throbber

`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`
`__________
`
`
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`
`__________
`
`
`
`MEDTRONIC, INC. and MEDTRONIC VASCULAR, INC.
`Petitioners
`
`v.
`
`MARITAL DEDUCTION TRUST
`Patent Owner
`
`
`__________
`
`
`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`__________
`
`
`
`PATENT OWNER’S RESPONSE
`TO PETITION FOR INTER PARTES REVIEW
`
`
`
`
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`TABLE OF CONTENTS
`
`I.
`
`II.
`
`INTRODUCTION ........................................................................................... 1
`
`BACKGROUND ............................................................................................. 2
`
`A. Dr. Rhodes’s Early Artery Repair Work ............................................... 2
`
`B.
`
`C.
`
`D.
`
`The Rhodes ’154 Patent ........................................................................ 2
`
`The ’417 Patent Under Review ............................................................. 4
`
`The Industry’s Use of Dr. Rhodes’s Ideas and the Implicit
`Acknowledgement of the Importance of its Benefits .......................... 17
`
`E.
`
`Level of Ordinary Skill in the Art ....................................................... 22
`
`III. CLAIM INTERPRETATION ....................................................................... 25
`
`A.
`
`B.
`
`“At Least One Surface” and “Said At Least One Surface”................. 25
`
`“Engagement With,” “Engaging,” and “To Tightly Engage the
`Interior Surface of the Vessel, Duct, or Lumen to Fixedly
`Secure Said Device in Place.” ............................................................. 26
`
`IV. KORNBERG DOES NOT ANTICIPATE CLAIMS 1, 2, 9, 10, AND
`13 OF THE ’417 PATENT. ........................................................................... 31
`
`A. Overview of Kornberg......................................................................... 31
`
`B.
`
`C.
`
`Fluid Flow Force Does Not Cause Kornberg’s Hooks to Engage
`at All, as They Fully Engage Upon Deployment. ............................... 34
`
`Kornberg’s Hooks do Not “Tightly Engage” Because They
`Puncture the Artery Wall. .................................................................... 36
`
`V.
`
`RHODES ’154 AND KORNBERG DO NOT RENDER CLAIMS 1,
`2, 9, 10, AND 13 OBVIOUS. ........................................................................ 39
`
`A. Kornberg is a Deficient Secondary Reference. ................................... 40
`
`B.
`
`There is No Motivation or Other Legally Sufficient Rationale to
`Combine Rhodes ’154 and Kornberg. ................................................. 42
`
`Response
`
`
`
`Page ii
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`C.
`
`D.
`
`E.
`
`F.
`
`The Understanding of the Art Taught Away From the Proposed
`Modification. ....................................................................................... 46
`
`The Obviousness Contention is Tainted by Impermissible
`Hindsight. ............................................................................................ 47
`
`Replacement of Rhodes ’154’s Protuberances With Kornberg’s
`Hooks Would Render Rhodes ’154 Unsatisfactory for its
`Intended Purpose and Change its Principle of Operation. .................. 49
`
`Secondary Considerations Demonstrate the Nonobviousness of
`the Invention. ....................................................................................... 51
`
`1.
`
`2.
`
`3.
`
`Recognition of a Problem, Long-Felt But Unmet Need,
`and Failure of Others ................................................................ 51
`
`Commercial Success ................................................................. 55
`
`Commercial Acquiescence and Licensing ................................ 55
`
`VI. CONCLUSION .............................................................................................. 56
`
`
`
`
`
`
`
`Response
`
`
`
`Page iii
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`TABLE OF AUTHORITIES
`
`Cases
`
`Atofina v. Great Lakes Chem. Corp., 441 F.3d 991 (Fed. Cir. 2006) ..................... 39
`
`Bettcher Indus., Inc. v. Bunzl USA, Inc., 661 F.3d 629 (Fed. Cir. 2011) ................ 38
`
`Bicon, Inc. v. Straumann Co., 441 F.3d 945 (Fed. Cir. 2006) ................................. 26
`
`Depuy Spine, Inc. v. Medtronic Sofamor Danek, Inc., 567 F.3d 1314 (Fed.
`Cir. 2009) ............................................................................................................ 46
`
`Elekta Instrument S.A. v. O.U.R. Sci. Int’l, Inc., 214 F.3d 1302 (Fed. Cir.
`2000) ................................................................................................................... 27
`
`In re Fine, 837 F.2d 1071 (Fed. Cir. 1988) ............................................................. 39
`
`In re Gordon, 733 F.2d 900 (Fed. Cir. 1984) .......................................................... 51
`
`Graham v. John Deere Co., 383 U.S. 1 (1966) ....................................................... 47
`
`Gubelmann v. Gang, 408 F.2d 758 (C.C.P.A. 1969) ............................................... 38
`
`Hewlett-Packard Co. v. MCM Portfolio, LLC, IPR2013-00217, Paper 10
`(Sept. 10, 2013) ................................................................................................... 44
`
`In re Kahn, 441 F.3d 977 (Fed. Cir. 2006) .............................................................. 42
`
`KSR Int’l Co. v. Teleflex Inc., 550 U.S. 398 (2007) .......................................... 42, 48
`
`Monroe Auto Equip. Co. v. Heckethorn Mfg. & Supply Co., 332 F.2d 406
`(6th Cir. 1964) ..................................................................................................... 48
`
`Oakley, Inc. v. Sunglass Hut Int’l, 316 F.3d 1331 (Fed. Cir. 2003) ........................ 44
`
`Tempo Lighting, Inc. v. Tivoli, LLC, 742 F.3d 973 (Fed. Cir. 2014) ...................... 29
`
`Trintec Indus., Inc. v. Top-U.S.A. Corp., 295 F.3d 1292 (Fed. Cir. 2002) .............. 38
`
`
`
`
`
`
`
`Response
`
`
`
`Page iv
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`
`
`Statutes
`
`35 U.S.C. § 102 ........................................................................................................31
`
`35 U.S.C. § 103 ........................................................................................................39
`
`35 U.S.C. § 316(a)(8) ................................................................................................. 1
`
`
`Rules
`
`37 C.F.R. § 42.120 ..................................................................................................... 1
`
`37 C.F.R. § 42.65 .....................................................................................................44
`
`
`Other Authorities
`
`MPEP § 2143(G) ......................................................................................................43
`
`MPEP § 2143.01(V) .................................................................................................51
`
`MPEP § 2143.03 ......................................................................................................39
`
`
`
`Response
`
`
`
`Page v
`
`

`

`I.
`
`INTRODUCTION
`
`Pursuant to 35 U.S.C. § 316(a)(8) and 37 C.F.R. § 42.120, the Patent Owner
`
`hereby provides a Response to the Petition for Inter Partes Review filed on
`
`November 6, 2013 and last corrected on November 12, 2013, and the challenges
`
`therein for which trial has been instituted by the Board’s March 25, 2014
`
`Institution Decision (Paper 15).
`
`The only challenges for trial are whether (1) U.S. Patent No. 4,562,596 (Ex.
`
`1006, herein “Kornberg”) anticipates claims 1, 2, 9, 10 and 13 of the ’417 Patent
`
`and (2) U.S. Patent No. 5,122,154 (Ex. 1008, herein “Rhodes ’154” or “’154
`
`Patent”) in view of Kornberg render the same claims obvious. As explained herein
`
`and in the accompanying declaration of Dr. James Silver, Ph.D. (Ex. 2002, herein
`
`“Silver Decl.”), both challenges fail. The anticipation challenge fails for two
`
`reasons. First, Kornberg fails to disclose that fluid flow forces cause projections to
`
`engage at all after deployment, as required by independent claim 1; instead,
`
`Kornberg’s graft’s projections fully engage to their intended position upon
`
`deployment. Second, Kornberg’s projections do not satisfy the “tightly engage the
`
`interior surface of the vessel . . .” limitation because those projections perforate all
`
`the way through the vessel wall. The obviousness challenge also fails for many
`
`reasons, including, for example, lack of motivation or other legally sufficient
`
`rationale to combine the teachings of Rhodes ’154 and Kornberg. The only
`
`Response
`
`
`
`Page 1 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`motivation put forward by the Petition is prevention of downstream migration, but
`
`that problem was unrecognized in the art until after the ’417 Patent. As a result,
`
`the obviousness challenge is tainted by hindsight, as well as other flaws.
`
`Secondary objective factors also weigh in favor of nonobviousness in this case.
`
`II. BACKGROUND
`
`A. Dr. Rhodes’s Early Artery Repair Work
`
`Dr. Rhodes, the inventor of the ’417 Patent, was a thriving vascular surgeon
`
`in Bricktown, New Jersey from the 1960s - 1990s. He spent many years of his
`
`career performing carotid endarterectomy procedures, which involved making a
`
`longitudinal incision from the common to the internal carotid artery to remove
`
`plaque from the inside of the artery. (Silver Decl. ¶ 35.) Based on his experience
`
`performing this procedure, he noticed that closure of the incision allowed for the
`
`possibility of narrowing of the artery from fibrointimal hyperplasia, resembling the
`
`stenosis that the surgery was originally performed to address. (See Ex. 2003,
`
`Rhodes, Valentine J., “Expanded Polytetrafluoroethylene Patch Angioplasty in
`
`Carotid Endarterectomy,” J. Vascular Surgery 1995, 22:724-31.)
`
`B.
`
`The Rhodes ’154 Patent
`
`By the late 1980s, Dr. Rhodes had perfected a carotid artery patching
`
`technique that minimized the effect of fibrointimal hyperplasia and restenosis,
`
`Response
`
`
`
`Page 2 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`maintaining arterial lumen diameter after the procedure. (See id.) After his
`
`successful experience in the surgical repair of carotid disease, Dr. Rhodes
`
`developed the idea for an innovative stent graft design for intraluminal treatment of
`
`carotid disease that would eventually become the Rhodes ’154 Patent. The ’154
`
`Patent focuses on solving fundamental structural problems of prior endovascular
`
`graft devices, such as being able to use a graft over long sections of tortuous
`
`human arteries to revascularize segments of stenosis or occlusion, or to bypass an
`
`aneurysm.
`
` (’154 Patent 2:67-3:2, 3:47-52, 3:62-68; Silver Decl. ¶ 37.)
`
`Specifically, the ’154 Patent teaches an endovascular bypass graft with multiple
`
`stents along its length that serve to provide structural support, while at the same
`
`time, providing improved flexibility over earlier fully-stented endovascular grafts
`
`due to the separate and spaced-apart stents. (’154 Patent Fig. 1; Silver Decl. ¶ 37.)
`
`The graft described in the ’154 Patent was intended to be secured in place against
`
`the wall by the multiple spaced apart stents, but also contemplated the use of other
`
`features to aid in holding the graft in place. (’154 Patent 7:18-36; Silver Decl. ¶
`
`39.) These other features include the use of small protuberances that project
`
`slightly outward from the surface of the graft to act as small pressure points that
`
`impact the vessel wall, which would help maintain the graft’s position, or
`
`alternatively, use of small pieces of mesh placed on the outer surface of the graft to
`
`Response
`
`
`
`Page 3 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`grow into contact with the vessel wall. (See ’154 Patent 7:18-36, Silver Decl. ¶
`
`39.).
`
`C. The ’417 Patent Under Review
`
`Dr. Rhodes continued to address problematic areas in the stent graft field
`
`after he received his ’154 Patent, shifting his focus to improvements in how the
`
`then-existing stent graft devices were being secured to the vessel wall. As Dr.
`
`Rhodes began developing what would become the subject matter of the ’417 Patent
`
`in 1995, he noticed that the prior art devices were designed with attachment
`
`mechanisms that could ultimately lead to significant problems, such as tearing of
`
`the patient’s arterial wall or perforation of the arterial wall that could potentially
`
`cause hemorrhaging or damage to surrounding organs. (’417 Patent 3:53-57;
`
`Silver Decl. ¶ 40.) Dr. Rhodes sought to provide an improved anchoring
`
`mechanism in order to prevent the problem of graft migration after deployment,
`
`while at the same time avoiding damage to the vessel wall and surrounding organs.
`
`(’417 Patent 7:17-32, 8:12-30; Silver Decl. ¶ 40.)
`
`The ’417 Patent specification specifically discusses the problem the stent
`
`graft industry was facing at the time of Dr. Rhodes’s invention and how his
`
`development would solve it. (’417 Patent 3:21-58; Silver Decl. ¶ 41.) Generally,
`
`in conventional vascular bypass surgery (the precursor to today’s intraluminal
`
`procedures) grafts are secured in place by suturing the ends of the graft to the walls
`
`Response
`
`
`
`Page 4 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`of the vessel. (Silver Decl. ¶ 41.) However, in an intraluminal endovascular
`
`procedure (like anticipated by Dr. Rhodes in the ’154 Patent), the medical
`
`practitioner inserts the graft percutaneously and does not have open surgical access
`
`to the vessel to allow for suturing. This restriction led to a need for a graft to not
`
`only be placed in the lumen of the blood vessel, but for the graft to remain where
`
`the practitioner inserted it. (’417 Patent 1:64-2:14; Silver Decl. ¶ 41.)
`
`When the first endovascular grafts were being developed, developers often
`
`believed that grafts could be held in place by tissue ingrowth into the graft and
`
`around the stents. (Silver Decl. ¶ 42.) Other practitioners believed that radial
`
`outward pressure and friction generated from a stent, or a series of stents, located
`
`along the length of the graft were adequate to hold the graft in place. This concept
`
`is referred to as passive fixation. (Id.) Still other developers believed that some
`
`type of anchoring was required which utilizes various types of staples, barbs or
`
`hooks at one or more locations along the graft to affix the graft in place. (Id.) This
`
`concept is referred to as active fixation. (Id.)
`
`The specification of the ’417 Patent specifically recognizes that various
`
`active fixation mechanisms for anchoring the stent graft to the patient’s artery wall
`
`existed prior to Dr. Rhodes’s invention:
`
`Various U.S. Patent Nos. have disclosed devices for
`intraluminar location and securement, which devices
`
`Response
`
`
`
`Page 5 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`include plural projections for effecting such securement,
`such as: 5,167,614 (Tessman et al.); 5,207,695 (Trout
`III); 5,275,622 (Lazarus et al); 5,306,286 (Stack et al.);
`5,383,892 (Cardon et al.); 5,387,235 (Chuter); 5,397,345
`(Lazarus); and 5,423,885 (Williams).
`
`(’417 Patent 3:28-34.) Representative figures from these exemplar prior art
`
`references are shown below:
`
`
`
`Response
`
`
`
`Page 6 of 56
`
`

`

`Case No. |PR2014-00100
`Case No. IPR2014-00100
`Patent No. 5,593,417
`Patent No. 5,593,417
`
`
`
`
`
`5,275,622 (Lazarus et al.) 5,306,286 (Stack et aL)
`
`165 Mi
`
`% €52
`
`my. 12
`
`
`
`“[E]acl1 of the hook-like
`elements 161 has been
`
`“More particularly. the present
`invention is directed to an
`
`
`
`
`5,387,235 Chute-1'
`
`provided with a barb 162 which
`extends outwardly fiom the
`main body .
`.
`.
`(EX. 2006 at
`
`absorhable
`
`stent for placement at the locus
`of a stenotic lesion .
`.
`.
`(Ex.
`
`10 :56-60).
`
`2007 at 2:22-28).
`
`5,383,892 Cordon et all.
`
`
`
`
`
`“The spring assembly 6 on the
`“By defomling so—outward
`distal (upstream) end of the
`opening- the ends sink
`graft 1 has small surgical baths
`slightly into the wall of the
`10 firmly attached to the spring
`parts of the body Where the
`assembly 6." (Ex- 2009 at 10:
`stent is implanted- hence
`nnpro‘t-‘ing the anchoring.“
`20-22).
`
`(Ex. 2008 at 2:62-65).
`
`Response
`Response
`
`
`
`Page 7 of 56
`Page 7 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`
`
`These prior art references disclose a multitude of shapes, including “hook-
`
`like projections,” “barbs,” “tine-like members with sharp points,” “protrusions”
`
`and “teeth.” However, Dr. Rhodes saw an opportunity for improvement over these
`
`devices, stating in his patent, “While such anchoring means are believed to be
`
`effective for their intended purpose, they never the less appear to be amenable to
`
`improvement insofar as graft retention is concerned.” (Id. 3:24-27.)
`
`All of these prior art anchoring mechanisms, derived from the concept of
`
`suturing a graft into position, have one thing in common: Each focused on the
`
`problem of embedding anchors in the wall of the artery immediately upon
`
`deployment of the device in order to affix the grafts in place. (Silver Decl. ¶ 44.)
`
`Anchors designed in this fashion utilize manual forces exerted by the practitioner,
`
`Response
`
`
`
`Page 8 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`such as the radial outward force exerted by a balloon, and/or forces exerted by the
`
`stents themselves, such as a self-expanding stent, for the anchor to engage and
`
`penetrate the vessel wall. (Id.) However, these anchoring mechanisms did not
`
`take into account the influence that body forces would exert on the device after
`
`deployment. (Id.) Sometimes, these anchors worked against the natural forces
`
`within the vasculature, and could either result in the anchors damaging or tearing
`
`the vessels as they were pushed deeper into the tissue, or could result in fatigue
`
`fracture of the anchors due to pulsatile and shear flow forces acting on the graft.
`
`(Id.)
`
`In the ’417 Patent, Dr. Rhodes set out not to improve the fixation of the stent
`
`graft device to the vessel wall upon deployment but instead focused on a new
`
`method of attachment that would utilize natural migration forces to keep the device
`
`in place. As explained in the specification of the ’417 Patent, Dr. Rhodes
`
`recognized that while a stent graft may be secured in place within a blood vessel by
`
`passive fixation from the stents or active fixation from anchors at the time of
`
`deployment, the stent graft could still experience caudal (i.e., downstream)
`
`migration after it had been deployed due to the shear forces from the blood flow
`
`exerted on the graft and the pulsatile expansion and contraction of the vessel.
`
`(’417 Patent 3:36-38; Silver Decl. ¶ 45.) Specifically, in the “Background of The
`
`Invention” section of the specification, Dr. Rhodes described this problem in the
`
`Response
`
`
`
`Page 9 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`stent graft field, noting that “[n]otwithstanding the foregoing, a need exists for
`
`means for ensuring good retention from migration for intraluminal grafts.” (’417
`
`Patent 3:35-37.) Dr. Rhodes further explained how the invention described in the
`
`’417 Patent would utilize the blood flowing through the graft to exert longitudinal
`
`and radial forces onto his designed anchors for providing a tight engagement with
`
`the wall so that unwanted migration could be prevented:
`
`As will be appreciated by those skilled in the art, with
`blood (or some other fluid) flowing through the device 20
`in the direction of arrow 46 a force will be applied by
`that flow to the interior surface of the sleeve 24, and from
`there through the stents to the projections 40. The force
`applied to the projections 40 will have a force component
`directed in the direction of the fluid flow, and a force
`component perpendicularly
`thereto,
`i.e., extending
`radially outward. Thus, the flow of fluid, e.g., blood,
`through the device 20 will tend to force the projections
`40 into good engagement with the wall 12 of the vessel,
`duct, or lumen. In the embodiment shown herein the
`projections penetrate or burrow slightly into the artery
`wall, as shown clearly in FIG. 4. Such penetration may
`not be necessary for good resistance to migration of the
`device. If some penetration is deemed desirable the
`height of the projections is selected so that their
`penetrating points do not penetrate too deeply into the
`
`Response
`
`
`
`Page 10 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`artery wall. In this regard, the height of the projections is
`selected so that they do not penetrate into the adventicial
`or medial layers of the artery wall, but can penetrate its
`intima. It is anticipated that for applications within the
`very largest arteries, such as the abdominal aorta, that the
`height of the projections will be in the range of
`approximately 1.0 mm to 1.5 mm. For intermediate
`arteries, the height of the projections will be in the range
`of approximately 0.75 mm to 1.0 mm, and for small
`arteries, the height of the projections will be in the range
`of approximately 0.5 mm to 1.0 mm.
`
`As should be appreciated by those skilled in the art the
`number of projections used in any device will also be a
`considerable factor in the amount of securement against
`migration provided
`thereby.
` Thus, as a general
`proposition, the more projections utilized
`the
`less
`“penetration” or “burrowing” will necessary for good
`securement against migration.
`
`(Id. 8:11-45.)
`
`This development was a critical advancement, especially when compared to
`
`the contemporary alternatives available at the time. (Silver Decl. ¶ 46.) A crucial
`
`difference between Dr. Rhodes’ invention and the prior art is the use of the body’s
`
`natural fluid forces to cause the anchoring that would prevent migration,
`
`Response
`
`
`
`Page 11 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`specifically in relation to the direction of the blood flowing through the vessel.
`
`(Id.)
`
`In developing his novel design, Dr. Rhodes recognized that it would not be
`
`possible to control the depth of penetration of the anchors with perfect specificity
`
`due to a number of variables.* However, by focusing on the location, direction,
`
`orientation (angle), and height of the anchors, Dr. Rhodes designed anchors that
`
`would achieve the desired amount of engagement and penetration for securement
`
`of the device against migration but that would only penetrate the vessel wall to the
`
`degree necessary to counter the caudal downstream migration forces being exerted
`
`on the device after deployment. (’417 Patent 8:12-45; Silver Decl. ¶ 46.) For
`
`instance, if the device only experiences minimal migration forces, the anchors may
`
`only contact or slightly penetrate the vessel wall. (Silver Decl. ¶ 46.) However, if
`
`
`* Such variables include variances in the amount of force applied to the barbs from
`
`the blood flow, the thickness of the vessel wall between different patients and in
`
`different locations within the vessel, and the thickness and hardness of plaque
`
`that has built up at different locations on the interior of the vessel wall, as well as
`
`the curvature of the vessel and the pulsatile expansion and contraction of the
`
`vessel. (Silver Decl. ¶ 46.)
`
`Response
`
`
`
`Page 12 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`the device experiences more significant migration forces, the anchors may burrow
`
`more deeply into the vessel wall to counter the larger forces. (Id. ¶ 46.) Yet, in
`
`either case, the anchors are designed such that they do “not pose a significant risk
`
`of perforating the tissue of the vessel, duct, or lumen.” (’417 Patent 3:53-57.)
`
`As will be discussed in further detail below, the prior art references cited in
`
`the Petition focus on anchors that fully engage the vessel wall to provide the
`
`necessary attachment to fix the device in place immediately upon completion of
`
`the device’s deployment, as would be the case with sutures. These anchors
`
`undergo full attachment during deployment and would not experience any further
`
`gain in fixation or migration resistance as a result of the blood flow forces. Rather,
`
`the blood flow forces would either have no effect on the amount of securement
`
`provided by the anchors or force the anchors to further penetrate the vessel such
`
`that there would be an increased risk of damage to the vessel wall, or an exertion of
`
`force on the anchor that could result in damage to the anchor itself. (See Ex. 2012
`
`(“the Malina Article”); Rowe Dep. Tr. (Ex. 2023) at 76:5-81:25; Gupta Dep. Tr.
`
`(Ex. 2021) at 52:9-14, 132:16-25.)
`
`For example, if the anchor is forced to penetrate more deeply into the vessel
`
`wall, it risks damage to the integrity of the structure of the vessel wall, which could
`
`lead to dissection, as well as hemorrhaging or damage to surrounding organs.
`
`(Silver Decl. ¶ 44, 47.) If the anchor has reached, or is near, its full level of
`
`Response
`
`
`
`Page 13 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`penetration upon deployment, additional downstream forces could cause the vessel
`
`wall to rupture, due to the creation of a longitudinal rip or tear in the vessel wall.
`
`(Id.) If the vessel wall does not tear under the additional downstream shear forces
`
`from the blood flow, a torque would be exerted on the anchor that could cause the
`
`anchor to invert or fracture. (Id.) In any of these cases, the anchor would no
`
`longer be effective for its purpose of securing the device in place.
`
`Dr. Rhodes specifically hoped to improve on these problems with his ’417
`
`Patent by focusing on the effect anchoring mechanisms had on the vessel wall after
`
`deployment. (’417 Patent 7:16-32, 8:12-38; Silver Decl. ¶ 49.) Dr. Rhodes
`
`demonstrated the importance of this concept throughout the claims and the
`
`prosecution history of the ’417 Patent. For example, in claim 1, the ’417 Patent
`
`requires “…whereupon the force applied to said tubular member by the fluid
`
`flowing through said passageway produces on each of said projections a force
`
`component to cause said at least one surface to tightly engage the interior surface
`
`of the vessel, duct, or lumen to fixedly secure said device in place.” (’417 Patent
`
`9:40-46 (emphases added).) Dr. Rhodes contemplated that his novel anchoring
`
`projections, by using the natural forces provided by the blood flowing through the
`
`graft, would allow the graft to secure itself post-deployment and further help
`
`prevent migration, if needed. (’417 Patent 7:16-32, 8:12-38; Silver Decl. ¶ 49.)
`
`Response
`
`
`
`Page 14 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`The prosecution history further demonstrates Dr. Rhodes’ intent. In an
`
`amendment in May of 1996, Dr. Rhodes indicated that each and every projection
`
`of his design must be oriented in an acute angle in the downstream direction, as
`
`well as appropriately shaped and sized in order to achieve the desired result, i.e., to
`
`have at least one surface of the trailing portion of the projection be caused to
`
`tightly engage the interior surface of the vessel to fixedly secure the device in place
`
`as a result of the forces applied from fluid flowing through the device:
`
`
`
`(Ex. 1002 at 46).
`
`In this same amendment, Dr. Rhodes explained how this development
`
`distinguished itself over the prior art, demonstrating the need for the fluid force to
`
`help secure the graft in place versus previous stent graft attachment methods:
`
`Response
`
`
`
`Page 15 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`
`
`
`
`(Ex. 1002 at 48-49).
`
`The concept of using the natural forces of blood flowing through the stent
`
`graft to prevent migration after deployment would specifically be a targeted benefit
`
`Response
`
`
`
`Page 16 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`by medical device manufacturers in the industry, however, without giving credit to
`
`Dr. Rhodes. (Silver Decl. ¶ 52.)
`
`D. The Industry’s Use of Dr. Rhodes’s Ideas and the Implicit
`Acknowledgement of the Importance of its Benefits
`
`In 1997, two years after Dr. Rhodes filed his ’417 Patent, the first report on
`
`endovascular graft migration was published. (See Ex. 2025; see also Ex. 2026 at 2
`
`(2005 article citing to Ex. 2021 as “[t]he first report on device migration”).) A
`
`year later, in 1998, the Malina Article (Ex. 2011) was published in the Journal for
`
`Endovascular Surgery regarding the benefits of using anchors to prevent migration.
`
`The Malina Article discussed the importance of using hooks and barbs to properly
`
`anchor the graft into the wall to prevent migration after deployment. Specifically,
`
`the article mentioned:
`
`Thus, attempts to improve endovascular graft anchoring
`by excessive oversizing of smooth stents might be futile,
`especially because stents may cause pressure ulcers and
`perforations of the vessel walls. Rather than being
`pushed into the aortic wall by the radial force of the
`stent, the hooks and barbs engage the aortic wall
`when the stent-graft
`is pulled distally by the
`bloodstream. The angle between the stent and its
`hooks and barbs is important for this action.
`
`
`
`Response
`
`
`
`Page 17 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`(Id. at 6 (emphasis added).) This realization, three years after Dr. Rhodes’ first
`
`developed it, mirrors what is stated in the ’417 Patent:
`
`. . . said trailing portion including at least one surface
`preferentially oriented to extend at an acute angle to the
`first direction, whereupon the force applied to said
`tubular member by the fluid flowing through said
`passageway produces on each of said projections a force
`component to cause said at least one surface to tightly
`engage the interior surface of the vessel, duct, or lumen
`to fixedly secure said device in place.
`
`(’417 Patent 9:38-45.)
`
`While others desired the benefits of Dr. Rhodes’ innovative design relative
`
`to the angle of barb penetration, they did not arrive at that conclusion until much
`
`later than Dr. Rhodes. (Silver Decl. ¶¶ 94-103.) For instance, other patents
`
`disclose the dangers and problems associated with preventing stent graft migration
`
`that Dr. Rhodes had identified years before:
`
`Migration can be a significant problem in the placement
`of expandable stents and other intraluminal devices,
`particularly when placed in the vascular system where
`the prosthesis is subject to the forces of blood, especially
`on the arterial side. Nowhere is the prevention of
`migration more important and more challenging than
`when placing a stent graft to repair an abdominal aortic
`
`Response
`
`
`
`Page 18 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`aneurysm (AAA) where downstream migration of the
`device can result in the aneurysm no longer being
`excluded. If the aneurysm is no longer intact or
`subsequent rupture were to occur, the patient would then
`face an increased risk of death. Unlike surgically placed
`grafts which are sutured into place, only the radial forces
`of the stent would be available to hold the prosthesis into
`place.
`
`the problem of migration, stent graft
`To address
`manufacturers sometimes place a series of barbs or hooks
`that extend outward from
`the main body of
`the
`prosthesis, typically at its proximal end, either by
`attaching them to the stent frame with solder or by some
`other bonding technique, or to the graft material,
`typically by suturing. It has been observed that sutures
`attaching barbed stents to the graft material are subject to
`breakage due in part to the flexibility of the graft material
`and the considerable pulsatile forces of arterial blood
`acting on the device. These forces have been known to
`directly contribute to the detachment between the graft
`portion and anchoring stent.
`
`It has also been observed that barbs soldered or otherwise
`attached to the stent frame are subject to fracture,
`detachment, or other failure, especially when the forces
`become concentrated at a particular location along the
`stent graft. Unfortunately, simply making the barbs
`
`Response
`
`
`
`Page 19 of 56
`
`

`

`Case No. IPR2014-00100
`Patent No. 5,593,417
`
`
`
`stronger to prevent fracture can result in increased
`damage to the anchoring tissue. Furthermore, adding
`rigidity
`to
`any
`outward-projecting
`barbs may
`compromise the ability of the device to be compressed

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket