`Instructions for Use
`Bedienungsanleitung
`Instructions d’utilisation
`Istruzioni per l'uso
`Instrucciones de uso
`Bruksanvisning
`Instruções de Utilização
`Gebruiksaanwijzing
`√‰ËÁ›Â˜ ÃÚ‹Û˘
`
`GB
`
`D F I E S P N
`
`L
`GR
`J
`CHI
`
`Page 2
`
`

`

`9' 0CR
`
`//’
`
`h:\
`",0.
`{ECHDFmon
`imumhindsum
`(E mm,
`mum R
`sninmr. 876 "Hz

`
`1
`
`5
`
`2
`
`6
`
`Position 1
`Position 1 \
`
`Position 2
`Position 2
`
`-
`
`~
`
`9
`
`10
`
`
`
`13
`
`Page 3
`
`Page 3
`
`

`

`3
`
`7
`
`4
`
`8
`
`
`
`(cid:3)
`
`(cid:2)
`
`(cid:4)
`
`11
`
`12
`
`Page 4
`
`Page 4
`
`

`

`GB
`
`Harmony • Instructions for Use
`
`Table of Contents
`
`1.0 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
`
`1.1 Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
`
`1.2 Patient Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
`
`2.0 Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
`
`2.1 Replacement Parts and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
`
`3.0 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
`
`4.0 Installing the Harmony 4R134, 4R135, 4R144 and 4R150. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
`
`4.1 Installing the Barbed Socket Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
`
`4.2 Attaching Socket and Adapter to Harmony . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
`
`4.3 Attaching Pylon to Harmony . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
`
`4.4 Connecting the Vacuum Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
`
`5.0 Installing the Harmony 4R133 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
`
`6.0 Adjusting the Harmony 4R144 and 4R150 Elastomer Rod Shock Absorber . . . . . . . . . . . . . . . . . . . . . . . . . 5
`
`7.0 Adjusting the Air Shock in the 4R133, 4R134, 4R135.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
`
`8.0 Monitoring the Elevated Vacuum Environment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
`
`8.1 Verifying the System if Vacuum Loss Occurs & Flow Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
`
`9.0 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
`
`10.0 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
`
`Otto Bock HealthCare
`
`1
`
`Page 5
`
`

`

`GB
`
`Harmony • Instructions for Use
`
`1.0 Application
`The Harmony Pump is a component of the Harmony Elevated Vacuum System, which also includes a Total Surface
`Weight Bearing Socket, Polyurethane Liner, Air Wick, and Sealing Sleeve. The Harmony is to be used solely for the
`prosthetic fitting of the lower limbs.
`For Patient Weights of 50 to 100 kg. (110 to 220 lbs.)
`• Harmony 4R133 Pump for use in some prosthetic feet (Fig. 5)
`• Harmony 4R134 34mm Tube Clamp Adapter (Fig. 3)
`• Harmony 4R135 Female Pyramid Receiver Adapter (Fig. 4)
`• Harmony 4R144 Integrated Male Pyramid Connector and 30 mm Distal Tube Clamp Adapter (Fig. 1)
`For Patient Weights of 100 to 150 kg. (220 – 330 lb.s)
`• Harmony 4R150 Female Pyramid Receiver Adapter (Fig. 2)
`
`1.1 Safety Considerations
`• The Harmony Pump should be examined for proper function during routine check-ups by qualified personnel.
`• The Harmony Pump may only be opened and repaired by an authorized Otto Bock Technician.
`• The patient should be properly instructed in the use of the Harmony Pump and Harmony System as outlined
`in the Patient Information Section.
`
`1.2 Patient Information
`• Handle the Harmony Pump carefully and examine often for any signs of damage.
`• Work with your prosthetist to make sure you understand how the Harmony System feels when functioning
`properly, when loss of vacuum occurs, and what the overall effects of applying an elevated vacuum to
`the limb.
`• Read section 8.0 of this manual to understand how to perform basic system checks in case of loss of
`the vacuum.
`
`IMPORTANT: Proper maintenance and care require flushing 1 to 2 ounces of distilled water (about 5 to 10 pumps)
`through the Harmony® system after contact with salt water or perspiration.
`
`2.0 Components
`The Harmony 4R144 and 4R150 pumps are delivered with the following components
`1. Harmony Pump
`2. Vacuum Gauge Assembly
`3. Vacuum Line, Neoprene Boot, (2) Tube Collars and (1) Straight- and (1) 90˚-Barbed Stainless Steel Socket
`Connectors
`4. Instructions for Use SL=2900-4005
`The Harmony 4R133, 4R134, 4R135 pumps are delivered with the following Components
`1. Harmony Pump
`2. Hand Pump
`3. Tube Collar and (1) Straight- and (1) 90˚-Barbed Stainless Steel Socket Connectors
`4. Instructions for Use SL=2900-4005
`
`2
`
`Otto Bock HealthCare
`
`Page 6
`
`

`

`GB
`
`Harmony • Instructions for Use
`
`2.1 Replacement Parts and Accessories
`Harmony components can be ordered individually from Otto Bock using the following information.
`Pc.
`Description
`Otto Bock TEC Part No.
`1
`Air Spring Pump Kit (for all units) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Y352
`1
`Vacuum Line, 5 feet (for all units) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Y309
`10pk
`Tube Collars (for all units) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Y310
`1
`Tube Clamp Bar (for 4R134) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Y323
`1
`Tube Clamp Screw, M5 x 22 Grade 12.9 (for 4R144) . . . . . . . . . . . . . . . . . . . . . . . 501Z2=M5X22
`1
`34-30 mm reducing Split spacer (for 4R134). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Y324
`1
`Valve Kit (for 4R134, 4R135) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4R138=5
`5pk
`Valve Kit, rubber air valves (for 4R144, 4R150) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Y351-5
`1
`Elastomer Rod, 90 shore A, yellow (for 4R150) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Y348
`1
`Elastomer Rod, 80 shore A red (for 4R144). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Y347
`1
`Intake Valve with screen (for 4R144, 4R150) (Fig. 1, Pos.1) . . . . . . . . . . . . . . . . . . . . . . . . . 4Y346
`1
`Expulsion Valve (for 4R144, 4R150) (Fig.1, Pos.2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Y345
`1
`1/8˝ Barbed Socket Connector/Filter (all units) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Y344
`1
`1/8˝ Barbed Socket Connector 90˚ Elbow Connector/Filter (all units) . . . . . . . . . . . . . . . . . . . 4Y350
`3pk
`Harmony Protective Boot (for 4R144, 4R150) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Y319=3
`3pk
`Harmony Protective Boot (for 4R134, 4R135) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Y319
`3pk
`Harmony Protective Boot (for 4R133) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Y320
`1
`Harmony Complete Installation Kit (4R133, 4R134, 4R135) . . . . . . . . . . . . . . . . . . . . . . . . . . 4R130
`1
`Harmony Testing Kit, 4R133, 4R134, 4R135 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4R131
`1
`Harmony 4R144, 4R150 Service Kit
`3.0 Description
`The Otto Bock Harmony Pump combines the function of a vacuum pump and the function of a damping shock
`absorber in one unit. The 4R144 and 4R150 include the additional features of axial rotation and torsional shock
`absorption.
`Function of the Vacuum Pump
`The Harmony Pump is designed to generate an elevated vacuum to increase the linkage between user and
`prosthesis, and requires a sealed environment between the user’s leg/liner, the socket, and the sealing sleeve.
`Once the environment is sealed, the Harmony Pump, which is compressed and released during normal walking,
`draws air from the socket through one-way valves and evacuates it through the pump. After multiple cycles of
`compression an elevated vacuum of 15-25 in. Hg. (508-847 millibars) is created in the sealed socket environment.
`The Harmony Elevated Vacuum Socket System:
`• Improves fluid exchange in the residual limb and helps to control daily volume fluctuations
`• Increases proprioception
`• Increases suspension to improve gait and overall function
`• Reduces forces to the residual limb and liner
`• Reduces moisture buildup on the residual limb
`Function of the Shock Absorber
`The Harmony shock absorption functions helps to reduce vertical and torsional forces in the extremities during
`walking. The result is an improvement for both the function and loading situation of the prosthesis wearer.
`Otto Bock HealthCare
`
`3
`
`Page 7
`
`

`

`GB
`
`Harmony • Instructions for Use
`
`Vertical and Rotational Shock Absorption in the 4R144 and 4R150
`Vertical shock absorption can be adjusted by the prosthetist by increasing or decreasing the compression of the
`elastomer rod. Rotational shock absorption is made possible by elastomer plates, which are not field serviceable.
`Vertical Shock Absorption in the 4R133, 4R134 and 4R135 can be adjusted by increasing or decreasing the
`amount of air pressure in the unit.
`4.0 Installing the Harmony 4R134, 4R135, 4R144 and 4R150
`Note: Failure to use the Harmony with a properly fitting Harmony socket system can compromise the limb, liner,
`and the Harmony pump.
`
`4.1 Installing the Barbed Socket Connector
`Note: To fabricate an airtight, laminated socket, blister-form a 3 mm PETG plastic sheet (575=1) as the first layer
`in the definitive socket lay-up.
`To ensure an airtight system you must create an airtight channel between the Barbed Socket Connector and the
`interior of the socket. We recommend the following technique.
`Prepare a channel in the socket by drilling a 3/8” (10 mm) hole in either the distal end or the posterior socket wall.
`Fill with five minute epoxy. Once epoxy has hardened, drill a 7/32 (5.5 mm) hole through the middle of the epoxy
`(Fig. 6). Use the Barbed Socket Connector to tap the hole, then remove and apply 5 minute epoxy to the threads,
`reinsert and hand tighten.
`
`4.2 Attaching Socket and Adapter to Harmony
`Pyramid Adapter style: Harmony 4R144
`Tighten the set screws of the receiver onto pyramid and torque to 15NM (or to the manufacturer’s specification if
`not an Otto Bock component). Use Loctite® 242 on screws for final adjustment.
`4-Bolt Pattern Style: Harmony 4R134, 4R135 and 4R150
`The Harmony 4R134, 4R135 and 4R150 pumps have a 4-bolt pattern on top for connecting the appropriate male
`or female adapters to the socket (i.e. 4R77, 4R23 or 4R51). Use the provided Cap Head Screws and connect the
`adapter to the Harmony. For final adjustment, apply Loctite® 242 to screws and torque to 115 in−lbs (13NM).
`
`4.3 Attaching Pylon to Harmony
`30mm Tube Clamp (4R144): Insert the 2R2 pylon into tube completely. Torque the tube clamp screw to 88 in-lbs
`(10 Nm) for the final adjustment.
`Female Pyramid Receiver Style (4R135 and 4R150): Tighten the set screws of the pyramid receiver onto the
`attachment and torque to 133 in-lbs (15Nm.). Use Loctite® 242 on screws for final adjustment.
`34mm Tube Clamp (4R134): Insert 2R76 pylon into tube clamp completely. Torque the tube clamp screw to 44
`in-lbs (5 Nm) for the final adjustment.
`
`4.4 Connecting the Vacuum Line
`To connect the Vacuum Line to the socket, cut line to length proximally, moisten end of line, slide Retention Collar
`over Line, attach Line to Barbed Socket Connector, and slide Retention Collar into place (Fig. 7).
`5.0 Installing the Harmony 4R133
`5.1 To remove original shock, remove the cap head screw from the housing of the foot with a 4mm Allen wrench.
`5.2 Push the pin through the shock and remove shock.
`5.3 Repeat steps 5.1 and 5.2 for the other end of the shock.
`5.4 Remove the original shock from the housing of the foot.
`5.5 Remove the washers from the original shock.
`4
`
`Otto Bock HealthCare
`
`Page 8
`
`

`

`GB
`
`Harmony • Instructions for Use
`
`5.6 Remove the bushings from the original shock and insert bushings in the Harmony 4R133.
`5.7 Re-use the original washers on the proximal end of the Harmony 4R133.
`5.8 Place provided washers supplied with Harmony 4R133 on both sides of the distal bushings.
`5.9 Install the Harmony 4R133 in the foot housing.
`5.10 Push the pins through the bushings.
`5.11 Reinstall the cap head screws in the pin and tighten with a 4 mm Allen wrench.
`5.12 To connect the Vacuum Line to the socket, see section 4.4.
`6.0 Adjusting the Harmony 4R144 and 4R150 Elastomer Rod Shock Absorber —
`PRACTITIONERS ONLY
`The preload on the elastomer rod in these units can be adjusted to accommodate different patient weights. This
`function should be adjusted for the patient’s normal gait. In order for the vacuum pump to function properly, the
`Harmony 4R144 and 4R150 must be adjusted so that the shaft extends completely when unweighted (with no air
`pressure in the air chamber), and so that it compresses between 5 and 7 mm when weighted.
`To replace the Elastomer Rod:
`1. Locate Blue Cup inside shaft (Fig. 8). Screw in completely using a 3/8” (10mm) Allen wrench
`2. To set Elastomer Rod tension, release pressure on the Rod by backing out the Blue Cup. See suggested
`starting points below. Because the resistance of the elastomer rods can change slightly over time, we
`recommend it be readjusted during routine checkups.
`4R144 only (Red Elastomer Rod)
`Patient Weight (lbs / kgs)
`# of turns out (counterclockwise) from fully inset position
`120 lb. / 50kg. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5
`140 lb. / 60 kg. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.0
`160 lb. / 70 kg. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.5
`180 lb. / 80 kg. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.0
`200 lb. / 90 kg. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5
`220 lb. / 100 kg . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.0
`4R150 only (Yellow Elastomer Rod)
`Patient Weight (lbs / kgs)
`# of turns out (counterclockwise) from fully inset position
`220 lb. / 100kg. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5
`240 lb. / 110 kg. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.0
`260 lb. / 120 kg. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.5
`280 lb. / 130 kg. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.0
`300 lb. / 140 kg. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5
`320 lb. / 150 kg. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.0
`
`Important: Do not unscrew more than 5 turns, as this may prevent the shaft from extending completely and
`compromise the vacuum pump function.
`
`7.0 Adjusting the Air Shock in the 4R133, 4R134, 4R135 — USERS AND PRACTITIONERS
`Resistance in the Air Shock can be modified with a hand pump when high-level activities are encountered (Fig. 9).
`Please be aware that increased resistance in the Air Shock may limit the travel of the pump piston, requiring more
`compression cycles for you to attain the desired elevated vacuum during normal walking.
`Note: When adjusting pressure in the Air Shock, we recommend that you begin at 0 psi as a starting point. Please
`see section 9.0 for recommended air pressure.
`Otto Bock HealthCare
`
`5
`
`Page 9
`
`

`

`GB
`
`Harmony • Instructions for Use
`
`8.0 Monitoring the Elevated Vacuum Environment
`A Vacuum Gauge Assembly is included with the 4R144 and 4R150 Pumps.
`We recommend purchasing the 4R130 Harmony Complete Installation Kit, or the 4R131 Harmony Testing Kit for
`monitoring and repair needs for the 4R133, 4R134, 4R135 units.
`To verify that the Harmony System can maintain an elevated vacuum environment, it is necessary to temporarily
`install a vacuum gauge. To do this, remove the Vacuum Line from the Harmony Intake Valve (Fig. 10). Next, take
`the Vacuum Gauge Assembly and attach the Vacuum Line (short side) to the Inlet Valve (Fig.1, Pos. 1) and
`connect the socket-side Vacuum Line to the open barb on Vacuum Gauge Assembly (Fig.11).
`While the patient is standing and wearing the properly fitting Harmony System, cycle the Harmony Pump to
`generate a vacuum between 15-25 in. Hg. (508-847 millibars). Stop cycling the pump and verify that the vacuum is
`maintained when it is not being cycled. If the vacuum is maintained, the system is functioning correctly. If vacuum
`loss does occur, it will be necessary to check the system as shown in Section 8.1.
`
`8.1 Verifying the System if Vacuum Loss Occurs – See Flow-Chart
`To determine if the vacuum loss exists in Zone 1, 2 or 3 (Fig. 12) install gauge as described in section 8.0 (Fig. 11).
`Cover the hole on the inside of the socket with a piece of tape that is not air-permeable (such as PVC tape). Cycle
`the pump to reach 15-25 in. Hg. (508-847 millibars). Check to see if vacuum level is maintained when the pump is
`not being cycled.
`
`Continue to the Flow-Chart
`
`6
`
`Otto Bock HealthCare
`
`Page 10
`
`

`

`GB
`
`Harmony • Instructions for Use
`
`FLOW CHART
`
`Vacuum maintained
`means Zone 2 and 3 are OK
`
`Vacuum NOT maintained
`
`Examine Zone 1 (socket and sleeve)
`for any leaks or damage.
`
`Problem exists in Zone 2 or Zone 3
`
`With T-Connector removed, attach gauge directly to socket side of intake valve (Fig.13) and cycle pump until 15-25 in.
`Hg. (508-847 millibars) is reached. Determine if vacuum level is maintained when the pump is not being cycled.
`
`Vacuum maintained means Zone 3 is OK
`
`Vacuum NOT maintained
`
`Intake valve is OK. Problem is probably in Zone 2
`(barbed fitting on socket, line, or channel in socket).
`Examine for leaks or damage.
`
`Problem exists in intake Valve, line to Harmony pump, or in
`Harmony Pump.
`
`Exchange Intake Valve and check again.
`
`Vacuum maintained. Replace valve.
`
`Vacuum NOT maintained
`
`Replace Tubing and test again by cycling pump until 15-25 in. Hg. (508-847 millibars) is reached. Determine if vacuum
`level is maintained when the pump is not being cycled.
`
`Vacuum maintained. Replace tubing.
`
`Vacuum NOT maintained
`
`Problem with Harmony Pump. Please call Otto Bock support Staff.
`
`Note: To clean the Barbed Socket Connector: Disconnect the tubing from the Barbed Socket Connector and gently
`blow air through the fitting back into the socket.
`
`Otto Bock HealthCare
`
`7
`
`Page 11
`
`

`

`GB
`
`Harmony • Instructions for Use
`
`9.0 Technical Data
`Harmony-4R133
`Weight: .68 lbs/309 grams
`Weight Limit: 220 lbs/100 kgs
`Air spring operating psi: 40-120 psi/1.35-4.06 bar
`Material: Aluminum
`Harmony-4R134
`Weight: 1.6 lbs/726 grams
`Height: 5 1⁄8 in/130 mm
`Diameter: 3.2 in/81mm
`Weight Range: 110-220 lbs/50-100 kgs
`Air spring operating psi: 0-20 psi/0-.68 bar
`Tube clamp screw: 44 in/lbs (5 Nm)
`Adapter 4-Hole pattern: 120 in/lbs (13 Nm)
`Material: Aluminum/Titanium
`Harmony-4R135
`Weight: 1.6 lbs/726 grams
`Height: 4 5⁄8 in/117 mm
`Diameter: 3.2 in/81mm
`Weight Range: 110-220 lbs/50-100 kgs
`Air spring operating psi: 0-20 psi/0-.68 bar
`Receiver Set Screw Torque: 133 in/lbs (15 Nm)
`Adapter 4-Hole pattern: 120 in/lbs (13 Nm)
`Material: Aluminum/Titanium
`Harmony-4R144
`Weight: 1.1 lbs/520 grams
`Height: 5 1⁄8 in/130.6 mm
`Weight Range: 110-220 lbs/50-100 kgs
`Tube Clamp Screw Torque: 88 in/lbs (10 Nm)
`Air spring operating psi: 0-20 psi/0-.68 bar
`Material: Aluminum/Stainless Steel
`Harmony-4R150
`Weight: 1.4 lbs/640 grams
`Height: 5.0 in/127 mm
`Weight Range: 220-330 lbs/100-150 kgs
`Receiver Set Screw: 133 in/lbs (15 Nm)
`Adapter 4-Hole pattern: 120 in/lbs (13 Nm)
`Air spring operating pressure: 0-40 psi/0-1.36 bar
`Material: Aluminum/Stainless Steel
`10.00 LIMITED WARRANTY
`Otto Bock HealthCare (Otto Bock) warrants all of its products, to the original purchaser, to be free from defects in
`materials and workmanship. This warranty applies, subject to normal wear and tear, when the products are used as
`intended, without unapproved modifications, following all Otto Bock instructions and requirements; and when they
`are fitted by or under the direct supervision of certified/licensed practitioners who meet all Otto Bock product-
`specific training requirements for a given product.
`
`8
`
`Otto Bock HealthCare
`
`Page 12
`
`

`

`GB
`
`Harmony • Instructions for Use
`
`The duration of our Limited Warranties varies by product, but all are effective from the date of fitting to a patient.
`The duration of the Limited Warranty for some components may be doubled when incorporated into a finished
`device in which all of the components are manufactured by Otto Bock and installed in conformance with all Otto
`Bock recommendations. Otto Bock’s sole obligation under this Limited Warranty shall be to repair or replace at no
`charge or to refund the cost of the system or component thereof to the original purchaser, at Otto Bock’s sole
`discretion.
`The Limited Warranty for the Harmony P2 is 12 months, and qualifies for extension to 24 months when in finished
`devices utilizing all Otto Bock components.
`THE EXPRESS WARRANTIES SET FORTH HEREINABOVE ARE IN LIEU OF ALL OTHER WARRANTIES
`EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES OF MERCHANTABILITY OR
`FITNESS FOR A PARTICULAR PURPOSE, AND ALL SUCH OTHER WARRANTIES ARE HEREBY DISCLAIMED
`AND EXCLUDED BY OTTO BOCK.
`IN NO EVENT SHALL OTTO BOCK’S LIABILITY OF ANY KIND INCLUDE ANY SPECIAL, INCIDENTAL OR
`CONSEQUENTIAL DAMAGES, EVEN IF OTTO BOCK SHALL HAVE BEEN ADVISED OF THE POSSIBILITY OF
`SUCH POTENTIAL LOSS OR DAMAGE.
`Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation
`may not apply to you.
`
`Otto Bock HealthCare
`
`9
`
`Page 13
`
`

`

`Harmony HD addendum to the Instructions for Use
`
`1.2 Patient Information – Proper maintenance and care requires flushing 1-2 ounces of distilled water
`(about 5-10 pumps) through the Harmony system after contact with salt water or perspiration.
`
`2.0 Components – New components included in the Harmony HD are filtered straight- and 90˚ elbow
`barbed socket connectors, and a vacuum gauge assembly. Please see the testing directions in the
`Instructions for Use manual, page 6.
`
`2.1 Harmony HD Replacement parts
`Additions:
`• 4Y352 Air Spring Pump Kit
`• 4Y351-5 5 pack of replacement rubber air valves
`• 4Y350 90 degree filtered socket connection barb
`Changes:
`• 4Y348 is a Yellow elastomer rod and was article number SL=2300-7027
`Deletions:
`• SL=2300-7012 was an Elastomer cup
`• SL=2300-7023 was a1/4” gray washer
`
`5.1.1 Creating channel in the socket for the Harmony HD – please use 1/4 - 28 tap.
`
`5.5 Adjusting the Harmony HD air shock is not required. The compression of the Harmony HD
`unit is made through the elastomer rod adjustment. Adjusting the Harmony HD’s air spring is only for
`patients who require additional compression resistance during short periods of increased activity
`(possibly basketball, long-distance running, etc). The air spring pump kit article number is 4Y352.
`
`Patented: 5258037, D.429335 (U.S.), 2098945 (Canada), Patents Pending. An ISO 9001/EN 46001 Certified Company
`Harmony® is a registered trademark of Otto Bock HealthCare LP. ©2004 Otto Bock HealthCare LP. • Printed in the USA • 04030405.2 • 3/04
`
`USA 800.328.4058 • Fax 800.962.2549
`Canada 800.665.3327 • Fax 800.463.3659
`www.ottobockus.com • www.ottobock.ca
`usa.customerservice@ottobock.com • info@ottobock.canada.de
`
`Page 14
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket