throbber
SPINE Volume 30, Number 5, pp E118 –E122
`©2005, Lippincott Williams & Wilkins, Inc.
`
`Clinical Measurements of Cement Injection Pressure
`During Vertebroplasty
`
`Jo¨rg Krebs, DVM,* Stephen J. Ferguson, PhD,* Marc Bohner, PhD,† Gamal Baroud, PhD,‡
`Thomas Steffen, PhD, MD,§ and Paul F. Heini, MD储
`
`Study Design. Clinical study of injection pressure dur-
`ing vertebroplasty.
`Objectives. To investigate the range of injection pres-
`sures during conventional vertebroplasty interventions
`and to study the influence of syringe design and cement
`polymerization time on injection pressure.
`Summary of Background Data. Vertebroplasty is an
`efficient procedure for the treatment of painful vertebral
`fractures. However, cement leakage is a potentially seri-
`ous complication. Although injection pressure has been
`suggested as a factor for extravasation risk, to date, there
`are only anecdotal reports of pressure data for cement
`augmentation in the clinic.
`Methods. Using a syringe holder instrumented with
`load and displacement transducers,
`injection pressure
`and volume were recorded in vivo during conventional
`cement augmentation. Wide (3 mm opening) and normal
`(1.8 mm opening) syringes were alternated such that each
`type was evaluated for early (300 –500 seconds postmix-
`ing) and late (⬎500 seconds postmixing) cement poly-
`merization time. The influence of syringe type and poly-
`merization time on injection pressure was evaluated
`using a multifactorial analysis of variance followed by
`Scheffe´ post hoc comparison.
`Results. The maximum peak injection pressure mea-
`sured was 3215 kPa. The average pressure peaks for nor-
`mal and wide syringes were 1693 ⫾ 653 kPa and 1727 ⫾
`597 kPa, respectively. No statistically significant differ-
`ences were found between injection pressures with wide
`and normal syringes. Higher injection pressures were ob-
`served later in the polymerization process.
`Conclusions. High injection pressures approaching 20
`atmospheres are reached during conventional vertebro-
`plasty. Widening the syringe tip diameter did not signifi-
`cantly change injection pressures, whereas elapsed time
`did. Further research is needed to improve injection
`equipment and materials for vertebroplasty.
`Key words: vertebroplasty, osteoporotic compression
`fractures, polymethyl methacrylate,
`injection pressure.
`Spine 2005;30:E118 –E122
`
`From the *M. E. Mu¨ ller Institute for Surgical Technology and Biome-
`chanics, Bern, Switzerland, the †Robert Mathys Stiftung, Bettlach,
`Switzerland, ‡Laboratoire de Biomecanique, De´partement de Ge´nie
`Me´canique, Universite´ de Sherbrooke, Sherbrooke, Que´bec, Canada,
`§Orthopaedic Research Laboratory, McGill University, Montreal,
`Canada, and the 储Department of Orthopaedic Surgery, Inselspital,
`Bern, Switzerland.
`Supported by AO Spine International, Du¨ bendorf, Switzerland.
`Acknowledgment date: March 9, 2004. First revision date: June 30,
`2004. Acceptance date: July 23, 2004.
`The manuscript submitted does not contain information about medical
`device(s)/drug(s).
`Foundation funds were received in support of this work. No benefits in
`any form have been or will be received from a commercial party related
`directly or indirectly to the subject of this manuscript.
`Address correspondence and reprint requests to Dr. Jorg Krebs, MEM
`Research Center for Orthopaedic Surgery, Institute for Surgical Tech-
`nology and Biomechanics, University of Bern, Stauffacherstr. 78, 3014
`Bern, Switzerland; E-mail: jorg.krebs@MEMcenter.unibe.ch
`
`E118
`
`Vertebroplasty, the percutaneous augmentation of ver-
`tebral bodies with polymethyl methacrylate (PMMA), is
`reported to be an effective surgical procedure for provid-
`ing pain relief for patients suffering from osteoporotic
`compression fractures.1–3 The technique is therefore
`gaining increasing popularity with both physicians and
`patients.4 The reported rate of serious complications is
`low, but there is the concern that complications are un-
`derreported.5 The main complication is cement leakage
`into adjacent structures,3,6,7 which can lead to spinal
`cord or nerve root compression1,8,9 or pulmonary embo-
`lism.10 –12 The consequences of these leaks may be se-
`vere, i.e., paraplegia13 or death.14,15
`The reported incidence of cement leakage is approxi-
`mately 20% to 71% of augmented osteoporotic verte-
`bral bodies.1,2,6,16 –18 Cement leakage has occurred into
`paravertebral tissue, neural foramens, epidural space,
`and perivertebral veins. The risk of cement leakage has
`been observed to increase with the volume injected.6 Fur-
`thermore, the incidence of leakage was higher when lev-
`els above T7 were augmented. Other factors determining
`leakage risk include the choice of injection equipment,
`injection pressure,6,19,20 and cement viscosity.21,22 Al-
`though injection pressure has been suggested to affect
`extravasation, to date, only pressure data of isolated
`cases have been reported for cement augmentation in the
`clinic.23
`To better understand the vertebroplasty process and
`to facilitate the design of improved injection equipment
`and augmentation materials, the primary goal of this
`study was to determine the range of injection pressures
`during conventional vertebroplasty interventions. The
`secondary goal of the study was to determine the influ-
`ence of syringe design and cement polymerization time
`on injection pressure.
`
`Methods
`Patient Population. Twenty patients (17 female, 3 male; aged
`54 – 84 years, mean 71.7) were randomly allocated for injection
`pressure measurements. Candidates were selected from pa-
`tients admitted to the Orthopedic Department of the Inselspital
`Bern for vertebroplasty due to osteoporotic vertebral compres-
`sion fractures. Patients selected for pressure measurements had
`sustained between 1 and 3 (mean 1.67) vertebral compression
`fractures.
`
`Surgical Technique. Vertebroplasty was carried out accord-
`ing to the technique described by Heini et al.2 Unipedicular
`augmentation of vertebral bodies (3–7 vertebral bodies per pa-
`tient, mean 4.77) was performed using standard 2 mL (no mea-
`surements) or 5 mL (pressure measurements) syringes contain-
`
`STRYKER EXHIBIT 1012, pg. 1
`
`STRYKER CORPORATION v. ORTHOPHOENIX, LLC
`
`IPR2014-01433
`
`

`

`Cement Injection Pressure • Krebs et al
`
`E119
`
`Figure 1. Syringe holder instrumented with load and displacement
`transducers (arrow) connected to the data acquisition system and
`palm computer.
`
`ing Vertebroplastic™ cement (DePuy AcroMed, Raynham,
`MA). Cement was injected through standard bone marrow bi-
`opsy needles (8 gauge, MD Tech, Gainesville, FL) under fluo-
`roscopic control. Cement injection was commenced following
`visual appraisal of cement viscosity. Cement injection was per-
`formed until a satisfactory vertebral body filling was achieved
`or until limited by the viscosity changes throughout the poly-
`merization process or by any signs of extravasation. Not only
`collapsed but also intact adjacent vertebral bodies were aug-
`mented.2,24 –27
`
`Measurements. Using a syringe holder instrumented with
`load and displacement transducers (Figure 1), the injection
`pressure and cement injection volume were recorded in vivo
`during conventional manual cement augmentation of 50 osteo-
`porotic, nonfractured vertebral bodies (T7–L5). For most ver-
`tebral bodies, several measurements were taken, with pauses
`due to leakage (Figure 2) or changing of syringes. Data were
`collected at a frequency of 4 Hz with a data acquisition system
`(MyCorder, DAS-1206, Datastick Systems, Santa Clara, CA)
`and stored on a Palm handheld computer (Palm III, Palm
`Corp., Milpitas, CA). Data were later transferred to a PC.
`
`Figure 3. Injection pressure (circles) and volume injected (solid
`line) during PMMA augmentation of a vertebral body.
`
`The influence of the syringe tip diameter was evaluated by
`expanding the standard 5 mL syringe tip (diameter 1.8 mm) to
`a diameter of 3 mm using a sterile drill before filling. In each
`patient, injection pressures during the augmentation of at least
`2 vertebral bodies were measured. During the augmentation of
`each vertebral body, we attempted to take measurements for
`both syringe types. However, this was not possible in all cases
`due to the clinical situation (e.g., termination due to extrava-
`sation, complete filling, technical error). As cement viscosity
`increases with polymerization time, wide and normal syringes
`were alternated such that each type was equally evaluated for
`the early (300 –500 seconds postmixing) and late (⬎500 sec-
`onds postmixing) polymerization phase.
`Statistical Analysis. Based on previously determined calibra-
`tion curves, force and displacement measurements were con-
`verted to maximum pressure and total injection volume. Data
`were calculated and presented as mean ⫾ standard deviation of
`the mean. The influence of syringe type (wide and normal),
`polymerization phase (early and late), and flow rate on injec-
`tion pressure was evaluated using a multifactorial analysis of
`variance (ANOVA) followed by Scheffe´ post hoc comparison,
`with a significance level of P ⫽ 0.05.
`
`Results
`
`Representative curves for the injection of 1 syringe into a
`vertebral body are presented in Figure 3. In a typical
`injection process, a fairly constant injection pressure is
`
`Figure 2. Fluoroscopic images
`(left lateral view/right anteropos-
`terior view) showing extravasa-
`tion of cement (arrow).
`
`STRYKER EXHIBIT 1012, pg. 2
`
`

`

`E120 Spine • Volume 30 • Number 5 • 2005
`
`Table 1. Injection Pressure Data of Normal and Wide
`Syringes During Vertebroplasty
`
`Polymerization Time
`
`Phase 1 (300–500 s)
`
`Phase 2 (501–800 s)
`
`Average
`Pressure
`(kPa)
`
`1391 ⫾ 520
`1394 ⫾ 388
`1943 ⫾ 654
`1968 ⫾ 622
`
`Peak
`Pressure
`(kPa)
`
`2493
`2142
`3215
`3178
`
`Syringe
`
`Normal
`Wide
`Normal
`Wide
`
`N
`
`24
`23
`29
`27
`
`The values are given as the mean ⫾ the standard deviation.
`N ⫽ number of measurements.
`
`produced using a hand-held syringe, and a constant flow
`rate of cement is obtained. Resting periods during the
`injection can be recognized. Injections were performed at
`an average flow rate of 0.09 ⫾ 0.04 mL/s (range 0.03–
`0.2 mL/s). There was no correlation between flow rate
`and injection pressure. During each measuring period,
`an average of 1.9 ⫾ 0.9 mL cement was injected. The
`average elapsed time after mixing during the early (300 –
`500 seconds) and later (501– 800 seconds) polymeriza-
`tion phase was 450 ⫾ 43 seconds and 616 ⫾ 89 seconds,
`respectively.
`Injection pressure data are summarized in Table 1.
`The average pressure peaks for all measurements (n ⫽
`103) was 1698 ⫾ 624 kPa, and the maximum injection
`pressure recorded during the study was 3215 kPa, which
`was measured with a normal syringe. The highest pres-
`sure recorded with wide syringes was 2178 kPa. The
`average pressure peak for measurements of normal (n ⫽
`53) and wide (n ⫽ 50) syringes were 1693 ⫾ 653 kPa and
`1704 ⫾ 597 kPa, respectively. No statistically significant
`differences were found between injection pressures of
`wide and normal syringes. Higher injection pressures
`were observed later in the polymerization process (P ⬍
`0.001). For injections during the early and late phase of
`cement polymerization, the average pressure peaks were
`1392 ⫾ 455 kPa and 1955 ⫾ 633 kPa, respectively.
`In 2 atypical cases, only a limited volume of cement
`could be injected (⬍2 mL) despite high injection pres-
`sures exceeding 2000 kPa.
`Discussion
`
`During conventional vertebroplasty injection, pressure
`peaks approached, on average, 2000 kPa (approximately
`20 atmospheres). Pressure peaks as high as 3215 kPa
`were observed. Absolute injection pressure peaks for
`wide syringes were lower compared to normal syringes;
`however, the differences were not statistically significant.
`There was no difference in average pressure peaks be-
`tween normal and wide syringes. Injection pressure in-
`creased with polymerization time.
`Until now, only injection pressure data of isolated
`cases were available for vertebroplasty in the clinical sit-
`uation.23 In the present study, injection pressure was
`equivalent to approximately 40 kg being applied onto
`the 5 mL syringe plunger (area: ⬃200 mm2). For the 2
`
`mL syringe (plunger area: ⬃130 mm2), which is rou-
`tinely used for vertebroplasty, the measured pressure
`would be equivalent to approximately 26 kg. This force
`is necessary to push the cement through the syringe and
`the cannula into the bone. Injection pressure data pro-
`vided by the present study will aid in the development of
`automated injection equipment or devices for vertebro-
`plasty that employ a mechanical advantage. However, it
`remains to be determined how well injection pressure
`correlates with intravertebral pressure. It was not feasi-
`ble in the present study to measure intravertebral pres-
`sure. Baroud et al28 reported that intravertebral pressure
`was 2 orders of magnitude lower than injection pressure
`in vitro due to the friction between the cement and the
`cannula wall. Using this ratio, intravertebral pressures in
`the present study may have approached 150 mm Hg. An
`experimental study of vertebroplasty in nonosteoporotic
`sheep reported intravertebral pressures averaging 400
`mm Hg during cement injection into intact nonosteopo-
`rotic vertebrae using conventional 3 mL syringes without
`cannulas.30
`It has been suggested that the risk of leakage increases
`with high injection pressures and consequently increased
`intravertebral pressures using current cements.6,20 How-
`ever, there may be no correlation between extravasation
`risk and intravertebral pressure. Extravasation is mostly
`determined by the viscosity of the cement and the perme-
`ability of the vertebral body at and around the injection
`point.21 High injection pressures and consequently in-
`creased intravertebral pressures during vertebroplasty
`may also increase the extravasation of bone marrow con-
`tents, leading to embolization in the lungs20,31 as re-
`ported during arthroplasty.32–34 In arthroplasty, drilling
`a vent hole with or without additional vacuum has been
`shown to allay intraosseous pressure increase and there-
`fore embolic events.32–35 During vertebroplasty, a bi-
`opsy needle in the contralateral pedicle may be used to
`decompress the vertebral body during injection.29,31 The
`preformation of a cavity for injection has also been sug-
`gested to reduce the required injection pressure20; how-
`ever, laboratory testing and computational models have
`demonstrated that the bottleneck for PMMA cement in-
`jection is the geometry of conventional syringes and can-
`nulas used for injection, rather than the inherent perme-
`ability of the vertebral body itself.36,37 It appears that the
`best way of decreasing the risk of cement leakage is to use
`high viscosity cement.21,22 The injection of high viscosity
`cement will involve higher injection pressures. Sugges-
`tions have been made to use screw-driven injection sys-
`tems to mechanically exert the extremely high pressures
`required for cement injection.19 Due to the inherent in-
`ertia of the cement and the deformation of the injection
`system, however, the cement flow cannot be immediately
`halted.21 The use of automated injection equipment or
`devices that employ a mechanical advantage should
`therefore be reviewed with caution. However, the in-
`jection of high viscosity cement may increase the ex-
`travasation of bone marrow contents (i.e., fat) from
`
`STRYKER EXHIBIT 1012, pg. 3
`
`

`

`the vertebral bodies.22 Secondly, there is a clear need
`for injectable materials that maintain a constant and
`optimal viscosity throughout the duration of the injec-
`tion process.
`No significant differences between wide and normal
`syringes could be demonstrated in the present series of
`measurements. This is in contrast to laboratory experi-
`ments that have demonstrated a trend towards lower
`injection pressures with wider syringes (Bohner et al,
`unpublished data). In these experiments, the inner part
`of the Luer lock of a 5 mLsyringe was removed by
`drilling a 4.5-mm hole into the frontal part. Using the
`modified syringe, the cement flowed directly from the
`syringe into a 7-gauge (⬇4.5 mm; present study: 8
`gauge ⬇ 4.1 mm) cannula, whereas with the nonmodi-
`fied syringe, it first flowed into the syringe tip (diameter
`⬃2 mm) and then into the cannula. The difference in
`injection pressure between modified and nonmodified sy-
`ringes was significant at P ⬍ 0.03. In the present study,
`the difference in diameter between the normal and wide
`syringe (difference: 1.2 mm) may have been too small to
`have a substantial influence on injection resistance and
`thus pressure. Different factors influence cement viscos-
`ity and thus injection pressure. Laboratory experiments
`have shown that time after mixing had the largest effect
`on the injection force (Bohner et al, unpublished data).
`The present study therefore analyzed injection pressure
`according to elapsed time after mixing. However, cement
`viscosity is also influenced by the ambient temperature,
`cement composition, liquid to powder ratio, and mixing
`time.38,39 Cement was used and mixed according to the
`manufacturer’s instructions. However, the mixing pro-
`cess (intensity and duration) in the clinical situation is
`variable. Baroud et al40 have shown that the variability
`in the viscosity for vertebroplastic cement is 30% higher
`after manual mixing compared to oscillatory mixing.
`Furthermore, the ambient temperature was not con-
`trolled in the present study. This may have masked a
`potential difference in injection pressure between normal
`and wide syringes. It was not feasible to measure cement
`viscosity in the present study.
`In conclusion, high injection pressures (approaching
`20 atmospheres) are reached during manual augmenta-
`tion of vertebral bodies with cement. Widening the sy-
`ringe tip diameter did not significantly influence injection
`pressures, whereas polymerization time did. There is a
`clear need to improve injection equipment and injectable
`materials for vertebroplasty to allow better control of the
`injection process and to minimize the risk of cement ex-
`travasation. Widening the cannula may decrease injec-
`tion pressure. Injection devices can provide the force
`needed to injection high viscosity cement; however, it is
`imperative that the operator is able to stop the advance
`of cement immediately at impending leakage. Injectable
`materials should maintain a constant and optimal viscos-
`ity throughout the duration of the injection process.
`
`Cement Injection Pressure • Krebs et al
`
`E121
`
`Key Points
`
`● Peak injection pressures during conventional
`vertebroplasty approached on average 2000 kPa
`(20 atmospheres).
`● The syringe tip diameter did not significantly af-
`fect injection pressure.
`● Injection pressure increased with cement poly-
`merization time.
`● Further research is needed to improve injection
`equipment and injectable materials for
`vertebroplasty.
`
`References
`
`1. Grados F, Depriester C, Cayrolle G, et al. Long-term observations of verte-
`bral osteoporotic fractures treated by percutaneous vertebroplasty. Rheuma-
`tology (Oxford) 2000;39:1410 – 4.
`2. Heini PF, Walchli B, Berlemann U. Percutaneous transpedicular vertebro-
`plasty with PMMA: operative technique and early results. A prospective
`study for the treatment of osteoporotic compression fractures. Eur Spine J
`2000;9:445–50.
`3. Zoarski GH, Snow P, Olan WJ, et al. Percutaneous vertebroplasty for osteo-
`porotic compression fractures: quantitative prospective evaluation of long-
`term outcomes. J Vasc Interv Radiol 2002;13:139 – 48.
`4. Murphy KJ, Lin DD. Vertebroplasty: a simple solution to a difficult problem.
`J Clin Densitom 2001;4:189 –97.
`5. Hardouin P, Grados F, Cotten A, et al. Should percutaneous vertebroplasty
`be used to treat osteoporotic fractures? An update. Joint Bone Spine 2001;
`68:216 –21.
`6. Ryu KS, Park CK, Kim MC, et al. Dose-dependent epidural leakage of poly-
`methylmethacrylate after percutaneous vertebroplasty in patients with osteo-
`porotic vertebral compression fractures. J Neurosurg 2002;96:56 – 61.
`7. Amar AP, Larsen DW, Esnaashari N, et al. Percutaneous transpedicular
`polymethylmethacrylate vertebroplasty for the treatment of spinal compres-
`sion fractures. Neurosurgery 2001;49:1105–14.
`8. Harrington KD. Major neurological complications following percutaneous
`vertebroplasty with polymethylmethacrylate: a case report. J Bone Joint Surg
`Am 2001;83:1070 –3.
`9. Ratliff J, Nguyen T, Heiss J. Root and spinal cord compression from meth-
`ylmethacrylate vertebroplasty. Spine 2001;26:E300 –2.
`10. Padovani B, Kasriel O, Brunner P, et al. Pulmonary embolism caused by
`acrylic cement: a rare complication of percutaneous vertebroplasty. AJNR
`Am J Neuroradiol 1999;20:375–7.
`11. Scroop R, Eskridge J, Britz GW. Paradoxical cerebral arterial embolization
`of cement during intraoperative vertebroplasty: case report. AJNR Am J
`Neuroradiol 2002;23:868 –70.
`12. Jang JS, Lee SH, Jung SK. Pulmonary embolism of polymethylmethacrylate
`after percutaneous vertebroplasty: a report of three cases. Spine 2002;27:
`E416 – 8.
`13. Lee BJ, Lee SR, Yoo TY. Paraplegia as a complication of percutaneous
`vertebroplasty with polymethylmethacrylate: a case report. Spine 2002;27:
`E419 –2.
`14. Childers JC, Jr. Cardiovascular collapse and death during vertebroplasty.
`Radiology 2003;228:902–3.
`15. Mathis JM, Wong W. Percutaneous vertebroplasty: technical considerations.
`J Vasc Interv Radiol 2003;14:953– 60.
`16. Yeom JS, Kim WJ, Choy WS, et al. Leakage of cement in percutaneous
`transpedicular vertebroplasty for painful osteoporotic compression frac-
`tures. J Bone Joint Surg Br 2003;85:83–9.
`17. Cortet B, Cotten A, Boutry N, et al. Percutaneous vertebroplasty in the
`treatment of osteoporotic vertebral compression fractures: an open prospec-
`tive study. J Rheumatol 1999;26:2222– 8.
`18. Hodler J, Peck D, Gilula LA. Midterm outcome after vertebroplasty: predic-
`tive value of technical and patient-related factors. Radiology 2003;227:
`662– 8.
`19. Al Assir I, Perez-Higueras A, Florensa J, et al. Percutaneous vertebroplasty: a
`special syringe for cement injection. AJNR Am J Neuroradiol 2000;21:159 –
`61.
`20. Phillips FM, Todd WF, Lieberman I, et al. An in vivo comparison of the
`potential for extravertebral cement leak after vertebroplasty and kyphop-
`lasty. Spine 2002;27:2173– 8.
`
`STRYKER EXHIBIT 1012, pg. 4
`
`

`

`E122 Spine • Volume 30 • Number 5 • 2005
`
`21. Bohner M, Gasser B, Baroud G, et al. Theoretical and experimental model to
`describe the injection of a polymethylmethacrylate cement into a porous
`structure. Biomaterials 2003;24:2721–30.
`22. Breusch S, Heisel C, Muller J, et al. Influence of cement viscosity on cement
`interdigitation and venous fat content under in vivo conditions: a bilateral
`study of 13 sheep. Acta Orthop Scand 2002;73:409 –15.
`23. Baroud G, Bohner M, Heini P, et al. Injection biomechanics of bone cements
`used in vertebroplasty. Biomed Mater Eng 2004;14:487–504.
`24. Berlemann U, Ferguson SJ, Nolte LP, et al. Adjacent vertebral failure after
`vertebroplasty. A biomechanical investigation. J Bone Joint Surg Br 2002;
`84:748 –52.
`25. Baroud G, Nemes J, Heini P, et al. Load shift of the intervertebral disc after
`a vertebroplasty: a finite-element study. Eur Spine J 2003;12:421– 6.
`26. Polikeit A, Nolte LP, Ferguson SJ. The effect of cement augmentation on the
`load transfer in an osteoporotic functional spinal unit: finite-element analy-
`sis. Spine 2003;28:991– 6.
`27. Heini PF, Orler R. Vertebroplasty in extreme osteoporosis. Experience with
`multilevel injections. Orthopade 2004;33:22–30.
`28. Baroud G, Heini P, Bohner M, et al. Drop in pressure at injection and
`infiltration in vertebroplasty. Presented at: 13th Interdisciplinary Research
`Conference on Biomaterials (GRIBOI); 2003; Baltimore, Maryland.
`29. Baroud G, Vant C, Giannitsios D, et al. Effect of vertebral shell on injection
`pressure and intra-vertebral pressure in vertebroplasty. Spine 2005;30:68 –
`74.
`30. Aebli N, Krebs J, Schwenke D, et al. Pressurization of vertebral bodies during
`vertebroplasty causes cardiovascular complications: an experimental study
`in sheep. Spine 2003;28:1513–20.
`
`31. Aebli N, Krebs J, Schwenke D, et al. Cardiovascular changes during multiple
`vertebroplasty with and without vent-hole: an experimental study in sheep.
`Spine 2003;28:1504 –11.
`32. Herndon JH, Bechtol CO, Crickenberger DP. Fat embolism during total hip
`replacement. A prospective study. J Bone Joint Surg Am 1974;56:1350 – 62.
`33. Tronzo RG, Kallos T, Wyche MQ. Elevation of intramedullary pressure
`when methylmethacrylate is inserted in total hip arthroplasty. J Bone Joint
`Surg Am 1974;56:714 – 8.
`34. Kallos T, Enis JE, Gollan F, et al. Intramedullary pressure and pulmonary
`embolism of femoral medullary contents in dogs during insertion of bone
`cement and a prosthesis. J Bone Joint Surg Am 1974;56:1363–7.
`35. Pitto RP, Koessler M, Draenert K. The John Charnley Award. Prophylaxis of
`fat and bone marrow embolism in cemented total hip arthroplasty. Clin
`Orthop 1998;355:23–34.
`36. Baroud, G, Beckman, L, Heini, P, et al. Clinical and laboratory analysis of
`the pressure at injection in a vertebroplasty. Presented at: International So-
`ciety for the Study of the Lumbar Spine (ISSLS); 2003; Vancouver, Canada.
`37. Baroud G, Wu JZ, Bohner M, et al. How to determine the permeability for
`cement infiltration of osteoporotic cancellous bone. Med Eng Phys 2003;25:
`283– 8.
`38. Farrar DF, Rose J. Rheological properties of PMMA bone cements during
`curing. Biomaterials 2001;22:3005–13.
`39. Lewis G. Properties of acrylic bone cement: state of the art review. J Biomed
`Mater Res 1997;38:155– 82.
`40. Baroud G, Matsushita C, Samara M, et al. Influence of oscillatory mixing on
`the injectability of three acrylic and two calcium-phosphate bone cements for
`vertebroplasty. J Biomed Mater Res 2004;68:105–11.
`
`STRYKER EXHIBIT 1012, pg. 5
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket