throbber
1
`
`3M COMPANY 2023
`Mylan Pharmaceuticals Inc. v. 3M Company
`IPR2015-02002
`
`

`
`808
`
`H.D.C. Smyth / Advanced Drug Delivery Reviews 55 (2003) 807-828
`
`816
`
`820
`821
`
`822
`822
`
`824
`
`.......................................................................................... ..
`5.2.2. HFA systems ........................................................... ..
`5.2.3. Vapor pressure effects in terms of mechanisms of atomization ....... ..
`5.3. Drug concentration and drug substance
`5.4. Surfactants .......................................... ..
`5 .4. 1 . Suspension stability.
`5.4.2. Solution stability ...................................................... ..
`5.5. Inverse micelles and disperse systems .................................. ..
`5 .6. Electrostatics ...................................... ..
`
`.......................................................................................... ..
`5.7. Other propellants ............................................................... ..
`.......................................................................................... ..
`5.8. Microbiological growth ...................................................... ..
`6. Influence of device variables ...... ..
`
` .......................................................................................... ..
`
`
`
`6.1. Valve and container design
`6.2. Actuator design...
`7. Conclusions ............. ..
`References .................................................................................... ..
`
`
`
`.......................................................................................... ..
`.......................................................................................... ..
`
`
`
`1. Introduction
`
`Aerosolized beta agonists and anti-allergic com-
`pounds were first
`formulated as pharmaceutical
`aerosols in 1956 using chlorofluorocarbons (CFCs).
`CFC propellants possessed several desirable charac-
`teristics that prompted their use in propellant-driven
`metered dose inhalers (pMDIs) that
`include non-
`toxicity, inertness, and high vapor pressures. Gener-
`ally, metered dose inhaler formulations containing
`CFCS combine a blend of propellants, excipients, and
`drug substance such that formulation factors and
`device characteristics combine to generate an effi-
`cient spray for delivery to the lungs. However, the
`manufacture of CFC propellants was eliminated after
`the signing of the Montreal Protocol [1]. The phase-
`out of CFCs was in response to concern over
`possible detrimental effects on the ozone layer
`originally raised in the 1970s [2]. Since this original
`hypothesis, stratospheric ozone depletion has been
`demonstrated over the Antarctic [3]. In the United
`States, the Food and Drug Administration has recent-
`ly set standards for
`the use of ozone—depleting
`substances and so—called ‘essential—use’ deterrnina—
`
`tions [4]. Specifically, for propellant-driven metered
`dose inhalers determined not
`to have continued
`
`(1)
`the following points were made:
`essentiality,
`products
`that are no longer marketed, and (2)
`products containing ozone—depleting substances mar-
`keted after January 1, 2005, may be proposed to be
`non-essential,
`(3) a moiety can remain on the
`essential—use list until a non—ozone—depleting product
`is marketed (same route of administration, indication,
`
`and convenience), has sufficient supplies to meet
`patient needs, and has sufficient post-marketing data,
`and (4) a CFC-MDI will not be removed until at
`least
`two non—ozone—depleting alternative products
`are marketed under more than one new drug applica-
`tion.
`
`Accordingly, alternative systems for the delivery
`of inhaled medications have been an active research
`
`focus in recent years. Several different approaches
`have been adopted and include altemative propellant
`systems, propellant—free liquid methods, and dry
`powder inhaler-based devices [5—7]. This article will
`outline research efforts that have focused on the
`reformulation of CFC—based metered dose inhalers
`
`using alternative propellants. Predominantly this
`relates to the use of hydrofluorocarbon (HFC) pro-
`pellants and the accompanying formulation strategies
`that have developed alongside this substitution effort.
`These formulation factors, including device charac-
`teristics, are reviewed with respect to the perforrn—
`ance of MDIS.
`
`2. Metered dose inhaler design
`
`types of devices are used to deliver a
`Several
`metered dose of aerosolized medication to the respi-
`ratory tract. However, pMDIs are specifically recog-
`nized as those devices that incorporate a propellant,
`under pressure,
`to generate a metered dose of an
`aerosol through an atomization nozzle. MDIS are the
`most widely used respiratory drug delivery device,
`with an estimated 800 million units produced in 2000
`
`2
`
`

`
`H.D.C. Smyth / Advanced Drug Delivery Reviews 55 (2003) 807-828
`
`809
`
`‘T Canister
`
`
`
`Spacer Chamber
`
`Fig. 1. Basic components of a pMDI system.
`
`
`
`Actuator
`
`Atomization
`orifice
`
`Drug Substance (suspension
`or solution)
`Liquified propellants /
`excipients
`
`Metering Valve
`
`[8]. MDIs consist of several components (Fig. 1): the
`active substance formulated with propellant and
`excipients, a container, a metering valve crimped
`onto the container, an actuator that connects the
`metering valve to an atomization nozzle, and a
`mouth piece. Additionally, holding chambers or
`spacers may also form part of the delivery system by
`connection to the actuator mouthpiece. A metered
`volume (typically between 20 and 100 pl) of the
`drug/ eXcipient/ propellant blend is expelled from the
`canister via the valve and quickly passes through the
`actuator orifice where primary atomization occurs.
`Characterization of aerosol clouds emitted from
`
`pMDIs is difficult due to the dynamics of the
`atomization process [9]. Individual droplets and the
`surrounding environment of the droplets change
`rapidly in terms of size, velocity, and position after
`being emitted from the actuator orifice. The inter-
`action of this dynamic aerosol plume with the
`geometry of the mouth and airways determines the
`extent of oral and lung deposition. Mechanisms of
`deposition include inertial impaction, sedimentation,
`diffusion, interception, and electrostatic precipitation
`[10]. The relative importance of each deposition
`mechanism on an individual droplet therefore de-
`
`pends on a multitude of factors such as droplet size,
`velocity, evaporation rates, and the anatomical fea-
`tures of the patient’s airways.
`
`3. Measures of the performance of MDIs
`
`3.1. Performance of current pllfll devices
`
`from CFC to HFC
`The transition of pMDIs
`systems has provided an opportunity for the pharma-
`ceutical industry to re-evaluate inhaler performance.
`From one point of view, the task might be seen as
`showing equivalence between the CFC systems and
`their replacements. This approach to the transition is
`less problematic in terms of regulatory approval and
`is therefore more cost effective in the short term.
`
`However, recent phenomenological reports compar-
`ing CFC systems to HFA systems suggest that a
`‘detuning’ of the HFA formulations is necessary to
`attain regulatory equivalence between the systems
`[11]. Thus,
`the opportunity has arisen to improve
`inhaler performance using a systematic approach of
`evaluating the effect of formulation and device
`factors on pMDI performance.
`
`3
`
`

`
`810
`
`H.D.C. Smyth / Advanced Drug Delivery Reviews 55 (2003) 807-828
`
`Inhaled therapy for treatment of asthma is targeted
`such that
`the active drug substance is delivered
`topically. In terms of bronchodilators this has the
`primary advantage of rapid onset of action, but also
`minimizes the risk of unwanted systemic side—ef—
`fects, which is particularly important
`in corticos-
`teriod therapy. However, simply targeting the lung as
`an organ may not result in optimal therapy, as the
`target receptors for a specific drug may be located in
`particular regions of the lung [12]. Furthermore,
`none of the available anti-asthma drugs are metabo-
`lized in the lungs
`[13]. Consequently,
`the drug
`delivered to the lung will eventually reach the
`systemic circulation and contribute to systemic ac-
`tivity. Thus, the effectiveness of inhaled therapy for
`topical diseases such as asthma depends on the
`device to deliver the correct dose of active drug
`substance to the site of action with minimal deposi-
`tion to other regions that may contribute to unwanted
`side—effects. As a result, measurement and assess-
`ment of inhaled drug delivery is very complex due to
`the multitude of variables that contribute to varia-
`
`tions in delivery such as drug formulation, delivery
`device, administration skill, breathing pattern, and
`lung pathology/ anatomy [14]. Also,
`the optimal
`outcome may not be defined identically between
`scientists, clinicians, patients, regulatory agencies,
`and those paying for the cost of treatment [14].
`
`3.2. In vivo determination of pMDI performance
`
`Measures of respiratory drug delivery performance
`include pharrnacokinetic/pharmacodynamic investi-
`gations,
`in vivo measures,
`in vitro tests, and also
`mathematical models of deposition. All measures are
`important
`ir1 optimizing the development of drug
`formulations and delivery devices for the appropriate
`clinical outcome. Pharmacokinetic/pharrnacodynam-
`ic studies involve measuring plasma drug/physiolog-
`ical marker concentrations and correlating them with
`clinical efficacy and toxicity. This approach allows
`elucidation of the relationship between drug delivery
`and efficacy/toxicity, but may be complicated by
`assay limits,
`local metabolism and gastrointestinal
`absorption. Used in conjunction with in vitro and
`scintigraphic
`studies,
`specific
`characteristics of
`formulation and device development can be related
`to the clinical outcome. Scintigraphic studies allow
`
`visualization of regional lung deposition by incorpo-
`ration of a garnma-radiating nuclide into the formula-
`tion [15]. The site and quantitative amount of
`deposition can be calculated, making this technique
`integral to device and formulation development as
`well as being useful in drug targeting studies [15].
`Used alone, however, this technique does not mea-
`sure clinical efficacy or
`toxicity. A number of
`clinical outcomes are also essential for evaluation of
`
`inhaled therapy. Spirometry, peak expiratory flow
`(PEF), bronchoprovocation testing, measurement of
`inflammation markers, toxicity, quality of life mea-
`sures, and epidemiological studies have all been
`widely used to evaluate the performance of forrnula—
`tions and devices
`in asthma therapy [14]. The
`performance of formulations and devices for bron-
`chodilator delivery to the lung can be assessed using
`changes in lung function (such as spirometry and
`PEF). However, there are no such immediate mea-
`sures to determine the performance of devices con-
`taining anti-inflammatory agents for the prevention
`of asthma. In these cases, the clinical response to
`regional lung delivery may be assessed using tissue
`biopsy and bronchoalveolar lavage samples or in-
`direct measures that
`include urinary and serum
`cortisol levels [16,17].
`
`3.3. Models and mechanisms of deposition within
`the airways
`
`The importance of in vitro measurements in pMDI
`development and optimization can be appreciated
`with an understanding of the mechanisms of particle
`deposition in the lungs. As already mentioned,
`inertial impaction, sedimentation, diffusion, intercep-
`tion, and electrostatic precipitation are the significant
`mechanisms that dictate where a droplet or particle
`will deposit in the airways. Inertial impaction in the
`lung occurs when particles of sufficient momentum
`(a product of mass and velocity) are unable to follow
`the curved streamlines of air within the airways
`during inhalation due to significant centrifiigal forces
`[18]. Sedimentation of particles within the airways is
`related to particle mass and residence times [19].
`Deposition via diffusion is often a small but some-
`times significant mechanism, particularly when par-
`ticle size is sufficiently small [10]. Significant depo-
`sition via interception occurs when the dimensions of
`
`4
`
`

`
`H.D.C. Smyth / Advanced Drug Delivery Reviews 55 (2003) 807-828
`
`811
`
`the anatomic spaces of the airways become compar-
`able to those of the particle [10]. Finally, electro-
`static precipitation may sometimes occur when
`charged particles, typically charged during atomiza-
`tion, are electrostatically attracted by a charge of
`opposite sign. Such charges may already be present
`on the walls of the device or airways or may be
`induced by the charged particle itself [10]. Data
`indicate that small particles (<1 pm) are most
`influenced, but the overall significance is unknown
`for pMDIs [18]. The extent and location of deposi-
`tion of pMDI aerosols can be modeled empirically or
`by using mechanistically based approaches
`[19].
`Despite being over-simplifications of aerosol deposi-
`tion, these models are very useful in demonstrating
`the effect of certain variables influencing deposition
`in the lung. Although there are a multitude of models
`available in the literat11re [l9,20], each predicts that
`particle diameter and airway diameter have the
`potential
`to influence particle deposition the most
`[19]. Thus, given that particle diameter in under
`more control
`in terms of formulation and device
`
`selection, it is generally accepted that particle size is
`the single most important parameter in pharmaceu-
`tical aerosol delivery.
`
`3.4. In vitro measures ofpMDI performance
`
`Given the complexity of the interactions of aerosol
`clouds emitted by pMDIs and the airways,
`it
`is
`difficult
`to define simple measures of device and
`formulation performance. However, in vitro methods
`have been developed that measure particle size and
`emitted dose characteristics of aerosol drug delivery
`devices. Both theoretical calculations derived from
`models such as those mentioned above and ex-
`
`perimental deposition studies using stable monodis—
`perse aerosols suggest particle size data can be used
`as an estimate of aerosol deposition efficiency [10].
`These predictions indicate that particles greater than
`around 6 pm will be deposited in oropharyngeal
`regions and will not enter the lung [10], while
`particles that traverse the pharynx and upper airways
`are generally less than 6 pm [21]. In addition to
`estimation of deposition characteristics, particle size
`measurements provide simple measures of quality
`control
`for pMDIs and enable a comparison of
`devices and formulations [22—25]. The most com-
`
`mon measure used is the mass median aerodynamic
`diameter (MMAD), which is a statistical measure of
`the aerodynamic size of the aerosol and represents
`the diameter that divides the particle size distribution
`into two halves with respect to mass (i.e. 50% of the
`mass lies in particles above and below the MMAD).
`MMAD is most often measured using inertial impac-
`tion particle size techniques such as cascade impac-
`tion or via multistage liquid impingement analytical
`devices. These methods have been adopted as phar-
`macopoeial standards for various devices, including
`the pMDI [26,27]. Various systems exist for these
`types of particle size techniques, which often yield a
`different quantity and quality of information. Conse-
`quently, data must be carefully analyzed for mean-
`ingful interpretations to be made [28]. In addition to
`MMAD, other parameters obtained from these par-
`ticle sizing techniques that may assist in prediction
`and assessment of
`lung deposition include the
`geometric standard deviation (GSD),
`fine particle
`fraction (FPF), and others
`such as non—balistic
`fraction (NBF) [29].
`Other in vitro methods have been employed to
`evaluate pMDI performance. These include spray
`pattern and geometry measurements
`[30], plume
`velocity calculation [31], spray temperature determi-
`nations [32], and spray force measurements [32].
`
`3.5. Relevance of in vitro and in vivo measures of
`performance
`
`Analyses of lung deposition of particles from a
`wide variety of sources have confirmed that
`the
`extent of particle deposition within the respiratory
`tract
`is related to aerodynamic particle size [33].
`However, there is also evidence to suggest that fine
`particle fractions and MMADs alone do not provide
`sufficient information to predict the respirability of
`aerosols [29,34]. With pMDIs, significant deviation
`of in vitro results from in vivo experiments appears
`to result from the unrealistic inlet tubes or throats
`
`that are used in most cascade impaction and liquid
`impinger systems [34]. It has been suggested that in
`vitro and in vivo correlations can be improved by
`using anatomic throats [35], various breathing pat-
`terns [14], and also by incorporation of other factors
`such as non—ballistic fraction and GSD into the
`
`model rather than MMADs alone [29]. Depending on
`
`5
`
`

`
`812
`
`H.D.C. Smyth / Advanced Drug Delivery Reviews 55 (2003) 807-828
`
`the goal of the particle size testing these correlations
`may be relevant. Particle size measurements are
`important for quality control, comparisons of forrnu—
`lations and devices, and can also be used to indicate
`relative amounts of lung deposition. They cannot be
`used, however,
`to demonstrate clinical efficacy.
`Ultimately pharmacodynamic and clinical studies
`will be required to demonstrate equivalence or
`efficacy and safety of reformulated products with
`dosage regimen changes.
`
`3.6. Strategies for pMDI optimization and
`improving performance
`
`Unfortunately, a great deal of device design and
`formulation development of pMDIs has been per-
`formed empirically. A satisfactory clinical response
`matched with low apparent systemic adverse effects
`have been interpreted as successful delivery markers
`for anti-asthma medications delivered via the inhala-
`
`tion route. As a consequence, pMDIs are not opti-
`mized systems. Often—cited disadvantages of this
`type of device are the low efficiency of lung
`deposition (often only 20% of the emitted dose
`reaches the lungs) [36] and poor inhaler technique
`[37]. These outcomes are inter-related and must both
`be addressed to improve the performance of pMDI
`device/ formulation combinations. What is required is
`a systematic evaluation of the influence of formula-
`tion factors and device parameters on clinical out-
`comes. However, direct in vivo evaluations of the
`effect of these formulation and device factors would
`
`involve large-scale and costly clinical experiments.
`The multitude of formulation variables and arrays of
`different device parameters would mean that even
`efliciently designed statistical experiments would
`involve massive matrices of controlled experiments.
`Thus more investigations are required that bridge
`different levels of testing such as in vitro measure-
`ments and in vivo outcomes, and in vivo assessments
`and clinical outcomes.
`
`This has been highlighted in recent phenomeno-
`logical studies relating clinical effect of inhaled
`corticosteroids to formulation factors [38]. Reforrnu—
`lation of beclomethasone diproprionate as a solution
`formulation in HFA propellants has resulted in
`greater peripheral lung deposition of the medication
`relative to the CFC equivalents that were formulated
`
`lung
`as suspensions [38]. The greater peripheral
`deposition and reduced oropharyngeal deposition,
`however, does not correspond to an equivalent
`increase in therapeutic effect [39]. The HFC forrnu—
`lation resulted in approximately six times more
`peripheral drug deposition than the CFC product, but
`only twice the therapeutic effect, raising the issue of
`required particle size and target region of the lung
`for optimal inhaled corticosteriod therapy. The par-
`ticle size that maximizes the therapeutic ratio of a
`molecule is likely to be different for a beta-adren-
`ergic agonist than for an inhaled corticosteriod. A
`greater understanding of this relationship will be
`required if we are to achieve improved drug targeting
`with future inhalers [40].
`Although particle diameter and efficiency of dis-
`persion are the primary variables under the control of
`the forrnulator of pMDIs for directing particle depo-
`sition in the airways, inspiratory flow rate is also an
`important factor to consider [41]. Inspiratory flow
`rate can influence the dose emitted firom an inhaler,
`amount
`inhaled, oropharyngeal deposition, and re-
`gional lung deposition of inhaled medications [41].
`Future designs of pMDIs should account for this and
`integrate systems that emit an aerosol at the correct
`time and flow rate. Some systems already incorpo-
`rate breath monitoring and breath-activated aerosols
`[42,43]. In addition, other patient-related factors such
`as inhalation techniques, compliance, and misuse
`need to be addressed. With current pMDI systems,
`effective use is strongly technique dependent [44,45].
`Development of a more ‘forgiving’ formulation and
`device may help to reduce the strong dependence of
`delivery success on patient-related factors
`[38]
`(Table 1).
`
`4. Physicochemical characteristics of HFA
`propellants
`
`Initial screening of alternative propellants iden-
`tified HFA propellants as
`likely candidates
`for
`replacing CFCs. They appeared to have the neces-
`sary physical properties: do not deplete ozone, non-
`flarnmable, sufficient vapor pressures, and,
`impor-
`tantly, they appeared to be as non—toxic as the CFC
`counterparts
`[46]. Although toxicological
`studies
`demonstrated the equivalency of HFAs to CFCs [47],
`
`6
`
`

`
`H.D.C. Smyth / Advanced Drug Delivery Reviews 55 (2003) 807-828
`
`813
`
`Table 1
`In vivo and m vitro endpoints of pMDI performance
`
`Measure
`type
`
`In vivo
`
`In vitro
`
`Endpoint
`
`Methods
`
`Pharmacokir1etic/
`pharrnacodynarnic
`investigations
`
`Scintigraphic
`studies
`
`Clinical
`measurements
`
`Particle size
`measurement
`
`Spray
`characterization
`
`Formulation
`performance
`
`Plasma drug concentrations
`Physiological markers
`Toxicity measures
`
`Gamma scintigraphy
`Single photon emission computed
`tomography (SPECT)
`Positron emission tomography (PET)
`
`Spirometry and peak expiratory
`flow (PEF)
`Bronchoprovocation testing
`Inflammation marker measurement
`Tissue biopsy
`Bronchoalveolar lavage
`Toxicity
`Quality of life measures
`Epidemiological studies
`
`Inertial methods
`Optical methods
`
`Spray force and velocity
`Plume temperature
`Spray patterns and geometry
`
`Suspension/solution stability
`Emitted dose
`
`Refs.
`
`[17,l29,130]
`
`[131]
`
`[132]
`[133]
`
`[134,135]
`[136,137]
`[138]
`[139]
`[140]
`[141]
`[142]
`[143]
`
`[144,145]
`[31,58]
`
`[32]
`[31]
`[30]
`
`[59,84]
`[86]
`
`it was quickly realized that HFA propellants were
`not ‘drop in’ replacements for CFCS in pMDIs [48].
`While CFCs were good solvents for a number of
`drug candidates [46], HFA propellants are generally
`poor solvents for many anti-asthma drugs and excipi-
`ents currently available for use in pMDIs [5]. These
`
`changes in the physicochemical characteristics of the
`propellant systems impact pMDI design and formula-
`-
`-
`-
`tion. A summary of_ the_phys1cochem1cal nature of
`HFA propellants, primarily 1,1,1,2-tetrafluoroethane
`(HFA 134a), is presented here.
`The structures of common propellants are shown
`in Fig. 2. The physical properties of these propellants
`and other alternatives have been widely reported in
`recent years [5,46,48—5l]. The physical properties of
`each propellant (Table 2) can be related to details of
`the chemical structure and experimental
`investiga-
`tions of propellant physicochemical behavior.
`HFA 134a and 227ea both have high vapor
`pressures and low boiling points. These properties
`
`CI
`
`C'—C—F
`(Ll
`CFC 11
`
`F
`
`F
`\
`CH -CZF
`2
`l
`F
`
`HFA 134a
`
`F
`
`0'
`
`F
`C
`I
`CI
`CFC 12
`
`F
`
`F
`
`F—C—C—F
`('3'
`[I
`CFC 114
`
`F
`
`F
`\C/F
`F\C/
`\
`/ HC/ \
`F
`
`F
`
`F
`HFA 227ea
`
`Fig. 2. Chemical structures of common CFC and HFA pro-
`pellants.
`
`are caused by the strong electronegative repulsive
`interactions between HFA molecules rather
`than
`
`from non-polar characteristics [5]. In fact, HFAS are
`relatively polar. HFA 134a contains two electroposi—
`
`7
`
`

`
`814
`
`H.D.C. Smyth / Advanced Drug Delivery Reviews 55 (2003) 807-828
`
`Table 2
`Physicochemical properties of pMDI propellants
`
`Property
`
`CFC ll
`
`CFC 12
`
`CFC 114
`
`HFA 134a
`
`HFA 227ea
`
`Thermodynamic
`Boiling point (°C)
`Vapor pressure (kPa)
`Enthalpy vap. (kJ/mol)
`
`Polarity
`Dielectric constant
`Dipole moment
`Induced polarization
`(m3 mol’2X105)
`Solubility parameter
`(Hildabrand units)
`Kauri—Butanol value
`Log P (oct/water)
`Water solubility (ppm)
`
`Liquid phase
`Density (g/cms)
`Viscosity (mPa~s)
`Surface tension (mN/m2)
`
`24
`89
`25.1
`
`2.3
`0.45
`
`2.8
`
`7.5
`60
`2.0
`130
`
`1.49
`0.43
`18
`
`-30
`566
`17.2
`
`2.1
`0.51
`
`2.3
`
`6.1
`18
`2.2
`120
`
`1.33
`0.20
`9
`
`4
`182
`22.1
`
`2.2
`0.58
`
`3.2
`
`6.4
`12
`2.8
`110
`
`1.47
`0.30
`11
`
`-26
`572
`18.6
`
`9.5
`2.1
`
`6.1
`
`6.6
`9
`1.1
`2200
`
`1.23
`0.21
`8
`
`-16
`390
`19.6
`
`4.1
`1.2
`
`6.1
`
`6.2
`13
`2.1
`610
`
`1.42
`0.27
`7
`
`tive protons due to the strongly electron—withdrawing
`effects of the multiple fluorine atoms attached to the
`carbon backbone [5]. Similarly, HFA 227ea has one
`asymmetrical electropositive proton. This polar na-
`ture, relative to CFCs,
`is reflected in the dipole
`moments
`and
`dielectric
`constants.
`The
`
`polarizabilities of HFAs relative to the CFC pro-
`pellants are much lower. This illustrates the strong
`electronegative nature of the fluorine atoms and the
`strength with which associated electrons are held [5].
`An increase in polarizability is associated with
`increased intermolecular attraction. As a conse-
`
`quence, HFAs are relatively polar but have very low
`intermolecular attractive forces when compared with
`CFCs systems.
`impact of these differences
`The most apparent
`between HFAs and CFCs is insufficient capacity to
`solubilize traditional surfactants [48,49]. Thus, it is
`surprising that relatively few published studies have
`investigated the solubilization capacity of HFAs
`[49,50,52,53]. Using co-solvents such as ethanol or
`other solubilizing agents,
`the solubility of surfac-
`tants, drug substances, and excipients in the HFA
`propellants can be increased [5,53,54]. However, if
`surfactants are required in a co—solvent—free system
`(for stability of a suspension formulation), alternative
`surfactants need to be used [5,55].
`
`Many pMDl formulations contain multi-compo-
`nent mixtures of propellant blends and co-solvents.
`Therefore,
`it is important to consider the physico-
`chemical characteristics of these systems in order to
`understand the effects of formulation variables on
`
`product performance. Density, molar volume, and
`vapor pressure are thermodynamic parameters that
`assess intermolecular forces within mixtures and are
`
`readily measurable in pMDls [56]. These can be
`evaluated by comparing theoretical calculations of
`ideal mixtures with the experimental measurements
`of real solutions. Propellant systems that have been
`studied with regard to propellant-driven pMDls
`include HFA l34a/ ethanol, HFA 227ea/ ethanol, and
`HFA 134a/HFA 227ea mixtures [5,53,56]. These
`three miscible components may allow the forrnulator
`to select appropriate densities (for suspension stabili-
`ty), solubility characteristics (for solution and sus-
`pension formulations), and also modify the emitted
`particle size via non-volatile composition effects
`[5,53]. Tzou [56]
`recently demonstrated that
`the
`observed densities of HFA l34a/ ethanol and HFA
`
`134a/HFA 227ea mixtures closely matched ideal
`mixture predictions. Vervaet and Byron [5] reported
`similar result for HFA l34a/ ethanol, HFA 227ea/
`ethanol, and HFA 134a/HFA 227ea mixtures. How-
`ever, when vapor pressure behavior was investigated,
`
`8
`
`

`
`H.D.C. Smyth / Advanced Drug Delivery Reviews 55 (2003) 807-828
`
`815
`
`positive deviations from Raoult’s law were observed
`with HFA 134a/ethanol and HFA 227ea/ethanol
`
`mixtures [5,53,56]. Blends of HFA 134a and HFA
`227ea did not show any significant deviation from
`theory [56,57]. Vervaet and Byron explain the appar-
`ent atypical observation of deviations from Raoult’s
`law with no accompanied changes in density as a
`surface phenomenon [5]. HFA propellants have a
`higher affinity for
`the gas—liquid interface than
`ethanol, which,
`in turn,
`is surrounded by HFA
`molecules.
`In addition, positive deviations
`from
`Raoult’s law with HFA/ ethanol mixtures indicate
`
`that the intermolecular forces between the compo-
`nents of the mixture (i.e. between ethanol and HFA
`molecules) are less than that between molecules of
`the pure constituents [53,56]. It was suggested that
`this positive deviation in vapor pressure will allow
`forrnulators to use higher concentrations of ethanol
`(for improved solubility) than would be predicted by
`ideality without detrimental effects on droplet size or
`aerosolization [5]. However, recent evidence sug-
`
`Table 3
`Common marketed pMDIs and their general composition
`
`Therapeutic
`group
`
`Bronchodilators
`Maxiair
`Maxair Autohalor
`Proventil
`Proventil HFA
`Tomalate
`
`Ventolin
`Ventolin HFA
`
`Corticosteriods
`Aerobid
`
`Drug
`
`Pirbuterol acetate
`Pirbuterol acetate
`Albuterol sulfate
`Albuterol sulfate
`Bitolterol mesylate
`
`Albuterol sulfate
`Albuterol sulfate
`
`Flunisolide
`
`Azmacort
`
`Triamcinolone acetonide
`
`Beclovent
`Becotide 100
`Flovent 44, 110, 220
`QVAR 50, 100
`QVAR Autohaler 50, 100
`Vanceril
`
`Beclomethasone dipropionate
`Beclomethasone dipropionate
`Fluticasone propionate
`Beclomethasone dipropionate
`Beclomethasone dipropionate
`Beclomethasone dipropionate-
`trichlorofluoromethane clathrate
`
`Other anti-inflammatory
`Intal
`Tilade
`
`Cromolyn sodium
`Nedocromil sodium
`
`gests that increasing ethanol concentrations by 10%
`w/w will have a significant impact on MMAD and
`droplet size, as discussed below [53,58].
`
`5. Influence of formulation variables
`
`The influence of formulation variables will now be
`
`discussed in the context of these physicochemical
`characteristics and the above-mentioned measures of
`
`performance.
`
`5.1. Solution or suspension formulations?
`
`The active drug substance in pMDIs is either
`suspended or dissolved in the propellant or pro-
`pellant mixture (Table 3). Partial solubility of sus-
`pended drug is undesirable as it
`leads to crystal
`growth via a process known as Ostwald ripening
`[59]. As a consequence, particle size changes and
`irregular emitted doses may result. However, if a
`
`Surfactants /
`excipients
`
`Propellant
`system
`
`Formulation
`type
`
`Sorbitan trioleate
`Sorbitan trioleate
`Oleic acid
`Oleic acid
`Ascorbic acid,
`saccharin, menthol
`Oleic acid
`None
`
`Sorbitan trioleate,
`menthol
`
`Oleic acid
`Oleic acid
`
`CFC 11, CFC 12
`CFC 11, CFC 12
`CFC 11, CFC 12
`HFA 134a, ethanol
`38% w/w ethanol,
`CFC 11, CFC 12
`CFC 11, CFC 12
`HFA 134a
`
`CFC11, CFC 12,
`CFC 114
`CFC 12,
`1% W/W ethanol
`CFC 11, CFC 12
`CFC 11, CFC 12
`CFC 11, CFC 12
`HFA 134a, ethanol
`HFA 134a, ethanol
`CFC 11, CFC 12
`
`Suspension
`Suspension
`Suspension
`Suspension
`Solution
`
`Suspension
`Suspension
`
`Suspension
`
`Suspension
`
`Suspension
`Suspension
`Suspension
`Solution
`Solution
`Suspension
`
`Sorbitan trioleate
`Sorbitan trioleate
`
`CFC 11, CFC 12
`CFC 11, CFC 12
`
`Suspension
`Suspension
`
`9
`
`

`
`816
`
`H.D.C. Smyth / Advanced Drug Delivery Reviews 55 (2003) 807-828
`
`solution formulation is chosen the drug must have
`sufficient solubility to allow therapeutic doses to be
`delivered with a few actuations (usually two) of the
`device. Thus, careful selection of drug form and
`propellant system from solubility studies and com-
`patibility investigations is required in pMDI develop-
`ment. In suspension systems, the predominant factor
`limiting the minimum emitted droplet size is the size
`of the suspended particles that will be contained
`within aerosol droplets [60—62]. In solution systems,
`however, the droplet size is primarily governed by
`factors
`such as
`the non-volatile fraction, vapor
`pressure, actuator design, and the physicochemical
`nature of the liquid formulation [62]. With a limited
`range of propellant systems available, the selection
`of a solution or a suspension system may be by
`necessity rather than by choice. However, it should
`be recognized that each system has distinct forrnula—
`tion requirements and may have certain advantages
`for a given drug substance. Suspensions require very
`low solubility of the drug in the formulation. This
`usually results in good chemical stability of the drug
`[63]. However, aggregation and rapid flocculation of
`suspension systems may require addition of stabiliz-
`ing excipients such as surfactants. In solution sys-
`tems, high solubility and good stability of drug in the
`propellant system is required. Most current forrnula—
`tions are suspension systems containing micronized
`drug with particle sizes between 2 and 5 um [64].
`
`5.2. Eflect of vapor pressure
`
`5.2.1. CFC studies
`
`One of the first investigations of the influence of
`formulation factors on particle size was performed
`by Polli and co—workers
`[65]. Using suspension
`systems in CFC propellant blends, the effect of a
`number of formulation factors, including vapor pres-
`sure, was studied. Increased vapor pressure resulted
`in decreased aerosol particle size, ultimately to the
`size of the suspended particle when the vapor
`pressure of the formulation was 77 psig. When
`temperature was used to modify vapor pressure,
`similar trends were als

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket