`DECLARATION OF WAYNE C. McDANIEL, Ph.D.
`
`I.
`
`BACKGROUND AND QUALIFICATIONS
`
`
`
`(1) My name is Wayne Charles McDaniel. I am currently Associate Director of
`
`the Technology Management and Industry Relations office at the University of Missouri-
`
`Columbia. I am also an Adjunct Professor of Electrical and Computer Engineering at the
`
`University of Missouri. During my career, I have worked extensively in biomedical
`
`engineering research involving cardiac therapy and defibrillation. I am expert in the areas of
`
`internal atrial and ventricular defibrillation, external ventricular defibrillation, experimental
`
`methods for defibrillation research, and cardiac safety of stun guns.
`
`(2)
`
`I hold a Bachelor of Arts in Biology, a Master of Science in Electrical
`
`Engineering with a biomedical engineering emphasis, and a Doctorate of Philosophy in
`
`Electrical Engineering with a biomedical engineering emphasis from the University of
`
`Missouri-Columbia.
`
`(3)
`
`From 2001 to 2011, I held the position of Senior Licensing and Business
`
`Development Associate of the Technology Management and Industry Relations office at the
`
`University of Missouri-Columbia. From 1987 to 2001, I was a Research Assistant Professor
`
`of Cardiothoracic Surgery at the University of Missouri-Columbia. From 1993 to 2001, I was
`
`
`
`1
`
`LIFECOR212-1003
`
`
`
`US Patent 6,047,212
`
`Acting Director of Cardiothoracic Surgery Laboratory Investigation at the University of
`
`Missouri.
`
`(4)
`
`I have over 35 years of experience in the biomedical engineering field and
`
`have published extensively in electrical ventricular defibrillation. For over twenty-five years,
`
`I have conducted and received numerous grants for research relating to cardiac therapy and
`
`defibrillation.
`
`(5)
`
`I was one of the pioneers of the biphasic waveform that is now used in
`
`virtually all automatic implantable cardioverter defibrillators and virtually all transthoracic
`
`defibrillators, including automatic external defibrillators.
`
`(6)
`
`I have authored or co-authored 34 published articles relating to cardiac
`
`therapy and defibrillation, including articles titled “Transthoracic Defibrillation of Dogs with
`
`Edmark, Biphasic, and Quadriphasic waveforms,” “Double-pulse transthoracic defibrillation
`
`in the calf using percent fibrillatory cycle length as spacing determinate,” and “Relationship
`
`between efficacy and frequency domain characteristics of defibrillatory shocks.”
`
`(7)
`
`I have given 46 presentations at national or international meetings, including
`
`a presentation entitled “Multiphasic truncated exponential waveforms require less peak
`
`current for atrial defibrillation than optimal biphasic waveforms” to the 22nd Annual Scientific
`
`Sessions of the North American Society for Pacing and Electrophysiology, in Boston,
`
`Massachusetts in May of 2001 and a presentation entitled “Comparison of the Efficacy of
`
`Two Transthoracic Biphasic Waveform Defibrillators” to the Europace 2003 Congress in
`
`
`
`2
`
`
`
`US Patent 6,047,212
`
`Dec. 2003 in Paris, France. I have presented at 33 colloquiums and symposiums, including
`
`presentations on ventricular and atrial defibrillation.
`
`(8)
`
`I am the sole inventor on U.S. Patent No. 6,738,664 entitled “Method of and
`
`apparatus for atrial and ventricular defibrillation or cardioversion with an electrical waveform
`
`optimized in the frequency domain,” which issued on May 18, 2004.
`
`(9)
`
`A copy of my C.V. is attached as Appendix A.
`
`II.
`
`STATUS AS AN INDEPENDENT EXPERT WITNESS
`
`(10)
`
`I have been retained in this matter by Fish & Richardson P.C. to provide
`
`various opinions regarding U.S. Patent No. 5,735,879 (“the ‘879 patent); U.S. Patent No.
`
`5,749,905 (“the ‘905 patent”); U.S. Patent No. 6,047,212 (“the ‘212 patent); U.S. Patent No.
`
`5,607,454 (“the ‘454 patent); U.S. Patent No. 5,836,978 (“the ‘978 patent”); U.S. Patent No.
`
`5,749,904 (“the ‘904 patent”); U.S. Patent No. 5,593,427 (“the ‘427 patent”); and U.S. Patent
`
`No. 5,803,927 (“the ‘927 patent”) (collectively, “the Philips Waveform Patents”). I am being
`
`compensated at the rate of $300 per hour for my work. My fee is not contingent on the
`
`outcome of this matter or on any of the opinions I provide below.
`
`(11)
`
`I have been advised that Fish & Richardson represents ZOLL Lifecor Corp. in
`
`this matter. I have no financial interest in ZOLL Lifecor Corp.
`
`(12)
`
`I have been advised that Philips Electronics North America Corp. (“Philips” or
`
`“Patent Owner”) owns the Philips Waveform Patents. I have no financial interest in Philips
`
`Electronics North America Corp. or in the Philips Waveform Patents.
`
`
`
`3
`
`
`
`US Patent 6,047,212
`
`III.
`
`MATERIALS CONSIDERED
`
`
`
`(13)
`
`In arriving at the opinions set forth herein, I have reviewed the Philips
`
`Waveform Patents and relevant portions of their respective file histories.
`
`(14) Additional materials that I have reviewed and relied upon in arriving at the
`
`opinions set forth herein are: (1) Bell (Appendix B); (2) Pless (Appendix C); (3) Kroll
`
`(Appendix D); (4) Schuder (Appendix E); (5) Swanson (Appendix F); (6) De Coriolis
`
`(Appendix G); (7) Ideker (Appendix H); (8) Fain (Appendix I); (9) Baker (Appendix J); (10)
`
`Packer (Appendix K); (11) Hahn (Appendix L); (12) Bach (Appendix M); (13) Adams
`
`(Appendix N); (14) Herleikson (Appendix O); (15) Cameron (Appendix P); and (16) Morgan
`
`(Appendix Q).
`
`IV.
`
`DESCRIPTION OF THE RELEVANT FIELD AND RELEVANT TIMEFRAME
`
`(15) Based on my review of the Philips Waveform Patents and the materials listed
`
`in Appendices B-P, I conclude that the relevant field of the Philips Waveform Patents for
`
`purposes of my testimony is waveforms used for defibrillation, and apparatus and
`
`techniques for generating and delivering such waveforms. I have been advised that the
`
`relevant timeframe is August 1993, which is the date that the applications that lead to the
`
`Philips Waveform Patents were filed.
`
`(16) As described in Section I above, I have extensive experience in the field of
`
`defibrillation waveforms, and apparatus and techniques for generating and delivering such
`
`waveforms.
`
`
`
`4
`
`
`
`US Patent 6,047,212
`
`
`
`V.
`
`PERSON OF ORDINARY SKILL IN THE ART
`
`(17)
`
`I have been advised that “a person of ordinary skill in the relevant field” is a
`
`mythical person to whom an expert in the relevant field could assign a routine task with
`
`reasonable confidence that the task would be successfully carried out. Here, the relevant
`
`field is waveforms used for defibrillation, and apparatus and techniques for generating and
`
`delivering such waveforms. Because these devices are used to deliver a shock to a
`
`patient’s heart, people engaged in developing these devices and related methods need to
`
`have a high level of skill. Based upon my experience in this area, one of ordinary skill in the
`
`art in this field at the relevant time frame would have had an advanced (post-Bachelor’s)
`
`degree in electrical engineering, biomedical engineering, or some closely related field, with
`
`at least 5 years of work experience in one or more of these fields, and at least 5 years of
`
`experience in developing (e.g., designing or implementing) medical devices for defibrillation,
`
`pacing, and/or cardiac medical devices (which experience could have overlapped in whole
`
`or part with the at least 5 years of experience in the fields of electrical engineering or
`
`biomedical engineering), or the equivalent of such experience. The person of ordinary skill
`
`in the art also must have been intimately familiar with the design of, theory behind,
`
`principles of operation of, and intended use of defibrillators, as well as the principles of
`
`human physiology that underlie the indications of use for defibrillators (cardiac arrest and
`
`
`
`5
`
`
`
`US Patent 6,047,212
`
`ventricular fibrillation), and the theories as to why the delivery of certain shocks may be
`
`useful to correct these conditions.
`
`(18) Based on my experiences, I have a good understanding of the capabilities of
`
`a person of ordinary skill in the relevant field. Indeed, in addition to being a person of at
`
`least ordinary skill in the art, I have worked closely with many such persons over the course
`
`of my career, and I have regularly taught material fundamental to the art in my role as
`
`professor and researcher over the past 35 years.
`
`VI.
`
`BACKGROUND OF THE RELEVANT TECHNOLOGY
`
`(19) Sudden cardiac death is the most common mode of death in our adult
`
`population, accounting for an estimated 30 percent of all natural deaths. Sudden cardiac
`
`arrest occurs when the heart stops beating in an organized way and instead begins
`
`fibrillating in a random manner. Many of the hearts of sudden cardiac death victims appear
`
`essentially normal at the time of autopsy, causing some investigators to refer to them as
`
`hearts that were “too good to die.” One major cause of sudden cardiac death is a
`
`phenomenon called ventricular fibrillation (VF) which occurs in structurally good hearts over
`
`900 times a day in the U.S. alone in out-of-hospital patients. In VF, the individual muscle
`
`fibers of the heart no longer contract in unison, but rather there are waves of contraction
`
`that run randomly through the heart. A direct consequence of VF is the inability of the heart
`
`to pump blood, which means the patient will suffer irreversible brain damage and then death
`
`if not treated promptly.
`
`
`
`6
`
`
`
`US Patent 6,047,212
`
`(20) Electrical defibrillation is a treatment of choice for ventricular fibrillation and
`
`consists of delivering a therapeutic dose of electrical energy to the patient’s heart, which
`
`depolarizes a critical mass of the heart muscle. This depolarization terminates the
`
`dysrhythmia, allowing the patient’s normal sinus rhythm to be reestablished. Defibrillators,
`
`which were first developed in the mid-1900s, are devices that restore normal contractions in
`
`the heart muscle by delivering a powerful shock via electrodes attached to the patient. The
`
`goal of a defibrillation shock is to deliver the appropriate amount of current to the patient to
`
`reestablish normal sinus rhythm, while minimizing damage to the patient’s heart. The shock
`
`is generally delivered by charging an energy source, such as a capacitor over time, and
`
`then closing a switch to release the charge. A shock takes just a fraction of a second.
`
`(21) The flow of electrical current released during the shock has a shape that can
`
`be characterized by a time versus voltage graph that shows the “waveform” of the shock.
`
`For example, if a shock is delivered freely, the “waveform” is simply an exponentially
`
`decaying shape that approaches zero after a relatively long period of time. For successful
`
`defibrillation, the amount of current delivered to the heart varies between patients on
`
`account of a given patient’s body mass, temperature and diaphoresis—collectively referred
`
`to as patient impedance. To account for patient impedance, the waveform of the shock can
`
`be modified, or shaped, by altering the initial and/or terminal voltage used in delivering the
`
`shock, as well as the duration of the shock. Such shaping allows one to modify the amount
`
`of current released into the patient.
`
`
`
`7
`
`
`
`US Patent 6,047,212
`
`(22) Modern defibrillators can either be external units, where electrodes are placed
`
`on the patient’s torso to deliver a shock to the patient’s heart, or internal devices, in which
`
`case a small electrical impulse generator is implanted into the body of a patient at risk of
`
`cardiac arrhythmia. The implantable device monitors the patient’s cardiac rhythms and
`
`automatically delivers a therapeutic shock if dysrhythmia is detected. Because external
`
`defibrillators indirectly deliver therapeutic shocks to the heart (i.e., through layers of fat,
`
`tissue, and skin), it is important to be able to modify different aspects of the waveform for
`
`patients of different sizes and body types.
`
`Brief History of Defibrillators
`
`(23) The earliest recorded defibrillation in humans with internal electrodes was
`
`accomplished in 1947. The first electrical waveform used for defibrillation was 60 Hz
`
`alternating current (AC), which is also used as standard household current. Similarly, the
`
`first recorded defibrillation of a human with external electrodes was accomplished in 1956,
`
`also by an AC waveform defibrillator.
`
`(24) AC waveform defibrillators were replaced in the 1960s by the development of
`
`the Lown and Edmark waveform defibrillators. These defibrillators comprised a resistor (R),
`
`an inductor (L), and a capacitor (C), and were therefore referred to as RLC defibrillators.
`
`RLC defibrillators were considered to be an advancement over AC defibrillators in that they
`
`were portable and did not need to be tethered to a power line. Both the Lown and Edmark
`
`waveforms delivered predominantly monophasic waveforms. However, under certain
`
`
`
`8
`
`
`
`US Patent 6,047,212
`
`patient resistances, one or more negative phases would be observed. Therefore, both
`
`Edmark and Lown waveform defibrillators would generate multiphasic waveforms under the
`
`right circumstances.
`
`(25)
`
`In 1966, John Schuder et al. published a study of the transthoracic
`
`defibrillation efficacy of triangular and trapezoidal waveforms in dogs. This study followed a
`
`similar study of monophasic square waveforms. This study found that a long slow decay on
`
`a waveform would reduce the efficacy, whereas truncating the descending triangular
`
`waveform yielded a superior waveform. This study laid the foundation for capacitor
`
`discharge defibrillators not containing an inductor.
`
`(26)
`
`In 1971, Schuder et al. extended this work to true capacitive discharge
`
`waveforms (as opposed to triangular waveforms above) and found that truncation of long-
`
`duration waveforms significantly improved defibrillation success. The Schuder lab
`
`pioneered this monophasic truncated exponential (MTE) waveform (which is sometimes
`
`called the Schuder waveform) for defibrillation. One advantage to the MTE waveform was
`
`the reduction of the peak current, which was believed to cause cardiac damage with the
`
`Edmark and Lown waveforms. Another advantage to the MTE waveform (when
`
`implemented in clinical devices) was that one could compensate for differing thoracic
`
`impedance values seen with different patients by delivering a constant value of delivered
`
`energy across a wide range of patient impedances. This was done by monitoring a patient-
`
`related electrical parameter and terminating the shock based on that measured value.
`
`
`
`9
`
`
`
`US Patent 6,047,212
`
`(27) Around the same time, the Schuder lab published a research study of the first
`
`implantable defibrillator and incorporated the MTE waveform in that defibrillator. The MTE
`
`waveform was particularly well suited for the implantable defibrillator, in that it could be
`
`generated without the use of an inductor, which would be too large and heavy for use in an
`
`implantable device.
`
`(28)
`
`In the early 1980s, John Schuder, in collaboration with Janice Jones,
`
`theorized that a biphasic waveform would cause less post-shock dysfunction than the MTE
`
`waveform for defibrillation. Biphasic waveforms differ from monophasic waveforms in that
`
`the shock delivered to the patient’s heart in a monophasic shock is delivered in one
`
`direction only from one electrode to the other. In biphasic waveform defibrillation, the shock
`
`passes in one direction from one electrode to the next and then reverses direction, traveling
`
`back to the original electrode. That is, in a biphasic waveform the pulse alternates between
`
`positive and negative polarities.
`
`(29) The Schuder lab began a series of studies of the biphasic waveform with the
`
`first study involving what are now called biphasic truncated exponential (BTE) waveforms.
`
`The first study involving implantable electrodes was promising (published in 1981), so the
`
`Schuder lab modified its high power research defibrillator to enable it to generate biphasic
`
`waveforms for external defibrillation of human-sized animals (device made public in 1982).
`
`The first study with the modified research defibrillator involved symmetric rectangular
`
`biphasic waveforms (equal constant currents and equal phase durations), which was
`
`
`
`10
`
`
`
`US Patent 6,047,212
`
`published in 1983. Then a study of asymmetric rectangular biphasic waveforms (unequal
`
`constant currents and equal phase durations) was published in 1984.
`
`(30) The studies of Schuder and Jones attracted the attention of both an
`
`implantable defibrillator manufacturer (Cardiac Pacemaker Inc.) and an external defibrillator
`
`manufacturer (Physio Control). Cardiac Pacemakers Inc. (CPI) and Physio Control worked
`
`collaboratively with the Schuder lab, the Jones lab, and other labs such as Ray Ideker’s lab.
`
`About once per year during this time period, representatives of each of the labs and each of
`
`the companies met for a research conference. It was known to all researchers involved at
`
`the time that the biphasic waveforms showed promise to dramatically improve both the
`
`implantable defibrillators made by CPI and the transthoracic defibrillators made by Physio
`
`Control.
`
`(31)
`
`In 1984 the Schuder lab published its study of asymmetric biphasic truncated
`
`exponential (BTE) waveforms (exponentially decaying currents and equal phase durations)
`
`used in the transthoracic defibrillation of 100 kg calves. This study examined the efficacy of
`
`several waveforms that were capable of being generated by a single capacitor bank.
`
`Specifically, when a single capacitor BTE waveform is generated clinically, a single
`
`capacitor bank is used, and the voltage and current delivered to a patient decays as the
`
`capacitor discharges. At some point during this discharge, the shock is interrupted and the
`
`polarity is reversed before reinitiating the shock. Then at some later point, the shock is
`
`terminated.
`
`
`
`11
`
`
`
`US Patent 6,047,212
`
`(32) By 1985 it was known that biphasic waveforms were an improvement over
`
`monophasic waveforms for both internal defibrillation and external defibrillation. It was
`
`further known that the best known method of generating biphasic waveforms for both
`
`internal and external defibrillation involved the use of a single capacitor biphasic truncated
`
`exponential waveform. The Schuder lab’s 1984 BTE waveform study further established
`
`that the individual pulse durations ranging from 2.8 to 7.54 ms (overall pulse durations of 5.6
`
`to 15.08 ms) showed positive results. The study also showed that the ideal capacitance
`
`values for human defibrillation were in the range of 64 to 196 microfarads. The general
`
`recognition of the biphasic waveform as a more effective defibrillation waveform was again
`
`illustrated in a 1988 paper from the Schuder lab entitled “General Superiority of Biphasic
`
`Over Uniphasic Shocks in Cardiac Defibrillation.” This paper summarized all of the studies
`
`to date of biphasic waveforms for defibrillation and concluded that biphasic waveforms were
`
`more effective for defibrillation than monophasic waveforms.
`
`(33) Following these advancements, further methods of modifying the biphasic
`
`waveform have subsequently been employed in defibrillators. For example, internal and
`
`external defibrillators containing an output circuit having four legs arranged in the form of
`
`the letter “H” have been developed. “H-bridge” circuits were employed in internal and
`
`external defibrillators to conduct a range of defibrillation pulse energies. Selectively
`
`switching on pairs of the switches in an H-bridge circuit allows the pulse generated by a
`
`single capacitance to alternate between positive and negative polarities. Another way
`
`
`
`12
`
`
`
`US Patent 6,047,212
`
`researchers modified biphasic waveforms was by adjusting the truncation or “tilt” of the
`
`waveform. Tilt specifically refers to the percentage difference between the ending voltage
`
`value and the initial voltage value. Modifications to the tilt of the waveform allow for control
`
`of the amount of energy delivered to the patient, thereby reducing the energy requirements
`
`of the defibrillator and preventing myocardial damage from overexposure.
`
`(34)
`
`In sum, by August of 1993 it was well known that one could compensate for
`
`differing patients’ impedances by (a) varying the predetermined duration of time over which
`
`discharge occurred, (b) measuring voltage decay (or equivalent electrical parameter) and
`
`stopping discharge when it reached a predetermined level, or (c) various combinations of
`
`both. At the same time, by 1993 it would have been obvious to apply knowledge and
`
`techniques learned from monophasic waveform research and implementations to biphasic
`
`waveform implementations, and to apply knowledge and techniques learned from
`
`implantable defibrillator research and implementations to external defibrillator
`
`implementations.
`
`VII.
`
`THE ‘212 PATENT
`
`
`
`(35) The ‘212 Patent is entitled “External Defibrillator Capable of Delivering Patient
`
`Impedance Compensated Biphasic Waveforms,” and its disclosure relates to an external
`
`defibrillator device having electrodes that are in “electrical communication” with the exterior
`
`of a patient. The defibrillator includes a controller that controls the application of electrical
`
`energy from an energy source to the electrodes by (i) generating a timing signal
`
`
`
`13
`
`
`
`US Patent 6,047,212
`
`corresponding the electrical energy; and (ii) selectively connecting the energy source to the
`
`electrodes in a first and second polarity. (‘212 Patent at 6:51-60).
`
`(36) The ‘212 Patent claims priority to an application filed August 6, 1993. The
`
`stated main difference between earlier electrotherapy apparatus and methods and the ‘212
`
`Patent is the disclosure in the ‘212 Patent of an external defibrillator and defibrillation
`
`method that automatically compensates for patient impedance differences by changing the
`
`nature of the delivered electrotherapeutic pulse. (Id. at 2:37-41, 2:55-59). Automatically
`
`compensating for patient impedance differences maximizes therapeutic efficacy across an
`
`entire population of patients. (Id. at 2:30-34). Methods and apparatus for automatically
`
`compensating for patient impedance differences were well-known by the time of the ‘212
`
`Patent’s priority date.
`
`(37) Claim 1 of the ‘212 Patent includes features relating to a means for selectively
`
`connecting the energy source to the electrodes in a first polarity and a second polarity. The
`
`‘212 Patent discloses the associated structure for this limitation as the circuit including five
`
`switches. (Id. at 6:61-7:53, FIG. 11).
`
`(38) Claim 2 includes features relating to a means for measuring an electrical unit
`
`of the energy source applied across the electrodes, and claim 6 includes features relating to
`
`a threshold means for determining whether the measured electrical unit signal is less than a
`
`predetermined threshold value. The ‘212 Patent discloses the associated structure for
`
`these limitations as a comparator. (Id. at 7:28-31).
`
`
`
`14
`
`
`
`US Patent 6,047,212
`
`(39) Claim 10 includes features relating to a means for limiting voltage applied
`
`across one of the plurality of electronic switches. The ‘212 Patent discloses the associated
`
`structure for this limitation as the fifth switch interposed between the second terminal of the
`
`energy source and the third and fourth switches. (Id. at 7:42-47, FIG. 11).
`
`(40) de Coriolis discloses a defibrillator circuit that delivers defibrillation shocks as
`
`biphasic truncated exponential waveforms to the heart. (de Coriolis at 4:31-33, 5:6-13,
`
`5:14-17, FIGs. 4 and 5). The circuit includes a capacitor 39 and a series of switches 40, 42,
`
`44, and 46 that open and close in response to signals from a timer circuit to deliver
`
`defibrillation shocks to the heart. (Id. at 5:6-13). Current is conducted along lead 33,
`
`through the heart, and along lead 26, which are all connected in series across the capacitor
`
`bank. (Id. at 7:66-8:6, FIG. 5). During the first phase of discharge, the voltage across the
`
`capacitor bank is compared to a reference voltage V2, and the capacitors are allowed to
`
`discharge to the reference voltage V2. (Id. at 5:31-42, 6:32-48, 6:63-64, 8:7-15, 9:48-49).
`
`In the second phase, voltage V2 is applied to the heart until the capacitors have discharged
`
`down to voltage V3, at which point the second phase is truncated. (Id. at 5:42-51, 9:1-3,
`
`9:58-59).
`
`(41)
`
`In my opinion, the series of switches 40, 42, 44, and 46 is equivalent to the
`
`structure associated with the “means for selectively connecting the energy source to the
`
`electrodes in a first polarity and a second polarity,” as recited in claim 1 of the ‘212 Patent.
`
`
`
`15
`
`
`
`US Patent 6,047,212
`
`(42)
`
`In my opinion, the comparator circuit that compares the voltage across the
`
`capacitor bank to the reference voltage is equivalent to the structures associated with the
`
`“means for measuring an electrical unit of the energy source applied across the electrodes,”
`
`as recited in claim 2 of the ‘212 Patent, and the “threshold means for determining whether
`
`the measured electrical unit signal is less than a predetermined threshold value,” as recited
`
`in claim 6 of the ‘212 Patent.
`
`(43) de Coriolis discloses a zener diode 120 that serves to limit the voltage applied
`
`between the source and gate of MOSFET 100 so that it is not damaged. (Id. at 7:43-45). In
`
`my opinion, the zener diode is a standard electrical component for limiting voltage and is
`
`equivalent to the structure associated with the “means for limiting voltage applied across
`
`one of the plurality of electronic switches,” as recited in claim 10.
`
`(44) Hahn discloses a defibrillator for applying a biphasic truncated exponential
`
`waveform from a capacitor through electrodes to a patient’s heart. (Hahn at Abstract). The
`
`defibrillator includes a controller having a timer that generates timed signals to open and
`
`close switches to control the application of electrical energy from the capacitor to the
`
`electrodes 24 and 26. (Id. at 6:22-44). The switches are arranged in an “H-bridge” circuit
`
`configuration for delivering electrical energy in a first polarity and a second polarity. (Id. at
`
`6:22-44). The H-bridge circuit includes two switches 14 and 16 and two thyristors 22 and
`
`30, which are a type of a switch. (Id.). An output sense circuit 28 monitors the voltage
`
`output across the electrodes 24 and 26 during discharge, and when the output voltage
`
`
`
`16
`
`
`
`US Patent 6,047,212
`
`across electrodes 24 and 26 falls to a predetermined level, switch 14 and thyristor 22 are
`
`shut off, truncating the first phase of the waveform. (Id. at 6:58-63). Electric switch 16 and
`
`thyristor 30 is subsequently turned on and then turned off after a second phase duration.
`
`(Id. at 6:22-44).
`
`(45)
`
`In my opinion, the H-bridge circuit is equivalent to the structure associated
`
`with the “means for selectively connecting the energy source to the electrodes in a first
`
`polarity and a second polarity,” as recited in claim 1 of the ‘212 Patent.
`
`(46)
`
`In my opinion, the output sense circuit that monitors the voltage output is
`
`equivalent to the structures associated with the “means for measuring an electrical unit of
`
`the energy source applied across the electrodes,” as recited in claim 2 of the ‘212 Patent,
`
`and the “threshold means for determining whether the measured electrical unit signal is less
`
`than a predetermined threshold value,” as recited in claim 6 of the ‘212 Patent.
`
`(47)
`
`In my opinion, a standard electrical switch that limits voltage is equivalent to
`
`the structure associated with the “means for limiting voltage applied across one of the
`
`plurality of electronic switches,” as recited in claim 10.
`
`(48) As noted above, de Coriolis and Hahn disclose measuring voltage during
`
`discharge. (de Coriolis at 6:32-48, 63-64, 8:7-14; Hahn at 8:7-15). It was well known that
`
`voltage across a capacitor in a resistor-capacitor circuit is associated with current through
`
`the resistive load as given by Ohm’s law. Accordingly, based upon my experience and
`
`
`
`17
`
`
`
`US Patent 6,047,212
`
`knowledge, it is my opinion that a person of ordinary skill in the art would recognize that
`
`measuring voltage is the functional equivalent of measuring current, as recited in claim 5.
`
`(49)
`
`I have also been asked to review the U.S. Patent No. 5,275,157 to Morgan
`
`(Morgan, Appendix Q). Morgan describes a defibrillator including a circuit for producing a
`
`biphasic waveform. (Morgan, Abstract, 5:8-13). The circuit includes a well-known H-bridge
`
`configuration including four electronic switches to control the various phases of the
`
`waveform. (Id. at 5:21-25). The circuit also includes a fifth switch operable to open and
`
`close the circuit between the H-bridge and a capacitor such that no current will flow through
`
`the H-bridge when the switch is open. (Id. at 27-28). FIG. 9 of Morgan also shows this
`
`configuration. Based on my review, I have concluded any of the switches of the H-bridge of
`
`Morgan may close only when the fifth switch is closed and may re-open only when the fifth
`
`switch is open, thereby teaching the corresponding limitation of claim 11 of the ’212 patent.
`
`(50)
`
`In my opinion, a person of ordinary skill in the art would have combined de
`
`Coriolis or Hahn with Morgan because the combination amounts to combining prior art
`
`elements according to known methods to yield predictable results. de Coriolis and Hahn
`
`are directed to a defibrillation method performed by an H-bridge circuit. (de Coriolis at 3:12-
`
`48; Hahn at 5:19-22). Similarly, Morgan teaches a defibrillation method performed by an H-
`
`bridge circuit with a fifth switch added in series with the H-bridge to control the flow of
`
`current. (Morgan at 5:7-6:10). Accordingly, it is my opinion that a person of ordinary skill in
`
`the art would have been motivated to modify the circuit of de Coriolis or Hahn to include the
`
`
`
`18
`
`
`
`US Patent 6,047,212
`
`series switch of Morgan in order to more precisely control the delivered waveform through
`
`the H-bridge circuit. The results of modifying the circuit of de Coriolis or Hahn in this way
`
`would have been predictable at least because Morgan contained such a circuit.
`
`(51)
`
` I have been asked to compare claim 1 of the ‘212 Patent with claim 29 of
`
`U.S. Patent No. 5,607,454 (Cameron; Appendix P; “the 454 Patent”). Both claims relate to
`
`external defibrillators. The only substantive differences between the two claims are that (a)
`
`claim 1 recites a genus that covers the species recited in claim 29, and (b) claim 1 recites
`
`that the controller includes a “timer for generating a timing signal corresponding to the
`
`application of electrical energy from the energy source to the electrodes,” whereas claim 29
`
`recites that the controller includes a “means for delivering a multiphasic waveform without
`
`the use of an inductor.” However, based upon my review of the ‘454 Patent, I have
`
`concluded that the structure corresponding to the “means” recited in claim 29 includes the
`
`timer shown in FIG. 3 of the ‘454 Patent with energy source 32, connecting mechanism 34,
`
`electrodes 36, and resistive load (the patient’s torso) 37. Thus, the controller recited in
`
`claim 29, in my opinion, also includes a “timer for generating a timing signal corresponding
`
`to the application of electrical energy from the energy source to the electrodes.”
`
`(52)
`
`I have been asked to compare claim 8 of the ‘212 Patent with claim 17 of the
`
`‘454 Patent. The only substantive differences between the claims are (a) claim 8 recites
`
`“electronic switches,” whereas claim 17 recites “switches,” and (b) claim 8 recites a
`
`defibrillator whereas claim 17 recites an external “apparatus for administering electrotherapy
`
`
`
`19
`
`
`
`US Patent 6,047,212
`
`to a patient’s heart.” In my opinion, based upon my experience, these differences are not
`
`meaningful.
`
`(53) Based upon my experience and knowledge in this field, electronic switches
`
`were commonly and typically employed as switches in medical devices employing
`
`electrotherapy at the time, as evidenced by de Coriolis, which describes a defibrillator circuit
`
`that includes four electronic switches arranged in an H-bridge configuration, and a fifth
`
`switch connecting the H-bridge circui