throbber
Trials@uspto.gov
`Tel: 571.272.7822
`
`
`
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`_____________
`
`
`PATENT TRIAL AND APPEAL BOARD
`
`_____________
`
`
`PRAXAIR DISTRIBUTION, INC.
`Petitioner
`
`v.
`
`MALLINCKRODT HOSPITAL PRODUCTS IP LTD.,
`Patent Owner
`
`_____________
`
`
`Case: IPR2015-00889
`Patent: 8,573,209 B2
`
`_____________
`
`PETITIONER’S REPLY TO PATENT OWNER’S RESPONSE
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Mail Stop PATENT BOARD
`Patent Trial and Appeal Board
`United States Patent and Trademark Office
`P.O. Box 1450
`Alexandria, VA 22313-1450
`
`
`
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`
`I.
`II.
`
`III.
`
`TABLE OF CONTENTS
`
`Introduction ...................................................................................................... 1
`The Problem Purportedly Solved by the ‘209 Patent Was Already Well-
`Known Prior To Its Filing ............................................................................... 1
`PO Misinterprets The Instituted Combinations ............................................... 4
`A.
`PO’s Complaints Regarding the Independent Claims ........................... 5
`B.
`PO’s Complaint Regarding Claim 2 ..................................................... 7
`IV. The Instituted Combinations are Proper Because PO’s Arguments About
`The References Lack Merit ............................................................................. 8
`A.
`The ‘083 Patent ..................................................................................... 9
`1.
`PO’s Nitric Oxide-Specific Arguments Are Inaccurate ............. 9
`2.
`The ‘083 Patent Obtains Concentration Data from Different
`Sources ...................................................................................... 11
`The ‘510 Patent ................................................................................... 12
`1.
`The ‘510 Patent Teaches Wireless Transmission ..................... 12
`2.
`The ‘510 Patent Suggests Using Stored Data for Control ........ 13
`3.
`PO Misrepresented the ‘510 Patent in Arguing the
`Open/Close Data is Never Used for Real-Time Tasks ............. 14
`The FR ‘804 Publication ..................................................................... 15
`C.
`V. Mr. Heim’s Testimony Should Be Accorded No Weight ............................. 19
`VI. PO’s FDA Arguments Are Incorrect And Immaterial .................................. 21
`A.
`The FDA 510(k) Process is Merely a Path to Marketability .............. 21
`B.
`If Relevant, the FDA Records Support a Finding of Obviousness ..... 22
`VII. Conclusion ..................................................................................................... 24
`
`
`B.
`
`
`
`
`
`i
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`
`PETITIONER’S UPDATED LIST OF EXHIBITS
`
`Ex. 1001 U.S. Patent No. 8,573,209 (“‘209 Patent”)
`
`Ex. 1002 Declaration of Robert T. Stone, Ph.D
`
`Ex. 1003 Curriculum Vitae of Robert T. Stone, Ph.D
`
`Ex. 1004 U.S. Patent No. 7,114,510 (“‘510 Patent”), filed May 15, 2003, issued
`
`October 3, 2006
`
`Ex. 1005 U.S. Patent No. 5,558,083 (“‘083 Patent”), filed November 22, 1993,
`
`issued September 24, 1996
`
`Ex. 1006
`
`French Publication No. 2 917 804 (“FR ‘804 Publication”), published
`
`December 26, 2008
`
`Ex. 1007
`
`ISO/IEEE 11073-30300, “Health informatics -- Point-of-care medical
`
`device communication -- Part 30300: Transport profile -- Infrared
`
`wireless,” ISO, IEEE, published December 15, 2004 (“IR Standard”)
`
`Ex. 1008 U.S. Patent No. 6,811,533 (“‘533 Patent”), filed January 22, 2001,
`
`issued November 2, 2004
`
`Ex. 1009 Assignment History of the ‘083 Patent
`
`Ex. 1010 Reserved
`
`Ex. 1011 Air Liquide OptiKINOX Brochure, dated 2009
`
`Ex. 1012
`
`“Guidance Document for Premarket Notification Submissions for
`
`ii
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`
`Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen
`
`Dioxide Analyzer,” (“FDA Guidance”) document issued January 24,
`
`2000 by the U.S. Department of Health and Human Services, Food
`
`and Drug Administration
`
`Ex. 1013 Reserved
`
`Ex. 1014 Center for Drug Evaluation and Research, Application Number: NDA
`
`20845, INOMAX®, Final Printed Labeling, (“INOMAX Label”)
`
`available
`
`at
`
`http://www.accessdata.fda.gov/drugsatfda_
`
`docs/nda/99/20845_inomax_prntlbl.pdf (August 9, 2000)
`
`Ex. 1015 Reserved
`
`Ex. 1016
`
`Prosecution History of U.S. Patent No. 8,573,209
`
`Ex. 1017 Reserved
`
`Ex. 1018 Reserved
`
`Ex. 1019 Reserved
`
`Ex. 1020 Declaration of translator Claudine Joly-King under 28 U.S.C. § 1746
`
`regarding Ex. 1006
`
`Ex. 1021 Resume
`
`of Warren P. Heim, P.E.
`
`downloaded
`
`from
`
`http://www.teammedical.us/images/WP%20Heim-
`
`Medical%20Device%20and%20R&D%20Expert.pdf
`
`iii
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`Ex. 1022
`
`Transcript of February 2, 2016 Deposition of Warren P. Heim, P.E.
`
`Ex. 1023
`
`INOvent Delivery System – Operation and Maintenance Manual
`
`(CGA Variant), Dated 02/08/00
`
`Ex. 1024
`
`510(k) Summary for INOmax DS (Delivery System), submitted
`
`December 18, 2009, published April 15, 2010
`
`
`
`iv
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`I.
`
`INTRODUCTION
`The Petition for Inter Partes Review (“Petition”) (Paper 1) explained why
`
`the prior art combinations supporting the instituted Grounds render the ‘209 Patent
`
`claims obvious.1 The Patent Owner Response (“PO Response”) (Paper 30) does
`
`not undermine the prima facie obviousness case presented in the Petition and
`
`adopted in the Board’s Decision to Institute Trial (“Decision”) (Paper 14).
`
`II. THE PROBLEM PURPORTEDLY SOLVED BY THE ‘209 PATENT
`WAS ALREADY WELL-KNOWN PRIOR TO ITS FILING
`
`PO argues that “Petitioner Has Failed to Show that a POSA Was Aware of
`
`the Problem Addressed by the ‘209 Patent Claims.” (PO Resp. at 38-40; see also
`
`id. at 55-56). In this vein, PO’s expert, Mr. Warren Heim, testified that “the risk of
`
`connecting a gas cylinder with the incorrect gas type had already been controlled
`
`by using CGA 626 connections, and thus no requirement remained that would have
`
`led to including the teachings of FR ‘804.” (Ex. 2021 at ¶ 153; see also PO Resp.
`
`at 46-48). It follows, PO argues, that a person of skill would not have been
`
`motivated to provide additional safety mechanisms, such as those in the FR ‘804
`
`Publication, when designing a NO delivery system. (PO Resp. at 48).
`
`PO’s argument is based on selective quotation of the FDA Guidance.
`
`(PO Resp. at 47). The PO Response omitted the following text:
`
`1 All emphasis herein added unless otherwise indicated.
`
`1
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`
`Plans for commercial distribution of nitric oxide in the United States
`include the use of only a single concentration of nitric oxide; the
`availability of only a single concentration renders the use of a
`compressed gas cylinder containing an incorrect concentration of
`compressed nitric oxide in nitrogen unlikely.
`(Ex. 1012 at 8). In the omitted text, the FDA Guidance itself notes that it is
`
`unlikely (not impossible) that the wrong container could be connected. (Id.). It
`
`further notes that even assuming a single concentration was to be used, the use of
`
`the standard gas-specific fitting “will control the risk of incorrect drug
`
`administration,” not eliminate it. (Id.). PO’s expert conceded that the CGA 626
`
`fittings are the same ones used for different concentrations of NO. (Ex. 2021 at
`
`¶137 (“[t]hese connectors are unique to therapeutic nitric oxide mixtures”). Where
`
`multiple concentrations of nitric oxide in nitrogen were available (as PO’s original
`
`labeling for iNO indicated was true by August, 2000, see Ex. 1014 at 6-7), the
`
`FDA Guidance indicates the use of a CGA standardized connector would be
`
`inadequate. (Ex. 1012 at 8). Accordingly, the FDA itself had previously identified
`
`the problem PO avers was unrecognized until the ‘209 Patent.
`
`The FR ‘804 Publication, assigned to a well-known NO supplier in France
`
`(see Petition at 17) similarly identifies the problem PO asserts was “unknown.” It
`
`states “purely mechanical solutions which involve the use of connection
`
`types…and connections, depending on different types of gas” were known but that
`
`2
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`nonetheless, “there are no such foolproofing mechanical systems for a certain
`
`number of compatible gases.” (Ex. 1006 at 17-18).
`
`PO’s INOvent system2 is the system originally cleared by the FDA for use to
`
`deliver NO gas. (Ex. 1014 at 6). The “Operation and Maintenance Manual” for
`
`the INOvent system (“INOvent Manual”) (Ex. 1023) also identifies the very
`
`problem PO and its expert allege was unknown until it was solved by the allegedly
`
`“novel and unique safety check” of the ‘209 Patent. (Ex. 2021 at ¶¶ 53, 54). It
`
`discloses that while the “INOvent delivery system is factory-set for an 800 ppm
`
`cylinder concentration,” “[a]lternate cylinder concentrations may be available by
`
`special request to Datex-Ohmeda.” (Ex. 1023 at 42). Appendix F deals with
`
`“Alternate Cylinder Concentrations.” (Id. at 175-79). Because of the possibility of
`
`connecting NO cylinders with different concentrations, the manual requires that the
`
`operator “[c]heck the therapy gas cylinders for the correct product identity labels
`
`and NO concentrations. The NO concentration must match that shown in the setup
`
`menu for ‘Cylinder Concentration.’” (Id. at 36; see also id. at 42, 106, 156, 157).
`
`The INOvent Manual identifies and warns against this risk despite the system’s use
`
`of the CGA 626 connectors. (Id. at 36, 142, 145, 166-167, 169).
`
`
`2 In discussing existing devices, PO does not even mention, let alone discuss the
`
`features of, its own prior art INOvent system. (PO Resp. at 58-59).
`
`3
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`
`Contrary to PO’s arguments for patentability, the problem allegedly solved
`
`by the ‘209 Patent was identified in the FDA Guidance, PO’s competitor’s patent
`
`documents, and PO’s own product documentation before the ‘209 Patent was filed.
`
`III. PO MISINTERPRETS THE INSTITUTED COMBINATIONS
`The PO Response spends only a handful of pages arguing about alleged
`
`deficiencies of the combination set forth in the Petition. (PO Resp. at 32-36). The
`
`instituted combinations rely on four primary references: the ‘083 Patent, the
`
`‘510 Patent, the FR ‘804 Publication, and the IR Standard. (Petition at 21-27).3
`
`The Petition explains the resulting system, for example stating:
`
`The result of incorporating the ‘510 Patent and the FR ‘804
`Publication in the ‘083 Patent’s delivery system is that gas data
`(which can include gas concentration in the cylinder per the ‘083
`Patent) is read from the carrier on the gas cylinder using an
`appropriate sensor (such as a bar code reader) and stored in the valve
`memory. This data would then be provided to the CPU of the ‘083
`Patent for use in the comparisons discussed above. (Ex. 1002 ¶ 94.)
`To provide that data to the CPU of the ‘083 Patent, a person of skill in
`the art would have considered at least known wireless medical device
`communication standards published by the ISO/IEEE. (Ex. 1002 ¶¶
`105-111.)
`
`3 The Petition, Decision, and PO Response use different shorthand for the prior art
`
`references. To avoid confusion, Petitioner relies on the shorthand in the Petition.
`
`4
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`(Petition at 24-25). The Petition also explains why the combinations disclose each
`
`limitation of the challenged claims. (See, e.g., id. at 27-50). PO’s argument that
`
`the Petition fails to “clearly demarcate…which elements it is proposing would be
`
`combined from each reference, and where it is relying on modifications of the
`
`references to reach the claim limitations at issue” is incorrect. (PO Resp. at 32-33).
`
`A.
`PO’s Complaints Regarding the Independent Claims
`PO identifies a single limitation of the independent claims allegedly not met
`
`by the proposed combination. It argues that the Petition does not identify why the
`
`prior art teaches communicating gas data to a “control module that controls gas
`
`delivery to a subject and to verify one or more of the correct gas, the correct gas
`
`concentration and that the gas is not expired.” (PO Resp. at 33). It relatedly
`
`argues the Petition does not show why this control module receives “gas through
`
`the valve” and that the claimed “verification” does not occur (Id.).
`
`Contrary to PO’s argument, the Petition explained why this limitation is
`
`satisfied when the CPU of the ‘083 Patent (part of the control module) performs
`
`the safety check of the FR ‘804 Publication prior to the ‘083 Patent’s control
`
`module delivering gas received from the ‘510 Patent’s valve:
`
`[T]he therapy gas exits the valve and flows into a fluid circuit, such as
`that disclosed in the ‘083 Patent. (Ex. 1005 at 3:61-5:59, Figs. 1, 2.)
`The fluid circuit and the control of that circuit by CPU 56 as disclosed
`
`5
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`
`in the ‘083 Patent is an example of the claimed control module. (Ex.
`1005 at 5:60-6:19.)
`(Petition at 28; see also id. at 29-33). The Petition identifies a valve (from the
`
`‘510 Patent) that allows gas to flow to a control module (from the ‘083 Patent),
`
`which the ‘209 Patent itself relies on as enabling disclosure of the claimed “control
`
`module.” (Id. at 29). It notes that “[w]hen the valve of the ‘510 Patent and the FR
`
`‘804 Publication are incorporated into in the ‘083 Patent, the combination discloses
`
`storing gas concentration data in the memory 22 for subsequent communication to
`
`the CPU of the ‘083 Patent’s control module. (Ex. 1004 at 5:43-6:2.)” (Id. at 31).
`
`“[W]hen the FR ‘804 Publication’s verification functionality is performed by the
`
`CPU of the ‘083 Patent, as in the combination relied on herein, the prior art
`
`discloses this limitation. (Ex. 1002 ¶¶ 116-117.)” (Id. at 32).
`
`PO also argues that “FR ‘804 does not disclose storing gas concentration
`
`data in a memory, and none of the references, alone or in combination, discloses
`
`communicating gas concentration data from the memory” to the CPU of the
`
`‘083 Patent. (PO Resp. at 35). This argument is unavailing based on the claims.
`
`Each claim of the ‘209 Patent recites a Markush group that can be satisfied by gas
`
`data selected from the group consisting of “gas identification, gas expiration date
`
`and gas concentration.” (Ex. 1001 at 16:29-31 (claim 1), 16:52-54 (claim 3),
`
`17:10-12 (claim 5), 18:7-9 (claim 6)). PO does not dispute that the proposed
`
`6
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`combination discloses gas identification data, so its argument does not overcome
`
`the prima facie obviousness case in the Petition. (Petition at 29). The Petition also
`
`shows why in the instituted combinations, “gas concentration data” of the ‘083
`
`Patent is transmitted for use by the control module:
`
`[T]he ‘083 Patent teaches that one of the gas data characteristics that
`can be stored and used to trigger alarms is gas concentration. (Ex.
`1005 at 5:60-6:4.) The ‘083 Patent also teaches that the concentration
`of the gas in the cylinder can be used to “verify that the proper supply
`is being utilized.” (Ex. 1005 at 6:5-8.) Accordingly, the combination
`of references…discloses that the valve memory stores gas data as
`required…(Ex. 1002 ¶ 100.)
`(Petition at 30). PO’s argument is thus incorrect because it misreads the claims
`
`and because it fails to account for the ‘083 Patent’s specific teaching of using gas
`
`concentration data to perform safety checks prior to delivering NO to a patient.
`
`PO’s Complaint Regarding Claim 2
`
`B.
`The only other complaint PO makes about the sufficiency of the art when
`
`combined relates to claim 2. Specifically PO argues that the “control module 300”
`
`of the FR ‘804 Publication “must be located at the valve memory—not elsewhere
`
`in the system” and “thus cannot meet the claim limitations.” (PO Resp. at 34-35).
`
`However, PO’s argument is incorrect because it fails to understand the teachings of
`
`the FR ‘804 Publication in view of the ‘510 Patent. As stated in the Petition:
`
`7
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`
`The FR ‘804 Publication discloses using bar codes encoding gas data
`on gas containers. (Ex. 1006 at 20-21.) Further, the FR ‘804
`Publication discloses using an appropriate sensor 110 (e.g., a bar code
`scanning device) to read the bar code and provide the read data to a
`control module 300. (Ex. 1006 at 20-21.) The ‘510 Patent discloses
`that gas data is inputted into the valve memory 22 via a user operated
`transfer device 44. (Ex.1004 at 5:61-6:2.)
`(Petition at 33). The combination of the FR ‘804 Publication and the ‘510 Patent
`
`discloses a sensor that reads gas data contained in a bar code and inputs that
`
`information into the valve memory; this disclosure satisfies every additional
`
`limitation of claim 2.
`
`Since PO’s only two arguments about the application of the instituted
`
`combinations to the claims of the ‘209 Patent are without merit, PO cannot rebut
`
`the Petition’s prima facie case that claims 1-7 of the ‘209 Patent are obvious.
`
`IV. THE INSTITUTED COMBINATIONS ARE PROPER BECAUSE
`PO’S ARGUMENTS ABOUT THE REFERENCES LACK MERIT
`
`The majority of PO’s substantive arguments deal with the alleged
`
`impropriety of the Petition’s combinations. (PO Resp. at 40-53). Since the FR
`
`‘804 Publication itself provides the motivation to communicate data stored in the
`
`valve of the ‘510 Patent to the ‘083 Patent’s CPU (see, e.g., Petition at 21-23; Ex.
`
`2020 at 111:22-122:18), PO’s assertions regarding combinability of the ‘510 Patent
`
`and the ‘083 Patent, on their own, is irrelevant. (PO Resp. at 40-45). Because the
`
`8
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`FR ‘804 Publication teaches the very “pre-use safety check” PO uses as a
`
`shorthand to describe the allegedly patentable feature of the claims of the ‘209
`
`Patent (see, e.g., Ex. 1006 at 19), the FR ‘804 Publication is an example of a
`
`reason and a way for the valve of the ‘510 Patent to communicate with the CPU of
`
`the ‘083 Patent to perform the safety check of the FR ‘804 Publication.
`
`Particularly since the FR ‘804 Publication was not considered during examination,
`
`PO cannot rebut the Petition’s prima facie case of obviousness.
`
`A. The ‘083 Patent4
`PO argues that the ‘083 Patent cannot be combined as proposed in the
`
`Petition because it discloses measuring NO and NO2 concentration at the point of
`
`inspiration by the patient, and therefore is not amenable to combination with the
`
`FR ‘804 Publication, which discloses using data stored on a carrier on the cylinder
`
`as an initial safety check. (See, e.g., PO Resp. at 13-14, 17). PO’s teaching away
`
`argument is contrary to common sense and the express teaching of the ‘083 Patent.
`
`PO’s Nitric Oxide-Specific Arguments Are Inaccurate
`
`1.
`PO argues that using data read from a carrier on a cylinder is superior to the
`
`technique of the ‘083 Patent, where sensed gas concentration at the point of
`
`
`4 According to PO, the ‘083 Patent describes “Patent Owner’s NO delivery
`
`systems.” (PO Resp. at 7; see also Ex. 2002 at 3 (marked with ‘083 Patent)).
`
`9
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`inspiration is used as feedback to the control system. (PO Resp. at 44-45).
`
`According to the FDA, “[n]itrogen dioxide is a toxic gas formed by reaction
`
`of nitric oxide with oxygen.” (Ex. 1012 at 9). 5 The ‘083 Patent confirms that the
`
`longer nitric oxide dwells in a delivery circuit, the higher the likelihood of
`
`formation of toxic nitrogen dioxide. (Ex. 1005 at 1:40-44).6 To control this risk,
`
`the FDA counsels that “[t]he administration device should include provision for
`
`nitrogen dioxide gas analysis with alarms. The breathing circuit location for
`
`sampling should sample gas which is representative of the inspired gas.” (Ex.
`
`1012 at 10). Thus, according to the FDA’s own guidance regarding nitric oxide
`
`delivery, sensors (akin to those disclosed in the ‘083 Patent) should be used to
`
`ensure appropriate levels of therapeutic and toxic gases are delivered to the patient.
`
`
`5 PO’s argument about “rebound pulmonary hypertension” is irrelevant. The
`
`“sudden cessation” PO cites could only occur if NO was being delivered to the
`
`patient and thereafter stopped; in this scenario, the data on the cylinder carrier is
`
`far less useful than the sensed NO/NO2 concentration at the inspiration point.
`
`6 This property of nitric oxide casts substantial doubt on the veracity of Mr. Heim’s
`
`testimony, in which he guessed wrong and testified that nitrogen dioxide forms in
`
`the presence of nitrogen, not oxygen (Ex. 1022 at 104:17-105:1) and alternatively
`
`that NO2 will be formed without any other reactant present (id. at 106-24:107-2).
`
`10
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`
`The record does not provide any support for the idea that using data on the
`
`cylinder is superior to measuring gas concentration at the point of inspiration.
`
`Incorporating the safety check of the FR ‘804 Publication is therefore an additional
`
`check, and not an alternative check, to those provided in the ‘083 Patent.
`
`2.
`
`The ‘083 Patent Obtains Concentration Data from Different
`Sources
`
`The only way PO can make its sensor-based argument is by ignoring
`
`alternative embodiments in the ‘083 Patent. (See, e.g., PO Resp. at 13-14). PO
`
`does not dispute that the ‘083 Patent’s CPU performs algorithms to control NO
`
`delivery (including ceasing NO delivery) based on inputs indicating NO
`
`concentration. (See, e.g., id. at 44-45). It teaches that “NO sensor 65 could, of
`
`course, be eliminated if the NO cylinder 10 is always constant or by keying into
`
`the NO sensor in the gas sensing bench 52.” (Ex. 1005 at 6:11-13). This is
`
`precisely the basis for combinability presented in the Petition: “[s]uch alternative
`
`gas data sources are expressly contemplated by the ‘083 Patent and would have
`
`been obvious to one of ordinary skill in the art. (Ex. 1005 at 6:11-15; Ex. 1002 ¶
`
`98-101.).” (Petition at 24). While Dr. Stone stressed the importance of this
`
`teaching in his deposition (Ex. 2020 at 92:11-93:13, 95:22-96:16) 7, Mr. Heim did
`
`7 PO argues that Dr. Stone’s testimony is insufficient because it does not render a
`
`conclusion on the ultimate issue of obviousness. (PO Resp. at 20-32). This is
`
`11
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`not address it at all (see, e.g., Ex. 2021 at ¶ 58-66).
`
`B.
`
`The ‘510 Patent
`1.
`Mr. Heim opined about the wired protocol disclosed in the ‘510 Patent.
`
`The ‘510 Patent Teaches Wireless Transmission
`
`(See, e.g., Ex. 2021 at ¶71). However, he did not dispute that the ‘510 Patent also
`
`discloses wireless transceivers “to transmit the data to a remote recording device at
`
`intervals or on command, as desired.” (Ex. 2021 at ¶79; see also, e.g., Ex. 1004 at
`
`7:1-4; Petition at 3, 25, 31). Accordingly, there is no dispute that the ‘510 Patent
`
`
`misplaced. The Board has held that “expert testimony is not required in every
`
`case.” Valeo, Inc. v. Magna Elec., Inc., Case No. IPR2014-00227 and IPR2014-
`
`00228, Paper 13 at p. 19 (May 29, 2014); see also Black Swamp IP, LLC v. VirnetX
`
`Inc., Case No. IPR2016-00167, Paper 12 at p. 4 (Feb. 4, 2016). And the Federal
`
`Circuit recently affirmed a Board decision of anticipation where Petitioner
`
`proffered no expert testimony and Patent Owner proffered extensive expert
`
`testimony. See Cutsforth, Inc. v. MotivePower, Inc., Appeal No. 2015-1315 (Dec.
`
`17, 2015); see also MotivePower, Inc. v. Cutsforth, Inc., Case No. IPR2013-00270,
`
`Paper 36, p. 2 (referencing testimony of PO expert Dr. Keim). Petitioner presents
`
`expert testimony here; it properly addresses technical issues not immediately
`
`apparent from the face of the references themselves.
`
`12
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`discloses that in some embodiments, data is wirelessly transmitted via a transceiver
`
`to a remote transceiver. This emphasizes the propriety of using the ‘510 Patent in
`
`the manner suggested in the Petition. (Petition at 31).
`
`The ‘510 Patent Suggests Using Stored Data for Control
`
`2.
`The Abstract of the ‘510 Patent discloses that:
`
`[R]elevant information, such as cylinder fill date, cylinder I.D.
`number, batch number, and patient name or account number may also
`be logged in the memory module. The log of the events and the
`corresponding dates and times may be used to prepare invoices for
`billing gas treatments, for inventory control, and for other record-
`keeping and control functions.
`(Ex. 1004 at Abstract). Dr. Stone testified that this discloses using data stored in
`
`the valve handle memory can serve a role in controlling therapy. (Ex. 2020 at
`
`124:16-125:6; see also id. at 128:23-129:1).8 Neither Mr. Heim nor the PO
`
`8 PO argues that the data stored in the ‘510 Patent valve memory is not the claimed
`
`“gas data” because it is not “gas type, gas concentration, or gas expiration date.”
`
`(PO Resp. at 16). The Petition explained why this is incorrect (Petition at 29) and
`
`even Mr. Heim conceded that inventory control requires knowledge of the gas type
`
`(Ex. 1022 at 78:20-24). Regardless, storing certain gas data is at least a motivation
`
`to look to storing other kinds of gas data, such as IDb of the FR ‘804 Publication
`
`(which PO concedes is the claimed “gas data”, see PO Resp. at 19) or gas
`
`13
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`Response addresses this testimony (or the clear disclosure of “control” in the ‘510
`
`Patent’s Abstract). Instead, Mr. Heim testified about column 7 of the ‘510 Patent
`
`(Ex. 2021 at ¶69) and implied that “control” in the ‘510 Patent deals with
`
`“inventory control.” However, the Abstract itself distinguishes “inventory control”
`
`from “other record-keeping and control functions.” (Ex. 1004 at Abstract).
`
`3.
`
`in Arguing the
`PO Misrepresented the ‘510 Patent
`Open/Close Data is Never Used for Real-Time Tasks
`
`PO argues that with regard to the ‘510 Patent, “[a]ll of the logging and
`
`tracking functions supported by the valve are contemplated for later analysis (e.g.,
`
`preparing invoices) as opposed to any real-time application, such as a safety check
`
`or verification of operating parameters, when the device delivers gas to a patient.”
`
`(PO Resp. at 15 (citing Ex. 2021 at ¶85)). Mr. Heim testified that this is
`
`inaccurate—at least displaying the open/closed status of the valve is a “real-time
`
`application.” (Ex. 1022 at 142:10-17). The operator can use the real-time display
`
`to assess the ongoing therapy, including the therapeutic impact on the patient. (Ex.
`
`1004 at 3:65-4:4, 5:31-42).9 Thus, the ‘510 Patent suggests that open/close data is
`
`concentration of the ‘083 Patent, in the valve memory. (Petition at 29-30).
`
`9 The Petition relied on the FR ‘804 Publication as disclosing controlling gas
`
`delivery upon verifying that the gas is the expected gas. The aspects of the ‘510
`
`Patent addressed here relate to the combinability with the FR ‘804 Publication.
`
`14
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`stored and used in real-time, and thus at least suggests that this data is amenable to
`
`use as the comparison data in the FR ‘804 Publication.
`
`C. The FR ‘804 Publication
`PO’s only real complaint regarding FR ‘804 Publication concerns
`
`combinability (PO Resp. at 45-53).10
`
`First, PO argues that the FR ‘804 Publication would not be combined as
`
`suggested because it is not directed to medical gas delivery devices. (PO Resp. at
`
`18-19). Mr. Heim admitted that the disclosed system “can be used for other
`
`operating circuit purposes besides filling bottles.” (Ex. 1022 at 144:13-14). One
`
`such use is in a “health facility.” (Ex. 2021 at ¶97; Ex. 1006 at 19). PO cannot
`
`dispute that Air Liquide (the assignee of the FR ‘804 Publication) was a company
`
`those of skill understood was in the business of providing inhaled NO in the prior
`
`art timeframe. Indeed, PO relies on this fact to make its secondary consideration
`
`arguments. (PO Resp. at 9-10, 58-59; Petition at 16-17; Ex. 1002 at ¶¶ 61-64).11
`
`
`10 PO’s argument about Air Liquide allegedly not including the features of the FR
`
`‘804 Publication are both tenuous from an evidentiary perspective (they are based
`
`on a single marketing brochure for a single product, see Ex. 2017) and irrelevant,
`
`as the combination involves the FR ‘804 Publication, not a prior art device.
`
`11 PO’s secondary considerations argument (PO Resp. at 58-59) fails to establish
`
`15
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`
`Second, PO argues that the FR ‘804 Publication is “inherently incompatible”
`
`with the ‘510 Patent because the “FR ‘804’s mechanism of operation relies on an
`
`automatic valve that defaults to a closed position and only opens when the
`
`controller signals a match in gas type,” while “[b]y contrast, [the ‘510 Patent]
`
`discloses a valve that is specifically designed for manual operation.” (PO
`
`Response at 49; see also id. at 49-53). PO’s argument is flawed for several reasons.
`
`Mr. Heim testified as follows:
`
`Q. …Prior to the opening of valve (20), is the gas from the
`cylinder occupying the elongated vertical rectangle between
`box (20) and box (12)?
`
`
`what features the prior art products and the INOmax DSIR had and did not have,
`
`and certainly does not perform a claim-by-claim analysis of these products. (PO
`
`Resp. at 8-10; Ex. 2021 at ¶¶ 111-13). Finally, PO fails to explain whether the lack
`
`of claimed features was due to failure of others to conceive or implement these
`
`features or some other regulatory, commercial, or logistical considerations). See
`
`AstraZeneca LP v. Breath Ltd., 88 F.Supp.3d 326, 391 (Fed. Cir. 2011) (no indicia
`
`of non-obviousness based on failure of others where no evidence of previous,
`
`unsuccessful attempts to achieve that which is claimed).
`
`16
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`
`A.
`
`There’s nothing in FR ‘804 that I recall that describes what
`goes on there, but it would be reasonable to expect that there
`would be gas in that chamber or line.
`(Ex. 1022 at 148:20-149:3). That is, the area of Fig. 1 highlighted in green below
`
`contains gas prior to the opening of the valve (20):
`
`
`
`Despite Mr. Heim being unable to fathom a single way this connection of the bottle
`
`might occur (Ex. 1022 at 149:4-150:12), such connection reasonably involves
`
`manual activity (Ex. 1006 at 20). The FR ‘804 Publication thus discloses that a
`
`valve beyond the detector 12 can be automatically actuated to either deliver or not
`
`deliver gas to the circuit 1. (Ex. 1006 at 20; Ex. 1022 at 148:20-149:3). The ‘083
`
`Patent discloses a system that operates precisely the same way as the FR ‘804
`
`Publication: gas flows from a cylinder into the circuit, where it hits an “on-off
`
`shutoff valve 14” that is solenoid operated. (Ex. 1005 at 3:61-67). That shutoff
`
`valve 14 allows the CPU to “discontinue the use of the NO to the patient by
`
`17
`
`

`
`Case: IPR2015-00889
`U.S. Patent No. 8,573,209
`
`shutting off the shutoff valve 14…” (Id. at 8:8-10).
`
`Third, PO’s argument is flawed because it asks the ‘510 Patent to disclose
`
`more details of a gas delivery system than is required of the instituted combination.
`
`The ‘510 Patent focuses on the valve and smart handle attached to a cylinder. (Ex.
`
`1004 at Abstract). It discloses that “[t]ypically, the outlet port 20 of the valve 10 is
`
`connected to a delivery device, such as a ventilator (not shown), which is used to
`
`adjust the concentration and flow rate or to mix gases administered to t

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket