throbber
ECTROCHIMICA
`ACTA
`“
`[PART B: ATOMIC SPECTROSCOl;f§T“'<~:=;.:
`Vol. 40B, No. 4, 1985
`
`Energetiq Ex. 2081, page 1 - |PR2015-01377
`
`Energetiq Ex. 2081, page 1 - IPR2015-01377
`
`

`
`U1 up 1 r\u\,r111v.uLt-1 AL 1 A
`
`PART B: ATOMIC SPECTROSCOPY
`EDITOR-IN-CHIEF
`Dr. P. W. J. M. Boumans, Philips Research Laboratories, 5600 JA Eindhoven, The
`Netherlands
`7
`EDITORS
`Dr. Walter Slavin, Perkin—Elmer Corporation, Main Avenue, Norwalk, CT 06856, USA
`Dr. H. W. Werner, Philips Research Laboratories, 5600 JA Eindhoven, The Nether/ands
`ASSISTANT EDITORS
`Dr. J.. A. C.’ Broekaert, Institute of Spectrochemistry and Applied Spectroscopy (15343),
`P.0. Box 778, 4600 Dortmund, Federal Republic of Germany
`Dr. R. J. Decker, Rhodesia University, Chemistry Department, P.0. Box MP l69, Harare,
`Zimbabwe
`EDITORIAL ADVISORY BOARD
`Chairman: J. D. Winefordner, Gainesville, FL
`C. Th. J. Alkemade, Utrecht, The Netherlands
`S. F. A. Kettle, Norwich, UK
`R. M. Barnes, Amherst, MA
`S. R. Koirtyohann, Columbia, MO
`L. S. Birks, Washington, DC
`K. Laqua, Dortmund, GFR
`R. F. Browner, Atlanta, GA
`B. V. L’voV, Leningrad, USSR
`C. L. Chakrabarti, Ottawa, Canada
`F. J. M. J. Maessen, Amsterdam,
`K. Dittrich, Leipzig, GDR
`The Netherlands
`H. Falk, Berlin, GDR
`S. L. Mandelshtam, Moscow, USSR
`V. A. Fassel, Ames, 1.4
`J. M. Mermet, Vernaison, France
`K. Fuwa, Tokyo, Japan
`N. Omenetto, Ispra, Italy
`L. de Galan, Delft, The Netherlands
`J. M. Ottéway, Glasgow, UK
`Y. Gohshi, Tokyo, Japan
`E. Pl§l<o,' Bratislava, CSSR
`D. Hercules, Pittsburgh, PA
`J. Robin, Villeurbanne, France
`G. M. Hieftje, Bloomington, IN
`F. Riidenauer, Vienna, Austria
`G. Horlick, Alberta, Canada
`M. Someno, Nagoaka City, Japan
`A. M. Huber, Orsay, France
`A. Strasheim, Pretoria, RSA
`H. Kawaguchi, Nagoya, Japan
`A. Walsh, Brighton, Australia
`M. Kent Wilson Washington, DC
`K. Zimmer, Budapest, Hungary
`Publishing, Subscription and Advertising Ofiicesx Pergamon Press Inc., Maxwell House, Fairview Park, Elmsford,
`NY 10523, U.S.A.; Pergamon Press Ltd., Headington Hill Hall, Oxford OX3 OBW, U.K. (Oxford 64881).
`Payments must be made in advance. Subscription enquiries from customers in North America should be sent
`to: Pergamon Press Inc., Maxwell House, Fairview Park, Elmsford, NY 10523, U.S.A., and for the remainder
`of the world to: Pergamon Press Ltd., Headington Hill Hall, Oxford OX3 OBW, U.K.
`One volume per annum. Published monthly.
`Subscription rates: For libraries, university departments, government laboratories, industrial and other multi-
`reader institutions (I985), U.S. $325.00; two-year rate (1985/86) U.S. $617.50 (including postage and insurance).
`Specially Reduced Rates to Individuals
`In the interests of maximizing the dissemination of the research results published in this important international
`' umal we have established the following price structure. Any individual, whose institution takes out a library
`subscription, may purchase a second or additional subscription for personal use at a much reduced rate of U.S.
`$75.00 per annum.
`Micrafortrr Sub.n:n'prr‘ans and Back Issues
`Back issues of all previously published volumes are available in the regular editions and on microfilm and microfiche. Current subscriptions are
`year.
`available on microfiche simultaneously with the paper edition and on microfilm on completion of the annual index at the end of the subscription
`Copyright © 1985 Pergamon Press Ltd.
`
`'
`
`
`
`It is a condition of publication that manuscripts submitted to this journal have not been published and will not be simultaneously submitted or
`published elsewhere. By submitting a manuscript, the authors agree that the copyright for their article is transferred to the publisher if and when
`
`permission in writing from the copyright holder.
`or transmitted in any form or by any means, electronic, electrostatic, magnetic tape, mechanical, photocopying, recording or otherwise. without
`U.S. Copyrighl Law Applicable to Users in the U.S.A.
`P/mtnrnpying Informationfor users in the U.S.A.; The Item-Fee Code for this publication indicales that authorization to photocopy items for internal
`or personal use is granted by the copyright holder for libraries and other users registered with the Copyright Clearance (‘enter (CCC) Tmnsuclional
`
`written permission must be obtained from the publisher for such copying. in case of doubt please contact your nearest Pergamon office.
`The Item-Fee Code for tlrir publication is: 0584-8547/85 $3.00 + .00.
`
`PERGAMON PRESS
`HEADINGTON HILL HALL, OXFORD OX3 OBW, ENGLAND
`MAXWELL HOUSE, FAIRVIEW PARK, ELMSFORD, NEW YORK 10523
`
`Energetiq Ex. 2081, page 2 - |PR2015-01377
`
`Energetiq Ex. 2081, page 2 - IPR2015-01377
`
`

`
`This material may be protected by Copyright law (Title 17 U.S. Code)
`
`Energetiq Ex. 2081, page 3 - IPR2015-01377
`
`

`
`
`
`666
`
`DAVID A. CREMERS et al.
`
`atmospheric pressures or above. The discharge resides near the focus of the laser be;
`independent ofany physical support and does not require a gas flow to stabilize the plasma,
`some other sources. Because the discharge is maintained using optical frequencies (30 TH:
`the plasma is called a “continuous optical discharge” (COD).
`
`small, about 1 mm in diameter, and appeared as a very bright white light. The plasma“,
`initiated by the spark produced by a focused Q-switched Nd :YAG laser pulse superimposed,
`on the focal volume of the cw-beam because the cw powers used to maintain the plasma we
`insufficient to induce optical breakdown. Typically, the focused powers of cw-CO2
`105-107 W/cmz, several orders of magnitude smaller than the 108-109 W/cmz
`
`produced on the surface of a material which is temporarily introduced into the focal volu
`of the cw-beam [15]. Once started, the plasma operates continuously as long as suffici’
`intensity is supplied to the focal volume.
`
`
`
`Fig. 1. Photograph of the COD produced inside a small quartz tube containing Xe at 2000 torr. The ~
`cw-CO2 laser beam entered the cell from the right. The plasma was photographed for 1 s, atf8.6, with
`a neutral density filter (ND) = 3 in front of the camera lens. Then the plasma was turned off and the
`cell photographed under room lights with the ND filter removed.
`
`[19] G. I. KozLov,V.A.Kuzr~;ETsov and V.A. MAsvUKOv,Zh.Tekh.Fiz. 49, 2304 (1979) [Sov. Phys. Teclrl’/1 '
`1283 0979)].
`[20] D. C. SMITH and M. C. FowLi-IR, Appl. Phys. Len. 22, 500 (1973).
`[21] D. R. KEEFER, B. B. HENRIKSEN and W. F. BRAERMAN, J. Appl. Phys. 46, 1080 (1975).
`
`identical gases and pressures [4, 10, 17]. The temperature of the COD depends upon theigas
`and at 1520 torr has been mmsured spectroscopically to be a maximum of 14000 Ki Xe
`
`Energetiq Ex. 2081, page 4 - |PR2015-01377
`
`Energetiq Ex. 2081, page 4 - IPR2015-01377
`
`

`
`Evaluation of the continuous optical discharge
`
`667
`
`3000 K in Ar, and 22000 K in N2 [16]. These are much higher than the temperatures
`31-acterizing other continuous excitation sources: arcs (4000-5000 K), ICP (60()0~8000 K)
`d a microwave discharge (5000—7000 K). The temperature of the COD can approach that
`fained by sparks (20 000 K. or higher) and is due to penetration of the high frequency optical
`ation into the core of the plasma [16]. At laser frequencies, which are typically above the
`sma frequency [22], absorption occurs mainly via free—free transitions (inverse
`femsst1‘ahlung) associated with electron—ion collisions [23]. For comparison, at radio and
`crowave frequencies, which are below the plasma frequency, plasma heating occurs
`771-ough direct plasma—electric field interactions characterized by larger absorption coef-
`clents. Consequently, only the outer layers of these plasmas are heated. The higher
`imperaturc of the COD is also related to its greater input energy density (p) compared to
`re conventional discharges. For example, for a 0.1-cm dia. plasma, a laser power of 45 W,
`‘d assuming 50% absorption by the plasma [7], p = 5 x 103 W cm”, vs p = 188 W cm‘ 3
`an ICP of l.5 kW and a volume of 8 cm3. The electron density of the Ar COD is
`out 10” cm”, at least an order of magnitude greater than other common continuous
`durces [4].
`tudics of the COD to date have concentrated mainly on measurements of its physical
`roperties in various gases under different conditions. Several theoretical models of the COD
`ave been developed to account for these properties [2429] and a few applications have
`em suggested [30—32]. An almost complete listing of the previous studies is presented in the
`eferences cited herein. To our knowledge, the COD has not previously been investigated as
`n excitation source for atomic emission spectroscopy. The high temperatures and electron
`ensity of this plasma should provide improved excitation for species diflicult to detect with
`ooler conventional cw sources. In addition, the COD is a single source which combines the
`h temperature of a spark with the continuous operation of a do are, the goal behind the
`evelopment of many other types of spectroscopic sources [33]. In this paper we present the
`ults of a preliminary examination of this plasma for spectrochemical analysis. Particular
`mphasis is placed on characteristics of COD generation, variation of plasma properties with
`perating parameters, and analytical performance.
`
`2. EXPERIMENTAL
`
`Apparatus
`schematic of the apparatus used to generate the COD and record the emission spectrum is shown in
`1g. 2. The experimental conditions are listed in Table l. The gas cell was evacuated to a pressure
`50 um before adding plasma gas. The cw—CO2 and pulsed Nd : YAG laser radiations were focused
`me the same volume of the cell at right angles as shown. The focal volume of the pulsed beam was
`usted to overlap that of the cw-CO2 beam by moving the glass imaging lens slightly using an X YZ
`
`
`
`] The plasma frequency (vp) is given by V; = (eznc)/(41rZe0mc), Where e, He, and me are the electronic charge,
`density, and mass, respectively, and so is the vacuum permittivity. At frequencies below vp, EM waves interact
`directly with the plasma to induce heating. At frequencies above vp, the waves only interact with individual
`electrons and ions to heat the plasma. For an electron density of8 X 10”‘ cm”, vp : 2.5 X 10” Hz, which is
`about an order of magnitude below the CO2 laser frequency of 30 TI-lz.
`’
`] T. P. HUGHES, Plasmas and Laser Light. John Wiley, New York (1975).
`] A. A. Kuunnov, T. YA. PoPovA and N. G. PREOBRAZENSKY, Pros. ofthe XIII Int. Conf. on Phenomena in
`Ionized Gases, 1977, Part II, p. 899. Berlin, GDR (1977).
`_
`25] G. I. KozLov and l. K. SELEZNEVA, Zh. Tekh. Fiz. 48, 386 (1978) [Son Phys. Tech. Phys. 23, 227 (1978)].
`] M. V. GERASIMENKO, G. I. Kozrov, V. A. KUZNFTSOV, Pis’ma Zh. Tekh. Fiz. 6, 485 (1980) [Son Tech. Phys.
`‘ Lett. 6, 203 (l980)].
`] Yu. P. RAZIER, Pis’ma Zh. Tekh. Fiz. 7, 938 (1981) [Sov. Tech. Phys. Lett. 7, 404 (198l)].
`] S. MULLER and J. Uurhwnuscn, Physica 112C, 259 (1982).
`19] G. G. GLADUSII and A. N. YAVUKHIN, Kuantovayu Eleklron. (Moscow) 10, 1399 (1983) [S01]. J. Quantum
`_
`Electron. 13, 908 (l983)].
`[30] R. W. THOMPSON, E. J. MANISTA and D. L. ALGER, Appl. Phys. Lett. 32, 610 (1978).
`i’ ] Laser Focus, Dec. 1977. p. 20.
`32] N. H. KEMP and R. G. Roor, J. Spacecraft 16, 65 (1979).
`] P. W. J. M. BOUMANS, Analytical Eniissimr Spectroscopy, Ed. E. L. GROVE, Part II, Chap. 6. Marcel Dekker,
`New York (1972).
`
`Energetiq Ex. 2081, page 5 - |PR2015-01377
`
`Energetiq Ex. 2081, page 5 - IPR2015-01377
`
`

`
`668
`
`DAVID A. CREMERS el al.
`
`
`
`S/GA/AL
`
`Fig. 2. Top view diagram of the apparatus used to generate the COD and monitor the emission
`spectrum. The large gas cell is shown here. The cw-CO2 laser beam was focused using a mirror (dashed
`line) or lens (shown here). The COD was sometimes started by replacing the side window by a pair of
`electrodes shown in the insert. PB: pulsed laser beam to start the COD; TS: translation stage to
`position the lens focusing the pulsed beam; W: quartz window; SW: salt window; C: chopper; QL;
`quartz lens; P: photomultiplier tube; M: monochromator; LIA: lock-in amplifier; R: chart recorder; E;
`electrode; B: vacuum bellows; G: pressure gage; V: vacuum pump; Xe: Xe gas supply. The tesla coil was
`connected to wire l-lV.
`
`
`
`translation stage. The mode of the cw-CO; laser was adjusted to be TEM00. The laser power was not
`stabilized via feedback control but it varied by less than l ‘X, over a 4-h period. The laser was operated
`broadband with a maximum available power of about 70 W. The pulsed laser was operated at about-
`I Hz and was turned off following ignition of the COD. The cell pressure reached a constant value
`within ~ 1 min after ignition.
`-
`For measurements of line intensities and line widths, the plasma light was mechanically chopped '
`imaged on the entrance slit of a scanning monochromator with unit magnification, and detected with a :
`photomultiplier tube (PMT). The PMT signal was processed by a lock-in amplifier (LIA). The“
`monochromator wavelength was scanned over spectral features and the LIA output recorded on a chart —
`recorder. The spectral response of the detection system from 350 to 960nm was determined using I
`standard lamp calibration. Phase sensitive detection with the LIA minimized noise introduced by
`slight regular fluctuation of the plasma intensity discussed below.
`A
`Some measurements of the stability of the COD were made by tuning the monochromator C
`wavelength to a spectral feature (emission line or background), storing the PMT current on a capacitor —'
`over an “exposure” period of 5 s, and then digitizing the accumulated voltage. This method was used ',
`
`because intensity data could be obtained over a much shorter period compared to scanning over th V
`,
`spectral line. Measurements were repeated 16 times to obtain statistical information.
`To spectrally resolve the intensity profile of the COD, the light at the rnonochromator exit slit was
`imaged with unit magnification on a vertically oriented photodiode array. The voltages from each ofthe —[
`512 photodiodes of the array were digitized and read into the computer memory for averaging.
`2.2. Gas cells
`
`
`
`lens or spherical mirror, as shown. The side windows were u.v. grade fused silica. These windows and the *
`rear flange of the cell were often replaced by various devices to test different analytc introduction
`techniques or nonlaser methods of starting the COD, some of which are discussed below. The fron,
`brass flange ofthc cell became very hot during prolonged operation and was water cooled to prevent the _.
`salt window from cracking. It was not necessary to cool the rest of the cell.
`The small gas cell is shown in detail in Fig. 3. The plasma was generated inside a quartz tube 3.3-Cm’
`long by 1.5-cm dia. to provide a small volume (~ 7 cm3) and to eliminate the need for separate windows
`to ignite the plasma and View the spectral emissions. These \vcre accomplished by imaging direcl1Y ‘
`through the tube walls. The 0.18-cm wall thickness of the quartz tube limited the gas pressure to ‘
`S 2800 torr. Analyte gases were introduced into the COD by injection either into a flowing gas stream .
`
`Energetiq Ex. 2081, page 6 - |PR2015-01377
`
`Energetiq Ex. 2081, page 6 - IPR2015-01377
`
`

`
`Evaluation of the continuous optical discharge
`
`669
`
`Table 1. Experimental apparatus and conditions
`
`A. COD generation apparatus
`(1) cw-CO2-laser
`wavelength
`power
`beam diameter
`beam divergence
`
`,
`
`(2) Nd':YAG pulsed laser (ignition)
`wavelength
`g
`pulse duration
`pulse energy
`repetition rate
`(3) Sample cells
`material
`size
`large cell (Fig. 2)
`small cell (Fig. 3)
`gas pressure measurements
`
`(4) Optics
`10.6 pm focusing lens
`
`1.06 pm focusing lens
`cw-C0; input window
`(both cells)
`side windows (large cell)
`B. Detection system
`(1) Spectrometer
`gratings
`
`slit width
`slit height
`second order filter
`
`(2) Phutomultiplier tubes
`
`(3) Chart recorder
`(4) Chopper
`(5) Photodiode array
`evaluation board
`
`C. Solid sample introduction
`(1) Nd:YAG pulsed laser
`wavelength
`pulse duration
`pulse energy
`repetition rate
`(2) Lens
`
`Advanced Kinetics MSL-100
`106 pm
`45 W typical
`6 mm
`6 mrad
`
`Quanta—Ray DCR—lA
`1.06 am
`15 ns
`250 ml/pulse
`1 Hz to initiate COD
`
`brass
`
`20L><10Hxl0Wcm
`9 L x 9 H x 9 W cm
`MKS 254 flow controller/Baratron,
`0~l0000 torr head
`
`2.54 cm dia., 2.54 cm fl, positive meniscus
`ZnSc, anti—reflection coated for 10.6 pm
`2.54 cm dia., 7.5 cm ii
`5 cm dia., 1.25 cm thick NaCl
`
`1 cm thick, 5 cm dia. quartz
`
`Spex industries 0.5 m Czerny Turner
`1200 lines mm”, 500 nm blaze, A 2 350 nm
`3600 lines mm 1, 300 nm blaze, .3. < 350 nm
`2 10 pm
`100 ,um—2 cm
`Corning CS 3—70, 1. 2 600 nm
`RCA C3l034, 14004600 V, l 2 350 nm
`Hamamatsu R955, 800 1200 V, /l < 350 nm
`Houston Instruments, Omniscribe
`Laser Precision CTX-534
`Reticon RL512S
`RC1024-S~2
`
`ILS NT-901-I
`1.06 pm
`15 ns
`75 mJ/‘pulse (attenuated)
`10 Hz
`10 cm H
`
`Ssing through the cell from front to rear or into a static gas fill using the rear latex seal shown in Fig. 3.
`_c seal was positioned out of view of the COD because it was melted by the large fraction of 10.6-pm
`radiation transmitted through the plasma volume. The front salt window flange and the rear analytc
`be was air cooled using convection and only a slight
`ection port were water cooled. The quartz tu
`azing ot‘ the tube directly above the COD was observed due to the high temperatures which reached
`°C at the outer wall. The inside tube diameter was restricted to a minimum of 1.5 cm to prevent the
`Sed Nd: YAG pulses from striking and damaging the tube walls.
`
`3. RESULTS
`
`n this section, results are presented describing some characteristics of plasma generation
`e using combined lower laser powers and lower gas pressures than previously reported.
`
` mat:ww>.~mw.
`
`Energetiq EX. 2081, page 7 - IPRZO15-01377
`
`Energetiq Ex. 2081, page 7 - IPR2015-01377
`
`

`
`670
`
`DAVID A. CREMERS er al.
`
`YGAS FLOW PORT
`
`
`
`Fig. 3. Diagram of the small gas cell, photographed in Fig. 1, used in analytical experiments with a
`static Xe fill. Gas samples were introduced through the injection port using a syringe. Solid samples
`were laser ablated from a metal strip positioned below the COD (see Section 3.3). WCC: water cooling
`coil.
`
`Xenon was used because of the difificulty of producing the plasma in Ar at the ‘low pressures
`(<3 atm) and low laser powers (< 70 W) available to this study. A minimum of 80W is
`needed to generate the COD in Ar at 3040 torr [I 0]. The lower operating pressures of the Xe
`COD also meant greater ease of sample introduction and less stringent materials
`requirements. The maintenance threshold of Xe is lowest ol'a1l stable inert gases because ofits
`relatively low ionization potential and thermal conductivity. The cw-CO2 laser beam was
`imaged into the gas cell horizontally because of the significantly greater cw maintenance
`thresholds required to generate the COD with a vertical beam
`Although Xe is expensive
`and would not normally be employed in a routine analysis, the results obtained here should
`indicate general trends applicable to a COD generated in other inert gases.
`"
`Initial attempts at generating the COD used a 2.8-cm focal length spherical copper mirror
`similar to those used in other studies [3, 4, 10]. The resulting plasma was unstable, as
`indicated by large irregular intensity fluctuations, and remained operating for at most 30 3;
`This behavior was attributed to the small diameter (6 mm) of the CO2 laser beam. The COD
`intercepted a significant portion of the central intense region of the incoming laser beam
`before it reached the mirror. This not only reduced the beam intensity reflected off the mirror
`back into the focal volume, but could also induce instability because energy was incident on
`the plasma from behind the focus. The plasma is spatially stable near the focus of the las’
`beam due to the decreasing intensity gradient as the plasma moves toward the focusing,
`
`In addition, with mirror focusing strong convection currents and density gradients from the;
`COD distort the incoming laser beam [17] and this effect will be more pronounced with
`smaller beam diameters. Instability in the COD noted here using mirror focusing was
`probably not as apparent in previous work because of the larger laser beam diameters, which
`were typically 2 cm or more [3, 7, 19]. In this study, a stable plasma was generated at lower X
`pressures and laser powers than previously reported in the literature using a 2.54-cm focal‘
`length best-form ZnSe lens. For example, at 1875 torr the COD was maintained with on
`25 W, whereas a minimum of about 80 W was reported previously [4]. The lower threshold is I
`
`probably due to beam focusing with a lens instead of a mirror. Zinc-selenide was used for lens.
`material because of its low absorption at 10.6 pm and absence of “thermal runaway
`exhibited by germanium at high cw laser powers.
`For a C02 laser beam power of 45 W, the plasma could be reliably maintained in Xe at:
`pressures between 1335 and 3200 torr using the large cell. With the small cell the minimum
`pressure was about ll50 torr. The lower maintenance threshold results from insulficien ‘
`absorption of the laser beam by the plasma to compensate for energy losses due to thermal
`
`
`
`Energetiq Ex. 2081, page 8 - |PR2015-01377
`
`Energetiq Ex. 2081, page 8 - IPR2015-01377
`
`

`
`
`
`Evaluation of the continuous optical discharge
`
`671
`
`ductivity, convection, and radiation [16]. The upper threshold results from increased
`ative losses at the higher gas densities.
`he COD was ignited in most cases using the laser spark generated by focusing a 2 MW
`' AG pulse within the focal volume ofthe 10.6-um beam. Careful alignment of the focal
`mes was necessary to produce ignition and movements of 0.03 cm were critical. Small
`ges in the alignment led to the observation of unique phenomena: one pulse would start
`COD but the next pulsc would turn it off and the third pulse would turn it on, and so
`rth. When the focal volumes were close, a bright light flash was observed from the chamber
`0 e COD ignited but quickly went out. Trial and error proved to be the best method of
`‘gning the focal volumes. Although most convenient, plasma ignition need not be
`mplished with a laser pulse. As noted above, the COD has been started in other studies
`mg the plasma formed on a tungsten wire introduced into the focal volume of the cw beam.
`is method did not work with the low laser powers used here. However, it was possible to
`rt the plasma using a tesla coil connected to the assembly shown in Fig. 2 consisting of a
`of electrodes positioned inside a bellows and flange assembly. The flange was mounted in
`ce ofa quartz side Window. One electrode was connected to a grounded metal tube passing
`
`ivated, a bright series of sparks was formed between the two electrodes. As the spark was
`nually swept through the focal volume of the C0; laser beam, the COD was easily ignited.
`e electrodes were then retracted out of the way.
`Unsuccessful attempts were made to generate the COD with up to 60W of 1.06-pm
`diation from a multimode cw-Nd:YAG laser. Because laser heating of a plasma via inverse
`Bremsstrahlung varies as A2 [23], the failure to form the COD was probably due to the 100
`es lower absorption of the plasma at 1.06 pm compared to 10.6 hm.
`
`. Plasma characteristics
`
`.2.1. Eflects of pressure and laser power. Figure 4 shows the Xe COD spectrum, not
`rected for detection system spectral response, over the 200—940 nm region at
`two
`ressures. Emissions from neutral and once-ionized Xe were observed having upper levels in
`he 9.6—l3.9 eV range. Because of the high background intensity, due to the high COD
`
`ed in Table 2. In general, Xel intensities of 100 or more on a scale in which the strongest Xel
`11615 104 [34], could be observed. Because of the high intensity of the COD, slit widths less
`han 10 um were used to prevent saturation of the detection system.
`0 he high electron density of the COD broadens emission line widths via the Stark elfect.
`
`Table 2. Relative intensities of some Xe lines observed in the COD*
`
`Wavelength (nm)
`
`Rel. Int. Wavelength (nm)
`
`Rel. Int.
`
`450.10
`452.47
`462.43
`467.12
`473.42
`480.70
`482.97
`484.33
`491.65
`492.32
`
`0.030
`0.018
`0.013
`0.033
`0.037
`0.013
`0.008
`0.016
`0.010
`0.003
`
`529.22 (Xe II)
`764.20
`823.16
`826.65
`828.01
`834.68
`840.92
`881.94
`895.23
`904.55
`
`0.006
`0.044
`0.839
`0.029
`0.259
`0.146
`0.070
`1.000
`0.278
`0.344
`
`*)(e pressure = 2000 torr; cw-C02 laser power = 45 W. Lines are
`from Xe 1 unless otherwise indicated.
`
`4] A. R. STRIGANOV and N. S. SVENTITSKII, Tables of Spectral Lines of Neutral and Ionized Atoms. 1191/Plenum,
`New York (1968).
`
`A
`
`
`
`Energetiq Ex. 2081, page 9 - |PR2015-01377
`
`Energetiq Ex. 2081, page 9 - IPR2015-01377
`
`

`
`
`
`672
`
`DAVID A. CREMERS et a1.
`
`Measurements of the half-width of the 450.10 and 834.68-nm Xe I and the 529.22-nm Xe,
`lines were obtained over the pressure range of 1402 to 2204 torr. The average widths we
`0.13, 0.17 and 0.13 nm, respectively. Only the 450.10«nm line showed a pressure dependeng
`decreasing from 0.14 to 0.12 as the pressure decreased. The instrumental resolution
`0.086 nm, determined using a low pressure Hg lamp. was not removed from the
`experimental line widths.
`_
`The COD resides near the focus of the CO2 laser beam that supplies it with energy, at
`position dependent upon several parameters including laser energy and gas pressu;
`Changes in these parameters moved the plasma along the axis of the focused laser beam. F
`example, decreasing the Xe pressure from 3175 to 1336 torr moved the plasma 0.5m
`toward the focus. Similarly, at a pressure of 2000 torr, changing the laser power from 25
`60 W moved the plasma away from the focus by 0.3 mm. These results are understood b
`considering the motion of the plasma in a parallel laser beam: the plasma moves toward‘!
`
`losses from thermal conduction, convection, and radiation. The absorption coefiicicnt varie
`with pressure (p) as 11 z p2 [16] so that as the pressure increases, AI also increases for a give
`laser power, and the plasma moves further away from the focus, into a region of reduce
`intensity, until absorption is again balanced by the loss mechanisms. In the same way, as th
`laser power is increased at constant pressure, AI increases and the plasma moves toward th
`focusing lens, as observed.
`Gas pressure and laser power also elfect the spectral excitation characteristics of the CO
`For example, pressure changes alter the intensities of the background signal and the Xe I an
`Xe II lines as shown in Figs 4 and 5. During these measurements, the position of the imagin
`A
`A
`
`
`
`
`
`2006- torr
`
`
`
`
`I404 torr
`
`200
`
`275
`
`350
`
`550
`
`7'50
`
`950
`
`Wavelength, nm
`Fig. 4. Spectrum from 200 to 950 nm of the Xe COD at two pressures, not corrected for detection
`system response. Different PMTs and gratings were used to monitor the regions above and below
`350 nm. The horizontal line under each spectrum indicates the baseline signal with the COD off.
`Emission lines marked by a triangle are off-scale. A filter was placed at the entrance slit beginning at
`the wavelength indicated by the arrows to eliminate the second order spectrum.
`
`Energetiq Ex. 2081, page 10 - |PR2015-01377
`
`Energetiq Ex. 2081, page 10 - IPR2015-01377
`
`

`
`Evaluation of the continuous optical discharge
`
`673
`
`(°~V)
`
`Ratios
`
`28
`
`32
`
`XeI.Xe11.B65Intensities,V
`
`16
`
`20 24
`
`|O“°'- Pressure.
`
`tort
`
`Fig. 5. Variation of the intensities of (1) 450.10 nm Xe I and (1') adjacent background at 448.00 mm
`(BG 1), and (2) 529.22 nm Xe II and (2') adjacent background at 527.00 nm (BG II), and the ratios (3)
`[Xe I/BG 1] and (4) [Xe 11/BG II], as a function of xenon gas pressure. The laser power was 46 W.
`
`ns was adjusted at each pressure to maintain the most intense portion of the plasma over the
`trance slit. As the pressure was decreased from 3175 to about 1900 torr the intensities from
`1 three sources increased. As the pressure was reduced further, the Xe 1 intensity increased
`ntil about 1700 torr and the Xe I1 intensity peaked at 1600 torr. The net result was that the
`tie of the emission line to background intensity steadily increased as the pressure was
`duced from 3175 to about 1500 torr. These observations agree with measurements reported
`elow, which indicate the temperature of the COD increases somewhat with decreasing
`ressure. Different behavior was observed by changing the laser power. As the power
`ccreased from 55 to 25 W, the intensities of Xel (450.10 nm), Xe 11 (529.22 nm), and the
`djacent backgrounds all increased slightly, but the ratios of these lines to background
`‘mained essentially constant.
`3.2.2. Temperature. Detailed measurements of the COD temperature have been reported
`reviously [8, 9, 11, 17]. The intention here was to determine how the temperature and
`ectron density varied as certain experimental parameters were changed to ascertain the
`ability of this source. The COD temperature was measured using the Saha equation and a
`oltzmann plot. Because of the asymmetry of the horizontal plasma, the resulting data were
`ot Abel inverted. However, only light from the center of the plasma image (10-um wide by
`00-pm high) was monitored by masking the monochromator slits. The resulting tempera-
`ures are therefore population averaged as discussed by BOUMANS [35]. The electron density
`as determined by seeding the Xe with about
`l.3—2.0 7,; hydrogen gas by volume and
`easuring the width of the H, line at 656.28 nm. The Hp line (486.13 nm), which yields more
`ccurate densities [36], was not used because of several Xe I interferences and the about
`ixfold weaker intensity of this line compared to H,,,. To minimize perturbation of the COD,
`' nly the smallest amounts of hydrogen needed to make a measurement were added. Above
`"/0,
`the plasma noticeably decreased in intensity and was extinguished above 5 ‘,’/,,
`oncentration. Using the formula of GRIEM [36], the electron density was deduced by an
`crative process with the Saha temperature. The 450.10-nm Xe I and 529.22-nm Xe 11 lines
`ere used for the Saha analysis. At the pressures of 2766, 2130 and 1601 torr, the temperature
`as measured to be 9630, 10200 and 10 375 K, respectively, using 45 W of laser power. The
`lectron density also varied slightly: at the respective pressures it was 7.39, 8.95 and 7.31
`X 10' 5 cm” 3, which corresponds to about 0.12 "/0 ionization ofthe Xe. The width of the H,
`
`
`
`
`
` 35] P. W. J. M. BOUMANS, Spectrochemical Excilulion. Hilger and Watts, London (1966).
`
`-VT—[36] H. R. GRIEM, Plasma Spectroscopy. McGraw-Hill, New York (1964). The formula used to calculate the
`‘
`electron density (he) is given on p. 305: ne = C(nc, T )A/13”. Here, A}. is the full Stark line width and C(ne,7‘) a
`coefficient tabulated on pp. 5384539.
`
` 5A(R)4O:1v‘I
`
`
`Energetiq Ex. 2081, page 11 - |PR2015-01377
`
`Energetiq Ex. 2081, page 11 - IPR2015-01377
`
`

`
`
`
`674
`
`DAVID A. CREMERS et al.
`
`line varied in the range 0.75-0.85 nm as the pressure varied. At 2130 torr, the COD,
`temperature changed by less than 2 “/0 and the electron density by less than 10 9,’, as the laser‘
`power increased from 32 to 48 W.
`To obtain a good measure of the Boltzmann temperature of the plasma, the spread in the
`upper energiesof the emitting levels should be at least 2 eV. Unfortunately, the majority 0f_
`Xel lines for which the oscillator strengths have been measured correspond to upper levejs,
`grouped between 11.4 and 12.0 eV [37]. However, when about 66 "/,§ Kr was added to the Xe
`COD at least 13 lines of Kr I were observed corresponding to upper levels spanning the range
`from 11.44 to 13.49 eV for which the oscillator strengths are known to 1 50 % (Table 3). The
`
`relative intensities of these lines were measured to produce the Boltzmann plot shown in F1’ *
`6. The integrated intensities were determined by tracing the lines on paper, cutting them on’
`Table 3. Data for Kr 1 lines used to construct Boltzmann
`plot*
`
`Wavelength (nm)
`427,40
`43514
`440.00
`445.39
`446.37
`587.09
`76015
`768.52
`769.45
`785.48
`810.44
`811.29
`850.89
`
`g1
`5
`1
`5
`5
`3
`5
`5
`1
`3
`3
`5
`7
`3
`
`10"‘ x A(s‘1)*
`0.026
`0.032
`0.020
`0.0078
`0.023
`0.018
`0.31
`0.49
`0.056
`0.23
`0.13
`0.36
`0.24
`
`Eu(eV)T
`12.82
`13.49
`13.46
`12.82
`12.81
`12.14
`11.55
`12.26
`11.53
`12.14
`1 1.44
`11.44
`12.10
`
`‘Data from: Wavelengths and Transition Probabilities for
`Atoms and Atomic Ions Part II, Transition Probabilities,
`NSRDS-NBS 68, US. Government Printing Office,
`Washington, DC (1980).
`1Here, g and E“ are the statistical weight and energy,
`respectively. of the upper emitting level corresponding to the
`transition having transition probability A, The estimated
`uncertainty of A is 1- 50 "/9.
`
` H
`
`I3
`
`I2
`
`Eu, eV
`
`Fig. 6. Boltzmann plot of Kr I lines in the Xe/Kr COD used to measure the plasma temperature.
`Only the center region of the plasma (10 um W x 100 um H) was monitored. The gas pressure was
`2000 torr and the laser

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket