throbber

`
`UNITED STATES PATENT AND TRADEMARK OFFICE
`__________________
`
`BEFORE THE PATENT TRIAL AND APPEAL BOARD
`__________________
`
`NUVASIVE, INC.
`Petitioner
`
`v.
`
`WARSAW ORTHOPEDIC, INC.
`Patent Owner
`
`
`Patent Number: 8,444,696
`Issue Date: May 21, 2013
`
`
`Case No. IPR2013-00395
`
`______________________________________________________
`
`
`SECOND DECLARATION OF DR. JOHN W. BRANTIGAN, M.D.
`
`Mail Stop “PATENT BOARD”
`Patent Trial and Appeal Board
`U.S. Patent and Trademark Office
`P.O. Box 1450
`Alexandria, VA 22313-145
`
`
`
`
`
`
`
`
`
`Page 1 of 34
`
`NUVASIVE 1017
`NuVasive v. Medtronic
`IPR2013-00395
`
`

`

`
`
`I, Dr. John W. Brantigan, M.D., of Shaw Island, WA, declare that:
`
`INTRODUCTION
`1.
`On June 24, 2013, I provided an initial Declaration in each of the IPR2013-
`
`00395 and IPR2013-00396 proceedings. See Ex. 1001 for IPR2013-00395 and Ex. 1101
`
`for IPR2013-00396. I reaffirm the opinions stated in that Declaration. I provide this Second
`
`Declaration in response to statements made in the April 11, 2014 Patent Owner’s Response
`
`and the accompanying April 10, 2014 Declaration of Dr. Charles L. Branch, Jr. (“Branch
`
`Declaration”) submitted in this proceeding. In this Second Declaration, I will respond to
`
`some of the technical errors and mischaracterizations of my prior testimony that are
`
`contained in the Patent Owner Response and supporting Branch Declaration.
`
`2.
`
`I have reviewed the '696 patent (Ex. 1002) and its prosecution history (Ex.
`
`1003) from the view point of an experienced spine surgeon. Additionally, I have reviewed
`
`the following patents and publications: (1) PCT Publication WO93/01771 to Senter et al.
`
`(“Senter,” Ex. 1007); (2) PCT Publication WO 89/09035 to Brantigan (“Brantigan '035,” Ex.
`
`1005); (3) U.S. Patent No. 5,192,327 to Brantigan (“Brantigan '327,” Ex.1006); (4) PCT
`
`Publication WO90/00037 to Michelson (“Michelson '037,” Ex. 1008); (5) U.S. Pat. No.
`
`5,443,514 to Steffee (“Steffee,” Ex. 1108); and (6) U.S. Patent No. 5,645,596 to Kim et al.
`
`(“Kim,” Ex. 1110). I also have reviewed the Corrected Petition for Inter Partes Review in
`
`each of the IPR2013-00395 and IPR2013-00396 proceedings, and my initial Declarations
`
`signed on June 24, 2013 (Ex. 1001 for IPR2013-00395 and Ex. 1101 for IPR2013-00396). I
`
`Page 2 of 34
`
`

`

`
`
`also have reviewed the Patent Owner’s Response in each of the IPR2013-00395 and
`
`IPR2013-00396 proceedings submitted on April 11, 2014, the accompanying Branch
`
`Declaration (Ex. 2005), the other accompanying exhibits (including Exhibits 2003 (the
`
`Brantigan ‘757 patent), 2004 (the Brantigan ‘772 patent), and 2009 (transcript of my April 7,
`
`2014 Deposition in this proceeding), the transcript of the April 23, 2014 Deposition of Dr.
`
`Branch (“Branch Depo.”) in this proceeding, and the December 20, 2013 Board Decision in
`
`each of the IPR2013-00395 and IPR2013-00396 proceedings.
`
`
`
`SENTER IN VIEW OF BRANTIGAN ‘035
`3.
`On pages 26-29 of the Patent Owner’s Response (IPR2013-00395), Warsaw
`
`argues that the implant of Senter does not provide the claimed “upper and lower bearing
`
`surfaces . . . .being convex.” See also Ex. 2005 at ¶ 35. This argument, however, appears
`
`to be based upon a narrower interpretation of claims 1 and 4 than my earlier analysis (first
`
`Brantigan Declaration at ¶ 29), and incorporates requirements into the claims that are not
`
`present in the claim language. I understand that the December 20, 2013 Board Decision
`
`stated that “the claim language does not require that the convexity be along the entire
`
`length of the implant.” See Board Decision, p. 9. Based on my reading of the claims, I
`
`agree with the Board’s interpretation that one of skill in the art in June of 1995 would not
`
`have interpreted the claims to require that the convexity extend along the entire length of
`
`the implant. As such, the plain meaning of the claim language (“upper and lower bearing
`
`Page 3 of 34
`
`

`

`
`
`surfaces . . . .being convex”) does not exclude the convexity shown in Senter. Additionally,
`
`a person of ordinary skill in the art in June 1995 would not have recognized the claimed
`
`convexity of the upper and lower bearing surfaces as being restricted to a specific degree of
`
`convexity (“convex curvatures conforming to the anatomic endplates”). See Patent Owner’s
`
`Response (IPR2013-00395) at pp. 16-17, 29; Ex. 2005, ¶ 35. If the claimed convexity of the
`
`upper and lower bearing surfaces recited in claims 1 and 4 was restricted to only “convex
`
`curvatures conforming to the anatomic endplates,” the requisite “curvature” of the bearing
`
`surfaces would be patient-dependent, unknown, and undefined because the curvature of
`
`anatomic endplates varies from patient to patient. Ex. 2005, ¶ 60 (describing the variations
`
`in endplate curvature); Branch Depo. at 89:10-22.
`
`
`MODIFICATION OF SENTER TO INCLUDE RATCHETINGS AS SUGGESTED BY
`BRANTIGAN ‘035
`4.
`On pages 33 to 34 of the Patent Owner’s Response (IPR2013-00395),
`
`Warsaw implies that modification of the implant of Senter to include ratchetings, such as the
`
`directional teeth 122 shown in Brantigan ‘035, on the convex upper and lower bearing
`
`surfaces of Senter would cause forward movement of the implant once the implant is put in
`
`place between two vertebrae. Additionally, in his deposition, Dr. Branch stated that
`
`modifying Senter to include the ratchetings of Brantigan ‘035 would cause the implant to be
`
`“propel[led]” toward the spinal cord once the implant is in place. See Branch Depo., 70:3-
`
`19; see also Ex. 2005, ¶¶ 87, 91. A person of ordinary skill in the art would have
`
`Page 4 of 34
`
`

`

`
`
`recognized that this contention is not consistent with the clinical realities of inserting a spinal
`
`implant before June 1995 (and even today).
`
`5.
`
`Contrary to pages 33 to 34 of the Patent Owner’s Response (IPR2013-00395)
`
`and Dr. Branch’s testimony (Deposition p. 70:3-19 and Branch Declaration at ¶¶ 87, 91, and
`
`others), such “ratchetings” as they are called in the Michelson ‘696 patent (which are more
`
`commonly called “teeth” or directional teeth), oriented with a primary purpose of resisting
`
`implant back-out as taught by the Brantigan ‘035 patent and other references, would not
`
`“propel” an implant forward. Instead, such directional “ratchetings” or teeth resist both
`
`implant back-out and forward movement of the implant, and a person of ordinary skill in the
`
`art before June 7, 1995 would have recognized this plain fact. Of course, with directional
`
`ratchetings or teeth, the resistance to forward movement of the implant can be less than the
`
`resistance to implant back-out, but such ratchetings or teeth nevertheless resist forward
`
`movement also. The resistance to movement in the forward direction is simply a matter of
`
`physics and anatomy. The angling of the forward-facing ratchet face, in combination with
`
`the presence of the ridge of the ratcheting which digs into the adjacent vertebrae end faces
`
`when the adjacent vertebrae are forced against the bearing surfaces of the implant, causes
`
`that resistance to the forward movement of the implant. See, e.g., Ex. 1005, 19:29-31
`
`(teaching that such ratchetings “provid[e] a roughened surface, biting into and gripping” the
`
`vertebral bone adjacent to the implant). This anchoring effect would be especially true for
`
`anteriorly inserted implants such as Senter, in which it was common practice to distract the
`
`Page 5 of 34
`
`

`

`
`
`vertebrae apart during insertion of the implant and then remove the distraction after the
`
`implant is finally seated to thereby cause the vertebrae to compress against the bearing
`
`surface of the implant and its teeth. See Ex. 2009 85:12 to 87:12 (stating that, when using
`
`anteriorly inserted interbody implants, “you generally distract the vertebral bodies apart and
`
`the endplates apart”); Ex. 1007, p. 16:20-22 (“The spine 20 is typically distended slightly
`
`during this final stage of insert . . . .”). Even in applications where forward resistance to
`
`movement of the implant may not be needed (for example, if the disc annulus at the far side
`
`of the disc space is still intact, thereby blocking forward movement of the implant out of the
`
`disc space), the common design option of ratchetings or directional teeth provides
`
`resistance to movement in the rearward (retraction) direction, in the forward (insertion)
`
`direction, and even in a side-to-side direction (after the teeth “bite” or otherwise anchor into
`
`the vertebral bone).
`
`6.
`
`Again responsive to pages 33 to 34 of the Patent Owner’s Response
`
`(IPR2013-00395) and Dr. Branch’s testimony (Deposition p. 70:3-19 and Branch
`
`Declaration at ¶¶ 87, 91, and others), in the process of advancing an implant (such as
`
`Senter’s implant) into the space between two adjacent vertebrae, it is not “ratchetings” or
`
`teeth on the implant that would somehow “propel” an implant forward into position. In other
`
`words, there is no “ratcheting” movement forward of an implant by the implant’s
`
`“ratchetings” in the typical sense of the word “ratcheting” (e.g., ratcheting teeth that interact
`
`with a pawl or other indexing structure). Rather, the implant would be advanced or pushed
`
`Page 6 of 34
`
`

`

`
`
`into position between the adjacent vertebrae by the surgeon using an inserter tool (applying
`
`an insertion force against the trailing end of the implant), as was conventional before June
`
`7, 1995 and as plainly described in prior art references cited above. See also My First
`
`Declaration in IPR2013-00395, Ex. 1001, ¶¶ 31, 38, and 43. In that sense, the label that the
`
`Michelson ‘696 patent gives to the feature of anti-migration implant “teeth”—in particular, the
`
`label of “ratchetings”—is a misnomer. The teeth or “ratchetings” on the implant bearing
`
`surfaces do not ratchet or propel movement in any direction in the typical sense of the word
`
`“ratcheting,” and a person of ordinary skill in the art before June 7, 1995 would have
`
`recognized this fact. That typical use of the term “ratcheting” is not what is at play with the
`
`so-called “ratchetings” or directional teeth on the implant having teeth shown in the ‘696
`
`patent (e.g., Figure 9), or any of the prior art implants of record that have “ratchetings” or
`
`teeth. See Ex. 1005 at FIGS. 18-19; 19:25 to 20:3; 20:30-33; and 21:1-5; Ex. 1108 at FIG.
`
`3; 2:54-59.
`
`7.
`
`A person of ordinary skill in the art before June 7, 1995 would also have
`
`understood that, during the procedure when an implant is implanted and put into place
`
`between two adjacent vertebrae, the patient is not in a standing or seated position, but
`
`rather is in a lying position, for example in a supine position (as with Senter’s anteriorly
`
`inserted implant), a prone position, or lateral decubitus (on one side) position. In other
`
`words, the full weight of the patient’s body is not causing the two adjacent vertebrae to
`
`compress the disc space where the implant is to be positioned. Also, in many cases
`
`Page 7 of 34
`
`

`

`
`
`(including Senter and other anteriorly inserted implants), the two adjacent vertebrae are
`
`distracted apart before and during the time that the implant is inserted. See My Prior
`
`Deposition Testimony, Ex. 2009, 85:12 to 87:12; Ex. 1007, p. 16:20-22. As such, contrary
`
`to pages 33 to 34 of the Patent Owner’s Response (IPR2013-00395) and Dr. Branch’s
`
`testimony (Deposition p. 70:3-19 and Branch Declaration at ¶¶ 87, 91, and others), a person
`
`of ordinary skill in the art before June 7, 1995 would have understood that the resistance to
`
`both forward and rearward movement is less during the time of initial implant insertion than
`
`it is later when distraction is removed (allowing the directional teeth or ratchetings of the
`
`implant “bite” into the adjacent vertebrae) and when the full force of the patient’s weight
`
`causes the two adjacent vertebrae to even further bear upon the implant. Ex. 1005, 19:29-
`
`31.
`
`8.
`
`In the case of an implant with convex bearing surfaces, where Senter’s
`
`convex bearing surface is secured within a concave groove formed in the vertebrae (Senter
`
`at p. 17:8-9), the convexity of the bearing surface also serves to resist forward migration of
`
`the implant. Therefore, contrary to pages 33 to 34 of the Patent Owner’s Response
`
`(IPR2013-00395) and Dr. Branch’s testimony (Deposition p. 70:3-19 and Branch
`
`Declaration at ¶¶ 87, 91, and others), putting “ratchetings” or other teeth on a surface of an
`
`implant convex bearing surface was considered by persons of ordinary skill in the art before
`
`June 7, 1995 to be a “belt-and-suspenders” approach to making sure the implant does not
`
`migrate from the position where it is initially implanted in the disc space. See Ex. 1007, p.
`
`Page 8 of 34
`
`

`

`
`
`17:5-13; Ex. 1009, FIG. 7, 5:64-65 and 7:24-28. Thus, a person of ordinary skill in the art
`
`before June 7, 1995 would have recognized that applying the traditional implant teeth
`
`shown in the Brantigan ‘035 patent (which again, are called “ratchetings” in the ‘696 patent)
`
`to the top and bottom convex bearing surfaces of the Senter implant (refer to ¶¶37-38 of the
`
`June 24, 2013 Brantigan Declaration) would also serve as a “belt-and-suspenders”
`
`approach to making sure the implant does not migrate (forward, backward, or side-to-side)
`
`from the position where it is implanted.
`
`9.
`
`In summary, and responsive to pages 33 to 34 of the Patent Owner’s
`
`Response (IPR2013-00395) and Dr. Branch’s testimony (Deposition p. 70:3-19 and Branch
`
`Declaration at ¶¶ 87, 91, and others), the convex bearing surfaces of Senter’s implant
`
`seated between the adjacent vertebrae would continue to assist in the prevention of forward
`
`migration of the implant toward the spinal cord once in place, even if modified to include the
`
`traditional feature of ratchetings or directional teeth that was known in field of implant design
`
`before 1995. In addition, a person of ordinary skill in the art before June 7, 1995 would
`
`have recognized that those ratchetings or directional teeth would, once the implant is in the
`
`surgeon’s selected position, “bite into” the concave surfaces of the adjacent vertebrae so as
`
`to prevent movement of the implant in both the forward direction and the reverse (or back-
`
`out) direction. See, e.g., Ex. 1005, 19:29-31. Additionally, even if any slight movement of
`
`the implant were to occur (in an unlikely scenario, for example, before fusion between the
`
`two adjacent vertebrae has fully formed), such movement would be minor for the reasons
`
`Page 9 of 34
`
`

`

`
`
`explained above (e.g., probably less than half the length of one ratcheting) and the angled
`
`configuration of the ratchetings of Brantigan ‘035 would bias the implant to move back into
`
`the original placed position such that the ratchetings would rest within the original “bite”
`
`grooves in the vertebrae created by the teeth/ratchetings after the initial placement of the
`
`implant. This coupling of the ratchetings with the original “bite” grooves in the adjacent
`
`vertebrae created by the ratchetings would be an additional means of accomplishing
`
`Senter’s stated goal of “lock[ing] the implant 50 or 74 into place [to] prevent dislocation of
`
`the implant.” Ex. 1007 17:6-7. One of skill in the art before June 1995 would have
`
`considered convex bearing surfaces and anti-migration directional teeth (called “ratchetings”
`
`in the ‘696 patent) as two traditional tools in an implant design toolbox before June 1995,
`
`and would not have considered these different implant migration tools inconsistent with one
`
`another such that they could not be used in the same implant.
`
`
`
`MODIFICATION OF SENTER TO INCLUDE A BONE FUSION OPENING AS
`SUGGESTED BY BRANTIGAN ‘035
`10. On pages 44 to 45 of the Patent Owner’s Response (IPR2013-00395),
`
`Warsaw contends that modifying Senter’s implant with a convex bearing surface to include
`
`one or more large openings for receiving fusion promoting material (a known prior art
`
`feature of Brantigan ‘035 and others) would cause bone filling inserted into the opening to
`
`“necessarily extend above and below the surfaces of the anterior platforms 56, 56a and the
`
`posterior ledges 60, 60a” and would therefore “not be protected by the leading end of the
`
`Page 10 of 34
`
`

`

`
`
`modified spinal disk implant 50 of Senter during insertion and would be subject to
`
`dislodgement via contact with the superior and inferior vertebrae.” Alternatively, Warsaw
`
`asserts that if “bone did not entirely fill the void created by the suggested opening, such
`
`bone could be protected by the leading end of the modified spinal disk implant 50 of Senter
`
`during insertion, but would be separated from the superior and inferior vertebrae 22a and
`
`22b by significant gaps” thereby preventing successful fusion. Patent Owner’s Response
`
`(IPR2013-00395) at pp. 44-45; see also Branch Decl. at ¶¶ 98-99. Based upon my
`
`knowledge and experience in this field, I believe that a person of ordinary skill in the art
`
`before June 1995 would have recognized that this alleged concern would not have been
`
`problematic for an ordinary surgeon in 1995, and is not a problem now.
`
`11.
`
`The dislodgement problem theorized by Warsaw’s Response and Dr. Branch
`
`ignores the traditional clinical practice for anteriorly inserted implants (including Senter’s
`
`implant) in which the two adjacent vertebrae are distracted apart before and during the time
`
`that the implant is inserted. See Ex. 2009 85:12 to 87:12; Ex. 1007, p. 16:20-22. Therefore,
`
`if dislodgement of some of the fusion promoting substance from a top part of the fusion
`
`opening might pose a legitimate risk, then an ordinary surgeon before June 1995 would
`
`have readily known how to address that concern by distracting the two adjacent vertebrae
`
`from one another so as to avoid such dislodgement. One of skill in the art would also have
`
`known before June 1995, as is known today, that spinal fusion implants do not need to be
`
`loaded to the very absolute top of the fusion aperture, in order to facilitate bone growth
`
`Page 11 of 34
`
`

`

`
`
`through the implant from one adjacent vertebra to the other, especially due to the bleeding
`
`that would occur to fill minor gaps on both the upper and lower surfaces of Senter’s implant
`
`during and after it is implanted. In addition, one of skill in the art would have known before
`
`1995, that if the fusion promoting substance included cancellous bone, such cancellous
`
`bone had a tendency to expand from its tightly packed state after the implant was placed
`
`between the vertebrae, thereby also filling minor gaps that might exist between the implant
`
`and the vertebral body. One of skill in the art would also have known before 1995, as would
`
`also be known today, that some minor dislodgement of fusion promoting material may be
`
`tolerated without compromising the ability of the fusion implant to carry out the fusion
`
`process.
`
`12.
`
`Additionally, if the alleged dislodgement problem (conventional large fusion
`
`“openings” through an implant with convex bearing surfaces) was in fact a legitimate
`
`problem for ordinary surgeons before June 7, 1995, then that is a problem for which the
`
`‘696 patent provides no solution. For example, although the ‘696 patent suggests an
`
`“opening” to an internal chamber of the convex implant embodiment 300 (FIGS. 13-17 and
`
`col. 9:48-51), it provides no guidance regarding the design of such “openings” that would be
`
`appropriate in an implant having convex bearing surfaces. In particular, the ‘696 patent
`
`discloses no protective structure on the leading end of an implant with convex bearing
`
`surfaces or other protective mechanism so as to prevent dislodgement of fusion promoting
`
`material from an opening in a convex surface during the implantation process. Further, the
`
`Page 12 of 34
`
`

`

`
`
`‘696 patent does not even describe where or how one or more “openings” would be
`
`included in the implant 300 with convex bearing surfaces. For example, the ‘696 patent
`
`does not disclose whether and how the then-conventional large openings like those shown
`
`in implant 400 of the ‘696 patent (see Figures 18-23) would be incorporated into an implant
`
`with convex bearing surfaces such as implant 300 of the ‘696 patent. Therefore, unless
`
`Warsaw and Dr. Branch are contending that the ‘696 patent fails to enable ordinary
`
`surgeons before June 1995 to overcome the alleged dislodgement problem, the lack of any
`
`solution in the ‘696 patent about how to avoid dislodgement of materials from the “opening”
`
`extending through the convex implant 300 supports the fact that avoiding this supposed
`
`dislodgement problem was well within the level of ordinary skill in the art in 1995.
`
`13.
`
` To further illustrate that the dislodgement problem theorized by Warsaw and
`
`Dr. Branch (Patent Owner’s Response (IPR2013-00395) at pp. 44-45; Branch Decl. at ¶¶
`
`98-99) is, in fact, not a problem, the Medtronic Clydesdale spinal implant has one very large
`
`opening for receiving fusion-promoting material, and that one large hole extends from a top
`
`convex bearing surface of the implant to a bottom convex bearing surface of the implant:
`
`Page 13 of 34
`
`

`

`
`
`
`See Ex. 2010, Appendix A. As shown in the diagram of the Medtronic Clydesdale implant
`
`shown directly above, the leading edge of the one large hole in the Clydesdale implant is
`
`not protected by any structure on the leading edge of the implant. As such, if dislodgment
`
`of bone fusion promoting substance from a large fusion hole during the insertion of a convex
`
`implant was an actual problem, that same problem would also be present with the
`
`Clydesdale implant design. I am not aware of Medtronic ever notifying the surgeon
`
`community that dislodgement of the bone growth promoting substance is a problem with its
`
`Clydesdale implant, and I think it unlikely that this is a problem.
`
`14.
`
`Therefore, the alleged dislodgment problem with modifying the spinal implant
`
`of Senter to include one or more large fusion openings as taught by the Brantigan ‘035
`
`patent would not have been a problem for practicing surgeons before June 1995 (especially
`
`when the disc space is distracted during implant insertion as taught by Senter), and would
`
`not have been a reason why an ordinary implant designer before June 1995 would have
`
`Page 14 of 34
`
`

`

`
`
`avoided adding a traditional bone fusion aperture to the Senter implant, especially given the
`
`bone fusion benefits described in Brantigan ‘035 and otherwise known by ordinary surgeons
`
`before June 1995.
`
`
`MODIFICATION OF MICHELSON ‘037 TO INCLUDE CONVEX BEARING SURFACES AS
`SUGGESTED BY WAGNER
`15. On pages 47-52 of the Patent Owner’s Response (IPR2013-00395), Warsaw
`
`argues that the image of the implant 10 in FIGS. 1a-c of Michelson ‘037 “implies its use for
`
`treatment” for a disc space that “lost its bi-convex appearance.” See also Ex. 2005, ¶¶ 60,
`
`104. In other words, Warsaw argues that Michelson ‘037’s implant is limited for use only in
`
`circumstances in which vertebral endplates are essentially flat or have otherwise lost their
`
`bi-concave appearance. Id. Based upon my knowledge and experience in this field, I
`
`believe that a person of ordinary skill in the art before June 1995 would have recognized
`
`that Michelson ‘037’s teaching is not so limited and that Michelson ‘037 expressly teaches
`
`the opposite. Namely, Michelson ‘037 teaches that the disclosed implant device “will fit any
`
`patient, anywhere throughout the spine, in any vertebral disc space, and without alteration
`
`of that interspace regardless of its natural size or shape.” Ex. 1008, p. 5:18-22. All of
`
`Warsaw’s arguments that limit the use of Michelson ‘037’s implant (including the
`
`contentions that it then would not be modified to provide convex bearing surfaces as
`
`suggested in Wagner) are inconsistent with the express teaching found in the Michelson
`
`Page 15 of 34
`
`

`

`
`
`‘037 patent and what a person of ordinary skill in the art before June 1995 would have
`
`understood from such a teaching.
`
`
`
`MODIFICATION OF MICHELSON ‘037 TO INCLUDE RATCHETINGS AS SUGGESTED
`BY BRANTIGAN ‘035
`16. On pages 52-55 of the Patent Owner’s Response (IPR2013-00395), Warsaw
`
`argues that, if the implant of Michelson ‘037 was modified to include convex bearing
`
`surfaces as suggested by Wagner, then the resulting combination must necessarily be
`
`limited to also including Wagner’s “pyramids” rather than any other conventional type of
`
`engagement features. See also Ex. 2005, ¶¶ 105-108. Based upon my knowledge and
`
`experience in this field, I believe that a person of ordinary skill in the art before June 1995
`
`would have recognized that Wagner’s teaching for engagement features is not so limited
`
`and that the available design options for conventional engagement features included
`
`directional teeth (as suggested, for example, by Brantigan ‘035). In particular, Wagner first
`
`suggests that the engagement features can be any “three-dimensional features” extending
`
`above the bearing faces, and then discloses that the depicted “embodiment” of such three-
`
`dimensional features includes the “pyramids.” Ex. 1009, 5:64-6:1. A person of ordinary skill
`
`in the art before 1995 would not have concluded that “pyramids” are the only option
`
`available (or even the best available option) for implementing three-dimensional
`
`engagement features on convex bearing surfaces. Furthermore, as described in my June
`
`24, 2013 Declaration (¶¶47-48), a person of ordinary skill in the art before 1995 would have
`
`Page 16 of 34
`
`

`

`
`
`recognized that the directional teeth or ratchetings suggested in Brantigan ‘035 were
`
`another traditionally available option for such “three-dimensional features” on the implant’s
`
`bearing surfaces and that such directional teeth or ratchetings can also provide an
`
`anchoring effect in multiple directions (forward, backward, and side-to-side) after “biting into”
`
`the vertebrae and provide increased resistance to the implant backing out. Ex. 1005, 19:25
`
`to 20:3, 21:1-5; FIGS. 18-19. The ordinary implant designer before June 1995 was more
`
`than capable of applying Wagner’s suggestion for convex bearing surfaces while also
`
`considering and implementing Brantigan’s suggestion for ratchetings on the bearing
`
`surfaces.
`
`17. On pages 54 to 55 of the Patent Owner’s Response (IPR2013-00395),
`
`Warsaw implies that implementing traditional ratchetings, such as the directional teeth 122
`
`shown in Brantigan ‘035, on the convex upper and lower bearing surfaces of the modified
`
`Michelson ‘037 implant would cause the implant to be “susceptible to forward and side to
`
`side movement.” Additionally, in his Declaration, Dr. Branch stated that implementing the
`
`traditional ratchetings of Brantigan ‘035 on the modified Michelson ‘037 implant would cause
`
`the implant to be “susceptible to forward and side to side movement.” See Ex. 2005,
`
`¶¶108-109. Again, for the same reasons described above in ¶¶ 6-11, this contention is not
`
`true and is not consistent with the clinical realities of inserting a spinal implant before June
`
`1995 (and even today). Additionally, although it is possible that such directional teeth or
`
`ratchetings can afford forward or side-to-side movement during the insertion process (when
`
`Page 17 of 34
`
`

`

`
`
`the implant is attached to the inserter tool and moved under manual force from the
`
`surgeon), it was known that these structures would provide an anchoring effect in multiple
`
`directions (forward, backward, and side-to-side) after reaching the final position and “biting”
`
`into the vertebral bone. See Ex. 1005, 19:25 to 20:3, 21:1-5; FIGS. 18-19.
`
`18. Contrary to pages 54 to 55 of the Patent Owner’s Response (IPR2013-00395)
`
`and Dr. Branch’s testimony (Deposition p. 70:3-19 and Branch Declaration at ¶¶ 108, 109,
`
`and others), such directional “ratchetings” as they are called in the Michelson ‘696 patent
`
`(which are more commonly called “teeth”), do not “propel” an implant forward. Instead, such
`
`directional “ratchetings” or teeth resist both implant back-out and forward movement of the
`
`implant, and a person of ordinary skill in the art before June 7, 1995 would have recognized
`
`this plain fact. See, supra, ¶¶ 7-11; see also Ex. 1005, 19:29-31 (teaching that such
`
`ratchetings “provid[e] a roughened surface, biting into and gripping” the vertebrae adjacent
`
`to the implant). This anchoring effect (resisting both forward, rearward migration, and side-
`
`to-side migration) would be especially true for anteriorly inserted implants such as
`
`Michelson ‘037, in which it was common practice to slightly distract the vertebrae apart
`
`during insertion of the implant and then remove the distraction after the implant is finally
`
`seated to thereby cause the vertebrae to compress against the bearing surface of the
`
`implant and any of its teeth. See Ex. 2009 85:12 to 87:12 (stating that, when using
`
`anteriorly inserted interbody implants, “you generally distract the vertebral bodies apart and
`
`the endplates apart”); Ex. 1008, p.9:8-10 (“mild distraction”).
`
`Page 18 of 34
`
`

`

`
`
`19.
`
`Also as previously described (refer above to ¶ 9), in the case of an implant
`
`with convex bearing surfaces (such as Michelson ‘037’s modified implant as described in ¶¶
`
`45-48 of the June 24, 2013 Brantigan Declaration), where the convex bearing surface
`
`conforms with a concave surface in an adjacent vertebra end face, the convexity of the
`
`bearing surface also serves to resist forward migration of the implant. Therefore, contrary to
`
`pages 54 to 55 of the Patent Owner’s Response (IPR2013-00395) and Dr. Branch’s
`
`testimony (Deposition p. 70:3-19 and Branch Declaration at ¶¶ 108, 109, and others),
`
`putting “ratchetings” or other teeth on a surface of an implant convex bearing surface was
`
`considered by persons of ordinary skill in the art before June 7, 1995 to be a “belt-and-
`
`suspenders” approach to making sure the implant does not migrate from the position where
`
`it is initially implanted in the disc space. See Ex. 1009, FIG. 7, 5:64-65 and 7:24-28;
`
`discussion herein supra, ¶¶ 8-9. A person of ordinary skill in the art before June 7, 1995
`
`would have recognized that applying the traditional implant teeth shown in the Brantigan
`
`‘035 patent (which again, are called “ratchetings” in the ‘696 patent) to the top and bottom
`
`convex bearing surfaces of the modified Michelson ‘037 implant (refer to ¶¶ 47-48 of my first
`
`Declaration) would also serve as a “belt-and-suspenders” approach to making sure the
`
`implant does not migrate from the position where it is implanted.
`
`
`
`MODIFICATION OF STEFFEE TO INCLUDE OPENINGS AS SUGGESTED BY
`MICHELSON ‘037
`
`Page 19 of 34
`
`

`

`
`
`20. On pages 34 to 38 of the Patent Owner’s Response (IPR2013-00396),
`
`Warsaw asserts that one of ordinary skill would not be motivated to modify the posterior
`
`lumbar interbody fusion (PLIF) insert-and-rotate implant of Steffee to include the fusion-
`
`promoting openings of Michelson ‘037 because Steffee teaches fusion between the vertebra
`
`accomplished by bone growth between two PLIF implants (allegedly not through the
`
`implants) and based on the false assertion that Steffee is not a spinal fusion implant. See
`
`also Ex. 2005, ¶¶ 77-78. Based upon my knowledge and experience in this field, I believe
`
`that a person of ordinary skill in the art before June 1995 would have recognized that
`
`Steffee’s implant is a type of spinal fusion implant in which bone growth occurs through the
`
`implant structure, and furthermore is generally analogous to other interbody fusion implants
`
`such as Michelson ‘037. For example, Steffee explicitly teaches “using the spinal implant to
`
`fuse together adjacent vertebrae of a spinal column.” Ex. 1108 1:7-8) (emphasis added).
`
`Additionally, Steffee expressly teaches that the implant can be equipped with openings
`
`through the implant “to provide for blood flow and bone growth from one side of the implant
`
`10 to the other side of the implant.” Ex. 1108, 2:64-67. Contrary to Warsaw’s and Dr.
`
`Branch’s assertion, it is not the direction of holes through an implant that make it a “fusion
`
`implant,” but instead a person of ordinary skill in the art before June 1995 would have
`
`recognized, as an ordinary surgeon would recognize today, that the Steffee implant and the
`
`Michelson ‘037 implant were both implants used in the interbody space for purposes of
`
`achieving spinal fusion between adja

This document is available on Docket Alarm but you must sign up to view it.


Or .

Accessing this document will incur an additional charge of $.

After purchase, you can access this document again without charge.

Accept $ Charge
throbber

Still Working On It

This document is taking longer than usual to download. This can happen if we need to contact the court directly to obtain the document and their servers are running slowly.

Give it another minute or two to complete, and then try the refresh button.

throbber

A few More Minutes ... Still Working

It can take up to 5 minutes for us to download a document if the court servers are running slowly.

Thank you for your continued patience.

This document could not be displayed.

We could not find this document within its docket. Please go back to the docket page and check the link. If that does not work, go back to the docket and refresh it to pull the newest information.

Your account does not support viewing this document.

You need a Paid Account to view this document. Click here to change your account type.

Your account does not support viewing this document.

Set your membership status to view this document.

With a Docket Alarm membership, you'll get a whole lot more, including:

  • Up-to-date information for this case.
  • Email alerts whenever there is an update.
  • Full text search for other cases.
  • Get email alerts whenever a new case matches your search.

Become a Member

One Moment Please

The filing “” is large (MB) and is being downloaded.

Please refresh this page in a few minutes to see if the filing has been downloaded. The filing will also be emailed to you when the download completes.

Your document is on its way!

If you do not receive the document in five minutes, contact support at support@docketalarm.com.

Sealed Document

We are unable to display this document, it may be under a court ordered seal.

If you have proper credentials to access the file, you may proceed directly to the court's system using your government issued username and password.


Access Government Site

We are redirecting you
to a mobile optimized page.





Document Unreadable or Corrupt

Refresh this Document
Go to the Docket

We are unable to display this document.

Refresh this Document
Go to the Docket