throbber

`
`NUMBER VVV anoupVAR'r UNIT
`
`PATENT
`‘
`V
`‘
`$1111:
`’
`1;?“
`NUMBER
` SUBCLASS
`‘ SERIAL NUMBER
`
`538122? , 553
`
`EBUS
`
`
`fifiVIEfi
`IZFIMEFIDN, éEfiTTLE,
`£4154; THUfié‘iS D. L'Y’STER, BBTHELL,
`I'M; Df-‘INIEL J.
`F‘DHERS, BIBIINBRIDEE ISLfi, N133; BRAE‘FDFIE E. ELINER, BELLE‘JUE, W31;
`1
`(CLINTON S. CDLE, ‘ SEQTTLE,
`143153: CARLTDN B. MURERN, EAIHBRIDGE ISLR, HQ.
`
`
`
`$=FCQNTINUINE DQTABF$2$=§=$=IHI=$$=:==II=§=:I=*4I:Ik-‘Iilfi-‘IHIHI:
`
`,
`
`,
`
`LIE?!)
`
`THIS FeF‘PLN IS {3: CIF’ BF
`
`881’153,837 fiBflZifil'BS
`
`*WFUREIEM/PCT EPPLIEATmNs444-44 4444-444
`fl?
`1
`_
`
`”V
`
`CERHHGATE
`Nov 11 1997
`W GDRREWON
`
`
`
`
`APPLICANTS
`
`
`
`
`
`FIBRE I EM F I L I NE L ICENSE fiRfiHTED [I81 1 3’ 1"34
`* * 4‘ ’4 4‘ SMALL ENT I TY 1‘ :4; *35' 1:
`STATE 0 R SHEETS
`.
`ATTORNEY’S
`FILING FEE
` Clycs
`Forolgn pvlomy clalmnd
`CLAIMS
`CLAIMS
`RECEIVED
`COUNTRV C RWGS.
`DOCKET NO.
`35 USC 119 commons met
`Ely
`
`
`
`
`
`Vermnd and Acknowiedgtd
`um nu's In I Ils
`
`37444ng 18 E’Hi-Ij
`
`
`HLfiRBTPF/WV‘ EWC'
`1 (Q1101 pavam. 1WguU/sumi400
`
`
`1:5,77‘7’7 V[jV/QSHi/J‘Q/ITJN 921191
`1
`
`
`ELECTRGTHERQFY METHDD FIND QPPARQTHS
`
`‘3 24143324339520
`
`U50“.MWL‘TMOIM—lemlo-Jl)
`
`
`
`
`
`,
`
`~
`
`
`
`
`
`
`
`
`PARTS OF APPLICATION ‘
`.
`,/
`V
`FILED SEPARATELY
`A- -lications Examiner
`
`:
`NOTICEOF ALLOWANCE MAILED
`CLAIMS ALLOWED
`I
`V
`TotalCIaims ,
`Print Claim
`
`
`
`(17-11%
`.
`95??”
`2%
`
`V
`1V

`ISSUE FEE
`~' m
`
`
`
`
`Amount Due
`Sheets Drwg. Figs. 9M
`
`
`V 1
`<3
`12. 5 ,913
`MARVIN M. LATEEF
`ISSUE
`supenwsonv PATENT EXAMINER
`BATCH
`
`
`
`eaoupason
`Primary Examiner
`
`
`
`NUMBER :B/(
`I
`PREPARED FOR ISSUE
`
`
`
`
`Wm PTO-435A
`1ev‘ 8/92)
`
`1
`
`L|FECOR454-1002
`
`1
`
`LIFECOR454-1002
`
`

`

`353100“ 1
`
`HIINIWllfllflflllflfll|llllflflml
`
`CONTENTS
`
`'
`
`1
`
`APPRQVED FOR LICENSE D
`
`fWWi
`
`A
`
`at';
`m
`3'
`'é?
`'0 Maniac};
`.y
`'35
`)1
`r‘v
`J’
`n
`t!)
`g
`C;
`r"
`
`J ‘E
`
`y*’«5p9¢
`
`
`
`Q'HVHZ .
`APR I
`l
`1995
`
`2
`
`

`

`
`
`'
`
`,
`
`WM m Exmr.
`6 ? n9
`
`jgma % ..
`
`
`
`if m
`
`SEARCHED ‘ SEARCH NOTES
`
`3
`
`

`

`
`
`—-
`
`1-—
`
`m-
`
`
`—--e
`——
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`
` 97 _ I. ~__
`
`
`
`
`
`
`
`Staple Issue Sl‘i‘p‘flllé‘re M V
`
`‘
`
`'
`
`'
`
`INDEX OF CLAIMS\
`
`
`Clalm
`Date
`-
`L.
`v
`a E #3
`°? {k 0E"
`IE
`.9 § ‘5,
`1
`g
`0 «1
`wi- V“ w
`
`V,
`
`
`
`
`=3
`
`, 2 2
`g a
`4
`5
`5'
`e 6
`1
`7
`
`a9
`
`_
`
`L
`i
`'
`21’ 63
`
`H
`
`33>—
`r
`L
`
`”\3.
`
`
`
`
`
`
`
`
`H
`._
`79
`80
`
`svmaoLs
`
`V
`
`
`
`
`+ nN "
`I
`..
`A ..
`O ..
`
`.-
`.—
`g
`
`_
`
`__'
`
`86 I
`87as
`
`39
`90‘
`
`1
`
`
`
`
`
`95
`
`I99
`100
`
`‘” I ,
`-
`
`,
`
`*
`
`H
`
`,‘“
`
`"
`
`E‘
`:
`%
`
`‘4
`7,
`.A‘
`
`i
`'
`
`q
`
`;
`
`z
`E
`
`
`
`
`
`__
`7
`
`
`
`
`
`
`[o 10
`\\ 11
`\1 12
`r
`*"
`
`
`4
`
`

`

`[11] Patent Number:
`[19]
`United States Patent
`5,607,454
`Cameron et a1.
`Mar. 4, 1997
`[45] Date of Patent:
`
`lllIlIlllllllllllllllllllllIlllllllllIlllllllllllllillllllllllllllllllllll
`U8005607454A
`
`[54] ELECTROTIIERAPY METHOD AND
`APPARATUS
`
`[75]
`
`Inventors: David Cameron. Beanie; Thomas D.
`Lyster, Bothell; Daniel J. Powers,
`Bainbridge Island; Bradford E. Gliner,
`Bellevue; Clinton S. Cole, Seattle;
`Carlton B. Morgan, Bainbridge Island.
`all of Wash.
`
`[73] Assignee: Heartstream, Inc., Seattle, Wash.
`
`[21] Appl.ND.: 227,553
`[22]
`Filed:
`Apr. 14,1994
`
`Related U.S. Application Data
`
`[63] Continuation-impart of Ser. No. 103,837, Aug. 6, 1993.
`[51]
`
` ...... AGIN 1/39
`[52] U.S. C1.
`........................ 607/5; 607/7; 607/6; 607/74;
`607/62
`
`. 607/2. 4. 5-7.
`607/62, 74
`
`field of Search .
`
`[58]
`
`[56]
`
`3,211,154
`3,241,555
`3,706,313
`3,782,389
`3,860,009
`3,862,636
`3,886,950
`
`References Cited
`U.S. PATENT DOCUMENTS
`10/1965 Becker :1 n1.
`.
`3/1966 Caywood et al. .
`12/1972 Miltmi et a].
`.
`..... 607/8
`
`1/1974 Bell ......
`607/8
`1/1975 Ballet al.
`.
`607/5
`
`1/1975 Bell etal.
`.
`6/1975 Ukkestad et
`
`(List continued on next page.)
`FOREIGN PATENT DOCUMENTS
`0281219
`9/1983 European Pat. 03', .
`0315368
`5/1989 European PM. Ofl’.
`.
`0353341
`2/1990 European P81. 03. .
`0437104
`7/1991
`European Pat. 01f. ,
`0507504 10/1992 European Pat. 01f. .
`2070435
`9/1931 United Kingdom .
`2083363
`3/1932 United Kingdom .
`W093I16759
`9/1993 WIPO .
`
`9/1994 WIPO.
`WON/21327
`WOW/22530 10/1994 WIPO.
`OTHER PUBLICATIONS
`
`, Alfemess at 8.1., “The influence of shock waveforms on
`defibrillation efficacy," IEEE Engineering in Medicine and
`Biology, pp. 25—27 (Jun. 1990).
`Anderson at 211,, “The efiicacy of trapezoidal wave forms for
`ventricular defibrillation," Chest, 70(2):298—300 [1976).
`Blilie et al., “Predicting and validating cardiothoracic cur-
`rent flow using finite element modeling," PACE, 152563,
`abstract 219 (Apr. 1992).
`(List continued on next page.)
`Primary Examiner—Marvin M. Lateef
`Assistant Examiner—Kennedy .l. Schaetzle
`Altamey, Agent. or Firm—Morrison & Foerster
`[57]
`ABSTRACT
`
`An electrotherapy method and apparatus for delivering a
`multiphasic waveform from an energy source to a patient.
`The preferred embodiment of the method comprises the
`steps of charging the energy source to an initial level;
`discharging the energy source across the electrodes to
`deliver electrical energy to the patient in a multiphasic
`waveform; monitoring a patient-depehdent electrical param-
`eter during the discharging step; shaping the waveform of
`the delivered electrical energy based on a value of the
`monitored electrical parameter, wherein the relative duration
`of the phases of the multiphasic waveform is dependent on
`the value of the monitored electrical parameter. The pre-
`ferred apparatus comprises an energy source; two electrodes
`adapted to make electrical contact with a patient; 3 connect-
`ing mechanism forming an electrical circuit with the energy
`source and the electrodes when the electrodes are attached to
`a patient; and a controller operating the connecting mecha-
`nism to deliver electrical energy from the energy source to
`the electrodes in a multiphasic waveform the relative phase
`durations of which are based on an electrical parameter
`monitored during delivery of the electrical energy. The
`preferred defibrillator apparatus weighs less than 4 pounds
`and has a volume less than 150 cubic inches. and most
`preferably, weighs approximately three pounds or less and
`has a volume of approximately 141 cu. in.
`
`59 Claims, 4 Drawing Sheets
`
`
`
`
`
`5
`
`

`

`5,607,454
`Page 2M
`
`US. PATENT DOCUMENTS
`
`.
`
`.......................... 607/5
`5/1977 Pantridge et a1.
`4,023,573
`511982 Charbcnnier et a1.
`4,328,808
`.
`4,419,998 12/1983 Heath .
`4,473,078
`9/1984 Angel .......................................... 607/6
`4,494,552
`1/1985 Heath .
`4,504,773
`3/1985 Suzuki et 81..
`4,574,810
`3/1986 Lemar: .
`4,595,009
`6/1986 Leinders .
`4,610,254
`9/1986 Morgan et al. .
`4,619,265 10/1986 Morgan et a1.
`.
`4,637,397
`1/1987 Jones et a].
`.
`4,745,923
`5/1983 Winstrorn .
`4,800,883
`1/1989 Winstrom .
`4,821,723
`4/1989 Baker, Jr. et al. .
`4,840,177
`6/1989 Charbcnnier et a1.
`4,848,345
`7/1989 Zenkich .
`.
`4,550,357
`7/1989 Bach, Jr.
`4,953,551
`9/1990 Mehra et a1.
`,
`4,998,531
`3/1991 130ch et al.
`.
`5,078,134
`1/1992 Heilman et a1.
`5,083,562
`1/1992 dc Coriolis et al. .
`5,107,834
`4/1992 Ideker etal..
`5,111,813
`5/1992 Charbnnuier et a].
`5,111,816
`5/1992 Pless etal..
`5,207,219
`5/1993 Adams etal”
`5,215,081
`6/1993 Ostmff .
`5,222,480
`6/1993 Couche et a1..
`5,222,492
`6/1993 Morgan et 111,,
`5,230,336
`7/1993 Fain et al. ................................... 607/7
`5,237,989
`8/1993 Morgan et a1.
`.
`5.249.573 10/1993 Fincko et a1.
`............................... 607/6
`5,275,157
`1/1994 Morgan et a1.
`.
`5,305,291
`4/1994 K1011 et a1.
`.
`5,334,219
`8/1994 Kroll .
`607/5
`5,352,239
`10/1994 Flees
`
`5,370,664 12/1994 Morgan e
`..
`5,372,606 12/1994 Lang et a1.
`.................................. 607/8
`OTHER PUBLICATIONS
`
`.
`
`.
`
`Chapman et al., "Non-thoracotomy internal defibrillation:
`Improved eflicacy with biphasic shocks," Circulation,
`762312, abstract no. 1239 (1987).
`Cooper et al., “Temporal separation of the two pulses of
`single capacitor biphasic and dual monophasic waveforms,"
`Circulation, 84(4):612, abstract no. 2433 (1991).
`Cooper et 311., "the efl’ect of phase. separation on biphasic
`waveform defibrillation,"mCE, 16:471—482 (Mar. 1993).
`Cooper et al., “The effect of temporal separation of phases
`on biphasic waveform defibrillation efficacy,” The Annual
`International Conference ofthe IEEE Engineering in Medi-
`cine and Biology Society, 13(2):0766-0767 (1991).
`Crampton et al,, “Low-energy ventricular defibrillation and
`miniature defibrillators," JAMA, 235(21):2284 (1976).
`Dahlhack
`et
`al.,
`“Ventricular
`defibrillation with
`square—waves," The Lancet (Jul. 2, 1966).
`Eeht et 31., “Biphasic waveform is more efficacious than
`monophasic waveform for transthuracic cardioversion,"
`MOE, 16:914. abstract no. 2.56 (Apr. 1993).
`Feeser et 31.. "Strength—duration and probability of success
`curves for defibrillation with biphasic waveforms," Circu-
`lation, “(822128-2141 (1990).
`Guse et al., “Defibrillation with low voltage using a left
`ventricular catheter and four cutaneous patch electrodes in
`dogs,” PACE, 143143451 (Mar. 1991).
`Jones et al., “Decreased defibrillatorwinduced dysfunction
`with biphasic rectangular waveforms," Am J. Physiol,
`247:H792—796 (1984).
`
`"Defibrillator wavesbape optimization,”
`31.,
`et
`Jones
`Devices and Tech. Meeting, NIH (1982).
`Jones et al., “Improved defibrillator waveform safety factor
`with hiphasic waveforms." Am, J. Phyriol., 245516065
`(1983).
`Jones at 211,, “Reduced excitation threshold in potassium
`depolarized myocardial cells with symmetrical biphasic
`waveforms," J. Mal. Cell. Cardiol, 17(39):XXVII, abstract
`no. 39 (1985).
`Jude et al., “Fundamentals of Cardiopulmonary Resuscita-
`tion," EA. Davis Company. Philadelphia PA, pp. 98-104
`(1965).
`Kerber et a1., “Energy, current, and success in defibrillation
`and cardioversion: Clinical studies using an automated
`impedance—based method of energy adjustment," Circula-
`tion. 77(5):1038—1046 (1988).
`Knickerbocker et al., “A portable defibrillator,”IEEE Trans:
`on Power and Apparatus Systems, 69: 1089—1093 (1963).
`Kouwenhoven, “The development of the defibrillator,"
`Annals af/ntemal Medicine, 71(3):449—458 (1969).
`Langer ct a1., “Considerations in the development of the
`automatic implantable defibrillator," Medical Instrumenta-
`tion, 10(3):163-167 (1976).
`Lennan et a1. “Current—based versus energy—based ventricu-
`lar
`deflbrillation:
`A
`prospective
`study,"
`JA CC,
`12(5):1259—1264 (1988).
`Lindsay et al., “Prospective evaluation of a sequential pac-
`ing and high—energy bi—directional shock algorithm for
`transvenous cardioversion in patients with ventricular tachy»
`cardia," Circulation, 76(3):601—609 (1987).
`Mirowski et al., “Clinical
`treatment of life threatening
`ventricular tachyarrhythmias with the aut0matic implantable
`defibrillator," American Heart Journal, 102(2):265—270
`(1981).
`‘Termination of malignant ventricular
`211,,
`Mirowski ct
`arrhythmias with an implanted automatic defibrillator in
`human beings," The New England Jaumal of Medicine,
`303(6):322—324 (1980).
`Podolsky, “Keeping the beat alive,“ US, News & World
`Report (Jul. 22, 1991).
`Product Brochure for First Medic Semi‘Automatic Defibril—
`lators (1994), Spaoelabs Medical Products, 15220 NE. 40th
`Street, P.O. Box 97013. Redmond, WA 98073-9713.
`Product Brochure for the Shock Advisory System (1987),
`Physio—Control, 11811 Willows Road Northeast, PO. Box
`97006, Redmond, WA 98073-9706.
`Redd (editor), “Defibrillation with biphasic waveform may
`increase safety,
`improve survival,” Met/lines, pp,
`1—2
`(Jun—Jul. 1984).
`Saksena et al., “A prospective evaluation of single and dual
`current pathways for transvenous cardioversion in rapid
`ventricular tachycardia," PACE, 10:1130—1141 (Sept—Oct.
`1987).
`Sakstma at £11.. “Developments for future implantable car-
`dioverters
`and
`defibrillators,"
`PACE,
`10:1342~1358
`(Nov—Dec. 1987).
`Schuder "The role of an engineering oriented medical
`research group in developing improved methods and devices
`for achieving ventricular defibrillation: The University of
`Missouri experience," PACE, 16:95—124 (Jan. 1993).
`of
`Schuder
`et
`a1.
`“Comparison
`of
`effectiveness
`relay—switched, one-cycle quasisinusoidal waveform with
`critically damped sinusoid waveform in transthoracic
`dcfibrillation of 100461on calves,” Medical Instrumen—
`tation, 22(6):281—285 (1988).
`
`6
`
`

`

`5,607,454
`Page 3
`
`Schuder et a1.. "A multielectrode—tirne sequential laboratory
`defibrillator for the study of implanted electrode systems,"
`Amer: Soc. Am]? Int. Organs, XVIII:514—519 (1972).
`Schuder et al., “Defibrillation of 100 kg calves with asym-
`metrical. bi—directional, rectangular pulses,“ Card. Res.
`18:419-426 (1984).
`Schuder ct
`511.. "Development of automatic implanted
`defibrillator.” Devices & Then. Meeting NIH (1981).
`Schuder et al., “One-cycle bi—directional rectangular wave
`shocks for open chest defibrillation in the calf," Abs. Am.
`Soc. Amf. Intern. Orgm. 9:16.
`Schuder et a1.. "llnnsthoracic ventricular defibrillatiou in
`the 100 kg calf with symmetrical one-cycle lat—directional
`rectangular
`wave
`stimuli,"
`IEEE
`Trans.
`EME,
`30(7):415—422 (1983).
`Schuder et a1., '"l'ransthoracic ventricular defibrillation with
`square—wave stimuli: Onewhalf cycle, one-cycle, and mul-
`ticycle waveforms," Circ. Rest, XV:258—264 (1964).
`Schuder et al., “Ultrahigh—cnergy hydrogen thyratronlSCR
`Iii—directional waveform defibrillator," Med. 3: Biol. Eng. &
`Compile. 20:419—424 (1982).
`Schuder et a1., “Waveform dependency in defibrillating 100
`kg Calves," Devices & Tech. Meeting NIH (1982).
`Schuder et a1., “Waveform dependency in defibrillation,"
`Devices & Tech. Meeting NIH (1981).
`Stanton ct 21., “Relationship between defibrillation threshold
`and upper limit of vulnerability in humans," PACE, 152563,
`abstract 22] (Apr. 1992).
`Tang et a1.. “Strength duration curve for ventricular defibril-
`lation using biphasic waveforms,“ PACE, 10: abstract no. 49
`(Aug. 1987).
`
`Tang et al., “Venuicular dcfibrillation using biphasic wave-
`forms of different phasic duration." PACE. 10: abstract no.
`47 (Mar—Apr. 1987).
`Tang et at, “Ventricular defibrillation using biphasic wave-
`forms: The importance of phasic duration,"
`JACC,
`13(1):207—214 (1989).
`Walcott et al.. “Comparison of monophasic, biphasic, and
`the edmark waveform for external defibrillation,” PACE.
`15:563. abstract 218 (Apr. 1992).
`Wathen et 31., “Improved defibrillation eflicacy using four
`nonthoracotomy leads for sequential pulse defibrillation,"
`PACE, 15:563. abstract 220 (Apr. 1992).
`Wetberbee et al., “Subcutaneous patch electrode—A means
`to obviate thoracotomy for implantation of the automatic
`implantable cardioverter defibrillation system?” Ciro,
`72:384. abstract no. 1536 (1985).
`'Winkle “The implantable defibrillator in ventricular arrhythe
`ntias." Hospital Practice. pp. 149—165 (Mar. 1983).
`Winkle et al, “Improved low energy defibrillation efficacy
`in man using a biphasic truncated exponential waveform,”
`JACC, 9(2):l42A (1987).
`Zipes, “Sudden cardiac death," Circulation. 85(1):160—166
`(1992).
`Product information for Model H MSA Portable Defibrilla-
`tor (Bulletin No. 1108—2); 4 pp.
`Product information for MSA Portable Defibrillator (Bulle-
`tin No. 1108—1); 4 pp.
`
`
`
`7
`
`

`

`
`
`U.S. Patent
`
`Mar. 4,1997
`
`_ Sheet 1 of4
`
`5,607,454
`
`
`
`
`VOLTAGE
`
`l 1
`
`VOLTAGE
`
`8
`
`

`

`US. Patent
`
`Mar. 4, 1997
`
`Sheet 2 of 4
`
`5,607,454
`
`
`
`<monmm—o
`
`v.0."—
`
`9
`
`
`

`

`US. Patent
`
`Mar. 4,1997
`
`Sheet 3 of 4
`
`5,607,454
`
`
`
`10
`
`10
`
`

`

`555555555
`
`
`
`
`
`11
`
`

`

`5,607,454
`
`1
`ELECTROTHERAPY METHOD AND
`APPARATUS
`
`CROSS REFERENCE TO RELATED
`APPLICATION
`
`'
`
`This application is a continuation-impart of U.S. patent
`application Ser. No. 08/103,837 filed Aug. 6, 1993, the
`disclosure of which is incorporated herein by reference.
`
`BACKGROUND OF THE INVENTION
`
`This invention relates generally to an electrotherapy
`method and apparatus for delivering an electrical pulse to a
`patient’s heart. In particular,
`this invention relates to a
`method and apparatus for shaping the electrical waveform
`delivered by the defibrillator based on an electrical param-
`eter measured during delivery of the waveform The inven-
`tion also relates to a defibrillator design meeting certain
`threshold size and weight requirements.
`Sudden cardiac death is the leading cause of death in the
`United States. Most sudden cardiac death is caused by
`ventricular fibrillation, in which the heart’s muscle fibers
`contract without coordination, thereby interrupting normal
`blood flow to the body. The only efiective treatment for
`ventricular
`fibrillation is electrical defibrillation. which
`applies an electrical shock to the patient’s heart.
`To be efl‘ective, the defibrillation shock must he delivered
`to the patient within minutes of the onset of ventricular
`fibrillation. Studies have shown that defibrillation shocks
`delivered within one minute afier ventricular fibrillation
`begins achieve up to 100% survival rate. The survival rate
`falls to approximately 30% if 6 minutes elapse before the
`shock is administered. Beyond 12 minutes, the survival rate
`approaches zero.
`One way of delivering rapid defibrillation shocks is
`through the use of implantable defibrillators. Implantable
`defibrillators are surgically implanted in patients who have
`a high likelihood of needing electrotherapy in the future.
`Implanted defibrillators typically monitor the patient‘s heart
`activity and automatically supply electrotherapeutic pulses
`directly to the patient‘s heart when indicated Thus,
`implanted defibrillators permit the patient to function in_a
`somewhat normal fashion away from the watchful eye of
`medical personnel. Implantable defibrillators are expensive.
`however, and are used on only a small fraction of the total
`population at risk for sudden cardiac death.
`to the
`External defibrillators
`send electrical pulses
`patient's heart through electrodes applied to the patient's
`torso. External defibrillators are useful in the emergency
`room. the operating room, emergency medical vehicles or
`other situations where there may be an unanticipated need to
`provide electrotherapy to a patient on short notice. The
`advantage of external defibrillators is that they may be used
`on a patient as needed, then subsequently moved to he used
`with another patient.
`However, because external defibrillators deliver their
`electrodrerapeutic pulses to the patient’s heart indirectly
`(i.e., from the surface of the patient’s skin rather titan
`directly to the heart), they must operate at higher energies,
`voltages and/or eta-rents than implanted defibrillators. These
`high energy. voltage and current, requirements have made
`existing external defibrillators large, heavy and expensive,
`particularly due to thelarge size of the capacitors or other
`energy storage media required by these prior art devices. The
`size and weight of prior art external defibrillators have
`
`10
`
`IS
`
`30
`
`35
`
`2
`limited their utility for rapid response by emergency medical
`response [381115.
`Defibrillator waveforms, i.e., time plots of the delivered
`current or volmge pulses, are characterized according to the
`shape, polarity, duration and number of pulse phases. Most
`current external defibrillators deliver monophasic current or
`voltage ‘electrotherapeutic pulses. although some deliver
`biphasic sinusoidal pulses. Some prior an implantable
`defibrillators, on the other hand, use truncated exponential,
`biphasic waveforms. Examples of biphasic implantable
`defibrillators may he found in U.S. Pat No. 4,82l,723 to
`Baker, Jr, et al.; U.S. Pat. No. 5.083.562 to de Coriolis et al.-,
`U.S. Pat. No. 4,800,883 to Winstrom; U.S. Pat. No. 4,850,
`357 to Bach, Jr; U.S. Pat. No. 4,953,551 to Mehra et al.; and
`U.S. Pat. No. 5,230,336 to Fain et al.
`Because each implanted defibrillator is dedicated to a
`single patient, its operating parameters, such as electrical
`pulse amplitudes and total energy delivered, may be efi'ec-
`Lively titrated to the physiology of the patient to optimize the
`defibrillator’s efi‘ectiveness. Thus, for example,
`the initial
`voltage, first phase duration and total pulse duration may be
`set when the device is implanted to deliver the desired
`amOunt of energy or to achieve a desired start and end
`voltage differential (i.e., a constant tilt). Even when an
`implanted defibrillator has the ability to change its operating
`parameters to compensate for changes in the impedance of
`the defibrillators leads and/or the patient’s heart (as dis-
`cussed in the Fain patent), the range of potential impedance
`changes for a single implantation in a single patient is
`relatively small.
`In contrast, because external defibrillator electrodes are
`not in direct contact with the patient’s heart. and because
`external defibrillators must be able to be used on a variety of
`patients having a variety of physiological differences. exter-
`nal defibrillators must operate according to pulse amplitude
`and duration parameters that will be eifective in most
`patients, no matter what
`the patient’s physiology, For
`example, the impedance presented by the tissue between
`external defibrillator electrodes and the patient’s heart varies
`from patient to patient, thereby varying the intensity and
`waveform shape of the shock actually delivered to the
`patients heart for a given initial pulse amplitude and dura-
`tion. Pulse amplitudes and durations effective to treat low
`impedance patients do not necessarily deliver effective and
`energy efiiclent treatments to high impedance patients.
`External defibrillators may be subjected to extreme load
`conditions which could potentially damage the waveform
`generator circuits. For example, improperly applied defibril-
`lator electrodes may create a very low impedance current
`path during the shock delivery, which could result in execs»
`sively high current within the waveform circuit, Thus, an
`external defibrillator has an additional design requirement to
`limit the peak Current to safe levels in the waveform circuit,
`which is not normally a concern for implanted defibrillators.
`Prior art defibrillators have not fully addressed the patient
`variability problem. One prior an approach to this problem
`was to provide an external defibrillator with multiple energy
`settings that could be selected by the user. A common
`protocol for using such a defibrillator was to attempt
`. defibrillata'on at an initial energy setting suitable for defibril-
`lating a patient of average impedance, then raise the energy
`setting for subsequent defibrillation attempts in the event
`that the initial setting failed. The repeated defibrillation
`attempts require additional energy and add to patient risk.
`Some prior art defibrillators measure the patient imped-
`ance, or a pararueter related to patient impedance, and alter
`
`45
`
`50
`
`55
`
`65
`
`12
`
`12
`
`

`

`5,607,454
`
`4
`BRIEF DESCRIPTION or THE DRAWINGS
`
`3
`the shape of a subsequent defibrillation shock based on the
`earlier measurement. For example, the implanted defibril-
`lator described in the Fain patent delivers a defibrillation
`shock of predetermined shape to the patient's heart
`in
`response to a detected arrhythmia. The Fain device measures
`the system impedance during delivery of that shock and uses
`the measured impedance to alter the shape of a subsequently
`delivered shock.
`Another example of the measurement and use of patient
`impedance information in prior art defibrillators is described
`in an article written by R. E. Kerber, et at, “Energy. current,
`and success in defibrillation and cardioversion." Circulation
`(May 1988). The authors describe an external defibrillator
`that administers a test pulse to the patient prior to adminis»
`tering the defibrillation shock. The test pulse is used to
`measure patient
`impedance;
`the defibrillator adjusts the
`amount of energy delivered by the shock in response to the
`measured patient impedance. The shape of the delivered
`waveform is a damped sinusoid.
`Prior art disclosures of the use of truncated exponential
`biphasic waveforms in implantable defibrillators have pro-
`vided little guidance for the design of an external defibril-
`lator that will achieve acceptable defibrillation or cardiover-
`siort
`rates across a wide population of patients. The
`defibrillator operating voltages and energy delivery require-
`ments affect the size, cost. weight and availability of com-
`ponents. In particular, operating voltage requirements aflcct
`the choice of switch and capacitor technologies. Total
`energy delivery requirements affect defibrillator battery and
`capacitor choices. Thus. even if an implantable defibrillator
`and an external defibrillator both deliver waveforms of
`similar shape, albeit with different waveform amplitudes, the
`actual designs of the two defibrillators would he radically
`different.
`
`SUMMARY OF THE INVENTION
`
`10
`
`15
`
`20
`
`25
`
`30
`
`35
`
`
`
`FIG. 1 is a schematic representation of a low~tilt biphasic
`cleatrotherapcutic waveform.
`FIG. 2 is a schematic representation of a high-tilt biphasic
`electrotherapeutie waveform.
`FIG. 3 is a block diagram of a defibrillator system
`according to a preferred embodiment of the invention.
`FIG. 4 is a schematic circuit diagram of a defibrillator
`system according to a preferred embodiment of this inven-
`tion.
`FIG. 5 is an external View of a defibrillator according to
`a preferred embodiment of this invention.
`FIG. 6 is a partial cutaway view of a defibrillator accord,
`ing to a preferred embodiment of this invention.
`
`DETAILED DESCRIPTION OF THE
`PREFERRED EMBODIMENT
`
`For any given patient and for any given defibrillator
`system design, whether implantable or external, there is an
`optimal biphasie waveform for treating a particular kind of
`arrhythmia. This principle is used when implanting defibril-
`lators; as noted ab0ve, implanted defibrillators are titrated to
`the patient at the time of implant. External defibrillators, on
`the other hand, must he designed to be efi'ective in a wide
`population of patients.
`For example. FIGS. 1 and 2 illustrate the patient-to-
`patient differences that an external defibrillator design must
`take into account. These figures are schematic representa-
`tions of truncated exponential biphasic waveforms delivered
`to two difierent patients from an external defibrillator
`according to the electrotherapy method of this invention for
`defibrillation or cardioversion. In these drawings, the verti-
`cal axis is voltage. and the horizontal axis is time. The
`principles discussed here are applicable to waveforms
`described in terms of current versus time as well.
`The waveform shown in FIG.
`1 is called a low-tilt
`waveform. and the waveform shown in FIG. 2 is called a
`high-tilt waveform, where tilt H is defined as a percent as
`follows:
`
`iAi—iDi
`“tn—“W
`
`As shown in FIGS. 1 and 2, A is the initial first phase voltage
`and D is the second phase terminal voltage. The first phase
`terminal voltage B results from the exponential decay over
`time of the initial voltage A through the patient, and the
`second phase tannins] voltage D results from the exponen-
`tial decay of the second phase initial voltage C in the same ,
`manner. The starting voltages and first and second phase
`durations of the FIG. 1 and FIG. 2 waveforms are the same;
`the differences in end voltages B and D reflect patient
`dilferences.
`We have determined that, for a given patient, externally-
`applied truncated exponential biphasic waveforms defibril-
`late at lower voltages and at lower total delivered energies
`than extemally-applied monophasic waveforms. In addition,
`we have determined that there is a complex relationship
`between total pulse duration, first to second phase duration
`ratio, initial voltage, total energy and total tilt in the delivery
`of an effective eardioversion waveform. Thus, it is possible
`to design a defibrillator and defibrillation method that is
`eifective not only for a single patient (as in most prior art
`implantable defibrillators) but is also efl'eetive for a broad
`
`
`
`4O
`
`45
`
`This invention provides a defibrillator and defibrillation
`method that automatically compensates for patient-to-pa-
`tient diferences in the delivery of electrotherapeutic pulses
`for defibrillation and cardioversion. The defibrillator has an
`energy source that may be discharged through electrodes to
`administer a truncated exponential biphasic voltage or cur-
`rent puise to a patient.
`The preferred embodiment of the method comprises the
`steps of charging the energy source to an initial
`level;
`discharging the energy source across the electrodes to
`deliver electrical energy to the patient in a multiphasie
`waveform; monitoring apatient-dependent electrical param-
`eter during the discharging step; shaping the waveform of 50
`the delivered electrical energy based on a value of the
`monitored electrical parameter. wherein the relative duration
`of the phases of the multiphasie waveform is dependent on
`the value of the monitored electrical parameter.
`The preferred apparatus comprises an energy source; two
`electrodes adapted to make electrical contact with a patient:
`a connecting mechanism forming an electrical circuit with
`the energy source and the electrodes when the electrodes are
`attached to a patient; and a controller operating the connect-
`ing mechanism to deliver electrical energy from the energy
`source to the elecuodes in a multiphasic waveform the
`relative phase durations of which are based on an electrical
`parameter monitored during delivery of the electrical energy.
`The preferred defibrillator apparatus weighs less than 4
`pounds and has a volume less than 150 cubic inches, and
`most preferably, weighs approximately three pounds or less
`and has a volume of approximately 141 cu. in.
`
`55
`
`60
`
`65
`
`13
`
`13
`
`

`

`
`
`5,607,454
`
`5
`population of patients. In addition, it is also possible to meet
`external defibrillator design requirements regarding the size,
`weight and capacity of the defibrillator energy source while
`still meeting the needs of a wide patient population.
`Up to a point, the more energy delivered to a patient in an
`electrodterapeutie pulse, the more likely the defibrillation
`attempt will succeed. Low-tilt biphasic waveforms achieve
`effective defibrillation rates with less delivered energy titan
`high-tilt waveforms. However,
`low-tilt waveforms are
`energy inefiicient, since much of the stored energy is not
`delivered to the patient. On the other hand, defibrillators
`delivering high-tilt biphasic waveforms deliver more of the
`stored energy to the patient than defibrillators delivering
`low-tilt waveforms while maintaining high efiieacy up to a
`certain critical tilt value. Thus, for a given capacitor, is given
`initial voltage and fixed phase durations, high impedance
`patients receive a waveform with less total energy and lower
`peel: currents but better conversion properties per unit of
`energy delivered, and low impedance patients receive a
`waveform with more delivered energy and higher peak
`currents.
`There appears to be an optimum tilt range in which high
`and low impedance patients will receive efi'ective and efii-
`cient therapy from an external defibrillator. An optimum
`capacitor charged to a predetermined voltage can be chosen
`to deliver an eifeetive and efficient waveform across a
`population of patients having a variety of physiological
`differences. For example, the defibrillator may operate in an
`open loop,
`i.e., without any feedback regarding patient
`parameters and with preset pulse phase durations which will
`be effective for a certain range of patients. The preset
`parameters of the waveforms shown in FIG. 1 and 2 are
`therefore the initial voltage A of the first phase of the pulse,
`the duration E of the first phase, the interphase duration G,
`and the duration F of the second phase. The terminal voltage
`B of the first phase, the initial voltage C of the second phase,
`and the terminal voltage D of the second phase are depen-
`dent upon the physiological parameters of the patient and the
`physical connection between the electrodes and the patient.
`For example, if the patient impedance (i.e., the total
`impedance between the two electrodes) is high, the amount
`of voltage drop (exponential decay) from the initial voltage
`A to the terminal voltage B during time E will be lower (FIG.
`1) than if the patient impedance is low (FIG. 2). The same
`is true for the initial and terminal voltages of the second
`phase during time F. The values of A, E, G and F are set to
`optimize defibrillation and/or cardioversion efficacy across a
`population of patients. Thus, high impedance patients
`receive a low-tilt waveform that is more etfective per unit of
`delivered energy, and low impedance patients receive a
`high-tilt waveform that delivers more of the stored energy
`and is therefore more energy efficient
`In order to ensure that the delivered shock will be within
`the optimum tilt range for an extended range ofpatients. this
`invention provides a defibrillator method and apparatus for
`adjusting the characteristics of the defibrillator waveform in
`response to a real-time measurement of a patient-dependent
`electrical parameter. FIG. 3 is a block diagram showing a
`preferred embodiment of the defibrillator system.
`The defibrillator system 30 comprises an energy source 32
`to provide a voltage or current pulse. In one p

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket