throbber
PATENT
`Heartstream Ref. 93-0O3—US1.C2
`
`mt)
`
`"Hewlett.-Packard Ref. 90980060-4
`_
`
`CERTIFICA'I'E.OF'-FACSIIVIILE TRANSMISSION UNDER 37 C.F.R. §1.'B D.
`that this correspondence is‘bdng SfiflV £55, [3'Kfl@-,
`P
`"
`
`.. - n -bar
`La?‘-t Us
`za maq °l‘i
`
`
`
`IN THE UNITED STA 7735 PA TENT AND TRADEMARK OFFICE
`
`
`
`
`
`4._-¢uvmwwm mmmmn\¢:mm~L-.-.-..
`
`In Re Application of:
`Bradford ~E. Gliner, et al.
`
`Group Art Unit
`
`
`
`
`
`
`Serial No.:
`
`08/946,843
`
`Examiner: K. Schaetzle
`
`Filing Date: 8 Oct 1997
`
`Title: ELECTROTHERAPY METHOD
`
`
`
`DECLARATION OF DANIEL]. POWERS
`UNDER 37 C.F.R. §. 1.132
`
`1, Daniel]. Powers, declare as follows:
`
`1.
`
`I am a Project Manager with the Hewlett-Packard Company and am
`
`responsible for developing defibrillators.
`
`2.
`
`I have been employed by the Hewlett-Packard Company since April of
`
`1998 (when Hewlett—Packard acquired Heartstream, Inc.). I began working at
`
`Heartstream in January of 1993.
`
`3.,‘ ,»
`Ann Arbor.
`
`Iphave a B.S. in Electrical Engineering from the University of Michigan at
`
`4.
`
`I have over 10 years of experience in the field of medical products,
`
`specializing in defibrillator design and development. I have a high degree of familiarity
`
`with defibrillator circuits.
`
`5.
`
`I am one of the inventors for the above-referenced patent application.
`
`6.
`Irhave reviewed the Office action dated 20: [an .1999, which was issued by '
`the U.S. Patent and Trademark Office in connection with the prosecution of this
`
`application. In reviewing the Office action, I also reviewed U.S. Patent 5,634,938 to
`
`
`
`Page ll of 4
`
`_ 1
`
`Philips Exhibit 2024
`_p pZoll_’Lifeco:r V.. bhilips
`.
`.
`._.>..~.»::;t‘.‘;'$o‘.=.m=‘i‘.5:*>.*.i;:
`IPR2013-0.0607
`
`Page 1 of 4
`
`Philips Exhibit 2024
`Zoll Lifecor v. Philips
`IPR2013-00607
`
`

`
`Swanson and European ‘Patent EP 0491649 to Pless which were cited by the Examiner in
`
`support of his argument.
`
`7.
`
`In the Office action dated 20 [an 1999, the Examiner contends that
`
`Swanson teaches an external biphasic defibrillator. The Examiner directs Applicants’
`attention to col. 6, lns. 47-65 of Swanson in support of his argument which provides:
`
`The system 10 . .. represents an implantable or external defibrillator
`system. In an implantable system, electrodes 18 and 20 may be
`implanted on or about the heart, one may be subcutaneous, and
`each may comprise one or more electrode, though only one is
`shown for each. In the case of an external defibrillator, the circuitry
`may be disposed external to the body of a patient but connect with
`implantable electrodes. Alternatively, the external unit may
`Connect
`at least two cutaneous electrodes placed on the body.
`
`Swanson, US. Patent 5,634,938, col. 6, lines 55-65 (emphasis added).
`
`8.
`
`As an electrical engineer in the engineering arts, I would be unable to
`
`build a high voltage, high energy biphasic external defibrillator based on these
`
`statements in combination with the remainder of the reference, which specifically
`
`teaches. how to build an implantable biphasic defibrillator.
`
`9.
`
`Even if higher voltage switches were used in the Swanson et a]. circuit, the
`
`circuit could not be used in an external defibrillator. The Swanson circuit depends on
`
`the load being stable and well controlled; this results because the electrodes are .
`
`attached to the heart internally. As a result, there is no means described in Swanson for
`
`protecting against a load fault. Since load faults may occur in the external application
`
`of defibrillator electrodes, any circuit that is used must provide a means for protecting
`
`the circuit against the load faults. Developing a circuit design that protects the circuit
`
`from load faults is a major design challenge.
`
`10.
`
`- Swanson contrasts its teaching against conventional devices, which use
`
`125 p.F capacitors charged to 780 volts, by teaching that:
`
`The lower voltage needed to charge the capacitor C3 simplifies the
`design of the pulse generator circuitry. This allows switches to be
`used with lower voltage and current ratings. Switches with lower
`2
`Heartstream Ref. 93-003—US1.C2
`HP Ref. 909800604
`
`Serial No. 08/946,843
`
`-‘..‘..:1<.;.-ac-2-.‘.»‘..‘;:.
`
`-.-
`
`Page
`
`of 4
`
`Page 2 of 4
`
`

`
`ratings are more readily available, and reliability is improved if
`peakvoltages and currents are reduced.
`
`Swanson, U.S. Patent 5,634,938, col. 7, lines 58-64 {emphasis added).
`
`Thus Swanson recognizes the advantage of using switches with lower voltages and
`
`current ratings. However, nowhere does Swanson address the problems associated
`with using switches with higher voltage and current ratings which would necessarily be
`required for a highvoltage, high energy external defibrillator.
`
`4 11.
`
`In order to build a high voltage, high energy biphasic defibrillator which
`
`is used externally, the. circuit must be designed so that it can withstand voltages in
`excesslof 1500 volts. This design must take into consideration the fact that readily
`available switches have an upper limit voltage rating of 1200-1400 volts. The circuit of
`
`Swanson would beunable to tolerate voltages in excess of 1500 volts.
`
`12.
`
`Because of the higher voltages present in the external defibrillator, a major
`
`design challenge becomes terminating phase one. This problem is further exacerbated
`
`by shorter phase.durations. Swanson teaches waveforms >15msec total duration (col.'6,
`
`line 43). Therefore the teaching of Swanson could not be easily adapted to be used in an
`
`external defibrillator capable of delivering a waveform shorter than 15msec.
`
`13.
`
`Pless measures the capacitor voltage until it reaches a threshold value.
`
`Once the threshold value is set, phase one is ended. Thus, Pless does not use a timer to
`
`control the duration of phase one.
`
`14.
`
`I would not be able to design and build a high voltage, high energy
`
`biphasic external defibrillator based on the teachings of Swanson, or Swanson in
`
`combination with Pless without overcoming the design challenges described above.
`
`I hereby declare that all statements made herein of my own knowledge are true
`
`and that all statements made on information and belief are believed to be true; and
`
`further that these statements were made with the knowledge that willful false
`
`statements and the like so made are punishable by fine or imprisonment, or both, under
`
`section 1001 of Title 18 of the United States Code, and that such willful false statements
`
`>1)
`
`4
`
`Page 3 of 4
`
`3
`
`'
`
`Heartstream Ref. 93-003—US1.C2
`HP Ref. 90980060-4
`
`Serial No. 08/ 946,843
`

`
`..
`
`.
`
`.
`
`_...:$..u¢:.>l.2a.s4:i:ls7§I{*.:i.:é:.n»-Gi-- ;.:.
`
`Page 3 of 4
`
`

`
`may jeopardize the valifiity of the application, any patent issuing thereon, or any patent
`to whicn this verified statefnent is directed.
`
`Dated:
`
`2
`
`
`
`Hewlett-Packard/Heartsh-eam
`2401 Fourth Avenue, Suite 500
`= Seattle Washington 98121
`Telephone: 206.664.7630
`Facsimile: 206.664.9694
`
`4
`
`Heartstream Ref. 93-003-US1.C2
`HP Ref. 909800604
`
`Serial No. 08/ 946,843
`
`..
`
`.-.-.».-:«-x~,~.'-.<.;.<:'.;‘a.‘.~.;,~.~;a.-L3.-=...L2....a-..
`
`'._- ....~;_';f;:....::e-.:.:.-.i.-
`
`_ _~'
`
`Page Zof 4
`
`Page 4 of 4

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket