`5.1
`
`THERMOGRAVIMETRIC ANALYSIS (TGA)
`
`and measuring the change in some physical property. The most important
`thermal methods for the study of solid-state chemistry are thermogravimet-
`ric analysis (TGA), differential scanning calorimetry (DSC), and thermal mi-
`croscopy (discussed in Section 4.4). Therrnogravimetric analysis measures the change
`in the mass of sample as the temperature is changed. Differential scanning calorimetry
`
`
`involves measuring the difference between the temperature of the sample and 3 refer-
`
`
`ence compound as the temperature of the system is changed, thus providing informa—
`
`
`tion on the enthalpy change of various solid-state processes. Thermal methods of
`
`
`analysis are important analytical tools for characterizing pharmaceutical solids. The use
`
`
`of TGA and DSC in conjunction with thermal microscopy (Section 4.4) can elucidate
`
`many behaviors of solids.
`
`Basically, a thermogravimetric instrument consists of a microbalance connected to a
`
`sample compartment situated in a small oven with computer-controlled temperature
`
`
`programming. A dry nitrogen atmosphere is most commonly used, however, other
`
`
`gases can be employed (the compostion and flow dynamics of the gas are important
`
`
`parameters.) This method measures the change in mass with temperature and is often
`
`
`used to study the loss of solvent of crystallization or other solid —) solid + gas reac-
`
`
`tions. A typical TGA trace is shown in Figure 5.1. In studies of solid-state chemistry,
`
`
`TGA is usually performed in one of three modes:
`
`
`1 . Isothermal mode——the temperature is kept constant.
`2. Quasi-isothermal mode—the sample is heated to a constant mass
`
`
`
`
`through a series of increasing temperatures.
`
`
`3. Dynamic mode—the temperature is raised at a known rate, typi—
`
`cally linear.
`
`Thermal Methods
`
`of Analysis
`
`‘ l
`
`hermal analysis generally refers to any method involving heating the sample
`
`
`
`81
`
`
`
`|PR2016-00006
`
`SteadyMed - Exhibit 1027 - Page 2
`
`IPR2016-00006
`SteadyMed - Exhibit 1027 - Page 2
`
`

`

`
`
`
`
`
`Chapter 5 Drugs as Molecular Solids
` 82
`TGA. Obvious]
`
`reaction and the
`
`
`
`mass loss
`
`
`’
`
`I,
`'
`
`time. These plot:
`also been used 1
`
`general, the kinei
`thermogravimetrj
`desolvation of or
`
`
`
`5.2 DIFFEREN'
`
`i
`
`*3J“
`
`
`
`
`
`
`
`
`s of the type shown in Equation 5.1 can
`
`be determined using
`
`|PR2016-00006
`
`SteadyMed - Exhibit 1027 - Page 3
`
`Differential scan
`energy (heat flu:
`DSC sample cor
`The result c
`
`
`
`
`
`state mass loss with T, (the transition temperature)
`
`ding to the point at which tangents to the original
`fl
`Figure 5.1 A typical TGA trace for a single-
`marked. The temperature correspon
`he transition temperature T,.
`baseline and to the slope of the tracing represents t
`
`
`
`emperature regions where no weight
`
`
`The last approach uses high heating rates in t
`here weight changes do occur, thus
`low rates in regions w
`
`
`changes are occurring and s
`g of peaks from overlapping
`avoiding transition temperature overshoot and blurrin
`
`
`
`
`transitions.
`
`
`ber of factors or conditions that affect TGA curves including the
`
`
`There are a num
`f the sample holder (pan), particle size of the
`
`
`heating rate, atmosphere, geometry 0
`t of the sample, thermal conductivity of the
`
`
`sample, nature of the reaction, treatmen
`
`
`sample, and sample weight. The effect of the heating rate has been extensively studied
`(Wendlandt, 1974).
`In general, as the heating rate is increased, the apparent starting
`temperature of the thermal event (Ti) increases. However, this condition can some-
`
`
`
`
`times be corrected by decreasing the sample size.
`For example, an
`The atmosphere can have a dramatic effect on the TGA curve.
`T,- or stop the reaction
`oduct gas can increase
`
`
`f the reaction, particu-
`atmosphere already containing the pr
`he atmosphere can change the course 0
`r the reactant. Knowledge
`completely. In addition, t
`
`
`r the products 0
`
`
`larly if the atmospheric gas reacts with eithe
`RH) is essential to
`
`
`onds to changes in relative humidity (
`
`
`of how the substance resp
`started. For these reasons,
`it is a
`dling of the sample before the scan is
`
`
`
`
`proper han
`ere of dry nitrogen when performing a study.
`prudent practice to use an atmosph
`
`
`the particle size of the sample has
`ton the reaction mechanism,
`Although dependen
`The smaller the particle size,
`the
`
`
`TGA curve in general.
`
`
`cause the smaller particle
`a predictable effect on the
`the lower the value of Ti. This is be
`faster the reaction and
`the nature of the reaction
`
`
`e of the product gas. Obviously,
`
`
`sizes allow more rapid escap
`r for more facile reactions.
`
`
`affects T; which will be lowe
`particular the extent of compression
`In addition, the treatment of the sample, and in
`pression will
`feet the Ti. For example, increased com
`of the sample, will obviously af
`11 have less opportunity to escape.
`
`
`increase T,- since the product gas wi
`ty of the sample will influence T,-. Anomalous ef-
`
`
`Finally, the thermal conductivi
`ature of the sample is not uniform because of poor
`fects may be obtained if the temper
`thermal conductivity.
`The rates of reaction
`
`
`.
`
`f
`
`
`
`
`
`
`
`Hall? 5.2 Cros:
`are [N
`
`
`
`
`
`
`. 5.3 A hyp
`sampl
`
`IPR2016-00006
`SteadyMed - Exhibit 1027 - Page 3
`
`

`

`
` 33
`5.2 Differential Scanning Calorimetry (DSC)
`
`
`TGA. Obviously, isothermal TGA traces can be used to determine the rate of the
`'mply plotting weight loss versus
`reaction and the rate law governing the reaction by 51
`
`
`
`
`time. These plots can then be analyzed as described in Chapter 3. Dynamic TGA has
`
`
`also been used to determine the rates of such gas—evolving reactions. However, in
`general, the kinetic data thus obtained should be substantiated by other data. Isothermal
`
`
`thermogravimetric analysis has been used extensively in our laboratory to study the
`
`
`
`
`desolvation of crystal solvates (Chapter 16).
`solid
`A
`'—-’ B
`
`
`5.2 DIFFERENTIAL SCANNING CALORIMET
`
`
`Differential scanning calorimetry (DSC) is a method which measures the difference in
`energy (heat flux or heat
`) and a sample (S). A typical
`
`
`flow) between a reference (R
`DSC sample compartment is shown in Figure 5.2.
`
`
`The result of a DSC analysis is a thermogram, a plot of AT = T, — T, (temperature
`
`
`solid
`
`+ C
`
`gas
`
`RY (DSC)
`
`outer lid
`
`(5.1)
`
`.u
`
`l
`
`i
`
`(the transition temperature)
`rich tangents to the original
`on temperature T,.
`
`:gions where no weight
`1t changes do occur, thus
`peaks from overlapping
`
`‘GA curves including the
`mm), particle size of the
`:rmal conductivity of the
`5 been extensively studied
`ised, the apparent starting
`this condition can some—
`
`curve. For example, an
`e T, or stop the reaction
`se of the reaction, particu—
`: the reactant. Knowledge
`ridity (RH) is essential to
`For these reasons,
`it is a
`)erforming a study.
`ticle size of the sample has
`naller the particle size, the
`>ecause the smaller particle
`y, the nature of the reaction
`
`it the extent of compression
`increased compression will
`to escape.
`fluence Ti. Anomalous ef-
`lOt uniform because of poor
`
`
`
`inner DSC
`cell lid
`
`
`
`therrnopile
`iunction
`
`
`
`silver
`furnace
`elements
`
`body
`
`
`to op amp
`
`
`The sample pan (S) and the reference pan (R)
`
`
`1996).
`Figure 5.2 Cross section of a Cahn® DSC 4000 cell.
`are positioned in the sensor (Cahn Instruments,
`
`
`
`
`
`
`
`l Exotherm
`
`Time
`
`
`
`re 5.3 A hypothetical DSC thermogram showing the changes that might occur upon heating a
`
`
`Figu
`sample.
`
`|PR2016-00006
`
`SteadyMed — Exhibit 1027 — Page 4
`
`5.1 can be determined using
`
`
`
`IPR2016-00006
`SteadyMed - Exhibit 1027 - Page 4
`
`

`

`84
`
`hi
`w
`M
`
`fl
`. M
`
`i
`I
`
`a
`1
`
`.
`
`
`
`
`
`Based on this wc
`of fusion rule
`
`
`
`Chapter 5 Drugs as Molecular Solids
`desolvation process
`
`5.3 shows an idealized DSC trace. The endotherms
`an important influen‘
`
`
`difference) versus T. Figure
`represent processes in which heat is absorbed, such as solvent loss, phase transitions,
`that affect the rate 01
`or melting. The exotherms represent processes such as crystallization or chemical
`
`
`has sublimed or m
`reactions where heat is evolved.
`In addition, the area under a peak is proportional to
`properties upon reh<
`the heat change involved. Thus this method, with proper calibration, can be used to
`Two definition:
`determine the enthalpies (AH)) of the various processes. The method can also be used
`
`
`as an accurate measure of the melting point and purity of the sample.
`In fact,
`the
`
`
`change of melting point is related to the mole fraction of impurities as given by Equa-
`
`
`
`
`tion 5‘2:
`2
`
`
`
`Ts = To _ Tu RX r
`(5.2)
`
`
`FAHf
`
`
`
`
`where T, is the sample temperature, T0 is the melting point of the pure compound, R is
`the gas constant, X,- is the mole fraction of the impurity, F is the fraction of the solid
`melted, and AHf is the enthalpy of fusion of the pure compound. According to the
`
`
`lope is proportional
`
`
`l/F should give a straight line whose s
`ars to fail when purity is less
`
`
`equation, a plot of T5 versus
`1991). However, the equation appe
`to X; (Brittain et al.,
`.
`.
`than 97%. Application of this equation is illustrated by the DSC thermograms shown
`
`
`in Figure 5.4.
`
`
`There are a number of factors other than purity that can affect the DSC curve in—
`eluding heating rate, atmosphere, sample holder, particle size, and sample packing.
`In
`
`
`general, agreaterheatingrate willcause a shift of the peaks to higher temperatures. A
`
`
`
`
`decreased heating rate also usually causes endotherms and exotherms to become
`:
`sharper The shape of the sample holder and whether it is open,
`totally sealed, or
`
`
`3,0
`.
`.
`.
`.
`
`
`J,
`contains a pin prick to vent gases can also affect a DSC curve. When a DSC experi-
`
`
`"
`ment is performed in a closed pan, the resulting atmosphere within the sample holder
`g DSC curve. Obviously, a tightly sealed sample holder
`hanism of a
`can greatly affect the rcsultin
`thereby changing the behavior or mec
`
`would not allow vapor to escape,
`
`
`ergies of polymorpl
`monotropic system
`temperature.
`In at
`(transition) tempera
`high temperature r:
`room temperature (
`cause confusion an
`system is enantioti
`termperature diagr
`reliable rules whi
`monotropic usingt
`l. The h
`dothei
`eratu
`fiotro
`the fcl
`forms
`”
`2. The l
`meltii
`relate
`
`99.6% ..
`
`.
`

`
`
`
`
`
`
`
`
`
`
`
`points but simila
`forty energy—tem
`much more wor]
`calculated the he:
`polymomhs( bas
`the applicability .
`DSC is alsr
`show the DSC st
`containi'tg mixtt
`the higher meltii
`5.6 shows pure
`this same mixtui
`form is convert:
`study of mixtur
`tures of polymt
`DSC thermogra
`DSC can be ust
`
`
`
`
`
`
`
`
`1 40,0
`
`1 45.0
`
`150.0
`
`155.0
`
`Temperature (°C)
`
`
`
`hree ethoxyearbonyl-3-phenylpropyl.L-alanine samples of varying
`iron, 1990).
`Figure 5.4 DSC thermograms of t
`purity from different manufacturers (G
`
`
`
`
`|PR2016-00006
`
`SteadyMed - Exhibit 1027 - Page 5
`
`IPR2016-00006
`SteadyMed - Exhibit 1027 - Page 5
`
`

`

`
`
`5.2 Differential Scanning Calorimetry (DSC)
`
`85
`
`desolvation processes. As with TGA, the particle size and packing of the sample has
`an important influence on reactions especially those of desolvation type. Any changes
`that affect the rate of heat transfer should also be taken into account. Thus a sample that
`has sublimed or melted and then recrystallized may show somewhat different DSC
`properties upon reheating.
`Two definitions are often used to describe the relationship between the relative en-
`ergies of polymorphs at different temperatures: monotropic and enantiotropic.
`In a
`monotropic system, one form is the thermodynamically stable form regardless of the
`temperature.
`In an enantiotropic system, one form is more stable below a certain
`(transition) temperature but another form is more stable above that temperature. Thus,
`high temperature recrystallization may lead to one form, whereas recrystallization at
`room temperature could lead to the other form. Enantiotropic systems can sometimes
`cause confusion and problems with crystallization.
`In general, to determine whether a
`system is enantiotropic or monotropic it would be helpful
`to construct an energy-
`terrnperature diagram. Burger and Ramberger (l979a—b) have construeteded two
`reliable rules which assist
`in determining whether a system is enantiotropic or
`monotropic using thermoanalytical results:
`1 . The heat (or enthalpy) of transition rule states that (a) if an en-
`dothermic transition is observed between the forms at some tem—
`perature it may be assumed that the two forms are related enan—
`tiotropically and (b) if an exothermic transition is observed between
`the forms at some temperature it may be assumed that
`the two
`forms are related monotropically.
`2. The heat (or enthalpy) of fusion rule states that if the higher
`melting form has the lower heat of fusion then the two forms are
`related enantiotropically, otherwise they are related rnonotropically.
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`The endotherms
`)hase transitions.
`ttion or chemical
`s proportional to
`1, can be used to
`can also be used
`ple.
`In fact.
`the
`[5 given by Equa-
`
`(5.2)
`
`e compound, R is
`action of the solid
`According to the
`me is proportional
`when purity is less
`ermograms shown
`
`the DSC curve in-
`ample packing.
`In
`:r temperatures. A
`therms to become
`, totally sealed, 0r
`'hen a DSC experi—
`tthe sample holder
`:aled sample holder
`or mechanism of a
`
`lanine samples of varying
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Based on this work. Grunenbcrg et al. (1996) expanded these rules with the entropy
`of fusion rule (particularly necessary for polymorphs with very different melting
`points but similar ethalpies of fusion) and a heat capacity rule. Since only about
`forty energy-temperature diagrams for pharmaceutical systems have been published,
`much more work needs to be done.
`In related studies. Behme and Brook (1991)
`calculated the heat of fusion of the lower melting of an enantiotropieally related pair of
`polymorphs( based on the heat of transition and the heat capacities) and demonstrated
`the applicability of thermodynamic calculations.
`DSC is also useful for studies of polymorphic mixtures. Figures 5.5 and 5.6
`show the DSC scans of propyphenazone. Figure 5.5 shows the DSC scans of batches
`containing mixtures of Forms 1 and 11 indicating that DSC can detect as little as 5% of
`the higher melting form in the mixtures (Giron-Forcst et al.. 1989). Trace A in Figure
`5.6 shows pure Form 1, trace B shows a mixture of Forms I and II. and trace C shows
`
`this same mixture after heating at 100°C for two days indicating that the higher melting
`
`form is converted to the lower melting form under these conditions. In a more extensive
`
`
`study of mixtures. (Giron, 1986) showed that DSC could be used to quantitatc mix-
`
`
`tures of polymorphs as shown in Figure 5.7. The left panel in Figure 5.7 shows the
`DSC thermograms of Forms I and II of a pharmaceutical; the right panel shows that
`
`
`DSC can be used to analyze mixtures of these two forms (Giron, 1986).
`
`
`
`
`|PR2016-00006
`
`SteadyMed - Exhibit 1027 - Page 6
`
`IPR2016-00006
`SteadyMed - Exhibit 1027 - Page 6
`
`

`

`5 mg 20 °Clmin
`
`hgure 5.7 Determination of l
`
`Thermal methods have
`
`
`
`-—>
`
`Endothermic
`13% Form ll
`
`5% Form llW...
`
`0% Form u
` j
`
`
`
`I
`100
`102
`105
`107
`110
`a
`
`
`
`Temperature (°C)
`Temperature ( C)
`
`
`Figure 5.5 Melting behavior of batches containing propyphenazone which are mixtures of Form 1
`with a small amount of Form 11 (Giron—Forest er al., 1989).
`
`
`
`
`
`
`
`
`
`
`
`
`
`mixture of Form I and Form II
`
`
`
`
`
`mixture 01 Form l and Form Il
`
`
`(same as above after Isothermal
`treatment for 2 ays at 100 “C
`
`
`
`
`____|__’__|___l__——l————
`
`
`
`1 009590 1 05 °C
` tures of Form I and Form [1 before
`
`
`ity (Giron, 1990). In this
`varying between 10:1 and
`of the heated samples is us
`and the results are compar
`should reflect the actual ;
`determine incompatibilities
`the DSC thermograms oi
`formation; thus, a change
`tive of a stability problem 1
`One advantage of DS
`thus, a study over a Wldt
`however, will have to be
`useful in the study of solic
`combined with other tech
`HPLC.
`
`Figure 5.6 DSC scans of propyphenazone pure Form 1 and mix
`and after isothermal treatment (Giron-Forest et al.. 1989).
`
`5.3 M ICROCALO RIMET
`
`Microcalorimetry is a vex
`given off or taken up by v
`try is used, for example,
`every transformation, eith
`of heat, this method has si
`baum and McGraw (1985
`different crystal forms ha
`However, the difference I
`solvents should remain t1
`diffeI‘Cnce is the heat of tr.
`
`|PR2016-00006
`
`SteadyMed - Exhibit 1027 - Page 7
`
`IPR2016-00006
`SteadyMed - Exhibit 1027 - Page 7
`
`

`

`
`5.3 Calorimetry
`
`87
`
`0 Form I
`(A): 41.9 mJ s"
`
`
`5 mg 20 °Clrnin
`a Form II (B): 4.2 mJ s"
`
`
`
`
`
`
`nperature (°C)
`
`zone which are mixtures of Form 1
`1989).
`
`4...,—
`105 °C
`
`id mixtures of Form 1 and For
`11., 1989).
`
`m 11 before
`
`‘
`'
`
`
`‘90
`no
`‘60
`110
`10 20 30 40 50 60 7O 80 90100%Formll
`
`70 60 50 40 30 20
`90 80
`10
`0 lVeForn'Il
`
`
`
`
`Figure 5.7 Determination of ratios of Forms 1 and II of a pharmaceutical (Giron, 1986).
`
`
`
`
`Thermal methods have been successfully used to study drug-excipient compatibil—
`
`
`ity (Giron, 1990). In this procedure, drug and excipient are intimately mixed in ratios
`
`
`varying between 10:1 and 1:10 and each mixture is analyzed by DSC. HPLC analysis
`of the heated samples is used to interpret any changes in the DSC profile of the mixture
`
`
`
`
`and the results are compared with those of the pure components. The ratios analyzed
`should reflect the actual proportions in the formulation; however,
`it is instructive to
`
`
`determine incompatibilities at other concentrations as well.
`It is important to note that
`
`
`the DSC thermograms of mixtures will show some changes simply from eutectic
`formation; thus, a change in DSC melting point for a drug and excipient is not indica-
`
`
`tive of a stability problem by itself.
`
`
`One advantage of DSC is that the sample is subjected to different temperatures;
`thus, a study over a wide temperature range can be rapidly carried. Most results,
`
`
`however, will have to be confirmed by using other methods. Thermal methods are
`
`
`useful in the study of solids but the power of these methods is greatly enhanced when
`
`
`combined with other techniques such as X-ray powder diffraction, microscopy, and
`I‘
`1
`HPLC.
`
`
` 5.3 MICROCALORIMETRY
`
`Microcalon'metry is a very sensitive calorimetrie technique that determines the heat
`
`
`given off or taken up by various processes. For pharmaceutical solids, microcalorime-
`try is used, for example, to measure heats of solution and degradation rates. Since
`
`
`
`
`every transformation, either chemical or physical, occurs with evolution or absorption
`
`
`. of beat, this method has significant potential for the study of transformations. Linden—
`haum and McGraw (1985) have used microcalorimetry to study drug forms. Because
`different crystal forms have different structures, they have different heats of solution.
`
`
`
`2 However, the difference between the heats of solution of two polymorphs in different
`
`solvents should remain the same (Table 5.1) if there is no solvate formation. This
`,.- erence is the heat of transition between the forms at that temperature.
`
`
`
`E
`3
`
`
`
`|PR2016-00006
`
`SteadyMed - Exhibit 1027 - Page 8
`
`IPR2016-00006
`SteadyMed - Exhibit 1027 - Page 8
`
`

`

`
`
`Chapter 5 Drugs as Molecular Solids
`88
`Table 5.1 Heats of Solution of Sodium Sulfathiazole
`Solvent
`AHs Form 1
`AHS Form 11
`Aflmm
`(kJ/mol, 25 °C)
`(kJ/mol, 25 °C)
`(kl/moi, 25 °C)
`
`Acetone
`1 1.94
`5.144
`6.798
`
`
`
`774.659 —1 1.47DMF 6.810
`' Lindenbaum and McGraw. 1985.
`
`Studies by Ip et al. (1986) on enalapn'l maleate give similar results showing that
`the heats of transition between the two forms determined by subtraction of the heats of
`solution in two different solvents are within the experimental error. With suitable
`calibration of known mixtures, this phenomenon can sometimes be the basis for
`analyzing mixtures of polymorphs or crystalline and amorphous forms of a compound.
`Of course these comparisons apply only to solids with the same composition (i.e.,
`when the resulting solutions are identical). Also, a hydrate and an anhydrate cannot be
`compared since the heat of the solution of water will be different in different solvents
`Isothermal microcalorimetry has also been used to determine the crystallinity of
`and thus the AH“ans will be different.
`mixtures of amorphous and crystalline antibiotics as shown in Figure 5.8 (Thompson et
`al., 1994). DSC could not be used since the samples decomposed prior to melting.
`In
`contrast to studies by Osawa and coworkers (1988) as well as Pikal and coworkers
`(1978), it was found that the heat of solution was not dependent on water content. The
`importance of initial water content is probably greatest when dealing with hydratable
`ionic species since sodium and quaternary ammonium salts have very high heats of
`Several important papers on the use of microcalorimetry for stability determina-
`hydration (see Figure 5.9).
`tions have appeared. Hansen et al. (1989) studied the kinetics of decomposition of
`lovastatin and other HMG—CoA reductase inhibitors using heat conduction calorime-
`try (the response of the instrument is directly proportional to the rate of heat produced
`in the sample cell). Heat conduction calorimetry has a substantial advantage over
`
`(J/g) -20 0
`
`10
`
`20
`
`30
`
`60
`50
`40
`Percent Crystallinity
`
`100
`
`HeatofSolution
`
`Figure 5.8 Heat of solution of antibiotic 302669 in 0.02 M NazHPO4 at 35 °C as a function of
`percent crystallinity (Thompson et LIL, 1994).
`
`
`
`
`
`
`
`Figure 5.9 The effect of w
`
`conventional microcalor‘
`,uW) can be detected. Tl
`determined after only a
`urement of degradation
`temperature. The rate 1
`calorimeters can also b1
`excipients and stabilize]
`rimetry to establish that
`atmospheres only a sn
`under oxygen atmospl
`atmospheres.
`Further
`change was about -—40
`oxidation. Bond ener;
`group would produce 2
`tion microcalorimetry.
`area of the sample has
`experiments, they Sh(
`produced under identi
`oxygen than others.
`that a single measuren
`used to predict the tot:
`conduction nlicrocalo
`cases and appears to l
`
`
`
`
`
`
`
`
`
`REFERENCESW
`
`Behme, Robert J. and D3
`of carbamazepine th
`analysis" J. Pharm.
`Brittain, Harry 0.. Susan
`Ann W. Newman
`963—973.
`
`Burger. A. and R. Raml
`crystals.
`1. Theory
`
`
`
`
`
`
`|PR2016-00006
`
`SteadyMed - Exhibit 1027 - Page 9
`
`IPR2016-00006
`SteadyMed - Exhibit 1027 - Page 9
`
`

`

`
`
`
`
`LIIIYIIS
`>1, 25 DC)
`t.798
`n.810
`
`llts showing that
`m of the heats of
`r. With suitable
`be the basis for
`5 of a compound.
`:omposition (i.e..
`hydrate cannot be
`different solvents
`
`the crystallinity of
`5.8 (Thompson et
`ior to melting.
`In
`(211 and coworkers
`vater content. The
`rig with hydratable
`very high heats of
`
`stability determina-
`f decomposition of
`duction calorime-
`te of heat produced
`tial advantage over
`
`80
`
`90
`
`1 00
`
`t 35 °C as a function of
`
`
`
`References
`
`89
`
`Sodium Ceiazolin
`Cephaloridine
`
`5 form
`a lorm
`
`
`AHskcal/mol
`
` Sodium Cephalothin
`Amorphous, lreeze dried
`
`
`
`6
`7
`4
`5
`
`
`
`Water, % (w/w)
`
`
`
`Figure 5.9 The effect of water content on the heats of solution of antibiotics (Pik'di et ul., I978).
`
`
`
`
`
`conventional microcalorimetric methods in that extremely small outputs of heat ($0.1
`
`
`HW) can be detected. The heat of decomposition and the kinetics of the process can be
`
`
`determined after only a very small percentage of reaction. This then allows the meas
`
`
`urement of degradation of the material in the early stages of the reaction even at room
`
`
`temperature. The rate law and the activation energy can also be determined. These
`
`
`calorimeters can also be used to study freshly formulated materials and the effects of
`
`
`
`
`excipients and stabilizers on degradation. Hansen et er a]. (1989) also used microcalo-
`
`
`rimetry to establish that oxygen was required for degradation of lovastatin since in inert
`
`
`atmospheres only a small amount of heat was produced whereas the heat produced
`under oxygen atmosphere was 20—90 times greater than that produced under inert
`
`
`atmospheres. Furthermore,
`they used the heat produced to estimate the enthalpy
`
`
`change was about —400 kJ mol'1 which is consistent with what one might expect for
`
`
`oxidation. Bond energy calculations show that reaction of oxygen with a methylene
`
`
`group would produce an enthalpy change of about —600 kJ mol". Using heat conduc—
`
`
`tion microcalorimetry, Hansen and coworkers were also able to show that the surface
`
`area of the sample has an effect on the rate of oxidation. as might be expected.
`In other
`
`
`
`experiments, they showed that
`there was significant
`lot—torlot variation in the heat
`produced under identical conditions. Some lots showed much greater reactivity with
`
`
`oxygen than others. One of the most significant results of this study was the finding
`
`
`
`that asingle measurement of the heat produced per gram of drug for each lot could be
`used to predict the total degradation of that lot under conventional stability testing. Heat
`
`
`conduction mierocalorimetry has been shown to have predictive capability in some
`
`
`cases and appears to be an important addition to other stability studies.
`
`
`
`wwWWWMWW
`REFERENCESWW
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Behme. Robert J. and Dana Brooke (l99l) “Heat of fusion measurement of a low melting polymorph
`of carbamazepine that undergoes multiple-phase changes during differential scanning calorimetry
`analysis" J. Pharm. Sci. 80 986—990.
`Brittain, Harry 0., Susan J. Bogdanowich. David E. Bugay, Joseph DeVincentis, Geoffrey Lewen, and
`Ann W. Newman (1991) “Physical characterization of pharmaceutical solids" lerm. Res. 8
`963—973.
`Burger, A. and R. Ramberger (1979a) “0n the polymorphism of pharmaceuticals and other molecular
`crystals.
`1. Theory of thermodynamic rules" Mikmc/iim. Arm ll 259727l.
`
`
`
`|PR2016-00006
`
`SteadyMed - Exhibit 1027 - Page 10
`
`IPR2016-00006
`SteadyMed - Exhibit 1027 - Page 10
`
`

`

`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Chapter 5 Drugs as Molecular Solids
`90
`
`
`hism of pharmaceuticals and other molecular
`
`Burger, A. and R. Ramberger (1979b) “On the polymorp
`crystals. 11. Applicability of Thermodynamic rules" Mikrochim. Acta 11 273—316.
`Cahn Instruments (1996) 5225 Verona Road, Bldg. 1, Madison WI 53711-4418.
`
`Giron, D. (1986) “Applications of thermal analysis in the pharmaceutical industry” J. Pharm. Biomed.
`Anal. 4 775—770.
`“Thermal analysis in pharmaceutical routine analysis” Acta Pharm. Jugosl. 40
`Giron, Daniele (1990)
`95457.
`Giron—Forest, D., Ch. Goldbronn, and P. Piechon (1989) “Thermal analysis methods for pharmaco-
`poeial materials” J. Pharm. Biomed. Anal. 7 1421—1433.
`Grunenberg, A., J.70. Henck, and H. W. Siesler (1996) “Theoretical deviation and practical application
`of energy/temperature diagrams as an instrument in performulation studies of polymorphic drug
`substances" Int. J. Pharm. 129 147—158.
`. Bergstrom, and Damaris DeGiaIt—
`Hansen, Lee D., Edwin A. Lewis, Delbert J. Eatough, Robert G
`duction calorimetry" Pharm. Res. 6
`Johnson (1989) “Kinetics of drug decomposition by heat con
`20—27.
`venson, Siegfried Lindenbaum, Alan W. Douglas, S.
`1p, Dominic P., Gerald S. Brenner, James M. Ste
`David Klein, and James A. McCauley (1986) “High resolution spectroscopic evidence and solution
`calorimetry studies on the polymorphs of enalapril maleate" Int. J. Pharm. 28 183—191.
`Lindenbaum, Siegfried and Scott E. McGraw (1985) “The identification and characterization of
`polymorphism in drug solids by solution calorimetry" Pharm. Manufacturing 27—30.
`of pharmaceuticals by high—
`Pikal, Michael J. and Karen D. Dellerman (1989) “Stability testing
`sensitivity isothermal calorimetry at 25 °C:
`cephalosporins in the solid and aqueous solution
`states" Int. J. Pharm 50 233—252.
`tive crystallinity determina-
`Pikal, M. J, A. L. Lukes, John E. Lang, and K. Gaines (1978) “Quantita
`tions for B-lactam antibiotics by solution calorimetry:
`correlations with stability” J. Pharm. Sci.
`67 767—773.
`Osawa, Takashi, Madhav S. Kamat, and Patrick P. DeLuca (1988) “Hygroscopicity of cefazolin
`sodium: application to evaluate the crystallinity of freeze-dried products" Pharm. Res. 5 421—425.
`Thompson, Karen C., Jerome P. Draper, Michael J. Kaufman, and Gerald S. Brenner (1994) “Charac-
`
`terization of the crystallinity of drugs: 802669, a case study” Pharm. Res. 11 1362—1365.
`
`Wendlandt, Wesley W. (1974) Thermal Methods of Analysis, 2nd ed; John Wiley and Sons: New
`
`York, NY: PP 9—13.
`
`
`
`
`
`
`
`
`
`
`
`
`potential for bioequivals
`
`are usually compounds
`
`disperse.
`Examples i
`
`digoxin,
`diphenylhyd:
`
`quinidine, and warfarin.
`
`to ensure that the Unite<
`
`
`
`
`
`
`
`Solubilit
`
`Testing
`
`he rate of disso
`
`important aspen
`and solubility (
`same drug can obviousl
`necessary for both pure
`have the proper dissoluti
`1991; Banakar, 1992).
`'
`dissolution testing and
`USP-NF (United States
`
`
`
`6.1 TESTING
`
`Dissolution tests are 3;
`individual drugs. For e:
`specification that 80%
`carbamazepine tablets m
`be dissolved in 60 m
`laboratories now measu
`variations. For dissolul
`variables (e.g., time prt
`chosen carefully.
`Dissolution tests a
`
`
`
`|PR2016-00006
`
`SteadyMed - Exhibit 1027 - Page 11
`
`IPR2016-00006
`SteadyMed - Exhibit 1027 - Page 11
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket