throbber
PrinterNOx
`
`Operating Manual – English
`Gebrauchsanweisung - Deutsch
`
`
`
`
`
`1
`
`Ex. 2027-0001
`
`

`
`Ex. 2027-0002
`
`
`
`
`
`
`
`Ex. 2027-0002
`
`

`
`
`
`Contents
`
`I.
`
`Important safety information – English ................................ 3
`Introduction ..................................................................... 4
`Medical Use ...................................................................... 5
`
`NO2 Production ................................................................. 6
`
`Environmental Safety ........................................................ 6
`
`Operation......................................................................... 7
`
` 1. Measurement ............................................................10
`
` 2. Calibration ................................................................17
`
` 3. Set Alarms................................................................23
`
` 4. Report to printer........................................................24
`
` 5. Status ......................................................................24
`
` 6. Setup .......................................................................25
`
` 7. Calculate Flow ...........................................................26
`
` Battery operation.............................................................27
`
`Paper Loading..................................................................28
`
`Battery Maintenance.........................................................29
`
`Cleaning Procedure...........................................................31
`
`Servicing.........................................................................31
`
`Symbols..........................................................................32
`
`Electrical Environment ......................................................32
`
`Specifications ..................................................................32
`
`II. Wichtige Sicherheitshinweise – Deutsch ..............................36
`Einleitung........................................................................37
`Medizinische Anwendung...................................................38
`NO2 Produktion ................................................................39
`Umweltschutz ..................................................................40
`Betrieb............................................................................40
` 1. Messung ...................................................................44
` 2. Kalibration ................................................................52
` 3. Alarmeinstellungen ....................................................58
` 4. Bericht an den Drucker...............................................59
`1
`
`
`
`
`
`
`
`
`
`
`Ex. 2027-0003
`
`

`
`
`
`5. Status .................................................................... ..59
` 5. Status ......................................................................59
`
`6. Setup ..................................................................... ..60
` 6. Setup .......................................................................60
`
`7. Durchfluss berechnen ............................................... ..61
` 7. Durchfluss berechnen .................................................61
`
`Batteriebetrieb .............................................................. . .62
` Batteriebetrieb ................................................................62
`Papier einlegen .............................................................. ..63
`
`Papier einlegen ................................................................63
`Batterie wartung ............................................................ ..64
`
`Batterie wartung ..............................................................64
`
`Reinigung........................................................................66
`Reinigung ...................................................................... ..66
`Wartung ........................................................................ ..67
`
`Wartung..........................................................................67
`Symbole ....................................................................... ..67
`
`Symbole .........................................................................67
`Elektrische Umgebung .................................................... ..68
`
`Elektrische Umgebung ......................................................68
`Technische Daten ........................................................... ..68
`
`Technische Daten .............................................................68
`
`Customer Contact Information ......................................... ..70
`Customer Contact Information ...........................................70
`
`
`
`
`
`
`2
`2
`
`Ex. 2027-0004
`
`Ex. 2027-0004
`
`

`
`
`
`
`
`PrinterNOx
`
`Operating manual
`
`
`I. Important safety information - English
`
`Nitric oxide is a highly toxic gas, exposure to levels as low as 200
`parts per million (ppm) may be fatal even after very short
`exposures, and prolonged exposure to much lower levels may be
`harmful. All users of nitric oxide must familiarise themselves with
`the appropriate safety regulations. When calibrating the PrinterNOx
`monitor it is important to ensure that calibration gas is adequately
`vented and safely exhausted. Nitric oxide reacts with oxygen in air
`to form nitrogen dioxide, an extremely toxic gas with a maximum
`allowable industrial ambient level of 5ppm.
`
`The PrinterNOx nitric oxide/nitrogen dioxide monitor has been
`designed to the highest standards and incorporates many inbuilt
`self-checks. As with any measurement device, however,
`malfunction is possible. When used in the administration of nitric
`oxide the PrinterNOx must not, therefore, be incorporated as a
`primary part of the NO delivery system, but rather as a safety check
`on the delivery system. The delivery system should be constructed
`in such a way as to minimise possible errors. CareFusion Ltd does
`not accept any liability for injury resulting from malfunction and
`failure to follow these guidelines. Failure to follow instructions in the
`remainder of this manual may also result in dangerously inaccurate
`measurements being obtained.
`
`3
`
`Ex. 2027-0005
`
`

`
`
`
`Introduction
`
`The PrinterNOx is a combined Nitric Oxide / Nitrogen Dioxide
`monitor based on electrochemical fuel cells and may be operated
`directly from the mains power or from the internal rechargeable
`NiCad battery pack. Fuel cells work through the reaction of the
`target gas with an electrolyte at one electrode and oxygen (from
`ambient air) at the other. This reaction generates an electrical
`current proportional to gas concentration. In comparison with other
`technologies (such as chemiluminescent detectors) electrochemical
`sensors are compact, inexpensive and much easier to use with zero
`warm up time and simple calibration. The PrinterNOx is designed to
`monitor delivered NO used as a therapy for adult respiratory distress
`syndrome and pulmonary hypertension on ventilated patients. The
`more toxic and non therapeutic by-product, NO2, is also monitored.
`The NO measurement has a resolution of 0.05 parts per million
`(ppm) with a full scale of 100ppm and the NO2 measurement has a
`resolution of 0.05ppm with a full scale of 20ppm. The gas to be
`monitored is sampled using a constant flow (250ml/min), low dead
`space, sidestream system. The system includes a water trap, with
`integral hydrophobic filter, and humidity conditioning circuit to
`protect the fuel cells against extremes of humidity and particulate
`contamination. The measurements are displayed on a large graphic
`liquid crystal display and can be printed as a chart, on the integral
`thermal printer, together with calibration due date and alarm
`settings. The last 24 hours of NO and NO2 measurements taken are
`recorded and stored in the internal battery backed memory and can
`be printed or uploaded on the RS232 interface for computer storage.
`The PrinterNOx includes a number of safety features and alarms.
`These include upper and lower alarms for NO and an upper alarm for
`NO2. An external alarm cable is available, and when connected to a
`NO delivery system, is used to trigger an alarm or initiate a safe
`operating mode if high concentrations of gas are detected. This
`concentration is 100ppm for NO, 9ppm NO2 or alarm level whichever
`4
`
`Ex. 2027-0006
`
`

`
`
`
`is the greater. The flow of gas in the sampling line is continuously
`monitored and an alarm sounded if the line is occluded or the water
`trap is full. A low battery condition is also alarmed.
`
`The PrinterNOx has been designed with ease of use and reliability as
`of foremost importance. To ensure accurate results, however, it is
`essential to familiarise yourself with the contents of this short
`manual.
`
`Medical Use
`
`Nitric oxide (NO) is evolving as a novel therapy used to reduce
`pulmonary arterial pressure and improve gas exchange in paediatric
`and adult intensive care settings. The NO is usually added to the
`ventilator circuit from a cylinder of NO in nitrogen with a
`concentration up to 1000ppm and diluted in the circuit to between 1
`and 40ppm.
`
`Therapeutic concentrations must be well controlled as levels as low
`as 100ppm may cause lung injury.
`
`For safety reasons sources with greater than 1000ppm NO should
`not be used. Lower concentrations carry less risk either to the
`patient in the case of accidental overdose, or to the hospital staff in
`the case of leakage. However, if moderately high doses of NO are to
`be administered then the decrease in the maximum FiO2 must be
`considered and is given by the relationship:
`
`FiO2 (max.) = (1 - Required NO Conc./Source NO Conc.) X
`
`100%
`
`Therefore, if 40ppm is required from a source of 500ppm NO in
`nitrogen then the maximum oxygen available would be:
`
`FiO2 (max.) = (1 - 40 / 500) X 100% = 92%
`
`5
`
`Ex. 2027-0007
`
`

`
`
`
`Methods for therapeutic administration of nitric oxide depend on the
`exact type of ventilator and ventilator circuit to be used and have
`not yet been standardised. It is therefore the responsibility of the
`user to pay due regard to current medical practice when using the
`unit to monitor concentrations of inhaled nitric oxide and nitrogen
`dioxide.
`
`NO2 Production
`
`Nitric oxide in the presence of oxygen undergoes the following
`chemical reaction to form nitrogen dioxide, (a more toxic gas than
`nitric oxide):
`
`2NO + O2 →→→→ 2NO2
`
`The rate of conversion is highly dependent on the NO concentration
`and this has some important implications for the NO delivery
`system.
`
`Firstly, sources of NO with excessively high concentrations should be
`avoided.
`
`Secondly, the administered concentration should be kept as low as is
`clinically acceptable.
`
`Finally, the administration system should mix the incoming NO
`quickly and effectively to rapidly reduce its concentration to the
`therapeutic level where NO2 production will proceed at a much
`slower rate.
`
`Environmental Safety
`
`To eliminate environmental pollution with NO and NO2 the
`PrinterNOx is supplied with two NOxSORB chemical scavengers for
`removing these gases from the exhaust. The NOxSORB works by
`converting NO to NO2, which is then chemically absorbed.
`6
`
`Ex. 2027-0008
`
`

`
`
`
`Connect a NOxSORB to the exhaust of the PrinterNOx, as shown
`overleaf, orientated such that the arrow is pointing away from the
`instrument. The NOxSORB should be replaced after 24 hours or
`when the mauve spheres turn dark brown. When re-ordering please
`quote 36-SORB100 for a box of 10 of 36-SORB200 for a box of 20.
`
`Environmental pollution with NO and NO2 may also arise when leaks
`are present in the connections to the source of NO.
`
`Operation
`
`Before operating the PrinterNOx, ensure that the unit has been fully
`charged using the AC adapter supplied. (See Charging Procedure.)
`The PrinterNOx may also be operated with the internal rechargeable
`batteries.
`
`Connect the water trap with integral hydrophobic PTFE filter and
`sampling line to the PrinterNOx as shown below:
`
`
`
`Water trap
`retaining
`screws
`
`Water
`Trap
`
`Sampling
`Line
`
`PrinterNOx
`
`NOxSORB
`
`Micro
`Medical
`
`Alarm
`
`1
`
`6
`
`2
`
`7
`
`3
`
`8
`
`4
`
`9
`
`5
`
`0
`
`PAPER
`
`C AN C EL
`
`C H AR G IN G
`
`EN T ER
`
`Exhaust
`port
`
`
`Switch on the PrinterNOx. Please note that the water trap must
`
`
`
`be used at all times during measurement to avoid
`
`contamination of the fuel cells.
`
`
`
`7
`
`Ex. 2027-0009
`
`

`
`
`
`The on/off switch is located at the rear of the unit:
`
`
`
`On/Off
`switch
`
`External
`alarm
`socket
`
`Contrast
`
`O
`
`1
`
`Power Input
` 500mA
` 12V
`
`External
` Alarm
`
` RS232/
`Analogue
` Output
`
`Display contrast
`adjustment
`
`AC Adapter
`socket
`
`RS232
`socket
`
`An introductory screen giving the software version will be displayed
`briefly:
`
`
`
` MICRO MEDICAL
`
` PrinterNOx
`
`
` V X.XX
`
`
`During this time the following information will be printed:
`
`
`
`12/04/95 16:14
`NO calibration due : 01/01/96
`NO2 calibration due : 07/01/96
`NO upper alarm : 15
`NO lower alarm : 5
`NO2 upper alarm : 4
`
`8
`
`
`
`Ex. 2027-0010
`
`

`
`
`
`This information can be checked and the date, time, calibration or
`alarm settings may be adjusted prior to performing a measurement.
`
`(See sections on Setup, Calibration, and Set Alarms)
`
`The calibration due date is set to occur thirty days after the last
`calibration. However, calibration may be performed at any time
`before this date to improve measurement accuracy. Measurements
`can be made after the calibration due date has passed but
`consideration must be given to the gas sensors which exhibit a slow
`decline in sensitivity over time. This loss of sensitivity is typically
`less than 2% per month but could be more if they are exposed to
`adverse conditions such as exposure to gas concentrations above
`the measurement range or to solvents such as alcohol.
`
`If the calibration due date has passed then the following reminder
`will be displayed both after the introductory screen and when
`commencing measurement:
`
` Calibration due!
`
` NO 12/12/95
` NO2 1/4/96
`
` Press ENTER to
` Continue
`
`
`
`Once ENTER is pressed the first menu is displayed and allows
`selection of several options:
`
`
`
`9
`
`Ex. 2027-0011
`
`

`
`
`
` Select Option
`1 Measurement
`2 Calibration
`3 Set Alarms
`4 Report to printer
`5 Status
`6 Setup
`7 Calculate Flow
`
`
`
`Whilst this display is active the sampling pump is turned off,
`measurements are not taken, and the gas level alarms are inactive.
`
`If no options are selected then the unit defaults to option 1
`“Measurement” after 45 seconds. This is done so that the unit
`defaults to monitoring gas levels with alarms activated in the event
`that the instrument was inadvertently left on the first screen or was
`switched off and on.
`
`During all other operations, a short warning alarm will occur, after
`45 seconds of inactivity, to remind the user that measurement mode
`is not operational.
`
`1. Measurement
`
`To obtain maximum accuracy from the measurement careful
`consideration must be given to the NO administration system in use.
`
`Gas measurement with ventilators producing cyclic flows may give
`an overestimate or under estimate of the average NO concentration.
`
`This is due to localised variation of NO concentration within the
`delivery system. Dependent on the measurement point, therefore,
`the sampled average may not correspond to the delivered average.
`It is therefore important to ensure that the NO delivery system is
`designed to avoid large variations in the level of delivered NO.
`
`10
`
`Ex. 2027-0012
`
`

`
`
`
`The PrinterNOx utilises a sidestream gas system, which is servo
`controlled to sample at a constant rate of 250 ml/min. By supplying
`the gas sensors with this constant flow, a high degree of accuracy
`and sensitivity can be achieved.
`
`To carry out NO / NO2 measurement, press “1”.
`
`The sampling pump will start and the following screen will be
`displayed:
`
`000
`000
`
`..
`
`NO
`NO
`
`15
` 5 p
` p
` 4 m
`2
` Press ENTER for menu
`
`
`
`The system should now be tested for leaks by occluding the
`sampling line. The pump will be heard to rise in pitch as the servo
`system tries to overcome the restriction.
`
`As the pump cannot maintain the required flow rate then the
`measurement will cease to be displayed, the pump will continue to
`run, and the following message will be displayed:
`
` Gas Line Occluded
` or water trap full
` Please clear blockage
` or empty water trap
` and filter.
`
`11
`
`
`
`Ex. 2027-0013
`
`

`
`
`
`When the occlusion is removed, the display will automatically revert
`to the measurement screen. Repeat this procedure by occluding the
`exhaust port.
`
`The sampling line should now be connected to the administered gas,
`normally as close to the patient as possible. The sampling port must
`be positioned in such a way that water cannot enter the tubing,
`causing rapid filling of the water trap.
`
`If the gas line appears to be continuously occluded, remove the
`water trap and empty any water into a suitable container. Ensure
`that there is no occlusion on the exhaust port.
`
`Under exceptional circumstances, water may pass through a
`damaged water trap and obstruct the inlet to the PrinterNOx. In this
`case turn the unit off and draw any liquid out of the inlet with a
`syringe connected directly to the inlet. Replace the water trap with
`a new one, turn the unit on, and return to the measurement screen.
`
`On the left-hand side of the NO legend are the upper and lower
`alarm levels for NO. On the left-hand side of the NO2 legend is the
`upper alarm level for NO2.
`
`On the bottom line, the rectangle will blink on or off every time a
`measurement is updated (every 2.6 seconds) to indicate correct
`functioning of the instrument.
`
`The gas concentrations are given in parts per million (ppm).
`
`If the printer option is turned on (see Setup section), then the
`following printout will be obtained during measurement:
`
`
`
`12
`
`Ex. 2027-0014
`
`

`
`
`
`12/04/95 16:14
`NO calibration due : 01/01/96
`NO2 calibration due : 07/01/96
`NO upper alarm : 15
`NO lower alarm : 5
`NO2 upper alarm : 4
`
`Patient :...............................
`
`Hospital :...............................
`
`Technician :...............................
`
` 5 NO 15 0 NO2 4
`
`1715
`
`1730
`
`
`Spaces for entering the patient name or ID, hospital name, and
`technician name are provided.
`
`
`
`This is followed by the axes for the NO and NO2 graphs.
`
`These graphs are auto scaled to the current alarm settings.
`
`A section of the graph will be printed once every 32 samples and
`includes a time stamp every 15 minutes on the left-hand side. Each
`dot on the graph represents the average of 4 measurements.
`
`13
`
`Ex. 2027-0015
`
`

`
`
`
`If the measured gas goes outside the range of the alarm settings
`then an audible alarm and a red flashing visual alarm will be
`generated. In addition, the relevant indicator on the liquid crystal
`display will flash. The graph is then replaced by a printout of each
`individual measurement together with a time stamp and an alarm
`indicator as shown below:
`
`12/04/95 16:14
`NO calibration due : 01/01/96
`NO2 calibration due : 07/01/96
`NO upper alarm : 15
`NO lower alarm : 5
`NO2 upper alarm : 4
`
`Patient :...............................
`
`Hospital :...............................
`
`Technician :...............................
`
` 5 NO 15 0 NO2 4
`
`1715
`
`1730
`
`1741 Alarm NO = 22.3 NO2 = 1.85
`1741 Alarm NO = 22.4 NO2 = 1.95
`1741 Alarm NO = 23.1 NO2 = 2.10
`1742 Alarm NO = 23.0 NO2 = 1.90
`
`
`If the NO measurement exceeds 100ppm or NO2 exceeds 9ppm then
`the external alarm, accessible at the rear of the instrument, is
`
`14
`
`Ex. 2027-0016
`
`

`
`
`
`activated. If the NO2 alarm level is set above 9ppm, then the
`external alarm comes into operation at the set level for NO2. The
`external alarm consists of a pair of normally closed contacts, which
`are latched open when the above condition occurs. This can be used
`to sound an external alarm or switch a NO delivery system into safe
`operating mode in systems equipped with this facility.
`
`The display will change to:
`
`15
` 5 p
` p
` 4 m
`
`148
`.
`987
`.
`
`NO
`NO2
`
` CUT OUT: PRESS ENTER
`
`
`
`Measurements will continue but it should be noted that the
`measurement range would be limited to just over 100ppm NO and
`20ppm NO2 and that the actual concentrations could be significantly
`higher.
`
`Press enter to clear the external alarm when the reading drops
`below the external alarm levels.
`
`When this condition occurs, the cause should be immediately
`investigated. However, caution should be observed when shutting
`down the supply of NO, as the sudden cessation of NO can be more
`detrimental to the patient than a temporary overdose of NO.
`
`Ideally, the external alarm should be used to switch the delivery
`system into a ‘safe’ operating mode or switch in a backup supply.
`
`15
`
`Ex. 2027-0017
`
`

`
`
`
`Should the water trap fill during the course of the measurement, it
`will overflow into the integral hydrophobic PTFE filter, cause an
`obstruction, and the following message will be displayed:
`
` Gas Line Occluded
` or water trap full
` Please clear blockage
` or empty water trap
` and filter.
`
`
`
`Unscrew the two retaining screws, and remove the water trap from
`the PrinterNOx.
`
`Empty the accumulated water into a suitable container by
`connecting a syringe (20ml minimum size) with the adapter supplied
`and pushing air through the assembly as shown below.
`
`20
`
`15
`
`10
`
`5
`
`Syringe
`adapter
`
`16
`
`
`
`
`
`
`
`Ex. 2027-0018
`
`

`
`
`
`Reconnect water trap to the PrinterNOx and secure with the two
`retaining screws. If the ‘Gas Line Occluded’ message reappears,
`disassemble and replace the water trap with a new one.
`
`Under exceptional circumstances, water may pass through a
`damaged water trap and obstruct the inlet to the PrinterNOx. In this
`case turn the unit off and draw any liquid out of the inlet with a
`syringe connected directly to the inlet. Replace the water trap with
`a new one, turn the unit on, and return to the measurement screen.
`
`The alarm may be muted for a 45-second period by pressing the
`CANCEL key.
`
`2. Calibration
`
`If the instrument is in use immediately before calibration is
`attempted, then the system should be purged with clean air. This
`may be done by disconnecting the water trap from the instrument
`and running the instrument in measurement mode for 5 minutes.
`
`To carry out sensor calibration press 2 and the following screen will
`be displayed:
`
`Disconnect instrument
` from gas sources.
`
`Press ENTER to start
`Press CANCEL to abort
`
`The PrinterNOx requires clean air free from any NO or NO2 so that
`the sensors can be zeroed.
`
`
`
`17
`
`Ex. 2027-0019
`
`

`
`
`
`With the water trap disconnected, press enter and the display will
`change to:
`
`
`
` Zeroing
`
`
`
` *..............
`
`Press CANCEL to abort
`
`The sampling pump will start and the PrinterNOx will monitor the
`output of both sensors until a stable reading is obtained.
`
`During this time, a line of asterisks appears on the screen to indicate
`correct operation.
`
`When zeroing is complete the readings for the NO and NO2 sensors
`are written into the battery backed memory and the display is
`changed to:
`
` Calibration
` -----------
` 1 Calibrate NO
` 2 Calibrate NO2
` 3 Exit
`
`
`
`18
`
`
`
`Ex. 2027-0020
`
`

`
`
`
`Press “1” to calibrate NO and the display will change to:
`
`Enter NO calibration
`gas concentration:
`
` 00.0 ppm
`
`
`
`
`Enter the required gas concentration using the DELETE key to
`correct any mistakes and press ENTER.
`
`The display will change to:
`
`Pass calibration gas
` through instrument
`
`Press ENTER to start
`Press CANCEL to abort
`
`
`
`CareFusion can supply calibration gas (25ppm NO in nitrogen and
`10ppm NO2 in air) complete with a control valve for simple and
`economic user calibration.
`
`The gas is supplied in convenient, disposable, aluminium canisters
`containing 58 litres of gas (Cat No 36-GAS100 for 25ppm NO in
`nitrogen and 36-GAS200 for 10ppm NO2 in air). A flow regulator
`pre-set to deliver 500ml/min is available separately, Cat No 36-
`GAS500. With careful use, approximately 45 calibrations can be
`obtained from each canister.
`
`19
`
`Ex. 2027-0021
`
`

`
`
`
`Screw the control valve on to the canister and connect to the
`PrinterNOx with the water trap connected as shown below:
`
`Pre-set
`
`Flow
`
`Regulator
`(500ml/min)
`
`Pressure
`Guage
`0
`psi 1000
`USE NO OIL
`
`One Way
`Relief Valve
`
`Control
`Valve
`
`25 ppm
`Nitric Oxide
`in Nitrogen
`
` OR
` 10 ppm
`Nitrogen Dioxide
`in air
`
`PrinterNOx
`
`Medical
`
`Micro
`
`1
`6
`PAPER
`
`2
`7
`CANCEL
`
`3
`4
`8
`9
`CHARGIN G
`
`5
`0
`ENTER
`
`Alarm
`
`Slowly turn the control knob fully anti-clockwise. This will then
`supply a flow of approximately 500ml/min. As the PrinterNOx
`samples at a rate of 250ml/min then 250ml/min will be vented
`through the one way valve.
`
`
`
`20
`
`
`
`Ex. 2027-0022
`
`

`
`
`
`Press ENTER on the keypad and the display will change to:
`
` Calibrating NO
`
` *.............
`
`Press CANCEL to abort
`
`
`
`The sampling pump will start and the PrinterNOx will monitor the
`output of the NO sensor until a stable reading is obtained. When
`stability has been achieved the calibration factor, time and date are
`written into the battery backed memory and the display returns to
`the previous screen. Remove the control valve by turning anti-
`clockwise to stop the gas flow. The NO2 sensor can now be
`calibrated using the same procedure if required.
`
`Within the lifetime of the sensors, zero drift of more than 10ppm
`(1ppm for NO2) or a calibration drift of more than 40% implies
`abnormal ageing of the electrochemical cell. If these errors are
`detected during the zeroing procedure or the calibration adjustment
`then the following will be displayed:
`
`------- ERROR -------
`Reading out of range:
`Check gas supply and
` try again.
` Refer to manual if
` error recurs.
`
` Press any key
`
`
`
`
`
`21
`
`Ex. 2027-0023
`
`

`
`
`
`The most likely reason for this message is an operational error in the
`calibration procedure rather than a faulty sensor.
`
`If this occurs during zeroing repeat the procedure and check that
`there is no nitric oxide, or nitrogen dioxide contamination, and that
`the water trap has been removed.
`
`If the calibration error message appears during sensitivity
`adjustment, repeat the calibration procedure and check:
`
`1. Is calibration gas within its warranty period?
`
`2. Has correct gas concentration been entered?
`
`3. Have the correct connections been made with no leaks?
`
`4. Is there sufficient calibration gas flow?
`
`5. Is there any reaction between nitric oxide and the
`connectors used?
`
`6. Are there any contaminants between nitric oxide supply and
`sensor?
`
`Nitric oxide is a very reactive gas and it is difficult to maintain a
`stable concentration within a cylinder. It is therefore advisable to
`have some independent means of checking the calibration gas used.
`
`After calibration is complete, selecting option “3 Exit” will return the
`display to the main menu screen.
`
`22
`
`
`
`Ex. 2027-0024
`
`

`
`
`
`3. Set Alarms
`
`Selecting option three will allow the NO and NO2 alarm levels to be
`set individually from the following menu:
`
` Set Alarms
`
` 1 Upper NO : 20
` 2 Lower NO : 5
` 3 Upper NO2 : 4
` 4 Exit
`
`The current alarm levels are displayed for reference and, for
`example, the upper alarm level for NO may be altered by selecting
`option 1 when the following screen will be displayed:
`
`
`
` Enter new alarm
` level :
`
` Upper NO :
`
`
`
`Type in the required alarm level and press ENTER.
`
`The lower NO and NO2 alarm levels are set with the same procedure
`with option 2 and 3 respectively.
`
`The valid range of entry is 0 to 99 for the NO upper and lower
`alarms and the upper alarm must be set higher than the lower
`alarm.
`
`23
`
`Ex. 2027-0025
`
`

`
`
`
`For NO2, the valid range of entry is zero to 19
`
`When the required alarms have been set, then selecting option 4 will
`return to the main menu.
`
`4. Report to printer (RS232)
`
`The last 24 hours of measurements are stored in battery backed
`memory and may be either printed or directed to the RS232 port at
`the rear of the instrument. The choice of recovery method depends
`upon whether the printer or RS232 port is currently active (see
`Setup).
`
`If the RS232 port is active then the data, in ASCII format, may be
`uploaded to a PC running standard communication software. A
`serial cable, Cat No 36-CAB1108, is available from CareFusion.
`
`Note: Please ensure that all PC’s and peripheral equipment comply
`to EN60950 ‘Safety of Information Technology Equipment’ and keep
`them out of reach of the patient when connected to the PrinterNOx.
`
`5. Status
`
`This option allows the current configuration and sensor information
`to be displayed.
`
`24
`
`
`
`Ex. 2027-0026
`
`

`
`
`6. Setup
`
`Selecting this menu item allows various options to be set and
`displays the screen:
`
` SET UP
` 1 Clear memory
` 2 Set date/time
` 3 Toggle output:
` *printer RS232
` 4 Toggle date format
` *DD/MM MM/DD
` 5 Exit
`
`
`
`The first option clears the stored 24 hours data from the memory.
`This is done when the currently stored data is of no interest or has
`previously been accessed so that future access of required data will
`be quicker.
`
`The second option is used to change the date and time stored in the
`real time clock.
`
`“Toggle output” is used to select whether data is sent to the printer
`or the RS232 port at the rear of the instrument. An asterisk appears
`against the selected output device. The initial few lines of
`calibration and alarm information is always printed regardless
`whether the printer is selected.
`
`The date format can be changed using option 4.
`
`Option 5 returns the display to the main menu screen.
`
`25
`
`
`
`Ex. 2027-0027
`
`

`
`
`7. Calculate Flow
`
`This facility is used to calculate the required NO flow to achieve the
`required therapeutic concentration for ventilators which employ a
`constant bias flow arrangement. When selected the following screen
`is displayed:
`
` Enter ventilator
` bias flow:
`
` L/min
`
`The valid range for the bias flow is 1-99 and once entered the
`display changes to:
`
`
`
` Enter NO gas
` concentration in
` bottle:
`
` ppm
`
`Gas concentrations of between 50 and 1000ppm may be entered.
`
`
`
`26
`
`
`
`Ex. 2027-0028
`
`

`
`
`
`Following this, the required NO concentration in the range 1 to 100
`should be entered:
`
` Enter required NO
` gas concentration
` to patient:
`
` ppm
`
`When all the data is entered the required NO flow and the maximum
`FiO2 is calculated and displayed on the following screen:
`
`
`
`Bias flow 70 L/m
`Bottle conc 650 ppm
`Patient conc 20 ppm
`
`Req NO flow 2.22 L/m
`Maximum FiO2 96.9 %
`
`Press key to continue
`
`
`
`Battery operation
`
`During patient transportation or mains power failure, the PrinterNOx
`may be powered from the internal rechargeable batteries for
`approximately 2 hours.
`
`27
`
`
`
`Ex. 2027-0029
`
`

`
`
`
`When the batteries are low the alarm will sound continuously and
`the following screen will be displayed:
`
` Battery low!
`
`Please connect charger
`and wait.
`
`
`
`
`Paper Loading
`
`The PrinterNOx requires thermal paper (Part Number 36-PSA1600 -
`Pack of 5 rolls) for the printer.
`
`A warning line appears when 30cm of the paper roll remains. A new
`roll should be loaded when this mark becomes visible.
`
`To load paper:
`
`a. Switch on the PrinterNOx.
`b. Cut the leading edge of the paper so that it is parallel to the
`roll shaft.
`c. Remove the paper cover.
`d. Align the paper so that it is straight when it is inserted into
`the paper inlet.
`e. Press the “PAPER” key to feed the paper until the leading
`edge of the paper passes the paper cutter.
`f. Replace paper cover.
`
`28
`
`
`
`Ex. 2027-0030
`
`

`
`
`
`Make sure that the specified paper (Part Number 36-PSA1600) is
`used as excessive wear on the thermal print head may occur if paper
`from other manufacturers is used. The following precautions should
`be observed when storing or handling the thermal paper rolls:
`
`• Do allow the printer to run without paper loaded.
`• Do not store paper rolls in places exposed to direct sunlight.
`• Do not use pastes containing organic solvents such as
`alcohol, esters or ketones.
`• Do not allow the paper t

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket