throbber
UNITED STATES PATENT AND TRADEMARK OFFICE
`
`
`
`
`
`
`
`
`
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`MEDTRONIC, INC.
`Petitioner
`
`v.
`
`NUVASIVE, INC.
`Patent Owner
`
`
`
`
`
`
`
`
`
`
`
`
`
`
`Case IPR2013-00506
`Patent 8,361,156
`
`
`
`
`
`
`
`
`
`
`NUVASVE INC.’S
`PATENT OWNER RESPONSE
`
`

`
`Patent No. 8,361,156
`Patent Owner Response
`
`
`
`TABLE OF CONTENTS
`
`Case IPR2013-00506
`Attorney Docket No: 13958-0116IP2
`
`I. 
`II. 
`III. 
`
`IV. 
`
`V. 
`
`VI. 
`
`INTRODUCTION ......................................................................................................... 1 
`BOARD’S DECISION TO INSTITUTE REVIEW ......................................................... 2 
`BACKGROUND ........................................................................................................... 4 
`A. 
`Background of modern implant design ............................................................ 4 
`B. 
`Background of implant sizing ........................................................................... 6 
`C. 
`Background of radiopaque implant markers .................................................. 12 
`D. 
`Background of the ‘156 patent ....................................................................... 14 
`SUMMARY OF THE PRIOR ART ............................................................................. 17 
`A. 
`Telamon (Ex. 1107 and 1108) ....................................................................... 17 
`B. 
`SVS-PR (Ex. 1106) ........................................................................................ 21 
`C. 
`Baccelli (Ex. 1104) ......................................................................................... 25 
`D. 
`Michelson (Ex. 1105) ..................................................................................... 27 
`E. 
`Saber (Ex. 1101, Appendix A) ....................................................................... 31 
`CLAIM 1 IS NOT RENDERED OBVIOUS BY THE PRIOR ART .............................. 32 
`A. 
`None of the prior art references discloses a posterior implant with a pair of
`radiopaque markers in the medial plane. ....................................................... 33 
`The cited references provide no motivation or reasoning to add a pair of
`radiopaque markers in the medial plane of posterior implants ...................... 36 
`It is not obvious to modify SVS-PR or Telamon to include a pair of radiopaque
`markers in the medial plane. .......................................................................... 37 
`CLAIM 5 IS NOT RENDERED OBVIOUS BY THE PRIOR ART .............................. 47 
`A. 
`SVS-PR and Telamon Are PLIF Implants Intended for PLIF Usage ............. 47 
`B. 
`Modifying the Posterior Implants in SVS-PR and Telamon to Exceed 40 mm
`Would Render Them Inoperable For Their Intended Purpose ....................... 49 
`i
`
`C. 
`
`B. 
`
`

`
`Patent No. 8,361,156
`
`Case IPR2013-00506
`Patent Owner Response
`
`Attorney Docket No: 13958-0116IP2
`VII.  CLAIM 9 IS NOT RENDERED OBVIOUS BY THE PRIOR ART .............................. 57 
`A. 
`SVS-PR and Telamon Are PLIF Implants Intended for PLIF Usage ............. 57 
`B. 
`Modifying the Posterior Implants in SVS-PR and Telamon to be 18 mm in
`Width Would Render Them Inoperable For Their Intended Purpose ............ 57 
`VIII.  CONCLUSION .......................................................................................................... 59 
`CERTIFICATE OF SERVICE ........................................................................................
`
`
`
`ii
`
`

`
`Patent No. 8,361,156
`Patent Owner Response
`
`Case Law
`
`
`
`TABLE OF AUTHORITIES
`
`Case IPR2013-00506
`Attorney Docket No: 13958-0116IP2
`
`KSR Int’l Co. v. Teleflex Inc., 550 U.S. 398, 418-19 (2007) ............................... 17, 36, 39, 44
`
`In re Kahn, 441 F.3d 977, 988 (Fed. Cir. 2006) ................................................................... 36
`
`W.L. Gore & Associates, Inc. v. Garlock, Inc., 721 F.2d 1540, 1553 (Fed. Cir. 1983) ... 38, 56
`
`In re Gordon, 733 F.2d 900, 902 (Fed. Cir. 1984) ................................................... 49, 55, 58
`
`DePuy Spine, Inc. v. Medtronic Sofamor Danek, Inc., 567 F.3d 1314, 1326-29 (Fed. Cir.
`2009) .................................................................................................................. 49, 58
`
`In re Ratti, 270 F.2d 810, 813 (CCPA 1959) ................................................................. 55, 56
`
`Decisions of the Patent Trial and Appeal Board
`
`Plas-Pak Industries, Inc. v. Sulzer Mixpac AG, 2014 WL 203101 ........................................ 50
`(Patent Tr. & App. Bd.)
`
`Ex Parte Shufflebotham, et al., 2013 WL 5409872 (Patent Tr. & App. Bd.) ......................... 50
`
`Ex Parte Landschaft, et al., 2013 WL 603815 (Patent Tr. & App. Bd.) ................................ 50
`
`Ex Parte Atsuumi, et al., 2013 WL 1331590 (Patent Tr. & App. Bd.) ................................... 50
`
`Statutes
`
`35 U.S.C. § 103 .................................................................................................................. 2, 3
`
`Rules and Regulations
`
`37 CFR § 42.107(c) ............................................................................................................... 2
`
`37 C.F.R. § 42.22(a)(2) ......................................................................................................... 44
`
`
`
`iii
`
`

`
`Patent No. 8,361,156
`Patent Owner Response
`
`
`
`EXHIBITS
`
`Case IPR2013-00506
`Attorney Docket No: 13958-0116IP2
`
`Exhibit Description
`Ex. #
`NUVA 2101 U.S. Patent No. 7,905,840 to Pimenta et al.
`NUVA 2102 Declaration in Support of the Pro Hac Vice Admission of Michael A. Amon;
`Declaration of Charles Forthaus (Not Filed)
`NUVA 2103 Vetebral Spacer-PR Brochure (Not Filed)
`NUVA 2104
`Forthaus Memo regarding Vertebral Spacer-PR Brochure (Not Filed)
`NUVA 2105 myspinetools.com front page (Not Filed)
`NUVA 2106 myspinetools.com Terms of Use (Not Filed)
`NUVA 2107
`Launch Binder (Not Filed)
`NUVA 2108 Webpage Toad MySQL (Not Filed)
`NUVA 2109 Reserved
`NUVA 2010 U.S. Patent 8,512,408 (Hynes) (Not Filed)
`NUVA 2011 Synthes SVS-PR Guide
`NUVA 2012 Medtronic Sofamor Danek Boomerang brochure
`NUVA 2013 Hynes Deposition Transcript
`NUVA 2014
`510(k) Summary Medtronic Sofamor Danek K122037
`NUVA 2015
`510(k) Summary Telamon® PEEK Spinal System K110562
`NUVA 2016 Synthes Vertebral Spacer – AR Brochure
`NUVA 2017 DePuy Spine Saber Surgical Technique Brochure
`NUVA 2018 Declaration of Barton L. Sachs, M.D. in IPR2013-00206
`NUVA 2019 Moro, et al., “An Anatomic Study of the Lumbar Plexus with Respect to
`Retroperitoneal Endoscopic Surgery”
`
`iv
`
`

`
`Patent No. 8,361,156
`Patent Owner Response
`
`
`
`
`Case IPR2013-00506
`Attorney Docket No: 13958-0116IP2
`
`NUVA 2020 Declaration of Dr. Hansen A. Yuan, M.D.
`NUVA 2021 Curriculum Vitae of Dr. Hansen A. Yuan, M.D.
`NUVA 2022 NuVasive - CoRoent XL Brochure
`NUVA 2023 Redacted Engineering Drawings
`NUVA 2024 Redacted Deposition of Steven DeRidder
`NUVA 2025 Clydesdale® Spinal System Brochure
`NUVA 2026 Clydesdale® Spinal System Images
`NUVA 2027 Medtronic Direct Lateral Interbody Fusion DLIF Surgical Technique
`Brochure
`NUVA 2028 Medtronic DILF-Direct Lateral Interbody Fusion Brochure
`NUVA 2029 Clydesdale® Spinal System Image
`NUVA 2030 Declaration of Patrick Miles
`
`
`
`
`v
`
`
`
`

`
`Patent No. 8,361,156
`Patent Owner Response
`I.
`INTRODUCTION
`Medtronic, Inc. (“Medtronic”) seeks inter partes review of U.S. 8,361,156 (“the ‘156
`
`Case IPR2013-00506
`Attorney Docket No: 13958-0116IP2
`
`
`
`
`patent”), which was asserted against Medtronic in a counterclaim of a lawsuit initiated by
`
`Medtronic. The claims of the ‘156 patent are tailored to a unique combination of features
`
`collectively provided in a spinal implant—features particularly suited for implant insertion in
`
`a lateral, trans-psoas surgical approach path. Yuan Decl. (Ex. 2020), ¶ 47; Ex. 1115, 5:29-
`
`31. The implant depicted in the ‘156 patent was the first lateral, trans-psoas fusion implant
`
`that achieved commercial success, and it pioneered a surgical method that changed the
`
`entire U.S. market for lumbar fusion procedures. Miles Decl. (Ex. 2030), ¶ 9; Yuan Decl., ¶¶
`
`48-49. Indeed, the lack of any anticipatory references in Medtronic’s Petition underscores
`
`the novelty of the claimed invention and the fact that this unique combination of features
`
`recited in the ‘156 patent claims were never previously implemented in a single implant.
`
`Petitioner’s obviousness arguments in this proceeding alleged that traditional
`
`posterior lumbar interbody fusion (“PLIF”) implants, such as the Telamon and SVS-PR
`
`implants, would be modified to borrow elements of Baccelli’s anterior cervical implant in a
`
`way that was never previously conceived or suggested merely because the SVS-PR, the
`
`Telamon, and the Baccelli implants are “from the same field of artificial spinal implants used
`
`in spinal fusion.” See, e.g., Ex. 1101, ¶¶ 68, 70, 72, 74, 99, 106. This is not enough.
`
`Petitioner’s position that it is obvious to create implants with virtually any combination of
`
`design features in any orientation ignores the basic fact that interbody fusion implant
`1
`
`
`
`

`
`Patent No. 8,361,156
`
`Case IPR2013-00506
`Patent Owner Response
`
`Attorney Docket No: 13958-0116IP2
`designers employ a time-consuming and rigorous process of research, testing, and analysis
`
`when choosing features, size, and shape for each commercial implant design—all centered
`
`on the intended use (insertion approach and final resting position for vertebral fusion) of the
`
`implant. See Yuan Decl., ¶ 40; Ex. 2013, ¶ 44:19-47:19. Consequently, ordinary implant
`
`designers do not modify, move, or add elements to commercial, FDA-approved interbody
`
`fusion implants haphazardly—they require a concrete and important reason to do so.
`
`Petitioner’s proposed obviousness grounds include no articulated reasoning with a rationale
`
`underpinning to modify the traditional PLIF block implants in a way that would result in the
`
`implant of claim 1.
`
`II.
`
`BOARD’S DECISION TO INSTITUTE REVIEW
`On February 13, 2014, the Board instituted inter partes review of U.S. 8,361,156
`
`(“the ‘156 patent”). Based on unrebutted declarant testimony only from Petitioner,1 the
`
`Board concluded that Petitioner had demonstrated a “reasonable likelihood” that it will
`
`prevail on the challenges and granted review on the following grounds:
`
` Claims 1–4, 7, 8, 10–14, 19, 20, 23, 24, 26, and 27 as obvious under 35 U.S.C. §
`
`103 by the combination of SVS-PR and Baccelli;
`
`
`
`1 37 CFR § 42.107(c) prohibits new testimony evidence in the preliminary response.
`
`
`
`2
`
`

`
`Patent No. 8,361,156
`Patent Owner Response
`
`
`
`
`Case IPR2013-00506
`Attorney Docket No: 13958-0116IP2
`
` Claims 5, 6, and 9 as obvious under 35 U.S.C. § 103(a) by the combination of SVS-
`
`PR, Baccelli, and Michelson;
`
` Claim 25 as obvious under 35 U.S.C. § 103(a) by the combination of SVS-PR,
`
`Baccelli, and Telamon;
`
` Claims 1–4, 7, 10–14, 19, 20, and 23–27 as obvious under 35 U.S.C. § 103(a) by the
`
`combination of Telamon and Baccelli; and
`
` Claims 5, 6, 8, and 9 as obvious under 35 U.S.C. § 103(a) by the combination of
`
`Telamon, Baccelli, and Michelson.
`
`Feb. 13, 2014 Decision in IPR2013-00506 (“Decision”), p. 19-20. Throughout the Decision,
`
`the Board recognized that Patent Owner only presented attorney argument and noted that
`
`there was no supporting evidence. For example, in response to Patent Owner’s argument
`
`that one proposed modification would render the SVS-PR implant inoperable for its intended
`
`purpose, the Board stated: “[i]t is, therefore, not known, without additional evidence, if the
`
`distance from the posterior to anterior edges of the disc space in SVS is less than, equal to,
`
`or greater than 40 mm.” Id. at 13. The Board also noted that Patent Owner did not provide
`
`evidence “to show that using an implant that is greater than 40 mm in length would” injure
`
`the patient. Id. at 13-14 (“without additional evidence”).
`
`The Board will find the positions in this Response fully supported by evidence,
`
`including the evidence identified as lacking in the Decision.
`
`
`
`3
`
`

`
`
`
`
`Patent No. 8,361,156
`Patent Owner Response
`III.
`BACKGROUND
`A.
`Background of modern implant design
`Spinal fusion implants are created via a time-consuming and rigorous process of
`
`Case IPR2013-00506
`Attorney Docket No: 13958-0116IP2
`
`research, testing, and analysis. Yuan Decl., ¶¶ 39-41; see also Hynes Depo. (Ex. 2013),
`
`45:9-46:4, 46:16-47:2. This is most certainly true for commercially available, FDA-approved
`
`interbody spinal fusion implants, which have specific sizes, surface features, apertures,
`
`tooling engagement structures, and other features specifically associated with the
`
`designer’s intended use of the spinal implant. Yuan Decl., ¶¶39-40. In fact, prior to 2004
`
`and still to this day, nearly all commercially available spinal implants were provided with
`
`surgical “guide” literature so as to instruct spinal surgeons on the intended use of the
`
`commercial implant. This was certainly true for the Telamon implant and the SVS-PR
`
`implant cited in the Petition. See Ex. 1108 (Telamon “guide”); Ex. 2011 (SVS-PR “guide”).
`
`Even slight changes to the design of a commercially available interbody fusion implant,
`
`especially the dimensions and marker orientation/positioning, can have a dramatic effect on
`
`the functionality and intended use of the implant (and possibly FDA scrutiny), and any
`
`ordinary implant designer understands this fact. Yuan Decl., ¶ 40.
`
`It is not only important that the implant be designed just right, but it is critical that the
`
`implant also be used just right. Yuan Decl., ¶¶ 38-40. Interbody spinal fusion implants and
`
`spinal access equipment are designed to be used by average surgeons in specific a
`
`medical procedure according to very a specific technique. Yuan Decl., ¶ 39; see also
`4
`
`
`
`

`
`Patent No. 8,361,156
`
`Case IPR2013-00506
`Patent Owner Response
`
`Attorney Docket No: 13958-0116IP2
`Hynes Depo. 44:9-11. Spinal fusion implants are designed not as universal tools usable by
`
`ordinary surgeons to freelance in new and exciting ways. Yuan Decl., ¶ 39. They are
`
`designed according to a specific use that is safe, effective, and reproducible in the hands of
`
`all surgeons—not just the most highly skilled and experimental group of surgeons, but also
`
`the ordinary surgeon that is faithful to the surgical “guide” and other related literature. Yuan
`
`Decl., ¶ 39. Surgeons receive extensive training using spinal implants (including the
`
`intended approach path to the spine) and spinal access equipment such that every surgeon
`
`can follow the procedure to achieve the identical, reproducible result that is (hopefully) safe
`
`and very effective. Yuan Decl., ¶¶ 39, 52. This reproducibility is such a critical aspect of
`
`interbody fusion implant design because it means not that the procedure might be safe
`
`sometimes but that it will be safe in the hands of ordinary surgeons (not just a small group
`
`of the most highly skilled spine surgeons). Yuan Decl., ¶¶ 39, 52.
`
`Even small design modifications to the dimensions and marker
`
`orientation/positioning of an interbody fusion implant make a big difference from a clinical
`
`perspective. Yuan Decl., ¶ 40. For interbody fusion implants, it is critically important that
`
`the implant fits where it is supposed to fit along the intended insertion path, can be
`
`positioned as it is supposed to be positioned, supports the load that it is supposed to
`
`support, aligns the spine as it is supposed to align the spine, stays in place where it is
`
`supposed to stay, and allows the spine to fuse and heal the way the spine is supposed to
`
`fuse and heal. Yuan Decl., ¶ 41. The difference between meeting all of these requirements
`5
`
`
`
`

`
`Patent No. 8,361,156
`
`Case IPR2013-00506
`Patent Owner Response
`
`Attorney Docket No: 13958-0116IP2
`and meeting only some of them is the difference between having a great implant and having
`
`an implant that is unsuitable for its intended purpose. Yuan Decl., ¶ 41. Small design
`
`changes to the dimensions and marker orientation/positioning can and do have unintended
`
`effects that negatively affect the performance of the spinal interbody fusion implant. Yuan
`
`Decl., ¶ 40. This is why ordinary implant designers are so careful about such design
`
`changes to commercially available, FDA-approved implants, and would not be prompted to
`
`modify such a commercially available implant in a manner that would prevent or otherwise
`
`reduce the safety of its original intended use. Yuan Decl., ¶¶ 40, 103, 105.
`
`This background is important to understanding how and why persons of ordinary skill
`
`in the art design implants the way they do. There are different commercially available
`
`implant designs in the prior art (and in this proceeding) not because “any-old-design will
`
`work,” but because each particular design was determined to be the right design for that
`
`particular intended use set forth in the surgical “guide” literature. Yuan Decl., ¶ 39.
`
`Background of implant sizing
`B.
`The size of a spinal interbody fusion implant is extremely important for the
`
`performance of the implant and for the intended insertion path of the implant. Yuan Decl.,
`
`¶¶35, 39-41. In general, spinal fusion interbody implants need to be small—on the order of
`
`millimeters. See e.g. Ex. 1115, claims 5 and 9; Ex. 1105, 10:41-47; Ex. 1106, p. 1; Ex.
`
`1107, p. 2. Spinal fusion interbody implants are small for a number of reasons related to
`
`human anatomy in and near the spinal column. Yuan Decl., ¶¶ 34, 35. For example, spinal
`6
`
`
`
`

`
`Patent No. 8,361,156
`
`Case IPR2013-00506
`Patent Owner Response
`
`Attorney Docket No: 13958-0116IP2
`fusion interbody implants are designed to reside in the disc space between vertebrae and
`
`replace a vertebral disc, which are relatively small in humans (again, on the order of
`
`millimeters). See Ex. 1113, p. 4; Yuan Decl., ¶ 30, 34.
`
`Vertebral body, Ex. 1113, p. 3
`
`
`
`Critical anatomy exists near the surgical location, which can force further size
`
`constraints. Yuan Decl., ¶¶ 26-30, 34, 35. For example, the spinal cord passes through the
`
`spinal canal toward the posterior (rear) portion of the vertebrae. Yuan Decl., ¶ 26, 28.
`
`Damaging the spinal cord can paralyze a patient and is a very real risk during spinal fusion
`
`procedures. Yuan Decl., ¶ 34, 111. Accordingly, implants (and insertion tools) are sized to
`
`avoid contact with the spinal cord during and after implantation. Yuan Decl., ¶ 34, 35. This
`
`especially constrains implant size in posterior procedures, such as a PLIF (posterior lumbar
`
`interbody fusion) procedure. Yuan Decl., ¶ 35. PLIF procedures for two of the prior art
`
`implants (SVS-PR and Telamon) are illustrated below.
`
`
`
`7
`
`

`
`Patent No. 8,361,156
`Patent Owner Response
`
`
`
`
`Case IPR2013-00506
`Attorney Docket No: 13958-0116IP2
`
`
`
`
`SVS-PR Implant (Ex. 2011) and Telamon Implant (Ex. 1108)
`
`Such PLIF implants are narrow so that two such PLIF implants can be inserted in
`
`direct posterior paths on opposite sides of the spinal canal and positioned side by side in
`
`the disc space. Yuan Decl., ¶¶ 99, 111-113. The SVS-PR and Telamon implants are
`
`reproduced below at actual size2 to illustrate just how small such implants are.
`
`
`
`
`
`SVS-PR Implant (Exhibit 1106) and Telamon Implant (Ex. 1108) (annotated)
`
`
`
`2 These images are sized so as to be the actual dimension disclosed in Ex. 1106-1108
`
`when printed on 8 ½ by 11 inch paper.
`
`
`
`8
`
`

`
`Patent No. 8,361,156
`
`Case IPR2013-00506
`Patent Owner Response
`
`Attorney Docket No: 13958-0116IP2
`Critical anatomy also include nerves and nerve roots near the vertebrae, including
`
`sensitive nerve roots passing through the psoas muscles lateral to (toward the sides of) the
`
`lumbar vertebrae. Yuan Decl., ¶ 26, ; Ex. 2019, Moro et al., An anatomic study of the
`
`lumbar plexus with respect to retroperitoneal endoscopic surgery, SPINE Vol. 28 (2003), pp.
`
`423, 428 (showing that Medtronic’s agent, Dr. Mathews, expressed the conventional
`
`wisdom in the prior art that “surgeons have not felt comfortable with dissecting the psoas
`
`because of the presence of the lumbar plexus”). Damaging such nerve roots can also
`
`paralyze or otherwise permanently harm a patient. Yuan Decl., ¶49. There are also blood
`
`vessels on different sides of the vertebrae, including particularly critical vessels anterior
`
`(front) of the vertebrae. Yuan Decl., ¶¶ 26, 34-35. Damaging one of those blood vessels
`
`puts a patient’s life at risk. Yuan Decl., ¶¶34-35, 98. Consequently, spinal fusion implants
`
`are sized small enough to avoid damaging critical blood vessels and spinal nerves during
`
`and after the implantation procedure. Yuan Decl., ¶¶34-35, 48-52, 98, 111.
`
`While it is desirable for implants to be large, the implant designs must still facilitate
`
`their intended use (PLIF procedure, ALIF procedure, TLIF procedure, etc.) within the
`
`anatomical/clinical constraints described above. Yuan Decl., ¶¶ 35, 40-41. Larger spinal
`
`fusion implants have a larger footprint and more material for structurally supporting and
`
`properly spacing adjacent vertebrae. Yuan Decl., ¶ 41; see also Ex. 1107, p. 2 (advertising
`
`“[h]igh ultimate strength” and providing load, fatigue, and stiffness data); Ex. 1106, p. 2
`
`(providing data on expulsion strength, compressive strength, fatigue); Ex. 1105, 6:43-57
`9
`
`
`
`

`
`Patent No. 8,361,156
`
`Case IPR2013-00506
`Patent Owner Response
`
`Attorney Docket No: 13958-0116IP2
`(touting the importance of length and width for creating a “much greater surface area of
`
`contact”). If an implant is too small, the implant can fail to properly support the vertebrae
`
`and cause pain (for example pinched nerves) that is even greater than the pain the spinal
`
`fusion procedure was intended to cure. Yuan Decl., ¶ 41.
`
`The conflict between the benefits of a large implant and those of a small implant
`
`force designers to undertake critical analysis, research, and testing when designing a spinal
`
`fusion implant. Yuan Decl., ¶¶ 39-41. These constraints, in addition to shape requirements
`
`that are necessary or desirable for a specific procedure or patient condition as well as other
`
`implant features causes the process of designing an implant to be complex and rigid in that
`
`even slight changes to the implant dimensions and radiopaque marker structure/positioning
`
`can have significant clinical problems. Yuan Decl., ¶40. Determining superior implant
`
`dimensions can be an extremely difficult process, especially when attempting to design
`
`implants for relatively new spinal fusion procedures that have no conventional implant
`
`design to emulate. Yuan Decl., ¶¶ 48-52. There is no guarantee that a designer will get it
`
`right. Yuan Decl., ¶¶ 50-52. Indeed, as described below for purposes of a laterally inserted
`
`implant, the Michelson patent illustrates several examples of theorized lateral implant
`
`designs that were improperly sized and shaped, recognized in the art as unsafe, and never
`
`implemented for insertion into a live human patient.
`
`In 2004, spinal fusion via a lateral (often called “direct lateral” or “extreme lateral”)
`
`access path was one of those procedures without a conventional implant design. Yuan
`10
`
`
`
`

`
`Patent No. 8,361,156
`
`Case IPR2013-00506
`Patent Owner Response
`
`Attorney Docket No: 13958-0116IP2
`Decl., ¶¶ 48-49. While lateral fusion procedures had been performed by a handful of highly
`
`skilled surgeons for years, such procedures were relatively rare and were not performed
`
`(especially in the lumbar spine) by ordinary spine surgeons. See, e.g., Yuan Decl., ¶51.
`
`This was, in part, due to reluctance by surgeons to traverse the psoas muscle for fear of
`
`damaging nerve roots that extend therethrough. Yuan Decl., ¶49; Ex. 2019, p. 428
`
`(“surgeons have not felt comfortable with dissecting the psoas because of the presence of
`
`the lumbar plexus”). Consequently, there was no appreciable market for lateral spinal
`
`fusion implants, and importantly, there was no conventional or standard design at that time.
`
`Yuan Decl., ¶¶ 49-52.
`
`Dr. Gary Michelson, the inventor of the Michelson patent (Ex. 1105) was one of
`
`those people trying to design lateral implants. See Ex. 1105; Yuan Decl., ¶50. There are a
`
`number of implant designs illustrated in Michelson (see Ex. 1105, Figs. 1-20) but they were
`
`not commercialized, and it is well known to both parties here that not even Dr. Michelson
`
`inserted such an implant in a live human patient. Michelson discloses what Dr. Michelson
`
`mistakenly believed to be appropriate sizes for lateral implants for use in the lumbar spine
`
`for dowel implants (see Ex. 1105, 7:21-26) and for more rectangular implants (see Ex. 1105,
`
`10:41-47). But Dr. Michelson’s proposed overall length and width combinations, along with
`
`the shapes, for his lateral lumbar implants were recognized in 2004 and earlier as being
`
`wrong, unsafe, or both, which is one reason why those proposed sizes are not ultimately
`
`practiced in the industry. Yuan Decl., ¶50. The Michelson patent never illustrated any
`11
`
`
`
`

`
`Patent No. 8,361,156
`
`Case IPR2013-00506
`Patent Owner Response
`
`Attorney Docket No: 13958-0116IP2
`implant design that was safe and feasible for lateral insertion, so even in the “post-
`
`Michelson” (a new term used repeatedly by the Petitioner) time frame before 2004, the
`
`industry was left with nothing more than wrong answers and a general avoidance by
`
`average spine surgeons of any lateral, trans-psoas surgeries. Dr. Michelson’s sizes are
`
`also not the sizes disclosed in the ‘156 patent for a laterally insertion implant, as will be
`
`discussed below. Yuan Decl., ¶¶ 90-91. Michelson is, therefore, one example of the
`
`reality that knowing an intended direction of insertion for a spinal fusion implant does not
`
`make it easy (or obvious) to select appropriate implant dimensions. Yuan Decl., ¶¶50-52.
`
`Background of radiopaque implant markers
`C.
`Radiopaque markers are called “radiopaque” because they are “opaque” when
`
`viewed via X-ray imaging. Yuan Decl., ¶¶42-46; Ex. 1115, FIG. 21. Patent Owner does not
`
`dispute that radiopaque markers have long been used in implantable medical devices to
`
`show positioning of such devices, both during and after the implantation procedure.
`
`Radiopaque markers are traditionally positioned at the extremities of spinal implants to
`
`show the location of those extremities. Yuan Decl., ¶45. For example, elongated implants
`
`such as the Telamon and SVS-PR would traditionally include a radiopaque marker at each
`
`lengthwise end. Yuan Decl., ¶¶ 78-82 and 83-86. These two markers allow the surgeon to
`
`ensure that each end is at the appropriate position with respect to the vertebrae, usually
`
`along a hardened bone region near the rim of the vertebrae. Yuan Decl., ¶¶ 81 and 86.
`
`
`
`12
`
`

`
`Patent No. 8,361,156
`
`Case IPR2013-00506
`Patent Owner Response
`
`Attorney Docket No: 13958-0116IP2
`Dr. Hynes also confirmed that putting “one on one end and one on the other end” of
`
`a PLIF implant is what he would do and “what’s commonly done.” Hynes Depo., 138:16-22.
`
`Dr. Hynes explained the purpose of putting radiopaque markers at each lengthwise end of a
`
`PLIF implant as follows: “if you have one marker on the other end, then you can follow the
`
`distal or the leading edge in and see where it is and then you have a marker on the trailing
`
`end and see where it ends up.” Hynes Depo., 142:1-5. Dr. Hynes further explained that
`
`ordinary surgeons would also “pick the front and the back in the direction of insertion. It's
`
`just intuitive, but it's also safer for the patient.” Hynes Depo., 142:8-12.
`
`The location of implant markers is very intentional to convey specific information to
`
`the implanting surgeons. Yuan Decl., ¶46. While it is technically possible from a
`
`manufacturing standpoint to add radiopaque markers to a PEEK implant in virtually an
`
`infinite number of different positions, the positions actually selected are very deliberate and
`
`purposeful based upon the original intended use (insertion path and final resting position) of
`
`the implant. Yuan Decl., ¶¶ 45-46. Because of this, extra markers are simply not added
`
`without a specific purpose. Id. The testimony evidence here from both parties shows the
`
`agreement that excess markers positioned in an implant at unnecessary locations can
`
`clutter and confuse x-ray imaging viewed by the surgeon. Yuan Decl., ¶¶ 45-46, 98-99, and
`
`102; Hynes Depo., 163:23-164:25. Indeed, when asked about adding additional markers
`
`across the medial center of a traditional PLIF implant, Dr. Hynes also agreed that such
`
`
`
`13
`
`

`
`Patent No. 8,361,156
`
`Case IPR2013-00506
`Patent Owner Response
`
`Attorney Docket No: 13958-0116IP2
`additional markers could create redundancy or confusion. Hynes Depo., 163:23-164:25.
`
`Dr. Hynes explained:
`
`What happens is the wrong marker, the size of the marker, direction of the
`marker, and the wrong place actually creates confusion sometimes. You'd
`see less. Sometimes less is better. We have to look at it.
`And so there's all kinds of combinations you could use and we could make
`them different sizes, different shapes, but then you start getting too much
`information and it actually gives you a less information.
`Hynes Depo., 164:10-19 (emphasis added). In order to convey specific information without
`
`creating confusion, designers carefully and critically select the number and location of
`
`radiopaque markers in each implant based upon the implant’s intended clinical use, not
`
`merely because a metallic marker could be manufactured anywhere into a plastic structure.
`
`Yuan Decl., ¶ 46. The number and position of radiopaque markers in a radiolucent implant
`
`is dictated and limited by the intended clinic use of that implant, not by manufacturing
`
`capabilities. Petitioner, and its alleged expert Dr. Hynes, appear to ignore this reality of
`
`interbody fusion implant design. Hynes Depo., 140:23-141:13 (relying on the possible
`
`manufacturing simplicity, not the clinical purpose of such markers, and stating “[t]hey’re [the
`
`markers are] easy to put in….You can have any choice you want.”)
`
`Background of the ‘156 patent
`D.
`The claims of the ‘156 patent do not broadly claim all lateral fusion implants. They
`
`do not claim all radiolucent implants having multiple radiopaque markers. They do not claim
`
`
`
`14
`
`

`
`Patent No. 8,361,156
`
`Case IPR2013-00506
`Patent Owner Response
`
`Attorney Docket No: 13958-0116IP2
`any type of PLIF implant or anterior implant (and in fact recite features that are not, and
`
`would not be, collectively provided in any PLIF implant or anterior implant). To be clear, the
`
`‘156 patent claims are specifically and reasonably tailored to a unique combination of
`
`features collectively provided in a spinal implant for insertion in a lateral, trans-psoas
`
`surgical approach path. Yuan Decl., ¶ 47. The implant depicted in Figs. 2-6, 20, and 21 of
`
`the ‘156 patent is NuVasive’s CoRoent XL implant. Yuan Decl., ¶ 53. The CoRoent XL
`
`implant is not just illustrated, but also has all of the features claimed in claim 1 and many of
`
`the dependent claims of the ‘156 patent. Yuan Decl., ¶¶ 53-63. The CoRoent XL implant
`
`has a unique combination of features including the size, shape, and orientation of the
`
`support structure and alignment markers, and was the first lateral, trans-psoas fusion
`
`implant that achieved commercial success, and indeed changed the entire U.S. market for
`
`lumbar fusion procedures. Yuan Decl., ¶¶ 47 and 53-63; Miles Decl., ¶¶ 7-10 and App. A.
`
`Prior to Patent Owner (“NuVasive”) developing the implant depicted in the ‘156
`
`patent, virtually all lumbar fusion surgeries actually performed in the U.S. used implants that
`
`approached the spine from the rear (posterior or posterolateral approaches) or from the
`
`front (anterior or anterolateral approaches)—approach paths that avoided the risks of the
`
`psoas muscle and the required implant structures of smaller sizes and different shapes.
`
`Yuan Decl., ¶¶ 48-49; see also Ex. 2019, p. 428 (“surgeons have not felt comfortable with
`
`dissecting the psoas because of the presence of

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket