Model EQ-77
`LDLS™
`Laser-Driven Light Source
`
`
`
`
`Operation Manual
`
`Revision 2 December 2015
`Part Number DOC-7070
`
`
`
`
`
`
`
`
`
`
`Energetiq Ex. 2056, page 1 - IPR2015-01362
`
`

`
`
`
`
`
`Copyright © 2015 Energetiq Technology Inc. All rights reserved.
`
`Energetiq products are covered by the following patents: US 7435982, 7786455, 8525138,
`8969841, 9048000; Japan 5410958, 5628253; Korea 10-1507617; UK GB2450045. All technical
`information, including drawings, schematics and specifications contained in this manual are the
`property of Energetiq and shall not be reproduced in whole or in part without the written
`consent of Energetiq. The content of this manual is subject to change without notice.
`
`Energetiq Technology Inc.
`7 Constitution Way, Woburn, MA 01801 USA
`Tel. +1 (781) 939-0763
`
`
`
`
`Energetiq Ex. 2056, page 2 - IPR2015-01362
`
`

`
`
`Fax +1 (781) 939-0769
`E-mail: support@energetiq.com
`http://www.energetiq.com
`
`
`
`
`Declaration of Conformity
`
`We, the manufacturers
`
`
`Energetiq Technology Inc.
`7 Constitution Way
`Woburn, Massachusetts USA
`
`hereby declare that the product family
`
`
`LDLS™ Laser-Driven Light Source
`Model EQ-77 High Brightness Broadband Light Source
`
`
`is in conformity with the requirements of the following standards
`
`
`EN61010-1 (2010)
`
`CISPR 11:2009
`
`EN61326-1:2013
`
`EN61000-3-2:2014
`
`EN61000-3-3:2013
`
`EN60825-1 (2007)
`
`
`
`
`
`Safety Requirements for Electrical Equipment for Measurement,
`Control and Laboratory Use: Part 1 – General Requirements
`Industrial, scientific and medical equipment - Radio-frequency
`disturbance characteristics - Limits and methods of
`measurement
`Electrical equipment for measurement, control and laboratory
`use. EMC requirements. General requirements
`Electromagnetic compatibility (EMC) - Part 3-2: Limits - Limits
`for harmonic current emissions (equipment input current less
`than or equal to 16 A per phase)
`Electromagnetic compatibility (EMC) - Part 3-3: Limits -
`Limitation of voltage changes, voltage fluctuations and flicker in
`public low-voltage supply systems, for equipment with rated
`current less than or equal to 16 A per phase and not subject to
`conditional connection
`Safety of laser products - Part 1: Equipment classification and
`requirements
`
`
`in accordance with the provisions of
`
`2006/95/EC
`2 004/108/EC
`2011/65/EU
`
`EU Low Voltage Directive
`EU Electromagnetic Compatibility Directive
`EU RoHS2 Directive
`
`
`
`
`
`
`
`
`
`
`
`13 July 2015
`Date
`
`
`
`
`Signed,
`
`
`
`
`
`
`
`Paul Blackborow, CEO
`
`at Woburn, Massachusetts USA
`
`Energetiq Ex. 2056, page 3 - IPR2015-01362
`
`

`
`Energetiq Ex. 2056, page 4 - |PR2015-01362
`
`
`
`
`
`Energetiq Ex. 2056, page 4 - IPR2015-01362
`
`

`
`
`
`TABLE OF CONTENTS
`
`
`
`
`

`Chapter 1 ...................................................................................................................................................... 1 
`General Information .................................................................................................... 1 
`Safety ............................................................................................................................................... 1 
`Chapter 2 ...................................................................................................................................................... 7 
`Description .................................................................................................................... 7 
`General ........................................................................................................................................... 7 
`Specifications ................................................................................................................................. 8 
`System Description .................................................................................................................... 10 
`Power Supply Controller ........................................................................................................... 11 
`Lamp House ................................................................................................................................ 13 
`Chapter 3 .................................................................................................................................................... 17 
`Installation................................................................................................................... 17 
`Unpacking .................................................................................................................................... 17 
`Connections ................................................................................................................................. 17 
`Installation Procedure ................................................................................................................ 21 
`Chapter 4 .................................................................................................................................................... 23 
`Operation .................................................................................................................... 23 
`Starting .......................................................................................................................................... 23 
`Stopping ....................................................................................................................................... 23 
`Serial Interface ............................................................................................................................ 24 
`Chapter 5 .................................................................................................................................................... 25 
`Troubleshooting ........................................................................................................ 25 
`Fault Indicator Block Diagram ................................................................................................ 25 
`Appendix A ................................................................................................................................................ 27 
`Engineering Drawings .............................................................................................. 27 
`Appendix B ................................................................................................................................................ 29 
`Revision History ........................................................................................................ 29 
`
`
`
`
`
`Energetiq Ex. 2056, page 5 - IPR2015-01362
`
`

`
`
`
`Energetiq Ex. 2056, page 6 - |PR2015-01362
`
`Energetiq Ex. 2056, page 6 - IPR2015-01362
`
`

`
`Safety
`
`
`
`
`C h a p t e r 1
`
`GENERAL INFORMATION
`
`WARNING
`
`This unit emits ultraviolet (UV) radiation that is
`harmful to humans. Avoid exposure to the
`direct or reflected output beam. Make certain
`that the appropriate output beam shields and
`optics are in place prior to energizing the unit.
`All
`interlocks must be satisfied prior
`to
`operation; failure to do so may
`lead to
`hazardous conditions.
`
`CAUTION
`
`levels of UV
`The EQ-77 emits dangerous
`radiation. Even short exposures to skin or eyes
`may cause burns. Ensure that only authorized
`personnel are in the vicinity of source during
`operation. Personnel in vicinity of operating
`source should wear protective eyewear, clothing,
`and gloves. Lighted UV warning lights and signs
`posted on doors to lab areas may help prevent
`accidental exposure.
`
`
`
`
`
`
`
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`1
`
`Energetiq Ex. 2056, page 7 - IPR2015-01362
`
`

`
`WARNING
`
`The EQ-77 utilizes an internal Class 4 IR laser
`capable of causing severe injury to eyes or skin.
`Do not open or attempt to service this unit.
`Contact Energetiq regarding any problems with
`the unit.
`
`
`
`
`
`
`
`General Precautions
`The output beam from the EQ-77 should be blocked when not in use with an electronic
`shutter or other appropriate beam blocking device. Due to the possibility of generating ozone
`with some models of EQ-77 when ambient oxygen is exposed to short wavelength light, the
`beam should always be enclosed in an appropriate beam pipe, tube, or enclosed space. We
`suggest purging any beam transport space with dry nitrogen gas.
`
`The EQ-77 source must also be cabled correctly and connected to a power source with a
`protective earth ground prior to operation.
`
`Refer to the Installation section of this manual in Chapter 3 for details of the facilities
`connections.
`
`There are no user-serviceable parts inside the EQ-77. For any problems encountered during
`operation, please contact Energetiq Technology for assistance. If there is a component
`failure, do not attempt to open the Power Supply Controller or Lamp House enclosure
`of the EQ-77.
`
`The EQ-77 utilizes a quartz lamp containing a high-pressure gas fill. Explosion of the
`lamp and possible injury from flying fragments can occur if the lamp is mishandled.
`
`Do not open the enclosure of either the Lamp House or the Power Supply Controller.
`Dangerous invisible infrared laser beams and hazardous voltages exist inside the
`Lamp House. Opening the chassis both voids the warranty and exposes the user to
`dangerous radiation and hazardous voltages.
`
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`
`
`
`2
`
`Energetiq Ex. 2056, page 8 - IPR2015-01362
`
`

`
`CAUTION
`
`Use of controls or adjustments or performance of procedures
`other than those specified herein may result in hazardous
`radiation exposure.
`
`
`
`
`
`Laser Information
`The EQ-77 uses a patented* laser drive system to excite a plasma that radiates in the UV as
`well as the visible bands. A class 4 laser is located in the Lamp House enclosure. The optical
`configuration of the Lamp House ensures that the direct laser beam cannot exit the unit. The
`EQ-77 laser product is designated as Class 1 during all normal operation.
`
`The parameters of the non-accessible internal laser are given below in Table 1.
`
`Wavelength
`Emission Type
`Laser Power for classification
`
`974 nm
`CW
`<36 mW via 7mm measurement
`aperture
`Beam Diameter
`~25 mm at aperture
`Divergence
`>100 mRad
`Transverse Beam Mode
`Diffuse
`Table 1: Embedded Laser Parameters
`
`
`No regular maintenance is required for the EQ-77. Any service to the system must be
`performed only by factory authorized and trained technicians. To avoid injury, under no
`circumstances should the user open or modify the Lamp House or Power Supply Controller
`enclosure.
`
`The unit must not be operated if the covers are removed or it is defective in any way. Contact
`Energetiq if any problems with the equipment are suspected.
`
` *
`
` US 7435982, 7786455, 8525138, 8969841, 9048000; Japan 5410958, 5628253; Korea 10-1507617; UK
`GB2450045
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`3
`
`Energetiq Ex. 2056, page 9 - IPR2015-01362
`
`

`
`Labels and Safety Notification
`
`
`The following safety labels appear on the product. Figure 1 shows the location of each label
`on the EQ-77 system.
`
`UV Hazard warning label – indicates hazardous levels of UV
`light are present.
`
`
`
`Manufacturer’s identification label – gives the manufacturer’s
`name and address, and the model, serial number, and date of
`manufacture of the equipment.
`
`
`
`Explanatory label – states the classification of the laser product.
`Class 1 is the lowest hazard level classification.
`
`Certification label – states that the equipment has been tested
`and verified to meet the standards indicated.
`
`Non-interlocked housing label – notifies of a potential hazard
`when covers are removed.
`
`
`
`
`
`
`
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`
`
`
`
`
`4
`
`Energetiq Ex. 2056, page 10 - IPR2015-01362
`
`

`
`Explanatory
`Label
`
`Certification
`Label
`
`Non-interlocked
`Housing Label
`
`
`
`
`UV Hazard
`Warning Label
`
`
`
`
`
`Manufacturer’s
`Identification label
`
`Figure 1: Safety Label Locations
`
`
`
`Safety Interlocks
`The EQ-77 is equipped with interlocks to prevent operation of the device when any of the
`following conditions are present:
`1. An external interlock is open
`2. No bulb is installed
`
`External Interlock
`External interlock pins are provided for the customer’s use (see Chapter 3 for connection
`details). Any suitable normally-open contact or solid-state switch can operate the interlock
`circuit. The contact or switch should be rated for 80mA minimum at 5VDC.
`
`The interlock circuit must be connected to enable the operation of the unit. Should the
`interlock connection open during operation or standby, the source is immediately disabled, and
`all light output from the aperture ceases.
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`5
`
`Energetiq Ex. 2056, page 11 - IPR2015-01362
`
`

`
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`
`
`
`6
`
`Energetiq Ex. 2056, page 12 - IPR2015-01362
`
`

`
`C h a p t e r 2
`
`DESCRIPTION
`
`General
`The EQ-77 is a broad-band lamp system for use in a wide variety of applications. The lamp
`produces high brightness, broad-band light from DUV wavelengths through visible and
`beyond. The output is very stable, and has a long lifetime before any service is required. A
`simple control interface ensures ease of use.
`
`Some of the advantages of the EQ-77 include:
`
` Very high brightness across complete spectrum
`– 190nm through visible and beyond
` Eliminates need for multiple lamps (replaces D2/Tungsten/Xenon Arc)
`– Simplified optical system
` Excellent spatial stability
`– Repeatable measurements
` Superior short and long term power stability
`– Repeatable measurements
` Electrodeless operation for long life
`– Reduced consumable costs
`– Minimal recalibration of instrument
`
`
`
`The EQ-77 system consists of a Power Supply Controller unit, Lamp House unit, and
`interconnecting cable. Connection to AC power is required for operation. Connections to
`nitrogen purge gas and cooling water are required. See Chapter 3 for connection details.
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`7
`
`Energetiq Ex. 2056, page 13 - IPR2015-01362
`
`

`
`Specifications
`Optical Performance
` Typical output spectrum: see Figure 2.
`
`
`
`
`
`Relative EQ-77 Spectral Radiant Flux (Typical)
`
`1.6
`
`1.4
`
`1.2
`
`1.0
`
`0.8
`
`0.6
`
`0.4
`
`0.2
`
`0.0
`
`Relative Spectral Radiant Flux (a.u.)
`
`
`
`200 250 300 350 400 450 500 550 600 650 700 750 800 850
`Wavelength (nm)
`
`
`
`Figure 2: Typical Output Spectrum
`
`Physical Specifications
`Dimensions (H x W x D)
` Lamp House: 128 x 175 x 102 mm (5.0 x 6.9 x 4.0 in)
` Power Supply Controller: 152 x 250 x 132 mm (6.0 x 9.8 x 5.2 in)
`
`
`Weight
` Lamp House: 2.2 kg (4.9 lb)
` Power Supply Controller: 2.9 kg (6.5 lb)
`
`
`
`
`8
`
`
`
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`
`Energetiq Ex. 2056, page 14 - IPR2015-01362
`
`

`
`Utility Requirements
` Electrical: 100 – 240 V~, 50/60 Hz, 350 W
` Purge gas: clean dry nitrogen, filtered to 5um, 20 psig (0.14 MPa) supply pressure
`– Fittings: 4mm push-to-connect
` Cooling water: >= 1 liter/minute, 18 – 30 °C
`– Fittings: 1/4-inch Swagelok
`
`
`
`Remote Interface
`Digital Inputs
` Type: Optocoupler LED
` Logic: Active High
`
`Input voltage: 5VDC
`
`Input current: 8mA
`
`
`External Interlock Input Only (pin 13)
` Type: Relay Coil
` Logic: Active High
`
`Input voltage: 5VDC
`
`Input current: 80mA
`
`
`Digital Outputs
` Type: Open collector to ground (digital common)
` Logic: Active Low
` Voltage: 30VDC max.
` Sink current: 30mA max.
`
`
`User Power
` Voltage: 5VDC, referenced to digital common
` Current: 400mA maximum
`
`Serial Interface
` Type: RS-485 4-wire (full duplex)
` Connector: Male 9-pin d-sub
` Termination: 120 ohms across receiver input (pins 2 and 7)
`
`Interface protocol: see Chapter 4.
`
`
`
`
`
`Environmental Requirements
`Operating
` Ambient temperature: 15–35°C
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`9
`
`Energetiq Ex. 2056, page 15 - IPR2015-01362
`
`

`
` Relative Humidity: non-condensing, 80% max. for temperatures up to 31°C,
`decreasing linearly to 50% max. at 35°C.
` Pollution Degree 2 (normally only non-conductive pollution; occasional,
`temporary condensation possible)
`
`Installation Category II
`
`Indoor use only
`Transport
` Temperature: -5–70°C
` Relative Humidity: non-condensing, 95% max.
`
`
`
`System Description
`As shown in Figure 3 the EQ-77 system consists of a Power Supply Controller unit, Lamp
`House, Controller to Lamp house interconnect cable (not shown), and power input cable (not
`shown). I/O interface connections (also not shown) are provided by the user.
`
`The following sections provide descriptions of the system components and controls, and give
`an overview of their functions. Refer to the “Installation” section of this manual (Chapter 3)
`for more detailed information.
`
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`
`
`
`10
`
`Energetiq Ex. 2056, page 16 - IPR2015-01362
`
`

`
`Optical Output
`(opposite side)
`
`Purge gas inlet/outlet
`fittings
`
`Retroreflector
`assembly
`
`Power Input
`Connector
`
`RS-485
`Connector
`
`Power Supply
`Controller
`
`Lamp
`House
`
`
`
`Power Supply Controller
`The Power Supply Controller contains:
` Laser power supply
` Thermo-electric cooler control system for laser
` Control electronics
` Status indicator LEDs
`
`Interface connectors
`
`
`
`External features (refer to Figure 3):
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`Lamp House Control
`Connector
`
`Water inlet/outlet
`fittings
`
`Status
`Indicators
`
`I/O Interface
`Connector
`
`Lamp House
`Control
`Connector
`
`Figure 3: EQ-77 Lamp System
`
`
`
`11
`
`Energetiq Ex. 2056, page 17 - IPR2015-01362
`
`

`
` Status Indicator LEDs
`These five LEDs indicate the system status. The function of these indicators is shown below
`in
`
`Table 2.
`
`
`
`LED Label
`POWER ON
`LAMP ON
`LASER ON
`
`CONTROLLER FAULT
`
`LAMP MODULE FAULT
`
`Meaning (when lit)
`AC power is connected to the EQ-77 Power Supply Controller
`UV Light is on
`Laser power is ON and laser light is being produced within the Lamp
`House
`One of the following has occurred in the Power Supply Controller:
`1. External interlock open
`2. Controller internal temperature too high
`3. Laser power not reaching setpoint
`4. Laser temperature fault
`5.
`Internal power supply voltage low
`One of the following has occurred in the Lamp House module:
`1. Control Cable not connected properly
`2. Lamphouse internal temperature too high
`3.
`Ignition Failure
`
`
`
`Table 2: Status Indicator LED Functions
`
`
`
`Input/Output (I/O) Connector
`Provides access to control and status signals. See Chapter 3 for pin assignments and functions.
`
`This is the only operator interface to the EQ-77 – there are no local controls. Energetiq offers
`the EQ-99-RC Remote Control Module which connects to the Input/Output connector of
`the EQ-77 or EQ-99 and provides a means of local control. Contact Energetiq for additional
`information.
`
`12
`
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`
`Energetiq Ex. 2056, page 18 - IPR2015-01362
`
`

`
`Power Input Connector
`This is an IEC 320 inlet connector for AC power input. See Chapter 3 for detailed
`information.
`
`Lamp House Control Connector (21-pin mixed D-sub)
`Provides various power and control signals to/from the Lamp House module. No other
`connector or cable may be used with the EQ-77 other than the one supplied.
`
`RS-485 Connector (9-pin D-sub)
`Connector for optional RS-485 interface. See Chapters 3 and 4 for electrical details and
`commands.
`
`
`
`Lamp House
`The Lamp House assembly contains:
` Lamp
`
`Igniter
` Laser
` Thermoelectric cooler
`
`IR pumping optics
` Output windows
` Retroreflector assembly
` Laser ON indicator
`
`Interface connectors
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`13
`
`Energetiq Ex. 2056, page 19 - IPR2015-01362
`
`

`
`
`
`Lamp
`Window
`(opposite side)
`
`Nitrogen Purge
`Inlet
`
`Retroreflector
`Assembly
`
`Nitrogen Purge
`Outlet
`
`Laser ON
`Indicator
`
`Control
`Connector (to
`Controller)
`
`
`
`Cooling Water
`Inlet
`
`Cooling Water
`Outlet
`
`Figure 4: Lamp House Assembly
`
`
`
`External features (refer to Figure 4):
`
`Lamp Windows
`The lamp windows at the optical output provide protection from the high pressure bulb inside
`the Lamp House. An internally-threaded SM1 adapter is provided for easy connection of
`optical hardware.
`
`Nitrogen Purge Inlet / Outlet
`These are the fittings for the required nitrogen purge gas. With no purge, ozone will form
`from atmospheric oxygen and attenuate the light output in the 220 – 280nm band. In
`addition, atmospheric oxygen and water vapor will attenuate the output below 200nm.
`
`Cooling Water Inlet / Outlet
`These fittings are for connection of cooling water required by the Lamp House.
`
`14
`
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`
`Energetiq Ex. 2056, page 20 - IPR2015-01362
`
`

`
`Laser On Indicator
`This LED is illuminated when the laser is ON.
`
`Control Connector (21-pin mixed D-sub)
`Provides various power and control signals to/from the Power Supply Controller. No other
`connector or cable may be used with the EQ-77 other than the one supplied.
`
`
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`15
`
`Energetiq Ex. 2056, page 21 - IPR2015-01362
`
`

`
`
`
`16
`
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`
`Energetiq Ex. 2056, page 22 - IPR2015-01362
`
`

`
`C h a p t e r 3
`
`INSTALLATION
`
`Unpacking
`Upon arrival, start by inspecting all parts of the system for completeness and any damage
`incurred in shipping. The EQ-77 shipping box should contain:
`
`1) EQ-77 Power Supply Controller unit
`1) EQ-77 Lamp House unit
`1) Black interconnecting cable from Lamp House to Power Supply Controller (21-
`pin mixed D-sub).
`
`OPTIONAL
`1) EQ-99-RC Remote Control Module with interlock connector
`1) I/O cable with 15 pin D- connector at each end
`
`If any part is missing or appears damaged, contact Energetiq immediately. Do not attempt to
`substitute any parts. There are no user-serviceable parts inside the EQ-77 Lamp House or
`Power Supply Controller unit.
`
`Connections
`Installation of the EQ-77 consists of connecting electrical, water, and purge gas supplies, and
`connecting the Lamp House module to the user’s equipment. Read "Installation Procedure"
`first before making any connections.
`
`Electrical Power
`The EQ-77 requires 100 – 240 V~, 50/60 Hz input voltage. Power consumption is 350 W
`maximum during normal operation. Some OEM versions have higher current requirements.
`
`Purge Gas
`Purge gas required for operation. With no purge, ozone will form from atmospheric oxygen
`and attenuate the light output in the 220 – 280nm band. In addition, atmospheric oxygen and
`water vapor will attenuate the output below 200nm.
`
`Connect a source of nitrogen purge gas to the port on the Lamp House. The fitting is a push-
`to-connect type, sized for 4mm O.D. tubing.
`
`Clean and dry nitrogen from either a Dewar or research-grade N2 bottle is recommended. Do
`not use any other purge gas. Grade 6 or better gas purity is recommended to maintain
`cleanliness of the optics, and gas should be filtered to <5um. Supply pressure should be 20
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`17
`
`Energetiq Ex. 2056, page 23 - IPR2015-01362
`
`

`
`psig (0.14 MPa). With a 20 psig inlet pressure, the EQ-77 will consume approximately 0.5 slm
`of flow.
`
`Cooling Water
`Connect a source of cooling water to the 1/4-inch Swagelok fittings on the Lamp House. A
`flow rate of at least 1 liter/minute is required. Inlet temperature should be between 18°C and
`30°C
`
`Optical Interface
`See drawing OTD-7111 in Appendix A for mechanical layout of the Lamp House.
`
`Signal Connections
`The EQ-77 is controlled through the remote I/O connector.
`
`Table 3 gives the pin assignments and functions for this interface. Connect to the user’s
`control system using a suitable cable. Mating connector is a standard high-density 15-pin d-
`sub male (for example, Amp part no. 748364-1 with contacts 1658670-2).
`
`Optionally, connect a model EQ-99-RC remote control module to the I/O connector using
`the supplied cable
`
`User I/O can be powered either by the EQ-77 internal isolated power supply, or an external
`supply.
`
`Figure 5 shows connection schematics for both configurations.
`
`Lamp House I/O
`The black 21-pin mixed D-sub cable is used to connect the Lamp House to the Power Supply
`Controller. This cable should be connected before AC power is applied to the Power Supply
`Controller.
`
`RS-485 Interface
`Connector J2 is a standard 9-pin male d-sub. Mating connector is a standard 9-pin female d-
`sub (for example, AMP part no. 205203-3 with contacts 1-66504-0). See Table 4 for pin
`connections.
`
`18
`
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`
`Energetiq Ex. 2056, page 24 - IPR2015-01362
`
`

`
`Description
`Commands (Inputs)
`LAMP OPERATE
`
`EXTERNAL INTERLOCK
`
`
`
`Status Indicators (Outputs)
`LAMP ON
`LASER ON
`LAMP MODULE FAULT
`CONTROLLER FAULT
`ISOLATED +5V SUPPLY
`DIGITAL COMMON
`RESERVED
`RESERVED
`
`Pin #
`
`
`Details
`
`
`12
`
`13
`
`OPERATE REQUEST, apply +5V (referenced to digital
`common) to initiate start cycle
`EXTERNAL INTERLOCK, apply +5V (referenced to
`digital common) to close interlock and allow operation.
`Current draw is 80mA at 5VDC.
`
`
`
`
`Pulled to digital common when ON
`1
`Pulled to digital common when ON
`2
`Pulled to digital common when OK, float on FAULT
`3
`Pulled to digital common when OK, float on FAULT
`4
`400mA maximum, referenced to digital common
`5
`6,7,8,9 Galvanically isolated from system
`10, 11
`Do not connect
`14,15
`Do not connect
`
`
`
`
`
`
`
`Description
`TRANSMIT A (–)
`TRANSMIT B (+)
`RECEIVE A (–)
`RECEIVE B (+)
`GROUND
`RESERVED
`
`Table 3: I/O Connector Pin Assignment
`
`Pin #
`8
`3
`7
`2
`5, 9
`1, 4, 6
`
`Details
`From EQ-77, connect to host RECEIVE A
`From EQ-77, connect to host RECEIVE B
`To EQ-77, connect to host TRANSMIT A
`To EQ-77, connect to host TRANSMIT B
`Galvanically isolated from system
`Do not connect
`
`Table 4: RS-485 Interface Pin Assignment
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`19
`
`Energetiq Ex. 2056, page 25 - IPR2015-01362
`
`

`
`Contact or
`solid-state switch
`
`Contact or
`solid-state switch
`
`Load
`
`Contact or
`solid-state switch
`
`Contact or
`solid-state switch
`
`Load
`
`+-
`
`User
`5VDC supply
`
`N/C
`
`N/C
`
`100
`
`5V Power
`Pin 5
`
`499
`
`0.1uF
`
`Input
`Pin 12
`
`Input
`Pin 13
`
`Outputs
`Pins 1, 2, 3, 4
`
`5V Return
`Pins 6, 7, 8, 9
`
`100
`
`5V Power
`Pin 5
`
`499
`
`0.1uF
`
`Input
`Pin 12
`
`Input
`Pin 13
`
`Outputs
`Pins 1, 2, 3, 4
`
`5V Return
`Pins 6, 7, 8, 9
`
`0.1uF
`
`Isolated
`5VDC supply
`
`+-
`
`0.1uF
`
`Isolated
`5VDC supply
`
`+-
`
`
`Figure 5: Remote Interface Schematic
`
`
`
`20
`
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`
`Energetiq Ex. 2056, page 26 - IPR2015-01362
`
`

`
`
`Installation Procedure
`1. Place the Power Supply Controller on a stable surface. The Power Supply Controller
`must be placed within 2 meters of the Lamp House due to the length of the
`interconnecting cable. See drawing OTD-6881 in Appendix A for dimensional details.
`2. Connect the Lamp House unit optical output to the user equipment (after removing
`the window cover). The beam should always be either directly coupled to a fiber optic
`cable, or enclosed in an appropriate beam pipe, tube, or enclosed space and purged
`with nitrogen. Operating the source without any output target or beam transport is
`not recommended, and may lead to unsafe operating conditions. Consult Energetiq for
`applications information and suggested configurations.
`The Lamp House should be mounted in the orientation shown in Figure 4, with the
`output window on the side and the lamp mounting surface oriented horizontally. The
`lamp has been factory aligned in this position. Mounting the lamp in a different
`orientation will cause the plasma position inside the bulb to shift slightly and may
`cause performance variations.
`Setup the Lamp House unit with appropriate ultraviolet safety measures and laser light
`safety measures in place. It is recommended that any enclosure or aperture-blocking
`hardware utilize switches wired to the EQ-77 external interlock circuit.
`See drawing OTD-7111 in Appendix A for mechanical details of the Lamp House.
`3. Connect the black 21-pin mixed D-sub interconnect cable from the Power Supply
`Controller to the Lamp House.
`4. Connect nitrogen purge gas to the Lamp House. Refer to “Connections” above.
`5. Connect cooling water to the Lamp House. Refer to “Connections” above.
`6. Connect user’s control system to the I/O interface connector per Table 3 and Figure
`5.
`7. Alternately, if using the EQ-99-RC Remote Control Module, place it on a clean rigid
`surface. Install the supplied 15-pin cable from the Power Supply Controller to the
`EQ-99-RC.
`8. Connect AC input power source to the Controller Unit.
`
`
`The system is now ready to operate.
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`21
`
`Energetiq Ex. 2056, page 27 - IPR2015-01362
`
`

`
`
`
`22
`
`
`
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`
`Energetiq Ex. 2056, page 28 - IPR2015-01362
`
`

`
`C h a p t e r 4
`
`OPERATION
`
`Starting
`Once the lamp is set up properly, verify that all personnel that will be in contact with the lamp
`system are aware of the potential hazards involved. It is the responsibility of the user to verify
`that the lamp is being used safely.
`
`This example assumes the use of the EQ-99-RC Remote Control Module to provide local
`control. If using a custom control system, substitute the appropriate digital input and output
`lines from Table 3 for the switches and LEDs described below.
`
`1. With the EQ-99-RC Remote Control Module connected properly, review the status
`LEDs on the Power Supply Controller. The POWER ON LED should be lit, and
`neither the CONTROLLER FAULT nor LAMP MODULE FAULT LEDs should
`be lit.
`2. Turn on the OPERATE Switch (place switch in UP position).
`3. Within several seconds the LASER ON LED will light. Laser light is now present in
`the Lamp House.
`4. In approximately 20-150 seconds the igniter will be turned on automatically and the
`plasma will ignite. The LAMP ON LED will be lit. The duration of the warm up time
`(20-150 seconds) will depend on the temperature and previous operating condition of
`the EQ-77. The EQ-77 will automatically detect when the unit has reached the
`optimum conditions for ignition.
`5. If a bulb fails to ignite within 150 seconds after the OPERATE switch was activated,
`the LASER ON LED will go out, the LAMP FAULT LED will be lit, and LAMP ON
`LED will remain off. This is very unusual. However, if this occurs, turn the
`OPERATE switch to the OFF position (down) and begin at Step 1 again. If this
`occurs multiple times, contact Energetiq service
`
`Stopping
`To turn the LAMP off, simply turn the OPERATE Switch to the OFF position. If the lamp
`will not be used for some time, the AC supply can be turned off.
`
`To minimize wear on the ignition components, it is best to avoid frequently starting and
`stopping the lamp. It is recommended to run the lamp continuously if long off-periods are
`not required.
`
`EQ-77 Operation Manual Rev. 2
`
`
`
`
`23
`
`Energetiq Ex. 2056, page 29 - IPR2015-01362
`
`

`
`Serial Interface
`The RS-485 serial interface is provided for user adjustment of laser operating power, in order
`to increase or decrease light output. The EQ-77 is shipped with the laser power set to a factory
`default value of 100% of full scale power. Laser power can be increased or decreased in
`increments of 0.5% of full scale. The following describes the serial commands and their
`functions.
`
` Commands consist of a single ASCII character, case-sensitive. This can be transmitted to the
`EQ-77 via a terminal emulation program, or the user's control system. Response from the
`EQ-77 will be a string of ASCII characters, format depending on the command issued.
`
`
`
`Command
`character
`U
`
`D
`
`Q
`
`F
`
`B
`
`?
`
`Increases the present laser power
`setpoint by 0.5% of full scale
`Decreases the present laser
`power setpoint by 0.5% of full
`scale
`Queries the value of the present
`laser power setpoint (in % of full
`scale)
`Resets laser power setpoint to
`factory default value
`Saves present settings to flash
`memory. This command should
`be issued after the settings are at
`their desired values. If not,
`changes will be lost if power to
`the EQ-77 is interrupted.
`Displays a help menu listing the
`available commands
`
`Function
`
`Reply from EQ-77
`
`Power = XX.X%
`
`Power = XX.X%
`
`Power = XX.X%
`
`Power = XX.X%
`
`Calibration data copied to FLASH
`Memory.
`
`EQ77 firmware v02d (FRM-7117 R005)
`build date 11/18/15

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket