throbber
SPINE Volume 26, Number 6, pp E122–E129
`©2001, Lippincott Williams & Wilkins, Inc.
`
`Load-Sharing Between Anterior and Posterior Elements
`in a Lumbar Motion Segment Implanted With an
`Artificial Disc
`
`Andrew P. Dooris, BS,* Vijay K. Goel, PhD,* Nicole M. Grosland, PhD,*
`Lars G. Gilbertson, PhD,† and David G. Wilder, PhD*
`
`Study Design. A nonlinear three-dimensional finite el-
`ement model of the osteoligamentous L3–L4 motion seg-
`ment was used to predict changes in posterior element
`loads as a function of disc implantation and associated
`surgical procedures.
`Objectives. To evaluate the effects of disc implantation
`on the biomechanics of the posterior spinal elements
`(including the facet joints, pedicles, and lamina) and on
`the vertebral bodies.
`Summary of Background Data. Although several arti-
`ficial disc designs have been used clinically, biomechani-
`cal data—particularly the change in loads in the posterior
`elements after disc implantation—are sparse.
`Methods. A previously validated intact finite element
`model was implanted with a ball-and-cup–type artificial
`disc model via an anterior approach. The implanted
`model predictions were compared with in vitro data. To
`study surgical variables, small and large windows were
`cut into the anulus, and the implant was placed anteriorly
`and posteriorly within the disc space. The anterior longi-
`tudinal ligament was also restored. Models were sub-
`jected to either 800 N axial compression force alone or to
`a combination of 10 N-m flexion– extension moment and
`400 N axial preload. Implanted model predictions were
`compared with those of the intact model.
`Results. Facet loads were more sensitive to the antero-
`posterior location of the artificial disc than to the amount
`of anulus removed. Under 800 N axial compression, im-
`planted models with an anteriorly placed artificial disc
`exhibited facet loads 2.5 times greater than loads ob-
`served with the intact model, whereas posteriorly im-
`planted models predicted no facet loads in compression.
`Implanted models with a posteriorly placed disc exhibited
`greater flexibility than the intact and implanted models
`with anteriorly placed discs. Restoration of the anterior
`longitudinal
`ligament reduced pedicle stresses,
`facet
`loads, and extension rotation to nearly intact levels.
`Conclusions. The models suggest that, by altering
`placement of the artificial disc in the anteroposterior di-
`rection, a surgeon can modulate motion-segment flexural
`
`From the *Iowa Spine Research Center, Departments of Biomedical
`Engineering and Orthopaedics, University of Iowa, Iowa City, Iowa,
`and the †Musculoskeletal Research Center, Department of Orthopae-
`dic Surgery, University of Pittsburgh, Pittsburgh, Pennsylvania.
`Acknowledgment date: March 7, 2000.
`First revision date: July 14, 2000.
`Acceptance date: August 22, 2000.
`Supported by the Sofamor Danek Group, Memphis, TN, The Iowa
`Spine Research Foundation, Iowa City, IA, and The Whitaker Foun-
`dation, Rosslyn, VA.
`The authors thank Dwight T. Todd and Malcolm H. Pope, PhD, for
`their assistance.
`Device status category: 7,9.
`Conflict of interest category: 16.
`
`E122
`
`stiffness and posterior load-sharing, even though the
`specific disc replacement design has no inherent rota-
`tional stiffness. [Key words: artificial disc, biomechan-
`ics, facets, finite element model, lumbar spine] Spine
`2001;26:E122–E129
`
`In many developed countries, the management of low
`back pain is a prevalent problem to the clinician.44 Dis-
`ability of patients with low back pain costs several billion
`dollars in the United States annually because of lost pro-
`ductivity and treatment costs,22 affecting millions of in-
`dividuals.19 Almost three quarters of the United States
`population has at one time experienced back pain, and
`approximately 4% of the population requires surgical
`intervention. As the population ages, the problem will
`certainly grow.12,26,43
`Fusion of painful segments (spinal arthrodesis) is of-
`ten the goal of spinal surgery (25% of 280,000 surgical
`interventions). Fusion typically is done to prevent or cor-
`rect deformity, to stabilize the spine after trauma or
`pathologic destruction, or to eliminate painful move-
`ment of the spinal segments. Despite advances in surgical
`techniques, fusions are neither fully successful nor with-
`out problematic side-effects (e.g., pseudoarthroses and
`donor-site pain). Because fusion— by design—limits mo-
`tion of the fused segment(s), some investigators16,30 be-
`lieve that fusion may induce degenerative changes in the
`neighboring segments, often necessitating additional fu-
`sion surgery.
`To avoid this cascade, some investigators are pursuing
`alternative treatment measures. Tissue engineering ap-
`proaches include regeneration of the nucleus through in-
`troduction of growth factors or gene therapy, as well as
`cell transplantation and scaffolding. Another approach,
`already used clinically, is to mimic normal disc function
`through artificial mechanical means to circumvent bio-
`mechanical changes normally attributed to rigid instru-
`mentation and fusion. Several artificial disc designs have
`been proposed in an effort to reinstate stabilized physio-
`logic motion in the lumbar spine to manage various ail-
`ments that normally indicate spinal fusion.
`A few disc designs have undergone clinical trials with
`limited published results.11,23,31 Likewise, some of the
`disc designs have been evaluated biomechani-
`cally.3,8,25,28 These designs vary greatly in their ap-
`GLOBUS MEDICAL, INC.
`proach to restoring normal disc joint function. One re-
`EXHIBIT 1013
`curring difficulty with each implant lies in determining
`IPR2015-to be assigned
`(Globus v. Flexuspine)
`1 of 9
`
`

`
`SPINE Volume 26, Number 6, pp E122–E129
`©2001, Lippincott Williams & Wilkins, Inc.
`
`Load-Sharing Between Anterior and Posterior Elements
`in a Lumbar Motion Segment Implanted With an
`Artificial Disc
`
`Andrew P. Dooris, BS,* Vijay K. Goel, PhD,* Nicole M. Grosland, PhD,*
`Lars G. Gilbertson, PhD,† and David G. Wilder, PhD*
`
`Study Design. A nonlinear three-dimensional finite el-
`ement model of the osteoligamentous L3–L4 motion seg-
`ment was used to predict changes in posterior element
`loads as a function of disc implantation and associated
`surgical procedures.
`Objectives. To evaluate the effects of disc implantation
`on the biomechanics of the posterior spinal elements
`(including the facet joints, pedicles, and lamina) and on
`the vertebral bodies.
`Summary of Background Data. Although several arti-
`ficial disc designs have been used clinically, biomechani-
`cal data—particularly the change in loads in the posterior
`elements after disc implantation—are sparse.
`Methods. A previously validated intact finite element
`model was implanted with a ball-and-cup–type artificial
`disc model via an anterior approach. The implanted
`model predictions were compared with in vitro data. To
`study surgical variables, small and large windows were
`cut into the anulus, and the implant was placed anteriorly
`and posteriorly within the disc space. The anterior longi-
`tudinal ligament was also restored. Models were sub-
`jected to either 800 N axial compression force alone or to
`a combination of 10 N-m flexion– extension moment and
`400 N axial preload. Implanted model predictions were
`compared with those of the intact model.
`Results. Facet loads were more sensitive to the antero-
`posterior location of the artificial disc than to the amount
`of anulus removed. Under 800 N axial compression, im-
`planted models with an anteriorly placed artificial disc
`exhibited facet loads 2.5 times greater than loads ob-
`served with the intact model, whereas posteriorly im-
`planted models predicted no facet loads in compression.
`Implanted models with a posteriorly placed disc exhibited
`greater flexibility than the intact and implanted models
`with anteriorly placed discs. Restoration of the anterior
`longitudinal
`ligament reduced pedicle stresses,
`facet
`loads, and extension rotation to nearly intact levels.
`Conclusions. The models suggest that, by altering
`placement of the artificial disc in the anteroposterior di-
`rection, a surgeon can modulate motion-segment flexural
`
`From the *Iowa Spine Research Center, Departments of Biomedical
`Engineering and Orthopaedics, University of Iowa, Iowa City, Iowa,
`and the †Musculoskeletal Research Center, Department of Orthopae-
`dic Surgery, University of Pittsburgh, Pittsburgh, Pennsylvania.
`Acknowledgment date: March 7, 2000.
`First revision date: July 14, 2000.
`Acceptance date: August 22, 2000.
`Supported by the Sofamor Danek Group, Memphis, TN, The Iowa
`Spine Research Foundation, Iowa City, IA, and The Whitaker Foun-
`dation, Rosslyn, VA.
`The authors thank Dwight T. Todd and Malcolm H. Pope, PhD, for
`their assistance.
`Device status category: 7,9.
`Conflict of interest category: 16.
`
`E122
`
`stiffness and posterior load-sharing, even though the
`specific disc replacement design has no inherent rota-
`tional stiffness. [Key words: artificial disc, biomechan-
`ics, facets, finite element model, lumbar spine] Spine
`2001;26:E122–E129
`
`In many developed countries, the management of low
`back pain is a prevalent problem to the clinician.44 Dis-
`ability of patients with low back pain costs several billion
`dollars in the United States annually because of lost pro-
`ductivity and treatment costs,22 affecting millions of in-
`dividuals.19 Almost three quarters of the United States
`population has at one time experienced back pain, and
`approximately 4% of the population requires surgical
`intervention. As the population ages, the problem will
`certainly grow.12,26,43
`Fusion of painful segments (spinal arthrodesis) is of-
`ten the goal of spinal surgery (25% of 280,000 surgical
`interventions). Fusion typically is done to prevent or cor-
`rect deformity, to stabilize the spine after trauma or
`pathologic destruction, or to eliminate painful move-
`ment of the spinal segments. Despite advances in surgical
`techniques, fusions are neither fully successful nor with-
`out problematic side-effects (e.g., pseudoarthroses and
`donor-site pain). Because fusion— by design—limits mo-
`tion of the fused segment(s), some investigators16,30 be-
`lieve that fusion may induce degenerative changes in the
`neighboring segments, often necessitating additional fu-
`sion surgery.
`To avoid this cascade, some investigators are pursuing
`alternative treatment measures. Tissue engineering ap-
`proaches include regeneration of the nucleus through in-
`troduction of growth factors or gene therapy, as well as
`cell transplantation and scaffolding. Another approach,
`already used clinically, is to mimic normal disc function
`through artificial mechanical means to circumvent bio-
`mechanical changes normally attributed to rigid instru-
`mentation and fusion. Several artificial disc designs have
`been proposed in an effort to reinstate stabilized physio-
`logic motion in the lumbar spine to manage various ail-
`ments that normally indicate spinal fusion.
`A few disc designs have undergone clinical trials with
`limited published results.11,23,31 Likewise, some of the
`disc designs have been evaluated biomechani-
`cally.3,8,25,28 These designs vary greatly in their ap-
`proach to restoring normal disc joint function. One re-
`curring difficulty with each implant lies in determining
`
`2 of 9
`
`

`
`Load-Sharing in an Implanted Lumbar Motion Segment • Dooris et al
`
`E123
`
`the proper indications and contraindications for disc re-
`placement. Although some investigators view disc re-
`placement as a therapeutic treatment, others see it as a
`safeguard against further degenerative changes.
`After four decades of disc design investigation and
`experimentation, authors have proposed several sets of
`criteria encompassing clinical and nonclinical (e.g., bio-
`mechanical and biomaterial) issues.3,24,31 For example,
`because the replacement disc might last decades, the de-
`sign must minimize wear debris, possibly even more than
`total hip replacement designs.25,28 However, not all of
`the biomechanical parameters can be evaluated by using
`experimental protocols. Some parameters, such as bone
`stresses or joint forces, are difficult to analyze in cadaver
`studies. Cadaver studies involve an array of widely vary-
`ing parameters, many of which are difficult to control,
`characterize, or even measure.
`Not only should healthy posterior elements be re-
`quired before implantation, a disc replacement design
`must also preserve the posterior elements for long-term
`success. Degeneration of the posterior elements, espe-
`cially of the facet joints, has been found to be a source of
`low back pain by various clinical, biologic, and biome-
`chanical studies.1,2,4,5,9,10,13–15,18,32–37,39 – 42 If it is true
`that degeneration of the posterior elements results from
`disc disease in most joint degeneration sequelae,7 then it
`appears reasonable that current protocols would contra-
`indicate implantation with degenerated posterior ele-
`ments.11,17,23 Reports also suggest that in addition to
`preoperative degeneration, the posterior elements may
`be abnormally stressed by disc replacement, and there-
`fore implantation should be avoided.31
`In the present study, a computational investigation
`into the biomechanical changes brought on by disc im-
`plantation in a lumbar spine was undertaken. Because
`flexion and extension are primary displacement modes, a
`half-sagittal plane finite element model of the implanted
`motion segment was built for this analysis. Comparisons
`were made between the intact and the implanted func-
`tional spinal unit (FSU) models. To address issues relat-
`ing to surgical technique, model parameters were varied
`to simulate different anulus window sizes and anteropos-
`terior artificial disc placement.
`The disc implant studied in this investigation was de-
`signed by the Sofamor Danek Group (Sofamor Danek,
`Memphis, TN). This implant uses a tested tribologic de-
`sign adapted from hip replacements, including two,
`matching, ceramic ball and cup components (Figures 1
`and 2). Because the cup is shallow, and its curvature
`decreases near its perimeter, the joint is slightly “slop-
`py.” The contacting surfaces are made of polycrystalline
`alumina and are surrounded by a titanium sleeve, which
`is grooved and beaded for bone in-growth. With this
`implant design and computational model, the authors
`hypothesized that the implanted FSU would resemble the
`intact FSU under sagittal-plane loading, but that surgical
`technique would affect segment mechanics, including
`facet loading, rotational stiffness, and bone stresses.
`
`Figure 1. The Sofamor Danek (Sofamor Danek, Memphis, TN)
`artificial intervertebral disc. Ball and cup components are made of
`polycrystalline ceramic with a grooved and beaded titanium
`sleeve.
`
`Methods
`
`A two-part investigation was undertaken. An in vitro study21
`quantified changes in load-displacement behavior after disc im-
`plantation in fresh ligamentous spine specimens. A finite ele-
`ment (FE) model with the implanted disc was loaded in a man-
`ner similar to that used in the in vitro study protocol, and FE
`model predictions were compared with in vitro results for FE
`model validation. Subsequently, the implanted FE model was
`loaded under more strenuous conditions, and the results of
`loading the implanted and intact models were compared.
`
`In Vitro Experimentation. Seven fresh cadaver spines (L1
`through S1) were cleaned of muscle and connective tissue and
`prepared for loading by fixing the sacrum and inserting rigid
`crossbeams through the L1 vertebra. Three infrared light-
`emitting diodes (LEDs) were attached noncolinearly at each
`vertebral level and connected to an active optical tracking sys-
`tem (Selspot II System, Partille, Sweden). Pure bending mo-
`ments, from 0 N-m to 6 N-m, in increments of 1.5 N-m, were
`applied to L1 via a loading frame attached to the crossbeams.
`For each specimen, the intact spine was loaded in flexion and
`extension, and the three-dimensional displacements of each
`vertebral level at each loading increment were recorded
`simultaneously.
`
`Figure 2. Geometry of the Sofamor Danek (Sofamor Danek, Mem-
`phis, TN) artificial disc. The ball is eccentrically located to the
`posterior of the implant.
`
`3 of 9
`
`

`
`E124 Spine • Volume 26 • Number 6 • 2001
`
`Figure 3. Three-dimensional, nonlinear, finite element half-model
`of the intact L3–L4 motion segment. Moment loads were applied to
`a pair of crossbeams fixed to the superior endplate of L3. Nodes
`at the bottom of L4 were fixed.
`
`Surgery was performed to excise the anterior longitudinal
`ligament at L4 –L5, the anterior portion of the anulus, and the
`nucleus. The joint was distracted, and the ball and cup compo-
`nents of the Sofamor Danek artificial disc were inserted. The
`load-displacement characteristics then were recorded in flexion
`and extension to 6 N-m, as was done previously with the intact
`spine. The average and standard deviation of the motions were
`computed. The data for the implanted spine were compared
`
`with the those of the intact spine to assess the ability of the
`artificial disc to restore normal motion.
`
`Finite Element Analysis. The FE model used in the present
`investigation expands on a previously validated intact model
`produced from computed tomography scans.20,27 Three-
`dimensional geometry was made by layering border designs of
`1-mm–spaced scans. The model (Figure 3) included one FSU,
`L3–L4 with the entire L3 vertebral body, intervertebral disc,
`ligaments, and the superior portion of L4. All seven spinal
`ligaments were represented by nonlinear cable elements, and
`cortical and cancellous bone was represented by eight-node
`continuum elements. Softened, contact, unidirectional ele-
`ments mimicked cartilaginous facet joints. The anulus con-
`sisted of layers of fiber-reinforced continuum elements, with
`fiber orientations at 30° and 120° to the horizontal, and fiber
`modulus changing from the center to the periphery of the disc
`space. Finite element model details are provided in Table 1.
`The model was altered to mimic the anterior surgical ap-
`proach used to implant the Sofamor Danek disc. This required
`removal of the anterior longitudinal ligament and the anterior
`portion of the anulus. Nucleus pulposus elements also were
`also removed. Implant geometry was obtained from PRO/
`Engineer CAD drawings (PTC, Needham, MA) imported into
`MSC PATRAN (MSC Software, Los Angeles, CA) for meshing,
`obtaining accurate geometric proportions in a refined mesh.
`The mesh (Figure 4) then was exported to ABAQUS/Standard
`(HKS, Pawtucket, RI) for analysis. Surfaces were created for
`ball and socket contact and for contact in overextension and
`flexion (along the flat faces).
`Specific assumptions and approximations facilitated model
`generation and analysis. Although the actual implant has
`
`Table 1. Summary of Finite Element Models Listing Element Type, Quantity, and Properties for the Model Entities. The
`Changes in These Parameters Due to Surgery are Also Listed
`
`Implanted
`
`Element Type
`
`E (MPa), v
`
`100, 0.3
`1200, 0.3
`100, 0.3
`1200, 0.3
`3500, 0.25
`3500, 0.25
`
`0–20, 0.3
`0–20, 0.3
`0–19.5, 0.3
`0–12, 0.3
`0–15, 0.3
`0–33, 0.3
`0–58.7, 0.3
`
`1E–6, 0.5
`Ground Substance 4.5, 0.45
`Fibers 357–550, 0.3–0.5
`
`380, 0.26†
`114, 0.32‡
`
`4 of 9
`
`Total nodes
`Total elements
`Bone*
`L3 Cancellous
`L3 Cortical
`L4 Cancellous
`L4 Cortical
`L3 Pedicle
`Lamina
`Ligaments*
`Anterior longitudinal
`Posterior longitudinal
`Ligamentum flavum
`Interspinous
`Supraspinous
`Capsular
`Intertransverse
`Intervertebral Disc*
`Nucleus
`
`Annulus
`
`Implant
`Aluminum oxide
`Titanium
`Socket component
`Ball component
`
`Intact
`
`3805
`3765
`
`832
`368
`520
`200
`56
`29
`
`40
`24
`8
`14
`4
`28
`10
`
`392
`
`896
`
`0
`0
`0
`0
`
`7451
`5661
`
`832
`368
`520
`200
`56
`29
`
`0
`24
`8
`14
`4
`28
`10
`
`0
`
`8 node continuum
`
`2 node cable
`
`8 node continuum
`
`560 (A⫺ model)
`656 (A⫹ model)
`
`8 node continuum with 2 node
`rebar
`
`1856
`800
`1658
`998
`
`8 node continuum
`
`Material property references denoted by superscript: *27; †38, ‡29.
`
`

`
`Table 2. Comparison of FE Predictions and In Vitro Results. Predictions Generally Fell Within One Standard Deviation
`After Accounting for Uncontrolled Implantation Procedures
`
`Load-Sharing in an Implanted Lumbar Motion Segment • Dooris et al
`
`E125
`
`Rotation Displacement Under Load Flexion ⬎ 0, Extension ⬍ 0; (°)
`
`In Vitro
`
`AD&A⫺
`
`AD&A⫹
`
`PD&A⫺
`
`PD&A⫹
`
`Applied
`Moment
`(N-m)
`
`⫺6
`⫺4.5
`⫺3
`⫺1.5
`0
`1.5
`3
`4.5
`6
`
`⫺4.4 ⫾ 1.1.8
`⫺3.1 ⫾ 1.7
`⫺2.1 ⫾ 0.7
`⫺0.8 ⫾ 0.3
`0
`1.5 ⫾ 1.4
`2.8 ⫾ 1.5
`3.7 ⫾ 1.4
`5.1 ⫾ 1.5
`
`⫺6.0
`⫺4.8
`⫺3.6
`⫺2.0
`0
`1.6
`2.5
`3.4
`4.0
`
`⫺4.2
`⫺3.4
`⫺2.5
`⫺1.3
`0
`1.6
`2.5
`3.3
`4.0
`
`⫺6.0
`⫺3.6
`⫺4.8
`⫺2.0
`0
`3.0
`4.9
`4.9
`5.8
`
`⫺4.1
`⫺3.3
`⫺2.3
`⫺1.2
`0
`2.6
`4.5
`4.5
`5.2
`
`grooves and beads for osseointegration with the vertebral end-
`plates, this complex surface was simulated as a flat plane, with
`implant nodes fixed to the endplate nodes, thereby assuming
`perfect bone ingrowth. In addition, the implant was scaled to
`match the spine model’s disc space height, resulting in the disc
`occupying 40% of the segment’s cross-sectional area in the
`transverse plane. Distraction of the segment to restore disc
`height in a clinical restoration was not simulated, because the
`intact model had normal disc height initially. The alumina por-
`tion was assigned appropriate elastic material values (E ⫽ 380
`GPa, ␯ ⫽ 0.26),38 as was the titanium portion (E ⫽ 114 GPa, ␯
`⫽ 0.32).29 Friction was assigned a value of 0.09, which is the
`value reported in literature for wet alumina–alumina sliding
`contact.6 The nodes on the bottom of the L4 vertebra were
`fixed in space, whereas the nodes on the midsagittal plane were
`fixed in the lateral direction (i.e., assuming symmetry across
`that plane).
`This FE investigation included two loading schemes, corre-
`sponding first to model validation and then model investiga-
`tion. Loading conditions for model validation included a small
`compressive force of 10 N (approximately the weight of the
`loading frame used in the in vitro investigation ⫹ superior spine
`segments) and up to 6 N-m flexion or extension moment (as
`was also used in the in vitro study). To simulate compressive
`loads, forces were applied to the nodes of the superior L3 end-
`plate. Model predictions from this set of conditions were com-
`pared to in vitro results. The second set of conditions included
`an axial compressive force of 400 N combined with flexion and
`extension moments up to 10 N-m. The effects of pure axial
`compression also were studied by applying forces of up to 800
`N to L3.
`Changing model parameters facilitated study of some seg-
`
`Figure 4. Finite element model of the artificial disc components.
`Contact surfaces were defined for the entire ball and socket
`interface. Geometry was imported from CAD designs.
`
`ment biomechanics dependencies on surgical technique. The
`artificial disc was placed as posteriorly in the disc space as
`allowed by the remaining anulus (“PD”; Figure 5A). In the
`present study, this meant shifting the prosthesis 8 mm anteri-
`orly, as far as it could be shifted without overhanging the end-
`plate (“AD”; Figure 5B). In another parameter, the smallest
`annular window possible was cut (“A⫹”; Figure 5A), as dic-
`tated by the width of the implant for anterior insertion,
`whereas in a complimentary case, fully one third of the anulus
`was removed (“A⫺”; Figure 5B). After removal, the remaining
`anulus was 42% of the disc space in the A⫹ case and 35% in
`the A⫺ case, the difference being entirely in the anterior third of
`the anulus. Lastly, the anterior longitudinal ligament was re-
`stored to its original form (not removed) to isolate the effects of
`anular and anterior longitudinal ligament resection.
`
`Results
`
`The data shown in Table 2 indicate an FE model valid for
`further analysis. By accounting for both surgical vari-
`ables in the in vitro procedures, FE predictions for rota-
`tions under applied moments either matched or strad-
`dled in vitro results. In flexion, FE model predictions
`were either larger or smaller than in vitro results, de-
`pending on model parameters. In extension, the anteri-
`orly placed disc results closely matched in vitro data. In
`general, predictions for segment rotations fell within one
`standard deviation of the in vitro data.
`In flexion and extension plus 400 N preload, the im-
`planted spine segment model exhibited less rotational
`stiffness than the intact model. In extension, segment
`rotations were sensitive primarily to the amount of re-
`maining anulus. Segments with less anulus remaining
`(AD/PD, and A⫺) rotated approximately 40% more in
`extension (⫺5.12°, ⫺5.23° vs. ⫺3.7°), but segments
`with more anulus remaining (AD/PD, and A⫹) rotated
`approximately 30% more than the intact model
`(⫺4.78°, ⫺4.79° vs. ⫺3.7°). Under flexion moments,
`however, disc placement and amount of anulus affected
`rotation values. Placing the disc anteriorly decreased
`flexion by 19% (4.8° vs. 5.88), but placing it posteriorly
`increased flexion by 44% in the PD/A⫺ and by 36% in
`the PD/A⫹ cases. Under axial compression, AD and
`A⫹/A⫺ cases extended (⫺1.5° at 800 N), and PD and
`A⫹/A⫺ cases flexed (2.4°and 4.4° for PD and A⫹/A⫺,
`
`5 of 9
`
`

`
`E126 Spine • Volume 26 • Number 6 • 2001
`
`Figure 5. Surgical variables as implemented in the finite element model. The artificial disc was placed
`anteriorly and posteriorly, and small and large amounts of anulus were removed. Posterior place-
`ment of the disc was limited by the anulus, and anterior placement was limited by the end plate.
`PD&A⫺ ⫽ posterior disc with a large amount of anulus removed; PD&A⫹ ⫽ posterior disc with a
`small amount of anulus removed; AD&A⫺ ⫽ anterior disc with a large amount of anulus removed;
`AD&A⫹ ⫽ anterior disc with a small amount of anulus removed.
`
`respectively). In contrast, the intact segment’s rotation
`was negligible under compression.
`Forces across the facet joints increased with artificial
`disc implantation (Figure 6) under most loading condi-
`tions. Only with a posteriorly placed artificial disc at low
`loads (2 N-m or less) were facet loads lower in the im-
`planted FSU. Placing the artificial disc anteriorly in-
`creased the load shift to the facets to more than that in
`the intact case and in the posteriorly placed case (at loads
`less than 10 N-m). The load shift also occurred earlier,
`such that facet loads were generated before any exten-
`sion moment. Total facet unloading with the anteriorly
`placed artificial disc was not achieved until more than 2
`N-m moment was applied in flexion. Under 800 N com-
`pression, the facet joint was unloaded completely with
`the posterior artificial disc placement. With anterior
`placement, however, loads across the facet joint in-
`creased to more than 150% of those in intact case (38 N
`vs. 98 N).
`Changes in the von Mises stresses in the L3 pedicle
`and lamina caused by implantation reflected the rota-
`tional changes. In the PD/A⫺ and PD/A⫹ models under
`10 N-m, L3 pedicle stresses increased by approximately
`94% and 76% from those in the intact model, respec-
`tively, under flexion moment, and by approximately
`50% under extension moment. The anteriorly placed ar-
`tificial disc model predicted less of a change, approxi-
`mately 30% higher pedicle stresses under 10 N-m exten-
`sion moment, but 15% decreased stresses under 10 N-m
`flexion moment. In pure compression (800 N), pedicle
`stresses in the posterior disc cases decreased to half those
`of the intact case, but doubled with anterior artificial disc
`placement, as compared with those in the intact case. In
`the lamina, von Mises stresses in the PD/A⫹ and PD/A⫺
`cases were nearly twice their intact values in flexion and
`50% more in extension. Functional spinal units with
`anteriorly placed implants exhibited a 25% increase in
`lamina stresses in extension, but a 15% decrease in
`flexion.
`
`Anulus fiber stress patterns also changed with implan-
`tation. Generally, the analyses predicted large decreases
`in fiber stresses. Maximum fiber stresses in the anterior
`(A⫹ cases), lateral, and posterior anulus ranged from as
`low as 20% to 40% of the intact model values in 10 N-m
`extension. In flexion, however, maximum fiber stresses
`in the posterior anulus equaled those in the intact case at
`10 N-m. Increasing or decreasing the amount of remain-
`ing anulus (i.e., the size of the annular window) pro-
`duced small changes in maximum fiber stresses. In pure
`compression, the measured stress in the anulus fibers
`with posterior placement of the artificial disc was just
`10 –30% of the stress measured in the intact case, with
`slightly higher stresses observed with anterior placement.
`In the implanted FSU, anular bulge decreased, and the
`disc space did not compress under axial load.
`Stresses in the implanted spine model’s L3 cortical and
`cancellous and L4 cancellous bone elements increased by
`10 –30% under applied flexion and extension moments.
`L4 cortical bone stresses, however, did not follow the
`
`Figure 6. Facet contact forces as a function of applied moment,
`disc placement (AD and PD), and amount of remaining anulus (A⫹,
`A⫺). Disc placement affected posterior loads more than anulus
`resection. Facet contact forces in implanted cases were higher
`than those in intact conditions in segments with moments greater
`than 4 N-m; facet contact forces for moments less than 4 N-m
`were lowest in segments with posteriorly placed discs.
`
`6 of 9
`
`

`
`Load-Sharing in an Implanted Lumbar Motion Segment • Dooris et al
`
`E127
`
`Table 3. Effect of Restoring the Anterior Longitudinal
`Ligament. Restoration Significantly Reduced Segmental
`Rotation in the AD&Aⴙ/ⴚ Cases and Reduced Both
`Facet Loads and Rotation in the PD&Aⴙ/ⴚ Cases Nearly
`Back to Intact Levels for 10 N-m Extension with 400 N
`Compressive Preload
`
`Condition
`
`Facet Load (N)
`
`Rotation (°)
`
`Intact
`AD&A⫹/⫺
`AD&A⫹/⫺ & ALL
`PD&A⫹/⫺
`PD&A⫹/⫺ & ALL
`
`166
`221/222
`220/261
`211/230
`163/169
`
`⫺3.70
`⫺4.8/⫺5.1
`⫺3.9/⫺3.7
`⫺4.8/⫺5.2
`⫺3.8/⫺3.9
`
`same general increase: L4 cortical bone stresses continu-
`ously decreased with extension. Whereas the PD/A⫺ L4
`maximum cortical stresses were more than twice those of
`the intact model under 10 N-m applied flexion moment,
`they decreased to 7% less than those in the intact model
`in 10 N-m extension. Similarly, in the PD/A⫹ case, max-
`imum L4 cortical stresses decreased from 62% more
`than those in the intact model at full flexion to just 5%
`more than those in the intact model at full extension. In
`compression, the average von Mises L4 cortical bone
`stress decreased from 4.9 MPa in the intact model to 1.76
`MPa in the PD/A⫹ case, but increased to 5.4 MPa in the
`AD/A⫹ case. Similar trends were seen for the L4 cancel-
`lous bone in compression.
`Restoring the anterior longitudinal ligament signifi-
`cantly affected segment mechanics in extension, as is
`shown in Table 3. The ligament had no effect on the
`segment under flexion because of its inability to transmit
`compressive loads. It did, however, marginally reduce
`rotation under compression for the AD/A⫾ cases be-
`cause of the segment’s tendency to extend under com-
`pressive load. Rotation, L3 pedicle stresses, and facet
`contact forces all returned to nearly intact values under
`extension with posterior artificial disc placement. Lam-
`ina stresses decreased from 50% more than intact values
`to just 20% more. L4 cortical stresses increased in exten-
`sion between the anterior longitudinal ligament-restored
`cases and their implanted counterparts.
`
`Discussion
`
`For artificial disc replacement to be successful, biomate-
`rial, biomechanical, and clinical requirements all must be
`met. Biomechanically, extreme alterations in the load-
`sharing among the different spinal structural elements
`(such as the disc anulus and the facet joints) caused by
`artificial disc replacement could lead to instability, soft
`tissue destruction, abnormal bone resorption/growth, or
`fracture. Therefore, analysis of changes in load-sharing
`relations may aid in evaluating the indications and con-
`traindications for disc replacement. Because of the diffi-
`culties in addressing these issues through in vitro exper-
`iments, complementary approaches are mandated.
`In the present study, a computational approach (with
`experimental validation) was used to assist in evaluation
`
`of the biomechanics of an artificial disc replacement. An
`FE model of a spinal motion segment incorporating a
`ball-cup–type artificial disc design was created.
`The experimental protocol and the corresponding FE
`model had several differences worth enumerating here. It
`was not practical to quantify the degree of degeneration
`in each specimen used for in vitro testing, nor was it
`practical to simulate those parameters in the FE model.
`During the experiments, the disc implant size was se-
`lected based on the disc dimensions of a specimen and the
`size of the implants available to the authors. Also, during
`insertion, it was not practical to control precisely the
`location of the implant with respect to the disc morphol-
`ogy. The FE model assumed perfect osseointegration,
`also a departure from the in vitro test condition. (This
`may, however, compensate for the distraction induced
`across the

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket