`__________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`__________
`
`STRYKER CORPORATION
`Petitioner,
`
`v.
`
`ORTHOPHOENIX, LLC,
`Patent Owner
`_________
`
`Case IPR2014-01433
`Patent 6,241,734 B1
`
`SECOND DECLARATION OF MARY E. JENSEN, M.D.
`
`
`
`STRYKER EXHIBIT 1041, pg. i
`
`STRYKER CORPORATION v. ORTHOPHOENIX, LLC
`
`IPR2014-01433
`
`
`
`Second Declaration of Mary E. Jensen, M.D.
`
`
`
`1.
`
`I previously provided a declaration for the above-captioned inter
`
`partes review, entitled Declaration of Mary E. Jensen, M.D. (Stryker Exhibit
`
`1002, “First Declaration”), relating to U.S. Patent No. 6,241,734 (“734 patent”).
`
`As I stated in my First Declaration, I have been retained as an expert by Stryker
`
`Corporation (“Stryker”) for that IPR. My background and qualifications remain
`
`the same as those set forth in my First Declaration.
`
`2.
`
`Since providing the First Declaration, I have reviewed the Board’s
`
`Decision instituting IPR, the Patent Owner Response, and the Declaration of
`
`Gamal Baroud, Ph.D., each relating to the 734 patent. I also reviewed the exhibits
`
`submitted by Orthophoenix and referenced by Dr. Baroud in this IPR, as well as
`
`the transcript of Dr. Baroud’s deposition dated July 30 and 31, 2015, as well as the
`
`exhibits referenced in this declaration. While I disagree with many of the
`
`statements made by Dr. Baroud in his declaration and during his deposition, I
`
`respond to only a few issues raised below.
`
`3.
`
`As I previously stated in my First Declaration, it was understood by
`
`August 1998 that bone cement could be delivered at “low pressures” including
`
`pressures no greater than about 360 psi. (¶ 39.) Such pressures were often
`
`preferred by physicians so the cement delivery could be adjusted and controlled.
`
`(¶ 39.) Indeed, typical vertebroplasty procedures at the time of the invention used,
`
`- 1 -
`
`STRYKER EXHIBIT 1041, pg. 1
`
`IPR2014-01433
`
`
`
`Second Declaration of Mary E. Jensen, M.D.
`
`
`
`for example, syringes with a volume of 1 cc or greater, resulting in delivery
`
`pressures less than about 360 psi. (¶ 40.)
`
`4.
`
`Dr. Baroud states that, at times, certain syringes could have delivery
`
`pressures above 360 psi. (Ex. 2021 at ¶¶ 19-27.) Specifically, he states that “the
`
`delivery pressures of PMMA or cement in a 2 or 3-cc syringe into bone will greatly
`
`exceed 360 psi. Therefore, the pressures in Deramond are not necessarily at or
`
`below 360 psi.” (¶¶ 26-27.) I disagree with Dr. Baroud’s opinion. The 734 patent
`
`provides a specific definition of low delivery pressure: “low delivery pressure, i.e.,
`
`a pressure no greater than about 360 psi” and explains “[a]s used herein, a ‘low
`
`delivery pressure’ is equivalent to the pressure at which liquid is expressed from 1
`
`cc syringe by the application of moderate force to the syringe piston, which
`
`amounts to a pressure that is no greater than about 360 psi.” (734 patent, at 3:7-10,
`
`2:6-10; see also 16:65-66 (“[t]he application of low pressure (i.e., no greater than
`
`360 psi), which is uniformly applied by the syringe 104 . . . .”).) In other words,
`
`the 734 patent equates “a pressure that is no greater than about 360 psi” to “the
`
`pressure at which liquid is expressed from a 1 cc syringe by the application of
`
`- 2 -
`
`STRYKER EXHIBIT 1041, pg. 2
`
`IPR2014-01433
`
`
`
`Second Declaration of Mary E. Jensen, M.D.
`
`
`
`moderate force to the syringe.”1 As I explained previously, the patent specifically
`
`and repeatedly teaches the ordinarily skilled artisan that the claimed delivery
`
`devices include conventional syringes of 1 cc or larger.2 (¶¶ 45-47; 734 patent at
`
`2:6-10, 10:33-35, FIGS. 5, 24-25, 27, 33.) In fact, a conventional syringe is the
`
`preferred delivery device identified in the 734 patent. Thus, consistent with my
`
`prior opinion, the ordinarily skilled artisan would understand that “a delivery
`
`device to convey the material at a delivery pressure of no greater than about 360
`
`psi” includes such conventional syringes, i.e., syringes of 1 cc or greater. (¶¶ 44-
`
`45, 40, 65, 69 n.5.)
`
`5.
`
`Thus, the fact that such syringes at times could have pressures above
`
`
`1 This is consistent with references that I previously discussed (paragraph 40),
`
`which show, e.g., that 1 cc syringes have mean maximum pressures of about 360
`
`psi (with larger syringes having even lower pressures). (Ex. 1011 (Hayward) at
`
`Table 1.)
`
`2 As I previously explained (¶ 40), as the size of the syringe increases, the delivery
`
`pressure of the syringe decreases. Thus, 2-, 3-, and 5- cc syringes deliver
`
`materials at lower pressures than 1-cc syringes. I note that Dr. Baroud agreed
`
`with my conclusion (Ex. 2021, ¶¶ 19, 23, 24; Baroud Tr. at p. 241, lns. 3-15).
`
`- 3 -
`
`STRYKER EXHIBIT 1041, pg. 3
`
`IPR2014-01433
`
`
`
`Second Declaration of Mary E. Jensen, M.D.
`
`
`
`360 psi – a concept not mentioned or accounted for by the 734 patent – is
`
`irrelevant to whether the syringes can and do deliver material at pressures no
`
`greater than about 360 psi under normal operating conditions. In fact, Dr.
`
`Baroud’s declaration (paragraphs 19 and 23) confirms that 2, 3, and 5 cc syringes
`
`(such as those disclosed in Deramond) can and do deliver materials at a pressure
`
`no greater than about 360 psi under normal conditions.
`
`6.
`
`I further note that with respect to 5-cc syringes in particular, Dr.
`
`Baroud has stated that the normal operating pressure of a 5 cc syringe is in the
`
`range of 161 to 334 psi (paragraph 19), which is well under 360 psi. Moreover,
`
`Dr. Baroud testified in his deposition that 5-cc syringes will always operate at
`
`pressures of less than 360 psi. (p. 267.) In my First Declaration, paragraph 39, I
`
`referenced the Krebs article that states that “[h]igh injection pressures approaching
`
`20 atmospheres [294 psi] are reached during conventional vertebroplasty” and
`
`reporting average pressures of 246 and 250 psi for normal and wide 5-cc syringes.
`
`Krebs discloses that the average for a 5cc syringe is 1698 kPa, which is 246 psi.
`
`(Ex. 1012, p. E120.) Moreover, the Krebs article shows the 5-cc syringe always
`
`operating at pressures much less than 360 psi, including during the first phase of
`
`injection. (Ex. 1012, Fig. 3, Table 1.)
`
`7.
`
`As Dr. Baroud acknowledged in his deposition (p. 278, lines 5-9),
`
`- 4 -
`
`STRYKER EXHIBIT 1041, pg. 4
`
`IPR2014-01433
`
`
`
`Second Declaration of Mary E. Jensen, M.D.
`
`
`
`there are physical limits on the pressures that a physician can manually generate
`
`using a syringe. Some of Dr. Baroud’s high-end pressure calculations for 1-, 2-,
`
`and 3-cc syringes are at odds with statements made regarding delivery pressures
`
`that can be generated by a physician. For example, in his declaration, Dr. Baroud
`
`states that “the delivery pressure that a physician can generate in 2 or 3-cc syringe .
`
`. . can reach pressures of up to 984.45 psi” and “the delivery pressure generated in
`
`the 1-cc syringe is in the range of 1640.76 +/- 575.24 psi.”3 (Ex. 2021, ¶¶ 23-24.)
`
`However, U.S. Patent Publication No. 2007/0185496, which was marked as
`
`Exhibit 1039 and identifies Dr. Baroud as an inventor, states that the 6900 KPa
`
`injection pressure, which is about 1000 psi, is “well beyond the limit of what of the
`
`surgeon can manually generate to inject cement with a standard syringe.”
`
`(Paragraph 9.) The application further states that “[p]ressures generated one-
`
`handed with a standard 2-cc syringe have demonstrated maximum obtained
`
`pressures in the order of 1700 KPa [247 psi].” (Paragraph 9.) This application
`
`later identifies that “the physical limit of the physician to be 1700 KPa,” which is
`
`247 psi. (Paragraph 54.) This is consistent with Krebs conclusion under “Key
`
`
`3 I note that, in paragraph 25, citing the Hayward reference, Dr. Baroud states that
`
`a 1-cc syringe can deliver materials as low as 216 psi but up to 728 psi.
`
`- 5 -
`
`STRYKER EXHIBIT 1041, pg. 5
`
`IPR2014-01433
`
`
`
`Second Declaration of Mary E. Jensen, M.D.
`
`
`
`Points”
`
`that “Peak
`
`injection pressures during conventional vertebroplasty
`
`approached on average 2000 kPa (20 atmospheres),” which is 290 psi. (Ex. 1012,
`
`p. E121.) Thus, Dr. Baroud’s estimates for theoretical high-end delivery pressures
`
`appear to suggest peak pressures well above what he has previously identified as
`
`the physical limits of the physician.
`
`8.
`
` In paragraph 22 of his declaration, Dr. Baroud states that “Krebs
`
`used an 8-gauge needle in his experiments” in connection with its syringes,
`
`apparently attempting to distinguish it from the needles that are attached to the
`
`syringes in Deramond. In my opinion, the size of the needle, e.g., 8-, 10-, or 15-
`
`gauge, attached to a syringe has no bearing on the delivery pressure of the syringe
`
`itself, which is the delivery pressure at issue in the claims. The claims state that
`
`the apparatus includes “a delivery device to convey the material . . . at a delivery
`
`pressure of no greater than about 360 psi.” This is supported by the 734 patent,
`
`which defines “low delivery pressure” (see above) to be independent of needle size
`
`and simply references delivery from the syringe itself. (734 patent at 2:6-10.)
`
`9.
`
`As I previously stated in my First Declaration, in my opinion, a person
`
`of ordinary skill in the art would have understood that material could be delivered
`
`into bone using the coaxial technique disclosed in Deramond (Fig. 4A), which was
`
`well known at the time of the invention. (¶ 71.) Dr. Baroud states that the coaxial
`
`- 6 -
`
`STRYKER EXHIBIT 1041, pg. 6
`
`IPR2014-01433
`
`
`
`Second Declaration of Mary E. Jensen, M.D.
`
`
`
`technique of Deramond “simply does not work” because “if the 15-gauge needle is
`
`inserted into the 10-gauge needle in a coaxial arrangement, the shorter 15-gauge
`
`needle cannot extend beyond the end of the longer 10-needle [sic, 10-gauge
`
`needle] (which would be required if the 15-gauge needle is to serve as a tamping
`
`instrument as discussed below).” (Ex. 2021 , ¶ 30.) Dr. Baroud apparently rests
`
`this conclusion on the needle lengths disclosed on page 285 of the Deramond
`
`reference. I disagree with Dr. Baroud’s conclusion. Figure 4A, reproduced from
`
`my First Declaration,
`
`illustrates
`
`the coaxial
`
`technique using a 15-gauge needle within a 10-
`
`gauge needle. Figure 4A clearly shows that the 15-
`
`gauge needle extends beyond the end of the 10-
`
`gauge needle. The fact that page 285 of Deramond
`
`identifies 10- and 15-gauge needles of particular
`
`lengths (10 to 15 cm and 5 to 7 cm, respectively)
`
`does not negate the clear depiction in Figure 4A,
`
`which shows a 15-gauge needle that is longer than a 10-gauge needle.4 Needles of
`
`
`4 In any event, the skilled artisan would not understand page 285 of Deramond to
`
`be a comprehensive list of every tool used in the procedures described in the
`
`
`
`- 7 -
`
`STRYKER EXHIBIT 1041, pg. 7
`
`IPR2014-01433
`
`
`
`Second Declaration of Mary E. Jensen, M.D.
`
`
`
`varying lengths within the same gauge were available off-the-shelf as of August
`
`1998. Thus, there were several 15-gauge needles available at the time of the
`
`invention including needles that were longer than 10-gauge needles. A person of
`
`ordinary skill in the art would have understood to select a needle of appropriate
`
`length for the coaxial technique and Deramond discloses that in Figure 4A.
`
`10. Dr. Baroud further states that the 15-gauge needle with stylet could
`
`not be “effectively” used to push material through the 10-gauge needle because
`
`there would be a space between the two needles. (Ex. 2021 at ¶¶ 31-33.) I
`
`disagree. I first note that there is nothing in the claims of the 734 patent that
`
`require any particular level of “effectiveness” for the tamping instrument. There is
`
`also nothing in the claims that precludes a space between the tamping instrument
`
`(including the nested stylet and nozzle of claim 12) and the subcutaneous cannula.
`
`Indeed, some type of space must exist between the tamping instrument and the
`
`access cannula or there could be no tamping. In this example, the space between
`
`the access cannula and the tamping instrument would not impede the ordinary
`
`skilled artisan from using the nested nozzle and stylet as a tamping instrument
`
`
`
`reference. For example, page 285 does not reference 5 cc syringes, yet 5 cc
`
`syringes are referenced elsewhere in the Deramond reference (e.g., page 287).
`
`- 8 -
`
`STRYKER EXHIBIT 1041, pg. 8
`
`IPR2014-01433
`
`
`
`Second Declaration of Mary E. Jensen, M.D.
`
`
`
`within the subcutaneous cannula. An ordinarily skilled artisan such as a physician
`
`would have understood that residual material could be cleared from the cannula by
`
`repeatedly pushing the tamping instrument (e.g., the 15-gauge needle with stylet)
`
`within the lumen of the subcutaneous cannula (e.g., the 10-gauge needle).
`
`11. Moreover, if the ordinary skilled artisan preferred a tighter fit between
`
`the cannula and tamping instrument, he or she could have selected a larger-
`
`diameter needle as a tamping instrument. Many different needle gauges (with
`
`stylets) were available off-the-shelf by the time of the invention. A 13-gauge
`
`needle (outer diameter about 2.4 mm) is an example of a larger-diameter
`
`instrument that would have been available off-the-shelf for use within a 10-gauge
`
`needle (inner diameter about 2.7 mm). (Ex. 2021, Machinery Handbook, p. 2520.)
`
`Indeed, in performing coaxial techniques such as that depicted in Deramond, the
`
`physician understood to select a particular gauge depending on the anatomical
`
`location of the specific procedure as well as the type and amount of material being
`
`delivered. For example, a smaller needle (e.g., 15-gauge) would have generally
`
`been preferred for the cervical vertebra to avoid damage to the carotid artery and
`
`other delicate vasculature. On the other hand, a larger needle (11-gauge) would
`
`have generally been preferred for procedures in the thoracic and lumbar vertebra to
`
`deliver a greater amount of filling material to these larger vertebrae.
`
`- 9 -
`
`STRYKER EXHIBIT 1041, pg. 9
`
`IPR2014-01433
`
`
`
`Second Declaration of Mary E. Jensen, M.D.
`
`
`
`12. As I previously stated in my First Declaration, Kuslich (Ex. 1009)
`
`describes delivering filling material to an expandable implant bag 40 that is
`
`inserted through a subcutaneous cannula 54 into a cavity in bone. (¶ 161.) Dr.
`
`Baroud states that “figures 5 and 6 show a picture of an intervertebral disk not a
`
`bone.” (Ex. 2021 at ¶ 35.) However, as I previously stated, while Kuslich does
`
`involve delivering filling material into an intervertebral disk, it also discloses
`
`delivering filling material into bone. This is shown in figures 4, 5, and 6 of
`
`Kuslich, which depict removal of bone of the vertebra (circled below) to expose
`
`cancellous bone to allow for fusion of the two vertebrae across the disk space.
`
`
`
`- 10 -
`
`STRYKER EXHIBIT 1041, pg. 10
`
`IPR2014-01433
`
`
`
`Second Declaration of Mary E. Jensen, M.D.
`
`
`
`This is further supported by claim 1 of Kuslich, which states that the cavity is
`
`formed by “boring an opening into an annulus of a disc, said disc including a
`
`nucleus and end plates, each end plate being adjacent a vertebra, excising out at
`
`least a portion of the nucleus and a portion of the end plates of said disc and bone
`
`of adjacent vertebrae to said disc to expose healthy bone.” (Ex. 1009 at claim 1.)
`
`Indeed, without removing bone, a physician could not fuse the two vertebrae and
`
`this would frustrate the entire purpose of Kuslich.
`
`13.
`
`In paragraph 37, Dr. Baroud references “physiological compressive
`
`loads exceeding 1000 psi.” As Kuslich clearly explains, this is the mechanical
`
`forces acting on the spinal disc, e.g., “when a heavy load is lifted,” and has nothing
`
`to do with the delivery pressure of the Kuslich device. (Ex. 1009 at 1:50-52.)
`
`14.
`
`In paragraph 38, Dr. Baroud states that the implanted bag of Kuslich
`
`“would leak” if filled with bone cement. This is irrelevant to claim 12 of the 734
`
`patent, which requires delivery of “material” and does not limit the material to
`
`bone cement (and also does not involve any bag to constrain bone cement). The
`
`734 patent describes many materials that can be selected for delivery into bone
`
`including autograft and allograft tissue, which is consistent with what is disclosed
`
`in Kuslich. (Ex. 1001 at 10:22-26; Ex. 1009 at 9:57-60.) In any event, Kuslich
`
`specifically states that the bag contains the material. (Ex. 1009 at 7:18-20.)
`
`- 11 -
`
`STRYKER EXHIBIT 1041, pg. 11
`
`IPR2014-01433
`
`
`
`STRYKER EXHIBIT 1041, pg. 12
`
`IPR2014-01433
`
`STRYKER EXHIBIT 1041, pg. 12