
`volume would
`then have dimensions of
`at
`least
`500 x 500 x 200 elements, orthogonally arrayed. We
`chose to examine ductal units whose axes are normal to
`the surface of the sample block and thus also parallel to the
`depth coordinate; this is the dimension of the data volume
`equalling approximately 200 elements.
`Using a confocal scanning laser microscope (CSLM), it is
`possible to collect a square array of fluorescence intensities
`(e.g. 256 x 256, 512 x 512, 1024 x 1024) at a single
`depth, then to descend to a greater depth and collect
`another square array, aligned precisely with the preceding
`one. For reasons explained later, however, each such
`square array of volume elements is physically too small to
`record the fluorescence pattern across an entire alveolar
`ductal unit cross-section. In practice, therefore, the whole
`data volume has to be built up by recording from 4 to 9
`adjacent series of 200 sections each, each series recording
`the entire available depth but only a fraction of the cross-
`section of the alveolar ductal unit. The lateral displace-
`
`221
`
`

`
`222 E . H. OLDMIXON & K . CARLSSON
`
`Fig. 1. Composite picture of four 256 x 256-pixel images which.
`together, record the entire cross-section at this depth of an alveo-
`lar ductal unit in dog lung parenchyma. In the text, the upper left
`image is I, upper right is m, lower right is n and lower left is 0; the
`entire image area (512 x 512 pixels) shown here is p. Each quad-
`rant image's content duplicates some content in the neighbouring
`images; thus, 1 shares content with m and o etc. These images are
`raw data fr3m the CSLM, except that for visibility in the figure the
`intensities of dim pixels were increased. The images I, m, n, o and p
`all are square, in reality; the vertical elongation and slight shearing
`seen were introduced by the video printer.
`
`these section series are not as yet
`ments between
`controllable at pixel dimensions, but have uncertainties of
`tens of micrometres. To enable subsequent lateral regis-
`tration of the whole series, each series is made to contain
`some of the same tissue features as its neighbours (see
`Fig. 1).
`Four problems are encountered when progressing from
`the separate section series to the single data volume
`containing the desired ductal unit:
`(1) illumination/detection characteristics of the microscope
`are not identical at every pixel location within a field of
`view;
`
`(2) the same structure is usually not recorded at the same
`intensity from one section to the next, an effect due to any
`of several causes, such as increasing absorption or
`scattering of incident and emitted light with depth, (mild)
`bleaching, or changes (made between sections) in instru-
`mental illumination and/or detection sensitivity;
`(3) the section series have to be adjusted to a common
`intensity level;
`
`(4) the section series containing different portions of the
`desired ductal unit cross-section must be fused into a single
`series containing a complete cross-section of the subject.
`
`Methods
`
`Specimen preparation
`
`Figure 1 shows dog lung parenchymal tissue. The lobe was
`obtained by approved procedure, inflated to a transpul-
`monary pressure of 30 cm HzO, then chemically and
`mechanically
`fixed by perfusion with glutaraldehyde
`[2.5% in sodium phosphate buffer, 0.075 M , pH 7.4, 350
`mOsm, 3% dextran (average molecular weight 72,000 Da)],
`tannic acid (4% w/v in sodium cacodylate buffer, 0.2 M, pH
`7.4, 350 mOsm, 52.5 mM sodium chloride, 3% dextran) and
`ethanol (linear gradient from 10% in H20 to 100%. 500 ml
`total), sequentially. Samples were cut with snapped, double-
`edged razor blades, immersed at once in absolute ethanol
`and dehydrated with gentle flow under vacuum overnight.
`Samples were submerged with gentle swirling in Lucifer
`Yellow CH fluorescent stain (0.001% in absolute ethanol) in
`vials (kept dark) for 24 h, rinsed with absolute ethanol (two
`changes, 30 min each), then processed through propylene
`oxide to Spurr's epoxy, to be cured at 55°C for 36 h. Each
`block was milled or lathed flat on one face, cutting into
`embedded tissue, sectioned with a glass knife and glued by
`the corners of this face to a flat-faced, ring-shaped plastic
`adaptor. This new facet was machined parallel to the first,
`then sectioned with a glass knife. The finished sample was
`carefully released from the adaptor, touched up and
`attached with a piece of adhesive putty to a microscope
`slide by either sectioned facet, exposing a smooth, level
`tissue section. Tissue prepared by this method is believed to
`retain in viva size and shape (Oldmixon et al., 1985).
`
`Confocal microscopy
`
`A CSLM developed at the Royal Institute of Technology in
`Sweden (Carlsson et al., 1985: Carlsson & Liljeborg. 1989)
`produces digitized images in square format, with up to
`1024 x 1024 square pixels
`in orthogonal rows and
`columns. Series of images obtained successively deeper
`within a sample are precisely aligned. Routinely, resolution
`in the plane of the section is closely comparable to that of
`the same microscope used in conventional transmission
`mode. The images, acquired in digital form, are suited to
`digital
`image processing and analysis and may be
`reconstructed in three dimensions.
`A characteristic of this microscope that is especially
`significant for this project is its low geometric distortion
`within each section, maximally 1.7% (Carlsson, 1990). We
`have elected to view this distortion as insignificant and have
`not attempted to compensate for it.
`
`

`
`An argon laser was used for illumination at 473 nm.
`with typically 0.06-0.1 mW actually transmitted to the
`specimen.
`
`Apparent depth resolution and field of view
`
`The depth resolution of the CSLM is such that an image
`obtained with a 63 x oil-immersion, planapochromatic
`objective (NA 1.4) resembIes subjectively a physically
`microtomed section of the same sample about 1.3 pm
`thick; with a 4 0 x oil-immersion, planapochromatic
`is to a 2.0-2.5-pm-thick
`objective, the resemblance
`physical section. The respective objectives produce images
`which cover nominal 161 x 161- or 256 x 256-pm2 fields
`of view. A 25 x objective (NA 0.64) provided a larger field of
`view, but unfortunately the optical sections were subjec-
`tively equivalent to 7-10-pm-thick physical sections.
`
`Limits on depth of observation
`
`In theory at least, sections may be obtained from the surface
`of the specimen down to a depth equal to the working
`distance of
`the objective. In practice, the maximum
`productive working depth may be decreased drastically by
`an optically dense specimen, which may
`introduce
`attenuation and degradation of resolution. If the refractive
`index of the specimen, n, differs from that, nl, of the medium
`interposed between the objective and the specimen (air,
`immersion oil, etc.) then the actual depth, dl, from which
`the image is obtained will not be the apparent depth, d,
`within the sample measured by the mechanical vertical
`displacement of the stage relative to the position at which
`the optics focus on the top of the specimen, but will be given
`by (Carlsson, 1991)
`
`increased by
`The effective working distance may be
`observing the specimen without a coverslip, if the physical
`properties of the specimen permit; the imaging depth may
`thus be increased to about 2 70 pm with the 63 x objective
`and deeper with the 40x objective. With epoxy-embedded
`lung samples, 200pm is currently the limiting depth, in
`practice.
`
`Sizes of physical and data volumes
`
`Combining the preceding considerations, the physical
`volumes that can be sampled from lung are 1 6 1 x 1 6 1 x
`200pm when using the 63x objective and 256 x 256x
`200 pm using the 40x objective. Converting these physical
`volumes to data volumes, however, must be influenced by
`such factors as the requisite resolution, anticipated pro-
`cessing techniques, and available hardware and software.
`
`LARGE C S L M DATA VOLIIMES 22 3
`
`In the present application, in-plane pixel dimensions of
`1 x 1 pm would be satisfactory to retain connective tissue
`cable profiles as groups of four or more distinctively high-
`intensity pixels. The 40 x (1.2 5 x auxiliary lens) objective
`filled this requirement, giving 256 x 256 pixel images with
`pixel dimensions of 1.0 x 1.0 pm2.
`intensity correction
`Experience with section series
`algorithms (the second problem, above) has shown that
`vertical or depth spacing between sections should be less
`than half the apparent corresponding physical section
`thickness, here, 2,512 = 1.25pm or less. Quite con-
`veniently, this spacing was chosen as 1.0 pm, giving voxel
`dimensions of 1.0 x 1.0 x 1.0 pm.
`The physical view volume, therefore, could be captured in
`200 sections each of 256 x 256 pixels. Each voxel, when
`recorded, is a 1-byte (8-bit) number, so the data volume
`from one stack would occupy 13.1 Mb. During processing.
`voxels are stored as 2-byte data, bringing the size to
`26.2 Mb per stack. Anticipating having to fuse together 4.
`5, 6, or even 9 stacks to portray the ductal unit segment,
`even a single instance of the data volume could eventually
`occupy 2 3 6 Mb.
`
`Computing hardware and software
`
`The growing availability of fast processors, generous
`memory and graphical interfaces makes it almost futile to
`specify a minimal system, but the work described here has
`been done comfortably with the following: Sun Microsys-
`tems (Sunnyvale, CA) 3/150 six-slot VME chassis with
`16 MHz Motorola 68020 processor, 19" grey -scale monitor,
`4-Mb memory, 160-Mb SCSI disk, 1/4" cartridge tape (QIC-
`24, 57Mb capacity), 1.2-Gb CDC (Minneapolis. MN) disk
`with Interphase (Austin. TX) ESMD controller, Vicom
`Systems (Fremont, CA) VME image processing peripheral
`(six VME slots, 1 5 Vicabus slots) with point processor,
`ensemble processor, 4-Mb image memory, 1-Mb display
`memory, display subsystem with 19" RGB monitor, and Sky
`(Lowell, MA) Warrior vector processor.
`Microscope control and image acquisition/archiving
`visualization software accompanies the CSLM; the com-
`bination is available from Molecular Dynamics (Sunnyvale,
`CA) as a commercial product. Vicom Systems VIPs
`software, particularly
`the interactive command inter-
`preter, was used for examining pixel values on displayed
`images and for correcting
`image sequences by
`the
`compensating image, for example. The desired operations
`can be programmed and the Unix awk utility used to create
`a filed sequence of commands retrievable from within the
`command interpreter. While this approach does not operate
`the hardware at its maximum speed, the time lost is small
`and the rapidity and flexibility of programming are more
`than compensatory. A cycle of correction by the com-
`pensating image takes about 3 s per serial section.
`
`

`
`Generating the compensating image using the local thin-
`plate spline program (Franke. 1982) in Fortran requires
`over 1 h on our system. We have not attempted to vectorize
`the code.
`Intensity adjustment section by section through a stack
`makes full use of both Vicom and Sky. Computation of
`adjustment factors takes about 5 s per serial section, and
`subsequent intensity adjustment about 3 s. As most of the
`time is spent transferring images to and from disk, we
`suspect that the Unix filing system creates a bottleneck here.
`Translating and fusing four 256 x 256-pixel sections
`followed by writing out the 512 x 512 result, for entire
`sequences, is handled by the Vicom command interpreter
`reading an awk-generated command file.
`
`Choices of pixel and section spacing
`
`For a variety of reasons concerned with the amount of
`installed memory, the underlying architecture of processing
`boards and display systems, and associated software, the
`image size most convenient for processing and analysis on
`our system is 512 x 512 pixels. Fortunately, this size often
`suffices to hold a complete cross-sectional view of our
`subject, the ductal unit.
`A ductal unit is, roughly, a cylinder c. 500pm in
`diameter. A cross-section of a ductal unit centred in a
`5 12 x 5 12 pixel image with some of the surrounding tissue
`for context could have pixels somewhat less than or equal to
`1.Opm on a side. The connective tissue cables which we
`wish to detect within the ductal unit typically have
`minimum caliper diameters of several micrometres or
`more; they often fully occupy groups of four or more pixels
`and thus are readily detectable.
`As mentioned above, pixels 1.0 pm2 can be produced by
`using a 40x objective to scan 256 x 256 pixel images
`(which are c. 2 56 x 2 5 6 pm in size). Since four such images
`arranged as quadrants cover a square c. 512 pm on a side,
`it is feasible in many cases to move each image toward the
`centre of the composite field, so that each overlaps its
`neighbours, and still cover the cross-section of the ductal
`unit and the surrounding context. Such a group of four
`series is used for illustration in this article, but other ductal
`units may require six (2 x 3) or nine (3 x 3) series to cover
`the ductal unit cross-section.
`
`Approaches to solutions of problems encountered in
`fusing multiple section series into single data volume
`
`Compensating for non-uniform illumination/detection within
`a section
`
`CSLMs which scan by deflecting a beam angularly neither
`illuminate a field of view perfectly evenly nor detect light
`emitted from each pixel position with equal efficiency. We
`
`compensated for this systematic unevenness in each section
`before proceeding further.
`The combined illumination and detection behaviour of
`the system can be determined empirically by scanning
`certain glasses which fluoresce uniformly and brightly upon
`their first exposure to laser light (e.g. an orange filter, Schott
`No. OF530). Intensity gradients seen systematically on
`individual scanned images are then due to the illumination/
`detection characteristics of the entire system. We sup-
`pressed high-frequency photon noise and extracted the
`gradients by averaging together eight images each scanned
`from fresh regions of the glass, using the same objective.
`pixel spacing and pixel number as for the section images.
`The plane of focus was 5-15 pm below the glass surface.
`Since emission intensity from the fluorescent glass falls
`I'apidly with repeated scans, laser power and detector
`sensitivity (photomultiplier voltage) were tested at distant
`locations on the glass before scanning the final calibration
`images. Since the fields to be scanned should not be
`illuminated beforehand, the stage must be moved just prior
`to scanning; there is, therefore, a slight chance that a
`previously exposed region would be encountered. The
`scanned images were scrutinized for suspect, dim areas,
`usually straight-sided, angular and crossing the image
`border; such flawed images were discarded and not included
`in the averaging.
`The averaged image, however, still contained random.
`high-frequency variations. We inspected the pixel values
`(now in 16-bit representation) in 8 x 8-pixel blocks in
`about 50 regions across the entire image and recorded the
`characteristic pixel intensity of each region. The object was
`to specify benchmarks to pass to a surface interpolation
`program which would produce a standard image with first-
`order continuity, while preserving all the important features
`of the illumination/detection behaviour. Additional pixel
`values were specified at the corners and along the sides:
`here values were chosen by extrapolating visually smooth
`curves through groups of pixel values at interior locations
`in the neighbourhoods of the desired border pixels. The local
`thin-plate spline surface interpolation method of Franke
`(1982) produced a 256 x 256 array of pixel values which
`when displayed appeared to be a noise-free rendering of
`the averaged image. The final compensating image C (see
`Fig. 2) was generated:
`C(i, j) = m * k/S(i, j ) ,
`where i, j are indices over the ranges of pixel locations in
`image rows and columns, respectively, C is the compensat-
`ing image, S the interpolated intensity of a standard image,
`m is the maximum representable pixel intensity (here
`+ 0.9995) and k is the lowest value in image S.
`Each scanned section was multiplied by image C to
`produce a new, compensated image in signed 16-bit
`representation.
`
`

`
`L A R G E CSLM D A T A VOLUMES 2 2 5
`
`reports no important changes to the intensities. In practice,
`this algorithm determines the factor needed to change the
`intensity of a section to that of its predecessor, and the
`factor applied to any section is the cumulative product of
`the preceding factors. This might result in some sections
`being brightened undesirably to the point of saturation and
`may be avoided by reducing the brightness of the first
`section by 15-20% or by scaling the factors so that the
`maximal pixel intensity in the adjusted series will be below
`saturation. Both variations have been used successfully.
`It may be of interest that once the images have been
`loaded from disk the computing system calculates the factor
`relating image-pair intensities in about 1 s. Mathematically,
`the code computes the eigenvector, but does so using the
`Vicom image processor to define regions of interest, add
`images, multiply them, compute their fractional powers,
`etc., taking a 'parallel processing' approach. These routines
`may be requested from the first author.
`Note that this serial-section intensity-correction algorithm
`succeeds or fails depending on changes in image content, the
`characteristic dimensions of fluorescent structures and voxel
`dimensions, particularly the depth increment. In our
`experience, it tends to fail on typical lung parenchymal
`image series if (other settings remaining the same) the depth
`increment is increased above 2.0 pm. (A failure is signalled by
`the appearance of saturated regions which spread and fuse.
`often eventually to obliterate the images.)
`
`Calculating displacements needed to form a composite image
`
`Consider a series of four 256 x 256-pixel images (I, m, n, o)
`collected in the pattern shown in Fig. 1. The four
`256 x 256-pixel
`images, arranged
`to abut, form a
`512 x 512-pixel image, p. The origin of each image is
`considered conventionally to be its upper left corner
`expressed as (row = 0, column = 0). Thus, the origin of
`image I is (0, O),, coincident with the origin of image p. The
`origin of image m on image p, though, is (0, 256),, or n,
`(256, 256), and of o, (256, O),.
`Relative displacements between I , m, n and o were
`determined by inspecting each neighbouring, overlapping
`pair of stacks along their depth, and looking for bright.
`distinctive, well-defined, shared features. Connective tissue
`cables shared by 1 and m (or by m and n, n and o or o and I)
`are often best. The pixel intensities in and around such
`features visible on two sections from the two different series
`were recorded on graph paper. Although the displacement
`between stacks will seldom be an integral number of pixel
`rows or columns so that the two pixel intensity patterns
`might be identical, the edge, centroid or maximal brightness
`can usually be located in both instances of the feature. The
`row and column displacements needed to move the landmark
`of the feature on image m to coincide with that on image 1
`were expressed in terms of coordinates on image p.
`
`Fig. 2. Compensation image for systematic intensity gradients within
`a field of view (see text). The image consists of 256 x 256 pixels, with
`a minimum value of 0.77490 at row 103, column 139 (row 1, col-
`umn 1 is the upper left corner), and maximum value of 0.99609 at
`row 256, column 1. The image was generated by performing a local
`thin-plate spline interpolation among over 50 landmarks selected
`manually from averaged and filtered images taken of a fluorescent
`glass standard, then computing reciprocal image and scaling.
`
`Equalizing intensities among images of a series
`
`During the time needed to scan a section series through the
`available depth range,
`the overall section
`intensity
`decreases due to the factors mentioned above. If this
`decrease jeopardizes the number of significant figures in
`the fluorescence intensity data, detector sensitivity may be
`adjusted between section scans so as to keep the maximum
`recorded intensity just below saturation. Since we do not
`know beforehand the optimal adjustment, even such
`section series may be improved by post-acquisition correc-
`tion of overall section intensities within each series. Our
`method of correcting uneven intensities among images in a
`depth and/or time series may be described simply. For each
`pixel location, the intensity on a section is plotted against
`the intensity on its immediate predecessor, forming a two-
`dimensional scatter plot or histogram. Any pixel pair in
`which one of the two intensities is very low (0.0-0.06 of
`full scale) or very high (1.0 of full scale) is deleted. This
`eliminates airspace with noise and saturated pixels,
`respectively. The slope of the best fit line (least sum of
`squared minimal distances from points to the line) through
`the remaining points is calculated, and the section is divided
`by the slope. The process is repeated through the stack,
`three times. On the third and final pass, the process typically
`
`

`
`226 E. H. OLDMIXON & K . CARLSSON
`
`This manual procedure is facilitated by displaying both
`sections on a monitor and magnifying to inspect the
`individual pixels. Usually there is no uncertainty by the
`operator as to the relative displacement; when there is, it is
`usually between two possible displacements differing by one
`pixel. In practice, when the matches made at 3-5 levels
`throughout the pair of stacks are reviewed, the proper
`displacement becomes clear.
`It should be possible to find those best matches by a
`computable criterion, perhaps using an algorithm similar to
`the intensity adjustment algorithm but varying displace-
`ment along rows and columns instead of intensity. We have
`not implemented such a procedure, however. The present
`manual matching procedure takes several hours of work
`per volume data set. The process was repeated for image
`series 1 and o; the displacement expressed the translations
`along rows and columns needed to move image o to
`coincide with image 1. Next, the process was repeated for n
`and m, then n and o; in these cases, the displacements were
`expressed as those needed to move n to coincide with m (or
`o) before m (or o) was translated to coincide with I. If we
`notate the row and column translations needed to move
`features on image b to coincide with those on image a as
`i
`b -+ a (r, c), then the translations described above are rn -+
`(r, c), o -+ 1 (r, c), n + m (r, c) and n -+ o (r, c). Note that
`a -+ b (r, C) = -b -+ a (r, C) = b -+ a (-r, -c).
`The four displacements (I -+ m -+ n + o -+ 1) should
`make a closed loop:
`rn --+ 1 (r, c) + n +m (r, c) + o -+ n (r, c) + I + o (r, c)
`
`If not, then the error has probably been introduced from
`two sources: (i) instrumental distortions of the fields of view
`and (ii) non-cancelling errors in selecting integral pixel
`translations to approximate actual translations between
`fields of view. Our experience has been that 1 -+ I (r, c)
`values are of the order of 1 or 2 pixels, a disparity readily
`eliminated by slightly changing one or more of the four
`component translations. The fact that the translation loop is
`nearly closed gives confidence in both the instrument and
`the stack-fusing technique.
`The preceding steps may leave the centre of the ductal
`unit's profile well away from the centre of image p; since the
`data volume may be conveniently rotated about its centre
`when making three-dimensional reconstructions, we try to
`bring the centre of the ductal unit to the centre of the data
`volume. The translation needed to shift the fused image p so
`that the tissue of interest is centred in the final 5 12 x 5 12
`image q is denoted by p -+ q (r, c).
`The final translations are
`
`Fig. 3. An example of four 256 x 256-pixel images (1, m. n, o)
`translated and fused to form 512 x 512-pixel image q (see text).
`Th translation operation has moved images I, o and n toward the
`centre of image q. h a g e m has not been shifted in this example;
`m -t q (r = 0, c = 0). The portions of I, n and o which were trans-
`lated across the original boundaries were discarded where they fell
`into a portion of the neighbouring image after translation, but kept
`in two places where the content of image I lies in an 'unoccupied'
`portion of the original o domain and where o does the same in n.
`
`o + q (r, C) = 0 -+ 1 (r, C) + P -+ 4 (rr c).
`The remaining question is how to deal with the regions of
`overlap among the images. A simple approach, which is
`usually satisfactory, is to retain each pixel of an image which
`remains within the original 256 x 2 56 domain of that
`image, and to retain a pixel shifted out only if its new location
`is unoccupied by a pixel from any other image (see Fig. 3).
`
`Equating intensities among stacks
`
`Before creating image q, the relative brightnesses of I , rn, n
`and o must be inspected and probably modified. Although
`
`Fig. 4. The final result; a surface-shaded reconstruction of canine
`lung alveolar ductal unit constructed from stacks containing the
`sections shown in Fig. 1. Tissue not belonging to this ductal unit
`has been removed.
`
`

`
`the intensities within stacks I, m, n and o have ostensibly
`been equalized, there is no guarantee that their intensities
`are equal, stack to stack. Having decided upon the
`translations, however, we have also defined the over-
`lapping regions of corresponding sections in each series.
`Taking the I-m pairs as one example, we generated a
`histogram from stack 1 of the tissue intensities in the region
`of overlap with stack m, then the same for stack m. The
`factor used to bring the tissue to a common intensity was
`determined by inspection. (The eigenvector method used to
`adjust pairwise intensities within a stack is also applicable.)
`The factors from m to I, n to m, o to n and I to o were found,
`then adjusted as required to bring the final intensities of all
`stacks to the desired level. Note that the product of these
`four factors should be unity.
`fusing four 256 x
`Figure 4 shows the result of
`256 x 192-voxel stacks into a 512 x 512 x 192-voxel
`stack by this process.
`
`Applications to lung structure
`
`Larger view volumes composed in the manner described
`prove useful in the investigation of lung parenchymal
`structure in several respects. A ductal unit is packed among
`other ductal units, and few portraits of one exist. The
`boundary of the composed ductal unit can be determined
`readily, and tissue beyond the boundary then deleted, with
`the result that a reconstruction of
`the edited volume
`shows the normally embedded ductal unit now isolated for
`study (Fig. 4). Qualitatively, it may be informative to see
`an anatomic structure which otherwise might remain an
`abstraction. The composed volumes promise to be useful
`in quantitative studies as well. The positions of tissue
`components (connective tissue condensations, for example)
`may be marked on the sections, since the internal structure
`of the tissue is visible there. It is also feasible to extract the
`voxels at the air-tissue interface, record their locations and
`determine the local orientation of the interface. The ductal
`unit axis may be established by viewing the entire volume
`and individual sections in turn. It is then possible, for
`example, to discern the radial distribution of connective
`tissue and air-tissue interface about the duct axis.
`
`Generality of the method and relationship to other
`techniques
`
`Techniques developed for the combination of the Royal
`Institute of Technology's confocal microscope and lung
`parenchymal samples should be applicable to other image
`acquisition methods and other samples, if the following
`qualifications are met.
`
`(1) The instrumental illumination/detection pattern within
`the field of view must be characterizable and reproducible.
`
`L A R G E CSLM DATA VOLUMES 2 2 7
`
`intensity equalization
`(2) To perform post-collection
`throughout a section series after unknown alterations
`to the instrument illumination/detection levels between
`section, the sample should contain
`three-dimensional
`structures which will appear in the same pixel locations
`on successive sections and which have the same intrinsic
`fluorescence in both instances. In the lung, examples
`of these structures which serve well with 1 x 1 x 1 pm
`voxels are connective tissue cables, nuclei and basement
`membranes.
`(3) Geometric distortion of the field of view has not been
`addressed, as it is judged not to be a problem with this
`CSLM. If distortion correction must be done, it should
`probably be performed after the within-section and within-
`series intensity adjustments have been made.
`
`The alignment of neighbouring section series as done
`with lung samples depends on small, bright, distinctive
`image

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket