throbber
3,782,389
`[111
`[191
`United States Patent
`
`
` Bell ' [45] Jan. 1, 1974
`
`
`
`
`
`[54] COMPUTER CONTROLLED
`DEFIBRILLATOR
`‘
`.
`Inventor. Davtd Bell, Omaha, Nebr.
`[75]
`[73] Assignee: Health Technology Labs, Inc.,
`Omaha, Nebr.
`
`.
`.
`Jan. 20’ 1972
`[22] Flled'
`[21] Appl. No.: 219,455
`
`.
`
`"""""""
`
`[52] U S C!
`128/419 D 128/421 128/423
`'
`'
`’
`’ 324/1 11’
`Int Cl
`[51]
`A6ln 1/36
`.
`[58] Field of Search
`......... 128/419 D, 419 P,
`128/419 R, 420, 421, 422, 423; 324/111
`
`
`
`[56]
`
`-
`
`References Cited
`
`UNITED STATES PATENTS
`9/197l
`Jaros et a1 ....................... 128/419 D
`3.605.754
`
`12“957 Albert -----
`324/1“
`23171817
`6/1966 sz """"""""""""""""" 1281419 D
`3'258'013
`12/1959
`Brown ................................. 324/111
`2,919,408
`FOREIGN PATENTS OR APPLICATIONS
`l,076,286
`2/1960 Germany ________________________ 123/419 D
`
`Primary Examiner—William E. Kamm
`Attorney—Zarley, McKee & Thomte
`
`.
`
`ABSTRACT
`[57]
`A computer controlled defibrillator comprising a set
`
`of electrodes which are engageable with a patient and
`which are connected to a source of electrical energy
`by a circuit means. The circuit means comprises stor-
`age capacitors, energy selector, computer, manual and
`reset switches. voltage monitor, current monitor, and
`output meter. The computer responds to certain exter-
`nal
`inputs, automatic and manual, and controls the
`output delivered to the patient. The energy selector
`permits the selection of the energy which is desired to
`be delivered to the patient. The sequence is started by
`
`closing the manual reset switch which zeroes the out-
`put meter and activates the power supply (electric en—
`ergy) at a voltage which is dependent on the energy
`selector. The energy drlved- from the power supply IS
`stored in the storage capacrtors. The energy selector,
`which is manually set to the energy desired, also feeds
`an input to the computer. When the manual switch is
`activated,
`the computer causes the stored energy
`source to be connected to the patient through the
`electrodes. The current monitor and voltage monitor
`feed instantaneous signals to the computer which
`computes the energy as a continuous integration pro-
`cess- When the computer energy equals the selected
`energy, the computer causes the energy source to be
`disconnected from the patient. The total energy delivc
`ered to the patient is indicated as a steady reading on
`the UMP“t meter.
`'
`
`4 Claims, 6 Drawing Figures
`
`8
`
`22
`
`2‘1
`
`y
`
`/¢
`
`.
`
`..
`
`QQ
`
`, Q
`
`\l
`
`1
`
`L|FECOR427-1005
`
`1
`
`LIFECOR427-1005
`
`

`

`PATENTEDJAH 11974
`
`'-
`
`,
`SHEEIlflFB
`
`_ 3.782.389
`'
`
`.é'nery
`coin/u er-
`3" 001729-0/
`
`fimpfirfli‘b
`
`[6/ eye
`
`2
`
`

`

`PATENTEU JAN 1 ‘974
`
`3.782.389
`
`sum 2 at 3
`
`
`
`I'i\lԤ\\|
`
`uoSQ\\
`
`3
`
`

`

`N\\v§§\
`
`Rxkqmgvxxg
`
`PATENTEDJAN 1 '974 .-
`
`3782.389 '
`
`SHEEI 3 [IF 3
`
`.\11{}-m\mnmw/,$M:W“V_.memxfiN
`
`vi.Mm:.,
`
`®aw
`
`‘A
`
`{NM
`
`mwfii34mm.
`
`VxQ\
`
`V‘QQ\
`
`0INNRAW
`
`WM
`
`4
`
`
`

`

`COMPUTER CONTROLLED DEFIBRILLATOR
`
`1 .
`
`3,782,389
`
`2
`FIG. 5 is a schematic view of more of the circuitry of
`the invention; and
`FIG. 6 is a schematic view of more of the electrical
`circuitry of the invention.
`The defibrillator of this invention is referred to gen-
`erally by the reference numeral 10' and comprises a
`portable housing 12 having a pair of electrodes or pad—
`dles 14 and 16 connected to the circuitry therein as will
`be described in more detail hereinafter. The electrodes
`or paddles l4 and 16 are engageable with the patient
`to deliver a predetermined energy output to the patient
`to momentarily stop the patient’s heart so that fibrilla-
`tion of the heart is stopped.
`The circuitry of the defibrillator is depicted in sche—
`matic form in FIG. 2 wherein the numeral 18 refers to
`a ”0 VAC power supply having a switch 20 associated
`therewith. The power supply 18 is electrically con-
`nected to the storage capacitors 22 which are adapted
`to store energy derived from the power supply 18. A
`“switch” mechanism 24 is connected to the storage ca-
`pacitors 22. Mechanism 24 is connected to the elec-
`trodes 14 and 16 as seen in FIG. 2 and to a voltage
`monitor means 26 and current monitor means 28. Man—
`ual switch 30 and reset switch 32 are connected to the
`energy computer and control means 34. Energy selec-
`tor 36 is also connected to the computer and control
`means 34 as is the output meter 38. Energy selector 36
`may be comprised of a conventional rotatable dial or
`the like for setting the energy to be delivered to the pa—
`tient.
`
`10
`
`15
`
`20
`
`25
`
`30
`
`The use of DC defibrillators in emergency resuscita—
`tion has become well established. Limitations due to
`weight have prevented more widespread use of the de-
`fibrillators. Most clinical defibrillators depend on the
`storage and discharge of energy through a stable RLC
`combination, thus requiring accurate capacitance, in-
`ductance and resistance. The conventional defibrilla-
`tors employ a pair of electrodes or paddles which are
`placed in contact with the patient‘s chest. A defibrilla-
`tion or electrical pulse is then applied to the patient,
`through the electrodes, to momentarily stop the heart
`so that fibrillation of the heart is stopped. Since time is
`critical in defibrillation techniques, it is extremely im-
`portant that a sufficiently large impulse be applied to
`the patient during the first attempt. A majority of the
`prior art devices employ some means for selecting the
`energy to be delivered to the patient. However, it has
`been found that
`these devices generally deliver a
`smaller or lower output to the patient than that which
`was selected. A further complication is that the resis-
`tance of the patients vary greatly. Thus, the operator
`could possibly determine that it was necessary to apply
`an impulse of 200 joules to the patient. Quite often, the
`variances in the defibrillator and the variable resistance
`of the patient will result in considerably less than 200
`joules being applied to the patient. If the pulse is insuf-
`ficient to momentarily stop the patient’s heart, the pa-
`tient could possibly die.
`Therefore, it is a principal object of this invention to
`provide an improved defibrillator.
`A further object of this invention is to provide a defi-
`brillator wherein the energy delivered to the patient
`substantially equals the selected energy.
`A further object of this invention is to provide a defi-
`brillator including a circuit means having an energy
`computer and control which computes the energy de-
`livered to the patient and causes the energy source to
`be disconnected from the patient when the computed
`energy substantially equals the selected energy.
`A further object of this invention is to provide a defi-
`brillator which delivers the selected energy to the pa-
`tient regardless of the resistance of the patient.
`A further object of this invention is to provide a defi-
`brillator which is light weight and portable.
`A further object of this invention is to provide a defi-
`brillator which is economical of manufacture, durable
`in use and refined in appearance.
`These and other objects will be apparent to those
`skilled in the art.
`This invention consists in the construction, arrange-
`ments, and combination of the various parts of the de-
`vice, whereby the objects contemplated are attained as
`hereinafter more fully set forth, specifically pointed out
`in the claims, and illustrated in the accompanying
`drawings, in which:
`FIG. 1 is a perspective View of the defibrillator of this
`invention.
`FIG. 2 is a block diagram of the electrical circuitry
`of the defibrillator.
`FIG. 3 is a block diagram illustrating the components
`of the energy computer and control and its relationship
`with other components of the device.
`FIG. 4 is a schematic view of a portion of the cir—
`cuitry of the invention.
`
`35
`
`40
`
`4s
`
`50
`
`55
`
`60
`
`65
`
`The energy computer and control means 34 is illus-
`trated in schematic form in FIG. 3. In FIG. 3, it can be
`seen that the current monitor 28 and voltage monitor
`26 are electrically connected to the Multiplier 40 and
`that the Multiplier 40 is connected to an Integrator 42.
`Integrator 42 is connected to an Analog Memory 44
`which is connected to the meter 38. The current moni-
`tor 28 and the voltage monitor 26 are also connected
`to a Time Out Comparator which is connected to the
`OR gate 48. The energy selector 36 is connected to the
`Time Out Comparator 46, Integral Comparator 50 and
`Voltage Comparator 52. The Integral Comparator 50
`is connected to the OR gate 48 and to the Integrator 42
`as depicted in FIG. 3. Voltage Comparator 52 is con-
`nected to the Voltage Reference 54 and to the Charge
`Logic 56. The Multiplier 40 is also connected to the
`Voltage Comparator 52.
`'
`The reset switch 32 is electrically connected to the
`Analog Memory 44 and to the Charge Logic 56 while
`the manual switch 30 is connected to the Delay-Start
`58 and to the Charge Logic 56.
`The heart of the control mechanism in the defibrilla-
`tor is the energy computer and control 34 which re
`sponds to certain external inputs, manual and auto-
`matic, and controls the output delivered to the patient.
`In operation, the manual reset 32 starts the sequence
`by zeroing the output meter 33 and activating the
`power supply 18 at a voltage which is dependent on the
`energy selector 36. Thus, if it were desired to deliver an
`impulse of 200 joules to the patient, the energy selector
`36 would be set at 200 joules. The energy derived from
`the power supply 18 is stored in the storage capacitors
`22. The energy selector 36, which is manually set to the
`energy desired, also feeds an input to the energy com-
`puter and control 34. The electrodes or paddles 14 and
`16 are then placed into contact with the patient and the
`
`5
`
`

`

`3,782,389
`
`4
`
`10
`
`15
`
`20
`
`25
`
`30
`
`35
`
`45
`
`3
`manual switch 30, located on either or both of the pad-
`dles l4 and 16, is activated.
`When the manual switch 30 is activated, the energy
`computer and control 34 causes the stored energy
`source to be connected to the patient.‘ The current
`monitor 28 and voltage monitor 26 feed instantaneous
`signals to the energy computer and control 34 which
`computes the energy as a continuous integration pro-
`cess. When the computed energy equals the selected
`energy, the energy computer and control 34 causes the
`energy source to be disconnected from the patient. The
`total energy delivered to the patient is indicated as a
`steady reading on the output meter 38.
`More specifically, the circuitry of FIGS. 4, 5 and 6
`operates as follows. The circuit of FIG. 4 is basically
`the power supply for the device. TP3 transformer feeds
`a full wave bridge rectifier to generate plus and minus
`DC voltage. The transistor and zener diodes regulate
`the DC to :15 v. and are of conventional design. The
`second set of diodes leading to the coils of K1 and K2
`supply power to operate these relays. Contacts K3 op-
`erates coil K2. K2 operates the contacts on FIG. 5. K3
`is operated off of control circuit FIG. 6. These devices,
`K2 and K3, control the main discharge from the firing
`circuit to the patient.
`K1 which is controlled by the voltage comparator 52
`and charge logic 56 switches l 10 VAC to transformers
`T1 and T2. This circuit supplies power to the capacitor
`bank 22 in FIG. 5 as required to maintain 1,400 VDC.
`The four rectifiers between T1 and T2 in FIG. 4 and
`the four capacitors 22 in FIG. 5 form two full wave
`voltage double circuits in cascade to generate 1,400 v.
`About 500 joules of energy are then stored in the ca—
`pacitor bank. Initially all four silicone controlled recti—
`fiers SCR are not conducting. The 150 K resistors
`around the SCRs are used to balance the off leakage
`current. The 0.05 mfd — 50 ohm networks around each
`SCR are to suppress switching transcients.
`Terminals 1, 2 and 3 are the monitor points. The volt-
`age between 1 and 2 is proportional to the stored volt—
`age and the voltage to the load. The voltage between
`1 and 3 is proportional to the current in the load. The
`5 ohm, 100 watt resistor serves the dual function of
`current shunt and crowbar protection.
`The remainder of this circuit can be best explained
`by a typical operating sequence. Initially the capacitors
`are charged and all SCRs are off. The cycle starts with
`the start input going to a positive 15 v. This starts the
`0.030 sec. timer 58. At the same time K2 relay begins
`to close. The timer delay is to allow K2 to close com-
`pletely. When the unijunction transistor in the timer
`fires, a large current pulse is fed to trigger transformers
`T1 and T2. These pulses turn on SCR 1 and 2 applying
`power to the load. The LED is turned on by the applied
`voltage and is optically coupled to the photo transistor
`in FIG. 6. This transistor starts timeout comparator Z9.
`When the comparator circuit determines the required
`energy has been delivered, a positive voltage is applied
`to the stop terminal. This fires the small 2N5062 SCR
`‘ generating a high current pulse in T3 and T4. This
`pulse fires SCR 3 and 4 which crow—bars the remaining
`energy in the capacitor bank.
`With respect to FIG. 6, amplifiers 21, 22 and Z3
`form two DC differential amplifiers. These amplifiers 65
`convert the essentially floating inputs 1, 2 and 3 to
`ground referenced signals. The two outputs are v(t)
`from Z2 and i (t) from ZS. These signals are fed to 40
`
`40
`
`which together with Z4 form an analog multiplier. The
`output of ZS in mathematical terms is v(t) x i(l)/K.
`This signal is proportional to the power being deliv-
`ered to the load at any instant of time. ZS is an integra-
`tor which integrates power with time to give energy.
`The AC coupling network on the output of Z3 removes
`the long term DC drift. The output at this point is ap-
`proximately an increasing ramp voltage. This ramp is
`compared to the setting of the pot 36, by comparator
`50. When these are equal the comparator sends the
`stop output high. The peak value of the ramp is stored
`on the 0.22 mfd capacitor in analog storage circuit 44.
`The four transistor amplifier has a gain of +1. This al-
`lows the energy delivered to be displayed on the meter.
`Comparator Z9 (46) performs a similar function to
`50 except it compares the pot 36 setting with time. In
`this way the output pulse width is limited to a maximum
`value for any given setting. This circuit does not affect
`operation for loads of less than 150 ohms.
`Comparator 26, controls the charging of the capaci-
`tor bank. 26 compares the output of amplifier Z2
`which is proportional to the bank voltage to a zener di-
`ode. A certain amount of positive feedback is used as
`controlled hysteresis to prevent chatter of relay K].
`The remaining transistors are used as switches to turn
`on or off certain functions when the manual switch 30
`is closed. For example, the voltage comparator Z6 is
`turned off and comparator 56 and 50 and analog mem-
`ory 44 are turned on.
`The operation described causes the selected energy
`to be delivered regardless of the variable patient load
`resistance. The limits are zero resistance at the patient
`(which will occur if the paddles are touched together)
`or an open or high (over 100 ohm) patient resistance.
`In case of the low limit, all of the energy stored in the
`capacitors would be dissipated within the defibrillator.
`In the case of the high limit, the defibrillator would at-
`tempt to deliver the selected energy but would take too
`long and the time out input will terminate the discharge
`before the selected value is reached. The meter 38 will
`indicate to the operator that the energy that was se-
`lected was not delivered and the load was abnormal.
`Thus it can be seen that a novel defibrillator has been
`provided which insures that the energy delivered to the
`patient will be substantially equal to the selected en-
`ergy regardless of the variable patient load resistance.
`The circuitry of the defibrillator permits a light weight
`and portable defibrillator to be provided so that the de—
`fibrillator can be easily transported to the patient. In
`summary, it can be seen that the defibrillator provides
`the following:
`I. Pro-selection of the energy to be delivered;
`2. Automatic Energy Control to deliver the energy
`selected independently of load; and
`3. Verification by computation of energy delivered
`and indication on an output meter.
`Thus it can be seen that the defibrillator accom-
`plishes at least all of its stated objectives.
`I claim;
`1. A defibrillator comprising in combination,
`an electrical power source,
`a set of electrodes engageable with a patient,
`circuit means connecting said power source to said
`set of electrodes comprising, a computer means, a
`storage capacitor means for storing energy derived
`from said power source, an energy selector means
`for selecting the energy to be delivered to the pa-
`
`50
`
`55
`
`60
`
`6
`
`

`

`6
`a set of electrodes engageable with a patient,
`circuit means connecting said power source to said
`set of electrodes comprising, a computer means, a
`storage capacitor means for storing energy derived
`from said power source, an energy selector means
`for selecting the energy desired to be delivered to
`the patient, a switch means for causing the stored
`energy to be connected to the patient, said switch
`means being operatively electrically connected to
`said computer means, a power monitor means for
`feeding signals to said computer means when said
`stored energy is delivered to the patient, said cir-
`cuit means having a viSual output means for indi-
`cating the energy delievered to the patient, said
`computer means computing the energy actually de-
`livered to the patient and causing said visual output
`means to indicate the energy actually delivered to
`the patient.
`
`*
`
`*
`
`‘*
`
`*
`
`3k
`
`5
`
`3,782,389
`
`tient, said energy selector also feeding an input to
`said computer means, a switch means for causing
`the stored energy to be connected to the patient,
`said switch means being operatively electrically
`connected to said computer means, a power moni-
`tor means for feeding signals to said computer
`means when said stored energy isdelivered to the
`patient, said computer means computing the en-
`ergy delivered to the patient and causing the deliv-
`ery of energy to the patient to be discontinued
`when the computed energy substantially equals the
`selected energy.
`2. The combination of, claim 1 wherein said power
`monitor means comprises a voltage monitor means and
`a current monitor means.
`3. The combination of claim 1 wherein said circuit
`means has a visual output meter which indicates the en—
`ergy delivered to the patient.
`4. A defibrillator comprising in combination,
`an electrical power source,
`
`10
`
`15
`
`20
`
`25
`
`30
`
`35
`
`40
`
`45
`
`50
`
`55
`
`60
`
`65
`
`7
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket