throbber

`
`‘
`‘
`
`Dissolution Theory,
`Methodology, and
`Testing
`
`Edited by
`Anthony Palmieri Ill
`
`
`
`
`
`
`Dissolution Technologies, Incorporated
`
`Hockessin, Dela ware
`
`ENDO - Ex. 2021
`
`Amneal v. Endo
`
`|PR2014-00360
`
`ENDO - Ex. 2021
`Amneal v. Endo
`IPR2014-00360
`
`

`

`
`
`DISSOLUTION THEORY, METHODOLOGY, AND TESTING
`
`Disclaimer
`
`No responsibility is assumed by the Publisher for any injury and/or damage to perat'ms or property as
`a matter of products liability, negligence or otherwise, or from any use or operation of any methods,
`products, instructions or ideas contained in the material herein. The laboratory procedures described
`herein may involve hazardous material. The authors, speaking for themselves. expressly disclaim any
`liability to users of these lab procedures and apparatus for consequential damages of any kind arising
`out of or connected with their use. Items of apparatus and test methodology described in this book are
`intended to illustrate proper techniques to obtain a quality analysis and are not to be considered official
`and/or required. Official compendia and regulatory requirements are to be considered the only author—
`itative source for such matters.
`'
`
`\
`
`,|_
`1
`
`.
`
`I
`
`
`
`
`
`
`
`All rights reserved. No part of this book may be reproduced or used in any form or by any means
`
`whatsoever without the permission of the publisher except in the case of brief quotations embodied in
`
`critical articles or reviews.
`
`First Edition 2007
`
`
`Library of Congress Control Number 2007929967
` Edited by Anthony Palmieri III.
`
`
`
`Copyright © 2007 by Dissolution Technologies. Inc.
`
`Printed in the United States of America
`
`
`
`Dissolution Theory, Methodology, and Testing
`Includes bibliographical references and index
`ISBN 0-9761519-1-X (hardcover)
`Cover design by Allen Press, Incorporated
`
`
`Published by Dissolution Technologies, Incorporated
`9 Yorkridge Trail, Hockessin, DE 19707 USA.
`www.dissolutiontech.com
`
`
`
`
`

`

`32
`
`20.
`
`21.
`
`22.
`
`23.
`
`24.
`
`25.
`
`26.
`
`27.
`
`28.
`
`29.
`
`30.
`
`31.
`
`32.
`
`DISSOLUTION THEORY, METHODOLOGY, AND TESTING
`
`Gray, V. A.; Dressman, J. B. Change of pH Requirements for Simulated In-
`testinal Fluid TS. Pharm. Forum 1996, 22 (1), 1943—1945.
`Noory, C.; Tran, N.; Ouderkirk, L.; Brown, 8.; Perry, 1.; Lopez, 1.; Colon, M.;
`Faberlle, M.; Henry, K.; Rorberg, J.-; Ali, S. N.; Shah, V. Rethinking the Use
`of Water as a Dissolution Medium. Dissolution Technol. 1999, 6 (4), 6—7.
`Leeson, L. J. Some Observations on ‘Rethinking the Use of Water as a Dis-
`solution Medium.’ Dissolution Technol. 2000, 7 (2), 16—17.
`Kaniwa, N. Japanese Perspectives on Pharmaceutical Product Release Rate
`Testing. Drug Inform. J. 2002, 36, 407—415.
`Morihara, M.; Aoyagi, N.; Kaniwa, N.; Kojima, S.; Ogata H. Assessment of
`gastric acidity of Japanese subjects over the last 15 years. Biol. Pharm. Bull.
`2001, 24 (3), 313—315.
`Acetaminophen and Codeine Phosphate Capsules. Pharm. Forum 1998, 24 (1),
`5436.
`
`Dissolution <711>. In United States Pharmacopeia XIX; The United States
`Pharmacopeial Convention, Inc.: Rockville, MD, 1975; p 651.
`Hofer, J. D.; Gray, V. A. Examination of Selection of Immediate-Release Dis-
`solution Acceptance Criteria. Pharm. Forum 2003, 29 (1), 335—340.
`Cabana, B. E.; O’Neill, R. FDA’s Report on Drug Dissolution. Pharm. Forum
`1980, 6(1), 71—75.
`Sarapu, A. C.; Lewis, A. R.; Grostic, M. F. Analysis of PMA Collaborative
`Studies of Dissolution Test Calibrators. Pharm. Forum 1980, 6 (2), 172—176.
`Pharmaceutical Research and Manufacturers of America (PhRMA),- Dissolution
`Test Working Group. USP Dissolution Calibrator Tablets: Recommendations
`for Reduced Calibration Testing. Pharm. Forum 1997, 23 (3), 4238—4242.
`Pharmaceutical Research and Manufacturers of America (PhRMA), Subcom-
`mittee on Dissolution Calibrators Dissolution Calibration. Recommendations
`
`for Reduced Chemical Testing and Enhanced Mechanical Calibration. Pharm.
`Forum 2000,26 (4), 1149— 1166.
`Mirza, T.; Grady, L. T.; Foster, T S. Merits of Dissolution System Suitability
`Testing: Response to PhRMA's Proposal on Mechanical Calibration. Pharm.
`Forum 2000, 26 (4), 1167—1169.
`
`
`
`Dissolution Equipment
`
`G. Bryan Crist
`
`DISSOLUTION APPARATUS OVERVIEW
`
`Over the past forty years, two basic techniques have evolved for in vitro dis-
`solution testing, the stirred beaker method and the flow-through procedure.
`The stirred beaker system places the test specimen and a fixed volume of fluid
`in a vessel, and stirring provides mechanical (hydrodynamic) agitation. This system
`was adopted as the official dissolution method in USP XVIII in 1970 and described
`as the rotating basket method, USP Apparatus 1.
`The rotating paddle method was adopted as an official dissolution method by
`the USP. several years later and became USP Apparatus 2. The origin of official
`equipment developed from a number of basket and stirring devices (I) is shown
`in Figure 1.
`Some of the needs for the flow--through type apparatus included a change of pH
`or any other change in the dissolution medium. Difficulties had also arisen for a
`number of sparingly soluble drugs, which were difficult to investigate with a limited
`volume of media. The flow-through system was first adopted by the Deutscher
`Arzneimittelcodex (German Pharmaceutical Codex, DAC) in 1981.
`Two flow-through apparatus were eventually added to the USP in 1990 to over-
`come some of the experimental difficulties from the use of the single vessel meth—
`odology. They became known as USP Apparatus 3, Reciprocating Cylinder, and
`USP Apparatus 4, Flow-Through Cell.
`The most common dissolution apparatus used throughout the world are the has-
`ket and the paddle. These methods are simple and robust and are generally flexible
`enough to allow dissolution testing for a wide variety of drug products. For this
`reason, Apparatus 1 and 2 should be used for dissolution method development
`unless shown to be unsatisfactory. Other in vitro dissolution apparatus such as the
`reciprocating cylinder and the flow-through cell system described in the USP may
`be considered, if needed.
`More drug release equipment, USP Apparatus 5 and 6, deal with transdermal
`systems. USP Apparatus 7 (Alza—type) was developed for the analysis of transder—
`
`33
`
`
`
`

`

`34
`
`DISSOLUTION THEORY, METHODOLOGY, AND TESTING
`
`Dissolution Equipment
`
`35
`
`Dissolution, Bloavallablllty and Bloequlvalence
`
`scribed in pharmacopeia throughout the world, but for purposes of discussion, ref-
`erences will be made to the USP regarding specifications and operation.
`
`Rotating Basket
`
`A general description of the rotating basket apparatus consists of shaft and basket
`component fabricated from 316 stainless steel. Unless otherwise specified in a lest
`method or official monograph, a 40-mesh basket is used, The basket shaft assembly
`containing the product is lowered into a IOOO-mL vessel and rotated at a specific
`speed within media. which is maintained at a specific temperature. The rotating
`basket method is routinely used for capsule formulations at an agitation speed of
`50—100 rpm. Rates outside a range of 50—150 rpm are generally unacceptable
`because of irreproducibility associated with the hydrodynamics below 50 rpm and
`turbulence above 150 rpm. High turbulence in the vessel leads to a loss of dis-
`criminatory power associated with the dissolution method.
`The vessel used for Apparatus 1 and 2 is typically a 1000-mL hemispheric—
`shaped vessel made of glass or suitably inert material. Vessel volume should be
`between 500 and 1000 mL with 900 mL used historically. One-liter vessels are
`tubing-based with dimensions of 98406 mm i.d. and loll—210 mm in height.
`Larger vessels have been developed over the years to allow more volume for poorly
`soluble compounds. The USP Z—L vessel has dimensions of 98—106 mm i.d. and
`280—300 mm in height. The USP 4—L vessel has dimensions of 145—155 mm i.d.
`and 280—300 mm in height.
`The official basket used for Apparatus 1 is a 40-mesh design, meaning there are
`40 openings per linear inch. Openings are equal in both directions producing a
`standard square weave. USP specifies that 40—mesh (40 X 40) screen be manufac-
`tured with wire having a 0.010” diameter. Dissolution baskets are fragile and require
`proper handling and care’. Attachment to or removal from Ihe basket shaft requires
`holding the upper rim. When not in use, baskets should be stored in protective
`cases. Baskets should be carefully inspected for damage or excessive wear since
`defective or misshaped baskets will affect test results. The standard 40-mesh basket
`with a 0.01" wire size results in a 0.381-mm square aperture. For comparison, the
`Japanese Pharmacopeia specifies U.Ul.|” wire diameter resulting in 0,425-mm
`square hole resulting in a 36-mesh basket. The baskets are not interchangeable and
`could result in a 24% difference in USP. El”. and JP baskets when testing the USP
`disintegrating calibrator, Prednisone.
`
`Current Physical Parameters and Tolerances:
`Wobble
`:1 mm basket lower rim
`Dimensions
`per USP
`Height
`25 i 2 mm
`Centering
`:2 mm center line
`Speed
`:4% of set speed
`Vessel Temp.
`37 i 0.5 °C
`Timepoints
`i2% of specified time
`
`Beaker-stirrer
`Lm-Nlpdluo)
`
`Paddle-flank
`Pool- (nu)
`
`Rotutin ballot
`hrn-mllg-Uootllu)
`
`U
`
`3:053"
`
`
`
`UCJUU
`
`
`Bout bath!
`rim," title?!)
`
`
`Planor we
`Stationary basket Magnetic basket
`Gum's.)
`Sunni.” ll(l.7!) mahonflalunb'rl)
`
`Figure 1. Difi‘erent designs of dissolution vessels and stirrers that have been uti-
`lized in major nonojj‘icial methods. (Reprinted with permission from ref 1. Copy-
`right 1982 Advanstar Communications, Inc.)
`
`mal systems as well as a variety of drug release systems such as osmotic pumps
`and Implants. Because of the diversity of delivery systems and the evolving nature
`of understanding in the area of drug release, different experimental modifications
`may be needed to obtain a suitable in vivo correlation with in vitro release data.
`Alternatives or modifications to established methodology should be considered on
`the basis of proven superiority for a particular product. If the release of the active
`drug substance from an individual drug product cannot be accommodated by one
`of the major compendial apparatus, appropriate modifications have to be developed.
`However, unnecessary proliferation of alternative dissolution apparatus should not
`be encouraged due to the reproducibility problems that plagued early dissolution
`equrpment; this will also hinder regulatory acceptance.
`
`USP APPARATUS 1 AND 2—ROTATING BASKET AND PADDLE
`
`USP Apparatus l and 2 are described officially in USP Physical Test Chapter
`<711> Dissolution (2). USP Apparatus 1 is called the rotating basket apparatus
`and USP Apparatus 2 is called the rotating paddle apparatus. They are also de-
`
`
`
`

`

`36
`
`DISSOLUTION THEORY, METHODOLOGY, AND TESTING
`
`Dissolution Equipment
`
`37
`
`Q
`
`6.3 to 6.5 or
`9.4 to 10.1 rrm
`
`Vent hole
`2.0 t 0.5 m diameter
`
`Retention spring with
`a lengeon 120° centers
`
`Clear opening
`202 :1: 0.1 mm
`
`37.01:
`-3.0 rrln
`
`.
`
`—r—
`27.0
`:t: 1.0 mm
`09°"
`
`Nate—Maximum allmable
`runoutal 'A' Is 11.0mmMIen
`the pertls rotated on anie
`with basket mounted.
`
`[
`Screen 0.D.
`
`22.2 t 1.0 mm
`Screen with welded seam:
`40 X 40 mesh. 025-mm wire
`diameter with wire openings of
`0.40 :r: 0.04 mm: where 20-
`mesh screen is specified. use
`20 X 20 mesh, GAO—mm wire
`diameter with wire openings of
`0.90 d: 0.09 mm. [Nore—
`Aller welding. the screen may
`be slightly altered]
`
`
`
`
`
`
`
`Figure 2. Various dissolution vessels.
`
`In addition to the current physical parameters, some pharmaceutical laboratories
`have adopted more stringent parameters gleaned from a PhRMA Subcommittee on
`Dissolution Calibration proposal to the USP (3) to maintain a higher degree of
`control over the rotating basket apparatus:
`
`Optional Parameters and Tolerances:
`Shaft wobble
`$0.5 mm total runout
`Basket wobble
`51.0 mm total runout
`Basket exam
`No defects at time of use
`
`Shaft verticality
`Speed
`Vibration
`
`Vertical using bubble level
`:2% set speed
`$0.2-mil displacement
`
`Typical products tested with the rotating basket are capsules, tablets, floaters,
`modified—release products, beads, and suppositories. Suppository testing may be
`performed with a slotted Palmieri basket with or without glass beads.
`Several allowable variations of the standard 40-mesh basket exist including a
`basket with a gold coating 2.5—um thick (0.0001-inch). Larger vessels accommo—
`dating up to two and four liters are now allowable variations in the USP. Such
`vessels are advantageous for poorly soluble drugs.
`Some non-official variations of the rotating basket include Teflon baskets, o-ring
`
`20211.0n'm
`
`'
`
`..
`
`-- 25.0: 3.0 mm
`
`Figure 3. Official basket used for USP Apparatus I.
`
`baskets, lO-mesh through 2300-mesh (5-micron), three-fin, mini, and bolus baskets
`for veterinary products. Examples of these baskets are shown in Figure 4.
`Occasionally a small volume apparatus may be required for low-dose, high—
`potency products. Such a variation consists of a mini-basket apparatus based on
`USP Apparatus 1 with 100- or 200-mL vessels. Small volume apparatus has a
`typical operational minimum volume of 30-mL. Shown in Figure 5 are a mini
`basket and an official rotating basket.
`
`
`
`

`

`‘38
`
`DISSOLUTION THEORY, METHODOLOGY, AND TESTING
`
`Dissolution Equipment
`
`39
`
`
`
`Figure 4. Examples of non-ofi‘icial variations of the rotating basket.
`
`Rotating Paddle
`
`General Apparatus 2 Description
`
`USP Apparatus 2, the rotating paddle method, followed the development of the
`rotating basket method with better stirring characteristics. The paddle blade is fixed
`to the bottom of the shaft and rotates at a height of 25 mm from the inner bottom
`of the vessel. The paddle consists of a metallic or suitably inert, rigid blade and
`shaft composing a single entity. The paddle blade and shaft may be coated with a
`suitable inert material.
`
`The paddle is lowered into a lOOO-mL vessel and rotated at a specific speed
`within media, which is maintained at a specific temperature. The rotating paddle
`method is routinely used at an agitation speed of 25 to 75 rpm. Rates outside a
`
`
`
`Figure 5. Mini basket and ofi‘icial rotating basket.
`
`
`
`Figure 6. Typical dissolution setup.
`
`range of 25 to 75 rpm are generally unacceptable because of irreproducibility of
`the hydrodynamic effects below 25 rpm and turbulence above 100 rpm. High tur-
`bulence in the vessel leads to a loss of discriminatory power associated with the
`method. Agitation rates around 25 rpm but less that 50 rpm are acceptable for
`suspensions. For solid dosage forms with excessive coning, rotational speeds
`around 75 rpm may be necessary to improve the data. As with any variance from
`normal operating parameters, atypical conditions must be supported with data from
`normally accepted conditions for justification of USP Apparatus 2. When disso-
`lution profiles exhibit inappropriately dissolving drug substance during method de-
`velopment, adjustments outside the normal rotational speed may be warranted.
`Any time a method references USP General Chapter <7ll>, the dosage unit
`must be allowed to settle to the bottom of the vessel before rotation of the paddle
`begins. To aid the dosage unit settling to the bottom of the vessel, a small, loose
`piece of stainless steel wire consisting of a few turns may be attached to a dosage
`unit that would otherwise float.
`
`
`
`

`

`38
`
`DISSOLUTION THEORY, METHODOLOGY, AND TESTING
`
`Dissolution Equipment
`
`39
`
`
`
`Figure 4. Examples of non-ofiicial variations of the rotating basket.
`
`Rotating Paddle
`General Apparatus 2 Description
`
`USP Apparatus 2, the rotating paddle method, followed the development of the
`rotating basket method with better stirring characteristics. The paddle blade is fixed
`to the bottom of the shaft and rotates at a height of 25 mm from the inner bottom
`of the vessel. The paddle consists of a metallic or suitably inert, rigid blade and
`shaft composing a single entity. The paddle blade and shaft may be coated with a
`suitable inert material.
`
`The paddle is lowered into a lOOO—mL vessel and rotated at a specific speed
`within media, which is maintained at a specific temperature. The rotating paddle
`method is routinely used at an agitation speed of 25 to 75 rpm. Rates outside a
`
`
`
`Figure 5. Mini basket and ofi‘icial rotating basket.
`
`
`
`Figure 6. Typical dissolution setup.
`
`range of 25 to 75 rpm are generally unacceptable because of irreproducibility of
`the hydrodynamic effects below 25 rpm and turbulence above 100 rpm. High tur-
`bulence in the vessel leads to a loss of discriminatory power associated with the
`method. Agitation rates around 25 rpm but less that 50 rpm are acceptable for
`suspensions. For solid dosage forms with excessive coning, rotational speeds
`around 75 rpm may be necessary to improve the data. As with any variance from
`normal operating parameters, atypical conditions must be supported with data from
`normally accepted conditions for justification of USP Apparatus 2. When disso-
`lution profiles exhibit inappropriately dissolving drug substance during method de-
`velopment, adjustments outside the normal rotational speed may be warranted.
`Any time a method references USP General Chapter <7ll>, the dosage unit
`must be allowed to settle to the bottom of the vessel before rotation of the paddle
`begins. To aid the dosage unit settling to the bottom of the vessel, a small, loose
`piece of stainless steel wire consisting of a few turns may be attached to a dosage
`unit that would otherwise float.
`
`
`
`

`

`40
`
`DISSOLUTION THEORY, METHODOLOGY, AND TESTING
`
`Dissolution Equipment
`
`41
`
`
`
`Figure 7. Examples of Apparatus 2 paddels.
`
`Since the construction of the sinker has such an impact on the hydrodynamics
`in the bottom of the vessel, individuals have sought to standardize the USP design.
`A unique standardization utilizing cork borers was presented in the Pharmacopeial
`Forum as a stimuli article (4). The guidance suggested a method to construct sink-
`ers by hand with the use of a cork borer which would minimize the variability
`resulting from different interpretations of the construction of a USP sinker.
`In addition to sinking floating dosage forms, sinkers may assist in keeping a
`dosage form from sticking to the vessel inappropriately as in the case with some
`film-coated tablets. Sinkers must be adequately described in laboratory standard
`operating procedures to eliminate hydrodynamic variation associated with different
`sinker devices. Sinkers of other descriptions may be used if properly validated.
`Many different sinkers have evolved, some of which are based on the wire helix
`design.
`
`
`Current Physical Parameters and Tolerances:
`Wobble
`Not specified in USP
`Dimensions
`per USP
`Height
`25 t 2 mm
`Centering
`:2 mm center line
`Speed
`:4% of set speed
`Vessel Temp.
`37 i 0.5 °C
`Timepoints
`:2% of specified time
`
`As mentioned in the previous section on physical parameters for basket, some
`
`9.4 to 10.1mm diameter
`before coating
`
`
`
`
`
`
`
`
`
`
`
`
`
`NOTES—
`") Shaft and blade material.-
`303 (or equivalent)
`stainless steel.
`(2) A and 8 dimensions are
`not to vary more than
`0.5 mm when part is
`rotated on 1 axis.
`(3) Tolerances an.I $1.0 mm,
`unless otherwise stated.
`
`
`
`
`
`4 I. 5 mm radius
`
`1.2 mm radii/5V
`
`4 01 1.0 mm
`
`L— 74.0 mm to 75.0 mm -—~[
`Figure 8. USP Apparatus 2 paddle specifications.
`
`laboratories have adopted more stringent parameters for the paddle apparatus to
`maintain a higher degree of control.
`
`Optional Parameters and Tolerances:
`Shaft wobble
`50.5 mm total runout
`Paddle exam
`No defects at time of use
`
`Shaft verticality
`
`Vertical using bubble level
`
`
`
`

`

`
`
`42
`
`DISSOLUTION THEORY, METHODOLOGY, AND TESTING
`
`Dissolution Equipment
`
`43
`
`
`
`
`Figure 10. Sinkers constructed using cork borers.
`
`
`
`Figure 11. Examples of sinker designs.
`
`Speed
`Vibration
`
`:2 RPM of set speed
`$0.2-mil displacement
`
`Typical products tested by the paddle method are tablets, capsules (with sinkers),
`hydrogel tablets, suspensions, powders, microparticles, and transdermals (paddle-
`over-disk method).
`
`Calibration of USP Apparatus 1 and 2
`
`In the early 19705, scientists began to evaluate significant apparatus-to-apparatus
`differences in dissolution results. Monograph specifications could not be enforced
`due to variability in results from lab to lab and apparatus to apparatus. In 1978,
`the USP established and issued the first official dissolution calibrator tablets and
`reference standards. The primary purpose was to control vibration since most other
`parameters could be controlled by mechanical measurements.
`Since that time, dissolution apparatus used under current Good Manufacturing
`Practices (cGMPs) should be challenged with an apparatus suitability test as out-
`lined in the USP Chapter <711> Dissolution. At the time of this printing, the
`suitability test must be conducted with USP dissolution calibrators of the disinte-
`grating and non-disintegrating type, Prednisone and Salicylic Acid tablets, respec-
`tively. The word calibration is somewhat of a misnomer since the “calibrator”
`tablets do not actually calibrate anything, as a weight would be used to calibrate
`an analytical balancej At this time there is no predefined period of calibration;
`however, our current Good Manufacturing Practices as outlined in 21 CFR Parts
`210 and 211 require calibration of analytical equipment according to prescribed
`schedules. These prescribed schedules are generally established by various orga-
`nizations to limit the liability originating from test results obtained on apparatus
`
`
`
`

`

`44
`
`DISSOLUTION THEORY, METHODOLOGY, AND TESTING
`
`Dissolution Equipment
`
`45
`
`that may fall out of calibration as a result of age, environmental factors, or relo-
`cation.
`
`In the event of obtaining an outside-of—target dissolution result or a suspect result
`during calibration or routine analysis, a detailed review of the equipment, method,
`materials, and analyst documentation should take place. The investigation should
`be thoroughly documented and include all observations and explanations for the
`aberrant result by showing a cause—and-effect relationship, corrective action, and
`eventually the retest. One change at a time should be made prior to retesting to
`isolate the cause of the aberrant result. Recheck all physical parameters after any
`adjustments are made and perform retesting on a set of six tablets.
`A laboratory review checklist consisting of the following areas should be im-
`plemented to perform a thorough evaluation: check calculations, reread samples
`that were non-conforming, examine spectrophotometer and any automatic sampling
`equipment, review sampling technique, review stande preparation, review media
`preparation, Documentation of the investigation should include a description of the
`failure with a full data summary, the laboratory review checklist, description of
`the findings, corrective actions, additional physical adjustment of the apparatus,
`specific reasons for the run to be invalidated such as a crack in the vessel, and the
`retest(s).
`
`Aberrant Dissolution Data Investigation
`The most common sources of error in dissolution testing for the rotating basket
`and paddle methods are deaeration, paddles and baskets conformance, condition of
`vessels, vibration and environmental issues, sampling technique, and filtering is-
`sues.
`
`Deaeration
`
`While numerous deaeration techniques have been utilized, some are better than
`others. The USP recommends heating media to approximately 41 °C followed by
`vacuum filtration through a 0.45pm filter under vigorous stirring. After filtration,
`continue to draw the vacuum for five additional minutes. In theory, the best way
`to remove dissolved gases is through boiling, but this is a waste of energy and
`time resources. However, heating media to 41 °C and applying a vacuum can
`achieve boiling at a lower temperature. The filter is simply used to provide a
`pressure gradient. Once the media passes through the filter, air is immediately and
`efficiently stripped out. Media is then measured and gently poured into the disso-
`lution vessel and allowed to equilibrate to 37 °C in the vessel. Alternate deaeration
`techniques have been used such as helium sparging or vacuum ultrasonication.
`Critical parameters for helium-sparging methods include gas flow rate, type of
`diffuser, and time per volume. The efficiency of alternate deaeration techniques
`must be demonstrated and documented through validation.
`Historically, media has been measured and gently added to the vessel with the
`aid of a cylinder or other calibrated “to deliver” device. Media should be delivered
`with an accuracy of :l%, which currently rules out most “class A” graduated
`
`cylinders because they are calibrated in lO-mL increments and are not capable of
`measuring the i9-mL tolerance for the typical 900-mL media volume. Alternately,
`media could be measured in a “to deliver” class A volumetric flask if it has been
`calibrated at the intended measuring temperature. An alternate media-measuring
`technique employed by many automated systems is gravimetric measurement.
`Dissolution media at a controlled elevated temperature may also be weighed by
`correcting for pre-determined density. This is how many automated delivery sys-
`tems measure and transfer media to the vessel.
`
`Vibration
`
`For dissolution equipment to operate correctly, the area must be maintained free
`from excessive vibration from sources such as centrifuges, vacuum pumps, fume
`hoods, shakers, ultrasonic cleaners, and unstable bench top and construction. All
`such sources of external vibration must be eliminated. Internal sources of vibration
`may be caused by tension or dirt on drive belt, worn parts and bearings, and
`turbulence in the water bath. Make sure the deflector shield is in place in the water
`bath.
`
`Preferably use a vibration meter during calibration periods to obtain a baseline
`measurement. If suspect 01' aberrant dissolution results are obtained on the appa—
`ratus, current vibration measurements may be compared to the level obtained during
`calibration to see if this could have contributed to the suspect data. Vibration mea-
`surements should be part of the routine physical calibration of the apparatus to
`detect vibration from unwanted sources prior to obtaining data.
`
`Sampling and Filtering
`The filter is essential to stop the dissolution test by removing undissolved drug
`product as well as particulate matter and turbidity from the sample. Filters must
`be tested for drug adsorptivity to show that they do not bind drug substance. Filters
`should also be evaluated for efficiency to demonstrate that drug substance did not
`pass through the filter and continue to dissolve. Separate, clean, dry filters and
`glassware must be used when sampling each vessel. Generally, the first several
`milliliters should be discarded prior to sample collection for analysis but the specific
`amount discarded must be determined through validation. Sample aliquots must be
`filtered immediately after the sample is drawn, otherwise the dissolution process
`continues.
`
`Baskets and Paddles
`
`Basket and paddle stirring elements must be checked for USP conformance.
`Routine physical observations of baskets and shafts should be conducted to ensure
`integrity of the stirring element. Physical observation of basket and shafts should
`include that they are straight and roll evenly on a flat surface. Any Teflon coating
`must not be chipped or peeling, which adds to the turbulence in the vessel. Check
`surfaces for corrosion or discoloration due to prolonged exposure to hydrochloric
`acid. Stainless steel, while resistant to rust and corrosion, will be attacked by chlo—
`
`
`
`

`

`48
`
`DISSOLUTION THEORY, METHODOLOGY, AND TESTING
`
`Dissolution Equipment
`
`49
`
`SIationary Tablet Basket Configuration Dlagrarn
`
`
`
`Figure 13. Small-volume dissolution apparatus.
`
`purpose was to improve the mixing characteristics within the vessel, eradicate con—
`ing, and produce better stirring from a low energy system that will not cause
`particle shear. The mega paddle may also be useful for Z-L and 4-L vessels where
`greater fluid movement is required. This modified paddle has not been as widely
`accepted as the peak vessel for improving hydrodynamics within the vessel.
`
`Stationary Basket
`
`\IIIII”will!!!
`
`:'""
`
`"'
`
`The stationary basket assembly is used with traditional rotating paddle apparatus.
`This modification suspends the dosage unit contained in a basket device Just above
`
`Figure 15. Stationary basket design.
`
`
`
`Figure 14. Mega paddle.
`
`the rotating paddle. Several variations of the stationary basket have eVolved. One
`system utilizes a basket held in place by a disk with clips similar to the rotating
`basket apparatus with the exception that a 0.25-inch shaft mounts the stationary
`basket to the evaporation cover. In addition to a standard USP 40-mesh basket, 10-
`and 20-mesh baskets have been used. Another variation was introduced in the USP
`
`26 First Supplement for a Felodipine monograph. This design utilizes a quadran-
`gular basket of stainless steel wire gauze, which is suspended 1 mm above the
`rotating paddle.
`
`Qualification of Non-Compendial Equipment
`
`While no USP specifications or calibration procedure are available for small and
`large volume vessels, peak vessels, mega paddles, or stationary baskets, the per-
`tinent physical characteristics should be measured. Detailed specifications of the
`modified equipment must be documented and reliable sources for the equipment
`should be available. Apparatus should be calibrated with one—liter vessels to indi-
`cate that the apparatus is suitable under standard conditions. Physical parameters
`of height, centering, speed, wobble, and temperature need to be measured against
`current USP criteria and documented.
`
`
`
`

`

`46
`
`DISSOLUTION THEORY, METHODOLOGY, AND TESTING
`
`ride ions, which will cause pitting in the surfaces and a reduction in the wire
`diameter used in the baskets. A basket will maintain specifications of 40-mesh
`unless it is misshapen, but the basket micron rating will change over its lifetime
`due to corrosion. A non-lustrous pewter appearance is an indication that the basket
`integrity is failing. Particles may fall out of the basket too early causing lower
`results if the wire diameter is significantly reduced due to corrosion. Gold coating
`up to 2.5—p.m thick is an allowable variation for baskets to inhibit corrosion asso-
`ciated with a stainless steel basket. The basket surface must be smooth, not wrin-
`
`kled or misshapen, and must not have a frayed appearance; the basket should be
`replaced, if necessary.
`USP basket clips must be tight since loose clips impart excessive wobble. Non-
`USP baskets such as o-ring attachments without clips must be validated to show
`that there is no change in test results. USP Prednisone calibrator tablets run in
`o-ring baskets have exhibited up to a 10% suppressed result over the USP clip-
`type basket.
`Regarding air bubbles, several observations should be made when starting a
`dissolution run. Bubbles occasionally form underneath the disk and sometimes hold
`a tablet in the upper portion of the basket and do not allow one side of the dosage
`unit to contact the media for several minutes. If this is not noticed, a non-disinte-
`
`grating dosage form similar to the salicylic acid calibrator tablet will produce results
`on the low side of the expected range. Bubbles that form under a basket will alter
`the performance of the basket and cause failures since dissolution media will not
`circulate through the basket properly. Bubbles forming in the mesh will alSo change
`the characteristics of the basket by blocking the openings and virtually changing
`the mesh of the basket. The latter condition is usually a result of poor media
`deaeration.
`/
`
`Vessels
`
`Dissolution vessels should be serialized or numbered and maintained in their
`
`original positions. This reduces the opportunity for a defective vessel to be moved
`from one apparatus to another, causing random failures. The inner surface of vessels
`should be routinely checked for irregularities, scratches, cracks, pits, and uneven-
`ness or surface aberrations. Vessels should be acquired from a reputable manufac-
`turer since the inner surface must have a defect—free hemispheric bottom. Vessels
`manufactured with poor quality control will exhibit shallow, protruding, or asym-
`metrical bottoms, which will greatly affect the dissolution results due to increased
`turbulence.
`
`Vessels need to be checked for cleanliness. Scum, film, or sticky residues build
`up over time and can greatly affect dissolution rates. Vessels must be scrupulously
`clean. Tablets are to be dropped into non-rotating medium and allowed to settle to
`the bottom of the vessel prior to starting the rotation of paddles. Observe the dosage
`unit after introduction and record unusual observations such as sticking, floating,
`bubbles, and irregular shape of the cone.
`
`
`
`Dissolution Equipment
`
`47
`
`
`
`Fig

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket