throbber
(Delivery System)
`
`Operation Manual
`(800 ppm INOMAX® (nitric oxide) for inhalation)
`
`Part No. 20003 Rev - 01
`
`INO THERAPEUTICS EXHIBIT 2002
`Praxair Distribution v. INO Therapeutics
`IPR2015-00891
`
`

`
`(Delivery System)
`
`Operation Manual
`(800 ppm INOMAX® (nitric oxide) for inhalation)
`
`Part No. 20003 Rev - 01
`
`Part No. 20010 Rev-05
`2012-03
`
`

`
`User Responsibility
`This Product will perform in conformity with the
`description contained in this operating manual
`and accompanying labels and/or inserts, when
`assembled, operated, maintained and repaired
`in accordance with the instructions provided. This
`Product must be checked prior to use following the
`Pre-Use Checkout procedure described in section
`3, and periodically as required. A defective Product
`should not be used. Parts that are broken, missing,
`visably worn, distorted or contaminated should be
`replaced immediately.
`
`replacement become
`repair or
`Should such
`necessary, INO Therapeutics LLC d/b/a Ikaria®
`recommends that a telephone request for service
`advice be made to the nearest Field Technical
`Service Support Center. This Product or any of
`its parts should not be repaired other than in
`accordance with written instructions provided by
`Ikaria. The Product must not be altered without the
`
`prior written approval of Ikaria Quality Assurance
`Department. The user of this Product shall have the
`sole responsibility for any malfunction which results
`from improper use, faulty maintenance, improper
`repair, damage, or alteration by anyone other than
`Ikaria.
`
`Caution: U. S. Federal and Canadian law restrict
`this device to sale by or on the order of a licensed
`medical practitioner. Outside the U. S. A. and
`Canada, check local laws for any restrictions
`that may apply.
`
`Inhaled Nitric Oxide mixtures must be handled
`and stored in compliance with federal, state and
`local regulations.
`
`Ikaria products have unit serial numbers with coded
`logic which indicates the year of manufacture and a
`sequential unit number for identification.
`
`
`
`
`
`20051234
`
`10007
`
`The first four numeric digits indicate the year of product manufacture, and
`the next 4 digits are the sequential unit number produced.
`
`INOmax DSIR® part number
`
`Open Source Software
`A CD-ROM is available upon request containing the full source code to the open source software used within this product.
`Portions of this software are copyright © 1996-2002 The FreeType Project (www.freetype.org). All rights reserved.
`
`Korean fonts Baekmuk Batang, Baekmuk Dotum, Baekmuk Gulim, and Baekmuk Headline are registered trademarks owned by Kim Jeong-Hwan.
`
`©2012 Ikaria, Inc.
`INOMAX®, INOmax DSIR
`®, INOblender®, INOcal® and INOvent® are registered trademarks of INO Therapeutics, LLC.
`
`No license is conveyed, either expressed or implied, with the purchase hereof under U.S. Patent 5,485,827 and U.S. Patent 5,427,729 and their
`foreign equivalents.
`
`U.S. Patent 5,558,083 and foreign equivalents.
`
`Part No. 20010 Rev-05
`2012-03
`
`

`
`Contents
`1. General Information ..............................................................................................................1
`
`
`Introduction to this Manual................................................................................................2
`
` Theory of Operation ........................................................................................................14
`
` Environmental Effects .....................................................................................................18
`
`2. Setup
`
` ..............................................................................................................................19
`
`3. Pre-Use Checkout...............................................................................................................21
`
`4. Patient Application ..............................................................................................................29
`
`
`INOblender® Operation ...................................................................................................31
`
` Backup NO Delivery .......................................................................................................32
`
` Transport Regulator/Cap Assembly Application .............................................................33
`
` Changing INOMAX® Cylinders and Purging the Regulatory Assembly ..........................37
`
` Oxygen Dilution Chart.....................................................................................................39
`
` Duration Chart INOMAX Cylinder 88-Size ......................................................................40
`
` Duration Chart INOMAX Cylinder D-Size .......................................................................41
`
` Monitoring the Environment ............................................................................................45
`
` Entering Patient Information ...........................................................................................46
`
` Connection to Various Breathing Systems .....................................................................48
`
`
` Bagging Systems While Using the Injector Module ...................................................49
`
`
` Bunnell Life Pulse High Frequency Ventilator Circuit.................................................52
`
`
`
` Connecting INOmax DSIR® Sample Tee to the Bunnell Life Pulse Circuit ............53
`
`
`
` Connecting INOmax DSIR Injector Module to the Bunnell Life Pulse Circuit ........53
`
`
` Circle Anesthesia System ..........................................................................................54
`
`
` Fisher/Paykel Healthcare Bubble CPAP ....................................................................56
`
`
` Hamilton Arabella Nasal CPAP ..................................................................................57
`
`
`
`ICU Ventilator Circuit..................................................................................................58
`
`
` Sensormedics 3100A/B High Frequency Oscillatory Ventilator
`
`
`
` with a Filtered Circuit ............................................................................................59
`
`
` Sensormedics 3100A/B High Frequency Oscillatory Ventilator
`
`
`
` with a Rigid or Flexible Circuit ..............................................................................60
`
`
` Spontaneous Breathing Patient on a Mask Circuit ....................................................61
`
`
` Spontaneous Breathing Patient on a Nasal Cannula ................................................62
`
`
` Teleflex Medical Comfort Flo Humidification System .................................................63
`
`
` Transport Ventilator Circuit ........................................................................................64
`
`
` Vapotherm 2000i ........................................................................................................66
`
`
` Vapotherm Precision Flow .........................................................................................67
`
`
` Viasys Infant Flow CPAP System; Cardinal Airlife nCPAP System .............................68
`
`
` Viasys Infant Flow SiPAP ...........................................................................................70
`
`
`INOmax DSIR Patient Circuit Disposables ......................................................................72
`
`Part No. 20010 Rev-05
`2012-03
`
`

`
`5. Alarms ..............................................................................................................................73
`
`6. Troubleshooting ..................................................................................................................77
`
`7. Calibration ...........................................................................................................................85
`
` Low Range Calibration ...................................................................................................86
`
` Oxygen Sensor High Range Calibration .........................................................................88
`
` NO Sensor High Range Calibration ................................................................................90
`
` NO2 Sensor High Range Calibration ..............................................................................93
`
`8. Maintenance .......................................................................................................................97
`
` Cleaning..........................................................................................................................98
`
` Replacing the O2, NO, and NO2 Sensors .....................................................................101
`
` Replacing the H2O Separator Cartridge & INOMAX® Reg. Tip .....................................103
`
` Cylinder Leak Check.....................................................................................................104
`
`
`INOblender® Used as a Stand-Alone Device ................................................................105
`
` Preventative Maintenance ............................................................................................106
`
` Parts and Accessories ..................................................................................................107
`
`9. Product Specifications .....................................................................................................109
`
`Part No. 20010 Rev-05
`2012-03
`
`

`
`Information
`
`General
`
`(Delivery System)
`
`1/ General Information
`
`Part No. 20003 Rev - 01
`
`Part No. 20010 Rev-05
`2012-03
`
`

`
`(Delivery System)
`
`Information
`
`General
`
`1/ General Information
`
`Part No. 20010 Rev-05
`2012-03
`
`

`
`1/ General Information
`• The INOmax DSIR® (delivery system) delivers
`INOMAX® (nitric oxide for inhalation) therapy gas into
`the inspiratory limb of the patient breathing circuit
`in a way that provides a constant concentration of
`nitric oxide (NO), as set by the user, to the patient
`throughout the inspired breath. It uses a specially
`designed injector module, which enables tracking
`of the ventilator waveforms and the delivery of a
`synchronized and proportional dose of NO. It may
`be used with most ventilators.
`
`• The INOmax DSIR provides continuous integrated
`monitoring of inspired O2, NO2 , and NO and a
`comprehensive alarm system.
`
`• The INOmax DSIR incorporates a battery that provides
`up to 6 hours of uninterrupted INOMAX delivery in the
`absence of an external power source.
`
`• The INOmax DSIR includes a backup NO delivery
`capability that provides a fixed flow of 250 mL/min of
`NO which along with user supplied 10 L/min of oxygen
`provides 20 ppm in the gas flow to a patients breathing
`circuit. It may also use the INOblender for backup.
`
`• The target population is controlled by the drug labeling
`for INOMAX and is currently neonates. The primary
`targeted clinical setting is the Neonatal Intensive Care
`Unit (NICU) and secondary targeted clinical setting is
`the transport of neonates.
`
`Important:
`
`Before using the INOmax DSIR, read through this manual.
`
`Read through the manuals for the ventilator, humidifier
`and any other accessory items used. Follow the manual
`instructions and obey the Warnings and Cautions.
`
`Keep this manual readily available to answer questions.
`
`Read the User Responsibility statement on the inside front
`cover of this manual; it describes what the user must do to
`maintain proper use and functioning of this product.
`
` WARNING:
`Warnings tell you about dangerous conditions
`that can cause injury to the operator or the
`patient if you do not obey all of the instructions
`in this manual.
` Caution:
`Cautions tell you about how to properly use the
`equipment and conditions that could cause damage
`to the equipment.
`Read and obey all warnings and cautions.
` Note:
`Notes provide clarification or supplemental information.
`
`Blue arrow denotes required user action.
`
` WARNING:
`• If an alarm occurs, safeguard the patient
`first before troubleshooting or repair procedures.
`• Use only pharmaceutical grade NO/N2.
`• The INOmax DSIR must only be used in
`accordance with the indications, usage,
`contraindications, warnings and precautions
`described in the INOMAX (nitric oxide) drug
`package inserts and labeling. Refer to this
`material prior to use.
`• The use of devices which radiate high intensity
`electrical fields may affect the operation of
`the INOmax DSIR. Constant surveillance of
`all monitoring and life support equipment is
`mandatory whenever interfering devices are
`in operation on or near a patient.
`• INO Therapeutics does not recommend that
`the INOmax DSIR be utilized with helium/
`oxygen mixtures in any situation. The
`INOmax DSIR is intended to deliver INOMAX
`therapy gas only in conjunction with the
`delivery of air and oxygen.
`• Do not connect items, which are not specified
`as part of the system.
`
`1
`
`Part No. 20010 Rev-05
`2012-03
`
`

`
`Introduction to this Manual
`Definitions and abbreviations
`% v/v
`% volume/volume
`Control wheel
`Rotary control used to change and confirm settings
`Display
`The electronic information panel on the front of the delivery system
`HFOV
`High frequency oscillating ventilator
`Menu
`A list of available choices for an operation
`N2
`Nitrogen
`NO
`INOMAX® (nitric oxide) for inhalation
`NO2
`Nitrogen dioxide
`NO/N2
`Nitric oxide (NO) and nitrogen (N2) gas mixture
`O2
`Oxygen
`ppm
`Parts Per Million
`Resolved alarm
`An alarm condition that has been corrected
`Set NO
`The dose of INOMAX set by the user
`Touch screen
`A display screen sensitive to touch used to select a function
`
`1
`
`2
`
`3
`
`4
`
`11
`
`12
`
`13
`
`14
`
`15
`
`16
`
`9
`
`10
`
`22
`
`8
`
`7
`
`6
`
`5
`
`21
`
`20
`
`19
`
`18
`
`17
`
`1. Sample Line Inlet
`2. Main Power Indicator
`3. Display Screen
`4. Alarm Speaker (under front label)
`5. Backup Switch
`6. Control Wheel
`7.
`Injector Module Tubing Outlet
`
`Figure 1-1 INOmax DSIR® Front View
`
`Injector Module Cable Inlet
`8.
`9. Water Trap Bottle
`10. Purge Port
`11. INOMAX Gas Inlets
`12. Blender Gas Outlet
`13. Ethernet Port
`14. Infrared Connector
`
`15. USB Port
`16. Water Separator Cartridge
`17. Water Trap Bottle
`18. Sample Gas Outlet Port
`19. Power Cord Inlet
`20. Ground
`21. ON/Standby Switch
`22. RS232 Port
`Figure 1-2 INOmax DSIR Rear View
`
`Part No. 20010 Rev-05
`2012-03
`
`2
`
`

`
`1
`
`2
`
`5
`
`7
`
`8
`
`3
`
`4
`
`6
`
`9
`
`® Mounting Post
`INOmax DSIR
`1.
`INOMAX® Regulator (2)
`2.
`3. Small Part Bin
`4.
`INOMAX Cylinder
`5. Cylinder Holding Bracket
`6. Cylinder Mounting Strap
`7. Oxygen Cylinder Bracket
`8. Caster Lock Lever
`9. Caster (4)
`
`Figure 1-3 INOmax DSIR and Cart
`
`3
`
`Part No. 20010 Rev-05
`2012-03
`
`

`
`Navigating the Display Screens
`There are five screens that can be displayed on the INOmax DSIR®.
`
`Note:
`
`
`The specific level is identified by the highlighted card on the Menu Button.
`The red arrows indicate going back to a previous screen.
`
`Main Screen (first level)
`
`Alarm History Screen
`(second level)
`
`Patient Information Screen
`
`Menu Screen
`(second level)
`
`Calibration Screen
`(third level)
`
`Part No. 20010 Rev-05
`2012-03
`
`4
`
`

`
`1
`
`2
`
`3
`
`4
`
`5
`
`19
`
`18
`
`17
`
`16
`
`15
`
`14
`
`13
`
`12
`
`11
`
`10
`
`9
`
`1. Alarm Silence Button
`2. Upper Alarm Limit Button
`3. Lower Alarm Limit Button
`4. Monitored Value
`5. Menu Button
`6. Text Message Area
`
`7. Monitor Area
`8. Graphical Area
`9. Patient Information Button
`10. Sample Line Icon
`11. Water Trap Icon
`12. Inspiratory Limb Icon
`
`Figure 1-4 Main Display Screen
`
`1
`
`2
`
`3
`
`4
`
`6
`
`7
`
`8
`
`5
`
`6
`
`8
`
`7
`9
`10
`1. Software Revision Field
`6. Calibration Due Dates
`2. Display Brightness Button
`7. NO Calibration Button
`3. Alarm Volume Button
`8. NO2 Calibration Button
`4. Return to Previous Level Button
`9. Oxygen Calibration Button
`5. Monitor Area
`10. Room Air Calibration Button
`Figure 1-5 Menu Screen (second level)
`
`Main Display Screen
`• On the Main Screen
`
`the user can view alarm
` messages, monitored
` values and graphical
`
`information.
`
`• By pressing the “Menu
` Button” on the touch
` screen (top right hand
` corner), the user can
` access the Menu Screen
`
`(see Figure 1-5).
`
`13. Injector Module Icon
`14. Delivery Line Icon
`15. Backup Line Icon
`16. Backup Switch Icon
`17. Delivery Setpoint Display
`18. NO Delivery Setpoint Button
`19. Cylinder Icon
`
`Menu Screen (second level)
`• On the Menu Screen the user
` can change the alarm volume
` and the display brightness
`
` and view the software version.
`
`• To review the Alarm History
` Screen (second level),
`
`refer to Section 5/ Alarms.
`
`
`• It also lets the user select
` different calibration options
`
`(third level) which will be
` covered in Section
` 7/ Calibration.
`
`5
`
`Part No. 20010 Rev-05
`2012-03
`
`

`
`Display and user controls
`
`The INOmax DSIR® has a color touch screen display and a control wheel for adjusting and entering user
`settings. The buttons on the touch screen and the control wheel perform a variety of functions using a three-
`step procedure (see “Setting and making changes on the INOmax DSIR”).
`
`Note: • If a button has been selected and no activity has been sensed within 20 seconds, the display will
` return to its previous condition. If a button is de-emphasized (grayed out), it is not accessible.
`• Position delivery system so user screen is unobstructed and the speaker is not covered.
`
`
`
`When a value is being changed, pressing the
`"Cancel Active Status" button during editing will
`stop the change and return the parameter to its
`original value (similar to the escape key on a
`computer).
`
`Dose settings
`
`Displayed dose settings are 1, 5, 10, 20, 40,
`60 and 80 ppm. Each click on the control knob
`corresponds to a known change in dose. The
`incremental dose per click corresponds to a value
`dependent upon the dose range in which the
`change is made, as illustrated in the table at right.
`
`Dose Setting Range Dose Change Per Click
`< 1 ppm
`0.1 ppm
`1 to 40 ppm
`1 ppm
`40 to 80 ppm
`2 ppm
`
`Part No. 20010 Rev-05
`2012-03
`
`6
`
`

`
`Setting and making changes on the INOmax DSIR®
`The touch screen buttons and control wheel are used to:
` • Set the concentration of delivered NO
` • Adjust alarm limits
` • Silence alarms
` • Calibrate the sensors
` • Review alarm history
` • Define setup options
` • Enter patient identifier
`
`1. SELECT
`
`(press) a button on the touch screen
` associated with the desired function.
`
`(an audible beep will sound when a button
`
`is selected, and the button will be displayed in
`
`inverse video)
`
`2. ROTATE
`
`the control wheel clockwise or
` counterclockwise to adjust the value.
`
`3. CONFIRM
`
`the selection by pressing the control wheel
` or the button associated with the desired
`
`function again.
`
`Note: After confirming a desired dose, the
`
`NO alarm setting (high and low) will
`
`automatically be set for the first setting
`
`only. Any other changes will require
`
`the high and low alarm settings to be
`
`adjusted. Also, a two minute lockout
`
`period will prevent monitoring
`
`alarms from occurring while measured
`
`values stabilize.
`
`7
`
`Part No. 20010 Rev-05
`2012-03
`
`

`
`Settings
`
`Access the menu screen (second menu level).
`
`Alarm Volume setting
`1. Push the alarm volume
` button on the touch screen.
`2. Turn the control wheel to indicate the volume
`
`level you want. Choices range from 1 (softest)
`
`to 5 (loudest).
`3. Push the control wheel to confirm your selection.
`4. If you are finished with the
` Menu Screen push the return
`
`to previous level button on the
`
`touch screen.
`
`Display Brightness setting
`1. Push the display brightness
` button on the touch screen.
`2. Turn the control wheel to indicate the display
` brightness level you want. Choices range from
` 1 (darkest) to 10 (brightest).
`3. Push the control wheel to confirm your selection.
`4. If you are finished with the
` Menu Screen, push the return
`
`to previous level button on the
`
`touch screen.
`
`Part No. 20010 Rev-05
`2012-03
`
`8
`
`

`
`Infrared Communication between the INOMAX® Cylinders
`and the INOmax DSIR®
`The INOmax DSIR has an interface using infrared (IR) technology which will allow the
`INOmax DSIR to communicate with the INOmeter® (which is mounted to each INOMAX cylinder).
`The INOmax DSIR checks the INOMAX cylinder for the correct expiration date and cylinder
`concentration. The INOmax DSIR also transmits a confirmed patient identifier to the INOmeter on
`any open INOMAX cylinder.
`
`The INOmax DSIR cart (PN 10018) has a cover ( ) with an infrared transceiver mounted
`1
`directly above each INOMAX cylinder. When INOMAX cylinders are loaded, communication
`will take place between the INOmax DSIR and the INOmeter ( ) (after the boot up phase of
`2
`the INOmax DSIR is complete). A cylinder icon will be displayed on the main screen when an
`INOMAX cylinder is recognized by the INOmax DSIR (see “Loading INOMAX Cylinders onto the
`INOmax DSIR Cart”, page 11).
`IR Communication Interference
`The INOmax DSIR transceiver is located under the cart cover and should be protected from
`outside IR sources. The INOmax DSIR cart was designed to protect the INOmeter from external
`light/IR energy sources. The INOmax DSIR transceiver transmits via a 30 degree transmission
`cone projecting towards the floor (see dotted lines in Figure 1-6). The specifications of the IR
`beam call for it to have a range of 20 cm (7.9 in), so based on these specifications it would not
`affect other devices in the vicinity of the INOmax DSIR.
`
`The INOmeter uses a lower energy source which results in a lower IR beam range than
`the INOmax DSIR cart. The INOmeter does not transmit IR signals unless it mounted on the
`INOmax DSIR cart.
`
`If there is interference with the INOmax DSIR/INOmeter communication, the cylinder icon on the
`user screen will not be displayed and a “Cylinder Not Detected” alarm will activate.
`
`If IR communication interference occurs, we recommend you take the following actions:
`• Move the external IR source
`• Move the INOmax DSIR cart to reduce the external IR source in the area of the INOmeter
`• Shield the INOmeter from the suspect IR source
`
`If the actions listed above do not remedy this issue, the Transport Regulator/Cap Assembly could
`be utilized.
`
`9
`
`Part No. 20010 Rev-05
`2012-03
`
`

`
`External Light Interference
`High frequency and/or high intensity light emission, in the area of the INOmeter, may interfere with
`communication between the INOmax DSIR® and the INOmeter on the INOMAX cylinder. If there is
`interference with the INOmax DSIR/INOmeter communication, the cylinder icon on the user screen will not
`be displayed and a “Cylinder Not Detected” alarm will activate.
`
`Test results have demonstrated susceptibility to unintended infrared energy from artificial light sources. Most
`notably, various compact fluorescent lighting fixtures that focus or reflect light, increasing the light intensity in
`the vicinity of the INOmax DSIR cart, could affect INOmeter communications.
`
`If external light interference occurs, we recommend you take the following actions:
`• Move the interfering light source
`• Move the INOmax DSIR cart to reduce the high intensity light in the area of the INOmeter
`• Shield the INOmeter from the suspect light source
`
`If the actions listed above do not remedy this issue, the Transport Regulator/Cap Assembly could be utilized.
`
`1
`
`2
`
`Figure 1-6
`
`Part No. 20010 Rev-05
`2012-03
`
`10
`
`

`
`Loading INOMAX® Cylinders onto the INOmax DSIR® Cart
`
`Note:
`
`
`Check the INOMAX gas cylinders for the correct product identity labels, cylinder concentration
`and expiration date.
`
`Ensure at least one INOMAX gas cylinder (with more than 200 psig) is available.
`Refer to Section 3 / Pre-Use Checkout prior to use.
`
`2
`
`4
`
`1
`
`3
`
`Loading a second
`INOMAX cylinder onto
`the cart will result
`in a second cylinder
`icon displayed on the
`screen .
`
`3 4
`
`Loading the first
`INOMAX cylinder
`on the cart will
`result in a cylinder
`icon displayed on
`the screen .
`
`1 2
`
`11
`
`Part No. 20010 Rev-05
`2012-03
`
`

`
`(Intentionally left blank)
`
`Part No. 20010 Rev-05
`2012-03
`
`12
`
`

`
`Symbols used in this manual or on the system
`
`Symbols replace words on the equipment and/or in
`this manual. These symbols include:
`
`NO Gas Inlet
`
`NO Gas Outlet
`
`On
`
`Purge Location
`
`Sample Gas Inlet Port
`
`Sample Gas Outlet Port
`
`
`
`
`
`
`
`
`
`
`
`
`
`Attention, consult accompanying
`documents!
`
`Alarm Silence
`
`Equipotential Stud
`
`Ethernet Port
`
`Fuse Rating
`
`Infrared Input/Output
`
`Injector Module
`
` SN
`
`Serial Number
`
`Low Range Calibration
`
`
`
`Standby
`
`Main Power Connected
`
` REF
`
`Stock Number
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Low Cal
`
`
`
`Type B Electrical Equipment
`
`USB Port
`
`
`
`
`
`
`
`NO Backup OFF
`
`
`NO Backup ON
`
`
`
`
`
`
`
`
`
`Water Separator Cartridge
`
`13
`
`Part No. 20010 Rev-05
`2012-03
`
`

`
`Theory of Operation
`The INOmax DSIR® provides a constant dose of
`INOMAX® into the inspiratory limb of the ventilator
`circuit. The INOmax DSIR uses a “dual-channel” design
`to ensure the safe delivery of INOMAX. The first
`channel has the delivery CPU, the flow controller and
`the injector module to ensure the accurate delivery
`of NO. The second channel is the monitoring system,
`which includes a separate monitor CPU, the gas
`sensors (NO, NO2, and O2 sensors) and the user
`interface, including the display and alarms. The dual-
`channel approach to delivery and monitoring permits
`INOMAX delivery independent of monitoring. This
`allows the monitoring system to shutdown INOMAX
`delivery, if it detects a fault in the delivery system. For
`example, INOMAX delivery will shut down should the
`monitored NO concentration become greater than 100
`ppm. (See Figure 1-7 for a schematic diagram).
`
`1. INOMAX drug is stored as a gas mixture of NO/N2
`in an aluminum cylinder at a nominal pressure of
`2000 psig.
`
`2. The cylinder is attached to a high pressure
`regulator, which incorporates a pressure gauge
`that indicates cylinder pressure when the cylinder
`valve is open. The cylinder regulator is attached via
`tubing to the INOmax DSIR using one of the two NO/
`N2 quick connect inlets on the back of the machine.
`3. The INOmax DSIR checks the INOMAX cylinder
`for the correct expiration date and cylinder
`concentration.
`
`4. The INOMAX enters the back of the INOmax DSIR,
`passes through a filter, then a safety shutoff valve,
`which is open under normal operation.
`
`5. An injector module is placed in the ventilator
`gas flow between the ventilator inspiratory outlet
`and the humidifier. Based on the ventilator flow, the
`INOMAX cylinder concentration and set INOMAX
`dose, the proportional solenoid valve delivers
`800 ppm INOMAX into the ventilator circuit via the
`injector module where it mixes with the breathing
`circuit gas flow to achieve the set dose. This allows
`the INOmax DSIR to deliver a constant dose of
`
`Part No. 20010 Rev-05
`2012-03
`
`14
`
`INOMAX regardless of the ventilator flow pattern
`or breath rate (see Figure 1-8).
`
`6. A flow sensor inside the INOmax DSIR also
`monitors the NO flow out of the machine. A
`check valve is included prior to the INOmax DSIR
`drug outlet to prevent pressure effects from the
`ventilator breathing circuit interfering with the
`NO flow sensor reading.
`
`7. Gas Monitoring - The INOmax DSIR gas
`monitoring system provides monitored values
`for inspired NO, NO2, and O2. The sample gas
`is withdrawn from the breathing circuit and goes
`through a water trap, a zero valve, a sample
`pump and finally a sample flow sensor to the
`gas monitoring sensors.
`
`• The zero valve allows the gas sensors to
`be zeroed (during low calibration) without
`having to disconnect the sample line from the
`breathing circuit.
`
`• The pump and sample flow sensor ensure
`a sample gas flow rate is maintained to the
`monitoring sensors.
`
`• The gas monitoring sensors are
`electrochemical; they are specific to each
`gas and provide an electronic signal which is
`proportional to the concentration of the gas
`present.
`
`8. Backup Delivery - If the delivery system does
`go into shutdown, the INOmax DSIR has an
`integrated backup function which provides a
`fixed flow of INOMAX (250 mL/min) into the
`injector module using a pneumatic on/off switch
`and a restrictor built into the delivery side of the
`system. This fixed flow of INOMAX will provide
`20 ppm of NO when the continuous ventilator
`gas flow is 10 L/min. The backup is only for
`short term use until a replacement delivery
`system can be obtained. An alarm will warn
`the user if the backup system is turned on while
`the main delivery system is in use for INOMAX
`delivery. The INOblender® can also be used as
`a backup.
`
`

`
`Regulator
`
`Infrared
`

`INOMAX
`Cylinder
`800 ppm
`
`DSIR Link CPU
`
`Monitor
`CPU
`
`Sample Gas Outlet Port
`
`F
`
`F
`
`NO
`Backup
`ON/OFF
`
`Restrictor
`
`Sensors
`
`O
`
`2
`
`NO
`
`NO
`
`2
`
`Sample Flow
`Sensor
`
`Sample Pump
`
`Zero Valve
`
`Water Trap
`
`Ventilator
`
`Sample
`Line
`
`Patient
`
`Expiratory Limb
`
`Breathing Circuit
`
`Sample Tee
`
`Inspiratory Limb
`
`Delivery
`CPU
`Pressure
`Switch
`
`Pressure
`Shut off valve
`
`Proportional
`Valve
`
`INOMAX Flow
`Sensor
`
`Injector
`Tubing
`
`Injector
`Module
`Cable
`
`Figure 1-7 Schematic Diagram of INOmax DSIR®
`
`Delivered Nitric Oxide (NO)
`Concentration using INOmax DSIR
`Delivery System
`
`Ventilator Flow
`
`Injector
`Module
`
`INOMAX
`
`Sample Gas
`
`Electronic
`
`Humidifier
`
`Figure 1-8 INOMAX injection method provides a constant NO concentration
`
`Time
`
`15
`
`Part No. 20010 Rev-05
`2012-03
`
`

`
`Effect of the INOmax DSIR® in a ventilator circuit
`Minute Volume
`There are two main effects of connecting and using
`the INOmax DSIR in a ventilator breathing circuit.
`When using volume ventilation with the INOmax DSIR,
`the measured tidal volume delivered to the patient
`• First, the INOmax DSIR adds NO/N2 gas to the
`shows small changes depending on the NO setting
` breathing circuit in proportion to the NO setting
`being used due to the addition and subtraction of
` and the ventilator flowrate. For example, at an 80
`gases by the delivery system. Some minor ventilator
` ppm NO setting (the maximum NO setting with
`adjustments to the minute volume may be required.
` an 800 ppm NO cylinder concentration) the
`The net result of the INOmax DSIR on the delivered
` INOmax DSIR adds 10% more gas to that delivered
`minute ventilation can be calculated as follows:
` by the ventilator, 5% more for a 40 ppm setting, etc.
`If the patient’s minute ventilation is 10 L/min
`• Second, the INOmax DSIR subtracts gas from the
` (500 cc X 20 breaths/min)
` breathing circuit via the gas sampling system at a
` nominal flow rate of 0.23 L/min.
`The additional minute volume due to the INOMAX
`can be calculated as follows:
`These two effects of adding and subtracting gas
`from the ventilator breathing circuit have the
`following effects:
`Oxygen Dilution
`The INOmax DSIR adds gas to the breathing circuit
`in proportion to the NO setting as described above.
`The NO/N2 mixture added to the ventilator gas
`dilutes the oxygen in proportion to the set INOMAX®
`dose. At an INOMAX dose setting of 80 ppm with a
`cylinder concentration of 800 ppm NO, the added
`gas is 10%. Thus, the O2 concentration is reduced by
`10% of its original value. For example, if the original
`O2 concentration is 50% v/v, then the value after
`injection, at the 80 ppm setting, should be 45% v/v.
`
`INOMAX dose x Minute Volume
`Cylinder Concentration -- INOMAX Dose = Additional INOMAX volume
`added per minute
`For a dose of 20 ppm the additional volume would be
` (20 X 10 / 800 – 20) = 0.25 L/min
`To calculate the net change in minute volume:
` 0.25 L/min INOMAX added - 0.23 L/min removed
`
` (sample system) = 0.02 L/min (net change)
`This formula may be used when calculating the
`changes to continuous flow on continuous flow
`ventilators as well (using the continuous flow in
`place of minute ventilation).
`Trigger Sensitivity
`The addition and subtraction of gases by the
`INOmax DSIR may affect the trigger sensitivity of
`the ventilator when using synchronized modes of
`ventilation. This may cause the ventilator to auto-
`trigger in ventilators which have flow trigger modes,
`especially where the trigger flow is set to less than
`1 L/min. The trigger sensitivity of the ventilator
`should be checked after connecting the
`INOmax DSIR delivery system.
`
`INOMAX Dose (ppm)
`80
`40
`20
`
`Oxygen Dilution % v/v
`10
`5
`2.5
`
`Part No. 20010 Rev-05
`2012-03
`
`16
`
`

`
`Circle Anesthesia Ventilator Systems
`
`The effect of the INOmax DSIR® on circle anesthesia
`ventilator systems (which use volume ventilation)
`is to cause small changes in the delivered minute
`volume as noted previously (see Minute Volume).
`Recirculation of INOMAX® in circle breathing systems
`should be avoided. The gas in the ventilator bellows
`may also contain undesirable levels of NO2 which
`may not be removed by the C

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket