TRANSTHORACIC DEFIBRILLATION OF 100 KG CALVES
`WITH BIDIRECTIONAL TRUNCATED EXPONENTIAL SHOCKS
`
`J. C. Schuder, W. C. McDaniel, and H. Stoeckle
`
`Jones and
`High-intensity electric field stimulation causes periods of arrest in cultured myocardial cells.
`Jones have shown that the duration of arrest caused by stimulation with a unidirectional rectangular pulse may be
`reduced by the addition of a low-amplitude reverse polarity pulse to the waveform1 and that the ratio of field in—
`tensity required to produce a given duration of arrest to that needed for cellular excitation tends to be larger for
`damped sinusoidal waveforms having a reverse current component than for unidirectional waveformsz. These
`findings suggest that the inclusion of a reverse current component in a defibrillatory waveform might reduce post—
`shock arrhythmias without having an adverse effect upon the level of successful defibrillation. On the contrary,
`Niebauer and colleagues have compared the efficacy and safety of unidirectional and bidirectional low-droop
`trapezoidal (nearly rectangular) waveforms in defibrillating isolated perfused canine hearts and concluded that
`symmetrical bidirectional waveforms offer no significant efficacy or safety advantage over unidirectional pulsesa.
`A comparison of the results of studies by our group of the effectiveness of one-cycle, symmetrical, bi—
`directional, rectangular wave shocks4, one—cycle, asymmetrical, bidirectional, rectangular wave shocks5, and
`unidirectional rectangular wave shocks6 in the transthoracic defibrillation of 100 kg calves indicates that the most
`successful of the symmetrical and asymmetrical bidirectional waveforms are more successful and require less
`energy than the most successful unidirectional waveform.
`While rectangular wave shocks are very convenient for defibrillation research because the various parame—
`ters can be so easily and unambiguously specified, they are technologically difficult to generate in clinical sized
`defibrillators.
`In the present paper, we describe a study of the effectiveness of a category of symmetrical and
`asymmetrical truncated exponential waveforms, which can be implemented in clinical sized apparatus,
`in the
`transthoracic defibrillation of 90—110 kg calves.
`
`METHOIE
`
`Waveform Selection. A generalized representation of the category of waveforms considered in this paper
`is shown in Figure 1 . The leading and lagging portions of the waveform have identical time constants, r , and
`identical durations, T. The amplitude of the initial current of the leading half-cycle of the waveform, I, is 70 amp
`in all cases. We define the fractional undershoot associated with this category of waveforms as
`
`Ilh/l
`=
`ll3/|]|
`=
`f
`where the vertical lines denote absolute value.
`The energy in joules delivered by a waveform of current of the type shown in Figure 1 is given by
`2
`2
`2
`2
`- IZ)+(|3 '
`|~)]
`
`Equationl
`
`Equation2
`
`2 |
`
`U
`
`= 0.5RT [(I1
`
`which, together with Equation 1, yields
`
`u
`
`= 0.5 R 1 (I12 -122) (1+ f2)
`
`where R is the chest resistance in ohms, T is in seconds, and the 1's are in amperes.
`Solving Equation 3 for 1 gives
`
`Furthermore,
`
`1
`
`= ng—T
`R(Il -|2) (1+f)
`
`T
`
`=
`
`1 1n (ll/'2)
`
`Equation3
`
`Equation4
`
`Equation5
`
`From Equations 4 and 5, one can determine the time constant and duration required to deliver a specified energy
`level, U. to a chest having a given resistance, R, with a waveform having specified values of 11, 12, and f.
`
`From the Departments of Surgery and Child Health, University of Missouri, Columbia, Missouri.
`Supported by grant HL-l80h0 from the National Heart, Lung, and Blood Institute and a grant
`from Physio
`Control Corporation.
`
`Vol.XXX Trans Am Soc Artif Intern Organs WM
`
`520
`
`1
`
`L|FECOR427-1010
`
`1
`
`LIFECOR427-1010
`
`

`

`Schuder et a1. Transthoracic defibrillation
`
`The 11, 12, f. and U values used in the 8 waveforms evaluated in the present study were selected so that
`the results for each specific waveform could be compared easily with results obtained previously with bidirec—
`tional rectangular waveforms“,5 having amplitude. fractional undershoot, and delivered energy values of I1, f.
`U and 12, f, U-
`the rectangular waveform shown in Figure 2b has a leading half-cycle amplitude equal to
`As an example,
`the initial amplitude of the leading half—cycle of the exponential waveform sketched in Figure 2a:
`the waveform
`shown in Figure 20 has a leading half-cycle amplitude equal to the final amplitude of the leading half—cycle of the
`waveform shown in Figure 2a. The 3 waveforms in Figure 2 have identical undershoots and deliver the same
`energy.
`
`In this table, the dura-
`The specifications for the 8 waveforms which were evaluated are shown in Table I.
`tions mbulated in the fourth and eighth columns are calculated from Equations 4 and 5.
`In some cases, the levels
`of current listed in columns 6 and 7 are experimental target values which have been rounded to the nearest integer
`from their theoretical noninteger values.
`
`TABLE 1.
`
`DESCRIPTION OF BIDIRECTIONAL EXPONENTIAL WAVEFORMS
`
`Leading Half-Cycle
`Current
`Duration
`
`Fractional
`Undershoot
`
`Lagging Half-Cycle
`Current
`Duration
`
`Waveform
`No.
`
`1
`2
`3
`1+
`5
`6
`7
`8
`
`amp
`
`I
`
`1
`
`70
`70
`70
`70
`70
`70
`70
`70
`
`|
`
`2
`
`35
`35
`50
`50
`35
`35
`50
`50
`
`T
`msec
`
`3.77
`3.77
`2.80
`2.80
`7.511
`7.511
`5.61
`5.61
`
`f
`
`1/2
`l
`1/2
`l
`l/h
`1/2
`1/11
`1/2
`
`—|
`
`3
`
`-|h
`amp
`
`35
`70
`35
`70
`18
`35
`18
`35
`
`18
`35
`25
`50
`9
`18
`13
`25
`
`T
`msec
`
`3.77
`3.77
`2.80
`2.80
`7.511
`7.5%
`5.61
`5.61
`
`Delivered
`Energy
`
`.
`
`U
`oules
`
`250
`1100
`250
`1100
`MS
`500
`l125
`500
`
`Delivered energy values are based upon an assumed representative chest resistance of 20 ohms.
`
`Equipment. A bidirectional waveform research defibrillator, which has been described elsewhere7, was
`used to supply a low current shock for inducing fibrillation, the bidirectional exponential waveform being evaluated
`(or a screening bidirectional rectangular wave shock), and an effective follow—up shock to salvage the animal in
`case the waveform being evaluated did not yield defibrillation. Current and voltage waveforms of the shock being
`evaluated were displayed on a Tektronix model 5113 dual-beam storage oscilloscope. The chest resistance of the
`bidirectional exponential waveforms was calculated as the arithmetic mean of the voltage to current ratios evalu—
`ated near the midpoints of the 2 half-cycles. Lead 1] electrocardiograms were displayed on an oscilloscope and
`recorded with an electrocardiograph for later evaluation.
`Procedure. Before being entered into our detailed study, calves were screened by requiring that success-
`ful first—shock defibrillation be achieved in at least 19 out of 20 defibrillation attempts with a bidirectional rec—
`tangular wave shock having a leading half -cycle amplitude of 50 amperes, a leading half-cycle duration of 8 msec,
`a fractional undershoot of 1/2, and a nominal delivered energy of 500 joules. Fourteen calves passed the screening
`requirement and were used in our detailed study: one calf failed the screening test.
`The procedure used in our detailed study was almost identical to that previously describede. Briefly, the
`calves were anesthetized with 110 mg/kg glycerly guaiacolate and 4. 4 mg/kg thiopental sodium injected intrave—
`nously. intubated, and maintained with methoxyflurane in 50% N20 and 50% 02. Fibrillation was induced with a
`low current transthoracic shock and, after 30 secs,
`the shock being evaluated was applied.
`If defibrillation was
`achieved on the first trial, it was recorded as a success and the electrocardiogram recorded for 2-1/2 mins.
`Otherwise, the shock was recorded as a failure and a shock of known high effectiveness used to defibrillate the
`animal. The procedure was repeated with not less than 5 mins between the start of successive episodes. Each
`specific waveform was evaluated on the basis of 120 episodes in 90—110 kg calves with no animal being involved in
`more than 20 episodes in a given session or more than 20 episodes with a single waveform. Our detailed study
`encompassed a total of 120 x 8 = 960 fibrillation-defibrillation episodes.
`
`RESULTS
`
`The results are summarized in Tables 11 and III. The experimental chest resistance values tabulated in
`Table II indicate a limited variation in chest resistance for the shocks evaluated and serve to justify the use of 20
`ohms as a representative value of chest resistance.
`
`521
`
`2
`
`

`

`Schuder et a1. Transthoracic defibrillation
`
`TABLE II.
`
`BODY WEIGHT AND CHEST RESISTANCE 0F CALVES
`
`Waveform
`No.
`
`Body Weight
`kg
`
`Chest Resistance
`ohms
`
`1
`2
`3
`1;
`S
`6
`7
`8
`
`103:6
`10215
`102 t 3
`102111
`100:3
`10112
`101 ill
`102 i 11
`
`20.3i1.3
`20.31- 1.3
`19.9 1 2.0
`19.71: 1.7
`153.512.]
`19.3i2.2
`19312.1
`18.6 i 1.1+
`
`Entries for body weight and chest resistance are
`means 1 SD and based upon 120 episodes.
`
`TABLE III.
`
`RESULTS OF BIDIRECTIONAL WAVEFORM SHOCKS
`
`Waveform
`No.
`
`Successful
`Defibrillations
`No.
`14
`
`Time Required for Appearance
`of First Ventricular Complex
`secs
`
`Time Required To Return
`To Normal Sinus Rhythm
`secs
`
`1
`2
`3
`11
`5
`6
`7
`8
`
`108
`113
`109
`117
`101-1
`115
`119
`120
`
`90
`9‘1
`91
`98
`87
`96
`99
`100
`
`3.0 i 1.0
`6.1 i 3.8
`3.1 $1.4
`5.2 1 3.9
`8.3 i 8.3
`6.1i 5.0
`6.2 i 11.7
`9.7 i 9.6
`
`6.11
`9.11 :1:
`26.2 i 15.5
`12.7 i
`9.11
`28.11 117.14
`17.7 i 10.7
`16.7 1111.1
`9.8 1 SJ;
`21.8 i 17.0
`
`Entries for the appearance of the first ventricular complex and for the return of normal
`sinus rhythm are means 1: SD and based upon the indicated number of successful defibrillations.
`
`In Figure 3 is plotted percent success versus delivered energy for the 4 exponential waveforms in which
`the initial half—cycle current sweeps from 70 to 35 amperes.
`In Figure 4 are plotted the corresponding data for the
`4 waveforms in which the initial half—cycle current sweeps from 70 to 50 amperes. Two short line segments are
`associated with each of the 8 datum points. These segments represent the percent success observed previously
`for bidirectional rectangular wave shocks4’ 5 having the fractional undershoot of the exponential waveform and
`leading half-cycle current amplitudes equal to the initial and final values respectively of the leading half—cycle of
`the exponential waveform.
`In calves, there is often an appreciable period of standstill or of only p—waves in the electrocardiogram
`following a defibrillatory shock. Data from the fourth column of Table III, along with corresponding data from
`earlier studies with bidirectional rectangular waveforms4v 5 are plotted in Figures 5 and 6. Similar graphs relat-
`ing the time required for a return of normal sinus rhythm in the electrocardiogram could be constructed from
`data in the final column of Table III and our earlier studies4'5.
`
`DISCUSSION
`
`The predominant tendency for each of the percent success levels of the bidirectional exponential wave shocks
`under evaluation to fall within, on. or very close to the boundaries defined by the 2 associated bidirectional rec—
`tangular wave shocks, as illustrated in Figures 3 and 4, supports the proposition that such shocks are well behaved
`and that their performance is approximately predictable from that of the associated bidirectional rectangular wave
`shocks. This proposition is further strengthened by the data concerning the shock induced electrocardiographic
`disturbances as shown graphically in Figures 5 and 6 and by a comparison of the data in column 5 of Table III with
`corresponding rectangular wave data developed earlier4’ 5. The most successful of the bidirectional exponential
`waveforms evaluated (100%) was equivalent to the most successful bidirectional rectangular waveform5 (100%) and
`superior to the most successful unidirectional rectangular6 (93%) and unidirectional truncated exponential8 (94%)
`waveforms in achieving ventricular defibrillation in our 100 kg calves.
`Although there are inherent uncertainties in trying to extrapolate expierimental experience from the 100 kg
`calf to the human, we interpret the results of this and our earlier studies47 ’8 as warranting clinical trials of hi—
`directional truncated exponential wave defibrillators.
`
`522
`
`3
`
`

`

`Schuder et a1. Transthoracic defibrillation
`
`REFERENCES
`
`Jones JL, Jones RE. Decreased arrhythmias in cultured myocardial cells following high intensity
`electric field stimulation with biphasic rectangular waveforms.
`Fed Proc 42:lllh, 1983.
`Jones JL, Jones RE.
`Improved defibrillator waveform safety factor with biphasic waveforms.
`Physiol 2h5:H60, 1983.
`Niebauer MJ, Babbs CF, Geddes LA, Bourland JD. Efficacy and safety of the reciprocal pulse de-
`fibrillator current waveform. Med Biol Eng Comput 22:28, 198h.
`Schuder JC, Gold JH, Stoeckle H, McDaniel WC, Cheung KN. Transthoracic ventricular defibrillation
`in the loo kg calf with symmetrical one-cycle bidirectional
`rectangular wave stimuli.
`IEEE Trans
`Biomed Eng BME-30:4]S, 1983.
`Schuder JC, McDaniel WC, Stoeckle H. Defibrillation of lOO kg calves with asymmetrical, bidirec-
`tional,
`rectangular pulses. Cardiovasc Res
`(in press).
`JM. Transthoracic ventricular
`Gold JH, Schuder JC, Stoeckle H, Granberg TA, Hamdani SZ, Rychlewski
`defibrillation in the loo kg calf with unidirectional
`rectangular pulses. Circulation 56:7“5,
`1977.
`Schuder JC, Gold JH, McDaniel NC. Ultrahigh-energy hydrogen thyratron/SCR bidirectional waveform
`defibrillator. Med Biol Eng Comput 20:h19, 1982.
`Schuder JC, Gold JH, Stoeckle H, Granberg TA, Dettmer JC, Larwill MH. Transthoracic ventricular
`defibrillation in the loo kg calf with untruncated and truncated exponential stimuli.
`IEEE Trans
`Biomed Eng ENE-27:37,
`l980.
`
`Am J
`
`70A
`
`b
`ZII.I
`KK
`3
`0
`
`was
`
`1
`
`._
`
`4
`
`'12
`
`Figure l. Generalized representation of category of
`bidirectional
`truncated exponential waveforms studied.
`
`TIME
`
`TIME
`
`III/2
`
`,_
`2
`II]
`x
`(a) K
`3
`0
`
`35A
`
`T'3.77ml
`
`
`
`I;m
`
`xm
`
`0
`3
`
`Pz
`
`2
`x
`
`30
`
`
`
`
`(b)
`
`(c)
`
`TIME
`-I15
`III)
`
`
`40
`
`IOO
`
`90
`
`4
`a 80
`3
`3
`g 70
`m
`5
`3 ‘0
`u
`1
`a so
`
`III/2
`I
`
`in mm
`l
`I
`
`mm
`1
`70A
`4.6
`
`OEXPONENTIAL
`‘-RECTANGULAR
`
`Figure 2. Waveforms which have the same
`value of undershoot and which deliver the
`same energy to a given load.
`(a) First
`bidirectional
`truncated exponential wave-
`form evaluated.
`(b) Bidirectional
`rec-
`tangular waveform having current amplitude
`of leading half-cycle equal
`to the initial
`current value of leading half-cycle of ex-
`ponential waveform.
`(c) Bidirectional
`u
`.
`rectangular waveform having current ampli-
`tude of leading half-cycle equal
`to flnal
`value of current
`in leading half-cycle of
`exponential waveform.
`
`35A
`
`c 250
`450
`400
`350
`300
`NOMINAL DELIVERED ENERGY IN JOULES
`
`500
`
`Figure 3. Percent success of ventricular defibrilla-
`tion related to fractional undershoot and delivered
`energy for bidirectional
`truncated exponential wave-
`forms in which leading half-cycle current sweeps from
`70 to 35 amp and for bidirectional
`rectangular wave-
`forms in which the leading half-cycle is either 70 or
`35 amps.
`
`523
`
`4
`
`

`

`Schuder et al. Transthoracic defibrillation
`
`I00
`90
`
`.l
`3 soa)In
`'3
`g 70
`m
`
`'2' soUU
`5n. 50
`
`40
`
`t;I/2
`L/70A
`\50A
`
`Iii/4
`III
`l
`5OA—‘* 3/50A
`J
`“”1 ”~{13‘
`10A
`
`III/2
`t
`50A
`.{IgA
`
`o EXPONENTIAL
`— RECTANGULAR
`
`
`
`‘50
`400
`350
`300
`250
`NOMINAL DELIVERED ENERGY IN JOULES
`
`500
`
`K:
`
`ma 3
`
`’- 0
`
`:5
`0x a
`iufl
`g;
`:8 6
`Eu:
`:5 4
`1::
`8:-
`“I- 22
`In»
`;:>
`
`0
`
`Figure 5. Mean time required for appear-
`ance of first ventricular complex follow-
`ing a defibrillatory shock related to
`fractional undershoot and delivered
`energy for bidirectional
`truncated expo-
`nential waveforms in which leading half-
`cycle current sweeps from 70 to 35 amps
`and For bidirectional
`rectangular wave-
`forms in which the leading half-cycle is
`either 70 or 35 amps.
`
`III/2
`I
`
`fll
`l
`
`III/4
`i
`
`{II/2
`I
`
`Figure A. Percent success of ventricular
`defibrillation related to fractional under-
`shoot and delivered energy for bidirec-
`tional
`truncated exponential waveforms in
`which leading half-cycle current sweeps
`from 70 to 50 amps and for bidirectional
`rectangular waveforms
`in which the leading
`half-cycle is either 70 or 50 amps.
`
`f-I/Z
`l
`
`le III/4
`
`III/2
`
`O EXPONENTIAL
`— RECTANGULAR
`
`{70A
`35A
`:16
`
`
`
`250
`300
`350
`400
`4 50
`NOMINAL DELIVERED ENERGY IN JOULES
`
`500
`
`8 l2
`I—g
`g0
`E3 IO
`'65
`w)< a
`33
`(0.
`oz! 6
`Q8
`:5
`g 4
`“an"-E
`“Jr
`a: 2
`r->
`l
`
`O
`
`
`
`O EXPONENTIAL
`- RECTANGULAR
`
`50A
`
`70A
`
`Figure 6. Mean time required For appear-
`ance of first ventricular complex follow-
`ing a defibrillatory shock related to
`Fractional undershoot and delivered
`energy for bidirectional
`truncated expo-
`nential waveforms in which leading half-
`cycle current sweeps from 70 to 50 amps
`and for bidirectional
`rectangular wave—
`forms
`in which the leading half-cycle is
`either 70 or 50 amps.
`
`
`
`
`250
`300
`350
`400
`450
`500
`NOMINAL DELIVERED ENERGY IN JOULES
`
`524
`
`5
`
`

`

`DISCUSSION OF MANUSCRIPT #103
`
`DR. DeVRIES: What was the efficacy of the defibrillator after the time when you Fibrillated?
`How quickly after you actually got fibrillation did you actually defibrillate?
`DR. SCHUDER:
`For both the rectangular waves and the exponential waves,
`attempted at 30 secs (1 l sec) after the induction of defibrillation.
`
`the defibrillation was
`
`'1‘
`
`525
`
`6
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket