throbber
3.2.4.1
`
`'Masking'
`
`For many of the procedures detailed, it is necessary to 'mask' the data. In essence, this
`removes unwanted data from the image. In the example below, a printed dot is shown,
`masked, and urunasked, Figure 3-9.
`
`Figure 3-9 - unmasked
`
`Figure 3-9 - masked
`
`Figure 3-9 - Masked and unmasked images
`
`Masking may be performed automatically, by height, co-ordinate, or by defining
`either a specified shape (rectangle, circle etc) or by defining an irregular polygon, as
`required. The technique is used to isolate cells (see Appendix l), or to isolate dots
`(see section 4.4.2.4) as well as to remove rogue data from measurements. For the
`analysis of dots, polygons were created to isolate the dot. In the case of cells, either a
`polygon or a circle was used depending on the size of the cells, and the amount of
`space surrounding the detail in question.
`
`3.2.5 Cell Volume
`
`Three different interferometers were available for use, and were used as appropriate.
`The first is a large, laboratory based interferometer. The interferometer itself is based
`on an air-table, minimising vibration, and offers the highest flexibil ity. Multiple
`internal magnifications are available, and several lenses may be mounted on the
`measurement turret. This interferometer was used in the analysis of diamonds.
`Although it is the most flexible in terms of measurement options, it is not possible to
`
`46
`
`FAST FELT 2010 , pg. 61
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`mount a gravure cylinder under the measurement turret. The second is a portable unit.
`Utilising no internal magnification, and a CCD camera of the same resolution as the
`laboratory based unit this device was used primarily for measurements taken on
`cylinders (cells or surface finish). The third device is another portable device, utilising
`O.Sx internal magnification. This device uses a lower resolution CCD, offering a
`lower horizontal resolution, and also has a lower vertical resolution. This device was
`used where excess vibration occurs, as it is less sensitive to vibration than the higher
`resolution portable device.
`
`As previously indicated in 3.2.3, traditional volume measurements are taken by
`measuring the length and width of a cell using optical microscopy, and inferring a
`volume from these measurements and the stylus angle. Microscope images of cells are
`shown in Figure 3-10- below.
`
`Figure 3-10- Microscope image of cell
`
`If the stylus angle, e,
`formula;
`
`is known, the volume of the cell is calculated, using the
`
`I
`~ume=
`(
`
`height x width) ( width
`)
`x
`2
`2tan(B/2)
`
`l
`x-
`3
`
`Equation 3.1 -Traditional volume calculation
`
`Results from this formula will be compared with results generated from the method
`that was developed based on actual cell geometry measurement.
`
`As a starting point, measurement of engraved cells made with the system as supplied
`is shown in Figure 3-ll. Due to low reflectance, optical noise and an inability of the
`
`47
`
`FAST FELT 2010 I pg. 62
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`lens to collect light back from the sloped cell sides, approximately 99% of the data
`
`from inside the cell is missing (left image).
`
`Figure 3-1 1 - Interferometric measurement of gravure cells
`
`In the example shown above (Figure 3-11), black indicates missing data. The only
`
`information available for calculation is the apex of the cell and the base corresponding
`
`to the roll surface. There is a requirement to interpolate the remaining data to
`
`calculate the volume within the cell. Also in Figure 3-ll , the right hand image shows
`
`a measurement taken with an improved optical system. Now 99% of the data from
`
`inside the cell is present, and calculation of actual volumes based on measurement can
`
`be performed, rather than inferring it. The optical configuration was improved by
`
`using a higher magnification lens. This allowed the measurement of sloped surfaces
`
`(such as the inside of a cell) and by minimising the number of cells being measured,
`
`the amount of optical noise is reduced.
`
`Because the sides of the cells are sloped, most of the light is not reflected back up into
`
`the interferometer, but bounces off the other walls, causing optical noise. By
`
`increasing the intensity, and decreasing the number of cells, the opportunity for noise
`
`is reduced, and more light is reflected from the cell walls, allowing measurement
`
`(Figure 3-l l, right hand image). With this much higher data rate, this allows accurate
`
`calculation of volume rather than the inferences, which are characteristic of previous
`
`works.
`
`48
`
`FAST FELT 2010, pg. 63
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`Previously several different possible techniques had been attempted, including coating
`
`the surfaces with carbon to reduce the reflectance, and minimise noise, and by
`
`removing filters from within the interferometer, to reduce the coherence length, and
`
`thus improve the sensitivity of the interferometer to measure the sloped cell walls.
`
`However, these methods had not provided data of sufficient quality to facilitate actual
`
`volume measurement.
`
`An analysis package was supplied with the interferometer to allow calculation of the
`
`specific volume of anilox rolls. This may, according to documentation, also be used
`
`for analysis of gravure cells. However, several assumptions are embodied in the
`
`measurement and data reduction. The analysis package's volume measurement system
`
`was re-evaluated with the new optical system. A much higher data rate was achieved,
`
`covering just one cell, however as the automatic analysis needs at lest 16 cells,
`calculations could not be performed using the supplied analysis package. It was also
`
`discovered on analysis of the automatic function that incorrect values (for example, an
`
`unrealistic cell volume) are often calculated for reasons that could not be established.
`
`Therefore a new methodology was required.
`
`c
`
`Surface
`roughness
`
`A
`
`Figure 3-12 - Dia.gram of Surface and Cell
`
`To facilitate accurate volume measurement. roll surface roughness needs to be taken
`
`into account, see Figure 3-12. If the volume was calculated below level 'c' this would
`
`include volume in the land surface of the cylinder in addition to the cell volume.
`
`Therefore before a volume measurement can be calculated, a series of surface levels
`
`49
`
`FAST FELT 2010, pg. 64
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`must be defined, the volume below which can be calculated. The total volume
`required will be the sum ofvolume V 1 and volume V2 in Figure 3-12 - i.e. from the
`apex of the cell to level 'C'. It is necessary first however, to define the 'land level' -
`i.e. the modal height for the unengraved, top surface of the cylinder, level 'A' in
`Figure 3- 12 which can be simply established from analysis of the data. However,
`using this datum line for volume calculation includes volume from the top surface of
`the cylinder in the measurement, which is not part of the cell volume. As a result of
`this the datum for volume measurement was redefined to level B, by stating that:
`
`Datum height = Average land height - Y2 Rz
`
`Equation 3.2 - Datum height calculation
`
`Part of the cell is formed below this datum (Figure 3-12 - volume V 1), and that
`anything above it is merely a part of the cylinders own rouglmess - barring the space
`above the actual cell (Figure 3-12- volume V2). Data above the datum line (indicated
`by line 'B' in Figure 3- 12) must be removed for volume measurement. This removes
`the land area from the measurement, leaving only the cell. Should the measurement
`contain other cells or parts of cells, these must also be removed, using the masking
`facility.
`
`To compute measured volume, the area of a single pixel is calculated, and multiplied
`by the depth from the datum height in the measurement. This result is then integrated
`across the entire measurement field to give a total volume (V1) for the cell. This is an
`automated function, built into the analysis package supplied. The correctness of its
`working was confirmed via the use of a stepped surface7
`, and was found to be
`accurate.
`
`This volume does not yet represent the volume of the engraved cell. The top part of
`the cell has been removed in order to remove the land area (volume V2), and this
`
`7 The stepped surface used was an extremely smooth and accurately defined step, used for calibration
`of the equipment, As the dimensions of the step are well specified by the manufacturers, the volume
`can be calculated malhematically and by the measurement technique.
`
`50
`
`FAST FELT 2010 , pg. 65
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`(
`
`needs to be added back in to calculate the final cell volume. This section is calculated
`using the open area of the cell (section 3.2.6). This is multiplied by the Rz to give
`volume V2, Figure 3-12, which when added to the cell volume previously calculated
`(V J) gives a total volume for the cell. A further section is missing around the very top
`section of the cell, indicated by the space between the volume V2 and the surface
`roughness in Figure 3-12. This was calculated as being between 0.3% for a small cell,
`to 0.1% for a larger cell. This is negligible, and was excluded in the analysis.
`
`Primary work was undertaken to calculate the number of cells that must be evaluated
`in order to provide an acceptable average. An example of this type of calculation is
`given in section 3.6 (for the number of analysed dots). It was found that the analysis
`of four cells was required. For each analysis of each type of cell, five were measured
`the additional cell being included to ensure that enough measurements were obtained
`in the event of a problem with one of the measurements, and results were then
`averaged. An example of calculation of all parameters is given at the end of this
`section.
`
`3.2.6 Cell length, width depth and open area
`
`With the measurement area masked for the volume calculation (see 3.2.5), the length
`and width of the cell was also calculated. Length is determined as the distance
`between the point at which the diamond enters the copper to engrave the cell, to the
`point at which it leaves. This is indicated by length 'a' in Figure 3-13. The width is
`defined as the maximum distance between points in the cross-cylinder direction. This
`is indicated by length 'b' in Figure 3-13. It was taken to be a simple linear measure,
`the curvature of the cylinder making little difference to the length on these scales. The
`dimension 'A' is however, not determined by diamond geometry, but by the speed at
`which the diamond is pushed into the copper, and the speed of rotation of the cylinder
`during the engraving process.
`
`51
`
`FAST FELT 2010 I pg. 66
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`Figure 3-13 - Width and Length calculation
`
`If all data above the datum line is masked, as in 3.2.5 (volume calculation) the number
`of measurement points making up the cell can be measured. By multiplying the
`number of pixels by the area of each pixel, the open area of the cell may be calculated
`(Equation 3.3)
`
`Open_Area = LN11 xP"
`
`Equation 3.3 - Open area calculation
`
`The measured depth of a cell is the vertical distance from line 'A' to the base of the
`cell in Figure 3-12. The actual depth of the cell must also include the 0.5 R2 that was
`subtracted by the manual masking. (see section 3.2.5). Adding this 0.5 Rz on to the
`depth ensures that the top and bottom of the cell are m the same place for both volume
`and depth calculations.
`
`3.2. 7 Cell offset
`
`Due to the electromechanical engraving technique used in the production of the
`cylinder the lowest point in the cell is not precisely in the middle of the cell. The
`offset angle defines the angle between a vertical line from the bottom of the cell and
`the line joining the bottom of the cell to the geographical centre of the cell. The offset
`is directly related to the diamond and its holder used in the engraving. A typical
`
`52
`
`FAST FELT 2010, pg. 67
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`example of the type used is shown below (Figure 3-14). The diamond is rotated into
`the copper about the pivot point. As a result, the diamond tip does not execute a
`straight line as it enters the copper, but a curve, which shows itself as the measured
`offset value.
`
`Pivot Point
`
`Cylinder
`
`Path of diamond
`
`Figure 3-14 - Engraving Diamond
`
`This can be seen most clearly if the cell data is inverted, Figure 3-15.
`
`Figure 3-15 - Inverted cell, showing offset
`
`The offset is indicated by the manner in which the cell 'leans' to the right.
`
`3.2.8 Cell wall roughness
`
`Once a cell has been measured, it is possible to select only a small portion of the cell,
`and flatten it in the measurement plane through the application of a mask (section
`3.2.4.1 ). From this, it is possible to calculate the roughness of the wall inside the cell.
`
`53
`
`FAST FELT 2010 , pg. 68
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`An example is shown in Figure 3-16. A preliminary investigation was performed.
`Although
`it highlighted internal cell features,
`
`it did not affect cell volume
`significantly and so was not considered to be important in the context of this study .
`
`oP------------------------..
`
`"
`..
`"'
`"
`
`,.,
`,..,
`
`1<10
`
`. ...
`
`OJI)
`
`<lSI
`
`·I«<
`
`·I !I)
`
`•Uil
`
`·!&J
`
`Figure 3-16- Cell wall quadrant, showing roughness
`
`3.2.9 Example calculation
`
`A typical measurement is shown below, Figure 3-17. Several cells are included in the
`measurement, and the measurement must be masked to remove the additional cells.
`
`54
`
`FAST FELT 2010 , pg. 69
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`~ Welsh Centre for
`~ Printing and Coating
`
`Surface Data
`
`Mag 181 X
`Mode VSI
`
`Date IOIIIn002
`Tune 10 5645
`
`:m
`
`2SJ
`
`:Ill
`
`ISl
`
`0111
`
`Sl
`
`Swface Statistics:
`Ra· 4 'l7 urn
`Rq 5 05 urn
`Rz 20 IOurn
`Rt 10 46 urn
`Rku 3 02
`Rsk -1 13
`
`Set-up PanJM1US:
`S.u 480 X736
`Samphng $42 39 nm
`Pnocessed OplioftS:
`T enns Removed
`Ttlt
`Flltenng
`Median
`
`Title: db70.1
`Note:
`
`so
`
`JD
`
`10
`
`· 10
`
`-3J)
`
`.. J)
`
`·l Jl
`
`ofO
`
`•UD
`
`·IJO
`
`•1$1
`
`Sl
`
`m
`
`1lD
`
`:m
`
`:I!D
`
`Figure 3-17- Raw measurement of engraved surface
`
`Figure 3-18 shows the individual cell to be analysed. This has been masked, removing
`the extra cells,. The shape of the mask is not critical, and here a simple circle has been
`used to isolate the cell.
`
`55
`
`FAST FELT 2010 , pg. 70
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`~ Welsh Centre for
`~ Printing and Coating
`
`Surface Data
`
`Mag- 18 I X
`Mode- VSI
`
`Date 10/1112002
`Tsme 10 56 45
`
`31l
`
`:m
`
`2!ll
`
`2ll
`
`lSI
`
`1111
`
`<II
`
`Surface Statistics:
`Ra 381 urn
`Rq 473 urn
`Rz 19 04 urn
`Rt 1976 urn
`Rku 4 01
`Rsk -I 48
`
`S.t.ap Para.eters:
`SIU 480 X 736
`Semphng 542 39 nm
`Pncessecl Optilnas:
`Terms Removed
`TilL
`Flltenng
`Medtan
`
`Title: db70.1
`Note:
`
`Figure 3-18- Masked measurement of engraved surface
`
`A measurement screen was set up to allow calculation of depth and volume, Figure
`3-19. This incorporates both a 20 image of the cell, (top left), a volume chart (top
`middle - unused), the volume information (top right) and a histogram, (bottom).
`
`56
`
`FAST FELT 2010, pg. 71
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`Volume Calculations
`
`Results
`VaNd Points
`Pixel SIZe
`Volume
`
`rm· II.OO'f.ofP.V
`Pis Below: 110.00% of Tote I
`Vol: &.38e+004 um3 100.00% of Tote I
`
`Histogram
`
`8021 7
`542 39 nm
`63789 um3
`
`A
`
`...
`
`..
`
`ltcljll
`
`..
`
`Figure 3-19- Developed calculation screen
`
`The histogram (at the bottom of the screen) shows the number of points at any given
`height. The 'Volume' number is the number of cubic microns of fluid which would be
`required to submerge the sample to the highest point on it. This is equivalent to the
`volume described in section 3.2.5 as the volume below line 'C' (Figure 3-12) The
`'Valid Points' variable, gives the number of pixels in the measurement.
`
`Histogram
`
`Jl ...
`
`...
`
`.,
`
`...
`
`·•
`
`..
`
`Ktlg ll
`
`·•
`
`·•
`
`Figure 3-20- Histogram indicating peak
`
`57
`
`FAST FELT 2010 , pg. 72
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`rn Figure 3-20 the highest point indicates the most common height for pixels - in this
`case, relating to the height of the land area.
`
`..r
`~ r-- ()(30om ]-- - - -
`=
`
`y 2553 0
`
`-= =
`--= ---=
`= ---
`=
`--=
`.-
`
`...
`
`Histogram
`X 188um
`
`-----
`
`(
`
`lt •158 .....
`Y• 4j)
`
`----
`
`-
`
`T
`I
`
`\ I
`
`j
`
`--
`
`__ __,r
`
`-
`
`•tl
`
`. ..
`
`.. Htlglt
`
`·l
`
`Figure 3-21 - Histogram, indicating cell depth
`
`The height from the land area to the lowest data point gives the depth of the cell,
`mjnus the 1/ 2 Rz, which for this cylinder = 0 .8Jlm above the mean land level, thus the
`total depth of the cell is 19.6Jlrn, Figure 3-21 (18.8jlm as indicated, + 0.8Jlm =
`19.6Jlm)
`
`Htstogram
`
`O~h~ J -------------
`
`v
`
`....
`
`I'Ll ~ .... •,1
`
`------
`
`Figure 3-22 - Histogram indicating masking point and direction
`
`The cursor has now been moved 0.8Jlrn to the left (down), and a 'Mask Right' is
`about to be performed. This removes all data to the right (above) of the cursor, Figure
`3-22. This leaves the part of the cell below line 'B' m Figure 3-12. As stated above,
`
`0.8~ is equivalent to 'h Rz for this cylinder.
`
`58
`
`FAST FELT 2010 , pg. 73
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`Volume Calculallons
`
`Results
`Valid Points
`Pixel stze
`Volume
`
`nm·· 0.00% of P-V
`pta Below: 110.00% of Total
`Vol: 5.62e+D04 um3 110.00% of Total
`
`Histogram
`
`31144
`542 39 nm
`56225 um3
`
`·•
`
`·•
`
`..
`
`·• N•IJII
`
`Figure 3-23- Entire data for masked cell
`
`The volume (V 1) for the cell may now be calculated using data taken from the
`analysis package, Figure 3-23. In this case;
`
`• Volume = 56225um3
`
`• Valid Points = 3l l44 points
`
`The area of this open area must now be calculated. The total measurement area is
`equal to
`
`342 urn x 260um = 88920urn2
`
`using a camera of resolution 736x480 pixels. Thus I pixel equals;
`
`342 x 260 = 0.252 um2/pixel
`736 X 480
`
`59
`
`FAST FELT 2010 , pg. 74
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`Thus
`
`Total Volume= Volume of cell+ Volume of top section
`
`Total Volume= Volwne of cell+ (Open area x R,)
`
`Total Volume= Volume of cell + (Valid Points x Area/pixel x 1.6)
`
`Total Volume= 56225 + (31144 x 0.252 x 1.6)
`Total Volume= 68,782um3 for this cell
`
`The cell dimensions may be established using a similar strategy, Figure 3-24.
`
`\I !i
`
`X Profile
`x 103.8 um
`
`\
`--~.,..,__,. -- -..-,
`..
`..
`
`Y Profile
`lt 1238um
`
`..
`...
`
`1.!
`!!
`
`!l!!l!!
`e-
`1!1!111!
`br%1!
`""•
`
`-
`
`1 J
`
`""
`I. 0
`1 "' ~
`
`..
`
`!,!
`
`1.!
`1!
`b
`
`.,,_
`
`~jf ...
`l!E•
`:IH~
`·lfllt ...
`
`., .....
`
`?9 .. f.J ._
`
`~J!
`·2D.•
`-1!!.91-
`
`,..,_
`
`4J ....
`l'l!.
`'8:•
`-UJ-..
`
`-
`
`..;
`
`X
`y
`!rt
`
`Wll
`19"
`.. ,
`
`~
`Tide: db70.1
`Note:
`
`---
`
`~ u .....
`ca U2-
`l!!M
`l!l!lf
`~~~ •lr?J: :S
`
`""· -ma-
`
`Figure 3-24 - Measurement of depth and width
`
`Here the length and width of the cell may be calculated, by finding the north, south,
`
`east and west extremes of the cell (indicated by the triangles in the cell picture, top
`
`left). The width is defined as the measured distance in the 'X Profile' (103.8um) and
`
`the length as the measured distance in the 'Y Profile' (l23.8um).
`
`60
`
`FAST FELT 2010, pg_ 75
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`X Profile
`~.2~
`
`...
`
`Y Profile
`lt 123 8 um
`
`._,-
`··'·~
`
`U:~•
`:!" !!!l
`•lt,71-
`
`!S
`1&
`II.
`k
`b
`
`J ...
`;.:,
`0
`"t ~
`
`,.
`
`!:!!!! m.lli!-"
`c .....
`·1&.H ...
`tapy
`VtP
`·2J:I•
`laldl
`,mn-
`
`.....
`
`...
`
`J. ~
`c
`T 0.
`~
`
`.gJ ...
`
`1.)1-
`1'22•
`'2!.
`
`-9~-
`
`!5
`1&
`II
`k
`b
`
`!:!!!!
`c .....
`z..
`.....
`bUll
`
`Ul..&
`t l2 -
`Ita
`•l21-
`·) ) ) l f -
`
`-
`
`W81
`
`. -. -..
`
`Title: db70.1
`Note:
`
`Figure 3-25- Measurement of offset
`
`The offset is calculated in Figure 3-25. The top of the cell is the point at which the
`
`diamond enters the copper (engraving from top to bottom) in this example, so the blue
`
`line (Y Profile) is lined up on the northmost point of the cell. The lowest point in the
`
`cell is found, and the distance in the X direction (east-west) is calculated from this
`
`point to the line coming down from the diamond entry point, in this case 8.2um.
`
`It is also possible to measure the roughness of the internal cell walls. First, the first
`
`quadrant to be examined must be isolated, Figure 3-26. Once the quadrant is isolated,
`
`selecting the 'terms mask' allows the section isolated to be used as the measurement
`
`plane.
`
`61
`
`FAST FELT 2010 , pg. 76
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`"
`
`p.a
`::J
`<'~Joe~ Jc:-~wo
`lr p.,.
`r-....!l!lJ~l
`_J _,___j _J
`~_J....i.._jl
`~ ...,.E_ r
`P ... l;p .. II
`__J
`y...
`u .. .
`,- RM
`~01&0
`t"" D .. &. M•lY f'~
`r
`..;
`(' Ha1k
`
`Figure 3-26 - Cell wall roughness, isolating the quadrant
`
`Once the measurement area has been isolated, curvature must be removed, effectively
`
`rendering a level plane.
`
`Once this has been selected, the quadrant is levelled, and the rouglmess, in terms of
`
`Ra, Rq etc, may be read from the analysis screen, Figure 3-27.
`
`Slllface Statistics:
`Ra: 611 .41 run
`Rq: 837 33 run
`Rz: 4.98 urn
`Rt· 10. 10um
`Rku: 9.78
`Rsk: 1.04
`
`Set-vp Parameters:
`Size. 480 X 736
`Sampling: 54239 nm
`P.rocessed OptiDJtS:
`T enns Removed:
`Curvature & Tilt
`Filtering:
`Medi81l
`
`76
`
`6D
`
`5D
`
`4D
`
`JD
`
`2D
`
`ID
`
`DD
`
`-10
`
`-25
`
`Figure 3-27- Flattened quadrant
`
`(
`
`62
`
`FAST FELT 2010, pg. 77
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`3.2.10 Comparison with traditional methods
`
`A series of comparisons benveen different cell models was performed, to see if it was
`
`possible to use the microscopy measurements to establish the cell volume.
`
`Calculations were performed on an 80% cell (the largest cell with no channel present),
`
`engraved with a 130° stylus. Average cell dimensions were used (137.9um wide,
`
`153.3um long). The measured volume was compared with both the traditional,
`
`pyramid shape approximation, and a new, average circumscribing conical model
`
`where;
`
`1r (width+ length )
`VO ume =-X
`I
`3
`4
`
`2
`
`width
`X --:-----:-
`2 tan(B / 2)
`
`Equation 3.4- Conical model
`
`The possibilities of user error are much larger when using the image analysis
`
`techniques than in the measurement technique, described previously due to the
`
`necessity for user intervention when defining the width and length of cells. This may
`
`lead to inaccuracies ofup to approximately 3j.lm in each dimension, which will have a
`
`significant effect on the results obtained.
`
`In addition, the stylus angle must be assumed to be perfect for this calculation. Since
`
`it is shown in Appendix 1 that the diamond angle changes with wear, this adds a
`
`further inaccuracy to the calculation. Clearly this error in stylus angle is not relevant
`
`to the interferometric technique described as this measures the volume of the cell
`
`itself, without inferring the internal shape.
`
`3.2.11 Blade load measurement
`
`(
`
`The effect of blade load on ink release was explored with the project (see figure 3.1)
`
`and a means of quantifying this is required. Practically the doctor blade is loaded on
`
`the cylinder pneumatically. The load is controlled using a gauge on the press, which
`
`63
`
`FAST FELT 2010 I pg. 78
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`monitors the pneumatic pressure supplied to the pistons. Thus strain gauges were
`
`glued to the doctor blade, and the bending of the blade monitored. However this does
`
`not allow a direct measure of load due to the friction that is present between the blade
`
`holder and the press. This friction may be variable and therefore blade load may not
`
`be indicated consistently via the pneumatic supply pressure. The strain induced in the
`
`blade by the application of the blade to the cylinder was monitored with a PC based
`
`datalogger. This was used to ensure that the same load was applied for each trial,
`
`rather than establishing actual load levels. The latter is particularly difficult to achieve
`
`and requires the application of a calibration procedure coupled with some analysis of
`
`blade stressing [ 61] [ 62].
`
`In the case of the scumming press trials, gauges were mounted in the centre of each
`
`surface finish band on the cylinder. This gave confirmation that the same load was
`
`being applied to each surface, with no significant variation caused by either the blade,
`
`cylinder or mounting system.
`
`3.3 Running the Press Trials
`
`3.3.1
`
`Ink Release From Image Areas
`
`Several inks are available to gravure printers, but in the UK at least, almost all
`
`gravure inks are of one of two standards, either NC (Nitrocellulose) based or PVB
`
`(Polyvinyl Butyral) based. NC inks are the most common, largely because of their
`
`lower costs, and (in the opinion of most printers) lower scumming attributes. As a
`
`result, NC inks were used for this trial.
`
`Several different substrates were considered for the trial, from newsprint to
`
`cartonboard. The gravure industry is approximately evenly split between the use of
`
`paper and film substrates, but the packaging industry primarily uses film substrates.
`
`Due to the need to measure ink transfer, it was decided to use a plastic film. This
`
`(
`
`would not absorb the ink (as most papers would) but would leave the ink on the
`
`surface of the substrate, where the volume transferred could be measured, and thus
`
`64
`
`FAST FELT 2010, pg. 79
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`quantified. It was decided to use a film commonly used in the industry, and as such it
`was agreed to use Mobil 247, white OPP (oriented poly-propylene) substrate due to its
`widespread usage, and thus direct commercial and practical implications for this
`research.
`
`3.3.1.1 Trial Methodology
`
`Prints were made on a standard OPP substrate using a small commercial press. The
`appropriate ink was loaded into the ink tray, and the press was run up to speed
`(I OOm/min). Prints were made for 60s, producing approximately 1OOm of print I 160
`repeats of the image. Prints were collected at the end of the press on the take-up spool,
`and all prints were retained for analysis. Unprinted samples of substrate were also
`collected to allow analysis of the substrate. The speed of the press was confirmed
`using a handheld optical tachometer.
`
`A series of thermocouples was used to monitor operating temperatures. They were
`placed on either side of the ink tray, two were attached to the frame of the press, and a
`further thermocouple was used to monitor ambient temperatures.
`
`Strain gauges were attached to the doctor blade to accurately measure the applied load
`in three places along the length of the doctor blade, or over the width of the cylinder.
`Blade load measurements and thermocouple measurements were logged at a rate of
`0.5Hz over the entire period of the measurements. These were used to ensure that the
`blade load was maintained, constant over the duration of the trial, it having been
`shown previously that doctor blade load I angle can make a significant difference to
`the printed colour I density (39] (40].
`
`65
`
`FAST FELT 2010 I pg. 80
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`0
`
`c
`
`c
`
`c
`
`3.3.2
`
`Ink Release From Non-Image Areas
`
`3.3.2.1
`
`Ink Selection
`
`A variety of inks were selected for use. For the surface finish trial, both NC and PVB
`
`inks were used, whilst at the ink trial, three NC inks were used. NC inks were used
`
`because these are the most commonly used in the packaging industry [10] [63] [64]
`
`[65] [66]. It was decided to use standard commercial inks to ensure correlation
`
`between experimental results and practical experience.
`
`Inks were standard,
`
`commercial inks, supplied as they would be supplied to any printer, and thinned with
`a standard solvent. In the case of the second trial, a special 'low scum' ink was tested,
`but again, this was a standard ink available from the manufacturer.
`
`The surface finish trial used black inks, while the ink trial used red inks. In each case,
`
`the inks used were produced to the same initial colour strength. They were then
`
`thinned with solvent, thus lowering the viscosity, and also lowering the colour
`
`strength. By reducing each ink to the same viscosities, colour strength was maintained
`
`at approximately similar levels.
`
`3.3.2.2 Substrate selection
`
`A variety of substrates were selected for use. For the first trial, three distinct
`
`substrates were chosen. A white OPP (Mobil 247 - as with the ink release from image
`
`areas trial), a clear OPP and a metallised foil. These were selected as they make up
`
`approximately 90% of the output of the printers. It was suggested that almost all
`
`scumming problems occur on plastic substrates [9] [10] [63) [64) [65] [66], so it was
`
`decided not to examine the effects on papers in this experiment.
`
`For the second trial only two substrates were examined, the same white OPP and the
`
`same clear OPP. These were selected for the same reasons, as well as to allow direct
`
`(
`
`comparison between the two trials.
`
`66
`
`FAST FELT 2010, pg. 81
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`c
`
`c
`
`3.3.2.3 Trial Methodology
`
`Prints were made on commercial presses8, under conditions as close to standard
`
`
`production conditions as possible, and conditions were replicated over the two trials.
`
`Print speed was 250m/min, and impression pressure was set at 2 bar, gauge. Blade
`
`load was set at a maximum and reduced (trial one) or at a minimum and increased
`
`(trial 2). In each case, the load was incremented by 0.25 Bar (gauge). Thus load was
`
`indicated by pressure setting on the press, and was monitored using strain gauges.
`
`Once the press was up to speed, it was run for 45s, the blade load was incremented,
`
`the press was run for a further 45s, and the process was repeated until either
`
`scumming appeared to be eliminated, or until 0 Bar gauge pressure was being applied.
`
`The press was only stopped when necessary for ink or cylinder changes
`
`3.4 Print Measurement Techniques
`
`As detailed, interferometric analysis was used to quantify the features of the cylinder.
`
`Further application of this technique was used for the analysis of prints, allowing the
`
`calculation of precise dot volumes. Spectrophotometric techniques were also used,
`
`allowing the calculation of traditional values of tone gain I CIEL·a·b• colour.
`
`Measurement methodologies for colour and density measurement have been
`
`standardised, and these standard methods are described in [ 15]. The following
`
`sections describe the key elements.
`
`3.4.1 Printed Density and colour
`
`Measurements were taken with both handheld Gretag SPM-50 spectodensitometers
`
`and automated Gretag Spectrolino/Spectroscans. All results were taken using a D50
`
`illuminant, 065 filter, and using the ANSI T standard. Transmission measurements on
`
`8 In each case, I 0 units were available, with a printing width of approximately 1.2m. Press footprint
`was approximately 4m x 30m, with a height of 5m
`
`67
`
`FAST FELT 2010 , pg. 82
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`c
`
`c
`
`c
`
`(
`
`clear
`
`substrates were
`
`taken usmg a Gretag Spetroscan-T
`
`transmission
`
`spectrophotometer.
`
`3.4.1.1 Density calculation
`
`To produce prints of differing colour, it is necessary to vary the thickness of the
`
`printed ink film. Since the rotogravure process is not a contone process, it is necessary
`
`to use an alternate method. By varying the size of the cells, and thus of the printed
`
`dots, it is possible to vary the amount of ink transferred over a given area. These
`
`different sizes of dots are referred to as halftones [ 19]. The density of this area
`measures the intensity of the printed colour. It is defined as:
`
`density = log10
`
`1
`reflectance
`
`Equation 3.5 - Density calculation
`
`3.4.1.2 Optical Tone gain calculations
`
`When ink is transferred from the printing cylinder to the substrate, the printed dot will
`
`not in most cases be the same size as the cell from which it is printed. This change in
`
`size is referred to as the tone gain (or dot gain). Tone gain is the effect caused by the
`
`ink spreading out on the substrate, and thus giving a higher optical density than would
`
`otherwise be expected. Tone gains were calculated using the Murray-Davies equation
`
`(see Equation 3.6, below)
`
`Where:
`a = printed area, 0 1 = tonal density, Ds = solid density
`
`Equation 3.6- Tone gain calculation
`
`68
`
`FAST FELT 2010, pg. 83
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`0
`
`In certain circumstances, the printed dot may be smaller than the cell opening. In this
`
`case, this is still referred to as tone gain, (with a negative value) and generally only
`
`occurs in the highlight regions of t

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket