throbber
Trials@uspto.gov
`Tel: 571-272-7822
`
`
`Paper 46
` Entered: March 24, 2015
`
`
`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`
`____________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`____________
`
`
`MEDTRONIC, INC., and MEDTRONIC VASCULAR, INC.,
`Petitioner,
`
`v.
`
`MARITAL DEDUCTION TRUST and ENDOTACH LLC,
`Patent Owner.
`____________
`
`Case IPR2014-00100
`Patent 5,593,417
`____________
`
`
`
`Before JACQUELINE WRIGHT BONILLA, MICHAEL J. FITZPATRICK, and
`HYUN J. JUNG, Administrative Patent Judges.
`
`BONILLA, Administrative Patent Judge.
`
`FINAL WRITTEN DECISION
`35 U.S.C. § 318(a) and 37 C.F.R. § 42.73
`
`
`
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`I.
`
`INTRODUCTION
`
`A. Background
`
`
`
`Petitioner Medtronic, Inc. and Medtronic Vascular, Inc. (“Petitioner”) filed a
`
`corrected Petition (Paper 5, “Pet.”) to institute an inter partes review of claims 1,
`
`2, 9, 10, and 13 of U.S. Patent No. 5,593,417 (Ex. 1001, “the ’417 patent”).
`
`35 U.S.C. § 311. Patent Owner, the Marital Deduction Trust, and its exclusive
`
`licensee Endotach LLC (“Patent Owner”)1 did not file a Preliminary Response.
`
`We determined that the information presented in the Petition demonstrated that
`
`there was a reasonable likelihood that Petitioner would prevail in challenging
`
`claims 1, 2, 9, 10, and 13 of the ’417 patent as unpatentable. Paper 15 (“Dec. to
`
`Inst.”), 2, 15. Pursuant to 35 U.S.C. § 314, we instituted this proceeding on March
`
`25, 2014, to review whether Kornberg2 anticipates claims 1, 2, 9, 10, and 13 of the
`
`’417 patent under 35 U.S.C. § 102, and also whether those claims would have been
`
`obvious over Rhodes ’1543 and Kornberg under 35 U.S.C. § 103. Dec. to Inst. 15.
`
`After institution of trial, Patent Owner filed a Patent Owner Response.
`
`Paper 27 (“PO Resp.”). Petitioner subsequently filed a Reply to the Response.
`
`Paper 35 (“Reply”). An oral hearing was held on November 20, 2014. A
`
`transcript of the hearing has been entered into the record. Paper 44 (“Tr.”).
`
`
`1 Patent Owner indicates that the Marital Deduction Trust, created under the
`Valentine J. Rhodes Revocable Trust, is the owner of the ’417 patent, while
`Endotach LLC is the exclusive licensee of all substantial interests of the patent.
`Paper 10, 2; Paper 20, 2.
`2 Kornberg, U.S. Pat. No. 4,562,596, issued Jan. 7, 1986 (“Kornberg”) (Ex.1006).
`3 Rhodes, U.S. Pat. No. 5,122,154, issued June 16, 1992 (“Rhodes ’154”) (Ex.
`1008).
`
` 2
`
`
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`
`We have statutory authority under 35 U.S.C. § 6(c). This Final Written
`
`Decision is issued pursuant to 35 U.S.C. § 318(a). Petitioner has shown by a
`
`preponderance of the evidence that claims 1, 2, 9, 10, and 13 of the ’417 patent are
`
`unpatentable.
`
`B. Related Matters
`
`Petitioner indicates that Patent Owner has asserted the ’417 patent against it
`
`in Endotach LLC v. Medtronic, Inc., No. 5:13-cv-03292-EJD (N.D. Cal.). Pet. 1.
`
`In its Mandatory Notices, Patent Owner identifies two other cases that may affect
`
`or be affected by this proceeding: Endotach LLC v. Cook Medical Inc., No. 1:13-
`
`cv-1135 (S.D. Ind.) and Endotach LLC v. W.L. Gore & Associates, Inc., No. 3:12-
`
`cv-00308 (N.D. Fla.). Paper 10, 2–3; Paper 20, 2–3.
`
`On April 25, 2014, after we instituted a trial in the current case, Petitioner
`
`filed another Petition in Case IPR2014-00695, involving the same parties and same
`
`claims of the ’417 patent at issue in this proceeding. IPR2014-00695, Paper 1
`
`(“Second Petition”). Petitioner also filed a Motion for Joinder requesting “that the
`
`Second Petition be joined with IPR2014-00100.” IPR2014-00695, Paper 2, 2. The
`
`Second Petition reasserted two grounds of unpatentability previously asserted in
`
`this proceeding, as well as three new grounds relying on two additional references.
`
`Second Petition 10–11, 13–32. In a Decision dated September 25, 2014, we denied
`
`Petitioner’s Motion for Joinder, as well as Petitioner’s Second Petition. IPR2014-
`
`00695, Paper 18, majority op. at 9.
`
`C. The ’417 Patent (Ex. 1001)
`
`The ’417 patent relates to an intraluminal medical device, such as an
`
`endovascular graft or stent. Ex. 1001, 3:45–48. The patent discusses U.S. Pat.
`
`No. 5,122,154 (Ex. 1008, “Rhodes ’154”), also relating to an intraluminal graft.
`
`Ex. 1001, 2:64–3:27. The ’417 patent states the present graft device “is
`
` 3
`
`
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`constructed in accordance with the teachings of my aforementioned patent [Rhodes
`
`’154], except for the means for fixedly holding it in place within the vessel, duct,
`
`or lumen,” i.e., the “anchoring means.” Id. at 5:10–17.
`
`Figures 2, 3, 7, and 8 of the ’417 patent are reproduced below.
`
`
`
`
`
`
`
`
`
`
`
`
`
`Figure 2 depicts a portion of an endovascular bypass graft. Id. at 4:47–52.
`
`Figure 3 depicts an enlarged view of the portion in Figure 2 designated as “FIG. 3”
`
`with broken lines. Id. at 4:53–55. Figure 7 depicts another embodiment of a graft.
`
`Id. at 4:65–67. Figure 8 depicts an enlarged sectional view taken along line 8-8 of
`
`Figure 7. Id. at 5:1–2.
`
`In Figure 2, the graft comprises tubular member 24 having a plurality of
`
`expandable, ring-like, stent members 26. Id. at 5:54–59. Each stent member 26
`
`comprises a plurality of links 30, where each link is joined to another link by joint
`
` 4
`
`
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`32. Id. at 6:21–32. “In order to help hold or secure the graft in position in the
`
`artery (or lumen or duct) once the graft has been expanded,” the graft includes
`
`anchoring means comprising projections 40. Id. at 7:9–13. Figure 3 shows details
`
`of an embodiment of “arrow head” projections 40 on joint 32. Id. at 7:60–63.
`
`Each projection “includes a leading edge 42 defining the ‘tip’ of the ‘arrow-head,’”
`
`where “leading edge 42 extends upward at an acute angle to the exterior surface of
`
`the stent and terminates at the top surface 44 of the projection.” Id. at 7:63–67; see
`
`also Fig. 4. The projections also include trailing edges 48, 50, and 52, each of
`
`which “inclines upward in the direction of the blood flow to terminate at the top
`
`surface 44.” Id. at 8:2–6.
`
`In another embodiment, shown in Figures 7 and 8, projections 70 are
`
`“wedge” shaped. Id. at 8:54–56. Leading surface 72 defines “the ‘front face’ of
`
`the ‘wedge,’” and “extends upward at an acute angle to the exterior surface of the
`
`stent and terminates at the top surface 74.” Id. at 8:56–58. The projections also
`
`include “trailing surface 76 which inclines upward in the direction of the blood
`
`flow to terminate at the top surface 74 in a penetration edge 78,” and “are
`
`preferentially oriented at an acute angle to the direction of blood flow.” Id. at
`
`8:58–67.
`
`D. Illustrative Claim
`
`Claim 1, the only challenged independent claim, is reproduced below.
`
`1. An intraluminal medical device for securement within a vessel, duct, or
`lumen of a living being, the vessel, duct, or lumen having an interior surface,
`said device comprising a tubular member and anchoring means,
`
`said tubular member having a passageway extending therethrough and
`an outer periphery, said tubular member being arranged to have a body fluid
`flow through said passageway in a first direction when said device is located
`within the vessel, duct, or lumen, whereupon a force is applied to said
`tubular-member,
`
` 5
`
`
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`
`said anchoring means being located adjacent said outer periphery of
`said tubular member and comprising plural projections arranged for
`engagement with the interior surface of the vessel, duct, or lumen,
`
`each of said projections having a leading portion located in the
`upstream direction of the fluid flow and a trailing portion located in the
`downstream direction thereof, said trailing portion including at least one
`surface preferentially oriented to extend at an acute angle to the first
`direction,
`
`whereupon the force applied to said tubular member by the fluid
`flowing through said passageway produces on each of said projections a
`force component to cause said at least one surface to tightly engage the
`interior surface of the vessel, duct, or lumen to fixedly secure said device in
`place.
`
`Id. at 9:23–45 (paragraph indentation added).
`
`II. ANALYSIS
`
`A. Claim Construction
`
`In an inter partes review, “[a] claim in an unexpired patent shall be given its
`
`broadest reasonable construction in light of the specification of the patent in which
`
`it appears.” 37 C.F.R. § 42.100(b); see also In re Cuozzo Speed Tech., LLC, No.
`
`2014-1301, 2015 WL 448667, at *5–*8 (Fed. Cir. Feb. 4, 2015) (“Congress
`
`implicitly adopted the broadest reasonable interpretation standard in enacting the
`
`AIA,” and “the standard was properly adopted by PTO regulation”). There is a
`
`“heavy presumption” that a claim term carries its ordinary and customary meaning.
`
`CCS Fitness, Inc. v. Brunswick Corp., 288 F.3d 1359, 1366 (Fed. Cir. 2002); In re
`
`Translogic Tech., Inc., 504 F.3d 1249, 1257 (Fed. Cir. 2007). A patentee may
`
`rebut this presumption, however, by acting as his own lexicographer, providing a
`
`definition of the term in the specification with “reasonable clarity, deliberateness,
`
`and precision.” In re Paulsen, 30 F.3d 1475, 1480 (Fed. Cir. 1994). In the absence
`
` 6
`
`
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`of such a definition, limitations are not to be read from the specification into the
`
`claims. In re Van Geuns, 988 F.2d 1181, 1184 (Fed. Cir. 1993).
`
`1. “Projection,” “Leading Portion,” “Trailing Portion,” and “Stent”
`
`Petitioner offers constructions for the terms “projection,” “leading portion,”
`
`and “trailing portion” recited in claim 1, as well as “stent” recited in claims 9, 10,
`
`and 13. Pet. 9–12. Upon review of the Specification of the ’417 patent and the
`
`challenged claims, we construe “projection” to mean a structure that extends
`
`outward from a surface, such as a surface of a stent. Additional claim language
`
`indicates that the projection comprises a leading portion and a trailing portion,
`
`where the projection is capable of engaging with an interior surface of a vessel,
`
`duct, or lumen in a living being. Ex. 1001, 3:21–24, 7:9–59, claim 1. The “leading
`
`portion” of a projection is located in an upstream direction of fluid flow, while the
`
`“trailing portion” is located in the downstream direction of fluid flow. Id. at claim
`
`1; Pet. 10–11. For example, as shown in Figure 8 of the Specification, leading
`
`surface 72 corresponds to a leading portion (oriented upstream from the direction
`
`of blood flow), while trailing surface 76 corresponds to a trailing portion (oriented
`
`downstream from the direction of blood flow). In addition, we construe “stent” to
`
`mean any structure that provides structural support, such as for a blood vessel,
`
`duct, or lumen. Ex. 1001, 2:12–14, 3:45–48; Pet. 11.
`
`2. “At Least One Surface”
`
`Patent Owner contends that the phrase “at least one surface” in relation to
`
`the “trailing portion” in claim 1 refers to “a portion, part or surface of the trailing
`
`portion of a projection which is oriented at an acute angle to the fluid flow,” as
`
`recited in claim 1. PO Resp. 25–26. We note that claim 1 recites in relevant part
`
`that the trailing portion includes “at least one surface preferentially oriented to
`
`extend at an acute angle to the first direction.” In view of the Specification, we
`
` 7
`
`
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`interpret this limitation as proposed by Patent Owner, i.e., as referring to a
`
`structural limitation indicating direction. Id.; Tr. 37–38. Petitioner does not
`
`disagree with this construction.
`
`3. “Engagement With” and “Engaging”
`
`Patent Owner contends that “engagement with” and “engaging,” as recited in
`
`claim 1, require no construction, and the plain and ordinary meaning applies. PO
`
`Resp. 26. We generally agree but clarify that those terms mean contacting in some
`
`fashion. See, e.g., Transcript of Deposition of James Silver, Ph.D., Ex. 1014,
`
`22:13–20 (Dr. Silver testifying that “engage would mean that the device was in
`
`contact with the vessel wall”); Order on Claim Construction (S.D. Ind. 2013), Ex.
`
`2013, 34.
`
`4. “Tightly Engage”
`
`Patent Owner contends that “tightly engage” means more than “engage” in
`
`that it describes the degree of connection between the projections and interior
`
`surface of the vessel, duct, or lumen. PO Resp. 26. Patent Owner notes that claim
`
`1 recites that a “force from the fluid flowing through” the tubular member
`
`produces a force “to cause” a surface of the trailing portion “to tightly engage the
`
`interior surface of the vessel, duct, or lumen to fixedly secure said device in place.”
`
`Id. at 27; Ex. 1001, claim 1.
`
`Referring to this claim language, Patent Owner contends that being “fixedly
`
`secure” occurs as a result of forces from blood flow only, but not other forces, such
`
`as manual force exerted by a physician, or force from a balloon or stent itself. Id.
`
`at 30. Patent Owner also contends, citing certain portions of the Specification, that
`
`a surface of the trailing portion tightly contacts or penetrates, i.e., tightly engages,
`
`only the interior surface of the vessel wall, i.e., “the intima (inner) layer, but not
`
`into the medial (middle) or adventitial (outer) layers.” Id. at 27–29 (citing
`
` 8
`
`
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`Ex. 1001, 8:28–30, 3:53–57).
`
`
`
`We agree with Patent Owner that “tightly engage” means more than simply
`
`“engage.” The Specification of the ’417 patent does not define “tightly” per se.
`
`Rather, it states that a force of fluid, e.g., blood flow, causes projections “to tightly
`
`engage, e.g., burrow slightly into, the interior of the wall of the vessel, duct, or
`
`lumen to thereby fixedly secure the device in place.” Ex. 1001, 4:13–25, 34–39,
`
`7:17–32. The Specification also states that projections “tightly engage (and not
`
`necessarily penetrate) the interior of the wall of the vessel, duct, or lumen to
`
`fixedly secure the device in place against migration.” Id. at 9:1–17. It also
`
`indicates that projections can “penetrate or burrow slightly into the artery wall.”
`
`Id. at 8:19–23, 42–45. Thus, the Specification clarifies that “tightly engage”
`
`encompasses burrowing slightly into, penetrating, or not necessarily penetrating, a
`
`blood vessel wall.
`
`In relation to a particular embodiment, shown in Figure 4, the Specification
`
`states that “[i]f some penetration is deemed desirable the height of the projections
`
`is selected so that their penetrating points do not penetrate too deeply into the
`
`artery wall,” i.e., “the height of the projections is selected so that they do not
`
`penetrate into the adventicial or medial layers of the artery wall, but can penetrate
`
`its intima.” Id. at 8:22–31. Notwithstanding that discussion, we do not read the
`
`Specification as a whole to require that “tightly engage” precludes projections
`
`penetrating adventicial or medial layers of an artery wall. Instead, the
`
`Specification states that an object of the invention is to provide anchoring means
`
`that do “not pose a significant risk of perforating the tissue of the vessel, duct, or
`
`lumen.” Id. at 3:53–58.
`
`Thus, we construe “tightly engage” in claim 1 to mean that trailing portions
`
`of the recited projections are structurally capable of contacting, burrowing into, or
`
` 9
`
`
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`penetrating into an interior surface of a vessel, duct, or lumen in a manner that
`
`fixedly secures the device in place, without posing “a significant risk of perforating
`
`the vessel, duct, or lumen” altogether. Id. The projections and their trailing
`
`portions must be structurally capable of allowing a fluid flow force, e.g., blood
`
`flow, through the tubular member to help cause and maintain the tight engagement
`
`of the projections, and keep the device securely in place, even if other forces, such
`
`as manual force by a physician, are also involved in deploying the device initially.
`
`See, e.g., Ex. 1001, 7:9–33 (stating that “to help hold or secure the graft in position
`
`in the artery (or lumen or duct) once the graft has been expanded, the graft
`
`includes . . . anchoring means” comprising “plural protuberances or projections
`
`40”), 3:35–37, 8:11–45 (stating that “flow of fluid, e.g., blood, through the device
`
`20 will tend to force the projections 40 into good engagement” and “penetration
`
`may not be necessary for good resistance to migration of the device”); Declaration
`
`of James Silver, Ph.D., Ex. 2002 ¶ 45.
`
`B. Anticipation by Kornberg
`
`Petitioner argues that Kornberg anticipates claims 1, 2, 9, 10, and 13 of the
`
`’417 patent. Pet. 8, 19–22.
`
`1. Kornberg (Ex. 1006)
`
`Kornberg describes a tubular graft comprising “a plurality of struts or stays
`
`equipped with hooks for rapid and secure attachment within the desired location of
`
`the damaged artery.” Ex. 1006, 2:15–19. Figures 1 and 2 of Kornberg are
`
`reproduced below.
`
`
`
`10
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`
`
`
`
`
`
`
`
`Figure 1 depicts an aortic bifurcation graft equipped with a circumferential row of
`
`hooks 14 and a plurality of longitudinal struts 12. Id. at 2:23–27. Figure 2 depicts
`
`an enlarged view of the upper end of a single strut 12 with hook 14 having barb 15.
`
`Id. at 2:28–29.
`
`Kornberg states that graft 10 “must have support along its length, so that the
`
`blood flow does not dislodge it,” but the graft “is sufficiently flexible to be capable
`
`of conforming to the interior contour of the wall portion of the artery into which it
`
`is inserted.” Id. at 2:56–62. Graft 10 “is a generally cylindrical, hollow, bifurcated
`
`sleeve with longitudinal supporting and reinforcing members called struts 12
`
`running along the major axis of the cylindrical sleeve.” Id. at 2:62–65. For
`
`instance, as shown in Figure 1, struts 12 run along the length of legs (10A and
`
`10B) of graft 10 and “assure proper orientation of the graft within the artery.” Id.
`
`at 2:62–68; 4:17–20. As described in Kornberg, “the number of circumferential
`
`located strengthening struts or ribs 12 attached or formed in the wall of the graft
`
`may vary from a minimum of four up to twelve or more, preferably eight.” Id. at
`
`
`
`11
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`3:1–5. Graft 10 also includes flexible ring 16 at the upper end of the graft. Id. at
`
`4:6–9.
`
`Hooks 14 are located at the upper end of each strut 12, and a “row of hooks
`
`14 forms a ring around the outer circumference of graft 10 and are oriented
`
`downwardly at an angle of about 10º–45º C. with respect to the vertical.” Id. at
`
`3:60–65; see also id. at 4:28–30 (stating that hook 14 is at an angle of about 30º in
`
`Figure 2). Hook lengths are “typically 2 to 8 mm.” Id. at 4:40–42; Ex. 2002 ¶ 68.
`
`As described in Kornberg, “[e]ach hook 14 has a barb 15 located at the lower end
`
`of the hook so as to inhibit upward movement which might tend to dislodge the
`
`graft after it is positioned and attached to the aorta wall.” Ex. 1006, 3:66–4:1.
`
`Kornberg also describes, referring to Figure 9, that “downward flow of
`
`blood holds the distal graft limbs 10A and 10B in place so that no mechanical
`
`attachment is necessary distally,” and that the “flow mechanism of the blood keeps
`
`the graft open.” Id. at 6:20–30. In relation to Figure 10, Kornberg describes that
`
`the “aortic wall at the neck of the aneurysm 56 is pierced by the multiple radially
`
`placed hooks 14. These puncture wounds are then excluded from the blood flow
`
`by the snug fit provided by the flexible ring 16 at the top of the graft.” Id. at 6:40–
`
`47. Kornberg also contemplates that one could make its graft without ring 16. Id.
`
`at 4:6–16.
`
`2. Analysis—Claim 1
`
`Petitioner contends that Kornberg discloses each and every element of claim
`
`1, referring to annotated versions of Figures 1, 2, and 9 in Kornberg and
`
`disclosures in the reference, as well as a claim chart and a Declaration by Travis
`
`Rowe. Pet. 19–22, Appx. A2, 5–9; Declaration of Travis Rowe, Ex. 1003 ¶¶ 22–
`
`26. For example, Petitioner points to disclosure in Kornberg as corresponding to
`
`certain elements in claim 1 as follows:
`
`
`
`12
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`Element in claim 1
`
`
`
`Disclosure in Kornberg
`
`“intraluminal medical device”
`
`Graft 10
`
`“tubular member”
`
`Struts 12 and ring 16
`
`“anchoring means” comprising “plural
`projections”
`
`Hooks 14
`
`Pet. 20, Appx. A2, 5–7.
`
`In relation to the “leading portion” and “trailing portion” of the projections
`
`recited in claim 1, Petitioner provides annotated figures from Kornberg, including
`
`annotated Figure 2, reproduced below.
`
`
`
`Pet. 20. Annotated Figure 2 depicts Figure 2 of Kornberg (shown previously),
`
`with added designations including “Leading Portion,” “Upstream,” “Trailing
`
`Portion,” “Acute Angle,” “Downstream,” and an arrow indicating “Fluid Flow
`
`(‘First Direction’).” Id.
`
`Petitioner contends that hooks 14 are located adjacent to the outer periphery
`
`of the tubular member (struts 12 and ring 16) in Kornberg’s graft 10. Id.
`
`According to Petitioner, each hook 14 includes a “leading portion,” as designated
`
`in annotated Figure 2, located in the upstream direction of fluid flow, as recited in
`
`
`
`
`13
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`claim 1. Id. Petitioner further contends that each hook includes a “trailing
`
`portion,” as also designated in annotated Figure 2, located in the downstream
`
`direction of fluid flow, which includes a portion that is oriented to extend at an
`
`acute angle to the fluid flow, as also recited in claim 1. Id. at 20–21.
`
`In relation to the last “whereupon” clause in claim 1, Petitioner also
`
`contends, relying on the Rowe Declaration, that:
`
`force applied to the manually-anchored tubular member (“struts 12
`and ring 16”) of Kornberg by fluid flowing through the interior
`passageway thereof inherently produces on each of the projections
`(“hooks 14”) a force component that causes at least one surface of the
`trailing portion to tightly engage the interior surface of the vessel,
`duct, or lumen to fixedly secure the device (“graft 10”) in place.
`
`Id. at 21 (citing Ex. 1003 ¶¶ 22–26). Petitioner also points us to Kornberg’s
`
`Abstract, which describes “struts having angled hooks with barbs at their upper
`
`ends, the upper ends of the struts extending beyond the upper end of the tubular
`
`material, thus allowing the graft to be securely attached to the inside of the aorta.”
`
`Ex. 1006, Abstract; Pet. Appx. A2, 7–8.
`
`Patent Owner responds, relying on a Declaration by Dr. James Silver (Ex.
`
`2002). PO Resp. 1. First, Patent Owner contends that Kornberg discloses that its
`
`graft projections fully engage upon completion of the device’s deployment, and
`
`therefore, Kornberg fails to disclose that fluid flow forces cause projections to
`
`engage after deployment. Id. at 1, 13. Patent Owner contends that, when using
`
`Kornberg’s device, “blood flow forces would either have no effect on the amount
`
`of securement provided by the anchors or force the anchors to further penetrate the
`
`vessel such that there would be an increased risk of damage to the vessel wall,
`
`or . . . the anchor itself.” Id. at 13–14 (citing Ex. 2012; Ex. 2023, 76:5–81:25;
`
`Ex. 2021, 52:9–14, 132:16–25; Ex. 2002 ¶¶ 44, 47). Patent Owner also refers to
`
`
`
`14
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`the prosecution history of the ’417 patent, arguing that inventor Dr. Rhodes
`
`addressed “the need for the fluid force to help secure the graft in place versus
`
`previous stent graft attachment methods.” PO Resp. 15–17.
`
`Along these lines, Patent Owner contends that “[w]hile forces from the fluid
`
`flow acting on the graft 10 may inherently exist and keep the graft 10 open,
`
`Kornberg does not teach or suggest that such fluid flow forces are what causes the
`
`at least one surface of the trailing portion of the hooks 14 to tightly engage the
`
`vessel wall.” Id. at 35 (citing Ex. 2023, 80:1–7; Ex. 2002 ¶ 73). According to
`
`Patent Owner, Kornberg’s device is “fixedly secured in place against migration”
`
`by manual force exerted by a physician, such that trailing portions of hooks 14
`
`fully penetrate a vessel wall during deployment, and thereafter, the hooks “would
`
`not experience any further gain in fixation against migration resistance as a result
`
`of the blood flow forces.” PO Resp. 35–36 (citing Ex. 2023, 80:14–25, 81:8–25;
`
`Ex. 2002 ¶¶ 74, 76, 77).
`
`Petitioner establishes by a preponderance of the evidence that Kornberg’s
`
`graft device includes a “tubular member,” and that hooks 14 in Kornberg’s graft
`
`device correspond to an “anchoring means” comprising “projections” that each
`
`have a “leading portion,” located in the upstream direction of fluid flow, as well as
`
`a “trailing portion,” located in the downstream direction of fluid flow, as recited in
`
`claim 1. Pet. 20 (including annotated Figure 2 from Kornberg). Petitioner also
`
`establishes sufficiently that Kornberg’s projections (hooks 14 with barbs 15) have
`
`trailing portions with a surface “oriented to extend at an acute angle to the first
`
`direction,” as recited in claim 1. Pet. 20–21 (including annotated Figure 2 from
`
`Kornberg), Appx. A2, 6–7.
`
`Patent Owner does not contend otherwise. Nor does Patent Owner dispute
`
`that Kornberg’s hooks and barbs are structurally capable of penetrating into an
`
`
`
`15
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`interior surface of a vessel in a manner that fixedly secures graft 10 in place. PO
`
`Resp. 31–39; Pet. Appx. A2, 7–8 (citing Ex. 1001, Abstract, 3:60–62 (stating that
`
`“proximal attachment of the graft 10 to the inside wall of the aorta is accomplished
`
`by the hooks 14 which are located at the upper end of each strut 12”)). Rather,
`
`Patent Owner relies on the last “whereupon” in claim 1 as providing additional
`
`structural limitations, including that the projections are structurally capable of
`
`having blood flow force “cause” a relevant surface of hooks 14 to “tightly engage”
`
`a blood vessel and “fixedly secure” the device in place. PO Resp. 34–36.
`
`In this regard, Kornberg describes that “downward flow of blood holds the
`
`distal graft limbs 10A and 10B in place so that no mechanical attachment is
`
`necessary distally,” and that the “flow mechanism of the blood keeps the graft
`
`open.” Ex. 1006, 6:20–30; Reply 9–10. Because hooks 14 with barbs 15 are part
`
`of graft 10 in Kornberg’s device, we are persuaded by Petitioner’s position that
`
`blood flow through the interior passageway of Kornberg’s device “inherently
`
`produces” a force that helps cause hooks 14, including their trailing portion
`
`surfaces, to engage and maintain engagement in the vessel, which fixedly secures
`
`the device (“graft 10”) in place. Pet. 21; Reply 9–10.
`
`Based on descriptions in Kornberg that blood flow keeps graft 10 open (Ex.
`
`1006, 6:23–30), we are not persuaded that blood flow forces would have “no effect
`
`on the amount of securement provided by” hooks 14 located on the graft, as Patent
`
`Owner contends. PO Resp. 13, 36. Kornberg’s descriptions necessarily imply
`
`that, absent forces from blood flow, the graft would not stay open, and therefore,
`
`hooks 14 would disengage from the vessel. Thus, a preponderance of the evidence
`
`establishes that Kornberg inherently describes projections (hooks 14 with barbs
`
`15), including surfaces of trailing portions of the hooks, that are capable of being
`
`forced by blood fluid flow to cause engagement with the interior of the vessel to
`
`
`
`16
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`“fixedly secure” the device in place, even if manual forces by a physician put the
`
`device and hooks in place initially.
`
`We note that evidence submitted by Patent Owner further supports this
`
`finding. Patent Owner points us to the Malina article (Ex. 2012), discussing “the
`
`importance of using hooks and barbs to properly anchor the graft into the wall to
`
`prevent migration after deployment.” PO Resp. 17. Patent Owner quotes that
`
`article as stating that the “angle between the stent and its hooks and barbs is
`
`important” for the action of the hooks and barbs engaging “the aortic wall when
`
`the stent-graft is pulled distally by the bloodstream.” Id. (citing Ex. 2012, 6).
`
`Thus, Patent Owner’s own evidence indicates that it is the presence and angles of
`
`hooks and barbs, exactly as depicted in Kornberg, that allow blood flow to cause
`
`and maintain tight engagement of projections in the vessel and keep the device
`
`securely in place.
`
`Second, Patent Owner argues, in relation to the last “whereupon” clause in
`
`claim 1, that Kornberg’s hooks 14 do not “tightly engage” a blood vessel because
`
`Kornberg’s hooks pierce blood vessels, and Kornberg uses ring 16 to cover
`
`resulting punctures in the vessel wall. PO Resp. 36–39; Ex. 1006, 4:6–16. Thus,
`
`according to Patent Owner, Kornberg does not describe projections that do not
`
`pose a significant risk of perforating the vessel, as required in the last whereupon
`
`clause in claim 1.
`
`Patent Owner acknowledges, as Petitioner contends, that Kornberg discloses
`
`a range of hook angles (10°–45°) and hook lengths (2–8 mm), and that certain
`
`lower angles and smaller lengths within those ranges would result in hooks being
`
`less than the thickness of an aorta vessel wall. PO Resp. 37 (citing Ex. 1006, 3:62–
`
`65, 4:40–42; 2002 ¶ 68 (stating that the height of Kornberg’s anchors (hooks 14
`
`with barbs 15) “will be from about 0.35 to 5.66 mm, and because the aortic wall is
`
`
`
`17
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`typically only about 2 mm thick, the anchors will penetrate the aorta in the
`
`majority of cases”); Reply 7–9. Patent Owner argues, however, that because
`
`Kornberg does not describe a specific combination of hook angle and hook length,
`
`it does not describe a combination that “would necessarily result in a hook not
`
`perforating through the artery wall.” PO Resp. 37–38. Patent Owner further
`
`contends that Kornberg lacks “sufficient specificity,” as discussed in Atofina v.
`
`Great Lakes Chem. Corp., 441 F.3d 991, 999–1000 (Fed. Cir. 2006), because
`
`Kornberg does not disclose or point to any particular embodiment as desirable, but
`
`instead emphasizes perforation, as shown in Figure 10 and addressed by ring 16.
`
`PO Resp. 38–39.
`
`Atofina explains that a reference’s description of a broad genus range (e.g.,
`
`broad temperature range) does not anticipate necessarily every species within that
`
`range. Atofina, 441 F.3d at 999. Here, however, Kornberg discloses relatively
`
`narrow ranges of hook angles (10°–45°) and lengths (2–8 mm), as noted above,
`
`and specifically describes an angle of about 30º in the embodiment shown in
`
`Figure 2. Ex. 1006, 4:28–29. As Petitioner points out, any hook having a 10º
`
`angle (at any described length) and any hook being 2 mm length (at any described
`
`angle, including 30º shown in Figure 2) would not penetrate an aorta wall, which is
`
`about 2 mm thick. Reply 7–9; see also Ex. 2002 ¶ 68. Moreover, Kornberg
`
`expressly describes that its graft could be made without ring 16, indicating that
`
`Kornberg contemplates that ring 16 (used to cover “punctures in the aorta”) may be
`
`unnecessary in certain embodiments. Reply 9; Ex. 1006, 4:6–16.
`
`In addition, Kornberg clarifies that it is the presence of hooks with barbs, at
`
`the disclosed locations and angles, as shown in Figures 2 and 5, for example, that
`
`“inhibit[s] upward movement which might tend to dislodge the graft after it is
`
`positioned and attached to the aorta wall.” Ex. 1006, 3:60–4:1. In this context,
`
`
`
`18
`
`
`
`

`
`Case IPR2014-00100
`Patent 5,593,417
`
`
`although Kornberg describes and depicts hooks that pierce the aorta wall, the
`
`reference also discloses hooks that necessarily do not pierce the aorta wall in that it
`
`describes smaller angles and lengths for hooks 14, as well as ring 16 being
`
`optional. Id. at 3:60–4:46 (describing hook 14 angles and lengths, and attaching
`
`the graft to “the aorta wall”).
`
`The evidence before us does not persuade us that a “considerable difference”
`
`exists between possible (unrecited) ranges of angles and lengths

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket