throbber
FATIGUE PERFORMANCE OF NITINOL TUBING'WITH AEOF 25°C
`Tracy Lashley Lopes, Xiao-Yan Gong, and Christine 'Irépanier
`Nitinol Devices &. Components, 47533 Westinghouse Drive, Fremont, CA 94539
`
`ABSTRACT
`The purpose of this study is to assess the fatigue behavior of Nitinol tubing aged at different tem—
`peratures 850-500" C) and times to achieve a 25° C i2” C Af. Fatigue properties were determined
`in tension at a mean strain of 1.5% and alternating strain magnitudes of 0.2%, 0.3%, and 0.4%.
`Although there are insufficient data to draw statistical conclusions from this preliminary investiga-
`tion, it is important to point out that there did not appear to be any major differences in the samples *
`tested despite the differences in microstructure and therefore in plateau strength. Still, among the
`heat-treatment temperatures evaluated, 500°C produces the smallest amount or” R—phase with the
`most fatigue-stable microstructure, and therefore is the recommended choice for good reproducible
`device performance.
`
`KEYWORDS
`
`Tension, Fatigue, Mechanical Properties, Af, Aging
`
`INTRODUCTION
`Like many metallic biomaterials, Nitinol’s mechanical properties are determined by the thermome—
`chanical processing of the alloy. In addition, Nitinol’s mechanical behavior at body temperature is
`strongly dependent upon its transformation temperatures [1,2]. Several implant devices, such as
`self—expanding Nitinol stents, rely on the superelastic properties of the material, which are affected
`by the Af transformation temperature of the alloy [3]. As a result, optimization of the mechanical
`properties of Nitinol for medical device applications is achieved by using various aging treatments
`to target a specific Af transformation temperature. Aging treatments in the 350 to 500° C range are
`routinely used to adjust the transformation temperatures of Nitinol medical devices [1]. Small
`changes in time and temperature have shown to have little effect on the performance of the device.
`
`311
`
`
`
`
`Edwards Exhibit 1021, p. 1
`
`

`

`
`
`312
`
`Although Nitinol’s fatigue properties have been previously investigated [4], few studies have been
`conducted on the effect of aging'on fatigue properties. Therefore, the purpose of this study is to
`assess the fatigue behavior of Nitinol tubing aged at different temperatures and times to achieve a
`25° C i2“ C Af.
`
`EXPERIMENTAL METHODS
`
`-
`
`-
`-
`
`-
`
`-
`
`As—received Ni50_gTi49'2 Nitinol tubing of 0.072 in. Outer Diameter (OD) by 0.051 in. Inner
`Diameter (ID) was measured using the bendfifree recovery method [5] to determine the initial
`Af temperature prior to any heat treatments.
`6-in. pieces of the tubing were heat treated to the times and temperatures listed in Table 1.
`One sample of each group was then measured using the bend-free recovery method to
`determine the final Af temperature (see Table 1). Another sample from each group was further
`characterized by Differential Scanning Calorimetry (DSC) [6].
`To avoid sample breaking near or inside the grip during the fatigue test, the outside diameter of
`the tubes was electropolished such that the gauge section had a slightly smaller outside
`diameter of 0.071 inches [7].
`The mechanical testing was performed on a MTS 858 Mini Bionix test system. A 2000N
`pneumatic grip system with MTS serrated grip surfaces of dimension 38mm by 58mm were
`used in the test. The test consists of a loading and unloading sequence from 0 to 6% strain and
`then a fatigue test at 1.5% mean strain with cut-Off cycle counts of one million, i.e., if the
`samples do not break at one million cycle counts, the test is stopped. The 1.5% mean strain is
`consistent with the in viva mean strain that SMART® stents experience [8]. Alternating strain
`magnitude for fatigue tests Were chosen to be 0.2%, 0.3% and 0.4%. All mechanical testing
`was performed at 37° C using circulated warm air.
`
`Table 1 Temperature, Time and Average Final Afof the Samples Used in the Study
`
`
`
` Heat Treat Temperature C” C) Film? (min) Final Af(° C)
`
`350
`35
`23
`
`
`
`
`400
`
`8
`
`l
`
`27
`
`450
`5
`26
`
`
`25
`90
`500
`
`
`We evaluated the following properties: (1) Af temperature after each heat treatment using the bend-
`free recovery method, (2) transformation temperatures using DSC testing, and (3) fatigue life and
`loading/unloading plateaus after aging.
`
`‘
`
`
`
`i
`
`
`
`Edwards Exhibit 1021, p. 2
`
`

`

`FATLGUE PERFORMANCE OF NITINOL TUBING WITH AF 0F 25°C 313
`
`RESULTS AND DISCUSSIONS
`
`Mechanical and Thermal Properties
`
`
`600
`
`500
`
`1.5% Strain
`
`(MPa)
`
`Stress
`
`0
`
`J__
`
`I
`
`J—
`
`0%
`
`1%
`
`2%
`
`3%
`
`4%
`Strain
`
`5%
`
`6%
`
`7%
`
`Figure 1 Stress-strain graph osz'tinol tubing heat treated at 450° C tested at 37° C.
`
`The typical stress«strain responses of the tubing after the 450° C heat treatment process at 37° C is
`presented in Figure l. The 1.5% mean strain used in the fatigue experiments is indicated on the fig—
`ure. With this lower heat—treatment temperature, the tangential stiffness decreases at lower strain,
`which indicates the presence of a significant fraction of the R—phase. This is further confirmed with
`the DSC tests performed on the same heat-treated sample (Figure 2). There is a known variation
`between the bend-free recovery method and the DSC method when determining the Af temperature
`due to the influence of strain. It is important to note that the DSC results were used as reference
`only and that all Af temperatures were set using the bend~free recovery method.
`
`Samples heat treated at 500° C (Figure 3), have a noticeable increase in the tangential stiffness over
`those heat treated at 450° C. This indicates that the R—phase is less of a factor at this temperature as
`indicated in Figure 4. The 1.5% mean strain for this heat treatment corresponds to the approximate
`onset of the plateau stress.
`
`Figures 5 and 6 show the stress—strain responses of samples heat treated at 350° C and 400° C before
`and after the 106 cycle fatigue tests. As can be seen in these figures, it appears that the displacement
`cycling modified the tangential stiffness by eliminating the R—phase inflection. This indicates that
`the R—phase interacted with the fatigue-induced dislocations thereby stabilizing the structure. Addi—
`tional mechanical and DSC testing is required to verify these observations.
`
`The plateau stress as a function of the hea~ treat temperature is plotted in Figure 7. Clearly there is
`a nonlinear relationship between plateau stress and aging temperature used to obtain an Afof 25° C.
`It is interesting to note that these thermal treatments led to a plateau stress range from 300 MP3 to
`over 500 MPa.
`
`
`
`
`Edwards Exhibit 1021, p. 3
`
`

`

`
`
`314
`
`10
`
`S ,
`
`E o-
`a.
`,2#1
`
`-57
`
`m
`
`-10
`
`
`
`
`
`-15 ,
`450
`
`.
`-100
`
`.
`t
`O
`-50
`Temperature (”C)
`
`.
`50
`
`100
`
`Figure 2 DSC results ofNitinol tubing heat treated at 450° C.
`
`
`
`400
`
`350 ,
`
`300 ~
`
` 1.5% Strain I
`
`
`
`
`
`M01O
`
`
`
`Stress(MPa) MOO
`
`_x UTCD
`
`100 ,
`
`50
`
`O
`
`0%
`
`l__l_
`1%
`
`|
`2%
`
`l
`3%
`
`—L
`4%
`Strain
`
`_I_
`5%
`
`6%
`
`7%
`
`Figure 3 Stress—strain graph ofNitinol tubing heat treated 511500” C tested at 37° C.
`
`Fatigue Properties
`
`Fatigue testing 011 thin—walled tubes is nontrivial as pointed out by Tabanli at al. [9]. Sample fail-
`ures inside or ve1y close to the grips are still possible even with the proper gauge section reduction.
`Grip failures are normally attributed to excessive load from the glip; therefore these fatigue data
`
`Edwards Exhibit 1021, p. 4
`
`

`

`
`.
`100 -
`
`FATIGUE PERFORMANCE OF NITINOL TUBING WITH AF 0F 25°C 315
`
`
`
`HeatFlow(mW)
`
`—150
`
`-100
`
`—50
`
`0
`
`50
`
`100
`
`Figure 4 DSC results Nitinol tubing heat treated at 500° C.
`
`Temperature (°C)
`
`600
`
`500
`
`J:C} O
`
`Original
`
`After Fatigue Test
`
`/
`
`
`
`Stress(MPa) co0CD
`
`[‘0O C)
`
`0
`
`I
`
`I
`
`0%
`
`1%
`
`2%
`
`3%
`
`4%
`
`5%
`
`6%
`
`Strain
`
`7%
`‘
`
`Figure 5 Stress/strain graph ofNitz‘nol tubing heat treated at 350° C before and afterfatigue test.
`
`were not included in the analysis. Figure 8 summarizes the fatigue—tests results for samples that
`either broke at the test section or had run out at 106 cycle counts. There was no tubing failure with
`0.2% half—alternating strain for any of the aging temperatures. At 350°C and 400° C, there were
`two samples with run out at 0.4% and 0.2% half-alternating strain magnitudes, yet there was a
`
`‘
`
`i
`
`E
`
`
`
`1'
`
`1
`
`1
`
`Edwards Exhibit 1021, p. 5
`
`

`

`
`
`316
`
`
`
`600
`
`500 ,
`
`
`
`
`
`After Fatigue Test
`
`
`
`Original
`\
`
`
`
`Stress(MPa)
`
`AC)C
`
`0:OO
`
`MO0
`
`£00
`
`0
`
`/
`
`
`
`
`
`0%
`
`1%
`
`2%
`
`3%
`
`4%
`Strain
`
`5%
`
`6%
`
`7%
`
`Figure 6 Shem/strain graph ofNitinol tubing heat treated at 400° C before and afierfatigue test.
`
`600
`
`
`,.
`
`i
`
`01DO
`
`4:.CD0
`
`300 -
`
`
`
`
`
`
`
`
`
`LoadingPlateauStress(MP3)
`
`200
`
`
`'
`'
`'
`'
`'
`
`300
`
`350
`
`400
`
`450
`
`500
`
`550
`
`Aging Temperature ('C)
`
`Figure 7 Plateau Stress as a. function ofaging temperature for a constant Af = 25° C.
`
`break at 0.3% half-alternating strain. A repetitive test for an additional sample aged at 400° C also
`had run out at 0.2% half—alternating strain. At 450° C, two samples broke at 0.4% at the test section,
`but no break was observed at 0.2% and 0.3% half—alternating strains. At 500° C, there was a sample
`
`Edwards Exhibit 1021, p. 6
`
`

`

`I 500‘0 run out
`
`
`
`
`
`AlternatingStrainAmplitude
`
`0.35 -
`
`0.30 -
`
`0.25
`
`0.20 '
`
`0.15
`
`FATIGUE PERFORMANCE OF NlTINOL TUBING WITH AFOF 25°C 317
`
`
`
`0.45
`
`0.40
`
`O 350'C run out
`A 400'6 run out
`
`I450‘C run out
`
`
`
`104
`
`’IU6
`105
`Cycies to Failure
`
`107
`
`Figure 8 Fatigue results of all tested tube samples.
`
`with run out at 0.2% half—alternating strain; the remaining four samples broke at 0.3% and 0.4%
`half alternating strains. A summary of the total tested samples is listed in Table 2. Typical fatigue
`failures of the tested tubes are shown in Figure 9.
`
`It is very important to point out that 1.5% mean strain was chosen also to coincide with an onset of
`stress—induced martensite for 500" C heat-treatment conditions {in bending). However, at the lower aging
`conditions, 1.5% is still nominally on the linear—elastic portion of the stress—strain curve even though the
`Afwas set to 25° C for all samples. Surprisingly, these initial results indicate that there are no significant
`differences in fatigue properties even though there are differences in material microstmctures.
`
`Figures 10a and b show typical SEM photos of a broken section. These images show typical fatigue
`fracture surfaces. It is worth mentioning that the fracture surfaces are nearly identical for all the
`eight broken samples. All samples indicated that fatigue fractures initiate from the ID of the tubing.
`This implies that the electropolishing treatment on the OD of the tubing sample reduces the surface
`flaws and defects on the tubing 0D and therefore increased the fatigue resistance on the OD sur-
`face. Conversely, since the ID surface was not electropolished, it is possible that ID surface defects
`may have played a role in decreasing the fatigue life of the specimens and may have contributed to
`some of the scatter in the data.
`
`CONCLUSIONS
`
`The scatter in the fatigue data makes it difficult to compare the fatigue characteristics of the tubing
`at different heat—treatment temperatures. Yet the data indicate that 0.2% seems to be the one million
`cycle fatigue endurance limit for the tubing aged at any temperature between 350°C and 500° C,
`consistent with previously reported tension~tension fatigue results for Nitinol microtubing [9].
`Although there are insufficient data to draw statistical conclusions from this preliminary investiga-
`
`Edwards Exhibit 1021, p. 7
`
`

`

`
`
`318
`
`Table 2 Fatigue Testing Summary
`
`Heat Treat Temperature (° C)
`
`Alternating Sttain (%)
`
`Cycles to Failure
`
`
`
`
`
`
`
`
`Run out
`
`
`Run out——
`
`—_
`—_
`——
`
`
`9.7 x 105
`
`
`
`
`
`
`
`
`
`
`
`
`
`0.4
`
`0.2
`
`6.8 x 104
`
`4.3 x 104
`
`4.6 x 104
`
`9.9 x 105
`
`tion, it is noteworthy that there did not appear to be any major differences in the samples tested
`despite the differences in material’s microstructure and therefore in plateau strength. As such, it
`may be speculated that the magnitude of the alternating load is less important than the alternating
`strain as reported by Gong et al [10]. Furthermore, it is important to note that all the fatigue frac—
`tures occurred on the ID of the tube samples, which was not surface treated. This may have played
`a role in the data scatter that was observed. Due to the time limitation and the complexity of the
`nature of fatigue, it is too early to draw any further conclusions from the results presented in this
`paper. Additional tests are in progress and the results will be analyzed accordingly to address the
`fatigue tolerance of the tubing aged through different heat treatments.
`
`It is also interesting to point out the effects of cyclic loading on the mechanical properties as
`shown in Figures 5 and 6. This can lead to different responses of a device during its service life,
`which is clearly an important design consideration for implantable devices. Therefore among the
`
`Edwards Exhibit 1021, p. 8
`
`

`

`
`
`_ _.—— 7
`.QSD IDIJIIm once iE‘ 4s: 35]
`
`..
`1‘
`
`_
`._._‘ '7
`.9 sebum Euflfl’ihl
`
`7
`4-:- e-EI
`
`Figure 10 (a) SEM image offiucture surface ofthe 350° C sample and (17) SEM image offracture
`swface of the 350° C sample
`
`heat treatment temperatures evaluated, 500° C produces the smallest amount of R—phase with the
`most fatigue-stable microstructure and is the recommended choice for a good reproducible
`device performance.
`
`ACKNOWLEDGMENTS.
`
`The authors would like to thank Dung Pham, Mike Connally and Rose Marie Ramirez of Nitinol
`Devices and Components for their assistance in sample preparation and mechanical testing. Special
`thanks also to Dr. Alan R. Pelton for inspirational discussions and providing the scientific material
`background knowledge.
`
`REFERENCES
`
`1. AR. Pelton, J. DiCello, and S. Miyazaki, MITAT 9, no. 2 (2000), p. 107.
`
`2. T.W. Duerig and R. Zadno, in Engineering Aspects ofShape Memory Alloys, eds. T.W. Duerig
`et at. (1990), p. 369.
`
`3. T.W. Duerig, D.E. Tolomeo, and M. Wholey, in MITAT 9, no. 3—4 (2000), p. 235.
`
`
`i
`
`!
`
`I
`
`
`
`Edwards Exhibit 1021, p. 9
`
`

`

`320
`
`9024.9?"
`
`10.
`
`S. Miyazaki, in Engineering Aspects ofShape Memory Alloys, eds. T.W. Duerig er al. (1990),
`p. 394.
`
`F 2082-01, Annual Books ofASTM Standards.
`
`F 2004-00, Annual Books ofASTM Standards.
`
`E8M—Olel, Annual Books ofASTM Standards.
`
`X.»Y. Gong and AR. Pelton, in SMST-2003: Proceedings of the International Conference on
`Shape Memory and Snperelastic Technologies, eds. AR. Pelton and T. Duerig (Pacific Grove,
`Ca]if.: International Organization on SMST).
`
`R.M. Tabanli, N.K. Simha, and ET. Berg, Mater. Sci. and Eng. A 273—275 (1999), p. 644.
`
`X.-Y. Gong et al., in SMST—2003: Proceedings ofrhe International Conference on Shape
`Memory and Snperelostic Technologies (Pacific Grove, Ca1if.: International Organization on
`SMST), in press.
`
`
`
`
`
`
`
`Edwards Exhibit 1021, p. 10
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket