throbber
NOTICE: This standard has either been superseded and replaced by a new version or discontinued.
`Contact ASTM International (www.astm.org ) for the latest information.
`
`An American National Standard
`
`ID Designation: D 2923-95 (cid:9)
`
`Standard Test Method for
`Rigidity of Polyolefin Film and Sheeting’
`
`This standard is issued under the fixed designation D 2923; the number immediately following the designation indicates the year of
`original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
`superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
`
`1. Scope
`1.1 This test method describes two procedures for measur-
`ing the rigidity of polyolefin film and sheeting.
`1.2 Procedure A prescribes a procedure using high-voltage
`static eliminators and the use of TFE-fluorocarbon 2-coated
`plates to overcome the spurious effects of static electricity and
`friction.
`1.3 Procedure B prescribes the use of a fine powder on
`uncoated plates to achieve a similar effect.
`
`(cid:151)Although the two procedures are designed to achieve similar
`NOTE
`effect, they may not achieve the same results.
`1.4 The values stated in ST units are to be regarded as the
`standard. The values given in parentheses are for information
`only.
`1.5 This standard does not purport to address all of the
`safety concerns, f(cid:149) any, associated with its use. It is the
`responsibility of the user of this standard to establish appro-
`priate safely and health practices and determine the applica-
`bility of ’regulatory limitations prior to use.
`
`NOTE 2(cid:151)There is no similar or equivalent ISO standard.
`
`2. Referenced Documents
`2.1 ASTM Standards:
`D618 Practice for Conditioning Plastics and Electrical
`Insulating Materials for Testing’
`D 1898 Practice for Sampling of Plastics’
`
`3. Terminology
`3.1 Definitions:
`combination of thickness and inherent
`3.1.1 rigidity(cid:151)that
`stiffness of a polyolefin film or sheet which resists flexure.
`
`4. Summary of Test Method
`4.1 The resistance to flexure of the sample is measured by a
`strain gage affixed to the end of a beam, the opposite end of
`which flexes the sample by forcing it into a groove or slot in a
`
`This test method is under thejurisdiction ofASTM Committee D-20 on Plastics
`and is the direct responsibility of Subcommittee D20.19 on Film and Sheeting.
`Current edition approved Oct. 10, 1995. Published December 1995. Originally
`published as D 2923 (cid:150)70. Last previous edition D 2923 (cid:150)94.
`This edition includes the addition of Note 2 and Section 15, and a correction to
`the equation in 12.1.
`i’his test method is based on the use of Teflonfi, a registered trademark of E. I.
`duPont de Nemours & Co.
`Annual Book of ASTM Standards, Vol 08.01.
`
`horizontal platform upon which the sample rests. An indicating
`microammeter, wired to the strain gage, is calibrated in grams
`of load sensed by the strain gage. The rigidity is read directly
`from the meter and expressed as grams per centimetre of
`sample width.
`
`5. Significance and Use
`5.1 The rigidity of polyolefin web can affect its machinabil-
`ity, particularly on those packaging machines where a cut
`portion of a web is required to remain flat momentarily without
`being supported on all sides.
`5.2 Rigidity is not a simple property since it depends on two
`other properties of the sample: the thickness (gage), and the
`stiffness which is an inherent property of the material of which
`the film or sheet is made. The combined effect of these two
`factors is the rigidity that influences performance on converting
`machines.
`
`6. Interferences
`6.1 Static electricity has considerable influence on the
`measured rigidity. It contributes to poor precision and accu-
`racy, frequently giving results biased toward the high side.
`6.2 To a lesser extent, precision and accuracy is adversely
`affected by frictional properties of the sample, particularly
`when the coefficient of friction is close to 1.
`6.3 The interference caused by the level of static electricity
`and friction of the specimen is overcome in Procedure A by
`electrically destaticizing the sample and using TFE-
`fluorocarbon-coated plates and in Procedure B by dusting the
`platform with a fine powder at the start of the test.
`
`7. Apparatus
`7.1 Procedure A:
`7.1.1 Handle-O-Meter, or equivalent, with TFE-
`fluorocarbon-coated plates complete with calibrating and aug-
`menting weights.’
`7.1.2 Cutting Board or Template suitable for preparing 203
`by 203-mm (8.0 by 8.0-in.) specimens.
`7.1.3 High-Voltage Static Eliminator-5 and Generator.
`7.2 Procedure B:
`7.2.1 Handle-O-Meter, or equivalent, complete with cali-
`brating and augmenting weights.
`
`The Handle-O-Meter and auxiliary equipment is available from Thwing-Albert
`Instrument Co., Philadelphia, PA.
`Simco Shockless Bars are suitable for this test method.
`
`Copyright ' ASTM. 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959. United States.
`
`Copyright by ASTM Int’l (all rights reserved); Mon Jun tO 08:55:49 EDT 2013
`Downloaded/printed by
`Susan Hanrahan (Vorys, Sater, Seymour and Pease LLP) pursuant to License Agreement. No further reproductions authorized.
`
`PAGE 000001
`
`

`

`lffTh D 2923
`11.1.7 Upon completion of the cycle, rotate the specimen
`7.2.2 Cutting Board or Template suitable for preparing 203
`90(cid:176) and repeat 11.1.1 through 11.1.6 for measurement in the
`by 203-mm (8.0 by 8.0-in.) specimens.
`transverse direction.
`7.2.3 Pine-Particle Silica, starch, or equivalent, having an
`approximate nominal particle size of 50 i.tm or less.
`11.1.8 Turn the specimen over so that the opposite side of
`the film is being forced into the opening and repeat 11.1.1
`through 11.1.7.
`11.1.9 Repeat 11.1.1 through 11. 1.8 for the remaining speci-
`mens.
`11.2 Method B:
`11.2.1 Place a light "dust" layer of powder on the Handle-
`0-Meter platform.
`11.2.2 Follow 11.1.3 to 11.1.9.
`
`8. Sampling
`8.1 The material shall be sampled in accordance with
`Practice D 1898.
`
`12. Calculation
`12.1 Calculate the average Handle-0-Meter rigidity in the
`machine direction and transverse direction for each side as
`follows:
`
`R jff) = (r/n) I kl7 or RTD (cid:9)
`
`(r/n)/tV (cid:9)
`
`(I)
`
`where:
`R
`
`= average machine direction rigidity for Side I,
`g/cm,
`= average TD rigidity for Side 1, g/cm,
`= sum of meter readings for specimens tested, g,
`r (cid:9)
`= number of specimens tested, and
`n (cid:9)
`= specimen width along the slot, cm.
`W (cid:9)
`12.1.1 If the Run (or RTD) for Side I is equal to Rauji (or
`R7-) for Side 2, the average of the two may be taken as the
`R 111 (or RTD) of the specimen. The two R 1 ’s (or RTD’s) shall
`be judged equal if they agree with each other within – 10 %,
`the error limits of the test.
`
`13. Report
`13.1 Report the following information:
`13. 1.1 The average Handle-0-Meter rigidity in each direc-
`tion,
`13.1.2 Specimen width,
`13.1.3 Procedure used,
`13.1.4 Direction tested, and
`13.1.5 Sample thickness.
`
`NOTE 5 I the sides are dilTbrent, report values for each side tested.
`
`14. Precision and Bias
`14.1 The precision and bias statement is currently being
`prepared.
`
`15. Keywords
`15.1 film; olefins; rigidity; sheet
`
`9. Test Specimens
`9.1 Using a cutting board or template, cut a minimum of
`three 203 by 203-mm (8.0 by 8.0-in.) test specimens with the
`edges parallel to the machine and transverse direetiois of the
`film. Mark the specimens in the machine direction and trans-
`verse direction.
`9.2 The specimens shall be conditioned in accordance with
`Procedure A of Practice D 618.
`
`10. Calibration
`10.1 Calibrate the Handlc-0-Meter in accordance with the
`instructions described in Appendix Xl.
`10.2 Calibrate other equipment in accordance with the
`manufacturer’s instructions.
`
`Ii. Procedure
`11.1 MethodA:
`11.1.1 Turn on the static eliminator.
`11.1.2 Just before placing a test specimen on the instrument,
`slowly draw the specimen several times over the bar of the
`static eliminator.
`11. 1.3 With the penetrator beam raised above the slot, lay
`the test specimen on the platform of the instrument so that the
`length of the slot is at right angles to the machine direction of
`the specimen. For example, in measuring machine direction
`rigidity, the bottom edge of the penetrator beam, which is
`parallel to the slot length, will contact the film in a line
`perpendicular to the machine direction of the film.
`11.1 .4 Position the specimen on the platform so that three
`fourths of the specimen is to the right of the slot.
`11. 1.5 With the specimen in correct position and the toggle
`switch on RUN, press the momentary contact switch to start
`the test cycle.
`11. 1.6 Read the maximum value in grams (to the nearest 0.5
`g) indicated on the meter.
`
`Norc 3---- -Readings in excess of the maximum meter value require use
`of an augmenting weight in accordance with the manufacturer’s instruc-
`tions. The proper reading with a 40-g augmenting weight is 40 g plus the
`maximum meter value.
`Nyu- 4--If the total reading is in excess of the maximum value plus
`augmenting weights, reduce the specimen width (measured distance along
`the slot) until an on-scale reading is obtained.
`
`Copyright by ASTM int’l (all rights reserved); Mon Jun 1008:55:49 EDT 2023
`Downlouded/prinied by
`Susan Hanrahan (Vorys, Saier, Seymour and Pease LLP) pursuant to License Agreement. No further reproductions authorized.
`
`PAGE 000002
`
`(cid:9)
`

`

`D 2923
`
`APPENDIX
`
`(Nonmandatory Information)
`
`Xl. CALIBRATION
`
`Xl .1 The instrument must be located on a solid, vibration-
`free, level surface. Refer to Fig. Xl. 1 for the placement and
`names of the components of the Handle-O-Meter.
`
`X1.2 The opening through which the penetrator beam
`travels must be checked each time the instrument is used or
`once a week if the machine is in continual use. For routine
`work, a 5-mm (0.197-in.) opening will be used.
`X1.2.1 Place the centering gage shown in Fig. X1.2 in the
`opening with the penetrator beam resting in the upper slot of
`the centering gage. Fig. X1.3
`X1.2.2 The gage should fit snugly in the opening (no more
`than 0.025-mm (0.01-in.) clearance on each side).
`Xl2.3 Adjust the opening if the gage is loose by releasing
`the set screws holding the plate and sliding both sides until
`contact is made with the centering gage. Tighten the set screws
`holding the plates and remove the gage.
`
`XI.3 The following steps should be followed in consecu-
`tive order to check the calibration of the Handle-O-Meter.
`Refer to Fig. X1.i for placement of control.
`
`FIG. X1.2 Centering Gage
`
`X1.3.1 Put toggle switch in RUN position and press the
`momentary contact switch. The blade of the penetration beam
`will move into the slot to a depth governed by the position of
`the load cell anchor nut.
`Xl.3.2 Adjust the load cell anchor nut so that the blade will
`penetrate to a depth of 7.94 mm (0.3125 in.) below the surface
`of the platform. Turning off the nut clockwise will increase the
`depth setting and vice versa.
`
`1. Load cell
`2. Microanimeter
`3. Case
`4. Penetrator beam
`5. Load-cell bracket
`6. Specimen platform
`7. Cam
`8. Penetrator beam counterweight
`9. Counterweight
`
`HI-LO SWITCH
`
`10. Toggle switch
`11. Momentary toggle switch
`12. Blade
`13. Load-cell anchor nut
`14. Motor
`15. Range calibrating weight
`16. Range rheostat
`17. 40-g augmenting weight
`
`FIG. X1.1 Handle-O-Meter Components
`
`Copyright by ASTM Intl (all rights reserved); Mon Jun 10 08:55:49 EDT 203
`Downloaded/printed by
`Susan Hanrahan (Volys, Sater, Seymour and Pease LLP) pursuant to License Agreement. No further reproductions authorized.
`
`PAGE 000003
`
`

`

`(cid:149) D 2923
`
`40-g
`
`SLOT __N\ i
`
`r
`
`SPECIMEN
`
`rPENmr$MhboS
`
`SPE
`
`WIDTH \\
`
`0000
`
`
`
`L aIRECrICH
`STING TESTED
`FIG. X1.3 Specimen in Place for Test
`
`’SPECIMEN (cid:9)
`
`X1.3.3 With the toggle switch in the OFF position, adjust
`the microamrneter to the mechanical zero with the zero adjust
`screw on the front of the microammeter.
`X13.4 With the toggle switch in either the RUN or OFF
`position, the rnicroammcter should read zero. Adjust the
`penetration beam counterweight until a zero reading occurs.
`X1.3.5 Place the toggle switch in the RUN position and
`press the momentary contact switch until the cycle starts.
`When the blade reaches the level of the specimen platform,
`change the toggle switch to the TEST position. The microam-
`meter should continue to read zero. This was a test to determine
`if the microammeter reads zero when the blade first touches the
`specimen.
`XI .3.6 Reach through the side door on the right side of the
`
`case and hang the range calibrating weight in the groove of the
`counterweight stand.
`X1.3.7 With the weight in this position, the reading on the
`microammeter should be 40.0 g. If it is not correct, reach
`through the side door and adjust the range rheostat until the
`reading is 40.0 g.
`
`NOTE X1.1(cid:151) Handle-O-Meter of a capacity double that of No. 211-2 is
`available. This is accomplished by furnishing an electric load cell of
`higher capacity, a switch shunted across a meter for a 2 to I ratio of
`indications. There is a HI-LO switch on the face of the instrument and
`with the LO position the instrument operates the same as No. 211-2 and
`readings are taken directly from the meter; however, with the switch in the
`HI position, all meter readings must be doubled to reflect the shunting
`offset and therefore can be used for stiffer or higher capacity tests.
`
`Xl.3.8 Remove the calibrating weight and check that the
`zero reading is indicated on the microammeter. If the reading
`has shifted, repeat X1.3.5 to and including X1.3.7.
`Xl.3.9 Place the 40-g augmenting weight on the blade,
`while the 40.0-g range calibrating weight is hung on the
`counterweight stud. The microammeter reading should now be
`zero, since each weight is counteracting the other.
`Xl.3.10 Remove both weights before going on to the
`operating procedures.
`Xl.3.11 The actual weight of the calibration weight and the
`augmenting weight must be checked once a month and weigh
`40 – 0.1 g.
`
`The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection
`with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such
`patent rights, and the risk of infringement of such rights, are entirely their own responsibility.
`
`This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
`if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
`and should be addressed to ASTM Headquarters Your comments will receive careful consideration at a meeting of the responsible
`technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your
`views known to the ASTM Committee on Standards, at the address shown be/ow.
`
`This standard is copyrighted byASTM, 100 Barr Harbor Drive, P0 Box C 700, West Conshohocken, PA 19428-2959, United States.
`Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above address or at
`610-832-9585 (phone), 610-832-9555 (fax), or seiviceastin.org (e-mail); or through the ASTM website (www.astm.org).
`
`Copyright by ASTM Int’l (all rights reserved); Mon Jun tO 08:55:49 EDT 2043
`Downloaded/printed by
`Susan Hanrahan (Vorys, Sater. Seymour and Pease LLP) pursuant to License Agreement. No further reproductions authorized.
`
`PAGE 000004
`
`

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket