throbber
310
`
`J ENDOVASC SURG
`1998;5:31 0-317
`
`+ EXPERIMENTAL ____________________________________ _
`
`Endovascular AAA Exclusion: Will Stents With
`Hooks and Barbs Prevent Stent-Graft
`Migration?
`
`Martin Malina, MD, PhD; Bengt Lindblad, MD, PhD;
`Krasnodar lvancev, MD, PhD*; Mats Lindh, MD*;
`Janne Malina, MOt; and Jan Brunkwall, MD, PhD
`
`Departments of Vascular Surgery, *Radiology, and tPathology, Malmo
`University Hospital, Lund University, Malmo, Sweden
`
`·------------------------------------------------------+
`Purpose: To investigate if stents with hooks and barbs will improve stent-graft fixation in
`the abdominal aorta.
`Methods: Sixteen- to 24-mm-diameter Dacron grafts were deployed inside cadaveric aortas.
`The grafts were anchored by stents as in endovascular abdominal aortic aneurysm repair.
`One hundred thirty-seven stent-graft deployments were carried out with modified self-ex(cid:173)
`panding Z-stents with (A) no hooks and barbs (n = 75), (8) 4 5-mm-long hooks and barbs
`(n = 39), (C) 8 10-mm-long, strengthened hooks and barbs (n = 19), or (D) hooks only (n
`= 4). Increasing longitudinal traction was applied to determine the displacement force
`needed to extract the stent-grafts. The radial force of the stents was measured and correlated
`to the displacement force.
`Results: The median (interquartile range) displacement force needed to extract grafts an(cid:173)
`chored by stent A was 2.5 N (2.0 to 3.4), stent B 7.8 N (7 .4 to 1 0.8), and stent C 22.5 N ( 17.1
`to 27.9), p < 0.001. Both hooks and barbs added anchoring strength. During traction, the
`weaker barbs were distorted or caused intimal tears. The stronger barbs engaged the entire
`aortic wall. The radial force of the stents had no impact on fixation, while aortic calcification
`and graft oversizing had marginal effects.
`Conclusions: Stent barbs and hooks increased the fixation of stent-grafts tenfold, while the
`radial force Of stents had no impact. These data may prove important in future endograft
`development to prevent stent-graft migration after aneurysm exclusion.
`J Endovasc Surg 1998;5:370-317
`
`Key words: endovascular grafting, abdominal aortic aneurysm, migration, Z-stents
`•
`
`$
`
`Migration of stent-grafts used for abdominal
`aortic aneurysm (AAA) exclusion is a recog(cid:173)
`nized complication of aortic endografting. 1- 5
`At our center, one third of the AAA patients
`
`Address for correspondence and reprints: Martin Malina,
`MD, PhD, Department of Vascular Surgery, Malmo Univer(cid:173)
`sity Hospital, S-205 02 Malmo, Sweden. Fax: 46-40-33-73-
`35.
`
`treated with endografts containing Z-stents
`with four weak hooks and barbs presented
`with late migration of the proximal stent. 1
`Stent-graft dislodgment ranged from minor (5
`to 10 mm) to severe migration with complete
`descent of the stent-graft into the aneurysm
`sac, requiring late conversion to open surgery.
`Recently, other reports have appeared of simi(cid:173)
`lar problems with various stent-grafts, includ-
`
`Page 1
`
`IPR2014-00100 Pat. Owner Ex. 2012
`Medtronic v. Marital Deduction Trust
`
`

`

`J ENDOVASC SURG
`1998;5:31 0-317
`
`ANCHORING STENT-GRAFTS
`MALINA ET AL.
`
`311
`
`ing the Chuter bifurcated graft,2 the Corvita en(cid:173)
`dovascular graft (Dereume JP et al. 24th
`Annual Symposium on Current Critical Prob(cid:173)
`lems, New Horizons and Techniques in Vascu(cid:173)
`lar and Endovascular Surgery, New York, New
`York, USA, November 20-23, 1997), the Van(cid:173)
`guard system (personal observation), and the
`Malmo system.3.4 Late migrations have also
`been reported when balloon-expanded stents
`were used for graft anchoring. 5
`As most migrations are diagnosed ::::: 1 year
`after stent-graft insertion,4 biological incorpo(cid:173)
`ration of endovascular grafts seems insuffi(cid:173)
`cientto resistthe pulsatile forces present inside
`the aorta. These observations call for improved
`mechanical anchoring of stent-grafts. How(cid:173)
`ever, many new stent-grafts, such as the An(cid:173)
`euRx, Corvita, and Talent devices, are an(cid:173)
`chored by smooth stents without hooks and
`barbs or by small barbs only (Vanguard
`system).
`To our knowledge, only two groups have as(cid:173)
`sessed the force needed to dislodge endovas(cid:173)
`cular stent-grafts from cadaveric aortas. In the
`first study, 6 five different stent types were de(cid:173)
`ployed in healthy porcine aortas, which ob(cid:173)
`viated the assessment of stent function in
`atherosclerotic vessels. In the second experi(cid:173)
`ment? only Gianturco stents were used. The
`displacement force was exerted on both the
`distal and proximal stents of an endovascular
`stent-graft bridging an aneurysm. Therefore,
`no definite conclusions could be made about
`the specific anchoring of the proximal stent.
`Prompted by incidences of stent-graft mi(cid:173)
`gration1 and late proximal endoleaks8 in endo(cid:173)
`vascularly excluded AAAs, we designed this
`study to assess if stent-graft fixation can be
`improved mechanically by adding hooks and
`barbs to the graft-anchoring stents.
`
`METHODS
`
`Dacron grafts (Cooley Verisoft, Meadox
`Medicals, Oakland, NJ, USA) with diameters
`of 16, 20, and 24 mm were sutured to self-ex(cid:173)
`pandable Gianturco Z-stents (Cook Inc., Den(cid:173)
`mark), leaving a few millimeters of the stents
`protruding from the grafts proximally as is
`usually done for endovascular AAA exclusion
`
`(Fig. 1 ). The stents were 2.5 em long and 4.5
`em wide in their uncompressed state and con(cid:173)
`sisted of 10 bends of stainless steel wire.
`Four different modifications ofthe Gianturco
`stents were examined (Fig. 1 ): (A) smooth
`stents without any anchoring appendages; (B)
`stents equipped with 4 weak hooks and barbs
`(5 mm long and 0.25 mm thick) similar to those
`used by several groups4 ·9 in AAA repair; (C)
`stents with 8 stronger hooks and barbs (10 mm
`long and 0.3 mm thick); and (D) 8 reinforced
`hooks but no barbs. All barbs protruded out
`from the graft at an angle of about 30°.
`The aortas of 15 human cadavers were ex(cid:173)
`posed with minimal dissection and left in situ
`so as not to disrupt their attachment to sur(cid:173)
`rounding tissue. The vessels were transected
`about 5 em proximal to the aortic bifurcation,
`their proximal end thus mimicking an aneu(cid:173)
`rysm neck. Stent-grafts were deployed 2.5 em
`into the transected aortas, a distance corre(cid:173)
`sponding to the length of many AAA necks en(cid:173)
`countered in clinical practice (Fig. 2). In each
`case, the entire stent was completely con(cid:173)
`tained within the aorta, allowing all hooks and
`barbs to engage the vessel wall.
`The median age of the cadavers was 66
`years (range 46 to 84); half had been smokers
`and half had a previously diagnosed cardio(cid:173)
`vascular disease. The aortic walls were ma(cid:173)
`croscopically classified into five groups: (I)
`nonatherosclerotic, (II) soft intimal thickening,
`(Ill) calcified plaques engaging part of the aor(cid:173)
`tic circumference, (IV) circumferentially lo(cid:173)
`cated calcified plaques, and (V) completely
`calcified, incompressible aortas.
`The sample size of the different stent types
`was dictated by the experimental setting.
`Stentswithout hooks and barbs that did not dis(cid:173)
`rupt the aortic wall were assessed first. Then,
`in the same aortic segment, stents with weaker
`hooks and barbs were tested. Because these
`stents caused damage to the aortic wall, only a
`limited number of trials could be performed in
`each aortic segment. Finally, the stents with
`reinforced hooks and barbs and hooks only
`were tested until the aortic wall was severely
`damaged. Thereby, stents with extra long and
`thick anchoring appendages were always as(cid:173)
`sessed last, when the aortas offered the least
`anchoring support. In this manner, attachment
`of stents with hooks and barbs was not favored.
`
`Page 2
`
`IPR2014-00100 Pat. Owner Ex. 2012
`Medtronic v. Marital Deduction Trust
`
`

`

`312
`
`ANCHORING STENT-GRAFTS
`MALINA ET AL.
`
`J ENDOVASC SURG
`1998;5:310-317
`
`8 9 10 11
`
`Figure 1 +Three of the four types of Z-stents evaluated in this study: smooth stent (left), stent
`with four weak hooks and barbs (middle), and a stent with eight strengthened hooks and
`barbs (right). The barbs protrude through the grafts. Not pictured is the stent with eight rein(cid:173)
`forced hooks only.
`
`Longitudinal traction then was applied to the
`distal end of the grafts (Fig. 2) and measured by
`a tensiometer. The traction was increased
`gradually by steps of 0.5 N until the stent-grafts
`were dislodged from the aorta, thereby defin(cid:173)
`ing the displacement force. After extraction
`from the aorta, the stents were examined for
`deformation of the appendages and the aortas
`for intimal or transmural injury. The damaged
`portion then was excised and additional mea(cid:173)
`surements were taken more proximally. This
`was repeated until the level of the renal arteries
`was reached. In this fashion, the smooth stents
`were tested on 75 occasions, the design with 4
`hooks and barbs in 39 instances, the model with
`8 hooks and barbs 19 times, and the 8 hooks
`only were tested on 4 occasions.
`The radial force exerted by the stents was
`assessed with a tonometer (Ophtalmo Dyna(cid:173)
`mometre du Dr Bailliart, Guilbert & Routit,
`
`Paris, France) by measuring the force required
`to compress the stents 2 and 4 mm (Fig. 3).
`The measurements were carried out with the
`stents compressed inside grafts that mea(cid:173)
`sured 16, 24, and 36 mm wide. To avoid lateral
`bulging during measurement, the sides of the
`graft were supported.
`Nonparametric data were analyzed as me(cid:173)
`dian and interquartile range (lOR). The Wil(cid:173)
`coxon rank sum test was used for comparing
`two groups; three or more groups were com(cid:173)
`pared using the Kruskai-Wallis test. Differ(cid:173)
`ences were considered significant at p < 0.05.
`
`RESULTS
`
`Stent-grafts were implanted 137 times in(cid:173)
`side the 15 infrarenal aortas. The median
`force needed to extract the stent-grafts with
`smooth Gianturco stents was 2.5 N (lOR 2.0 to
`
`Page 3
`
`IPR2014-00100 Pat. Owner Ex. 2012
`Medtronic v. Marital Deduction Trust
`
`

`

`J ENDOVASC SURG
`1998;5:31 0-317
`
`ANCHORING STENT-GRAFTS
`MALINA ET AL.
`
`313
`
`3.4) (Fig. 4). The corresponding displacement
`force for the Gianturco stents with 4 weak
`hooks and barbs was 7.8 N (lOR 7.4 to 10.8),
`while the stents with 8 reinforced hooks and
`barbs required a median force of 22.5 N (lOR
`to 27.9) (p < 0.001 ). Z-stents with 8
`17.1
`strong hooks but no barbs had a displace(cid:173)
`ment force of 11.8 N (lOR 10.4 to 13.0) com(cid:173)
`pared to 22.5 N for stents with both hooks
`and barbs (p < 0.01 ).
`In every segment of cadaveric aorta, the
`stent-graft fixation was compared with the first
`and last time each type of stent was used. The
`median displacement force was slightly higher
`on the first occasion (p = 0.2). Hence, the fixa(cid:173)
`tion ofthe stents with hooks and barbs was not
`favored in this experimental setting because
`these stents were always assessed last.
`The weaker hooks and barbs engaged the
`thickened media but did not penetrate the en(cid:173)
`tire aortic wall. The dislodgment occurred by
`
`two mechanisms: upward distortion of the
`barbs upon traction until the angulation of the
`barbs was such that they could slide out of the
`aortic wall or by tearing of the intima. Often,
`pieces of atherosclerotic intima could be seen
`on the hooks and barbs of the extracted stents.
`In contrast, the reinforced hooks and barbs en(cid:173)
`gaged the entire aortic wall, frequently perfo(cid:173)
`rating it and emerging outside the vessel wall
`(Fig. 2). Upon powerful traction, these barbs
`would also become deformed or cause a
`transmural tear of the aorta different from the
`limited intimal injury produced by the weaker
`barbs.
`Oversizing the graft diameter increased
`slightly the displacement force but only for
`stents without hooks and barbs (p < 0.01 ).
`Stents with hooks and barbs retained their
`strength of fixation even when the grafts were
`undersized by 1 or 2 mm. Smooth stents had
`marginally stronger fixation in increasingly
`
`Figure 2 + Stent-grafts were deployed 2.5 em into transected cadaveric infra renal aortas (A)
`in situ to avoid disruption of the aortic attachment to surrounding tissue. Longitudinal traction
`then was applied to the distal end of the grafts (G). Upon traction, the enforced barbs (arrows)
`penetrated the aortic wall.
`
`Page 4
`
`IPR2014-00100 Pat. Owner Ex. 2012
`Medtronic v. Marital Deduction Trust
`
`

`

`314
`
`ANCHORING STENT-GRAFTS
`MALINA ET AL.
`
`J ENDOVASC SURG
`1998;5:310-317
`
`N
`
`Radial force (N)
`.4
`
`.2
`
`. 1
`
`~ .3
`
`16 mm
`
`24 mm
`
`36 mm
`
`0
`
`2
`
`4
`
`Stent depression
`(mm)
`
`Figure 5 +Radial force of stents contained within
`16-, 24-, and 36-mm grafts. Stents compressed to a
`diameter of 16 mm exerted a radial force almost
`twice that seen when compressed to 36 mm. The
`force required to make a 4-mm depression into the
`stents was twice that of making a 2-mm depression
`regardless oft he degree of compression. Each point
`is based on data from 6 registrations.
`
`Thus, the radial force of stents contained
`within 16-mm grafts was almost twice that of
`stents in 36-mm grafts. Similarly, the force
`needed to compress the stents 4 mm with the
`tonometer was almost twice that needed to
`compress them 2 mm, regardless of the diam(cid:173)
`eter of the graft containing the stent. By com(cid:173)
`paring the fixation of stents in aortas of differ-
`
`Displacement
`force (N)
`
`.. ...__ .. ..--......
`
`1 2
`
`8
`
`4
`
`o~-....-----.-........ ---.--
`1.2 1.4 1.6 1.8 2.0 Aortic
`diameter (em)
`
`for stent-grafts
`force
`Figure 6 +Displacement
`placed in aortas of various diameters and anchored
`by smooth stents (circles) or by stents with 4 hooks
`and barbs (squares). The stents (n = 72) exerted a
`greater radial force in small aortas, yet the displace(cid:173)
`ment force was equal at all aortic diameters and
`therefore unaffected by the radial force of the
`stents.
`
`Figure 3 +The radial force of stents at various states
`of compression was assessed by placing the stents
`inside grafts of different diameters: The force (N)
`needed to make a further 2- and 4-mm depression
`of the stents was measured by a tonometer. To
`avoid lateral bulging during measurements, the
`sides of the stent were supported.
`
`atherosclerotic aortas, while the displacement
`force for stents with hooks and barbs was un(cid:173)
`affected by the aortic calcification.
`The radial force of the Z-stents was greater
`the more the stents were compressed (Fig. 5).
`
`Displacement
`Force
`(N)
`30
`
`20
`
`1 0
`
`z
`0 ~-------------------------------
`Smooth
`
`8 strong
`hooks & barbs
`
`4weak
`hooks & barbs
`
`Figure 4 +The force required to displace 3 different
`Z-stent modifications from cadaveric aortas (n =
`75, 39, and 19, respectively). Medians, interquartile
`ranges (within box), and 1Oth and 90th percentiles
`(horizontal bars) are shown.
`
`Page 5
`
`IPR2014-00100 Pat. Owner Ex. 2012
`Medtronic v. Marital Deduction Trust
`
`

`

`J ENDOVASC SURG
`1998;5:310-317
`
`ANCHORING STENT-GRAFTS
`MALINA ET AL.
`
`315
`
`ent diameters, the impact of the stents' radial
`force on the displacement force could be as(cid:173)
`sessed (Fig. 6). The displacement force was
`unaffected by the aortic diameter; therefore,
`the radial force of the stents had no impact on
`the fixation of either Z-stent type. This analysis
`included only stent-grafts with a graft diame(cid:173)
`ter exceeding the aortic diameter by ::::: 2 mm
`(n = 72) to ensure that the stents would exert
`their radial force unimpeded.
`
`DISCUSSION
`Stents originally were designed to prevent
`occlusion of stenotic vessels. Today, however,
`they are widely used to anchor endovascular
`grafts, although little has been done to adapt
`the stents for this novel mission. The first com(cid:173)
`mercial aortic endograft, the EVT device, had
`hooks but no barbs for anchorage. Since then,
`numerous stent-graft systems have been de(cid:173)
`veloped worldwide, many without hooks or
`barbs to affix the proximal stent to the aortic
`wall
`(e.g., AneuRx, Corvita, and Talent).
`Among the second- and third-generation en(cid:173)
`dografts, only the Vanguard model has barbs,
`and they are a mere 2 mm long. Despite this,
`most centers report few cases of stent-graft
`10
`11 This may be due to short fol(cid:173)
`5
`migration. 1-
`•
`•
`low-up or careful selection of patients with fa(cid:173)
`vorable neck anatomy, thereby excluding >
`50% of AAA patients from endovascular re(cid:173)
`pair.
`At our institution, endovascular aneurysm
`repair was introduced in 1993; since then, up
`to 70% ofthe aneurysm repairs have been per(cid:173)
`formed endovascularly. Such liberal patient
`selection has unmasked the necessity for safer
`stent-graft anchoring. One third of our patients
`exhibit stent-graft migration. 1 That may be due
`to dilation or shortness of the aneurysmal
`neck, which often contains mural thrombus
`and calcification. However, migration has oc(cid:173)
`curred despite satisfactory neck anatomy,
`graft sizing, and stent-graft deployment. As
`most cases of migration were identified at :::::
`12 months, the healing of endovascular stent(cid:173)
`grafts seems insufficient to withstand the pul(cid:173)
`satile force within the aorta. Wall-penetrating
`stent hooks and barbs may provide improved
`fixation.
`The present study shows that simple stent
`appendages, such as hooks and barbs, areca-
`
`pable of improving the fixation of endovascu(cid:173)
`lar grafts tenfold in this experimental model.
`Both the hooks and the barbs improved attach(cid:173)
`ment. Short hooks and barbs (5 mm) engaged
`the thickened aortic intima only; the fragility
`of the intima then seemed to limit the force
`required to displace the stents. Therefore, the
`stents with longer and stronger barbs, which
`penetrated the entire aortic wall, provided a
`firmer fixation. However, a stent's radial force
`was of no significant importance for graft at(cid:173)
`tachment. Thus, attempts to improve endo(cid:173)
`vascular graft anchoring by excessive oversiz(cid:173)
`ing of smooth stents might be futile, especially
`because stents may cause pressure ulcers and
`perforations of the vessel walls. 12 Rather than
`being pushed into the aortic wall by the radial
`force of the stent, the hooks and barbs engage
`the aortic wall when the stent-graft is pulled
`distally by the bloodstream. The angle be(cid:173)
`tween the stent and its hooks and barbs is im(cid:173)
`portant for this action.
`The forces that cause endovascular aortic
`grafts
`to migrate
`in vivo are unknown.
`Transmural surgical sutures placed during
`open aneurysm repair can withstand the pul(cid:173)
`satile forces in the aorta. Therefore, long hooks
`and barbs that pierce the aortic wall (Figs. 2
`and 7) may also prove more durable, but there
`is a potential danger of penetration to juxta(cid:173)
`aortic structures, such as the duodenum and
`the renal vein. The long-term risks of these
`hooked and/or barbed stents have not been as(cid:173)
`sessed, but no damage to adjacent organs by
`aortic wall-piercing hooks has been demon(cid:173)
`strated in autopsy studies. 13 The metallic
`hooks and barbs must be strong enough to
`resist breakage or distortion by long-standing,
`repetitive trauma. Placing additional hooks
`and barbs on each stent will increase the vol(cid:173)
`ume of the devices and make them more diffi(cid:173)
`cult to introduce through slim sheaths.
`There are other ways to improve stent-graft
`anchoring without perforating the aortic wall.
`By placing additional, serially attached stents
`into the suprarenal aorta, more hooks and
`barbs can be used. This implies that vital aortic
`branches will be covered by stents. Although
`it has been reported that arterial branches cov(cid:173)
`ered by stents remain patent, 14- 16 various
`stents affect the branches differently. 17 Fully
`stented grafts with greater columnar strength
`
`Page 6
`
`IPR2014-00100 Pat. Owner Ex. 2012
`Medtronic v. Marital Deduction Trust
`
`

`

`316
`
`ANCHORING STENT-GRAFTS
`MALINA ET AL.
`
`J ENDOVASC SURG
`1998;5:310-317
`
`Figure 7 +Human specimen ofthe neck of an infra renal aortic aneurysm treated by endovascu(cid:173)
`lar exclusion 6 months previously. The caudally oriented barb of the proximal graft-anchoring
`Z-stent has perforated the aortic wall. No bleeding or injury to adjacent viscera was noted.
`The stent-graft was anchored securely without migration.
`
`may provide support for the proximal stent
`and make the graft "sit" on the aortic bifurca(cid:173)
`tion instead of being suspended from the an(cid:173)
`eurysm neck. However, migration of fully
`stented grafts has also occurred (Dereume JP
`et al. 24th Annual Symposium on Current Criti(cid:173)
`cal Problems, New Horizons and Techniques
`in Vascular and Endovascular Surgery, New
`York, New York, USA, November 20-23,
`1997). The extent to which the displacement
`force may be diverted from the proximal stent
`in fully stented grafts is unknown .
`It may be argued that the transected aortas
`in this experimental setting represent a subop(cid:173)
`timal model for aneurysm necks. However,
`one true aneurysm with a diameter of 8 em
`was included in this study. The displacement
`forces recorded at its neck corresponded
`
`closely to those seen in the nonaneurysmal
`aortas (data not presented).
`Today's challenge for those who develop
`endografts is to design a device that inhibits
`stent migration and the continuous dilatation
`2° Few endoleaks due
`ofthe aneurysm neck. 18
`-
`to neck enlargement have been reported so
`far. 1 However, present data suggest that the
`dilatation rate of the neck is about 1 mm per
`2° Considering the limited oversizing of
`year. 19
`•
`endovascular grafts, dilating necks might be a
`problem in the future.
`In summary,
`this experimental model
`showed that stent-graft fixation can be im(cid:173)
`proved by hooks and barbs. The data seem
`particularly valuable, as several devices with(cid:173)
`out stent appendages are being marketed for
`endovascular aneurysm repair. Excessive ra-
`
`Page 7
`
`IPR2014-00100 Pat. Owner Ex. 2012
`Medtronic v. Marital Deduction Trust
`
`

`

`J ENDOVASC SURG
`1998;5:310-317
`
`ANCHORING STENT-GRAFTS
`MALINA ET AL.
`
`317
`
`dial force by stents does not contribute to the
`strength of proximal attachment, suggesting
`that intentional oversizing should be avoided.
`
`REFERENCES
`
`1.
`
`3.
`
`4.
`
`lvancev K, Resch T, Brunkwall J, et al. Endo(cid:173)
`luminal repair of abdominal aortic aneurysms:
`A critical reappraisal after a three and half year
`experience. Sem
`lntervent Radiol 1998:in
`press.
`2. Chuter TA, Wendt G, Hopkinson BR, et al. Bi(cid:173)
`furcated stent-graft for abdominal aortic aneu(cid:173)
`rysm. Cardiovasc Surg 1997;5:388-392.
`lvancev K, Chuter TAM, Hopkinson BR, et al.
`Predictors of proximal stent migration follow(cid:173)
`ing endovascular AAA repair with a Gianturco
`Z-stent based stent-graft. (Abstr.) J Endovasc
`Surg 1997;4(Suppll):l-17.
`lvancev K, Malina M, Lindblad B, et al. Abdomi(cid:173)
`nal aortic aneurysms: Experience with the
`lvancev-Malmo endovascular system for aor(cid:173)
`tomonoiliac stent-grafts. J Endovasc Surg
`1997;4:242-251.
`5. White RA, Donayre CE, Walot I, et al. Regres(cid:173)
`sion of an abdominal aortic aneurysm after en(cid:173)
`dog raft exclusion. J Vase Surg 1997;26:
`133-137.
`6. Andrews SM, Anson AW, Greenhalgh RM, et
`al. In vitro evaluation of endovascular stents
`to assess suitability for endovascular graft fix(cid:173)
`ation. Eur J Vase Surg 1995;9:403-407.
`7. Chuter TAM, Green RM, Ouriel K, et al. Trans(cid:173)
`femoral endovascular aortic graft placement.
`J Vase Surg 1993;18:185-197.
`8. White GH, Yu W, May J, et al. Endoleak as a
`complication of endoluminal grafting of ab(cid:173)
`dominal aortic aneurysms: Classification, inci(cid:173)
`dence, diagnosis, and management. J Endo(cid:173)
`vasc Surg 1997;4:152-168.
`9. Yusuf SW, Whitaker SC, Chuter TA, et al. Early
`results of endovascular aortic aneurysm sur(cid:173)
`gery with aortouni-iliac graft, contralateral
`iliac occlusion, and femorofemoral bypass. J
`Vase Surg 1997;25:165-172.
`10. Blum U, Voshage G, Lam mer J, et al. Endolum-
`
`inal stent-grafts for infrarenal abdominal aor(cid:173)
`tic aneurysms. N Engl J Med 1997;336:13-20.
`11. White GH, Yu W, May J, et al. Three-year expe(cid:173)
`rience with the White-Yu Endovascular GAD
`Graft for transluminal repair of aortic and iliac
`aneurysms. J Endovasc Surg 1997;4:124-136.
`12. Malina M, Brunkwall J, lvancev K, et al. Late
`aortic arch perforation by graft anchoring
`stent: Complication of the endovascular exclu(cid:173)
`sion of a thoracic aneurysm. J Endovasc Surg
`1998;5:27 4-277.
`13. McGahan T J, Berry GA, McGahan SL, et al. Re(cid:173)
`sults of autopsy 7 months after successful en(cid:173)
`doluminal treatment of an infrarenal abdomi(cid:173)
`nal aortic aneurysm. J Endovasc Surg 1995;2:
`348-355.
`14. Malina M, Lindh M, lvancev K, et al. The effect
`of endovascular aortic stents placed across
`the renal arteries. Eur J Vase Endovasc Surg
`1997;13:207-213.
`15. Nasim A, Thompson MM, Sayers RD, et al. In(cid:173)
`vestigation of the relationship between aortic
`stent position and renal function. J Endovasc
`Surg 1995;2:90-91.
`16. Malina M, Brunkwall J, lvancev K, et al. May
`renal arteries be covered by aortic stents? Clin(cid:173)
`ical experience from endovascular grafting of
`aortic aneurysms. Eur J Vase Endovasc Surg
`1997;14:1 09-113.
`17. Desgranges P, Hutin E, Kedzia C, et al. Aortic
`stents covering the renal arteries ostia: An ani(cid:173)
`mal study. J Vase lnterv Radiol 1997;8:77-82.
`18. Edwards JM, Teefey SA, Zierler RE, et al. Intra(cid:173)
`abdominal para-anastomotic aneurysms after
`aortic bypass grafting. J Vase Surg 1992;15:
`344-353.
`19. Sonesson B, Malina M, lvancev K, et al. Dilata(cid:173)
`tion of the infra renal aneurysm neck after en(cid:173)
`dovascular exclusion of abdominal aortic an(cid:173)
`eurysm. J Endovasc Surg 1998;5:195-200.
`20. May J, White G, Yu W, et al. A prospective
`study of anatomico-pathological changes in
`abdominal aortic aneurysms following endo(cid:173)
`luminal repair: Is the aneurysmal process re(cid:173)
`versed? Eur J Vase Endovasc Surg 1996;12:
`11-17.
`
`Page 8
`
`IPR2014-00100 Pat. Owner Ex. 2012
`Medtronic v. Marital Deduction Trust
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket