throbber
UNITED STATES PATENT AND TRADEMARK OFFICE
`______________________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`______________________
`
`MEDTRONIC, INC., MEDTRONIC VASCULAR, INC.,
`and MEDTRONIC COREVALVE, LLC
`Petitioner
`
`v.
`
`TROY R. NORRED, M.D.
`Patent Owner
`______________________
`
`Case IPR2014-00395
`Patent 6,482,228
`______________________
`
`DECLARATION OF TIMOTHY CATCHINGS, M.D.
`
`NORRED EXHIBIT 2295 - Page 1
`Medtronic, Inc., Medtronic Vascular,
`Inc., & Medtronic Corevalve, LLC
`v. Troy R. Norred, M.D.
`Case No. IPR2014-00395
`
`

`
`I, Timothy T. Catchings, M.D., declare as follows:
`
`1.
`
`I have been asked by Troy R. Norred, M.D.
`
`to provide my
`
`opinions in this declaration in connection with the iizterparxer review of United States
`
`Patent No. 6,482,228 (the “‘228 patent”).
`
`In preparing this declaration,
`
`I have
`
`reviewed the ‘228 patent, the prosecution history of the ‘228 patent, the prior art cited
`
`by the Patent Trial and Appeal Board in instituting these proceedings, and other prior
`
`art that is or may be pertinent to the patentability of Dr. Norred’s invention. I have
`
`personal knowledge of the following facts and would and could testify competently
`
`regarding the following statements if called as a witness.
`
`2.
`
`In forming my opinions, I understand that the claims should be
`
`interpreted as they would be understood by a person of ordinary skill in the art of the
`
`patents.
`
`I also understand that
`
`they are ordinarily construed based on the plain
`
`meaning of the terms used in the claims, and also with reference to the specification,
`
`the patent drawings, and the prosecution history. In addition,
`
`I understand that
`
`although the specification should be consulted to aid in the process of interpreting the
`
`claims, the specific examples disclosed in the specification generally do not limit the
`
`scope of the claims. Finally, I also understand that claim interpretation may be aided
`
`by reference to other sources of information, such as dictionaries, textbooks, and
`
`literature or other patents in related fields,
`
`in order to determine the ordinary
`
`meanings of terms used in the claims.
`
`NORRED EXHIBIT 2295 - Page 2
`NORRED EXHIBIT 2295 - Page 2
`
`

`
`3.
`
`In my opinion, a person of ordinary skill in the art would possess
`
`a Doctor of Medicine from an accredited medical school plus at least three yea1's of
`
`residency in internal medicine or the equivalent in surgical residency, plus three years
`
`of cardiology fellowship or the equivalent in cardiovascular surgery.
`
`4.
`
`All of the opinions expressed in this declaration are my own,
`
`formed upon analysis of the materials listed above.
`
`Professional Background
`
`5.
`
`I am a citizen of the United States and a resident of Waycross,
`
`Georgia. My post office address is 1610 Screven Avenue, Waycross, Georgia 31501.
`
`6.
`
`I currently am in private practice in Waycross, Georgia as a board
`
`certified interventional cardiologist.
`
`7.
`
`As part of my practice,
`
`I see and treat patients suffering from
`
`valvular heart disease, including aortic stenosis.
`
`8.
`
`I attended college at Albany State College in Albany, Georgia
`
`from 1974 to 1978, earning a Bachelor of Arts in Chemistry.
`
`I attended medical
`
`school at Emory University School of Medicine in Atlanta, Georgia from 1974 to
`
`1.978, earning a Doctor of Medicine. Following graduation from medical school, from
`
`1978 to 1979,
`
`I worked as an Intern at Medical College of Georgia Affiliated
`
`Hospitals in Augusta, Georgia. Following my internship, from 1980 to 1982, I served
`
`as a Resident at
`
`the Naval Regional Medical Center
`
`in Portsmouth, Virginia.
`
`NORRED EXHIBIT 2295 - Page 3
`NORRED EXHIBIT 2295 - Page 3
`
`

`
`Following my residency, from 1982 to 1984, I was a Fellow in Pulmonary Diseases at
`
`the Naval Hospital in Portsmouth, Virginia. After that, fromjuly 1995 to May 1998, I
`
`was a Fellow in Cardiovascular Diseases at the National Naval Medical Center in
`
`Bethesda, Maryland.
`
`Then,
`
`from July 1999 to July 2000,
`
`I was a Fellow in
`
`Cardiovascular Diseases at the University Hospital of the University of Missouri.
`
`I
`
`was an Assistant Professor of Medicine and the Director of the Coronary Care Unit at
`
`the University Hospital at the University of Missouri from August 2000 until August
`
`2006. After that, I worked in private practice in \Williamsburg, Virginia until October
`
`2008, when I moved to my present position.
`
`9.
`
`I am a Retired Captain in the Medical Corps of the U.S. Navy
`
`Reserve. As part of my military service, I served on the USS Sylvania from 1979 to
`
`1980; at the Naval Regional Medical Center in Portsmouth, Virginia from 1980 to
`
`1982; at the Naval Hospital in Portsmouth, Virginia from 1982 though 1987; at the
`
`Naval Reserve Centers in Macon and Atlanta, Georgia from 1987 to 1995; and at the
`
`National Naval Medical Center in Bethesda, Maryland from 1995 to 1999.
`
`In
`
`addition, from 1990 to 1992,
`
`I was the Director of Medical Services at the Fleet
`
`Hospital 14 Headquarters injacksonville, Florida.
`
`10.
`
`I
`
`am or have been a member of
`
`several professional
`
`organizations, including the American College of Chest Physicians, the Society of
`
`NORRED EXHIBIT 2295 - Page 4
`NORRED EXHIBIT 2295 - Page 4
`
`

`
`Critical Care Medicine, the American Medical Association, the Bibb County Medical
`
`Society, the National Medical Association and the Virginia Thoracic Society.
`
`11.
`
`I hold a license to practice medicine in Georgia, and have
`
`previously held licenses in Virginia, Missouri and Oklahoma.
`
`12.
`
`I am at least a person of ordinary skill in the art as it pertains to
`
`the prosthetic aortic valve invented by Dr. Norred and described in the ‘228 Patent.
`
`T/53 Nafive Aorfic Vafve
`
`13.
`
`The aortic valve is located between the left ventricle and the aorta.
`
`14.
`
`The function of the aortic valve is to allow blood to flow in one
`
`direction, from the left ventricle to the aorta.
`
`15.
`
`The aortic valve opens to allow blood to flow into the aorta, and
`
`closes to prevent back flow into the left ventricle. It does this approximately 103,000
`
`times per day, and approximately 3.7 billion times in its lifespan.
`
`16.
`
`The aortic valve consists of three membranous leaflets and three
`
`aortic sinuses.
`
`17.
`
`The leaflets are the parts of the valve that most directly control
`
`blood flow.
`
`It once was thought that the leaflets were pushed open by the ejection of
`
`blood from the left ventricle. We know now that this View was overly simplistic and
`
`somewhat
`
`inaccurate.
`
`The leaflets are attached to the aortic wall
`
`through
`
`commissures. The commissures move outwardly during systole and inwardly during
`
`‘5'
`
`NORRED EXHIBIT 2295 - Page 5
`NORRED EXHIBIT 2295 - Page 5
`
`

`
`diastole. It is the outwardly movement of the commissures that causes the leaflets to
`
`open.
`
`18.
`
`Comrnissural movement is possible because of the elastic walls of
`
`the sinuses. The sinuses are cavities behind the leaflets, and represent dilations of the
`
`base of the aorta. The sinuses are continuous with the left ventricle at the lower
`
`margin, and form part of the ascending aorta at the upper margin.
`
`19.
`
`The area where each leaflet merges with the other two leaflets to
`
`close the valve is called the coaptation surface. The coaptation surface helps prevent
`
`aortic regurgitation.
`
`20.
`
`The leaflets, commissures and sinuses together form a unit that
`
`allows for the circumferential dispersion of force throughout the valve. This, in turn,
`
`contributes to the longevity of the valve by reducing the stress and strain on the
`
`leaflets themselves.
`
`21.
`
`A cardiac cycle involves the rapid opening of the valve, followed
`
`by the ejection of blood from the heart. The valve is maximally open in early systole,
`
`as the leaflets move into the sinuses. Vortices form between the leaflets and the sinus
`
`walls. As systole continues, the persistence of these vortices combined with the
`
`deceleration of aortic flow causes the leaflets to gradually come together. By the end
`
`of systole, the valve is almost closed. At that point, the leaflets close rapidly and the
`
`heart begins to fill with blood in preparation for the next cycle.
`
`NORRED EXHIBIT 2295 - Page 6
`NORRED EXHIBIT 2295 - Page 6
`
`

`
`Dr. Norredir Im,-emtiorz
`
`22.
`
`Dr. Norred first approached me about his prosthetic aortic valve
`
`in 1999, while we were both at the University of Missouri. At that time, surgical aortic
`
`valve replacement was the state of the art. A person of ordinary skill in the art was
`
`not aware of a prosthetic aortic valve that could be placed in a patient through non-
`
`surgical methods.
`
`23.
`
`Surgical aortic valve replacement
`
`is highly traumatic because,
`
`among other things, it requires that a patient’s heart be stopped and the patient placed
`
`on a cardiopulmonary bypass machine. This limits the types patients for whom
`
`surgical aortic valve replacement is a viable treatment option. Elderly and infirm
`
`patients often are unable to tolerate this surgery and therefore cannot receive a
`
`surgically placed aortic valve.
`
`24.
`
`Dr. Norred suggested that a prosthetic aortic valve could be
`
`placed percutaneously through a catheter held in place with a stent structure.
`
`I
`
`immediately saw the value in this concept, but was skeptical it would work.
`
`I did not
`
`believe the radial force exerted by the stent would be sufficient to anchor the device in
`
`the aorta without damaging the device or the aortic wall or both.
`
`I thought the device
`
`would dislodge or leak.
`
`25.
`
`From the fall of 1999 through the fall of 2000, Dr. Norred and I
`
`had regular discussions regarding the viability of and need for his invention.
`
`J-
`
`NORRED EXHIBIT 2295 - Page 7
`NORRED EXHIBIT 2295 - Page 7
`
`

`
`26.
`
`From my discussions with Dr. Norred, I learned that he was able
`
`to develop, in collaboration with others at the University of Missouri, a mathematical
`
`model to prove that an aortic valve could be held in place with a stent.
`
`27.
`
`From my perspective, this was critical because, absent such proof,
`
`it was unlikely Dr. Norred could garner support for the development and testing of
`
`the device.
`
`I know his objective was to test
`
`the device in live animals for his
`
`fellowship project.
`
`28.
`
`The prosthetic aortic valve invented by Dr. Norred has four
`
`characteristics that a person of ordinary skill in the art would consider functionally
`
`significant.
`
`29.
`
`First, the prosthetic aortic valve invented by Dr. Norred consists
`
`of components that can be collapsed to fit into a catheter. This is what allows the
`
`device to be placed percutaneously. This, primarily, is what distinguishes this device
`
`from prosthetic aortic valves that were on the market at
`
`the time Dr. Norred
`
`conceived his invention. The prosthetic valves that were on the market at that time,
`
`such as the Starr-Edwards valve, St. _]ude’s tilting disk valve and the Medtronic tilting
`
`disk valve, were comprised of rigid components that could never fit within a catheter.
`
`30.
`
`Second,
`
`the prosthetic aortic valve invented by Dr. Norred
`
`anchors in place by the interconnected rods of the stent system exerting radial force
`
`against the aortic wall. This eliminates the need for sutures. This is important
`
`'8‘
`
`NORRED EXHIBIT 2295 - Page 8
`NORRED EXHIBIT 2295 - Page 8
`
`

`
`because suturing cannot be done through a catheter.
`
`It was and remains a surgical
`
`procedure. Eliminating the need for sutures is thus imperative if the device is to be
`
`placed percutaneously.
`
`31.
`
`Third, the prosthetic aortic valve invented by Dr. Norred contains
`
`a stent stnlcture that extends into the ascending aorta. This extended stent structure
`
`allows for the circumferential dispersion of force, mimicking the manner in which the
`
`native valve disperses force through the leaflets, commissures and sinuses acting in
`
`concert with one another. This is important because it reduces the stress and strain
`
`on the leaflets, and contributes toward the longevity of the valve.
`
`In addition,
`
`this
`
`extended stent structure makes it significantly easier to place and correctly align the
`
`prosthetic valve.
`
`32.
`
`Finally,
`
`the prosthetic aortic valve invented by Dr. Norred
`
`contains a structural element—a ring member—to seal it against the root of the native
`
`aortic valve upon placement. This is important because if the device is not sealed
`
`against the root of the native aortic valve upon placement, the high pressures and high
`
`flow of blood within the aorta will cause perivalvular leaks.‘ A prosthetic aortic valve
`
`susceptible to perivalvular leaks would not be acceptable to a person of ordinary skill
`
`‘ As used herein, the term “perivalvular leak” means a leak between the aorta and the
`
`left ventricle.
`
`'9'
`
`NORRED EXHIBIT 2295 - Page 9
`NORRED EXHIBIT 2295 - Page 9
`
`

`
`in the art, because such leaks can lead to heart failure, hemolytic anemia, blood clots,
`
`stroke, infective endocarditis and death?
`
`33.
`
`Referring specifically to the claims of the ‘228 Patent, I would
`
`understand the term “ring member” as used in claims 16 and 20 to mean a ring made
`
`of pliable, biocompatible material. This definition derives from the specification,
`
`Named,
`
`():1—2, and also from the requirement
`
`that
`
`the ring member have a
`
`circumference “adapted” to seat about the aortic wall. The aorta is not a perfectly
`
`circular or cylindrical structure, but rather, has an irregular, oblong shape that differs
`
`from person to person. The adaptation that would allow the ring member to seat
`
`against this irregular, oblong surface is pliability. Further, the aortic wall is not static.
`
`Instead, as discussed elsewhere herein, it is constantly expanding and contracting.
`
`The adaptation that would allow the ring member to seat against this expanding and
`
`contracting surface is pliability. A pliable ring member could move in tandem with
`
`these expansions and contractions without becoming dislodged. Finally,
`
`I would
`
`understand the term “ring member” as used in claims 16 and 20 to mean a ring that
`
`2 Some early designs, such as Wolfe, US 4,030,142, published june 21, 1977, relied
`
`upon leaks and the subsequent formation of clots to maintain a seated position in the
`
`heart.
`
`ll’/oie, 3:61-68. This never would be acceptable to a person of ordinary skill in
`
`the art. These designs were never put into practice.
`
`'10‘
`
`NORRED EXHIBIT 2295 - Page 10
`NORRED EXHIBIT 2295 - Page 10
`
`

`
`seals against the root of the native aortic valve to reduce perivalvulat leaks.
`
`This
`
`definition is consistent with the specification.
`
`Nomad,
`
`6:1—4. Moreover,
`
`this
`
`functionality is vital in order for the prosthetic aortic Valve to be a viable replacement
`
`for the native aortic valve.
`
`34.
`
`I would understand the term “membrane” as used in claim '16 to
`
`be a thin, soft, pliable sheet or layer. This is the ordinary and customary meaning of
`
`“membrane” and one that is consistent with the specification. Norred, l:26~27, 4:16-
`
`19, 4:58-64, 5:40-44.
`
`35.
`
`I would understand the term “means for mounting” as used in
`
`claim 16 to mean hingedly attached or hingedly secured through a material fold, with
`
`the material functioning as a hinge. This is consistent with the specification and is
`
`depicted in Fig. 18.
`
`36.
`
`I would understand the term “means for maintaining” as used in
`
`claim 19 to mean the interconnected rods that form the stent system that extends into
`
`the ascending aorta. This is consistent with the specification. The specification
`
`provides that stent system 28 is made up of a small, slotted stainless steel tube or
`
`series of interconnected rods. Narred, 2:60-62. The stent system 28 is depicted in its
`
`entirety in Fig. 5. The inflow region of the stent system is depicted in Fig. 18. Rods
`
`104 are shown in Fig. 18 as an integrated part of the stent system 28. Nomad, 6:4-7.
`
`-11-
`
`NORRED EXHIBIT 2295 - Page 11
`NORRED EXHIBIT 2295 - Page 11
`
`

`
`37.
`
`I would understand claim 16, constnied in
`
`light of
`
`the
`
`specification, to describe a prosthetic aortic valve comprised of a membrane hingedly
`
`attached or hingedly secured to a ring member, where both the membrane and the
`
`ring member can be collapsed to fit within a catheter for percutaneous delivery,’ and
`
`where the ring member has a pliable circumference in order to seat about the aortic
`
`wall and seal against the root of the native aortic valve to reduce perivalvulat leaks.
`
`38.
`
`I would understand claim 19, construed in light of
`
`the
`
`specification, to mean the prosthetic aortic valve described in claim 16, where the ting
`
`member is attached to the interconnected rods of a stent system that extends into the
`
`ascending aorta and maintains the ring member against the aortic wall.
`
`39.
`
`I would understand claim 20, construed in light of
`
`the
`
`specification, to describe a tissue valve hingedly attached or hingedly secured to a ring
`
`3 I am aware the Board found these claims are not limited to a prosthetic aortic valve
`
`that can be placed percutaneously.
`
`I note, however, these claims specifically refer to
`
`the “placement” of the prosthetic valve. The only form of placement described in the
`
`specification is percutaneous placement. Norrea’, Title, Abstract, 1:7-9, 1:26-27, 1:32-
`
`33, 1:53-55, 1:58-60, 1:67-2:2, 2:55-3:18, 5:63-67. Accordingly, I would read into each
`
`of these claims the requirement that the prosthetic valve be capable of percutaneous
`
`placement.
`
`'12"
`
`NORRED EXHIBIT 2295 — Page 12
`NORRED EXHIBIT 2295 - Page 12
`
`

`
`member, where both the tissue valve and the ring member can be collapsed to fit
`
`within a catheter, where the ring member has a pliable circumference in order to seat
`
`about the aortic wall and seal against the root of the native aortic valve to reduce
`
`perivalvular leaks, and where the ring member is attached to the interconnected rods
`
`of a stent system that extends into the ascending aorta and maintains the ring member
`
`against the aor1:ic wall.
`
`Pa1Fem‘abz'Zz'_zjz qfDr. Narredis‘ Im2em'z'on
`
`40.
`
`I am aware of other prosthetic valves that were designed for or
`
`purportedly capable of percutaneous placement. However, each of these designs
`
`lacks one or more of the attributes that make Dr. Norred’s device viable.
`
`41.
`
`One example is Anderson, US 5,411,552, published May 2, 1995.
`
`Anderson is comprised of an elastically collapsible valve mounted on an elastical stent.
`
`Anderxorz, 2:30-37. This construction allows the device to be placed percutaneously
`
`and to anchor in place without sutures. Anderrozz, 2:56-68, 6:20-44. However,
`
`Anderson lacks a ring member to seal the device against the root of the native aortic
`
`valve. Anderson, 2:28-55, Figs 1 & 2. As a result, there is nothing to prevent blood
`
`from flowing between and around the loops of the stent. Perivalvular leaks will
`
`occur, which could cause immediate, catastrophic, procedural failure resulting in
`
`death. The same problem is found in Bailey, US 6,458,153, published October 1,
`
`2002; jayaraman, US 5,855,597, published January 5 1999; Ersck, US 3,657’,744,
`
`' I 3'
`
`NORRED EXHIBIT 2295 — Page 13
`NORRED EXHIBIT 2295 - Page 13
`
`

`
`published April 25, 1972; and Garrison, US 6,425,916, published July 30, 2002. It also
`
`is
`
`found in Figulla, German Patent App. No. DE 195-46-692, and Fraunhofer,
`
`German Patent App. No. 198-57-887. None of these designs contain a structural
`
`component to seal the device against the root of the native aortic valve.
`
`42.
`
`Dill/Iatteo, US 6,440,164, published August 27, 2002, incorporates
`
`a trellis surrounded by a biocompatible non-thrombogenic liner that is intended to
`
`facilitate tissue ingrowth over time, with this tissue ingrowth eventually creating a seal.
`
`Dzrl/lattes, 4:61-62,
`
`l1:‘l—4. This theory is fundamentally flawed. The pressures and
`
`blood flow in the aorta are too great for tissue ingrowth to occur. If a prosthetic aortic
`
`valve has a perivalvular leak upon placement, it will not seal. A perivalvular leak could
`
`cause immediate, catastrophic procedural failure, resulting in death of the patient.
`
`Because the device described in DiMatteo does not seal upon placement, it would not
`
`be acceptable to a person of ordinary skill in the art.
`
`43.
`
`Leonhardt, US
`
`5,957,949, published September
`
`28,
`
`1999,
`
`describes a valve comprised of a super—elastic spring stent fully enclosed by graft
`
`material coated with light—activated adhesive. L20}:/Jardt, 3:33—45, 6:4—7.
`
`If this device
`
`is placed within the native annulus, graft material must be cut from between the
`
`distensible fingers that are formed by the zig-zags of the stent in order to prevent the
`
`stent from blocking the coronary artery. Leon/Jardt, 6:9-23. The device then must be
`
`placed so that the distensible fingers are on either side of the coronary artery.
`
`Id.
`
`-14-
`
`NORRED EXHIBIT 2295 - Page 14
`NORRED EXHIBIT 2295 — Page 14
`
`

`
`There is no way to ensure such precise placement, even with radiopaque markers, as
`
`these only provide a two—dimensional depiction of the three—dimensional placement
`
`site.
`
`If one of the distensible fingers were to occlude the coronary artery, it would
`
`lead to a myocardial infarction and possibly death. A person of ordinary skill in the
`
`art would consider the risk of occlusion unacceptably high with this device.
`
`44.
`
`Further, because the distensible fingers must be placed on either
`
`side of the coronary artery,
`
`the entirety of the stent structure, except
`
`for
`
`the
`
`distensible fingers, must be below the coronary artery. L.eoz2/Jardt, 5:40-52, Fig. 2. The
`
`device cannot extend into the ascending aorta, and therefore cannot displace force
`
`citcumferendally in a manner mimicking the native valve.
`
`.-\s a result, this valve is
`
`likely to fail}
`
`4 Leonhardt is one of several prosthetic aortic valves that utilize a stent as a mere
`
`substitute for sutures. Because the purpose of the stent in these designs is to anchor
`
`the valve to the aortic wall, the stent typically occupies the same area as the native
`
`aortic valve and does not extend above the coronary artery into the ascending aorta.
`
`The performance of these designs is likely to parallel the performance of surgically-
`
`implanted valves such as the Hancock valve. These valves are unable to displace
`
`force circumferentially, so the leaflets undergo far greater stress and strain than is
`
`experienced by the leaflets in the native valve. This causes these valves to fail.
`
`-15-
`
`NORRED EXHIBIT 2295 - Page 15
`NORRED EXHIBIT 2295 — Page 15
`
`

`
`45.
`
`Additionally, the distensible fingers on the inflow portion of the
`
`Leonhardt valve will interfere with the subvalvular structures of the aorta, such as the
`
`mitral valve, the chordae tendinae, and the membranous septum. Interference with
`
`the mitral valve leaflet and the chordae tendinae could induce mitral regurgitation,
`
`which in turn could result in immediate congestive heart failure.
`
`Interference with the
`
`membranous septum could disrupt the timing function of the heart.
`
`46.
`
`If the Leonhardt device is placed above the native annulus as
`
`depicted in Fig. 3 rather than within the native annulus, cutting graft material will not
`
`prevent the coronary artery from being blocked, because biological valve 22 will block
`
`the coronary artery. Leorz/Jardt, 3:59-61, 6:23-31. Biological valve 22 is a membrane
`
`that will not allow blood flow therethrough.
`
`47.
`
`Leonhardt also is problematic because it does not utilize a ring
`
`member to seal
`
`the device against
`
`the root of the native aortic valve to reduce
`
`perivalvular leaks.
`
`Instead, Leonhardt relies on the stent structure itself to create a
`
`temporary seal. See Lear:/Jardt, 4:55-60, 5:2-5, 5:48-50. This temporary seal allows the
`
`device to be removed or repositioned at will, even after full deployment.
`
`I..eou/Jardz,
`
`3:4~6, 3:27-30, 11:37-53. Once the device is positioned properly, a permanent seal
`
`purportedly is created through activation of the adhesive coating. Leorzfmrdr, 8:43-67,
`
`9:1-5. The adhesive coating purportedly seals the device until it ultimately degrades,
`
`and living tissue takes its place. Lem:/2am’t, 8:43-54, 8:66-67, 9:1. This design would not
`
`-16-
`
`NORRED EXHIBIT 2295 - Page 16
`NORRED EXHIBIT 2295 — Page 16
`
`

`
`be acceptable to a person of ordinary skill in the art. At the time Leonhardt was
`
`published, there were no light—activated adhesives that would work in a wet, high-
`
`pressure environment such as
`
`the aorta.
`
`I am aware of a substance known as
`
`poly(glycerol sebacate acrylate) or PGSA, a mix of glycerol (C31-1803) and sebacic
`
`acid (C101-H804) that is currently in development for this purpose, but clinical trials
`
`have not yet begun on this substance, and it remains unclear whether it could or
`
`would work with the Leonhardt device.
`
`48.
`
`Certain aortic valve designs incorporate stents that as constructed
`
`lead to sharp areas of protrusion. These sharp areas of protrusion are typically
`
`disclosed as barbs, hooks or flanges. These designs include Figulla, German Patent
`
`App. No. DE 195-46-692; Fraunhofer, German Patent App. No. 19867-887;
`
`jayaraman, US 5,855,597, published January 5, 1999; and Bailey, US 6,458,153,
`
`published October 1, 2002. These designs are placed through expansion of the stent.
`
`Expansion of the stent will cause these sharp areas ofprotrusion to come into Contact
`
`with delicate surrounding tissue. This will cause these sharp areas of protrusion to
`
`pierce this tissue, particularly when combined with the high pressures within the aorta
`
`and the Constant expansion and contraction of the aorta and heart. This,
`
`in turn,
`
`could lead to aortic dissection and/ or perforation. As a result, devices employing
`
`these sharp areas of protrusion would not be acceptable to a person of ordinary skill
`
`in the art.
`
`-17-
`
`NORRED EXHIBIT 2295 - Page 17
`NORRED EXHIBIT 2295 — Page 17
`
`

`
`49.
`
`Bailey, US 6,458,153, published October 1, 2002, anchors in place
`
`by means of proximal anchor flange 22, which consists of a plurality of stent struts
`
`and stent interstices that project radially outward from the central longitudinal axis of
`
`the valve stent 10 as depicted in Fig. 1. Bai/e , 9:61-65.
`
`If this device is placed in the
`
`native annulus as shown in Fig. 20H,
`
`these stent struts will
`
`interfere with the
`
`subvalvular structures of the aorta, such as the mitral valve leaflet,
`
`the chordae
`
`tendinae,
`
`the membranous septum and the left ventricular outflow tract.
`
`As
`
`discussed in connection with the Leonhardt device, interference with the mitral valve
`
`leaflet and the chordae tendinae could induce mitral regurgitation, which in turn could
`
`result in immediate congestive heart
`
`failure.
`
`Interference with the membranous
`
`septum could disrupt the timing function of the heart. If anchor flange 22 pierces the
`
`membranous septum, it will cause an immediate ventricular septal defect.
`
`If anchor
`
`flange 22 pierces the left ventricular outflow tract, it could cause exsanguination and
`
`death.
`
`50.
`
`Bailey proposes to avoid these issues by eliminating one or more
`
`stent struts to define an open region where these subvalvular structures are located.
`
`Baife , 9:65-10:14. This
`
`solution is unworkable for
`
`two reasons.
`
`First,
`
`these
`
`subvalvular structures completely surround the inflow portion of the valve, so unless
`
`all stent struts are eliminated, they will interfere with these structures. Second, even if
`
`these subvalvular structures did not completely surround the inflow portion of the
`
`‘1 8'
`
`NORRED EXHIBIT 2295 - Page 18
`NORRED EXHIBIT 2295 — Page 18
`
`

`
`valve, there is no way to place the device with sufficient precision to avoid these
`
`structures. Although radiopaque markets can be employed to assist with placement,
`
`these only provide a two—dimensional depiction of the three—dimensional placement
`
`site. There is an unacceptably high likelihood the device will be placed in a manner
`
`that interferes with or obstructs surrounding structures.
`
`1-"urther, removal of stent
`
`struts will lessen the ability of the flange 22 to anchor the device in place, which
`
`makes it more likely the device will leak or dislodge.
`
`51.
`
`The valve described in Schreck, US 6,454,799, published
`
`September 24, 2002, purports to anchor in place through the combination of a tubular
`
`member, which exerts radial force against the walls of the native aortic valve, and
`
`barbs, staples or flanges as necessary.
`
`Sc/Jreaé, 2:16-36, 13:12-31. The immediate
`
`problem with this design is that the tubular member does not extend into the aorta,
`
`even into the area where the new valve is seated (the new valve is seated directly
`
`above the “tissue engaging base,” which houses the tubular member).
`
`5:.‘/Jrecvé, Figs. 2
`
`& 7. As a result, there is nothing to maintain the new valve against the aortic wall.
`
`When the new valve is exposed to the high pressures within the aorta, the posts 146
`
`and 148 will cantilever towards the center of the device. This will pull the tissue-
`
`engaging base 104 and fabric skirt 110 away from the aortic wall, causing
`
`misalignment of the leaflets 32, perivalvular leaks, and ultimately dislodgmcnt and
`
`failure of the device. See 3:‘/Jl‘E6‘/(3, 8:10-13.
`
`'19‘
`
`NORRED EXHIBIT 2295 - Page 19
`NORRED EXHIBIT 2295 — Page 19
`
`

`
`52.
`
`I have no financial
`
`interest
`
`in D1’. N01'1'ecl’s
`
`invention or the
`
`outcome of these proceedings.
`
`I hereby declare under the penalty of perjury under the laws of the United States of
`
`..-'\me1'ie21 that the foregoing is true and flCCu1’:1tC to the heel’ of In}? knowledge and
`
`understanding.
`
`___‘__
`
`
`Do. +ae_
`
`:......‘._?a-....‘_...
`
`l\3o¥:;"¥L\"‘ ' };.b\ k c.
`
`Vi ‘ I Ht ’ 3.0‘: Li
`
`Notary Pub1ic,Piarca County. Georgia
`My Commission Expires Mar. 3, 2GI5
`
`NORRED EXHIBIT 2295 - Page 20
`NORRED EXHIBIT 2295 - Page 20

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket