throbber
' -0 ·
`
`Colour
`
`Fiker
`
`L
`
`Lens *~
`r "~ -~
`~'"""""'""'00
`
`Polensauon fik11r
`
`I
`
`Figure 2-3 - Densitometer
`
`As in a spectrophotometer, light is reflected from the substrate under test, and the
`reflected light is passed through a lens or prism, and collected on a CCD. The CCD
`applies four filters, to calculate the reflectance m the cyan, magenta, yellow and
`'black' regions of the spectrum. These filters will be applied electronically, and
`densitometers are also available to measure hexachrome and similar colour printing
`systems.
`
`The reflected light is filtered to only analyse light over a certain wavelength band
`Two examples are shown below, for cyan and yellow, Figure 2-4.
`
`•• ,.
`
`1.2
`
`'
`<IJ u
`c ••
`"' t) DO
`Q) ••
`<IJ
`0:: 11
`
`10
`
`I 4
`
`01
`
`- G)
`
`0~--~--~--~------~~====~
`480
`$»
`,.,
`tlO
`Wavelength (run)
`
`Figure 2-4 - Colour filters
`
`The curves shown in each case represent the amount of light reflected from a sample
`at each of the wavelengths indicated. The box indicates the area over which
`reflectance is measured to calculate the density of the colour.
`
`16
`
`FAST FELT 2010 I pg. 31
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`(
`
`(
`
`The amount of light from the illumination source is known (incident light). The
`
`amount of light reflected is also known. Reflectance is thus defined as;
`
`Reflectance = Reflected light
`
`Incident light
`
`Equation 2.1 -Reflectance
`
`Density is defined as the reciprocal of the reflectance, so that as the ink film thickness
`
`increases, and the reflectance decreases, the density increases. A logarithmic scale is
`
`used to improve the closeness of the results between density and ink film thickness.
`
`Thus density is defined as;
`
`Density= Log10 (!/Reflectance)
`
`Equation 2.2 - Density
`
`This work was extended by Murray [19), who defined a technique for the calculation
`
`of the size of dots. By comparing both the solid density (Ds) (the density measured at
`
`100% coverage) and the halftone density (D1) (the density measured at the coverage
`
`under investigation) a relationship between the optical density and the actual printed
`
`coverage was observed to be;
`
`Area=
`
`1-10-0
`'
`1-10- '
`0
`
`Equation 2.3- Murray Davies equation
`
`Close correlation was observed between densities measured with a densitometer and
`
`the fractional dot area (measured using an optical microscope). Comparing prints and
`
`proofs showed higher densities on the prints, caused by spreading of the ink into or
`
`over the substrate - tone gain. It was therefore demonstrated that tone gain may be
`
`17
`
`FAST FELT 2010, pg. 32
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`calculated from the optical densities.2 Yule and Nielson [20] modified the Murray(cid:173)
`Davies equation to include an empirically derived factor to compensate for light
`scattering within the substrate. This attempts to calculate a value for the physical tone
`gain observed in the sample, ie the degree of spreading observed by the ink on the
`substrate, making dots larger than was intended. This method requires a visual
`assessment of a 50% coverage dot, and as such requires a significant degree of user
`intervention and thus an additional level of variability. For the purposes of this work it
`is attempted to minimise user intervention, and to use properly defined techniques. As
`a result, this technique was not used for the calculation of physical dot gain. Image
`processing provides techniques for the calculation of physical dot gain over the range
`of tonal coverage [21].
`
`However, conventional colour measurement techniques were used, in evaluation of
`scumming, and
`these are discussed
`[ 15]. Using optical
`fully
`in Hunt,
`spectrodensitometry to measure colour, both in terms of the printed ink, and in the
`unprinted areas enabled identification of the levels of scumming between the printed
`substrate (in unprinted regions) and the bare, unprinted substrate.
`
`2.2.2 Surface Profiling and Volume measurement
`
`A gravure cylinder is an engraved, etched or ablated surface, and one focus of this
`work was to analyse and measure the engraving details on the cylinder. As such, this
`section will review the methods available for measuring surface topography, with
`particular reference to those methods that are most appropriate. [22]
`
`Many techniques are available for surface profiling, from the most complex (AFM,
`STM etc) to simply dragging a fingernail over a surface. Each method has its own
`advantages and disadvantages, but since the purpose of the use of surface profiling in
`this case is to investigate indentations, intentionally placed into a surface, rather than
`natural indentations and ridges, the deflection that must be measured may be large and
`
`2 Tone gain is defined as the calculated tonal coverage minus the specified coverage. Thus a 50"'o dot,
`measured to be of 58% coverage exhibits 8% tone gain.
`
`18
`
`FAST FELT 2010 I pg. 33
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`the sides may also be relatively steep. As a result, many of the methods traditionally
`used in engineering for surface profiling will not be applicable.
`
`(
`
`Traditionally, stylus measurement teclmiques are used in the analysis of surface
`profiles [23]. In these procedures a stylus is dragged across the surface being analysed
`whilst in contact with the measurement surface. The vertical displacement of the
`stylus is tracked, and a profile is generated. From this data, various rouglmess
`parameters may be calculated, eg. R3 , Rq, Rz, R, etc. [24]. This two-dimensional data,
`although useful for calculating the rouglmess of a surface, is of little use when
`attempting to calculate volumes, which require three-dimensional measurements,
`although a raster scan may be performed
`to produce a
`three-dimensional
`measurement. However, this is not suitable when measuring steep cell sides as under
`these conditions the stylus sides contact the indentation, and not the tip of the stylus.
`
`Several different three-dimensional methods have been proposed for analysis of
`gravure cells, many of which due to the nature of the cells are impossible to use. Kunz
`[25] simply cut cylinders up to examine them under an optical microscope. Due to the
`cost involved in this sort of destructive analysis, little other work has been published
`using this method, although it does allow reasonably detailed analysis. Scanning
`electron microscopy (SEM) was also abandoned due to the necessary destruction of
`the cylinder to obtain suitable sample sizes.
`
`Certain stylus teclmiques are applicable to three-dimensional measurements. Bowen
`et al [26] [27] discuss the uses and limitations of atomic force microscopy. Atomic
`force microscopy (AFM) uses a very fine stylus mounted on the end of a cantilever,
`held close enough to the measurement surface that inter-atomic forces cause the stylus
`to be deflected. This deflection is measured using a laser, reflected off the gold plated
`top surface of the cantilever, and received by a position sensitive detector. This
`deflection is interpreted into a height profile of the measurement area. It is possible to
`operate an AFM in several modes, including contact (where the stylus is held in
`contact with the surface), tapping (where the stylus is briefly brought into contact at
`each measurement point, but is not dragged between them) or non-contact (where the
`atomic forces are measured to keep the stylus and the substrate separate). The stylus is
`moved across the surface of the substrate in a series of linear measurements, which
`
`19
`
`FAST FELT 2010 , pg. 34
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`are all taken to produce a three-dimensional image. It is however, only applicable to
`
`maximum height deviations of±IO!-lm, due to the length of the stylus protrusion from
`
`the cantilever and thus not usable to examine cells, which may be up to I 00!-lm deep.
`
`Scanning tunnelling microscopy (STM) is another probe I stylus type measurement,
`
`where the probe is maintained close to the substrate, with either a constant 'tunnelling
`
`current' being passed between the stylus and the substrate (which must be electrically
`
`conductive) or with the stylus height being maintained, and the current being
`
`measured to infer the distance between the stylus and the substrate. This was excluded
`
`for similar reasons [28], and particularly time- it will require approximately 1 Y2 years
`
`to measure the area of a single large cell, although this will allow calculation of the
`
`position of each atom in the outer surface of the cylinder chrome surface.
`
`White light interferometry [24] is a recent development, and allows a non-contact
`
`examination to be performed. The principles of interferometry are well established,
`
`but the use of a white light source giving a much longer coherence length and the use
`
`of an L VDT controlled head movement allow much larger scan ranges, and with
`
`minor modifications to the optical systems (detailed in chapter 3.2) allow detailed
`
`analysis of the top surface and insides of gravure cells, including measurement of the
`
`depth, volume, open area, width, and length of the cells.
`
`In white light interferometry a beam of white light is split by a beamsplitter, with half
`
`being passed to a reference mirror, and half being passed to the substrate. These
`
`beams are then recombined, and the resultant interference pattern is captured by a
`
`CCD camera, Figure 2-5
`
`20
`
`FAST FELT 2010, pg. 35
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`CCD
`
`c
`
`A
`
`Illumination
`
`source
`
`Figure 2-5 - Interferometer
`
`Light comes from the iJiurnination source, (A) (normally a standard tungsten-halogen
`
`bulb) and is divided into two beams by the beamsplitter, (B). Half the light is sent to
`
`the reference mirror, and half to the substrate. The reflected light from the mirror and
`
`the substrate is then recombined, and passed to the CCD array, (C). The whole
`
`apparatus is then slowly moved closer to the substrate, and images are regularly
`
`captured for analysis by computer.
`
`It is possible to operate most interferometers in two modes, VSI (Vertical Scanning
`
`Interferometry) and PSI (Phase Shifting Interferometry). PSI is the older technique,
`having been used in interferometers since the early 1980s. It uses a series of
`
`measurements, where the apparatus is moved only Y4 of the wavelength of the light
`
`being used between each measurement. From this measurement, the relative height of
`
`each pixel is calculated. Due to this limited number of measurements, the speed of
`
`measurement is significantly higher than when taking VSr measurements. The
`
`limitation of these measurements is in terms of the maximum height differences
`
`which can be measured, 160nrn in PSI mode, SOOJliil in VSI mode. Vertical resolution
`is improved in PSI mode, typically 3A while with VSI measurements a resolution of
`1 run can be achieved with care. The latter is very appropriate in the current
`
`application and so VSI measurements were used throughout this investigation.
`
`21
`
`FAST FELT 2010, pg. 36
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`2.2.3
`
`Ink transfer measurement
`
`Since colour is a function of ink deposit, which is itself a function of print quality it is
`
`necessary to analyse the quantity of ink being transferred. Deposit is determined by
`
`transfer mechanisms and motivates this review.
`
`Kistler and Schweizer [29] describe several teclmiques for film thickness I ink transfer
`
`measurement. Absorption techniques, either using infrared or microwaves may be
`
`used to examine the thickness of transferred films, whereby the absorption of either
`
`infrared light or microwaves may be measured, and a thickness may be inferred from
`
`the consequent attenuation. This teclmique however, requires greater film thicknesses
`
`than those observed in the gravure process due to the accuracy of the system, and the
`
`need for enough of the film to be present to absorb the infrared/microwaves. X-ray
`
`fluorescence measures the emission of X-rays from suitable elements. Disadvantages
`
`of this method include extremely long sampling times, which are required to acquire
`
`accurate mean values, and the necessary use of unrealistic inks containing tracer
`
`elements for testing purposes.
`
`Capacitance gauging is also occasionally used for analysis of thin films. By placing a
`
`flat plate sensor above a grounded target plate, with the substrate between them, the
`
`capacitance of the parallel plates can be measured, and the thickness inferred from the
`
`change in capacitance between the unprinted substrate and a printed substrate with a
`
`known film thickness applied. It is however, impractical to use this teclmique with
`
`films of the thickness used in the gravure process due to the low precision of the
`
`teclmique. Each of these three teclmiques is only able to be used with even films, such
`
`as are obtained at 100% coverage. They are also only able to be used over relatively
`large areas, of the order of a few mm2 at least. Due to the need to measure much
`
`smaller areas for the analysis of individual dots, these methods were not usable.
`
`The Moire technique (also sometimes referred to as 'mechanical interferometry') uses
`
`interference techniques to measure film thickness, however it is limited to use on
`
`films of greater thickness than lO)lm, and only has a vertical resolution of IOJ.lm. By
`
`22
`
`c
`
`(
`
`(
`
`(
`
`FAST FELT 2010, pg. 37
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`superimposing two optical gratings, one undistorted, and one distorted by the film, it
`is possible to infer the thickness of the film. Again, this is only of use with
`significantly larger areas than are available and significantly thicker film deposits.
`
`(
`
`'Profilometers are typically not used for film thickness measurements' ([29] p226,
`7.3.2.1) because traditional profilometers use a stylus that is dragged across the
`surface of a sample. Optical profilometers however, are extremely suited to the
`measurement of films. Two primary techniques are available. The triangulation
`method uses a laser beam, focused on the surface at test and viewed at a fixed angle
`by a position detector, similar to the technique used for height displacement
`measurement as used in an atomic force microscope. As the height of the surface
`changes (i.e. as the thickness of the transferred film changes) the light spot on the
`position detector moves, thus giving a measurement of displacement of the test
`surface, relative to the detector. The method can use extremely high sampling rates
`(up to 40kHz), high resolution, large measurement ranges (up to I Omm) and a
`relatively high working distance (up to Scm). Unfortunately it is only practical to use
`this on thicker films than are transferred in the gravure process. White light
`interferometry [24] however provides a non-contact, non-destructive technique for the
`analysis of films. The teclmique is described in 2.2.2. By varying the optical
`configuration of the interferometer, suitable x-y measurement areas may be defined
`either for the measurement of individual dots (the gravure process not being a contone
`process) or for continuous films, whilst the vertical resolution of the interferometer
`allows for the measurement of the extremely thin films observed. This technique does
`have several drawbacks, primarily those of time and cost. Measurement of film
`thickness takes as little as 30s, but measurement of film thickness where a broken film
`occurs (for example, at the highlight end of the tonal range) takes approximately five
`minutes per sample. This is, however the only process which allows the measurement
`of such broken films (30].
`
`When calculating the quantity of ink released from the cells it is necessary to either
`know the amount of wet ink that is transferred, or to calculate the volume of dry ink,
`and the amount of solvent which is associated with this quantity of solids. Due to the
`extremely fast drying nature of the ink, these wet measurements are not possible, and
`as such dry analysis is required. A precise breakdown of the ink was supplied by the
`
`23
`
`FAST FELT 2010, pg. 38
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`manufacturer, and the quantity of solvent in the wet ink is known precisely, but the
`
`solvent content in the dry ink must still be measured. Traditionally, to identify the
`
`solvent component of a sample, gas-chromatography mass-spectrometry is used [31]
`
`[32] [33]. Thermal desorption heats the sample, driving off the volatile fraction into a
`moving, inert gas stream, and condenses it onto a cold surface [34] [35]. This surface
`
`is then rapidly heated; releasing the volatiles back into the gas stream, and passed
`
`onto the gas chromatograph - mass spectrometer. Clearly the thermal desorber is
`required to ensure that all the volatile components are injected into the gas
`
`chromatograph at the same time, rather than over a period of several seconds if the
`
`entire sample had to be heated. The sample is then passed through a column, which
`separates out the components. This is then passed on to the mass spectrometer, to
`
`identify the components that are included [31] [36). The mass spectrometer uses
`ionisation techniques to break up the component molecules, and then analyses the
`
`components by comparing the break-up pattern with those of known substances. This
`
`is performed automatically. Once the quantity of solvents in the sample has been
`measured and the volume of the dry dots has been calculated, it is possible to
`
`calculated the quantity of solvent originally related to the solids, and thus the volume
`of wet ink which was released from the cell.
`
`Gas chromatography has been used as an analytical tool since approximately 1950,
`although experimental tools have been used since 1940. The sample under test is
`
`injected into a moving stream of gas, which passes through an oven. Inside the oven is
`
`a long coiled tube, which the gas passes through. By the end of the tube the
`components have been separated by elution. They are distinguished by the different
`
`times they take to pass through the column - the retention time.
`
`Finally, the sample is passed to the mass spectrometer. This is the oldest part of the
`
`system, with mass spectrometry having been used in some form since 1898, and in
`
`conjunction with gas chromatographs since 1958, shortly after its introduction. A
`
`mass spectrometer takes the input molecules and bombards them with electrons to
`
`make ions, which spontaneously break up. This break-up pattern is specific to each
`molecule, and so a piece of software may analyse the patterns observed, and identify
`
`the most likely chemicals from the pattern. In addition, the frequency of each ion (or
`abundance) is measured, thus allowing a detailed breakdown of the components fed to
`
`24
`
`FAST FELT 2010 I pg. 39
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`the mass spectrometer. Results are produced as a 'total ion chromatograph' (TIC),
`which details the frequency of each ion, and each fragment identified from the break(cid:173)
`up of the molecules. Automated analysis follows, identifying each component in the
`sample, and its abundance from the TIC.
`
`There is no published literature indicating that this technique may be used on printed
`polymer films for this specific purpose. Similar trials have however, been performed,
`for example Garcia and Sanz [37] who used the procedure to analyse the volatile
`content of Origanum Vulgaris leaves. For the purposes of this research, a new
`technique was developed in conjunction with the manufacturers of the thennal
`desorber - gas chromatograph - mass spectrometer in use. [38]
`
`2.3 Experimental Results
`
`To consider the mechanisms of ink release from gravure cylinders, the review of
`previous experimental work relating to rotogravure printing is discussed in three
`sections. The first details the effects of altering inks, the second of changing
`substrates, and the third examines the effects of cylinder engraving.
`
`2.3.1 Inks
`
`Rotogravure inks are predominately solvent based. They consist of a series of
`components intended to adhere to the substrate, and are approximately 80% solvent
`by mass. Much of the rest of the ink is made up of pigment and binder. Many things
`may be added to the basic solvent blend, including adhesion promoter (itself another
`solvent), nitrocellulose or polyvinyl butyral, surfactants and many other chemicals
`specific to an
`individual
`ink. Changing any one of these components may
`significantly affect the ink, and thus its transfer.
`
`Bohan et al [39) investigated the effects of changing ink viscosity, demonstrating that
`ink viscosity has a large influence on the colour of the final printed copy with A.E
`values of up to six being recorded. These changes are significantly larger than those
`
`25
`
`FAST FELT 2010 I pg. 40
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`necessary to be detected visually, and point to the importance of ink viscosity It was
`
`found that subtle changes in viscosity at low viscosities are far more significant than
`
`(
`
`changes at higher viscosities. These experiments were also repeated, with the same
`
`findings in Bohan et at [40]. Kunz [25] examined the effects of viscosity on print, and
`
`observes that tone gain is significantly increased at low viscosities, while it is reduced
`
`at high viscosities. He also notes the effect of changing viscosity on the filling state of
`
`the cells, where high viscosity inks do not fill gravure cells as completely as low
`
`viscosity inks. The analysis technique used for this is not entirely clear, but it is
`
`known that he used an early interference technique to calculate the deflection between
`
`the top surface of the ink and the top surface of the cylinder.
`
`Further mention is made of ink temperature, and its effect on ink transfer, although
`
`this is primarily due to temperature affecting the viscosity of the inks under study. It
`
`should be noted that the inks Kunz used are of a significantly higher viscosity than
`
`those which would be used on a press today (36s, Zahn 2 - normal operating
`
`viscosities are of the order of 20s, Zahn 2).
`
`Elsayad et a! [ 41 ] investigated the effects of viscosity and resin content, discovering
`
`that the amount of ink transferred by the cylinder increased as the viscosity of the ink
`increased. It was also found that the type of ink resin used had a considerable effect
`
`on the amount of ink transferred. Benkreira et a! [29] [ 42] [ 43] [44) also investigated
`
`the effects of ink, indicating that it has little effect on the release.
`
`Pekarovicova et al [ 45] investigated the use of hot melt inks for gravure printing, but
`
`as their research was limited to inks which at best were several hundred times more
`
`viscous than gravure inks, not solvent based, and limited to draw-down testing of
`
`them, this has little relevance to the gravure process.
`
`2.3.2 Substrates
`
`Gravure printing may be performed on many different substrates, from extremely thin
`
`papers, such as those used in catalogue printing, to heavy papers as used in art books.
`
`From thin packaging films (which may themselves be metallised or coated in some
`
`26
`
`FAST FELT 2010 , pg. 41
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`other fashion) to thick cartonboard as used in cigarette packs. All the work in this
`investigation was performed on plastic substrates, due to the non-porous nature of
`plastic films, but even here, by changing the type of film, the surface roughness and
`specific surface energy may change significantly. The specific surface energy relative
`to the surface tension of the ink may well need to be changed, using (for example)
`corona treatment, to enable the ink to transfer and adhere.
`
`Kunz [25] indicates that the substrate is an extremely significant variable. He
`indicates that there is a "clear connection between smoothness and printability",
`suggesting that there is an operability window for smoothness, where substrates that
`are either too rough, or too smooth will have "negative effects"3
`. It is also observed
`that for printing to be achieved, it is necessary for the impression roller to force the
`substrate into contact with the ink, and that this is "problematic" due to the need to
`squeeze the substrate into the cells, as the meniscus of the ink lowers the top surface
`of the ink below the top surface of the cylinder. However, this work was performed
`using ink significantly more viscous than that used in the modern process (34s Zahn
`2) and at significantly lower speeds than those used on commercial presses.
`
`LaFaye et al (46] produced different papers and printed them on test equipment (some
`were also printed on commercial equipment) to test their printability. They show that
`both solids content and coating weight (both of which promote surface smoothness)
`have a positive effect on printability. This correlates with (25], where it is shown that
`generally smoother substrates print better. Heintze [ 47] also investigated the
`smootlmess of coated board on a GRI printability tester. He discovered that the wear
`of the doctor blade can affect the print quality, particularly in terms of the 'speckle' of
`the print. He acknowledged that the use of a lamella blade minimises this problem.
`
`Hansson and Johansson [ 48] [ 49] examined the printability of different substrates
`using stereo photometric techniques, using irradiance to infer a measurement of
`surface height and roughness. They also determined that areas of substrate
`significantly lower than the average surface tend to have missing dots. This method is
`
`3 The comparison is given to water on glass, which beads up and does not adhere
`
`27
`
`FAST FELT 2010 I pg. 42
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`however, not as accurate as white light interferometry, which was used in the conduct
`ofthis investigation.
`
`2.3.3 Cylinder analysis
`
`Tucker [50] examined the use of interferometry to evaluate anilox rolls, but since an
`anilox roll is an intaglio roller, rotating in an ink bath, the principles of measurement
`hold for measurement of gravure cylinders. Interferometry was considered to be a
`breakthrough in measurement technology, replacing the traditional methods of
`theoretical calculation, and
`liquid volume measurement. Theoretical volume
`measurement is the standard technique where an optical microscope is used to
`calculate the open area at the roller surface, and to infer the volume from these
`physical dimensions. It is limited by the operator's identification of edges, and the
`assumptions that must be made about the shape of the cells, which, for the purposes of
`calculation must be assumed to be ideal, despite the limitations which this imposes.
`Liquid volume measurement is an indirect technique, by which a known volume of
`liquid is applied to the cylinder, doctored, and then 'printed' onto a substrate. The
`area printed can then be calculated, and the volume inferred from this measurement.
`This again, is limited by the operator as the doctoring process is likely to influence the
`area over which the liquid is spread and calculation of the printed area is dependent
`on the users interpretation [51]. It also assumes that the cells are completely filled,
`which has already been shown by Kunz [25] not to be true.
`
`2.3.4
`
`Ink Transfer
`
`Several variables are believed to influence ink transfer. Ofthese, only engraving type
`and screen ruling have been investigated. This section reviews the work that has been
`performed in this area. This section also includes scumming as this can be considered
`as unwanted ink transfer.
`
`Little work has been performed in evaluating the different forms of engraving in
`terms of the ink release. Kunz [25] however completed the earliest work, and he
`
`28
`
`FAST FELT 2010 I pg. 43
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`examined the filled and emptied states of four different types of cells, establishing that
`
`laser engraved cells release the most ink (88%), followed by two types of etching,
`round dot (80%) and conventional (75%), with electromechanically engraved cells
`
`releasing the least ink (68%). This he attributed to the shape of laser engraved cells,
`
`which have very smooth bases and lack the pointed apex of electromechanically
`
`engraved cells. This was, again, established by the use of an early interferometric
`
`microscope, but under conditions far removed from the modem industrial process.
`
`Pulkrabek and Munter [44] examined the release of ink from a knurled roller,
`
`analogous to a gravure cylinder. Their work indicates that 59% of the volume of the
`ink in the roller grooves is transferred to a substrate, but that the release is inversely
`
`proportional to the fluid viscosity to "a small degree". It should be noted however,
`
`that their ink was highly shear thinning, with a rest viscosity of 5000cP
`(approximately 20,000s - Zahn 2) - or approximately 1 OOOx more viscous than
`
`normal process gravure ink, and a thinned viscosity of approximately lOOcP (43s (cid:173)
`Zahn 2) - or approximately twice as viscous as normal gravure ink. No indication is
`given of how the portion released was measured.
`
`Benkreira and Patel [ 42] measured transfer using infrared absorption techniques in
`
`reverse gravure coating. It was discovered that the amount of ink released from a cell
`
`appears to be consistent at approximately 32% of the transferring cell volume. By
`
`using three different types of cell engraving, it was also established that the release is
`independent of the type of cell, except at very low speeds, where variation was
`
`observed. Benkreira et al. [43] measured the transfer from both loaded and unloaded
`
`forward gravure cylinders, establishing that unloaded cylinders release approximately
`20% of their volume, but that loaded cylinders release up to 33% of their volume. It
`
`should be noted that no indication of the impression pressure was given, despite
`
`significant evidence that the impression pressure will affect the release from the
`
`cylinder. Benkreira and Cohu (52] continued this work, demonstrating that with
`
`unloaded forward gravure coating release is of the order of 15-20% of cell volume.
`
`This work also demonstrates that the thickness depends on operating parameters,
`
`including speed and cell geometries. It should be noted that all this work was carried
`
`out on test rigs, not on production scale presses. It should also be noted that the
`infrared analysis equipment used in all three of these studies provides an accuracy of
`
`29
`
`FAST FELT 2010 I pg. 44
`Owens Corning v. Fast Felt
`IPR2015-00650
`
`

`
`(
`
`±0.5um - or approximately 10% of the thinner film thickness measured. The
`technique used for analysis relies on infrared absorption over a large area, and
`averaging over a period of time. As a result, the technique is not applicable to the
`measurements being performed here. Furthermore it should be noted that all this work
`has only been performed on solid coverages, and not the halftones being examined in
`this work.
`
`Jeske [53] simulated the gravure process with an impulse driven press, where a loaded
`mass forced the substrate onto the printing plate. Although this is similar to the
`earliest gravure printing techniques, several problems occur with this technique.
`Firstly, the printing plate forces the substrate into a completely different contact
`regime than a cylinder. Secondly, the plate was doctored by hand, despite Kunz's [25]
`and others assertion that doctoring can, and will affect the filling state, and thus the
`release characteristics of the cells. Despite the criticism that "unrealistic inks" were
`employed in the research of other authors, Jeske used a highly viscous ink, mixed
`with oils to reduce the viscosity, rather than solvents. No indication is given of the
`size of the measurement area, but analysis of transfer was calculated by the
`differential mass method, the accuracy of which is low, particularly for practical
`conditions. It is also known that the surface tension of the ink affects release to some
`extent, as the ink is 'sucked' onto the substrate (Kunz [25]) yet the inks used here,
`have a significantly higher surface tension than inks made with solvents. It is also
`demonstrated here, through the use of photomicrographs that the cells are not
`completely filled, although the technique used to demonstrate the lack of filling does
`not allow quantification of the fill state.
`
`Kapur, Gaskell and Bates [54] investigated the release from a gravure cylinder in an
`offset gravure coating test rig. Tests

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket